The following symbols are used in this manual to prevent accidents which may occur as a result of incorrect use of
the instrument.
Denotes an instr uction regarding a safety warning or note.
Read the instruction carefully to ensure safe and correct use.
Denotes an instr uction regarding the risk of electric shock.
Read the instruction carefully to ensure safe and correct use.
Denotes an instr uction regarding the risk of re.
Read the instruction carefully to ensure safe and correct use.
Denotes a prohibited operation.
This operation must never be performed.
Denotes an instr uction.
This instruction must be strictly adhered to.
Denotes a prohibited operation.
Never disassemble the instrument.
Denotes an instr uction.
Be sure to disconnect the AC adapter from the AC outlet.
This symbol indicates A.C.
This symbol indicates D.C.
This symbol indicates class II protection against electric shock.
Trademarks
• Windows
Notes on this Manual
• Copying or reproduction of all or part of the contents of this manual without the permission of KONICA
MINOLTA is strictly prohibited.
• The contents of this manual are subject to change without prior notice.
• Every effort has been made in the preparation of this manual to ensure the accuracy of its contents.
• However, should you have any questions or nd any errors, please contact your retailer or a KONICA MINOLTA-
authorized service facility.
• KONICA MINOLTA will not accept any responsibility for consequences arising from the use of the instrument.
®
is a registered trademark of Microsoft Corporation in the United States and other countries.
Page 3
Safety Precautions
To ensure correct use of this instrument, read the following points carefully and adhere to them.
After you have read this manual, keep it in a safe place where it can be referred to anytime a question arises.
WARNING
Do not use the instrument in places where
ammable or combustible gases (gasoline
etc.) are present.Doing so may cause a re.
Always use the AC adapter supplied as a
standard accessory or the optional AC
adapter, and connect it to an AC outlet of
the rated voltage and frequency. If an AC
adapter other than those specied by
KONICA MINOLTA is used, it may result
in damage to the unit, re or electric
shock.
If the instrument will not be used for a
long time, disconnect the AC adapter plug
from the AC outlet. Accumulated dirt or
water on the prongs of the AC adapter’s
plug may cause a re and should be
removed.
Do not forcibly bend, twist, or pull the AC
adapter power cable. Do not scratch or
alter the power cable or place heavy
objects on it. Doing so may damage the
power cable and cause a re or electric
shock.
Insert the power plug fully and securely.
Incomplete insertion may cause re or
electric shock.
(Failure to adhere to the following points may result in death or serious
injury.)
Do not disassemble or modify the
instrument or the AC adapter. Doing so
may cause a re or electric shock.
Take special care not to allow liquid or
metal objects to enter the instrument and
the AC adapter. Doing so may cause a re
or electric shock. Should liquid or metal
objects enter the instrument, turn the
power OFF immediately, disconnect the
AC adapter plug from the AC outlet, and
contact the nearest KONICA MINOLTA-
authorized service facility.
The instrument should not be operated if it
is damaged or the AC adapter is damaged,
or if smoke or odd smells occur. Doing so
may cause a re. In such situations, turn
the power OFF immediately, disconnect
the AC adapter plug from the AC outlet
and contact the nearest KONICA
MINOLTA-authorized service facility.
Always hold the plug itself when
disconnecting the AC adapter plug from
the AC outlet. Pulling on the power cable
may damage it and cause a re or electric
shock.
Do not insert or disconnect the AC adapter
plug from an AC outlet with wet hands.
Doing so may cause electric shock.
CAUTION
Do not perform measurement with the
specimen measuring port directed towards
your eyes. Doing so may damage your
eyes.
Be careful not to get your hand caught in
the openable section of the instrument.
Doing so may result in injury.
When cleaning, unplug the power plug.
Leaving it plugged in may cause electric
shock.
(Failing to adhere to the following points may result in injury or damage
to the instrument or other property.)
Do not place the instrument on an unstable
or sloping surface.Doing so may result in
its dropping or overturning, causing injury.
Be careful not to drop the instrument when
carrying it.
Make sure that the AC outlet is located
near the instrument and that the AC
adapter plug can be connected to and
disconnected from the AC outlet easily.
E1
Page 4
Introduction
Thank you for purchasing the CM-5.
This is a precise, benchtop type spectrophotometer developed for color and color difference measurement of
reective/transmitted objective colors in various industries.
Packing materials of the product
Be sure to keep all packing materials used for shipping the product (cardboard box, cushioning material,
plastic bags, etc.).
The CM-5 is a precision measuring instrument.When transporting the instrument to a service facility for
maintenance or for other reasons, be sure to use the packing materials to minimize shock or vibration.
If the packing materials are lost or damaged, contact a KONICA MINOLTA-authorized service facility.
Notes on Use
Be sure to use this instrument properly. Use of this instrument in ways other than those specied in this manual
may result in risk of injury, electric shock, instr ument damage, or other problems.
Operating Environment
• Use the CM-5 at an ambient temperature between 13°C and 33°C and relative humidity 80% or less (at 35°C)
with no condensation.
Be sure to use the instrument within this range. Do not use it in areas of rapid temperature changes.
• Do not leave the CM-5 in direct sunlight or near sources of heat, such as stoves etc. The internal temperature
of the instrument may become much higher than the ambient temperature in such cases.
• Do not use the CM-5 in areas where dust, cigarette smoke or chemical gases are present. Doing so may cause
deterioration in performance or a breakdown.
• Do not use the CM-5 near equipment which produces a strong magnetic eld (such as speakers etc.).
• The CM-5 belongs to installation category I products (equipment which is powered by an AC adapter
connected to commercially available power).
• The CM-5 belongs to pollution level 2 products (equipment which may cause temporary electrical hazards due
to contamination or condensation or products which are used in such an environment).
• Do not use the CM-5 at altitudes higher than 2,000 m.
• The CM-5 and the AC adapter supplied as a standard accessor y have been designed exclusively for indoor use.
They should never be used outdoors because rain or other factors may damage the instr ument.
Measurement
• Make sure no dirt or dust get into the specimen measuring port.
• Use a blower to remove dust and dirt from the shutter before using the instrument after a long period of
disuse.
• When using the instrument for long periods of time, the measurement value may change depending on
changes in the environment. Therefore, in order to achieve accurate measurements, we recommend that white
calibration be done regularly using the White Calibration Cap.
White Calibration Plate
There are two types of white calibration plate for use with this instrument: the built-in White Calibration Plate
inside the shutter and an external White Calibration Plate as an optional accessory.
• The calibration data for the White Calibration Plate was measured at 23°C.
To achieve the highest accuracy when measuring absolute values (colorimetric values), calibration and
measurement should be performed at 23°C.
• Do not allow the White Calibration Plate (optional accessory) to get scratched or stained, or get ngerprints on
it.
• When the White Calibration Plate (optional accessory) is not in use, be sure to close the cap so that the White
Calibration Plate is not exposed to ambient light.
Target Mask
There are three types of Target Masks for use with this instrument: the standard ø30 mm mask, and optional ø3
mm and ø8 mm masks.
• The ø30 mm Target Mask must be used in combination with the instrument that bears the same pairing
numbe r.
• Do not touch the Target Mask's inner surface by hand, scratch it or make it dirty.
• When not in use, Target Masks should be stored in the packing used for shipment or in the optional accessory
case so that they will not be exposed to ambient light.
E2
Page 5
Power Source
• Make sure that the power switch is set to OFF (“”) when the CM-5 is not in use.
• This instrument can be powered from the AC adapter (AC-A305).
• Always use the AC adapter supplied as a standard accessory (AC-A305) and connect it to an AC outlet of the
rated voltage and frequency. Use an AC power supply of the rated supply voltage (within ±10%).
• Do not connect the AC adapter to an overloaded electrical circuit. In addition, do not wrap or cover the AC
adapter with cloth or other material while in use. Doing so may cause an electric shock or re.
System
• Do not subject the CM-5 to strong impact or vibration.Doing so may cause deterioration in performance or a
breakdown.
• Since the specimen measuring port and integrating sphere are extremely precise optical components, great
care should be taken to prevent them getting dirty or exposing them to impact.
• The CM-5 may cause interference if used near a television, radio, etc.
• When the instr ument is exposed to strong external static electricity, the LCD may go blank or the
measurement result may not be displayed properly. If the instrument is communicating with an external
device, the communication may be interrupted. In these cases, turn the power OFF and then turn it ON again.
If black smudges appear on the LCD, wait until they disappear naturally.
• When turning the power OFF and then ON again, wait several seconds after tur ning the power OFF.
Backup Battery
• Measured data and various settings are stored in the memory backed up by batteries. The backup batteries are
automatically charged during operation of this instr ument, and can retain the contents of the memory for three
months if they have been fully charged. At the time of purchase, the backup battery may not be fully charged.
To charge the backup battery, set the power switch to ON. The backup battery is fully charged in 72 hours
while the instrument is switched on, and there is no danger of overcharging.
• It is recommended to keep a backup of your important data on another recording medium using the optional
Color Management Software SpectraMagic™ NX (CM-S100w).
Notes
• The backup batter ies’ model number is ML2020 (3 V).
• Do not try to replace the backup batteries by yourself. Contact a KONICA MINOLTA-authorized service
facility.
E3
Page 6
Notes on Storage
• The CM-5 should be stored at temperatures between 0°C and 40°C, and at a relative humidity of 80% or less
(35°C) without condensation.Do not store the instrument in areas subject to high temperatures, high humidity,
sudden changes in temperature, or where freezing or condensation may occur, because these circumstances
may cause a breakdown.It is recommended to store the CM-5 with a drying agent at a temperature around
20°C.
• Do not leave the CM-5 inside a car such as in the cabinet or tr unk. Otherwise, the temperature and/or
humidity may exceed the allowable range for storage during midsummer or midwinter, resulting in a
breakdown.
• Keep the packing materials used for shipment and use them to transport the CM-5. This protects the
instrument from sudden changes in temperature, vibration, and shock.
• Do not store the CM-5 in areas where dust, cigarette smoke or chemical gases are present.Doing so may cause
deterioration in performance or a breakdown.
• Entry of dust into the specimen measuring port will hinder accurate measurement.When the instrument is not
in use, you must close the transmittance specimen chamber and cover the instrument with the supplied Dust
Cover to prevent the entry of dust into the integrating sphere.
• The White Calibration Plate (optional accessory) may become discolored if left exposed to light. Therefore,
make sure to close the cap when it is not in use so that the White Calibration Plate is not exposed to ambient
light during storage.
• The Target Masks may discolor if they are left exposed to light. When they are not in use, keep them in a safe
place to prevent exposure to light and to protect them from scratches and dust.
• Be sure to keep all packing materials (cardboard box, cushioning material, plastic bags, etc.). They can be
used to protect the instrument during transportation to the service facility for maintenance (re-calibration
et c.).
Notes on Cleaning
• If the CM-5 becomes dirty, wipe it with a soft, clean dry cloth.Never use solvents such as thinner or benzene.
• If the White Calibration Plate (optional accessory) becomes dirty, wipe it with a soft, clean dry cloth.If dirt is
difcult to remove, wipe it off with a cloth dampened with commercially-available lens cleaning solution.
Then remove the solution with a cloth dampened with water, and leave the plate to dry.
• If the inner surface of the Target Masks, the inside of the integrating sphere , or the built-in White Calibration
Plate get dirty, contact a KONICA MINOLTA-authorized service facility.
• Should the CM-5 break down, do not try to disassemble and repair it by yourself. Contact a KONICA
MINOLTA-authorized service facility.
Disposal Method
• Make sure that the CM-5, its accessories, and the packing materials are either disposed of or recycled
correctly in accordance with local laws and regulations.
This manual describes how to safely operate the CM-5 using a specic procedure to perform measurement.
• Page layout
Symbols used in this manual are explained below.
* Note that the page shown in the illustration is for explanatory purposes only, and is not an actual page from
this manual.
Memo
Provides useful
information,
supplementary
explanations, and
similar details.
Notes
Provides information that
you need to know in order
to operate the CM-5
correctly. Always read
this information before
operating the CM-5.
Procedure
Indicates an operating
procedure.
Settings
Describes ranges and
provides explanations
regarding settings for the
screen in question.
Condition Setting
The CM-5 requires condition se ttings (measurement condition, m easurement option, and color) before
measurement can be st arted.
To congure condition set tings, select “Meas. Cond ition” (measurement condit ions), “Meas. Option” (measurement
Memo
options) or “Color” (display condition s) from the <Conguration> s creen to open an appropr iate screen.
If any predened cond ition is selected, the cond ition setting (measure ment condition/measu rement option/color) is
Notes
disabled. Set the condit ion to OFF before starti ng condition setting.
Setting the Measurement Conditions
To set measurement conditions, select “Me as. Condition” from the <Congurat ion> screen.
You can select or specify the following thre e items as the measurement condit ions:
Meas. Type: Select the measurement t ype such as reectance measu rement or transmitt ance •
Meas. Area: When th e measurement type is reec tance measurement, select an me asurement area.•
SCI/SCE: When the measurem ent type is reectance meas urement, select the specular c omponent •
[Setting Procedure]
E50
measurement.
mode.
Press the [Menu] button.1.
The <Conguration> scre en is displayed.
Use the 2. or button of the cross key to move the
cursor to “Meas. Condition” and then press the [Ente r/
Edit] butt on.
The <Measurement Condition Set tings> screen is
displayed.
The • <Measurement Condition Sett ings> screen shows the
current settings.
After you set the measurement conditions , you can 3.
press the [BACK] button to return to the screen
displayed before you press the [Menu] button in Step 1.
To return to the <Cong uration> screen, press t he
Memo
[Menu] but ton.
Start screen
Indicates the initial
screen from which the
operation is commenced.
Screenshot
Indicates the state of the
screen when the
operation described in
the procedure to the left
is carried out.
For the version of the instrument rmware
The version of the instrument r mware can be conr med on the <Information> screen. For details, refer to
page 152 “Displaying the Instrument Information” in this manual.
E8
Page 11
Chapter 1
Before Using the Instrument
1
Before Using the Instrument
E9
Page 12
Accessories
Standard and optional accessories are available with the instrument.
Memo
Standard Accessories
Make sure that all the following items are present.
Target Mask: ø30 mm CM-A197
During reectance measurement, this target mask can be replaced
with the optional target mask (ø3 mm or ø8 mm) to change the
illumination area (specimen measuring port size) according to the
specimen.
Transmittance Mask ø20 mm CM-A200
Attached to the instrument.
Guides light to the Transmittance Specimen Chamber when
conducting transmittance measurement.
The shape of some products may be different from those shown.
AC Adapter: AC-A305
Used to supply power from an AC outlet to the instrument.
Input: 100-240 V
Output: 5 V
Plug design:
USB Cable (2 m): I F-A19
Used to connect the instrument to a personal computer (PC).
Functional Ground Cable: CM-A211
Used to prevent communication problems caused by static
electricity when connected to a PC.
Dust Cover: CM-A191
Used to protect the instrument by preventing dust from entering the
inside.
50/60 Hz 24-38 VA
2 A
Center-negative
E10
Page 13
Optional Accessories
You may purchase the following accessories if necessary.
Petri Dish Measurement Full Set CM-A205
Petri Dish: CM-A128
Zero Calibration Box: CM-A124White Calibration Plate:
Used to perform zero calibration.Can be used in place of the built-in
Mini Petri Dish Measurement Full Set CM-A164
Mini Petri Dish: CM-A157
Target Mask (for Petri Dish):
CM-A203
Used to perform reectance
measurement using a Petri Dish.
CM-A210
(with CD-ROM containing calibration
data and data-setting software)
White Calibration Plate to perform
white calibration.
Target Mask (for Mini Petri
Dish): CM-A158
Used to perform reectance
measurement using a Mini Petri
Dish.
1
Before Using the Instrument
Calibration Glass
(for Petri Dish): CM-A212
Used for calibration for reectance
measurement using a Petri Dish.
Accessory Case: CM-A193
Calibration Glass
(for Mini Petri Dish): CM-A159
Used for calibration for reectance
measurement using a Mini Petri
Dish.
Zero Calibration Box: CM-A124White Calibration Plate:
CM-A210
(with CD-ROM containing calibration
data and data-setting software)
Used to perform zero calibration.Can be used in place of the built-in
White Calibration Plate to perform
white calibration.
Notes
Instructions and notes regarding the use of the Mini Petri Dish CM-A157 and its accessories are the same as for the
Petri Dish CM-A128 unless otherwise specied.
Accessory Case: CM-A193
E11
Page 14
Transmittance Measurement Set CM-A206
Transmittance Zero Calibration
Plate: CM-A213
A light-shielding plate used for zero
calibration for transmittance
measurement.
Transmittance Specimen Holder:
CM-A96
10 mm Wide Cell Measurement Set CM-A207
Transmittance Specimen Holder (10 mm wide): CM-A198Transmittance Zero Calibration
Transmittance Specimen Holder
Attachment: CM-A199
Used to attach the transmittance
specimen holder to the receiving
window side.
Accessory Case: CM-A193
Plate (10 mm): CM-A204
Set of accessories used to perform transmittance measurement of liquid specimen using a commercially
available 10 mm-wide Cell.
Color Management Software SpectraMagic™ NX
CM-S10 0w
Used to operate the instrument from a PC for data processing and
le management.
Connecting Cable: CM-A58
Used to connect the instrument to a commercially available serial
data printer.
E12
Page 15
Target Mask
Target Mask: ø3 mm
CM-A195
Used to switch the illumination area (specimen measuring port size) according to the specimen.
Cell
Cell: 2 mm
CM-A97
A glass container to contain the liquid specimen during transmittance measurement.
Plastic Cell
Plastic Cell: 2 mm
CM-A130
Cell: 10 mm
CM-A98
Plastic Cell: 10 mm
CM-A131
Target Mask: ø8 mm
CM-A196
Cell: 20 mm
CM-A99
Plastic Cell: 20 mm
CM-A132
1
Before Using the Instrument
A plastic container to contain the liquid specimen during transmittance measurement.
Specimen Viewing Mirror CM-A202
Used to check the measurement points on the specimen for
reectance measurement.
Color Tiles (14 colors)
E13
Page 16
System Diagram
Color Tiles (14 colors)
Target Mask: ø3 m m
CM-A195
Target Mask: ø8 mm
CM-A196
Target Mask: ø30 m m
Dust Cover
CM-A191
CM-A197
Spectrophotometer
CM-5
Transmit tance Measurement /
Petri Dish Measurement Full Set CM -A 214
(with CD-ROM contain ing calibration data and data-
setting software)
• Zero Calibration Box CM-A124
• Target Mask (for Petri Dish) CM-A203
• Petri Dish CM-A128
• Calibration Glass (for Petri Dish) CM-A212
• Accessory Case CM-A193
Mini Pet ri Dish Measurement Full Set CM-A16 4
• W hite Calibration Plate CM-A 210
(with CD-ROM contain ing calibration data and data-
setting software)
• Zero Calibration Box CM-A124
• Target Mask (for Mini Petr i Dish) CM-A158
• Mini Petri Dish CM-A157
• Calibration Glass (for Mini Pet ri Dish) CM-A159
• Accessory Case CM-A193
10 mm Wide Cell Measurement Set CM-A207
• Transmittance Specimen Holder (10 mm wide)CM-A198
• Transmittance Zero Calibration Plate (10 mm)CM-A204
Connecting Cable CM-A58
Functional Ground Cable: CM-A 211
USB Memor y
(commercially available)
USB keyboard
(commercially
ava il able)
Cell
CM-A97
(2 mm)
Plastic Cell:
2 mm
CM-A130
(x 100)
(commercially available)
Color Management Software
Spe ctraM agic™ NX
CM- S100 w
Professional Edition, Lite Ed ition
Transmittance Mask ø20 m m
CM-A200
Specimen Viewing Mirror
CM-A202
AC Adapter AC -A 305
USB Cable (2 m) IF-A19
Personal computer
Cell
CM-A98
(10 mm)
Plastic Cell:
10 mm
CM-A131
(x 100)
Data Pr inter
(commercially
ava il able)
Cell
CM-A99
(20 mm)
Plastic Cell:
20 mm
CM-A132
(x 100)
E14
10 mm-wide Cell
(commercially available)
Standard accessories
Optional accessories
Page 17
Names and Functions of Parts
34
5678
1
2
9
1
Before Using the Instrument
101112 13
E15
Page 18
LCD screen
1
Control panel
2
Specimen measuring port
3
Target Mask
4
Illumination window
5
Transmittance Mask
6
Transmittance Specimen
7
Chamber
Transmittance Specimen
8
Chamber Cover
USB connection terminal
9
(A type)
Displays setting items, measurement results and messages.
Used to switch screens or select/determine/save setting items.
For details, refer to page 23 “Control Buttons”.
The port for measuring the specimen. The port size can be changed by
changing the Target Masks. This port opens its shutter only for
measurement. Inside the shutter is the built-in White Calibration Plate.
During reectance measurement, this target mask can be replaced with
the optional target mask ø3 mm (CM-A195) or ø8 mm (CM-A196) to
change the illumination area (specimen measuring port size) according
to the specimen.
The port for measuring the specimen.
Guides light to the Transmittance Specimen Chamber when conducting
transmittance measurement.
For transmittance measurement, set a specimen in this chamber.
Slide this cover to open/close the Transmittance Specimen Chamber.
Used to attach a USB memory device to the instrument. When a USB
memory device is properly connected to the instrument, the
memory device connected) icon is displayed.
A USB keyboard can also be used to input le names and sample data
names. The screen display does not change when a USB keyboard is
attached.
(USB
USB connection terminal
10
(B type)
Power switch
11
AC adapter terminal
12
RS-232C Connector
13
Used to connect the instrument to a PC with the supplied USB cable
(I F-A19).
Used to turn ON/OFF power. Setting this switch to “” turns the power
OFF, and setting it to “ ❙ ” turns the power ON.
When using the supplied AC adapter (CM-A305), connect the adapter’s
plug to this terminal.
Used to connect the instrument to the optional data printer with the
optional connecting cable (CM-A58).
E16
Page 19
Cleaning Parts
This section explains how to clean the Target Mask, White Calibration Plate, Zero Calibration Box, and the
inside of the integrating sphere.
Target Mask and Transmittance Mask (Standard/optional accessory)
Use a blower to remove dirt and dust from the Target Masks.
Notes
Do not touch the inside (edge) of the Target Mask with your
ngers. If it becomes dirt y, wipe it with a soft, clean dry cloth.
Edge
Edge
White Calibration Plate (Optional)
• When the White Calibration Plate becomes dirty, gently wipe the
dirt off with a soft dry cloth. If the dirt is difcult to remove,
wipe it off with a cloth dampened with commercially available
lens cleaning solution. Then remove the solution with a cloth
dampened with water, and leave the plate to dry.
• When parts other than the White Calibration Plate become dirty,
lightly wipe the dirt off with a cloth dampened with water or
soapy water. Never use solvents such as thinner or benzene.
Notes
Be careful not to scratch the White Calibration Plate.
White Calibration Plate
1
Before Using the Instrument
Zero Calibration Box (Optional)
Use a blower to remove dirt from the inside of the box.
If the inner surface is accidentally touched with hands or ngers,
wipe off the ngerprints or other traces gently with a soft dry cloth.
Zero Calibration Box
CM-A124
E17
Page 20
Inside the Integrating Sphere
1. Set specular component to SCE.
Memo
2. Ensure that there is nothing in the Transmittance
Specimen Chamber.
3. Block the receiving window of the Transmittance
Specimen Chamber so that no dust or dirt enters through
it.
For the procedure for switching the specular component,
refer to “SCI /SCE” on page 62.
Receivi ng
window
4. Open the shutter and blow off the dust and dirt in the
Integrating Sphere with a blower.
Memo
Notes
For the procedure for opening/closing the shutter, refer to
page 66 “Shutter Open/Close”.
Do not touch the white-coated inner surface of the
integrating sphere, wipe it with a cloth or put an object inside
it. If the inside is so dirty that dir t cannot be removed with a
blower, contact the nearest KONICA MINOLTA-authorized
service facility.
E18
Shutter
Page 21
Receiving window of the Transmittance Specimen Chamber
1. Set measuring diameter to ø3 mm.
Memo
For the procedure for switching measuring diameter, refer to
page 61 “Measurement Area”.
2. Use a blower to remove dirt and dust from the receiving
window.
Notes
Do not insert your nger through the receiving window to
touch the lens of the light-receiving optical system.
Connecting the Functional Ground Cable
1
Before Using the Instrument
Receivi ng
window
Notes
Attach the Functional Ground Cable (CM-A211) in order to avoid communication problems with the PC caused by
static electricity.
[Procedure]
1. Connect the Y terminal of the Functional Ground Cable
to the Connection Terminal Screw on the rear of the
instrument.
Connection
Terminal Screw
Functional
Ground Cable
Grounding
Ter mi nal
Screw
2. Connect the other Y terminal of the Functional Ground Cable to a reliably grounded object (e.g.
a 100-240 VAC earth terminal or a metal desk).
E19
Page 22
Connecting the AC Adapter
Notes
1. Make sure that power is OFF (Power switch is set to
2. Connect the AC adapter’s connector plug to the adapter’s
3. Insert the AC adapter power plug into an AC outlet (100
• To supply AC power to the instr ument, always use the AC adapter (AC-A305) supplied with the instrument.
• Before connecting or disconnecting the AC adapter jack or plug, make sure that the instr ument is tur ned OFF.
[Procedure]
“”).
terminal.
to 240 VAC, 50 -60 Hz).
Turning power ON/OFF
[Procedure]
Turning power ON
1. Turn the power switch to the “ ❙ ”.
The power will be tur ned ON (energized).
[Procedure]
Turning power OFF
1. Turn the power switch to the “”.
The power will be tur ned OFF.
E20
Page 23
Items You Must Know
Initial Settings of the CM-5
When the instr ument is turned ON, “Welcome to Easy Setup Wizard!” will appear automatically in English.
For normal measurement, you do not need to change the initial settings.
Nevertheless, you can check the date setting or change the display language before using the instr ument. To do
so, use the <Conguration> screen.
The display language can be selected from eight languages including English.
For details, refer to page 93 “Other Settings”.
Control Panel
The front of the CM-5 contains the LCD screen on which the instrument displays measurement results and
messages, and the control buttons which are used to set measurement options or to change displays.
LCD screen
Operation Panel
1
Before Using the Instrument
Screen Display (LCD Screen)
The LCD screen displays measurement settings, measurement results and messages. It also indicates the status
of the instrument with icons.
The basic screen layout is shown below.
Status Bar
Screen title
Tab
Displays the device
settings and status.
For more details, refer
to the next page.
Displays measurement
values and setting
items.
E21
Page 24
Status Bar
123456789
DisplayDescription (Status)Meaning
1
2
3
4
5
6
7
8
xxxx / Auto
/ / / None
None /
None /
None /
xx:xx:xx
Ref / Tra / Pet / Liq
30 mm / 8 mm / 3 mm
Active Target numberTarget number currently set in the inst rument/
Auto Target set to ON
CalibratingZero Calibration / White Calibration / User
Calibration not performed or Calibration is
complete
Use these buttons to set items or change screens according to the guide on the LCD screen.
1
2
[Target/Sample] button
1
[Detail/List] button
2
[CAL] button
3
[MENU] button
4
Cursor key
5
(
/ / / )
3
4
Switches between <Target> screen and <Sample> screen.
When the <Sample> or <Target> screen is displayed, use this button to
switch between the detail display and list display.
Displays the <Calibration> screen.
Displays the <Conguration> screen.
Holding down the [MENU] button displays the <Welcome to Easy Setup Wizard!> screen.
Moves the cursor on the screen or changes the selected value.
To determine the item indicated by the cursor, press the [OK /Edit]
button.
56
789
1
Before Using the Instrument
[OK/Edit] button
6
[Back] button
7
[MEAS] button
8
[PRINT] button
9
Memo
Enabling/Disabling Control Buttons
Buttons for this instrument can be enabled or disabled as desired using the optional Color Management Soft ware
SpectraMagic™ NX (CM-S100w), such as to enable just the [MEAS] button.
For more infor mation, refer to the information for the “Standalone Conguration” dialog box of the Color
Management Software Spect raMagic™ NX (CM-S100w).
Sets the item indicated by the cursor or saves the setting.
Cancels the setting or retur ns to the previous step on each setting screen.
Performs calibration or measurement.
Outputs the sample data to the connected serial printer.
E23
Page 26
The screen will change depending on the button operation.
<Conguration> screen
<Measurement Condition
Settings> screen
<Sample Data Function>
screen
<Target Data Function> screen
<Sample> detail screen
<Target> detail screen
<Sample> screen
<Sample> list screen
<Target> screen
<Target> list screen
E24
Page 27
Data Saving
Data used with this instrument are saved automatically.
1
Before Using the Instrument
E25
Page 28
E26
Page 29
Chapter 2
Preparation for Measurement
2
Preparation for Measurement
E27
Page 30
Flow of Measurement
nOptional settingsnBasic procedure
Turning power ON (Page 20)
Language selection (Page 104)* As necessary (after initialization, etc.)
Condition settings (Page 59)
Target Mask replacement, etc. (Page 42)
Zero calibration (0% Calibration)
(Page 52)
White Calibration (100% Calibration) (Page 55)
*When color difference is measured*When color difference is not measured
Target color settings (Page 76)
Target color selection (Page 132)
Measurement (Page 112)
* When required depending on the
measurement conditions
* Only when required
(after initialization, etc.)
End
E28
Page 31
Quick Setting Wizard
When you use the instrument for the rst time, you can set the measurement conditions and calibrate the
instrument step by step by following the instructions on the screen.
[Procedure]
1. Turn the instrument ON. The
<Welcome to Easy Setup Wizard!>
screen appears after the initial screen.
• The Easy Setup Wizard can also be
entered by holding the [Menu] button pressed for a long time.
• If you want to save condition
settings to a USB memory device,
connect a USB memory device to
the instrument before continuing.
Use the or button of the cross key to
2.
move the cursor to “Start Simple Wizard”
and then press the [OK/Edit] button.
3. The <Measurement Type Selection>
screen appears.
2
Preparation for Measurement
Memo
To move the cursor to
select a setting item, press
the or button of the
cr oss ke y.
To move the cursor
between “Exit”,
“Previous” and “Next”,
press the or button
of the cross key.
When you select “Exit”
and then press the [OK /Edit] button, the wizard is
inter rupted and the
<Sample> screen is
displayed.
4. Go to the appropriate step according to the measurement type.
<Reectance measurement> Refer to Steps 4 to 20 (Pages 30 to 33).
<Reectance measurement using Petri Dish> Refer to Steps 4 to 18 (Pages 33 to 36).
<Transmittance measurement> Refer to Steps 4 to 13 (Pages 37 to 39).
<Liquid transmittance measurement> Refer to Steps 4 to 12 (Pages 39 to 41).
When you select
“Previous” and then press
the [OK/ Edit] button, the
screen which was
displayed immediately
before the Quick Setting
Wizard is displayed.
E29
Page 32
<Reectance measurement>
4. Use the or button of the cross
key to move the cursor to
“Reectance” and then press the
[OK/Edit] button.
The cursor moves to “Next”.
5. Press the [OK/Edit] button.
6. The <Measurement Area Selection>
screen appears.
Use the or button of the cross
key to move the cursor to the desired
measurement area and then press
the [OK/Edit] button.
The cursor moves to “Next”.
7. Press the [OK/Edit] button.
8. The <Target Mask Setting> screen
appears.
9. Attach the Target Mask for the size
of the measurement area selected.
10. Check that the cursor is on “Next”,
and press the [OK/Edit] button.
Memo
For the procedure of attaching
the Target Mask, refer to page
42.
E30
Page 33
11. The <Specular Component Selection>
screen appears.
Use the or button of the cross
key to move the cursor to the desired
specular component and then press
the [OK/Edit] button.
The cursor moves to “Next”.
12. Press the [OK/Edit] button.
13. The <Meas. Condition Conrmation>
screen appears.
Check that the cursor is on “Next”,
and press the [OK/Edit] button.
2
Preparation for Measurement
14. The <Zero Calibration> screen
appears.
The following is the explanation when
zero calibration will be skipped.
15. Check that the cursor is on “Next”,
and press the [OK/Edit] button.
16. The <White Calibration> screen
appears.
Check that the cursor is on “White
Calibration”, and press the [OK/
Edit] button.
White calibration is performed by using
the built-in White Calibration Plate of
the instrument.
Memo
Since the CM-5 stores the data
of the zero calibration
perfor med at the factory, it is
unnecessar y to repeat the zero
calibration every time you
tur n ON the instrument. For
details of zero calibration,
refer to page 52.
Memo
For details of white
calibration, refer to page 55.
E31
Page 34
17. When the white calibration is completed,
the <Auto White Cal. Selection>
screen appears.
Settings:
“Automatic”: White calibration will be
performed automatically immediately
after instrument is switched on.
“Manual”: White calibration must be
performed manually before taking
measurements.
Memo
Automatic white calibration is
convenient if measurements
are almost always performed
without changing the
measurement type or other
measurement conditions each
time.
For details of automatic white
calibration, refer to page 55.
To change settings, use the
or button of the cross key to move the
cursor to the desired setting and then
press the [OK/Edit] button.
The cursor moves to “Next”.
18. Check that the cursor is on “Next”,
and press the [OK/Edit] button.
19. If a USB memory device is not
connected, the <Meas. Cond. Settings
completed> screen appears. Skip to
step 20.
If a USB memory device is connected,
the <Save Condition File> screen
appears.
If the condition le will not be saved:
Check that the cursor is on “Next”, and
press the [OK/Edit] button. The
<Meas. Cond. Settings completed>
screen appears. Skip to step 20.
To change the le name:
Use the or button of the cross key
to move the cursor to the current le
name and press the [OK /Edit] button.
The <Edit Name> screen will appear.
Use the cross key buttons to move the
cursor to the desired character and press
the [OK/Edit] button to conrm it.
Press the [CAL] button to erase
characters.
When the le name has been input, press
the [MENU] button to move the cursor
to “Save”, and press the [OK/Edit]
button to conrm the le name and
close the screen.
To change a le name using the USB
keyboard, remove the USB memory
device while on the name editing screen,
and input the name after connecting the
USB keyboard.
Memo
The USB memor y device must
be connected to the instrument
before saving the condition le
to a USB memory device. For
information on connecting a
USB memory device, refer to
page 141 “Connecting a USB
Me mo r y”.
Memo
A USB keyboard can also be
used to input le names. For
information on connecting a
USB keyboard, refer to page
143 “Connecting a USB
Keyboard”.
Notes
Keyboard input is not possible
if the cursor is in the le name
input eld. Use the or
button of the cross key to
move the cursor to the input
character before inputting with
the keyboard.
E32
Page 35
To save the condition le:
Verify that a USB memory device is
connected.
Use the
to move the cursor to
[OK/Edit] button.
After the condition le has been saved,
the <Meas. Cond. Settings completed>
screen appears.
20. Check that the cursor is on “Finish”,
and press the [OK/Edit] button.
The Easy Setup Wizard is completed
and the <Sample> screen appears.
or button of the cross key
and press the
<Reectance measurement using Petri Dish>
4. Use the or button of the cross
key to move the cursor to “Petri
Dish” and then press the [OK/Edit]
button.
The cursor moves to “Next”.
2
Preparation for Measurement
5. Press the [OK/Edit] button.
6. The <Measurement Area Selection>
screen appears.
Use the or button of the cross
key to move the cursor to the desired
measurement area and then press
the [OK/Edit] button.
The cursor moves to “Next”.
7. Press the [OK/Edit] button.
E33
Page 36
8. The <Target Mask Setting> screen
appears.
9. Attach the optional Target Mask
(for Petri Dish) if ø30mm was
selected or the optional Target Mask
(for Mini Petri Dish) if ø3mm was
selected in step 6 above.
10. Check that the cursor is on “Next”,
and press the [OK/Edit] button.
11. The <Meas. Condition Conrmation>
screen appears.
Check that the cursor is on “Next”,
and press the [OK/Edit] button.
12. The <Zero Calibration> screen
appears.
The following is the explanation when
zero calibration will be skipped.
13. Check that the cursor is on “Next”,
and press the [OK/Edit] button.
Target Mask
(for Petri Dish)
CM-A203
Memo
For the procedure of attaching
the Target Mask, refer to page
42.
Memo
Since the CM-5 stores the data
of the zero calibration
perfor med at the factory, it is
unnecessar y to repeat the zero
calibration every time you
tur n ON the instrument. For
details of zero calibration,
refer to page 52.
14. The <White Calibration> screen
appears.
Check that the cursor is on “White
Calibration”, and press the [OK/
Edit] button.
White calibration is performed by using
the built-in White Calibration Plate of
the instrument.
E34
Memo
For details of white
calibration, refer to page 55.
Page 37
15. When the white calibration is completed,
the <Auto White Cal. Selection>
screen appears.
Settings:
“Automatic”: White calibration will be
performed automatically immediately
after instrument is switched on.
“Manual”: White calibration must be
performed manually before taking
measurements.
To change settings, use the
or button of the cross key to move the
cursor to the desired setting and then
press the [OK/Edit] button.
The cursor moves to “Next”.
16. Check that the cursor is on “Next”,
and press the [OK/Edit] button.
17. If a USB memory device is not
connected, the <Meas. Cond. Settings
completed> screen appears. Skip to
step 18.
Memo
Automatic white calibration is
convenient if measurements
are almost always performed
without changing the
measurement type or other
measurement conditions each
time.
For details of automatic white
calibration, refer to page 55.
2
Preparation for Measurement
If a USB memory device is connected,
the <Save Condition File> screen
appears.
If the condition le will not be saved:
Check that the cursor is on “Next”, and
press the [OK/Edit] button. The
<Meas. Cond. Settings completed>
screen appears. Skip to step 18.
To change the le name:
Use the or button of the cross key
to move the cursor to the current le
name and press the [OK /Edit] button.
The <Edit Name> screen will appear.
Use the cross key buttons to move the
cursor to the desired character and press
the [OK/Edit] button to conrm it.
Press the [CAL] button to erase
characters.
When the le name has been input, press
the [MENU] button to move the cursor
to “Save”, and press the [OK/Edit]
button to conrm the le name and
close the screen.
To change a le name using the USB
keyboard, remove the USB memory
device while on the name editing screen,
and input the name after connecting the
USB keyboard.
Memo
A USB keyboard can also be
used to input le names. For
information on connecting a
USB keyboard, refer to page
143 “Connecting a USB
Keyboard”.
Notes
Keyboard input is not possible
if the cursor is in the le name
input eld. Use the or
button of the cross key to
move the cursor to the input
character before inputting with
the keyboard.
E35
Page 38
To save the condition le:
Verify that a USB memory device is
connected.
Use the
to move the cursor to
[OK/Edit] button.
After the condition le has been saved,
the <Meas. Cond. Settings completed>
screen appears.
or button of the cross key
and press the
18. Check that the cursor is on “Finish”,
and press the [OK/Edit] button.
The Easy Setup Wizard is completed
and the <Sample> screen appears.
E36
Page 39
<Transmittance measurement>
4. Use the or button of the cross
key to move the cursor to the
“Transmittance” and then press the
[OK/Edit] button.
The cursor moves to “Next”.
5. Press the [OK/Edit] button.
6. The <Meas. Condition Conrmation>
screen appears.
Check that the cursor is on “Next”,
and press the [OK/Edit] button.
7. The <0% Calibration> screen appears.
The following is the explanation when
0% calibration will be skipped.
8. Check that the cursor is on “Next”,
and press the [OK/Edit] button.
9. The <100% Calibration> screen
appears.
Check that the Transmittance
Specimen Chamber is empty and
that the cursor is on “100%
Calibration”, and press the [OK/
Edit] button.
100% calibration will be performed.
2
Preparation for Measurement
Memo
Since the CM-5 stores the data
of the 0% calibration
perfor med at the factory, it is
unnecessar y to repeat the 0%
calibration every time you
tur n ON the instrument. For
details of 0% calibration, refer
to page 52.
Memo
For details of 100%
calibration, refer to page
55.
E37
Page 40
10. When the 100% calibration is
completed, the <Auto 100% Cal.
Selection> screen appears.
Settings:
“Automatic”: 100% calibration will be
performed automatically immediately
after instrument is switched on.
“Manual”: 100% calibration must be
performed manually before taking
measurements.
Memo
For details of 100%
calibration, refer to page 55.
To change settings, use the
or button of the cross key to move the
cursor to the desired setting and then
press the [OK/Edit] button.
The cursor moves to “Next”.
11. Check that the cursor is on “Next”,
and press the [OK/Edit] button.
12. If a USB memory device is not
connected, the <Meas. Cond. Settings
completed> screen appears. Skip to
step 13.
If a USB memory device is connected,
the <Save Condition File> screen
appears.
If the condition le will not be saved:
Check that the cursor is on “Next”, and
press the [OK/Edit] button. The
<Meas. Cond. Settings completed>
screen appears. Skip to step 13.
To change the le name:
Use the or button of the cross key
to move the cursor to the current le
name and press the [OK /Edit] button.
The <Edit Name> screen will appear.
Use the cross key buttons to move the
cursor to the desired character and press the
[OK/Edit] button to conrm it.Press the [CAL] button to erase characters.
When the le name has been input, press
the [MENU] button to move the cursor to
“Save”, and press the [OK /Edit] button to
conrm the le name and close the screen.
To change a le name using the USB
keyboard, remove the USB memory device
while on the name editing screen, and input
the name after connecting the USB keyboard.
Memo
Automatic 100% calibration is
convenient if measurements are
almost always perfor med without
changing the measurement type
or other measurement conditions
each time.
For details of automatic 100%
calibration, refer to page 55.
Memo
The USB memor y device must
be connected to the instrument
before saving the condition le
to a USB memory device. For
information on connecting a
USB memory device, refer to
page 141 “Connecting a USB
Me mo r y”.
Memo
A USB keyboard can also be
used to input le names. For
information on connecting a
USB keyboard, refer to page
143 “Connecting a USB
Keyboard”.
Notes
Keyboard input is not possible
if the cursor is in the le name
input eld. Use the or
button of the cross key to
move the cursor to the input
character before inputting with
the keyboard.
E38
Page 41
To save the condition le:
Verify that a USB memory device is
connected.
Use the
to move the cursor to
[OK/Edit] button.
After the condition le has been saved,
the <Meas. Cond. Settings completed>
screen appears.
or button of the cross key
and press the
13. Check that the cursor is on “Finish”,
and press the [OK/Edit] button.
The Easy Setup Wizard is completed
and the <Sample> screen appears.
<Liquid transmittance measurement>
4. Use the or button of the cross
key to move the cursor to “Liquid”
and then press the [OK/Edit]
button.
The cursor moves to “Next”.
2
Preparation for Measurement
5. Press the [OK/Edit] button.
6. The <Meas. Condition Conrmation>
screen appears.
Check that the cursor is on “Next”,
and press the [OK/Edit] button.
7. The <0% Calibration> screen appears.
The following is the explanation when
zero calibration will be skipped.
8. Check that the cursor is on “Next”,
and press the [OK/Edit] button.
Memo
Since the CM-5 stores the data
of the 0% calibration
perfor med at the factory, it is
unnecessar y to repeat the 0%
calibration every time you
tur n ON the instrument. For
details of 0% calibration, refer
to page 52.
E39
Page 42
9. The <100% Calibration> screen
appears.
Place a specimen for 100%
calibration (a container containing a
liquid with high transmittance such
as distilled water) in the
Transmittance Specimen Chamber
10. Check that the cursor is on “100%
Calibration”, and then press the
[OK/Edit] button.
100% calibration will be performed.
11. When the 100% calibration is
completed:
If a USB memory device is not
connected, the <Meas. Cond. Settings completed> screen appears. Skip to
step 12.
Light
Cell or
Plastic Cell
Memo
For details of 100%
calibration, refer to page
55.
If a USB memory device is connected,
the < Save Condition File > screen
appears.
If the condition le will not be saved:
Check that the cursor is on “Next”, and
press the [OK/Edit] button. The
<Meas. Cond. Settings completed>
screen appears. Skip to step 12.
To change the le name:
Use the or button of the cross key
to move the cursor to the current le
name and press the [OK /Edit] button.
The <Edit Name> screen will appear.
Use the cross key buttons to move the
cursor to the desired character and press
the [OK/Edit] button to conrm it.
Press the [CAL] button to erase
characters.
When the le name has been input, press
the [MENU] button to move the cursor
to “Save”, and press the [OK/Edit]
button to conrm the le name and
close the screen.
To change a le name using the USB
keyboard, remove the USB memory
device while on the name editing screen,
and input the name after connecting the
USB keyboard.
Memo
A USB keyboard can also be
used to input le names. For
information on connecting a
USB keyboard, refer to page
143 “Connecting a USB
Keyboard”.
E40
Page 43
To save the condition le:
Verify that a USB memory device is
connected.
Use the
to move the cursor to
[OK/Edit] button.
After the condition le has been saved,
the <Meas. Cond. Settings completed>
screen appears.
or button of the cross key
and press the
12. Check that the cursor is on “Finish”,
and press the [OK/Edit] button.
The Easy Setup Wizard is completed
and the <Sample> screen appears.
Memo
The USB memor y device must
be connected to the instrument
before saving the condition le
to a USB memory device. For
information on connecting a
USB memory device, refer to
page 141 “Connecting a USB
Me mo r y”.
2
Preparation for Measurement
E41
Page 44
Setting a specimen
Reectance measurement
To use the instrument for reectance measurement of specimens other than powder or paste, attach the Target
Mask to the specimen measuring port and set a specimen on it. The Target Mask can be selected from three
types based on the illumination area (specimen measuring port size): ø30 mm, ø8 mm and ø3 mm, depending on
the specimen and application.
Notes
Required accessory
When taking reectance measurements, make sure that there is no object in the transmittance specimen chamber.
Target Mask: ø30 mm
CM-A197
For LAVFor MAVFor SAV
Notes
Do not touch the inside (edge) of the Target Masks with your
ngers. If it becomes dirt y, wipe it with a soft, clean dry cloth.
Target Mask: ø8 mm
CM-A196 (O pt ional)
Target Mask: ø3 mm
CM-A195 (O pt iona l)
Attaching the Target Mask to the instrument
1. Align the Target Mask with the positioning groove on
the instrument and push the mask in.
Edge
Holding example 1
E42
Holding example 2
Page 45
Petri Dish/Mini Petri Dish measurements
Use a Petri Dish or Mini Petri Dish to allow the instrument to measure the reectance of powder or paste
specimens. The Mini Petri Dish enables measurements with much smaller amounts of specimens.
Notes
Required accessory
When taking reectance measurements, make sure that there is no object in the transmittance specimen chamber.
Petri Dish Measurement Set CM-A205 (Opt ional)
Petri Dish:
CM-A128
Zero Calibration Box:
CM-A124
Target Mask for Petri Dish:
CM-A203
White Calibration Plate:
CM-A210
(with CD-ROM containing calibration
data and data-setting software)
Mini Petri Dish Measurement Set CM-A164 (Opt ional)
Mini Petri Dish:
CM-A157
Target Mask for Mini Petri Dish:
CM-A158
Calibration Glass for Petri Dish:
CM-A212
Accessory Case:
CM-A193
Calibration Glass for Mini Petri
Dish:
CM-A159
2
Preparation for Measurement
Zero Calibration Box:
CM-A124
White Calibration Plate:
CM-A210
(with CD-ROM containing calibration
data and data-setting software)
Accessory Case:
CM-A193
E43
Page 46
Notes
Be careful that the bottom of the Petri Dish or Mini Pet ri Dish or the Calibration Glass (for Petri Dish) or Calibration Glass
(for Mini Petri Dish) does not get scratched or stained, or get ngerprints on it. If it becomes dirty, wipe it with a soft, clean
dry cloth. If dirt is dif cult to remove, wipe it off with a cloth dampened with commercially-available cleaning solution. Then
remove the solution with a cloth dampened with water and allow the plate to dry. If the scratch or stain cannot be removed,
replace the Petri Dish, Mini Petri Dish, Calibration Glass (for Petri Dish), or the Calibration Glass (for Mini Petri Dish).
Notes
When the Zero Calibration Box is not in use, be sure to attach the cap to prevent the entry of dust. If dust enters the box, blow
it off with a commercially-available blower.
Be careful that the inner surface of the Zero Calibration Box (black-coated side) does not get scratched or stained. If the
scratch or stain cannot be removed, replace the Zero Calibration Box.
Notes
The White Calibration Plate (optional accessory) may become discolored if left exposed to light. Therefore, make sure to
close the cap when it is not in use so that the W hite Calibration Plate is not exposed to ambient light.
Do not allow the White Calibration Plate (optional accessory) to get scratched or stained, or get ngerprints on it. If it
becomes dirt y, wipe it with a soft, clean dry cloth. If dirt is difcult to remove, wipe it off with a cloth dampened with
commercially-available cleaning solution. Then remove the solution with a cloth dampened with water, and leave the plate to
dry. If the scratch or st ain cannot be removed, replace the White Calibration Plate.
When you replace the White Calibration Plate (optional accessory), use the data of the new White Calibration Plate to
overwrite the white calibration data stored in the instrument as user calibration data.
E44
Page 47
Attaching the Target Mask to the instrument
1. Align the Target Mask (for Petri Dish) or Target Mask
(for Mini Petri Dish) with the positioning groove on the
instrument and push the mask in.
Notes
Although automatic zero and white calibration can be performed when using the Petri Dish or Mini Petri Dish and
are suf cient for relative measurements comparing samples with a target, for the highest accuracy, zero calibration
and white calibration should be performed using the Zero Calibration Box CM-A124 and W hite Calibration Plate
CM-A210 respectively together with the Calibration Glass (for Petri Dish) CM-A212 or Calibration Glass (for Mini
Petri Dish) CM-A159.
<To perform zero calibration>
2. When using the Petri Dish: Insert the Calibration Glass
(for Petri Dish) into the recess in the Target Mask (for
Petri Dish).
When using the Mini Petri Dish, insert the Calibration
Glass (for Mini Petri Dish) into the recess in the Target
Mask (for Mini Petri Dish).
3. Align the Zero Calibration Box with the positioning
groove on the Target Mask (for Petri Dish) or Target
Mask (for Mini Petri Dish) and push the box in.
Target Mask (for
Pet ri Dis h)
CM-A203
2
Preparation for Measurement
Calibration Glass
(for Petri Dish)
CM-A212
Zero Calibration
Box
CM-A124
<To perform white calibration>
Normally, calibration is performed using the built-in White Calibration Plate inside the shutter, so there is no
need to set the calibration plate. Use the following procedure to perform white calibration as user calibration
using an optional White Calibration Plate.
2. When using the Petri Dish: Insert the Calibration Glass
(for Petri Dish) into the recess in the Target Mask (for
Petri Dish).
When using the Mini Petri Dish, insert the Calibration
Glass (for Mini Petri Dish) into the recess in the Target
Mask (for Mini Petri Dish).
3. Align the White Calibration Plate with the positioning
groove on the Target Mask (for Petri Dish) or Target
Mask (for Mini Petri Dish) and push the plate in.
Calibration Glass
(for Petri Dish)
CM-A212
White Calibration
Plate
CM-A210
E45
Page 48
<To measure a specimen>
2. When using the Petri Dish: Insert the Petri Dish
containing a specimen into the recess in the Target
Mask (for Petri Dish).
When using the Mini Petri Dish:
Insert the Mini Petri Dish containing a specimen into
the recess in the Target Mask (for Mini Petri Dish).
Then, carefully place the Zero Calibration Box over the
Mini Petri Dish containing the specimen, align the Zero
Calibration Box with the positioning groove on the
Target Mask (for Mini Petri Dish) and press down.
When using the Mini Petri Dish, the small amount of
specimen used may result in external light penetrating the
specimen and affecting measurements. For this reason, the
Zero Calibration Box should be used to cover the Mini Petri
Dish containing the specimen to block out external light
during measurements.
Petr i Dish
CM-A128
E46
Page 49
Specimen Viewing Mirror (Optional)
The Specimen Viewing Mirror allows you to check the measurement points for a specimen when using the
instrument for reectance measurement.
How to use the Specimen Viewing Mirror
1. Place a specimen on the Target Mask.
2. Open the Transmittance Specimen Chamber Cover and
set the Specimen Viewing Mirror on the Chamber.
3. Open the shutter.
For the procedure of opening/closing the shutter, refer to page
66.
4. Look into the Specimen Viewing Mirror from the Top
and adjust the position of the specimen.
Specimen
2
Preparation for Measurement
Specimen
Viewing
Mirror
Shutter
5. Close the shutter.
For the procedure of opening/closing the shutter, refer to page
66.
6. Remove the Specimen Viewing Mirror and close the
Transmittance Specimen Chamber Cover.
Specimen
Viewing
Mirror
Shutter
E47
Page 50
Transmittance Measurement and Liquid Measurement
To use the instrument for transmittance measurement, attach the Transmittance Mask on the illumination
window of the Transmittance Specimen Chamber and then set a specimen.
The Transmittance Measurement Set allows you to set a container containing a liquid specimen or a specimen in
the form of plate or lm easily and reliably.
To conduct measurement using a commercially-available 10 mm-wide Cell, using the 10 mm Wide Cell
Measurement Set allows you to set the cell easily and reliably.
Required accessory
Transmittance Mask
ø20 mm
CM-A200
Notes
Transmittance Measurement Set CM-A206 (Opt ional)
Transmittance Zero
Calibration Plate
CM-A213
Transmittance Specimen
Holder
CM-A96
Transmittance Specimen
Holder Attachment
CM-A199
Cell (O ptional)
Cell: 2 mm
CM-A97
Cell: 10 mm
CM-A98
Plastic Cell (Optional)
Plastic Cell: 2 mm
CM-A130
Plastic Cell: 10 mm
CM-A131
Do not touch the inner surface (edge) of the
Transmit tance Mask with your ngers or wipe it
with a cloth to remove dust.
Accessory Case
CM-A193
Cell: 20 mm
CM-A99
Plastic Cell: 20 mm
CM-A132
10 mm Wide Cell Measurement Set CM-A207 (Optional)
Transmittance Specimen Holder (10 mm wide) CM-A198Transmittance Zero Calibration
Plate (10 mm) CM-A204
E48
Page 51
Attaching the Transmittance Specimen Holder to the instrument
Notes
• Do not spill a specimen or other liquid on the instrument. Doing so may cause malfunction. If any liquid is
spilled on the instrument, wipe it off immediately with a soft, dry cloth.
• Limit the thickness of a specimen to be set on the Transmittance Specimen Holder (optical path length) to 22.5
mm. (The total thickness of the Cell (20 mm) or Plastic Cell (20 mm) and the glass will be 22.5 mm.) Using a
specimen or container exceeding this thickness may cause damage.
• To move the position of a specimen after it is set in the Transmittance Specimen Holder, open the specimen
xing plate beforehand to protect the surface of a specimen or the container of a specimen.
1. Open the Transmittance Specimen Chamber Cover.
2. As shown in the gure, align the mounting screw of the
Transmittance Specimen Holder with the screw hole on
the instrument for mounting the holder and tighten the
screw with a at-blade screwdriver.
Memo
Memo
Although the Transmittance Specimen Holder can be
attached facing either the illumination window side or the
receiving window side, attach it facing the illumination
window side for normal use. In this case, the geometry will
be diffuse illumination and 0° viewing angle (di: 0°, de:
0°). If the Transmittance Specimen Holder is attached
facing the receiving window side, the geometry is
approximately 0°:0° geometry. Although the geometry
does not strictly satisfy standards for 0°:0° geometry, this
presents no problems if the values are used as relative
values.
To attach the holder with the sample or cell on the
receiving window side, attach the CM-A199 Transmittance
Specimen Holder Attachment before the holder.
Transmittance
Specimen Holder
Attachment
CM-A96
2
Preparation for Measurement
Example when the
holder is at tached on
the illumination
window side
Transmittance
Specimen Holder
CM-A96
Example when
the holder is
attached on the
receiving
window side
Transmittance
Specimen Holder
CM-A96
E49
Page 52
Setting a lm specimen
Notes
Water drops for med on a specimen due to condensation or other causes prevent accurate measurement.
1. Open the specimen xing plate and secure a specimen
with the plate.
Notes
When setting the specimen on the illumination window
side, position it so that it fully covers the illumination
window. When setting the specimen on the receiving
window side, position it so that it fully covers the receiving
wi n do w.
Setting a liquid specimen
Notes
• As for the container for a liquid specimen, it is recommended to use the CM-A97 to 99 Cell or CM-A130 to 132
Plastic Cell provided as the optional accessor y. To use other cont ainers, use a colorless, transparent container
with surfaces which are parallel with the illumination window and receiving window of the instr ument.
• The Cell is fragile and it must be handled with care.
• Do not pour hot liquid in the Cell or Plastic Cell, and do not wash them with hot water.
• W hen cleaning the Cell or Plastic Cell with an ult rasonic cleaner, do not clean them for a long time to prevent
damage.
• The measurement values obtained with cells having different optical path lengths may not be compatible with
each ot her.
• Do not spill a specimen or other liquid on the instrument. Doing so may cause malfunction. If any liquid is
spilled on the instrument, wipe it off immediately with a soft, dry cloth.
• Do not measure inammable liquid.
• Ensure that there are no dust, stains, scratches, or nger prints on the specimen or on the inner and outer surfaces
of the container of a liquid specimen. Set a specimen by holding a section other than the measurement su rface to
avoid ngerprints.
• Bubbles in the container containing a liquid specimen prevent accurate measurement. Be sure to remove bubbles
before measurement. (Sometimes bubbles may be generated when a specimen is left for a while.)
• Water drops formed on a specimen or the container containing a liquid specimen due to condensation or other
causes prevent accurate measurement.
• The measurement accuracy for liquid containing ne par ticles may be unstable due to the deposition of the
particles.
Illumination
window
Memo
When measuring a transparent index such as Gardner, etc., use the Transmittance Specimen Holder (10 mm wide)
and a 10 mm-wide cell with an optical path length of 10 mm.
1. Pour a specimen in a transparent container.
2. Open the specimen xing plate and secure the container
with the plate.
Notes
• W hen setting the specimen on the illumination window
side, position it so that it fully covers the illumination
window. When setting the specimen on the receiving
window side, position it so that it fully covers the
receiving window.
• W hen using a cell or plastic cell, use the cell guide to
determine the center of the cell.
E50
Light
Cell or Plastic
Cell
Specimen
xing plate
Page 53
Measuring a liquid specimen by using a commercially-available 10 mm-wide cell
Notes
Memo
• As for the container for a liquid specimen, use a colorless, transparent container with surfaces which make
parallel with the illumination window and receiving window of the instrument.
• For 10 mm-wide cells, use cells which have an outside width dimension of 12.5 mm. Cells with an outside width
of 12 mm cannot be set securely.
• The measurement values obtained with cells having different optical path lengths may not be compatible with
each ot her.
• Do not spill a specimen or other liquid on the instrument. Doing so may cause malfunction. If any liquid is
spilled on the instrument, wipe it off immediately with a soft, dry cloth.
• Do not measure inammable liquid.
• Ensure that there are no dust, stains, scratches, or nger prints on the specimen or on the inner and outer surfaces
of the container of a liquid specimen. Set a specimen by holding a section other than the measurement su rface to
avoid ngerprints.
• Bubbles in the container containing a liquid specimen prevent accurate measurement. Be sure to remove bubbles
before measurement. (Sometimes bubbles may be generated when a specimen is left for a while.)
• Water drops formed on a specimen or the container containing a liquid specimen due to condensation or other
causes prevent accurate measurement.
• The measurement accuracy for liquid containing ne par ticles may be unstable due to the deposition of the
particles.
When measuring a transparent index such as Gardner, etc., use the Transmittance Specimen Holder (10 mm wide)
and a 10 mm-wide cell with an optical path length of 10 mm.
1. Open the Transmittance Specimen Chamber Cover.
2
Preparation for Measurement
2. As shown in the gure, align the mounting screw of the
Transmittance Specimen Holder (10 mm wide) with the
screw hole on the instrument for mounting the holder
and tighten the screw with a at-blade screwdriver.
3. Change the position of the cell holding plate to one of
the 10 mm, 20 mm or 50 mm groove on the
Transmittance Specimen Holder (10 mm wide)
according to the optical path length of the cell to be
used.
4. Pour a specimen in a transparent, 10 mm-wide
co nta i ner.
5. Set the container to the Transmittance Specimen
Holder (10 mm wide).
Notes
Be sure to set the cell so that the transparent surfaces of the
cell are or iented to the illumination and receiving
windows.
Transmittance
Specimen Holder
(10 mm)
CM-A198
10 mm
20 mm
50 mm
Light
E51
Page 54
Calibration
Zero Calibration (0% Calibration)
Since the CM-5 stores the data of the zero calibration (0% Calibration) performed at the factory, it is
unnecessary to repeat the zero calibration (0% Calibration) every time you turn ON the instrument. If, however,
the measurement conditions change greatly, or if the optional Target Mask (ø8 mm or ø3 mm), Petri Dish, Mini
Petri Dish, or Cell is used, you need to perform zero calibration (0% Calibration) before white calibration (100%
Calibration).
Memo
Notes
• The effects of stray light (i.e. light generated due to the are character istics of the optical system) will be
compensated for automatically by the zero calibration (0% Calibration) data.
• The amount of st ray light may change because of dust or dirt which has collected in the optical system, humidity,
repeated operation, or vibration and shock exerted on the instrument. In this case, performing zero calibration
(0% Calibration) periodically is recommended.
• If the inst rument is not used for long periods of time, the zero calibration (0% Calibration) data stored in the
instrument may be lost. If the data is lost, zero calibration (0% Calibration) must be performed again.
• If you performed zero calibration (0% Calibration) using the optional Target Mask (ø8 mm or ø3 mm), Petri Dish
or Cell, you must perform zero calibration (0% Calibration) again before conducting measurement using the
standard Target Mask.
[Procedure]
Although zero calibration (0% Calibration) is included in the Easy Setup Wizard, the following explains the
procedure to perform zero calibration (0% Calibration) from the measurement screen.
Start the procedure from the measurement screen.
1. Press the [CAL] button.
The <Measurement Condition Settings> screen is
displayed.
Notes
Memo
Conrm that the displayed measurement type,
measurement area and specular component setting match
with conditions of the measurement to be conducted.
The bottom of the <Measurement Condit ion Settings>
screen shows two items: Zero Calibration (0% Calibration)
and White Calibration (100% Calibration). The date of the
last calibration is displayed for each item. If no calibration
has been performed yet, “--” is displayed.
2. Use the or button and the or button of the
cross key to move the cursor to
Zero Calibration
(0% Calibration).
3. Go to the appropriate step according to the measurement type.
<Reectance measurement> Refer to Steps 3 to 4 (Page 53).
<Reectance measurement using Petri Dish> Refer to Steps 3 to 5 (Page 53).
<Transmittance measurement> Refer to Steps 3 to 4 (Page 54).
<Liquid transmittance measurement> Refer to Steps 3 to 4 (Page 54).
E52
Page 55
<Reectance measurement>
3. Clear an area of 1 m radius from the specimen
measuring port of the instrument.
Keep everything including light sources (such as uorescent
lamps and other illumination) and reective items (hands,
desks, walls etc.) more than 1 m away from the specimen
measuring port.
Memo
Use of the optional Zero Calibration Box CM-A182 ensures
proper zero calibration.
Within 1 m
4. Press the [OK/Edit] button.
Zero calibration will be performed.
The Xe lamp ashes three times during the calibration.
During zero calibration, the screen shows the number of
ashes performed.
When the zero calibration is completed, the cursor moves to
White Calibration.
<Reectance measurement using Petri Dish>
3. Insert the Calibration Glass for Petri Dish into the recess
in the Target Mask for Petri Dish.
4. Align the Zero Calibration Box with the positioning
groove on the Target Mask for Petri Dish and push the
box in.
2
Preparation for Measurement
Calibration Glass
(for Petri Dish)
CM-A212
Zero Calibration
Box
CM-A124
5. Press the [OK/Edit] button.
Zero calibration will be performed.
The Xe lamp ashes three times during the calibration.
During zero calibration, the screen shows the number of
ashes performed.
When the zero calibration is completed, the cursor moves to
White Calibration.
E53
Page 56
<Transmittance measurement>
3. Open the Transmittance Specimen Chamber Cover and
set the Transmittance Zero Calibration Plate so that it
fully covers the illumination window.
Notes
Set the Transmittance Zero Calibration Plate horizontally,
as shown to the right. If the Transmittance Zero Calibration
Plate is set ver tically, the Transmitt ance Specimen
Chamber Cover may not close.
4. Close the Transmittance Specimen Chamber Cover and
press the [OK/Edit] button.
0% calibration will be performed.
The Xe lamp ashes three times during the calibration.
During 0% calibration, the screen shows the number of ashes
performed.
When the 0% calibration is completed, the cursor moves to
100% Calibration.
<Liquid transmittance measurement>
3. Open the Transmittance Specimen Chamber Cover and
set the Transmittance Zero Calibration Plate so that it
fully covers the receiving window. (Refer to the gure
for the case of transmittance measurement.)
To use the 10 mm Wide Cell Measurement Set, set the
Transmittance Zero Calibration Plate (10 mm) so that it fully
covers the receiving window.
Transmittance
Zero Calibration
Plate
CM-A213
Transmittance Zero
Calibration Plate
CM-A204
4. Close the Transmittance Specimen Chamber Cover and
press the [OK/Edit] button.
0% calibration will be performed.
The Xe lamp ashes three times during the calibration.
During 0% calibration, the screen shows the number of ashes
performed.
When the 0% calibration is completed, the cursor moves to
100% Calibration.
Notes
• After zero calibration (0% Calibration), you need to
perfor m white calibration (100% Calibration). Go to step
3 of “White Calibration (100% Calibration)” on page 55.
• Even though you performed white calibration (100%
Calibration) after turning ON the power, you need to
perfor m it again if you performed zero calibration (0%
Calibration) after the rst white calibration.
E54
Page 57
White Calibration (100% Calibration)
White calibration (100% Calibration) must be performed prior to start of measurement after the power is turned
ON for the rst time after purchase at the current settings.
Memo
Notes
• The built-in W hite Calibration Plate of the instrument is provided with its own calibration data.
• W hen automatic white calibration (100% Calibration) is set to ON, white calibration (100% Calibration) will
star t automatically when the instrument is turned on.
(Except for liquid measurement or user calibration)
• The reading may uctuate slightly due to changes in the ambient temperature or due to heat generation caused by
repeated operation of the instru ment. In this case, make sure to perfor m white calibration (100% Calibration)
regularly.
• W hite calibration (100% Calibration) must be performed at the same temperat ure as the one at which
measurement will be performed.
• Before starting white calibration (100% Calibration), make sure that the instrument is sufciently acclimatized
to the ambient temperature.
2
Preparation for Measurement
[Procedure]
Although white calibration (100% Calibration) is included in the Easy Setup Wizard, the following explains the
procedure to perform white calibration (100% Calibration) from the measurement screen.
Start the procedure from the measurement screen.
1. Press the [CAL] button.
The <Measurement Condition Settings> screen is
displayed.
Notes
Memo
Conrm that the displayed measurement type,
measurement area and specular component setting match
with conditions of the measurement to be conducted.
The bottom of the <Measurement Condit ion Settings>
screen shows two items: Zero Calibration (0% Calibration)
and White Calibration (100% Calibration). The date of the
last calibration is displayed for each item. If no calibration
has been performed yet, “--” is displayed.
2. Use the or button and or button of the
cross key to move the cursor to
White Calibration
(100% cal ibration).
3. Go to the appropriate step according to the measurement type.
<Reectance measurement>Refer to Step 3 (Page 56).
<Reectance measurement using Petri Dish>Refer to Steps 3 to 5 (Page 56).
<Transmittance measurement>Refer to Steps 3 to 4 (Page 57).
<Liquid transmittance measurement>Refer to Steps 3 to 4 (Page 57).
E55
Page 58
<Reectance measurement>
3. Press the [OK/Edit] button.
White calibration is performed by using the built-in White
Calibration Plate of the instrument.
The Xe lamp ashes three times during the calibration.
During white calibration, the screen shows the number of
ashes performed.
When the white calibration is completed, the screen returns to
the one displayed before the <Measurement Condition Settings> screen.
<Reectance measurement using Petri Dish (1)>
(Using the built-in White Calibration Plate to perfor m white calibration)
3. Press the [OK/Edit] button.
White calibration is performed by using the built-in White
Calibration Plate of the instrument.
The Xe lamp ashes three times during the calibration.
During white calibration, the screen shows the number of
ashes performed.
When the white calibration is completed, the screen returns to
the one displayed before the <Measurement Condition Settings> screen.
<Reectance measurement using Petri Dish (2)>
(Using the optional W hite Calibration Plate to perform white calibration as user calibration)
3. Insert the Calibration Glass for Petri Dish into the
recess in the Target Mask for Petri Dish.
4. Align the White Calibration Plate with the positioning
groove on the Target Mask for Petri Dish and push the
plate in.
5. Press the [OK/Edit] button.
White calibration will be performed.
The Xe lamp ashes three times during the calibration.
During white calibration, the screen shows the number of
ashes performed.
When the white calibration is completed, the screen returns to
the one displayed before the <Measurement Condition Settings> screen.
Calibration Glass
(for Petri Dish)
CM-A212
White Calibration
Plate
CM-A210
E56
Page 59
<Transmittance measurement>
3. Open the Transmittance Specimen Chamber Cover and
empty the space between the illumination and receiving
windows. (No specimen, cell or Zero Calibration Plate
being set.)
Notes
100% calibration can be performed with the Transmittance
Specimen Holder installed.
4. Close the Transmittance Specimen Chamber Cover and
press the [OK/Edit] button.
100% calibration (air calibration) is performed.
The Xe lamp ashes three times during the calibration.
During 100% calibration, the screen shows the number of
ashes performed.
When the 100% calibration is completed, the screen returns to
the one displayed before the <Measurement Condition
Settings> screen.
<Liquid transmittance measurement>
3. Open the Transmittance Specimen Chamber Cover and
set the cell lled with distilled water or puried water
to more than two-thirds of the capacity in the
Transmittance Specimen Holder.
Notes
Be sure to use cells with the same optical path length for
the calibration and measurement.
Set the cell so that its transparent surfaces are oriented to
the illumination and receiving windows.
2
Preparation for Measurement
Cell or Plastic
Cell
Light
4. Close the Transmittance Specimen Chamber Cover and
press the [OK/Edit] button.
100% calibration (distilled water calibration) is performed.
The Xe lamp ashes three times during the calibration.
During 100% calibration, the screen shows the number of
ashes performed.
When the 100% calibration is completed, the screen returns to
the one displayed before the <Measurement Condition
Settings> screen.
Memo
If you change the measurement type or other measurement
conditions after white calibration (100% calibration), you
need to perform white calibration (100% calibration) again
under the new conditions.
E57
Page 60
User Calibration
You can perform calibration by using your own reference plate and calibration data instead of the white
calibration (100% calibration) data. The calibration data for user calibration can be specied by connecting the
instrument to a PC and using the optional Color Management Software “SpectraMagic™ NX”.
You can select whether to use the user calibration data for measurement on the <Calibration Option Setting> screen of the instrument.
[Procedure]
Start the procedure from the measurement screen.
1. Press the [MENU] button.
The <Conguration> screen is displayed.
2. Use the or button of the cross key to move the
cursor to
calibration options and then press the
[OK/Edit] button.
The <Calibration Option Settings> screen is displayed.
3. Use the or button of the cross key to move the
cursor to the desired calibration data and then press the
[OK/Edit] button.
The <Calibration Options> screen is displayed.
4. Use the or button of the cross key to move the
cursor to “User Calibration” and then press the [OK/
Edit] button.
Now you can perform white calibration (100% calibration)
using your own calibration data.
Notes
You need to prepare user calibration data and write it to the
instrument memory before you can set user calibration to
ON.
E58
Page 61
Condition Setting
The CM-5 requires condition settings (measurement condition, measurement option, and color) before
measurement can be started.
Memo
Setting the Measurement Conditions
To set measurement conditions, select “Meas. Condition” from the <Conguration> screen.
You can select or specify the following three items as the measurement conditions:
• Meas. Type: Select the measurement type such as reectance measurement or transmittance
• Meas. Area: When the measurement type is reectance measurement, select an measurement area.
• SCI/SCE: When the measurement type is reectance measurement, select the specular component
options) or “Color” (display conditions) from the <Conguration> screen to open an appropriate screen.
measurement.
mode.
The <Conguration> screen is displayed.
2
Preparation for Measurement
2. Use the or button of the cross key to move the
cursor to “Meas. Condition” and then press the [OK/
Edit] button.
The <Measurement Condition Settings> screen is
displayed.
• The <Measurement Condition Settings> screen shows the
current settings.
3. After you set the measurement conditions, you can
press the [Back] button to return to the screen
displayed before you press the [MENU] button in Step
1.
Memo
To return to the <Conguration> screen, press the
[MENU] button.
E59
Page 62
Mea sur e me n t Ty pe
Select the measurement type such as reectance measurement or transmittance measurement.
[Setting Procedure]
Start the procedure from the <Measurement Condition Settings> screen.
1. Use the or button of the cross key to move the
cursor to “Meas. Type” and then press the [OK/Edit]
button.
The <Measurement Type Settings> screen is displayed.
2. Use the or button of the cross key to move the
cursor to the desired item.
Settings
Reectance: Measure the reectance of object colors.
Transmittance: Measure the transmittance of specimens in
the form of plate or lm.
Petri Dish: Measure the reectance of powder or paste
specimens by using a Petri Dish.
Liquid: Measure the transmittance of liquid specimens.
3. Press the [OK/Edit] button.
The selection is conrmed and the screen returns to the
< Measurement Condition Settings> screen.
Notes
If you press the [Back] button without pressing the [OK/
Edit] button, you return to the <Measurement Condition
Settings> screen without changing the setting.
E60
Page 63
Measurement Area
When the measurement type is Reectance or Petri Dish, select the measurement area.
[Setting Procedure]
1. Use the or button of the cross key to move the
cursor to “Meas. Area” and then press the [OK/Edit]
button.
The <Measurement Area Settings> screen is displayed.
2. Use the or button of the cross key to move the
cursor to the desired item.
Settings
ø30 mm: Measurement area of 30 mm(LAV)
ø8 mm: Measurement area of 8 mm (MAV); using the
optional ø8 mm Target Mask (CM-A195)
ø3 mm: Measurement area of 3 mm (SAV); using the
optional ø3 mm Target Mask (CM-A196)
3. Press the [OK/Edit] button.
The selection is conrmed and the screen returns to the
<Measurement Condition Settings> screen.
Notes
If you press the [Back] button without pressing the [OK/
Edit] button, you return to the <Measurement Condition
Settings> screen without changing the setting.
Start the procedure from the <Measurement Condition Settings> screen.
2
Preparation for Measurement
E61
Page 64
Specular Component (SCI/SCE)
When the measurement type is reectance measurement, select the specular component mode.
When the Petri Dish measurement is selected, “SCE” is set automatically.
[Setting Procedure]
Start the procedure from the <Measurement Condition Settings> screen.
1. Use the or button of the cross key to move the
cursor to “SCI/SCE” and then press the [OK/Edit]
button.
The <Measurement Mode Settings> screen is displayed.
2. Use the or button of the cross key to move the
cursor to the desired item.
Settings
SCI: Specular component included
SCE: Specular component excluded
3. Press the [OK/Edit] button.
The selection is conrmed and the screen returns to the
<Measurement Condition Settings> screen.
Notes
If you press the [Back] button without pressing the [OK/
Edit] button, you return to the <Measurement Condition
Settings> screen without changing the setting.
E62
Page 65
Measurement Option Settings
To set measurement options, select “Measurement Options” from the <Conguration> screen.
You can select or specify the following three items as the measurement options:
• Auto Measurement (1-10): Specify the number of measurements for auto averaging.
• Manual Measurement (1-30): Specify the number of measurements for manual averaging.
• Shutter Open/Close: Open or close the shutter of the specimen measuring port.
[Setting Procedure]
1. Press the [MENU] button.
The <Conguration> screen is displayed.
2. Use the or button of the cross key to move the
cursor to “Meas. Option” and then press the [OK/Edit]
button.
The <Measurement Options> screen is displayed.
• The <Measurement Options> screen shows the current
settings.
3. After you set the measurement conditions, press the
[Back] button to return to the <Conguration> screen.
2
Preparation for Measurement
E63
Page 66
Auto Measurement (1-10)
Specify the number of measurements for auto averaging.
The average of the data obtained from the specied number of continuous measurements is determined as
sample data.
[Setting Procedure]
Start the procedure from the <Measurement Options> screen.
1. Use the or button of the cross key to move the
cursor to “Auto Measurement (1-10)” and then press the
[OK/Edit] button.
Above and below the cursor and are displayed.
2. Use the or button of the cross key to specify a
value.
Settings
1 to 10 times:
Specify the number of measurements, from 1 to 10 times, to
perform when using auto averaging.
3. Press the [OK/Edit] button.
The selection is conrmed.
Notes
If you press the [Back] button without pressing the [OK/
Edit] button, you return to the <Measurement Options>
screen without changing the setting.
E64
Page 67
Manual Measurement (1-30)
Specify the number of measurements for manual averaging.
The average of the data obtained from the measurements conducted with the presses of the [MEAS] button for
the specied number of times is determined as sample data.
[Setting Procedure]
1. Use the or button of the cross key to move the
cursor to “Manual Measurement (1-30)” and then press
the [OK/Edit] button.
Above and below the cursor and are displayed.
2. Use the or button of the cross key to specify a
value.
Settings
1 to 30 times:
Specify the number of measurements, from 1 to 30 times, to
perform when using manual averaging.
3. Press the [OK/Edit] button.
The selection is conrmed.
Notes
If you press the [Back] button without pressing the [OK/
Edit] button, you return to the <Measurement Options>
screen without changing the setting.
Start the procedure from the <Measurement Options> screen.
2
Preparation for Measurement
E65
Page 68
Shutter Open/Close
When cleaning the inside of the integrating sphere or to check the measurement points for a specimen using the
Specimen Viewing Mirror (Optional), you need to open or close the shutter of the specimen measuring port. Use
this option for the open/close operation.
During measurement, the shutter is automatically opened or closed according to the measurement type.
[Setting Procedure]
Start the procedure from the <Measurement Options> screen.
1. Use the or button of the cross key to move the
cursor to “Shutter Open/Close” and then press the
[OK/Edit] button.
Above and below the cursor and are displayed.
2. Use the or button of the cross key to specify either
to open or close the shutter.
Settings
Open: Opens the shutter
Close: Closes the shutter
3. Press the [OK/Edit] button.
The shutter opens or closes according to the selection.
Notes
If you press the [Back] button without pressing the [OK/
Edit] button, you return to the <Measurement Options>
screen without changing the setting.
E66
Memo
For the transmittance measurement, the geometry will be set to di:0° when the shutter is closed and will be set to
de:0° when the shutter is open and the Zero Calibration Box is set.
In the haze measu rement, the optional Color Management Software SpectraMagic™ NX (CM-S100w) allows haze
calculation after automatic selection of the geometry (shutter open/close).
Page 69
Setting the Display Conditions
To set display conditions, select “Color” from the <Conguration> screen.
You can select or specify the following ve items as the display conditions:
• Color Space: Select the color space to be displayed.
• Color Index: Select the index (WI, YI, etc.) to display.
• Observer: Select the observer angle: 2° or 10°.
• Illuminant 1: Select the illuminant used to measure colorimetric data.
• Illuminant 2: Select the secondary illuminant used for MI (metamerism index) calculation, etc.
[Setting Procedure]
1. Press the [MENU] button.
The <Conguration> screen is displayed.
2. Use the or button of the cross key to move the
cursor to “Color” and then press the [OK/Edit] button.
The <Color Settings> screen is displayed.
• The <Color Settings> screen shows the current settings.
2
Preparation for Measurement
3. After you set the display conditions, press the [Back]
button to return to the <Conguration> screen.
E67
Page 70
Color Space
Select the color space to be used or the index used for transmittance measurement.
[Setting Procedure]
Start the procedure from the <Color Settings> screen.
1. Use the or button of the cross key to move the
cursor to the desired “Color Space” and then press the
[OK/Edit] button.
The <Color Space Settings> screen is displayed.
2. Use the or button of the cross key to move the
cursor to the desired item.
• Use the or button of the cross key to move the cursor
to the digit on the left or right.
Settings
L*a*b*: L*a*b* color space
L*C*h: L*C*h color space
Hunter Lab: Hunter Lab color space
Yxy: Yxy color space
XYZ: XYZ color space
Munsell (C): Munsell color space (illuminant “C”)
Munsell (D65): Munsell color space (illuminant “D
Iodine Color Value: Transparent index, Iodine color scale
Ph. EU: European Transparent index, Pharmacopoeia
US Pharmacopeia: Transparent index, US Pharmacopeia
Memo
• A value for Munsell (C) is displayed only when observer
2° and illuminant “C” are selected.
• A value for Munsell (D65) is displayed only when
observer 2° and illuminant “D65” are selected.
• A value for Gardner is displayed only when observer 2°
and illuminant “C” are selected.
• A value for Hazen /APHA is displayed only when
observer 2° and illuminant “C” are selected.
• A value for Iodine is displayed only when observer 2°
and illuminant “C” are selected.
• A value for Ph. EU is displayed only when observer 2°
and illuminant “C” are selected.
• A value for US Pharmacopeia is displayed only when
observer 2° and illuminant “C” are selected.
E68
Page 71
3. Press the [OK/Edit] button.
If you select L*a*b* or L*C*h, <Difference Formula
Settings> screen is displayed.
If you select Ph. EU, <Ph.EU Hue setting> screen is
displayed.
If you select options other than above, the selection is
conrmed and the screen returns to the <Color Settings>
screen.
Notes
If you press the [Back] button without pressing the [OK/
Edit] button, you return to the <Color Settings> screen
without changing the setting.
4. If you select L*a*b*, L*C*h or Ph. EU, use the or
button of the cross key to move the cursor to the desired
item.
Settings (When L*a*b* or L*C*h is selected)
ΔE*ab: ΔE*ab (CIE1976) color difference formula
CMC: CMC color difference formula; parameters can be
changed.
ΔE*94: ΔE*94 (CIE1994) color difference formula;
parameters can be changed.
ΔE00: ΔE00 (CIE DE2000) color difference formula;
parameters can be changed.
Settings (When Ph. EU is selected)
Auto: Aut oma tic
B: Br own
BY: Brownish-Yellow
Y: Ye l l o w
GY: Greenish-Yellow
R: Red
5. Press the [OK/Edit] button.
The selection is conrmed and the screen returns to the
<Color Settings> screen.
Notes
If you press the [Back] button without pressing the [OK/
Edit] button, you return to the <Color Space> screen
without changing the setting.
2
Preparation for Measurement
E69
Page 72
Color Index
Select the index (WI, YI, etc.) to be used.
This screen is available when the user index has been set with the optional Color Management Software
SpectraMagic™ NX (CM-S100w) in advance.
[Setting Procedure]
Start the procedure from the <Color Settings> screen.
1. Use the or button of the cross key to move the
cursor to “Color Index” and then press the [OK/Edit]
button.
The <Color Index> screen is displayed.
A scroll bar to the right of the <Color Index> screen,
signies that some color indexes cannot be displayed on the
screen.
2. Use the or button of the cross key to move the
cursor to the desired item.
Settings
Move the cursor up from the top of the list, or down from the
bottom of the list, to display those indexes that cannot be
displayed on the screen.
WI ASTM E313-73: Whiteness index (ASTM E313-73)
WI ASTM E313-96: Whiteness index (ASTM E313-96)
WI Hunter: Whiteness index (Hunter)
YI ASTM E313-73: Yellowness index (ASTM E313-73)
YI ASTM E313-96: Yellowness index (ASTM E313-96)
YI ASTM D1925: Yellowness index (ASTM D1925)
B ASTM E313-73
ISO Brig htnes s
None
Memo
• A value for WI ASTM E313-73 is displayed only
when observer 2° and illuminant “C” are selected.
• A value for WI ASTM E313-96 is displayed only
when illuminant “C”, “D50” or “D65” is selected.
• A value for WI Hunter is displayed only observer
2° and illuminant “C” are selected.
• A value for YI ASTM E313-73 is displayed only
when observer 2° and illuminant “C” are selected.
• A value for YI ASTM E313-96 is displayed only
when illuminant “C”/“D65” are selected.
• A value for YI YI D1925 is displayed only when
illuminant “C” is selected.
• A value for B ASTM E313-73 is displayed only
when observer 2° and illuminant “C” are selected.
E70
Page 73
3. Press the [OK/Edit] button.
The selection is conrmed and the screen returns to the
<Color Settings> screen.
Notes
If you press the [Back] button without pressing the [OK/
Edit] button, you return to the <Color Settings> screen
without changing the setting.
2
Preparation for Measurement
E71
Page 74
Observer
Select the observer angle: 2° or 10°.
[Setting Procedure]
Start the procedure from the <Color Settings> screen.
1. Use the or button of the cross key to move the
cursor to the desired “Observer” and then press the
[OK/Edit] button.
The <Observer Settings> screen is displayed.
2. Use the or button of the cross key to move the
cursor to the desired item.
The selection is conrmed and the screen returns to the
<Color Settings> screen.
Notes
If you press the [Back] button without pressing the [OK/
EDIT] button, you return to the <Color Settings> screen
without changing the setting.
E72
Page 75
Illuminant 1
Select the illuminant used to measure colorimetric data.
[Setting Procedure]
Start the procedure from the <Color Settings> screen.
1. Use the or button of the cross key to move the
cursor to the desired “Illuminant 1” and then press the
[OK/Edit] button.
The <Primary Illuminant> screen is displayed.
2. Use the or button of the cross key to move the
cursor to the desired item.
Settings
AStandard illuminant A (Incandescent lamp, Color
temperature: 2856K)
CIlluminant C; (Daylight, The relative value of the
spectral distribution in the ultraviolet region is
small; Color temperature: 6774K)
D65Standard illuminant D65 (Daylight, Color
temperature: 6504K)
D50Standard illuminant D50 (Daylight, Color
temperature: 5003K)
ID65 Indoor daylight illuminant ID65 (Daylight after
passing through window glass, Color temperature:
6504K)
ID50 Indoor daylight illuminant ID50 (Daylight after
passing through window glass, Color temperature:
5003K
F2Cool white (uorescent lamp)
F6Cool white (uorescent lamp)
F7Color rendering A daylight white (uorescent
lamp)
F8Color rendering AAA natural white (uorescent
lamp)
F103-band type natural white (uorescent lamp)
F113-band type cool white (uorescent lamp)
F123-band type warm white (uorescent lamp)
2
Preparation for Measurement
E73
Page 76
3. Press the [OK/Edit] button.
The selection is conrmed and the screen returns to the
<Color Settings> screen.
Notes
If you press the [Back] button without pressing the [OK/
Edit] button, you return to the <Color Settings> screen
without changing the setting.
E74
Page 77
Illuminant 2
Select the secondary illuminant used for MI (metamerism index) calculation, etc.
[Setting Procedure]
1. Use the or button of the cross key to move the
cursor to the desired “Illuminant 2” and then press the
[OK/Edit] button.
The <Secondary Illuminant> screen is displayed.
2. Use the or button of the cross key to move the
cursor to the desired item.
Settings
The setting values are the same as those for “Illuminant 1”
and “None”.
3. Press the [OK/Edit] button.
The selection is conrmed and the screen returns to the
<Color Settings> screen.
Notes
If you press the [Back] button without pressing the [OK/
EDIT] button, you return to the <Color Settings> screen
without changing the setting.
Start the procedure from the <Color Settings> screen.
2
Preparation for Measurement
E75
Page 78
Color Difference Target Color Data Operation
Settings
To measure the color difference between two specimens, the color of one of the specimens must be set as the
target color. The CM-5 can store up to 1,000 target colors.
When using the instrument alone, measure a target specimen by following the procedure below and set the result
as the target color.
Memo
Notes
[Setting Procedure]
1. From the <Sample> screen, press the [Target/Sample]
• Target colors are stored with set ting numbers from 0001 to 1000 being assigned. Even when the data at some
midpoint is deleted, these setting numbers do not change. It is useful to group data by assigning numbers of
specic digits.
• To set more accurate target color data, use averaging to measure the target specimen. For details, refer to
“Average Measurement” on page 134.
• Be sure to perform white calibration before setting target colors.
• To ensure measurement accuracy, make sure to keep ambient conditions (temperature, etc.) constant.
button.
The <Target> screen is displayed.
• Pressing the [Detail/List] button switches between the
<Target> screen and <Target List> screen.
2. Use the or button of the cross key to move to a
desired target color No.
Notes
Unlike measured data, the numbers assigned to target color
data do not change automatically. When you measure
colors continuously to set target colors, you need to move
the cursor manually to set each data.
3. Place a specimen on the Target Mask or Transmittance
Specimen Chamber according to the measurement type.
Specimen
E76
Page 79
4. Make sure that (Ready to measure) is displayed or
the Ready lamp is green, and then press the measuring
button.
The specimen is measured and the result is displayed on the
screen.
Notes
If you selected a number to which target color data has
already been set, a message is displayed to conrm
overwriting.
5. To set more target colors, repeat steps 2 to 4.
2
Preparation for Measurement
E77
Page 80
Editing Target Color Data
The <Target Data Function> screen allows the following operations for target color data.
“Current” tab
• Save: Saves the current target color data to a USB memory device.
• Statistics: Performs statistical calculations using multiple target color data selected in a list screen.
• Edit Name: Name the target color data.
• Print: Print the target color data.
• Tolerance: Specify the tolerance (box tolerance) used for pass/fail judgment of measured data.
• Delete: Delete the target color data.
“All” tab
• Save: Save all target color data to the USB memory device. This option can be selected only
when a USB memory device has been connected.
• Tolerance: (Def.) Before setting tolerance for individual target color data, set default tolerance to be
applied to all target color data in common.
• Data Protect: Disable the deletion or editing of the target color data. When the data protection is set,
you cannot select “Edit Name”, “Tolerance”, “Delete” and “Delete All” on the <Tar get Data Function> screen.
• Delete All: Delete all target color data when the data protection is not set.
[Setting Procedure]
1. On the <Target> screen, press the [OK/Edit] button.
The <Target Data Function> screen is displayed.
2. Use the or button of the cross key to select the
tab, use the
or button to move the cursor to the
desired item to set or start, and then press the [OK/
Edit] button.
The setting screen for the selected item is displayed.
For the setting procedure of each item, refer to the following
sections.
3. On the <Target Data Function> screen, press the [Back]
button to return to the <Target> screen.
E78
Page 81
Saving the Current Target Color Data to a USB Memory Device
Save the current target color data to a USB memory device.
You will need to connect a USB memory device to the instr ument in advance. For information on connecting a
USB memory device, refer to page 141 “Connecting a USB Memory”.
[Operating Procedure]
Start the procedure from the <Target Data Function> screen
1. Use the or button of the cross
key to select the “Current” tab, use
or button to move the
the
cursor to “Save”, and then press the
[OK/Edit] button.
2. The <Save Target Data> screen is
displayed with a suggested le name
shown.
To change the le name, use the
button of the cross key to
and
move the cursor to the current le
name and then press the [OK/Edit]
button.
3. The <Edit File Name > screen is
displayed.
Use the or and or buttons
of the cross key to move the cursor
around the characters and then press
the [OK/Edit] button.
The selected character is added to the text
box.
To change a le name using the USB
keyboard, remove the USB memory
device while on the name editing screen,
and input the name after connecting the
USB keyboard.
4. Repeat Step 3 until you have entered
the necessary characters.
• The position where characters will be
added can be changed by using the
or
button of the cross key to
position the cursor in the text box. The
le name will then be modied at the
position of the cursor.
• To delete the character to the left of the
cursor in the text box, press the [CAL] button.
• The le name can be up to 20
characters long.
2
Preparation for Measurement
Memo
A USB keyboard can also be
used to input le names. For
information on connecting a
USB keyboard, refer to page
143 “Connecting a USB
Keyboard”.
E79
Page 82
5. When you are nished entering
characters, press [MENU] button,
move the cursor to “Save” and then
press the [OK/Edit] button.
The settings are conrmed and the screen
returns to the <Save Target Data>
screen.
Memo
If you press the [Back] button
during the setting, “Do you want to
save the changes?” is displayed. If
you then place the cursor on “OK”
and press the [OK/Edit] button, the
setting details are con rmed and the
screen retur ns to the <Save Target
Data> screen; if you place the
cursor on “Cancel” and press the
[OK/Edit] button, the setting
details are not changed and the
screen retur ns to the <Save Target
Data> screen.
6. Verify that a USB memory device is
connected. To accept the displayed le
name, use the
or button of the
cross key to move the cursor to and
then press the [OK/Edit] button.
The <Saving…> screen is displayed.
When the saving is complete, the screen
returns to the <Target> detail screen.
Memo
The USB memor y device must
be connected to the instrument
before saving the condition le
to a USB memory device. For
information on connecting a
USB memory device, refer to
page 141 “Connecting a USB
Me mo r y”.
Performing Statistical Calculations on Target Data
Performs statistical data on target data stored in the instrument. The data to be used for calculations can be
selected in a list screen. After calculations have been performed, the average of the targets selected for statistical
calculations can be saved.
To save the average as a new target data, be sure to select a target color No. for which no data has been set before
proceeding. If this procedure is performed when a target color No. which already contains target color data, the
existing target color data will be overwritten when save is performed.
1. Use the or button of the cross
key to select the “Current” tab, use
or button to move the
the
cursor to “Statistic calculation”, and
then press the [OK/Edit] button.
E80
Page 83
2. The <Target List> screen will appear.
3. Use the or button of the cross
key to select a target to use for
statistical calculations and then press
the [OK/Edit] button to conrm the
selection.
• A check mark will appear in the box to
the left of the target name when a target
is selected.
• Statistics can be performed using only
targets measured under the same
measurement conditions (measurement
type, measurement area, SCI/SCE
setting).
• To deselect a target with a check mark,
move the cursor to the target and then
press the [OK/Edit] button again.
• To select all targets shown on the
screen, use the
cross key to move the cursor to the
“Page” button and then press the [OK/Edit] button. Check marks will appear
to the left of all targets shown.
• Use the
key to move to other pages in
instrument memory. Selected data will
remain selected even if pages are
changed.
or button of the
or button of the cross
2
Preparation for Measurement
E81
Page 84
4. After all desired targets have been
selected, use the
or button of
the cross key to move the cursor to
the “OK” button and then press the
[OK/Edit] button.
The <Statistic calculation> screen
will appear.
• To save the average of the data, use the
or button of the cross key to
move the cursor to the [Save] button
and press the [OK/Edit] button. If data
already exist for the currently selected
target color No., an overwrite
conrmation message will appear. The
data will be saved under the name
“Averaged” and the <Statistic calculation> screen will close.
• To exit the <Statistic calculation> screen without saving, use the
button of the cross key to move the
cursor to the [Cancel] button and press
the [OK/Edit] button. The screen will
return to the <Target List> screen.
or
E82
Page 85
Edit Name
Name the target color data.
[Procedure]
Start the procedure from the <Target Data Function> screen.
1. Use the or button of the cross key to select the
“Current” tab, use the
or button to move the
cursor to “Edit Name”, and then press the [OK/Edit]
button.
The <Edit Name> screen is displayed.
Memo
A USB keyboard can also be used to input le names. For
information on connecting a USB keyboard, refer to page
143 “Connecting a USB Keyboard”.
2. Use and or and buttons of the cross key to
move the cursor around the characters and then press
the [OK/Edit] button.
• Up to 20 characters can be used.
• The selected character is displayed in the text box.
3. Repeat Step 2 until you have entered the necessary
characters.
• You can use the or button of the cross key to move
the blue box cursor on the text box.
In this status, you can move the character position and
modify the character by pressing the
cross key.
• To delete the character to the left of the cursor in the text
box, press the [CAL] button.
or button of the
2
Preparation for Measurement
4. When you are nished entering characters, use and
or and buttons of the cross key to move the
cursor to “Save” and then press the [OK/Edit] button.
The setting is conrmed and the screen returns to the
<Target> screen.
E83
Page 86
Memo
If you press the [Back] button dur ing the setting, “Do you
want to save the changes?” is displayed. When you place
the cursor on “OK” and press the [OK/Edit] button, the
setting details are con rmed and the screen ret urns to the
<Target> screen. When you place the cursor on “Cancel”
and press the [OK/Edit] button, the setting details are not
changed and the screen returns to the <Target Data
Functions> screen.
Print
Print the target color data.
You need to connect the instrument to a serial printer in advance. For the connection of the instrument and a
serial printer, refer to “Printer Connection” (Page 145).
Notes
• If the connection is not established properly, printing is disabled.
• Even if the connection has been correctly established, printing may fail for reasons such as the printer is turned
off when printing is attempted.
[Procedure]
Start the procedure from the <Target Data Function> screen.
1. Use the or button of the cross key to select the
“Current” tab, and use the
or button to move the
cursor to “Print”.
2. Press the [OK/Edit] button. The <Print Target Data>
screen is displayed and the data is printed from the
connected printer.
• When the printing is complete, the screen returns to the
<Target> screen.
Notes
The <Print Target Data> screen is also displayed for the
case where printing will not be processed properly such as
when the pr inter connection is incor rect.
E84
Page 87
Tolerance Setting
Specify the tolerance used for pass/fail judgment of measured data for each target color.
Memo
• Before tolerance is specied for individual target colors, the instrument is set with default tolerance. For details,
refer to page 89 “Default Tolerance Setting”.
• The tolerance setting items are the same as those for the color space or index cur rently selected for the display
conditions.
• With the optional Color Management Software “SpectraMagic™ NX”, you can set or use the color difference
tolerance easily.
[Setting Procedure]
Start the procedure from the <Target Data Functions> screen.
1. Use the or button of the cross key to select the
“Current” tab, use the
or button to move the
cursor to “Tolerance”, and then press the [OK/Edit]
button.
The <Tolerance> screen is displayed.
2. Use and or the and button of the cross key
to move the cursor to the desired item.
3. Press the [OK/Edit] button to change the details.
• When the selected item is not currently checked, pressing
the [OK/Edit] button causes the item to be checked, and
you can change the setting value.
• When the selected item is not currently checked, pressing
the [OK/Edit] button causes the item to be checked. Press
the
button of the cross key to move the blue box cursor
to the setting value area. When you press the [OK /Edit]
button, the cursor appears on the value. Press the
button of the cross key to change the value.
Memo
Settings
0.1 to 20.0
• You need to press the [OK/Edit] button to conrm the
You can move the cursor between the digits of the value by
pressing the or button of the cross key.
* Settings of x and y are 0.001 to 0.200.
change every time you change the setting of an item.
or
2
Preparation for Measurement
E85
Page 88
4. When you nish all settings, press the button or
[MENU] button of the cross key to move the cursor on
“Save”, and then press the [OK/Edit] button.
The setting is conrmed and the screen returns to the <Tar get
Data Function> screen.
Memo
If you press the [Back] button dur ing the setting, the
<Save Conrmation> screen is displayed. When you
place the cursor on “OK” and press the [OK/Edit] button,
the setting details are conr med. When you place the
cursor on “Cancel” and press the [OK/Edit] button, the
setting details are not changed. After that, the screen
returns to the <Target Data Function> screen.
Delete
Delete the target color data.
Notes
When the data is protected, you cannot select “Delete” on the <Target Data Function> screen.
[Procedure]
Start the procedure from the <Target Data Function> screen.
1. Use the or button of the cross key to select the
“Current” tab, use the
or button to move the
cursor to “Delete”, and then press the [OK/Edit] button.
The <Delete Target Data> screen is displayed.
2. Use the button of the cross key to move the cursor to
“OK” and press the [OK/Edit] button. The data is
deleted.
• When the deletion is complete, the screen returns to the
<Target> screen.
• When you place the cursor on “Cancel” and press the [OK/
Edit] button, the deletion is canceled and the screen returns
to the <Target Data Function> screen.
E86
Page 89
Save to USB memory device
Save all target color data stored in the instrument to the USB memory device.
The USB memory device must be connected to the instrument before saving the data. For the connection of a
USB memory device, refer to page 141 “Connecting a USB Memory”.
Memo
A USB keyboard can also be used to input le names. For information on connecting a USB keyboard, refer to page
143 “Connecting a USB Keyboard”.
[Setting Procedure]
Start the procedure from the <Target Data Functions> screen.
1. Use the or button of the cross key to select the
“All” tab, use the
or button to move the cursor to
“Save”, and then press the [OK/Edit] button.
2. The < Target List > screen appears, with all targets initially
selected (check marks shown in the box to the left of the target
name.
• To deselect a target with a check mark, move the cursor to
the target and then press the [OK/Edit] button.
• To deselect all targets shown on the screen, use the
button of the cross key to move the cursor to the “Page”
button and then press the [OK/Edit] but ton. The check
marks to the left of all targets shown will disappear.
• Use the
pages in instrument memory. Selected data will remain
selected even if pages are changed.
or button of the cross key to move to other
or
3. When selection/deselection of targets has been
completed, use the
or button of the cross key to
move the cursor to the “Save” button and then press the
[OK/Edit] button.
2
Preparation for Measurement
The <Save Target Data> screen is displayed with a suggested
le name shown.
4. To change the le name, use the or button of the
cross key to move the cursor to “File Name” and then
press the [OK/Edit] button.
The <Edit Name> screen is displayed.
Memo
Connect a USB keyboard to input le names using the USB
keyboard.
E87
Page 90
5. Use and or and buttons of the cross key to
move the cursor around the characters and then press
the [OK/Edit] button.
The selected character is displayed in the text box.
6. Repeat Step 5 until you enter necessary characters.
• You can use the or button of the cross key to move
the blue box cursor on the text box. In this status, you can
move the character position and modify the character by
pressing the <
or>button of the cross key.
• To delete the character to the left of the cursor in the text
box, press the [CAL] button.
• Up to 20 characters can be used.
7. When you are nished entering characters, use and
or and buttons of the cross key to move the
cursor to “Save” and then press the [OK/Edit] button.
The setting is conrmed and the screen returns to the <Save
Target Data> screen.
Memo
• If you press the [Back] button during the setting, “Do
you want to save the changes?” is displayed. When you
place the cursor on “OK” and press the [OK/Edit]
button, the setting details are con rmed and the screen
returns to the <Save Target Data> screen. When you
place the cursor on “Cancel” and press the [OK/Edit] button, the setting details are not changed and the
screen retur ns to the <Save Target Data> screen.
• Disconnect any USB keyboards connected to the
instrument.
8. Connect the USB memory device. To accept the
displayed le name, use the
key to move the cursor to
or button of the cross
“Save Target Data” and
then press the [OK/Edit] button.
• The <Saving…> screenis displayed.
• When the saving is complete, the screen returns to the
<Target> screen.
E88
Page 91
Default Tolerance Setting
This instrument allows tolerance setting for individual target color data. Before setting such individual
tolerance, the instrument is set with default tolerance. This section describes the procedure to set default
tolerance.
Memo
[Setting Procedure]
1. Use the or button of the cross key to select the
2. Use the and button of the cross key or to
• Before tolerance is specied for individual target colors, the instrument is set with default tolerance.
• W hen you change default tolerance after tolerance has been set for individual target colors, the tolerance
specied for these target colors will not be changed.
• With the optional Color Management Software “SpectraMagic™ NX”, you can set or use the color difference
tolerance easily.
• The default tolerance setting items are the same as those of the color space or index currently selected for the
display conditions.
Start the procedure from the <Target Data Function> screen.
“All” tab, use the
or button to move the cursor to
“Tolerance (Def.)”, and then press the [OK/Edit]
button.
The <Tolerance (Def.)> screen is displayed.
move the cursor to the desired item.
2
Preparation for Measurement
3. Press the [OK/Edit] button to change the details.
• When the selected item is not currently checked, pressing
the [OK/Edit] button causes the item to be checked, and
you can change the setting value.
• When the selected item is not currently checked, pressing
the [OK/Edit] button makes the item be checked. Press the
button of the cross key to move the blue box cursor to
the setting value area. When you press the [OK/Edit]
button, the cursor appears on the value. Press the
button of the cross key to change the value.
Memo
Settings
0.1 to 20.0
• You need to press the [OK/Edit] button to conrm the
You can move the cursor between the digits of the value by
pressing the or button of the cross key.
* Settings of x and y are 0.001 to 0.200.
change every time you change the setting of an item.
or
E89
Page 92
4. When you nish all settings, press the button or
[MENU] button of the cross key to move the cursor on
“Save”, and then press the [OK/Edit] button.
The setting is conrmed and the screen returns to the <Tar get
Data Function> screen.
Memo
If you press the [Back] button dur ing the setting, the
<Save Conrmation> screen is displayed. When you
place the cursor on “OK” and press the [OK/Edit] button,
the setting details are conr med. When you place the
cursor on “Cancel” and press the [OK/Edit] button, the
setting details are not changed. After that, the screen
returns to the <Target Data Function> screen.
E90
Page 93
Data Protection
You can specify data protection so that the saved target color setting will not be deleted or changed by accident.
When the data protection is set, you cannot select “Edit Name”, “Tolerance”, “Delete”, “Statistic calculation”
and “Delete All” on the <Target Data Function> screen.
[Setting Procedure]
1. Use the or button of the cross key to select the
“All” tab, use the
“Data Protect”, and then press the [OK/Edit] button.
The <Data Protect> screen is displayed.
2. Use the or button of the cross key to move the
cursor to the desired item.
Settings
ON: Protect the data.
OFF: Do not protect the data.
3. Press the [OK/Edit] button.
The selection is conrmed and the screen returns to the
<Target> screen.
Notes
If you press the [Back] button without pressing the [OK/
Edit] button, you return to the <Target Data Function>
screen without changing the setting.
Start the procedure from the <Target Data Function> screen.
or button to move the cursor to
2
Preparation for Measurement
E91
Page 94
Delete All
Delete all the target color data which have been set.
Notes
When the data is protected, you cannot select “Delete All” on the <Target Data Function> screen.
[Setting Procedure]
Start the procedure from the <Target Data Function> screen.
1. Use the or button of the cross key to select the
“All” tab, use the
or button to move the cursor to
“Delete All”, and then press the [OK/Edit] button.
The <Delete All Target> screen is displayed.
2. Use the button of the cross key to move the cursor to
“OK” and press the [OK/Edit] button. The data is
deleted.
• When the all data is deleted, the screen returns to the
<Target> screen.
• When you place the cursor on “Cancel” and press the [OK/
Edit] button, all data deletion is canceled and the screen
returns to the <Target Data Function> screen.
E92
Page 95
Other Settings
Setting Automatic White Calibration (100% Calibration)
to ON/OFF
Set whether to perform white calibration (100% Calibration) automatically when the power switch of the
instrument is set to ON.
[Setting Procedure]
1. Press the [MENU] button.
The <Conguration> screen is displayed.
2. Use the or button of the cross key to move the
cursor to “Calibration Options” and then press the
[OK/Edit] button.
The <Calibration Option Settings> screen is displayed.
The <Calibration Option Settings> screen shows the
current setting of each item.
2
Preparation for Measurement
3. Use the or button of the cross key to move the
cursor to “Automatic White Cal.” and then press the
[OK/Edit] button.
The <Auto White Calibration Settings> screen is displayed.
E93
Page 96
4. Use the or button of the cross key to move the
cursor to the desired item.
Settings
OFF : Disable automatic white calibration (100%
Calibration) the next time the instr ument is turned
ON (the power switch is set to ON).
ON : Enable automatic white calibration (100%
Calibration) the next time the instr ument is turned
ON (the power switch is set to ON).
Memo
When you set automatic white calibration to ON, white
calibration (100% calibration) is automatically performed
after the power switch of the inst rument is set to ON before
the <Welcome to Easy Setup Wizard!> screen appears.
If you have disabled the display of the <Welcome to Easy Setup Wi zard!> screen, the <Sample> screen is
displayed immediately after the instrument is turned ON
and white calibration (100% calibration) is performed. This
allows you to start measurement quickly. This function is
useful when you always perform measurement using the
same condition settings.
For the procedure of disabling the display of the <Welcome
to Easy Setup Wizard!> screen, refer to page 102.
5. Press the [OK/Edit] button.
The selection is conrmed and the screen returns to the
<Calibration Option Setting> screen.
Notes
If you press the [Back] button without pressing the [OK/
Edit] button, you return to the <Calibration Option
Settings> screen without changing the setting.
6. After you set the calibration options, press the [Back]
button to return to the <Conguration> screen.
E94
Page 97
Setting the Measurement Result Display
To set details about the measurement result display, select “Graph” from the <Conguration> screen.
[Setting Procedure]
1. Press the [MENU] button.
The <Conguration> screen is displayed.
2. Use the or button of the cross key to move the
cursor to “Graph” and then press the [OK/Edit] button.
The <Graph and Data Settings> screen is displayed.
2
Preparation for Measurement
3. After you set the graph display options, press the [Back]
button to return to the <Graph and Data Settings>
screen.
E95
Page 98
Setting the Custom Data Screen to ON/OFF
The measurement result display function allows you to switch between various screens including “A bsolut e ” or
“Difference”. You can add the “Custom Data” screen showing your desired items to these screens.This section describes the procedure to set whether to display this “Custom Data” screen.
Memo
To set the items displayed on the “Custom” screen, you need to use the optional Color Management Software
SpectraMagic™ NX. For details, refer to the instruction manual of SpectraMagic™ NX.
[Setting Procedure]
Start the procedure from the <Graph and Data Settings> screen.
1. Use the or button of the cross key to move the
cursor to “Custom Data View Settings” and then press
the [OK/Edit] button.
The <Custom Data View Settings> screen is displayed.
2. Press the [OK/Edit] button.
The <Show Custom Data Screen> screen is displayed.
3. Use the or button of the cross key to move the
cursor to the desired item.
Settings
OFF : The “Custom Data” screen is not displayed as the
measurement result screen.
ON : The “Custom Data” screen is displayed as one of
the measurement result screens.
4. Press the [OK/Edit] button.
The selection is conrmed and the screen returns to the
<Custom Data View Settings> screen.
Notes
If you press the [Back] button without pressing the [OK/
Edit] button, you return to the <Custom Data View
Settings> screen without changing the setting.
E96
Page 99
5. Press the [Back] button.
The screen returns to the <Graph and Data Settings>
screen.
Setting the Assess. Graph Screen to ON/OFF
The measurement result display function allows you to switch between various screens including “Absolute” or
“Difference”. You can add the “Assess. Graph” screen based on tone data to these screens.
This section describes the procedure to set whether to display this “Assess. Graph” screen.
2
Preparation for Measurement
[Setting Procedure]
Start the procedure from the <Graph and Data Settings> screen.
1. Use the or button of the cross key to move the
cursor to “Color Assessment Settings” and then press
the [OK/Edit] button.
The <Color Assessment Settings> screen is displayed.
2. Press the [OK/Edit] button.
The <Show Color Assess.> screen is displayed.
3. Use the or button of the cross key to move the
cursor to the desired item.
Settings
OFF : The <Assess. Graph> screen is not displayed as
the measurement result screen.
ON : The <Assess. Graph> screen is displayed as one of
the measurement result screens.
Memo
When the color assessment is set to ON, the <Assess.
Graph> screen is displayed instead of the <Diff.Graph>
screen. Moreover, color assessment is added to the
<Difference> screen.
E97
Page 100
4. Press the [OK/Edit] button.
The selection is conrmed and the screen returns to the
<Color Assessment Settings> screen.
Notes
If you press the [Back] button without pressing the [OK/
Edit] button, you return to the <Color Assessment
Settings> screen without changing the setting.
5. Press the [Back] button.
The screen returns to the <Graph and Data Settings>
screen.
Setting the Spectral Graph Data Display
The measurement result display function allows you to switch between various screens including “Absolute” and
“Difference”, and the “Spectral” screen which shows a spectral graph is one of them.
This section describes the procedure to set whether to show spectral data on this “Spectral” screen.
[Setting Procedure]
Start the procedure from the <Graph and Data Settings> screen.
1. Use the or button of the cross key to move the
cursor to “Spectral Graph Settings” and then press the
[OK/Edit] button.
The <Spectral Graph Settings> screen is displayed.
2. Use the or button of the cross key to move the
cursor to “Spectral Data Display” and then press the
[OK/Edit] button.
The <Spectral Data Display> screen is displayed.
E98
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.