Formal designations of application software used in this manual
(Designation in this manual) (Formal designation)
Bluetooth Bluetooth®
Windows, Windows 7 Microsoft® Windows® 7 Professional Operating System
Windows, Windows 10 Microsoft® Windows® 10 Pro Operating System
macOS, macOS Sierra Apple® macOS® Sierra
Excel Microsoft® Oce Excel
Trademarks
• “Microsoft”, “Windows”, “Windows 7”, and “Windows 10” are registered trademarks of Microsoft Corporation in the
U.S.A. and other countries.
• “Apple” and “macOS” are registered trademarks of Apple Inc. in the U.S.A. and other countries.
• “Intel” and “Intel Core” are registered trademarks or trademarks of Intel Corporation in the U.S.A. and other countries.
Other company names and product names mentioned in this manual are the registered trademarks or trademarks of their
respective companies.
Page 3
Safety Precautions
To ensure correct use of this instrument, read the following points carefully and adhere to them. After you have read this
manual, keep it in a safe place where it can be referred to anytime a question arises.
WARNING
Do not use the instrument in places where flammable or combustible gases (gasoline, etc.) are present.
Doing so may cause a fire.
Firmly push the cable plug completely into the outlet. Incomplete insertion may cause fire or electric shock.
Always hold the plug itself when disconnecting the cable from an outlet. Pulling on the cable may damage it
and cause a fire or electric shock. Also, do not insert or disconnect the plug with wet hands. Doing so may cause
electric shock.
Do not forcibly bend, twist, or pull the cords or cables. Also, do not scratch, or place heavy objects on the
cables. Doing so may damage the cable and cause a fire or electric shock.
Do not disassemble or modify the instrument. Doing so may cause a fire or electric shock.
Take special care not to allow liquid or metal objects to enter the instrument. Doing so may cause a fire or
electric shock. Should liquid or metal objects enter the instrument, turn the power OFF immediately, and
contact the nearest KONICA MINOLTA-authorized service facility.
(Failure to adhere to the following points may result in death or serious injury.)
The instrument should not be operated if it is damaged, or if smoke or odd smells occur. Doing so may cause a
fire. In such situations, turn the power OFF immediately, and contact the nearest KONICA MINOLTA-authorized
service facility.
Do not insert or disconnect the plug with wet hands. Doing so may cause electric shock.
CAUTION
When cleaning the instrument, disconnect the cable. Failure to do so may result in electric shock.
Do not place the instrument on an unstable or sloping surface. Doing so may result in the instrument dropping
or overturning, causing injury. Be careful not to drop the instrument when carrying it as well.
(Failure to adhere to the following points may result in injury or damage to the
instrument or other property.)
1
Page 4
Introduction
The CA-410 series is an instrument designed to measure the color, luminance, and flicker of various color displays. Please
read this manual carefully before using the instrument.
Packing materials of the product
Be sure to keep all packing materials used for shipping the instrument (cardboard box, cushioning material, plastic bags,
etc.). This instrument is a precision measuring instrument. When transporting the instrument to a service facility for
maintenance or for other reasons, be sure to use the packing materials to minimize shock or vibration. If the packing
materials are lost or damaged, contact a KONICA MINOLTA-authorized service facility.
Notes on Use
Operating Environment
• This instrument has been designed exclusively for indoor use. Outdoor use is prohibited.
• This instrument is composed of precision electronic components. Never disassemble the instrument.
• This instrument is a pollution level 2 product (equipment to be used primarily in manufacturing environments,
laboratories, warehouses, and similar locations). This instrument should be used in environments where exposure to
metallic dust or condensation is not a concern.
• This instrument is an overvoltage category I product (equipment for connection to circuits in which measures are taken to
limit transient overvoltage to an appropriately low level).
• Do not use the instrument at altitudes higher than 2,000 m.
• Take care to prevent foreign matter from entering the instrument. Using the instrument while subjected to intrusion of
water or metals is extremely dangerous.
• Using the instrument in direct sunlight or near heating equipment can cause the internal temperature of the instrument
to become much higher than the ambient temperature, resulting in malfunction. Do not use the instrument in such areas.
Also, use the instrument in a well-ventilated environment.
• Avoid subjecting the instrument to sudden temperature changes and condensation.
• Do not use the instrument in areas where dust, smoke, or chemical gases are present, or in extremely humid
environments.
• This instrument should be used in an environment with an ambient temperature between 10 and 35°C and a relative
humidity of 85% or less. Use of the instrument outside this range will result in unsatisfactory performance.
System
• Do not subject the instrument to strong vibrations or impacts.
• Do not pull, forcibly bend, or apply excessive force to the connected cables and cords. Doing so may cause the cable or
cord to break.
• The instrument should be connected to a power source with as little noise as possible.
• The instrument may cause interference if used near a television, radio, transceivers, etc.
• When restarting the instrument, wait several seconds after turning the power OFF before turning the power ON again.
• Do not turn the power o during writing to the memory such as calibration and measurement.
• The instrument may not operate properly due to the compatibility with the equipment connected to it such as a PC.
Please check whether the instrument operates properly by yourself.
• Make sure to disconnect the USB connection between the probe and the PC and to turn o the power when the
instrument is not in use.
• Periodic backup of important data and settings to another storage medium is recommended.
• Should the instrument break down, do not try to disassemble and repair the instrument. Consult with your nearest
KONICA MINOLTA-authorized service facility.
Lens
• Before measurement, make sure that the lens surface is not dirty. You may not be able to perform correct measurement if
there is dirt, dust, or finger mark on it, or if stain remains after cleaning.
• Do not touch the lens surface with hands.
• Avoid subjecting the instrument to sudden temperature changes under a high-humidity environment. It may fog the lens
and hinder correct measurement.
2
Page 5
Storing the Instrument
• This instrument should be stored at a temperature between 0 and 45°C with a relative humidity of 85% or less and no
condensation. Storing the instrument in an environment with high temperatures and high humidity will result in
unsatisfactory performance. Storing the instrument at or near room temperature and humidity is recommended.
• Make sure that the instrument is not subjected to condensation when stored. In addition, take care to prevent rapid
temperature changes to prevent condensation from occurring when transporting the instrument to the storage location.
• Storing the instrument in direct sunlight or near heating equipment can cause the internal temperature of the instrument
to become much higher than the ambient temperature, resulting in malfunction. Do not store the instrument in such
areas.
• Do not store the instrument in areas where dust, smoke, or chemical gases are present. Doing so may cause deterioration
in performance or a malfunction.
• Do not leave the instrument inside the cab or trunk of a vehicle. Otherwise, the temperature and/or humidity during
midsummer or midwinter may exceed the allowable range for storage, resulting in a malfunction.
• Before storing, make sure that the lens is attached with the cap supplied as a standard accessory.
Notes on Cleaning
• If the instrument becomes dirty, wipe it with a soft, dry cloth. Never use organic solvents (such as benzene or thinner) or
other chemicals for cleaning.
• If there is dust or dirt on the optical system of a probe, use a blower to blow it o and then gently wipe it with a soft, dry
cloth or with a lens-cleaning cloth. Never use organic solvents (such as benzene or thinner) or other chemicals for
cleaning.
• If you are unable to remove dirt from the instrument or if the instrument becomes scratched, contact a KONICA
MINOLTAauthorized service facility.
Notes on Transporting
• When transporting the instrument, be sure to use the packing materials to minimize shock or vibration.
• When sending the instrument in for service, package and send the instrument and all accessories.
Maintenance and Inspection
• To maintain measurement accuracy, the instrument should be inspected once a year. For information on inspection,
contact the nearest KONICA MINOLTA-authorized service facility.
Disposal Method
• Make sure that the instrument, its accessories, and the packing materials are either disposed of or recycled correctly in
accordance with local laws and regulations.
About This Manual
This manual is designed for those who possess basic knowledge of displays and relevant equipment.
Please read this manual carefully before using the instrument.
In the descriptions about probes in this manual, probes are described as "probe" where the type of probe is not specified.
For Those Who Want to Purchase Optional Accessories for This Instrument
This manual also explains how to use optional accessories available for this instrument.
If an explanation of how to use an optional accessory is given in this manual, its product name is also given.
Please read the explanation together with the manual supplied with the accessory.
Information ................................................... 76
5
Page 8
Configuration of the CA-410 Series
1. Configuration
The CA-410 series consists of the following three systems:
zProbes
Sensor units for measuring displays.
Can be connected to a PC to perform measurement (using PC software).
Can be connected to a data processor to perform measurement.
zData Processor CA-DP40
Can connect up to ten probes simultaneously to perform measurement.
Can be connected to a PC so that it can be controlled from the PC (using PC software).
zPC software CA-S40
Can control a probe connected to a PC to perform measurement.
Can control a data processor connected to a PC to perform measurement.
The instruction manual for probes and PC software (this manual) contains the following descriptions about probes and the
standard PC software, PC Software for Color Analyzer CA-S40:
• Handling procedure and specifications of probes
• How to install the PC software
• How to measure when the probe(s) or the data processor are connected to a PC
6
Page 9
2. System Configuration Diagram
[Probe]
Probe standard accessoryData Processor standard accessoryOptional accessory
Standard Probe
CA-P427
CA-P410
Miniature Probe
CA-MP410
* A lens hood and lens cap are
included as standard with all probes.
High Sensitivity Probe
CA-VP427
CA-VP410
(miniDIN)
BNC Conversion Cable IF-A35
(miniDIN)
Conversion Cable IF-A29
(Commercially available)
USB Cable for
Probe-PC (2 m)
IF-A28
(BNC)
(D-sub)
(Power Cable)
(BNC)
RS-232C cable
RS Cable for
Probe-DP (2 m)
IF-A30
[Data Processor]
CA-DP40
Lithium-Ion Battery
CM-A223
RS Cable for
Probe-DP
IF-A31 (5 m)
IF-A32 (10 m)
AC Adapter AC-A312F
PC software for
Color Analyzer CA-S40
SDK for Color Analyzer
CA-SDK2
User Memory .......................................................................... 26
1. About Calibration Channels ....................................................... 26
2. About Targets ........................................................................... 26
3. About User Calibration .............................................................. 27
Specifications of Each Probe .................................................... 28
Probe GuideProbe Guide
9
Page 12
Probes
1. About Probes
A CA-410 series probe can be used to measure the luminance, chromaticity, and flicker values of displays.
A probe can be controlled in the following three ways:
• You can connect a probe to a PC and control it from PC software to perform measurement. By using the standard
software, PC Software CA-S40 For Color Analyzers, you can select one of the probes connected to a PC via USB and
control it to perform measurement.
Display
PC
Probes (only one probe can be controlled)
USB cable connection
CA-S40
• You can connect a probe to a data processor and control it from the data processor to perform measurement. Multiple
probes can be controlled simultaneously to perform measurement.
Display
Probes (up to 10 probes can be controlled simultaneously)
Mini DIN cable connection
or USB cable connection
Data Processor
• You can connect a probe to a data processor, connect the data processor to a PC, and control it from PC software to
perform measurement. By using the standard software, PC Software for Color Analyzer CA-S40, you can select one of the
probes connected to the data processor and control it to perform measurement.
Display
Probes (only one probe can be controlled)
Mini DIN cable
connection or USB
cable connection
Data Processor
RS-232C
PLC or PC
10
USB
connection
BNC (sync/trigger)
PC
CA-S40
Page 13
2. Probe Lineup (All Probes are Provided with Lens Caps)
Eight types of probes are available according to your application:
ModelNameDescriptionDimensions
CA-P42727 ProbeProbe with a measurement area of 27 mmP. 14
CA-P427H27 Probe for high luminanceHigh-luminance probe with a measurement area of 27 mmP. 14
CA-P41010 ProbeProbe with a measurement area of 10 mmP. 15
CA-P410H10 Probe for high luminanceHigh-luminance probe with a measurement area of 10 mmP. 15
CA-MP41010 mini ProbeCompact probe with a measurement area of 10 mmP. 18
CA-MP410H 10 mini Probe for high luminance
CA-VP42727 High Sensitivity ProbeHigh-sensitivity probe with a measurement area of 27 mmP. 16
CA-VP41010 High Sensitivity ProbeHigh-sensitivity probe with a measurement area of 10 mmP. 17
Compact high-luminance probe with a measurement area of 10 mm
P. 18
3. Standard Accessories
zLight-shielding cylinders and lens caps
UseLight-shielding cylinderLens cap
ModelNameModelName
For a 27 probeCA-H12 Hood for 27 ProbeCA-H13 Lens Cap for 27 Probe
For a 10 probeCA-H14 Hood for 10 ProbeCA-H15 Lens Cap for 10 Probe
For a 10 mini probeCA-H16 Hood for 10 Mini ProbeCA-H17 Lens Cap for 10 Mini Probe
For a 27 high-sensitivity probe CA-H18 Hood for 27 High Sensitivity ProbeCA-H19 Lens Cap for 27 High Sensitivity Probe
For a 10 high-sensitivity probe CA-H20 Hood for 10 High Sensitivity ProbeCA-H21 Lens Cap for 10 High Sensitivity Probe
Probe Guide
zUSB Cable for Probe-PC IF-A28
Connects a probe to a data processor or a PC via USB.
How to connect: Refer to page 19.
zPC Software for Color Analyzer CA-S40
Controls a probe or data processor connected to a PC via USB to perform
measurement.
This is stored on the USB flash drive supplied with the product.
For how to use it, refer to the instruction manual for the PC software on
the USB flash drive.
zSDK for Color Analyzer CA-SDK2
A library for users to create programs on their own. This is stored on the
USB flash drive supplied with the product.
zInstruction Manual (this manual)
Please read before using the instrument.
* PC Software CA-S40, SDK CA-SDK2, and the latest version of the probe software are available for download from the
following product support webpage:
https://www.konicaminolta.com/instruments/download/
11
Page 14
Probes
4. Optional Accessories
zConversion Cable IF-A29
Connects a probe to a PC or PLC (sequencer) via RS-232C and inputs
synchronization signals to the probe.
How to connect: Refer to page 20.
zBNC Conversion Cable IF-A35
Inputs synchronization signals to the probe.
How to connect: Refer to page 20.
5. Description of Each Part of Probes
(5)
(2)
(3)
(4)(4)(4)(4)
ISO screwTripod mount screwISO screwTripod mount screw
(1)
zProbe other than minizMini probe
Note: CA-P427, P427H, P410, and P410H have another tripod mount screw
on the front side of the ISO screw. For details, refer to the dimensions.
Note: Mini probes have screws for jigs and
cable clamps. For details, refer to the
dimensions.
(1) Receptor
Make this part face perpendicular to a display to perform measurement.
(2)
(3)
(2) USB connector
For USB connection with a data processor or a PC, connect the USB cable IF-A28 supplied as a standard accessory.
(3) RS-232C connector
For RS-232C connection to a data processor, connect the mini DIN cable IF-A30/31/32 for the data processor to the
RS-232C connector.
For RS-232C connection with a PLC or PC, connect an optional conversion cable IF-A29.
Use a conversion cable IF-A29 or a BNC conversion cable IF-A35 to input the display's vertical synchronization signals, which
are used for measurement in “EXT” measurement synchronization mode.
12
Page 15
(4) Screw holes
There are two types of screws for securing a probe:
Tripod mount screw: Used for mounting the probe on a tripod. The depth is 6 mm.
ISO screw: Used for securing the probe to jigs. Use an ISO 5 mm screw and the depth is 6 mm.
• The distance from each screw to the end of the optical column is the same as that for a CA-310 series probe. (Excluding
mini probes)
• The position of the optical axis (in the height direction) is the same as that for a CA-310 series probe.
(5) Status indicator
Located on the rear of all probes to indicate the status of the probe.
No power supply: O
During measurement or zero calibration: O
Memory error: Blinking (On for 0.2 sec, then O for 0.2 sec)
Zero calibration not performed: Blinking (On for 1 sec, then O for 0.2 sec)
Measurement possible: On (even when measurement is not possible due to the setting of an invalid synchronization
frequency or integration time in INT, EXT, or MANUAL sync mode, without displaying that state)
Probe Guide
13
Page 16
Probes
6. Probe Dimensions
CA-P427Ø27 ProbeProbe with a measurement area of 27 mm
CA-P427H Ø27 Probe for High Luminance High-luminance probe with a measurement area of 27 mm
(unit: mm)
When a lightshielding cylinder
is mounted
ISO 5 mm screw (depth: 6 mm)
ISO 5 mm screw (depth: 6 mm)
Tripod mount screw
(depth: 6 mm)
Tripod mount screw
(depth: 6 mm)
14
Page 17
CA-P410Ø10 ProbeProbe with a measurement area of 10 mm
CA-P410H Ø10 Probe for High Luminance High-luminance probe with a measurement area of 10 mm
(unit: mm)
Probe Guide
When a lightshielding cylinder
is mounted
ISO 5 mm screw (depth: 6 mm)
ISO 5 mm screw (depth: 6 mm)
Tripod mount screw
(depth: 6 mm)
Tripod mount screw
(depth: 6 mm)
15
Page 18
Probes
CA-VP427Ø27 High Sensitivity ProbeHigh-sensitivity probe with a measurement area of 27 mm
Tripod mount screw
(depth: 6 mm)
ISO 5 mm screw (depth: 6 mm)
(unit: mm)
A
When a lightshielding cylinder
is mounted
ISO 5 mm screw (depth: 6 mm)
16
Viewed from A
Page 19
CA-VP410Ø10 High Sensitivity ProbeHigh-sensitivity probe with a measurement area of 10 mm
Tripod mount screw
(depth: 6 mm)
ISO 5 mm screw (depth: 6 mm)
(unit: mm)
Probe Guide
A
When a lightshielding cylinder
is mounted
ISO 5 mm screw (depth: 6 mm)
17
Viewed from A
Page 20
Probes
CA-MP410Ø10 Mini ProbeCompact probe with a measurement area of 10 mm
CA-MP410H Ø10 Mini Probe for High LuminanceCompact high-luminance probe with a measurement area of
10 mm
4 x ISO 3 mm screw (depth: 5.5 mm)
(unit: mm)
ISO 5 mm screw (depth: 6 mm)
3 (depth: 3 mm)
When a lightshielding cylinder
is mounted
ISO 3 mm screw (depth: 5 mm)
ISO 5 mm screw (depth: 6 mm)
Depth: 3 mm
Tripod mount screw
(depth: 6 mm)
18
Page 21
7. How to Set a Probe
Secure the display.
1
Secure the probe in a way to place its end 30 mm *1
2
Notes on setting
• When measuring displays with high view angle dependence, you can
• Use of the light-shielding cylinder supplied as a standard accessory not
•
away from the display surface.
Be sure to set it perpendicular to the display.
*1: L = 10 mm for mini probes.
obtain a higher measurement reproducibility by not changing the
setting angle θ for every measurement. In particular, the CA-VP410
high-sensitivity probe has a large aperture angle, which has a greater
eect on the viewing angle. When measuring displays with a high view
angle dependence, run user calibration and use the probe without
rotating it from the calibrated position.
only blocks ambient light but also makes it easier to achieve a proper
measurement distance and an angle of 90°.
•
For a 27 probe, the specification range of the length is 30 mm±10 mm.
•
For a 10 probe, the specification range of the length is 30 mm ±5 mm.
•
For a mini probe, the specification range of the length is 10 mm ±5 mm.
When strict low-luminance measurement is required, perform
measurement in a dark room with no external light. Even when measuring
in a dark room, the ambient light from the display itself may act as an
external light. In such situations, light shielding using a light-shielding
cylinder is recommended. If strict limitations are required, compare
measurement results from a general luminance meter such as the
CS-2000, and verify whether using a light-shielding cylinder has any eect.
L = 30 mm
*1
Display surface of the display
(when a light-shielding cylinder is not used)
(when a light-shielding cylinder is used)
2 mm (1 mm for mini probes)
Setting angle θ
Probe
Display
Probe Guide
8. How to Connect Cables
A probe has a USB connector for connecting a USB mini B plug and a connector for connecting an RS-232C mini DIN plug.
Before connection, make sure that power is not supplied to the probe and the external device. Be sure to connect a cable
plug to the specified connector.
• For USB connection with a PC, connect the USB mini B plug of the supplied USB cable IF-A28 to the USB connector of the
probe. Connect the USB plug on the other side of the cable to the USB connector of the PC.
USB connection (using IF-A28)
Mini B plug
IF-A28
USB plug
19
PC
Page 22
Probes
• For RS-232C connection with a PC or PLC, connect the mini DIN plug of an optional conversion cable IF-A29 to the
RS-232C connector of the probe, and the D-sub plug to the RS-232C connector of the PC or PLC via a cross cable. The
baud rate of the probe is set to 38,400 bps.
In this case, it is also necessary to use the cable's power supply line to supply power to the probe. As a power source,
5V/300 mA is required.
To control a probe from a PC via RS-232C, use the supplied SDK (CA-SDK2). For the specifications of the SDK and how to
use it, refer to the instruction manual for CA-SDK2.
RS-232C connection (using IF-A29)
PC
Mini DIN plug
RS-232C connector
RS-232C connector pin No. and functions
Pin No.Pin functionDetails
1ExSync or Trigger IN0 to 5 V, CMOS level, Max. 240 Hz
2Trigger OUT
3External power supply (5 V)Min. 4.9 V to Max. 5.5 V
4Tx D
5GND
6RTS
Mini-DIN 8-pin (female)
Trigger timing diagram
7RxD
8CTS
RS-232C connector
IF-A29
Cross
connection
BNC connector
External synchronization signal
Red
Black
t > 5 sect > 5 sec
+ −
5 V DC/300 mA
Trigger IN
Transition to trigger mode
Trigger OUT
Measurement start Measurement endMeasurement start Measurement end
• To input trigger signals or external synchronization signals to the probe, input the signals to the BNC plug of an optional
conversion cable IF-A29 or an optional BNC conversion cable IF-A35.
Input to the BNC plug
C-MOS logic level
Input trigger and external
synchronization signals
(Luminance/chromaticity measurement: 0.5 to 240Hz)
(Flicker measurement: 0.5 to 130 Hz)
BNC connector
(Operated
with 5 V)
20
Page 23
For connection with a CA-DP40 data processor, you can use a mini DIN cable or a USB cable.
• For RS-232C connection, connect the mini DIN plug of the RS mini DIN cable IF-A30 (or IF-A31 or IF-A32) for the data
processor to the RS-232C connector of the probe and the probe RS-232C connector of the CA-DP40 data processor.
Connection via a mini DIN cable (using IF-A30/31/32)
Mini DIN plugMini DIN plug
• For USB connection, connect the USB mini B plug of the supplied USB cable IF-A28 to the USB connector of the probe.
Connect the USB plug on the other side of the cable to the probe USB connector of the CA-DP40 data processor.
Connection via a USB cable (using IF-A28)
Mini B plug
IF-A30/31/32
IF-A28
USB plug
Data Processor
About Zero Calibration
Probe Guide
Zero calibration performs zero point adjustment while blocking entry of light into the probe's receptor.
Be sure to perform zero calibration whenever the instrument is started.
Notes on zero calibration
• Perform zero calibration if the ambient temperature has changed.
• During zero calibration, the shutter inside the probe automatically closes, but just in case, do not direct the end of the
probe toward an illuminant with luminance exceeding the upper limit of the measurement range.
1. How to Check Zero Calibration
To check whether zero calibration has been performed correctly, block entry of light into the end of the probe using a
blackout curtain, etc. so that the receptor is not exposed to light.
• Zero calibration has been performed correctly if zero is displayed for the Lv section on the controller's measured value
display. If a value other than zero is displayed, perform zero calibration again.
21
Page 24
Measurement Functions of Probes
1. Measurement Speeds
A probe has the following four modes (measurement speeds), which are available according to the purpose.
A probe outputs every measured value and a PC or data processor connected to it can retrieve all of them through
communication. Generally, however, not all measured values are displayed but values are displayed at an appropriate
interval.
FAST mode
Requires short measurement time, but may compromise measurement repeatability for low-luminance measurements.
SLOW mode
Repeats FAST mode measurements several times. Use this mode when you want to perform measurement with high
repeatability.
AUTO mode
Switches FAST, SLOW, and setting with a further longer integration time automatically depending on the luminance level.
Normally, this is the recommended setting.
LTD.AUTO mode
Switches between FAST and SLOW settings automatically depending on the luminance level.
Notes on selecting the measurement speed
• The currently selected measurement speed setting is stored in the probe.
The set measurement speed will be used at the next start-up.
22
Page 25
2. Measurement Synchronization Modes
Measurement synchronization modes are designed to select a measurement time (integration time) according to the
vertical scanning frequency of a display. A probe has the following six types of measurement synchronization modes. Select
a mode suitable for the type of the display to be measured.
Measurement
synchronization
mode
NTSCMode for measuring NTSC displays33.3 ms59.94 HzNot required
PAL
EXT
UNIV
INT
MANUAL
Mode for measuring PAL and SECAM
displays
Mode for measuring while synchronizing
with the display’s vertical
synchronization signal input (refer to
page 20 for how to input vertical
synchronization signals)
Mode used when the display’s vertical
scanning frequency is unknown or
cannot be input; a mode that can
measure any display
When the display’s vertical
synchronization signal is known, set that
information so that it can be used for
measurement.
Mode that allows you to specify the
integration time
Description of mode
Measurement time
(Integration time)
40.0 ms50 HzNot required
(1 vertical scanning
period)
×2
100 ms—Not required
(1 vertical scanning
period)
×2
Settable range
4.0 to 4000.0 ms
(For FMA:
7.7 to 4000.0 ms)
Vertical scanning
frequency
0.50 to
240.00 Hz
(For FMA:
0.50 to 130.00 Hz)
Settable range
0.50 to
240.00 Hz
(For FMA:
0.50 to 130.00 Hz)
—Not required
Display's vertical
synchronization
signal input
Required
Not required
Probe Guide
Memo
On a flickering LCD, the screen is fluctuating at half the frequency of the vertical scanning frequency. When you
measure a display in this state, the luminance and the chromaticity vary if you set the same integration time as the
vertical scanning frequency.
When measuring with the measurement synchronization mode set to MANUAL, set the integration time double (or
another even multiple of) the vertical scanning frequency to ensure stable LCD measurement.
23
Page 26
Measurement Functions of Probes
Relationship between measurement speed and measurement synchronization mode
The measurement time (integration time) is determined by the selected measurement synchronization mode.
Also, the measurement speed (number of measurements/outputs possible per second) is determined by the measurement
synchronization mode and the following conditions:
• Luminance of the display to be measured
• Measurement mode
• Data output method (RS-232C or USB)
• Baud rate (for RS-232C)
• Number of probe to be used
Typical measurement speeds are described in the specifications.
24
Page 27
3. Measurement Modes
A probe has the following measurement modes:
Measurement modeDescription of mode
xyLv modeMode for displaying and outputting chromaticity coordinates x, y and luminance Lv
TduvLv modeMode for displaying and outputting correlated color temperature T, color dierence from
blackbody locus duv, and luminance Lv
u'v'Lv modeMode for displaying and outputting u'v' chromaticity diagram coordinates (CIE1976 UCS
chromaticity diagram) and luminance Lv
Flicker modeDisplays the flicker amount in the contrast method (AC/DC). The unit is “%”.
Displays the flicker amount in the JEITA method. The unit is “dB”.
XYZ modeMode for displaying and outputting tristimulus values XYZ
λdPeLv modeMode for displaying and outputting dominant wavelength λd, excitation purity Pe, and
luminance Lv
* Under the initial settings, the mode is set to xyLv mode at the start-up.
Probe Guide
25
Page 28
User Memory
1. About Calibration Channels
A probe has calibration channels from CH00 to CH99 (100 channels in total).
The following can be set for each of CH00 to CH99:
(1) Correction coecient for user calibration
(2) Target
(3) ID
CH00 is a channel for calibration in accordance with the Konica Minolta calibration standard. It can only be used for setting
targets or IDs.
Users can use CH01 to CH99 for calibration or saving targets.
Under the initial settings, CH01 to CH99 are also provided with the same Konica Minolta calibration value as with CH00.
Each channel can also be assigned with an ID.
How the memory is used during user calibration
xyLv calibration value
for W
Measured value for W
Used for single-point calibration
Used for RGB+W matrix calibration
xyLv calibration
value for R
Measured value
for R
xyLv calibration
value for G
Measured value
for G
xyLv calibration
value for B
Measured value
for B
2. About Targets
Targets become the base color for measuring how dierent another measurable color is, which is required for color
dierence display. Targets can be set for each probe of each calibration channel.
Targets are set after the following operations:
(1) User calibration: After user calibration, the calibration value is set as a target at the same time.
(2) Setting/changing a target: To be performed in the following cases:
• To set a target in calibration channel CH00
• To set a dierent color from user calibration as a target in a user-calibrated calibration
channel
• To perform measurement without performing user calibration but in accordance with the
Konica Minolta calibration standard and to use color dierence display
• If a correction coecient for user calibration is input, a target is set at the same time, which erases the previously set
target.
• To change a set target, set/change a target. Changing a target does not change the correction coecient set for user
calibration.
In a same calibration channel, the target is saved to common channel, irrespective of measurement mode.
As a result, the target set last is saved, irrespective of measurement mode.
For one calibration channel and one probe number:
• The correction coecient for user calibration is common to xyLv mode, TduvLv mode, u'v'Lv mode, and XYZ mode.
• The target is common to all measurement modes.
26
Page 29
3. About User Calibration
• User calibration allows users to set their own correction coecients to the calibration channel of the probe by measuring
any display colors and setting calibration values (x, y, Lv) to the instrument.
Afterward, the value corrected with the specified correction coecients will be displayed/output for every measurement.
• Two types of user calibration can be performed for a CA-410 series probe: single-point calibration and RGB+W matrix
calibration.
• RGB+W matrix calibration has been performed when the instrument is shipped from the factory.
• User calibration is performed independently for each probe.
• User calibration can be performed for every calibration channel (except for CH00).
• When a probe is used for the first time after being shipped from the factory, measurement is performed in accordance
with the Konica Minolta calibration standard in all calibration channel. You can use user calibration to apply the following
corrections to measurements based on the specified correction coecients.
(1) Corrects the dierence in the reading due to the deviation of spectral response from the CIE 1931 color-matching
function.
(2) Corrects the dierence in the readings between probes when multiple probes are used for measurement.
• When user calibration is performed, the color is set to the same calibration channel as its target at the same time.
Targets become the base color for displaying how dierent another measurable color is, which is used as the base for
color dierence display.
Probe Guide
When multiple probes are used
When using multiple probes for measurement, you can correct the dierence in the readings by performing user calibration
as follows:
When the target color value for the display to be measured is known
Make a standardized color display on a standard display and perform user calibration for all probes.
When the target color value for the display to be measured is not known
(1) Determine one master probe.
(2) Set the measurement mode to xyLv mode, make the master probe face perpendicular to the display on which the target
color is displayed, and perform measurement.
(3) Record the measured value.
(4) Using the display on which the target color is displayed and the value displayed in (3), perform user calibration for other
probes.
27
Page 30
Specifications of Each Probe
Specifications of Each Probe (1)
CA-P427CA-P427H
Measurement areaØ27 mmØ27 mm
Acceptance angle±2.5°±2.5°
Accuracy guaranteed measurement distance30±10 mm30±10 mm
Display range Luminance0.0001 to 5000 cd/m
2
0.0001 to 30000 cd/m
ChromaticityDisplayed in 4 digitsDisplayed in 4 digits
LuminanceAccuracy guaranteed luminance range0.001 to 5000 cd/m
Accuracy (for white)
Repeatability
*1
(2σ)
AUTO0.001 - cd/m
*1, *3
0.001 - cd/m
0.01 - cd/m
0.1 - cd/m
1 - cd/m
10 - cd/m
100 - cd/m
0.01 - cd/m
0.1 - cd/m
1 - cd/m
10 - cd/m
100 - cd/m
2
2
2
2
2
2
2
2
2
2
2
2
±9%---
±2%±9%
±1.5%±2%
±1.5%±1.5%
±1.5%±1.5%
±1.5%±1.5%
10%---
1%10%
0.40%1%
0.10%0.40%
0.10%0.10%
0.10%0.10%
Chromaticity Accuracy guaranteed luminance range0.01 to 5000 cd/m
*6
Flicker
(Contrast)
Accuracy (for white)
*1, *3
0.01 - cd/m
0.1 - cd/m
1 - cd/m
10 - cd/m
100 - cd/m
Repeatability
*1
(2σ)
In 100 cd/m2
(for monochrome)
AUTO0.01 - cd/m
100 cd/m
*2
0.1 - cd/m
1 - cd/m
10 - cd/m
100 - cd/m
Measurement luminance range5 to 1500 cd/m
Range to be measured (flicker frequency)0.25 to 65 Hz0.25 to 65 Hz
Accuracy
30 Hz, AC/DC 10% sine wave
60 Hz, AC/DC 10% sine wave
Repeatability (2σ)
20 to 65 Hz,
2
2
2
2
2
2
2
2
2
2
2
±0.003---
±0.002±0.003
±0.002±0.002
±0.002±0.002
±0.002±0.002
±0.003±0.003
0.0035 ---
0.0015 0.0035
0.0004 0.0015
0.0003 0.0004
0.0002 0.0003
2
± 0.4%± 0.4%
± 0.7%± 0.7%
0.3%0.3%
2
2
0.01 to 30000 cd/m
0.1 to 30000 cd/m
30 to 9000 cd/m
AC/DC 10% sine wave
Flicker (CA-310 Mode)
Flicker
(JEITA)
Measurement luminance range5 to 1500 cd/m
Range to be measured (flicker frequency)0.42 to 65 Hz0.42 to 65 Hz
Accuracy
Repeatability (2σ)
30 Hz, AC/DC 4% sine wave
30 Hz, AC/DC 1.2% sine wave
30 Hz, AC/DC 4% sine wave
30 Hz, AC/DC 1.2% sine wave
±0.35 dB±0.35 dB
±0.35 dB---
0.1 dB0.1 dB
0.3 dB---
2
30 to 9000 cd/m
2
2
2
2
2
28
Page 31
Specifications of Each Probe (1)
CA-P427CA-P427H
*6
Flicker
(Contrast)
Measurement luminance range5 to 5000 cd/m
Range to be measured (flicker frequency)0.25 to 200 Hz0.25 to 200 Hz
Accuracy30 Hz,
± 1.2%± 1.2%
2
30 to 30000 cd/m
AC/DC 10% sine wave
60 Hz,
± 1.7%± 1.7%
AC/DC 10% sine wave
Repeatability (2σ)
20 to 65 Hz,
1.7%1.7%
AC/DC 10% sine wave
XYZ (Wide Frequency Mode)
Flicker
(JEITA)
Measurement luminance range5 to 4500 cd/m
Range to be measured (flicker frequency)0.42 to 200 Hz0.42 to 200 Hz
Accuracy30 Hz,
±0.35 dB±0.35 dB
2
30 to 27000 cd/m
AC/DC 4% sine wave
30 Hz,
±0.35 dB---
AC/DC 1.2% sine wave
Repeatability (2σ)
30 Hz,
0.4 dB0.4 dB
AC/DC 4% sine wave
30 Hz,
1.5 dB---
AC/DC 1.2% sine wave
Waveform Measurement luminance range1 to 5000 cd/m
Weight270 g (including mount)270 g (including mount)
Power supply5 V DC (input from USB bus
power line or RS communication
connector)
Operating temperature/humidity range
*5
Temperature: 10 to 35°C, Relative humidity: 85% or less with no
5 V DC (input from USB bus
power line or RS communication
connector)
condensation
Storage temperature/humidity range0 to 45°C, relative humidity 85% or less (at 35°C) with no
condensation
2
2
2
Probe Guide
Measured with Konica Minolta’s specified PC and probe connected directly, using the supplied measurement software.
*1: Measured under Konica Minolta's standard light source (6,500 K).
*2: The luminance for monochrome is measured when reading of luminance for white is 100 cd/m
2
.
*3: Temperature 23°C/±2°C, relative humidity 40%±10%
*4: In NTSC sync mode using USB with one probe
*5: Amount of changed indicated values (with indicated values of 23°C, 40% for reference): Range for luminance (white) is no more than
±2%; range for chromaticity (with measured luminance of 100 cd/m
2
) is no more than ±0.002 (white) / ±0.003 (monochrome)
*6: “Flicker (CA-310 Mode)” and “XYZ (Wide Frequency Mode)” are mode names for PC Software CA-S40. “XYZ (Wide Frequency Mode)”
can only be used when no CA-DP40 data processor is connected.
29
Page 32
Specifications of Each Probe (2)
CA-P410CA-P410H
Measurement areaØ10 mmØ10 mm
Acceptance angle±5°±5°
Accuracy guaranteed measurement distance30±5 mm30±5 mm
Display range Luminance0.0001 to 5000 cd/m
ChromaticityDisplayed in 4 digitsDisplayed in 4 digits
LuminanceAccuracy guaranteed luminance range0.01 to 5000 cd/m
Accuracy (for white)
Repeatability
*1
(2σ)
AUTO0.001 - cd/m
*1, *3
0.001 - cd/m
0.01 - cd/m
0.1 - cd/m
1 - cd/m
10 - cd/m
100 - cd/m
0.01 - cd/m
0.1 - cd/m
1 - cd/m
10 - cd/m
100 - cd/m
2
2
2
2
2
2
2
2
2
2
2
2
------
±2.5%---
±2%±2.5%
±2%±2%
±1.5%±2%
±1.5%±1.5%
------
2%---
0.60%2%
0.20%0.60%
0.10%0.20%
0.10%0.10%
Chromaticity Accuracy guaranteed luminance range0.01 to 5000 cd/m
*6
Flicker
(Contrast)
Accuracy (for white)
*1, *3
0.01 - cd/m
0.1 - cd/m
1 - cd/m
10 - cd/m
100 - cd/m
Repeatability
*1
(2σ)
In 100 cd/m2
(for monochrome)
AUTO0.01 - cd/m
100 cd/m
*2
0.1 - cd/m
1 - cd/m
10 - cd/m
100 - cd/m
Measurement luminance range15 to 3000 cd/m2 or more90 to 18000 cd/m2 or more
Range to be measured (flicker frequency)0.25 to 65 Hz0.25 to 65 Hz
Accuracy
30 Hz, AC/DC 10% sine wave
60 Hz, AC/DC 10% sine wave
Repeatability (2σ)
20 to 65 Hz,
2
2
2
2
2
2
2
2
2
2
2
±0.006---
±0.002±0.006
±0.002±0.002
±0.002±0.002
±0.002±0.002
±0.003±0.003
0.0070 ---
0.0020 0.0070
0.0008 0.0020
0.0005 0.0008
0.0003 0.0005
± 0.4%± 0.4%
± 0.7%± 0.7%
0.3%0.3%
2
2
AC/DC 10% sine wave
Flicker (CA-310 Mode)
Flicker
(JEITA)
Measurement luminance range15 to 3000 cd/m
Range to be measured (flicker frequency)0.42 to 65 Hz0.42 to 65 Hz
Accuracy
Repeatability (2σ)
30 Hz, AC/DC 4% sine wave
30 Hz, AC/DC 1.2% sine wave
30 Hz, AC/DC 4% sine wave
30 Hz, AC/DC 1.2% sine wave
±0.35 dB±0.35 dB
±0.35 dB---
0.1 dB0.1 dB
0.3 dB---
2
or more90 to 18000 cd/m2 or more
2
0.0001 to 30000 cd/m
0.1 to 30000 cd/m
0.1 to 30000 cd/m
2
2
2
30
Page 33
Specifications of Each Probe (2)
CA-P410CA-P410H
*6
Flicker
(Contrast)
Measurement luminance range15 to 5000 cd/m
Range to be measured (flicker frequency)0.25 to 200 Hz0.25 to 200 Hz
Accuracy30 Hz,
± 0.7%± 0.7%
2
90 to 30000 cd/m
AC/DC 10% sine wave
60 Hz,
± 1.1%± 1.1%
AC/DC 10% sine wave
Repeatability (2σ)
20 to 65 Hz,
1.0%1.0%
AC/DC 10% sine wave
XYZ (Wide Frequency Mode)
Flicker
(JEITA)
Measurement luminance range15 to 5000 cd/m
Range to be measured (flicker frequency)0.42 to 200 Hz0.42 to 200 Hz
Accuracy30 Hz,
±0.35 dB±0.35 dB
2
90 to 30000 cd/m
AC/DC 4% sine wave
30 Hz,
±0.35 dB---
AC/DC 1.2% sine wave
Repeatability (2σ)
30 Hz,
0.3 dB0.3 dB
AC/DC 4% sine wave
30 Hz,
0.9 dB---
AC/DC 1.2% sine wave
Waveform Measurement luminance range1 to 5000 cd/m
Weight200 g (including mount)280 g (including mount)
Power supply5 V DC (input from USB bus
power line or RS communication
connector)
Operating temperature/humidity range
*5
Temperature: 10 to 35°C, Relative humidity: 85% or less with no
5 V DC (input from USB bus
power line or RS communication
connector)
condensation
Storage temperature/humidity range0 to 45°C, relative humidity 85% or less (at 35°C) with no
condensation
2
2
2
Probe Guide
Measured with Konica Minolta’s specified PC and probe connected directly, using the supplied measurement software.
*1: Measured under Konica Minolta's standard light source (6,500 K).
*2: The luminance for monochrome is measured when reading of luminance for white is 100 cd/m
2
.
*3: Temperature 23°C/±2°C, relative humidity 40%±10%
*4: In NTSC sync mode using USB with one probe
*5: Amount of changed indicated values (with indicated values of 23°C, 40% for reference): Range for luminance (white) is no more than
±2%; range for chromaticity (with measured luminance of 100 cd/m
2
) is no more than ±0.002 (white) / ±0.003 (monochrome)
*6: “Flicker (CA-310 Mode)” and “XYZ (Wide Frequency Mode)” are mode names for PC Software CA-S40. “XYZ (Wide Frequency Mode)”
can only be used when no CA-DP40 data processor is connected.
31
Page 34
Specifications of Each Probe (3)
CA-MP410CA-MP410H
Measurement areaØ10 mmØ10 mm
Acceptance angle±5°±5°
Accuracy guaranteed measurement distance10±5 mm10±5 mm
Display range Luminance0.0001 to 5000 cd/m
ChromaticityDisplayed in 4 digitsDisplayed in 4 digits
LuminanceAccuracy guaranteed luminance range0.01 to 5000 cd/m
Accuracy (for white)
Repeatability
*1
(2σ)
AUTO0.001 - cd/m
*1, *3
0.001 - cd/m
0.01 - cd/m
0.1 - cd/m
1 - cd/m
10 - cd/m
100 - cd/m
0.01 - cd/m
0.1 - cd/m
1 - cd/m
10 - cd/m
100 - cd/m
2
2
2
2
2
2
2
2
2
2
2
2
------
±2.5%---
±2%±2.5%
±2%±2%
±1.5%±2%
±1.5%±1.5%
------
2.4%---
0.70%2.4%
0.25%0.70%
0.12%0.25%
0.10%0.12%
Chromaticity Accuracy guaranteed luminance range0.01 to 5000 cd/m
*6
Flicker
(Contrast)
Accuracy (for white)
*1, *3
0.01 - cd/m
0.1 - cd/m
1 - cd/m
10 - cd/m
100 - cd/m
Repeatability
*1
(2σ)
In 100 cd/m2
(for monochrome)
AUTO0.01 - cd/m
100 cd/m
*2
0.1 - cd/m
1 - cd/m
10 - cd/m
100 - cd/m
Measurement luminance range15 to 3000 cd/m2 or more90 to 18000 cd/m2 or more
Range to be measured (flicker frequency)0.25 to 65 Hz0.25 to 65 Hz
Accuracy
30 Hz, AC/DC 10% sine wave
60 Hz, AC/DC 10% sine wave
Repeatability (2σ)
20 to 65 Hz,
2
2
2
2
2
2
2
2
2
2
2
±0.006---
±0.002±0.006
±0.002±0.002
±0.002±0.002
±0.002±0.002
±0.003±0.003
0.0085 ---
0.0025 0.0085
0.0010 0.0025
0.0006 0.0010
0.0004 0.0006
± 0.4%± 0.4%
± 0.7%± 0.7%
0.3%0.3%
2
2
AC/DC 10% sine wave
Flicker (CA-310 Mode)
Flicker
(JEITA)
Measurement luminance range15 to 3000 cd/m
Range to be measured (flicker frequency)0.42 to 65 Hz0.42 to 65 Hz
Accuracy
Repeatability (2σ)
30 Hz, AC/DC 4% sine wave
30 Hz, AC/DC 1.2% sine wave
30 Hz, AC/DC 4% sine wave
30 Hz, AC/DC 1.2% sine wave
±0.35 dB±0.35 dB
±0.35 dB---
0.1 dB0.1 dB
0.3 dB---
2
or more90 to 18000 cd/m2 or more
2
0.0001 to 30000 cd/m
0.1 to 30000 cd/m
0.1 to 30000 cd/m
2
2
2
32
Page 35
Specifications of Each Probe (3)
CA-MP410CA-MP410H
*6
Flicker
(Contrast)
Measurement luminance range15 to 5000 cd/m
Range to be measured (flicker frequency)0.25 to 200 Hz0.25 to 200 Hz
Accuracy30 Hz,
± 0.9%± 0.9%
2
90 to 30000 cd/m
AC/DC 10% sine wave
60 Hz,
± 1.3%± 1.3%
AC/DC 10% sine wave
Repeatability (2σ)
20 to 65 Hz,
1.3%1.3%
AC/DC 10% sine wave
XYZ (Wide Frequency Mode)
Flicker
(JEITA)
Measurement luminance range15 to 5000 cd/m
Range to be measured (flicker frequency)0.42 to 200 Hz0.42 to 200 Hz
Accuracy30 Hz,
±0.35 dB±0.35 dB
2
90 to 30000 cd/m
AC/DC 4% sine wave
30 Hz,
±0.35 dB---
AC/DC 1.2% sine wave
Repeatability (2σ)
30 Hz,
0.3 dB0.3 dB
AC/DC 4% sine wave
30 Hz,
1.2 dB---
AC/DC 1.2% sine wave
Waveform Measurement luminance range1 to 5000 cd/m
Weight200 g (including mount)200 g (including mount)
Power supply5 V DC (input from USB bus
power line or RS communication
connector)
Operating temperature/humidity range
*5
Temperature: 10 to 35°C, Relative humidity: 85% or less with no
5 V DC (input from USB bus
power line or RS communication
connector)
condensation
Storage temperature/humidity range0 to 45°C, relative humidity 85% or less (at 35°C) with no
condensation
2
2
2
Probe Guide
Measured with Konica Minolta’s specified PC and probe connected directly, using the supplied measurement software.
*1: Measured under Konica Minolta's standard light source (6,500 K).
*2: The luminance for monochrome is measured when reading of luminance for white is 100 cd/m
2
.
*3: Temperature 23°C/±2°C, relative humidity 40%±10%
*4: In NTSC sync mode using USB with one probe
*5: Amount of changed indicated values (with indicated values of 23°C, 40% for reference): Range for luminance (white) is no more than
±2%; range for chromaticity (with measured luminance of 100 cd/m
2
) is no more than ±0.002 (white) / ±0.003 (monochrome)
*6: “Flicker (CA-310 Mode)” and “XYZ (Wide Frequency Mode)” are mode names for PC Software CA-S40. “XYZ (Wide Frequency Mode)”
can only be used when no CA-DP40 data processor is connected.
33
Page 36
Specifications of Each Probe (4)
CA-VP427CA-VP410
Measurement areaØ27 mmØ10 mm
Acceptance angle±2.5°±8.5°
Accuracy guaranteed measurement distance30±10 mm30±5 mm
Display range Luminance0.0001 to 3000 cd/m
ChromaticityDisplayed in 4 digitsDisplayed in 4 digits
LuminanceAccuracy guaranteed luminance range0.001 to 3000 cd/m
Accuracy (for white)
*1, *3
0.001 - cd/m
0.01 - cd/m
0.1 - cd/m
1 - cd/m
10 - cd/m
100 - cd/m
Repeatability
*1
(2σ)
AUTO0.001 - cd/m
0.01 - cd/m
0.1 - cd/m
1 - cd/m
10 - cd/m
100 - cd/m
Chromaticity Accuracy guaranteed luminance range0.01 to 3000 cd/m
Accuracy (for white)
*1, *3
0.01 - cd/m
0.1 - cd/m
1 - cd/m
10 - cd/m
100 - cd/m
Repeatability
*1
(2σ)
In 100 cd/m2
(for monochrome)
AUTO0.01 - cd/m
100 cd/m
*2
0.1 - cd/m
1 - cd/m
10 - cd/m
100 - cd/m
Flicker
*6
(Contrast)
Measurement luminance range-----Range to be measured (flicker frequency)-----Accuracy
30 Hz, AC/DC 10% sine wave
60 Hz, AC/DC 10% sine wave
Repeatability (2σ)
20 to 65 Hz,
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
±9%±9%
±2%±2.5%
±1.5%±2%
±1.5%±2%
±1.5%±1.5%
±1.5%±1.5%
10%7%
1%1%
0.25%0.25%
0.10%0.10%
0.10%0.10%
0.10%0.10%
2
±0.003±0.003
±0.002±0.002
±0.002±0.002
±0.002±0.002
±0.002±0.002
±0.003±0.003
0.0030 0.0020
0.0008 0.0008
0.0003 0.0003
0.0002 0.0002
0.0002 0.0002
------
------
------
AC/DC 10% sine wave
Flicker
(JEITA)
Flicker (CA-310 Mode)
Measurement luminance range-----Range to be measured (flicker frequency)-----Accuracy
Repeatability (2σ)
30 Hz, AC/DC 4% sine wave
30 Hz, AC/DC 1.2% sine wave
30 Hz, AC/DC 4% sine wave
30 Hz, AC/DC 1.2% sine wave
------
------
------
------
2
2
0.0001 to 3000 cd/m
0.001 to 3000 cd/m
0.01 to 3000 cd/m
2
2
2
34
Page 37
Specifications of Each Probe (4)
CA-VP427CA-VP410
Flicker
*6
(Contrast)
Measurement luminance range5 to 3000 cd/m
Range to be measured (flicker frequency)0.25 to 200 Hz0.25 to 200 Hz
Accuracy
30 Hz,
± 1.1% ± 0.4%
2
15 to 3000 cd/m
AC/DC 10% sine wave
60 Hz,
± 1.7%± 0.7%
AC/DC 10% sine wave
Repeatability (2σ)
20 to 65 Hz,
1.6% 0.3%
AC/DC 10% sine wave
Flicker
(JEITA)
XYZ (Wide Frequency Mode)
Measurement luminance range5 to 3000 cd/m
Range to be measured (flicker frequency)0.42 to 200 Hz0.42 to 200 Hz
Accuracy30 Hz,
±0.35 dB±0.35 dB
2
15 to 2000 cd/m
AC/DC 4% sine wave
30 Hz,
±0.35 dB±0.35 dB
AC/DC 1.2% sine wave
Repeatability (2σ)
30 Hz,
0.4 dB0.1 dB
AC/DC 4% sine wave
30 Hz,
1.4 dB0.3 dB
AC/DC 1.2% sine wave
Waveform Measurement luminance range1 to 3000 cd/m
2
1 to 2500 cd/m
Sampling frequency3 kHz3 kHz
Accuracy
guaranteed
measurement
*4
speed
LVxyAUTO0.001 - cd/m
0.15 - cd/m
2 - cd/m
2
2
2
1 time/sec1 time/sec
5 times/sec5 times/sec
20 times/sec20 times/sec
Flicker (Contrast)------
Flicker (JEITA)------
Measurement synchronization modeNTSC, PAL, EXT, UNIV, INT, MANU
Weight510 g (including mount)570 g (including mount)
Power supply5 V DC (input from USB bus
power line or RS communication
connector)
Operating temperature/humidity range
*5
Temperature: 10 to 35°C, Relative humidity: 85% or less with no
5 V DC (input from USB bus
power line or RS communication
connector)
condensation
Storage temperature/humidity range0 to 45°C, relative humidity 85% or less (at 35°C) with no
condensation
2
2
2
Probe Guide
Measured with Konica Minolta’s specified PC and probe connected directly, using the supplied measurement software.
*1: Measured under Konica Minolta's standard light source (6,500 K).
*2: The luminance for monochrome is measured when reading of luminance for white is 100 cd/m
2
.
*3: Temperature 23°C/±2°C, relative humidity 40%±10%
*4: In NTSC sync mode using USB with one probe
*5: Amount of changed indicated values (with indicated values of 23°C, 40% for reference): Range for luminance (white) is no more than
±2%; range for chromaticity (with measured luminance of 100 cd/m
2
) is no more than ±0.002 (white) / ±0.003 (monochrome)
*6: “Flicker (CA-310 Mode)” and “XYZ (Wide Frequency Mode)” are mode names for PC Software CA-S40. “XYZ (Wide Frequency Mode)”
can only be used when no CA-DP40 data processor is connected.
35
Page 38
Page 39
PC Software
CA-S40
CA-S40Overview
PC Software CA-S40 ................................................................38
Before using the software, please read this manual as well as the instruction manuals of
your instrument and personal computer (PC) carefully to ensure safe and correct use.
Introduction
PC Software CA-S40 For Color Analyzers is software that loads and saves measurements and data on a computer connected
with the probe or data processor of Display Color Analyzer CA-410 via a USB cable.
Please note that this manual assumes that the reader is familiar with the basic operations of Windows.
The USB flash drive, provided as a standard accessory, includes a PDF version of this instruction manual.
To view PDF files, you will need Adobe Acrobat Reader® from Adobe Systems Incorporated. The latest Adobe Acrobat
Reader® can be downloaded for free from the Adobe website.
Software License Agreement
The terms of the license agreement for this software are provided in the [Software License Agreement] dialog box displayed
during the installation process. This software can be used only if you agree to all the terms of the agreement.
KONICA MINOLTA holds the copyrights to the CA-410 USB driver and this manual stored on the USB flash drive.
Notes on Use
• This software is application software for use with Windows 7, Windows 10, and macOS Sierra. Please note that this
software does not include any OS.
• One of the above OS must be installed on the PC in order to use this software.
• When plugging the USB flash drive into the USB port of your PC, note the correct orientation and plug it in gently.
• To connect the CA-410 series probe or data processor to a PC, you need the dedicated USB driver.
• This software only controls the CA-410 series probe or data processor connected to the PC with a USB cable. It is not
possible to control the probe or data processor of the CA-410 series connected by anything other than USB with this
software. It does not support the CA-210 or CA-310 series.
• Make sure that the connector is oriented correctly and connected securely.
• Do not touch or otherwise dirty the terminal portion of the connector or subject the terminals to unreasonable force.
• Use a cord with a suitable length. Putting tension on the cable may cause connection failure or wire breakage.
• Disconnect the connection cord only by holding the plug or the connector. Also, do not pull the cord itself or bend it with
unreasonable force.
• Avoid exposing the USB flash drive to rapid temperature changes and condensation.
• Avoid leaving it in locations where it may be exposed to high temperatures from direct sunlight or heaters.
• Do not drop the USB flash drive or subject it to strong impact.
• Keep the USB flash drive away from water, alcohol, paint thinners, and other such substances.
• When removing the USB flash drive, follow the instructions on the manual for your PC.
Notes on Storage
• Keep the USB flash drive in a safe place.
• Avoid leaving the USB flash drive in locations where it may be exposed to high temperatures from direct sunlight or
heaters.
• Do not keep the USB flash drive in a humid environment.
Every eort has been made in the preparation of this product to ensure the accuracy of its contents. However, should you
have any questions or find any errors, please contact your retailer or a KONICA MINOLTA-authorized service facility.
About this manual
The screen capture shots in this manual are examples and may be slightly dierent from actual screen images. Windows
screens are shown in most cases. Menu items, button names, etc. and their respective functions are the same under macOS
even if their position on the screen is slightly dierent.
38
Page 41
Overview
System environment........................................................................................................ 40
Notes on measurement ................................................................................................... 43
1. Notes on hardware operation ................................................................................ 43
CA-S40Overview
39
Page 42
System environment
Operating requirements
OS Windows 7 Professional 32bit (with hotfix KB2964073 applied)
Windows 7 Professional 64bit (with hotfix KB2964073 applied)
Windows 10 Pro 32bit
Windows 10 Pro 64bit
macOS Sierra
• Hotfix KB2964073 for Windows 7 can be obtained from the following Microsoft® website:
https://support.microsoft.com/en-us/help/2964073/stop-error-7e-or-stop-error-d1-occurs-during-read-operations-on-a-usbs
• The PC’s system configuration must be either that recommended for the above OS or have the following specifications,
whichever is higher.
PC Computer equipped with Intel Core i series, or equivalent or higher processor
RAM 4 GB or more
Hard disk More than 100 MB of available space
Out of the above, there must be at least 50 MB of available space on the system drive (drive where
the OS is installed)
Display resolution 1280 x 768 or higher, 16-bit color or higher
Other USB port required to connect the USB flash drive
USB 2.0 or above required to connect the instrument
Double-click [setup.exe] on the supplied USB flash drive or in the Windows folder where the file was downloaded. Proceed
as instructed on the screen. When you agree to the license agreement, the installation process starts and the following
folders are created under the [Program Files (x86)] folder. The program and manual are installed, and the CA-S40 startup
menu is registered in the Start menu.
[Program Files (x86)] – [KONICA MINOLTA] – [CA-S40]
2. Uninstalling the software (Windows)
Open [Control Panel] - [Uninstall a program], select CA-S40 from the list, and double-click it or click [Uninstall].
3. Installing the software (macOS)
Double-click [setup.pkg] on the supplied USB flash drive or in the Mac folder where the file was downloaded. Proceed as
instructed on the screen. When you agree to the license agreement, the installation process starts and the following folders
are created under the [Application] folder. The program and manual are installed, and the CA-S40 startup menu is registered
in the Launchpad.
[Application] – [KONICA MINOLTA] – [CA-S40]
4. Uninstalling the software (macOS)
Right-click the CA-S40 application and click [Move to Trash].
40
Page 43
Major functions
Display (Color) Lv, x, y, Lv, u', v’, Lv, Tcp, duv, X, Y, Z, Lv, λd, and Pe values, chromaticity diagram, and trend graph
(JEITA) Flicker value (dB) and trend graph, spectral intensity value and graph, and waveform graph
(FMA) Flicker value (%) and trend graph
About displayed values
• Although the CA-S40 oers a higher calculation accuracy through internal calculations using more decimal places than
found in the displayed values, some calculations, such as rounding, and those during color space conversion may result
in a dierent minimum decimal place from the output of the instrument by one digit. In addition, plotted positions may
also dier from those obtained from the displayed values for the same reason.
Instrument controlSingle measurementContinuous measurement
Interval measurement
Color and FMA
Number of measurements: 1 to 4,000 times/Measurement interval: 0 to 7,259 sec
Number of measurements: 4,001 to 40,000 times/Measurement interval: 1 to 7,259 sec
JEITA
Number of measurements: 1 to 4,000 times/Measurement interval: 0 to 7,259 sec
Instrument settings Retrieval of data stored in the instrument
User calibration
Data processors cannot be connected when using ver. 1.1 or earlier of the CA-S40 for macOS.
CA-S40Overview
Target Writing of a target to the instrument
Data list Listing of measurement data
Edit (delete, copy)
External input/output Saving in text/CSV format, copying lists to the clipboard
41
Page 44
Operation flow
Connect the instrument
Start the software
(P.45)
(P.45)
When multiple instruments are connected
Set up the
instrument
Zero calibration
Measurement
Single measurement
Set the measurement
conditions
User calibration
Set the target
Continuous measurementInterval measurementPackage
Set up options
(P.54)
(P.49, P.60)
(P.65)
Select the instrument
(P.57)
(P.46)
Exit the software
Disconnect the instrument
(P.45)
Import or
save data
Import measurement data
Save list and graph
(P.49)
(P.47, P.50)
42
Page 45
Notes on measurement
1. Notes on hardware operation
• Use the USB cable to connect the instrument. This software can only control probes or data processors of the CA-410
series connected by USB. It is not possible to control the probe or data processor of the CA-410 series connected by
anything other than USB with this software. Also, the CA-210/310 series can not be used with this software.
• The CA-VP410 high-sensitivity probe has a large aperture angle, which creates a higher dependence on panel directivity.
When measuring panels with high directivity, run arbitrary calibration and use the probe without rotating it from the
calibrated position.
CA-S40Overview
43
Page 46
Operation Guide
Preparing and starting the software ................................................................................. 45
1. Connect the instrument and start the software ...................................................... 45
1. Setting the target .................................................................................................. 65
Data processor data ......................................................................................................... 66
1. Handling of stored data ........................................................................................ 66
2. Handling of Log data ............................................................................................. 67
44
Page 47
Preparing and starting the software
1. Connect the instrument and start the software
Connect the instrument
(1) Connect the instrument (CA probe or data processor) to the PC’s USB port. When connecting the instrument, connect
it to a USB port that can supply power to the probe. When a message appears prompting for the installation of the USB
driver, install the USB driver in the [Driver] folder.
(2) Check whether the probe’s LED indicator blinks.
(3) Go to [Start menu] - [KONICA MINOLTA] - [CA-S40] and click CA-S40 to start the software.
• When only one instrument is connected, the measurement screen appears. (See 1-1. Screen layout on P.46)
• When multiple instruments are connected, the instrument selection screen appears. Select one instrument. (See
2-2. Instrument selection CA Info. on P.57)
• When no instrument is connected, the error screen appears.
• When the software starts and the instrument to be used has been chosen, the information of the instrument is
displayed in the status bar, indicating that it is ready to be operated from the PC.
• If a probe that has not been calibrated to its zero point is directly connected to the PC, the internal shutter closes
automatically to start zero calibration.
• When zero calibration is completed, the LED indicator changes from blinking to lighting up steadily.
Notes: The probe operates on USB bus power and requires 5 V and 300 mA. It is recommended that an AC power
supply is used for the PC as necessary. It is also recommended that the instrument be disconnected whenever it
is not in use.
CA-S40
(4) Click [Settings] and set the measurement conditions.
Disconnect the instrument
Exit the software and disconnect the instrument from the PC.
Precautions for connecting the instrument
• When the instrument’s communication is controlled by another PC or other software, CA-S40 cannot control its
communication.
Make sure that the instrument’s communication is not controlled by other software. If it is, exit that software as necessary.
If the instrument is connected to another PC, disconnect that connection to ensure that it is only connected to the
intended controlling PC.
• Connection-related problems (*) may occur and an error message such as "CA device not connected" may be displayed
even after an instrument has been connected successfully and proper communication has been established. In this case,
restart CA-S40 after checking that the cable is connected and that the instrument (if a data processor is connected) is
turned on. Also, the instrument (data processor or probe) may not operate if the operating system on your PC is started
(or restarted) after the PC is connected to the instrument using a cable. In this case, exit CA-S40 and, if the instrument is a
data processor, turn it o and then on again, and if the instrument is a probe, disconnect and reconnect it to the PC before
restarting CA-S40.
* Ex. When CA-S40 is operated after one of the probes is disconnected from the data processor while the CA-S40 is
running with one or more probes connected to the data processor
When the [ESC] key of the data processor is pressed while CA-S40 is running
• If your PC enters power-saving mode while connected to a measurement instrument (such as a probe or data processor)
via USB, communication operations, such as for continuous measurement, may stop, or a forced termination of Windows
(blue screen) may occur. When using the PC in a high communication load situation, such as with continuous
measurement, change the power-saving mode settings in the PC control panel to prevent Windows from entering powersaving mode.
• A Firewall alert may be displayed when CA-S40 is started for the first time after installation. In this case, click “Allow
access” to grant access because CA-S40 does not connect to an external network.
• If you start the CA-S40 while the data processor startup screen is displayed, connection may not be possible. Please start
up the CA-S40 after the data processor starts up and zero calibration is completed.
Guide
Operation
45
Page 48
Measurement
1. Chromaticity measurement
Click the [Color] tab to display the following screen. On this screen, you can measure chromaticity and display the results in
the selected color space. User calibration and gamma measurement can also be performed.
Information about the instrument that is connected to the PC is displayed.
(3) Probe information
Probe information is displayed when a data processor is connected. Click here to select the probe.
(4) Measurement information
This section displays the synchronization mode, measurement speed, calibration mode, and measurement status.
(5) Help function
The instruction manual, saved on the PC when the software was installed, opens.
(6) License information
Click here to open a screen that shows the software name and version along with copyright information.
(7) Individual functions (pop-up)
Zero calibration, settings, data import from data processor, user calibration, gamma measurement
(8) Measurement start/pause/stop button (Operation with shortcut keys of Windows: Ctrl + M, macOS: Command + M is
possible)
(9) Color space selection tab
(10) Latest data display
This area shows the latest data in the color space specified in the color space selection tab above.
46
Page 49
(11) Chromaticity diagram
After selecting a graph area,
• Use the mouse wheel to zoom in and zoom out.
• Drag to move the display area.
• Double-click to return to the original scale.
• Right-click on the graph to display the following menu, which allows you to
copy the graph to the clipboard or a file.
Right-click to Clipboard Background Color: Black
Background Color: White
to File Background Color: Black
Background Color: White
* When the chromaticity diagram is enlarged or reduced during a continuous
measurement or interval measurement, the plotting may momentarily stop.
(Measurement is continuing)
(12) Data display toggling
Toggles between showing both the data list and trend graph, data list only, and trend graph only
(13) Data list (displayed when [List+Trend Graph] or [List] is selected for data display)
• Right-click on the list to display the following menu, which allows you to copy items and selected data to the
clipboard or delete selected measurements.
Right-click Title to Clipboard
to Clipboard
Line Clear
• [∆t] in the list shows the temperature change inside the probe from when zero calibration was performed.
• [ErrCode] in the list shows the following statuses: [ErrCode] in the list shows the following statuses: (Refer to page
76 for information on other error codes.)
0: Successfully processed.
1:
The currently selected calibration channel uses user calibration data or target data that has been set using a
dierent probe.
2: The temperature inside the probe has changed 6°C or more since zero calibration. It is recommended that zero
calibration be performed because an error has occurred due to shift of the zero point.
3: 1 and 2 have occurred simultaneously.
4: The brightness of the measurement target is lower than the measurable range of the instrument.
5: 2 and 3 have occurred simultaneously.
6: 2 and 4 have occurred simultaneously.
7: 1, 2, and 4 have occurred simultaneously.
CA-S40
Guide
Operation
(14) Trend graph (displayed when [List+Trend Graph] or [Trend Graph] is selected for data display)
Measured values are plotted on the graph in chronological order.
• Right-click on the graph to display the following menu, which allows you to copy the graph to the clipboard or a file.
Right-click to Clipboard Background Color: Black
Background Color: White
to File Background Color: Black Image Size: 50%
Image Size: 100%
Image Size: 150%
Image Size: 200%
Background Color: White Image Size: 50%
Image Size: 100%
Image Size: 150%
Image Size: 200%
(15) Color display window settings
Displays the RGB-specified color window
(16) Data save and clear
Click [Save to CSV] to save the data to a file. Click [Clear] to clear data. The data is saved in CSV format.
47
Page 50
Measurement
1-2. Measurement type
1) Select the measurement type.
Single: Single measurement
Continuous: Continuous measurement (graph scrolls)
* Displays up to 100 data sets and deletes data from the oldest.
Interval:
Package:
Interval measurement, which performs the specified number of measurements at the specified interval.
Package measurement minimizes the communications time between PC and data processor and maximizes
the measurement function when data processor is used. The times of measurements can be set up to 20.
Interval measurementPackage measurement
Time passageTime passage
PC
(Commercially available)
Communication
Communication
Communication
Communication
Communication
Communication
Communication
Communication
・
・
・
・
・
・
・
・
・
・
(Commercially available)
PC
Communication
Communication
Communication
Communication
Communication
Communication
・
・
・
・
・
Communication
Communication
2) Click to start measurement.
* When using continuous measurement or interval measurement, the button display changes to “Hold”, which you
click to pause or stop the measurement.
1-3. Data display
The most recent measurement results are displayed in the [Current Data] field and added to the data list and trend graph.
Use the tab in (9) to select the color space to be used for display.
The trend graph is also displayed in the selected color space.
Click the pop-up mark as shown on the left to display an enlarged view of “Current Data” in a pop-up
window.
1-4. Color display window
You can display the color that you defined using the RGB levels in a pop-up window to be measured using the instrument.
You can also change the RGB settings while having this window open.
RGB-specific settingsSet R, G, and B at the same timeSample color display
Show window
* Display example
48
Page 51
2. Common operations
1) On each measurement screen, you can toggle between “Data List” display and “Trend Graph” display.
2) Click to start the measurement with the settings applied.
3) Click [Save to CSV] to save the data to a file. Click [Clear] to clear data.
* The data is saved in CSV format.
4) “Right-click” functions
Right-click on the data list or graph to display a pop-up menu from which you can copy the data or graph to the
clipboard.
<Example>
You can also specify the background color (B/W) and size for the graph that you want to save.
3. Individual functions
CA-S40
Zero
calibration
SettingsData importUser
calibration
Gamma
measurement
Zero calibration [0-CAL]
• When you click [0-CAL], the shutter inside the probe automatically closes, and zero calibration starts.
• It is recommended that zero calibration be used after start-up or whenever there is a change in ambient temperature or a
temperature change inside the instrument due to using it for long periods.
Settings [Settings]
• The pop-up window for setting the measurement conditions appears.
• You can set the measurement speed and the calibrated channel.
Data import [Read Data]
• This function imports the data from the data processor.
User calibration [User Cal] (only for chromaticity measurement)
• The pop-up window for user calibration appears. (See User calibration on P.60)
• The calibration channel (CH00 to CH99) data is read before displaying the pop-up window.
Gamma measurement [Gamma] (only for chromaticity measurement)
• The pop-up window for gamma measurement appears. (See 3-1. Gamma measurement (only for chromaticity
measurement) on P.50.)
Guide
Operation
49
Page 52
Measurement
3-1. Gamma measurement (only for chromaticity measurement)
• Click [Gamma] among the individual functions to bring up the pop-up window for gamma measurement. Here, you can
measure the color output intensity for each color’s input tone.
• When measuring, configure the settings related to the color and tone to be measured.
• Click the measurement start button. The measurement results are displayed in the graph as gamma curve and chromatic
dierence curve.
• The obtained chromaticity data is displayed on the chromaticity measurement screen and can be saved.
Select color to
Gamma
curve
be measured
Chromatic
dierence
curve
Tone settingWait time after switching tone
Select color to
be displayed and
the color
dierence
Measurement
start button
Right-click on the gamma curve or chromatic dierence curve (chromatic dierence for each tone when the 255 gradation
value is set to 0) graph to display the following menu, which allows you to copy the graph and data to the clipboard or a file.
Right-click on a graph to Clipboard Plot data (for Excel)
Background Color: Black
Background Color: White
to File Plot data (for Excel)
Background Color: Black Image Size: 50%
Image Size: 100%
Image Size: 150%
Image Size: 200%
Background Color: White Image Size: 50%
Image Size: 100%
Image Size: 150%
Image Size: 200%
50
Page 53
4. Flicker measurement (JEITA)
Click the [JEITA] tab to bring up the following screen, where you can set the JEITA frequency resolution. Click [Measure] to
start measurement. The JEITA flicker measurement values, frequency characteristics, and waveform are displayed.
Measured
value (dB)
JEITA frequency resolution
Frequency
characteristics
Waveform
Right-click on the graph to display the following menu, which allows you to copy the graph and data to the clipboard or a
file. You can also change the scale on the Waveform graph.
The right-click functions accessible by right-clicking over the list or trend graph are the same as those for chromaticity
measurement.
Power Spectrum graph
Right-click to Clipboard Background Color: Black
Background Color: White
to File Background Color: Black Image Size: 50%
Image Size: 100%
Image Size: 150%
Image Size: 200%
Background Color: White Image Size: 50%
Image Size: 100%
Image Size: 150%
Image Size: 200%
Waveform graph
Right-click Change Scale
to Clipboard to Excel
Background Color: Black
Background Color: White
to File CSV
Background Color: Black Image Size: 50%
Image Size: 100%
Image Size: 150%
Image Size: 200%
Background Color: White Image Size: 50%
Image Size: 100%
Image Size: 150%
Image Size: 200%
CA-S40
Guide
Operation
51
Page 54
Measurement
Aliasing Noise
During JEITA flicker measurement, when measuring a light source that includes a harmonic component such as a
rectangular wave light source, aliasing noise may occur in the low-frequency range, which can be displayed as the JEITA
flicker value. Peaks that vary greatly on the Power Spectrum graph when changing the JEITA frequency resolution setting by
one step likely are being influenced by aliasing noise.
Set the sampling frequency for JEITA flicker measurement as follows (typical examples).
JEITA frequency resolution
Flicker (CA-310 Mode)*XYZ (Wide Frequency Mode)*
0.1 Hz204.8 Hz819.2 Hz
0.5 Hz1024 Hz1024 Hz
1 Hz1024 Hz1024 Hz
* “Flicker (CA-310 Mode)” and “XYZ (Wide Frequency Mode)” are mode names for PC Software CA-S40. “XYZ (Wide
Frequency Mode)” can only be used when no CA-DP40 data processor is connected.
Sampling frequency
5. Flicker measurement (FMA)
Click the [FMA] tab to bring up the following screen and click [Measure] to start measurement. The FMA flicker
measurement values are displayed.
When performing flicker measurement (FMA), make sure the measurement mode is set correctly (refer to page 54).
To measure waveforms with a vertical synchronization frequency that exceeds 240 Hz (flicker frequency exceeds 120 Hz),
set the INT setting frequency of measurement mode to 240 Hz.
Measured
value (%)
Right-click on the list or graph to copy the graph and data to the clipboard or a file.
The right-click menu is the same as that for chromaticity measurement.
52
Page 55
6. Waveform measurement
Click the [Waveform] tab to bring up the following screen where you can set various parameters. Click [Measure] to start
measurement. The flicker measurement values and other measurement results will be displayed in a waveform graph and
power spectrum graph.
(Also refer to “1. Measurement condition settings <Setting> tab”)
Graph display settings
CA-S40
Guide
Operation
Measurement result
Set Waveform sampling conditions according to Sync Mode
: When checked, sampling will be performed according to the input Sync Mode (refer to
page54). As such, configuration of Sampling Number or Decimation Interval will not be
possible.
Sampling Number : Sets the number of measurements to perform.
Decimation Interval : Sets the number of intervals for measurement.
The measurement time is defined as follows.
Measurement time = Flicker measurement speed × Sampling number × Decimation interval
The possible range of calculating the flicker value diers depending on the measurement time while adhering to the
following.
(1 / measurement time) to 200 Hz
Weighting Factor : Sets a filter for the measured value. (None/JEITA/ LowPass/User)
• No filter will be applied if “None” is selected. The graph will display original data only.
• When “JEITA” is selected, a filter adjusted to the frequency response characteristics of the human eye is used.
• When “LowPass” is selected, the user can set the LowPass filter to between 0 to 1500.
• When “User” is selected, the user can load a custom filter to be used for measurement from the [Weighting Factor] tab
in the [Settings] tab.
The data will be displayed as weighted data with the filter applied. Users can select to display the graph using the check
box in the graph display settings.
Sensor : Sets the sensor when “XYZ (Wide Frequency Mode)” is selected for the flicker measurement
operation mode. (X/Y(Lv)/Z)
When “Flicker (CA-310 Mode)” is selected, only sensor Y can be used.
Measurement results such as flicker measurement values and Power Spectrum graph data are subject to a filter and
considered Weighted Data.
53
Page 56
Measurement condition settings
<Settings> window
Click [Settings] among the individual functions to bring up the following pop-up window for setting the measurement
conditions and optional items.
Sync Mode : Select the mode for synchronized measurement. (NTSC/PAL/UNIV/INT/EXT/MANUAL)
Measurement
synchronization
mode
NTSCMode for measuring NTSC displays33.3 ms59.94 HzNot required
PAL
EXT
UNIV
INT
MANUAL
Mode for measuring PAL and SECAM
displays
Mode for measuring while
synchronizing with the display’s vertical
synchronization signal input
Mode used when the display’s vertical
scanning frequency is unknown or
cannot be input; a mode that can
measure any display
When the display’s vertical
synchronization signal is known, set
that information so that it can be used
for measurement.
Mode that allows you to specify the
integration time
Description of mode
Measurement time
(Sampling time)
40.0 ms50 HzNot required
(1 vertical scanning period)
x 2
100 ms—Not required
(1 vertical scanning period)
x 2
Settable range
4.0 to 4000.0 ms
(For FMA, 7.7 to 4000.0 ms)
Vertical scanning frequency
0.50 to 240.00 Hz
(For FMA, 0.50 to 130.00Hz)
Settable range
0.50 to 240.00 Hz
(For FMA, 0.50 to 130.00Hz)
Use the [SYNC Assist] tab to
search for the optimum
frequency.
—Not required
Vertical
synchronization
signal input
Required
Not required
54
Page 57
Measurement Speed : Select the measurement speed. (SLOW/FAST/LTD.AUTO/AUTO)
• Selecting FAST can shorten measurement time, but may compromise measurement repeatability for low-luminance
measurements.
• Selecting SLOW repeats FAST measurements several times to ensure measurement with high repeatability.
• When AUTO is selected, measurements are automatically selected between SLOW, FAST, and setting with a further
longer integration time depending on the luminance level.
• When LTD.AUTO is selected, measurements are automatically selected between SLOW and FAST depending on the
luminance level.
Notes: For the first measurement or when the brightness of the measurement target changes from the that in the
previous measurement, the switching of ranges may result in longer measurement time.
Averaging Times :
Flicker Output Sensor : Select the operation mode for flicker measurement.
• “Flicker (CA-310 Mode)” is the same operation mode as flicker measurement (JEITA screen, Waveform screen) in
conventional CA-310 and CA-S40 (up to Ver. 1.2) products. In this mode, the measurement range for the measurement
target (flicker frequency) is 0.25 to 65 Hz, and the flicker measurement speed is 33.333 s. This mode is selectable
only when probes other than CA-VP427 or CA-VP410 are connected.
•
In “XYZ (Wide Frequency Mode)” operation mode, the measurement range for the measurement target (flicker frequency)
is 0.25 to 200 Hz. The flicker measurement speed is 333.333 s. This mode can be selected when any probe is connected.
Set the number of measurements taken to average for obtaining a single measurement value. (1 to 20)
(Flicker (CA-310 Mode) / XYZ (Wide Frequency Mode))
CH ID : Select a calibrated channel. Measurement values are corrected using the calibration
coecient of the selected calibration channel. If you do not need such correction, select CH00.
The target saved to the selected calibration channel is displayed.
1-3. JEITA frequency characteristics settings [Weighting Factor tab]
CA-S40
Guide
Operation
Weighting Factor : Specify the frequency characteristics for JEITA flicker measurements.
• When Default (JEITA) is selected, a CA-310-compatible filter compliant with JEITA standards and adjusted to the
human eye is used.
•
When “User” is selected, the filter created by the user is used. In the text file (.txt) for the filter, include the starting frequency, end
frequency, frequency pitch, number of data sets, and the respective gains in tab-delimited format. Click [...] to navigate to the
reference folder, and load the created filter file. Click the [Load] button to draw the graph. This user-created filter will be applied to
flicker measurement (JEITA) results and Waveform measurement results when “User” is selected under Weighting Factor.
55
Page 58
Measurement condition settings
Text file content:Example of text file content:
<Starting frequency> <End frequency> <Frequency pitch> <Number of
data sets>
<Frequency> <Value>
· ·
· ·
· ·
· ·
· ·
Click [OK] to confirm the settings or click [Cancel] to cancel the edit and close the pop-up.
0.01 65 0.01 6500
0.01 1.0
0.02 1.0
· ·
· ·
· ·
65 0.0
1-4. Frequency search [SYNC Assist tab – Settings]
If “INT” is selected as the sync mode, and if the user has some idea of vertical synchronization frequency of the display
(measurement target), using the following procedure to set the internal synchronization frequency correctly up to the
second decimal place makes it possible to perform measurement with good repeatability.
(1) Set the number of measurements to perform for Sampling Number.
(2) Set the known vertical synchronization frequency for the display (measurement target) for Frequency, and click
[Measure]. During measurement, the standard deviation 2σLv is calculated from the measured values obtained from
the number of measurements for Sampling Number, and the results are plotted on the graph. The most recent 2σLv
value is also displayed as the Current Value.
It is possible to change Frequency during measurement.
(3) While changing Frequency, watch the graph to determine which frequency shows the smallest 2σLv value.
(4) When the optimum frequency has been found, stop measurement and return to the [Mode] tab. The obtained vertical
synchronization frequency will be displayed in the [INT] column of [Sync Mode].
If [Measurement Speed] in the [Mode] tab has been changed, it will be necessary to obtain the optimum frequency
again in the [SYNC Assist] tab after confirming the setting by clicking [OK].
56
Page 59
2. Option settings <Info> tab
Display settings
Instrument
information
CA-S40
2-1. Selecting the date display format and luminance unit
Application Settings
Date Format : Select a date display format.
Luminance Unit : Select a luminance unit.
If the selection of [Luminance Unit] is not displayed on the screen, exit the software, and restart the software after adding
“/u” as the startup option for the software execution file “CA-S40.”
2-2. Instrument selection CA Info.
This screen appears first when the software is started with multiple instruments connected to the PC.
Information about the selected instrument is displayed.
Device:
Among the instruments connected to the PC, select the one to be used in the drop-down menu.
Information about the selected instrument (firmware version and serial number) is displayed.
If a data processor is selected as the instrument to be used, then also select the probe connected to the data processor in the
[Probe] field on the right.
Information about the selected probe (firmware version and serial number) is displayed.
Type : Displays the type of instrument connected.
Version : Displays the firmware version of the selected instrument.
Serial No. : Displays the serial number of the selected instrument.
Guide
Operation
Click [OK] to confirm the settings or click [Cancel] to cancel the edit and close the pop-up.
57
Page 60
Measurement condition settings
2-3. Switching between data processor and probe
How to switch from "Type" on the Settings-Info screen
(1) Select the data processor in "CA Info" - "Device" - "Type".
(2) Click the [Probe] button on the "CA Info". (Select the device connected to the data processor.)
(3) Select an area that is not any probe on the Select Probe screen and click [OK] to switch to the data processor.
Select the probe you want to connect to on the Select Probe screen and click [OK] to switch to the probe.
Information about the probe connected
to the data processor
Select device
Select the probe connected to the data
processor
How to switch on the status bar
(1) Select the measuring device switching button on the status bar.
(2) Select an area that is not any probe on the Select Probe screen and click [OK] to switch to the data processor.
Select the probe you want to connect to on the Select Probe screen and click [OK] to switch to the probe.
Measuring device switching button
58
Page 61
3. Command communication <Command> tab
The user can check the operations of communication commands when developing original software.
The specifications of communication commands are available for download from the following webpage:
https://www.konicaminolta.com/instruments/download/
CA-S40
Command : Enter a command to be sent and click [Send] to send it to the instrument.
Return : Displays the command reception result from the instrument.
* This function is used for checking command communication operation. Restart CA-S40 after executing command
communication to update the CA-S40 settings to reflect the changes from command communication.
Guide
Operation
59
Page 62
User calibration
1. About user calibration <User Calib> tab
• User calibration allows users to set their own correction coecients to the calibration channel of the probe by measuring
any colors and setting calibration values to the probe. After setting, the values corrected with the specified correction
coecients can be displayed and output for every measurement.
• As user calibration, you can select single-point calibration or RGB+W matrix calibration.
• User calibration can be performed for every calibration channel (except for CH00).
• User calibration cannot be performed for CH00 because it is calibrated in accordance with the Konica Minolta calibration
standard.
• You can use user calibration to apply the following corrections to measurements based on the specified correction
coecients.
• Correct the dierence in the reading due to the inconsistencies in the spectral response of the light-receiving element
of the probe.
• Correct the dierence in the readings between probes used simultaneously.
2. User calibration procedure
• Click the [User Cal] button on the chromaticity measurement screen to open the pop-up window for user calibration.
• The calibration channel (CH00 to CH99) data is read from the probe before displaying the pop-up window.
• As calibration channel data, the ID, target probe, calibration probe, calibration mode, Lv,x,y, calibration date/time, and
comments are displayed.
• A user calibration coecient is written to the specified calibration channel. If a coecient is already written in that
calibration channel, it is overwritten.
• The target value used is written to a calibration channel as its target. If target data is already written in that calibration
channel, it is overwritten.
• Performing zero calibration by clicking the [0-CAL] button is recommended before running a calibration.
• Data written to the calibration channel can be saved and read as a file for each channel.
(1) Select the [Calibration] tab.
(2) Select the calibration channel to be calibrated. (CH01 to CH99)
• Click over the calibration channel data.
(3) As calibration mode, select [RGB+W Matrix]. Repeat the following steps, (4) through (6), for Red, Green, Blue, and
White.
• You can start from any color (Red, Green, Blue, or White) when performing steps (4) to (6). As an example, the
following explains how to measure Red and enter a target.
60
Page 63
(4) To measure Red, click the measurement button on the right of Red in the [Measured Value] field, which shows the color
display window for measuring Red.
(5) After you align the color display window and the instrument and click [OK] on the pop-up screen, the measured value
is displayed in the [Measured Value] field for Red. After measurement is completed, click the [Close] button in the color
display window to close the window.
CA-S40
(6) Enter the target values you want to correct and display in the [Calibration Value] field.
Guide
Operation
61
Page 64
User calibration
(7) After performing measurement and entering targets for all of Red, Green, Blue, and White, click the [Apply] button to
complete RGB+W matrix calibration.
• To stop the calibration, click [Close]. The change is not applied until you click the [Apply] button.
62
Page 65
2-2. Single-point calibration
(1) Select the [Calibration] tab.
(2) Select the calibration channel to be calibrated. (CH01 to CH99)
• Click over the calibration channel data.
(3) As calibration mode, select [Single-Point].
(4) Click the measurement button on the right of White in the [Measured Value] field to measure and obtain the value.
(5) In the [Calibration Value] field, enter the target values that you want corrected for display.
(6) Click [Apply] to complete the calibration.
• To stop the calibration, click [Close]. The change is not applied until you click [Apply].
2-3. ID setup
You can add an ID and comment to a calibration channel to make it easier to distinguish.
(1) Select the [Set ID] tab.
(2) Select a calibration channel. (CH00 to CH99)
• Click over the calibration channel data.
(3) Enter an ID in the ID text box. You can enter a comment in the Comment text box.
• You can enter up to 10 alphanumeric characters for an ID and up to 50 alphanumeric characters as a comment.
(4) Click [Apply]. The change is not applied until you click [Apply].
(5) The entered ID and comment are applied.
Correction coecients are used for all measurement modes (Lv xy, Lv u' v', Lv Tcp duv, XYZ, and Lvλd Pe).
You cannot perform user calibration for channel CH00.
(CH00 is a channel for performing measurement in accordance with the Konica Minolta calibration standard.)
Performing RGB+W matrix calibration for a preset channel erases the last-set value.
Performing RGB+W matrix calibration for a single-point-calibrated channel erases the correction coecients for the last
single-point calibration and sets correction coecients for RGB+W matrix calibration.
The target value used is written to a calibration channel as its target. If target data is already written in that calibration
channel, it is overwritten.
Guide
Operation
63
Page 66
User calibration
3. Specifying the calibration coecients <Settings> window
To perform measurement using your own correction coecients, specify which calibration channel's correction coecients
you want to use. Specify [CH00] to perform measurement without performing correction but in accordance with the Konica
Minolta standard.
3-1. Selecting a calibrated channel
(1) Click [Settings] to bring up the following pop-up window for setting the measurement conditions and optional items.
(2) In the [CH ID:] field under [Memory Info.] in the [Setting] tab, select a calibrated calibration channel.
The target saved to the calibration channel is displayed in the pane below for your review.
(3) Click [OK] to confirm the calibration channel for which the correction coecients are to be used. The pop-up closes.
Click [Cancel] to close the pop-up without confirming the calibration channel.
64
Page 67
Target settings
The target values used for user calibration are written as the target in a calibration channel. If you want to set a separate
target for color dierence assessment (this function is available with a data processor), you can set a target in a calibration
channel using [Set Target]. While any target data already written in that calibration channel is overwritten, the set correction
coecients remain the same.
If you do not want to lose the target values used for user calibration, it is recommended that you set them in a channel
dierent from the one you used for user calibration.
• Click the [User Cal] button on the chromaticity measurement screen to open the pop-up window for user calibration.
• The calibration channel (CH00 to CH99) data is read before displaying the pop-up window.
• As calibration channel data, the ID, target probe, calibration probe, calibration mode, Lv,x,y, calibration date/time, and
(1) Select the [Set Target] tab.
(2) Select the calibration channel in which the target is set. (CH01 to CH99)
• Click over the calibration channel data or select from the channel selection drop-down menu.
(3) Enter the target values in the [Measured Value] field or obtain the value by clicking the measurement button on the
right side. The values can be rewritten after measurement. Performing zero calibration by clicking the [0-CAL] button
is recommended before running a measurement.
(4) Click [Apply] to complete setting the target. The change is not applied until you click [Apply].
(2) Setting the ID
To set the ID or enter a comment, follow the same procedure as for setting the ID and entering a comment for user
calibration. See 2-3 on P.63 for instructions on how to set the ID. (ID is set for a calibration channel.)
65
Page 68
Data processor data
1. Handling of stored data
The stored data are the measurement results stored in the data processor.
You can measure a single data using up to 10 probes and 100 Data can be registered.
(1) When Data is selected in the Data list on the [Measurement Data] tab
• When you click the [File Save] button, all the measurement results of the probe used for measurement in the
selected Data are saved.
• Click [Data Delete] to erase all measurement results of the selected Data.
• Click [All Delete] to erase all Data.
(2) When Probe is selected in the Data list on the [Measurement Data] tab
• When you click the [File Save] button, the measurement results of all Data measured with the selected probe are
saved.
• Click [Data Delete] to erase Data on the selected probe.
• When you click [All Delete], all data will be deleted on the screen and in the data processor.
Stored dataLog dataSave fileDelete DataDelete all DataSelect Data or Probe
66
Page 69
2. Handling of Log data
One Log Data has the Data of the number of probes for the measured times.
Log No. / date and time measurement started / date and time measurement stopped / data number tabs are displayed on
the Log Data screen.
• Select the Log No. you want to save and click the [File Save] button to save the measurement result of the selected Log
No.
• When you click [All Delete], all data will be deleted on the screen and in the data processor.
Information ..................................................................................................................... 76
CA-S40Appendix
69
Page 72
Error message
The operation is not correct. Follow the instructions displayed immediately.
MessageDetails
--failed to write or read on memory
--reconnect the probe
--System error
--restart and application and a device
--SDK has problem
--restart SDK
Unknown error.An unknown error code has been returned from the SDK. Due to the
Cannot calibrate selected channel.The selected channel cannot be calibrated. If the channel in question is not
Cannot set target on
selected channel.
A memory error has occurred inside the probe. If this condition persists after
reconnecting, contact a service center to deal with the possible malfunction.
A program operation error has occurred inside the probe or SDK. If this
condition persists after restarting, contact a service center to deal with the
possible malfunction.
A program operation error has occurred inside the SDK. If this condition
persists after restarting, contact a service center to deal with the possible
malfunction.
possibility of having an incompatible version, re-installation is
recommended.
CH00, that channel may be locked. Contact a service center.
Target cannot be set for the selected channel. The channel may be locked.
Contact a service center.
70
Page 73
Caution
The setting or operation is not correct.
MessageDetails
Cannot open the manual
Make sure that a PDF reader is installed and
it is set as the default app for PDFs.
Cannot find the manual
Please check your app installation.
Failed to connect to selected probe.
Please check probe connection.
CA device not connected.
Please connect device.
[Max] is invalid.
Please set it to [Min] - 99999.0.
[Min] is invalid.
Please set it to -5 - [Max].
[Max] and [Min] cannot be the same.The maximum and minimum values may not be the same.
Failed to re-connect to device.
Please disconnect and connect the device
again.
Failed to connect to device.
Please check device connection.
[WHITE/RED/GREEN/BLUE Lv] has input
value error.
Please input 0.0001 to 99999.0
[WHITE/RED/GREEN/BLUE x] has input
value error.
Please input 0.0001 to 0.9999.
[WHITE/RED/GREEN/BLUE y] has input
value error.
Please input 0.0001 to 0.9999.
[WHITE/RED/GREEN/BLUE x] or [WHITE/
RED/GREEN/BLUE y] has input value error.
Make sure x+y <= 1.
Invalid memory channel ID.
ID is too long (Max of 10 characters only).
Invalid memory channel comment.
Comment is too long (Max of 50 characters
only).
Loaded file is invalid.
Please select .dat file.
PDF viewer software for displaying the instruction manual cannot be opened.
Check whether PDF viewer software is installed and set to default.
The instruction manual is not found. The manual may have been moved from
the correct folder path. Re-install CA-S40.
This message appears when connection to a probe fails while a data
processor is connected. Check the connection with the probe, and establish
the connection again after restarting CA-S40.
This message appears when CA-S40 is started without connecting a probe/
data processor. Restart CA-S40 after checking the connection.
This is an error in the manual setting of the vertical axis scale in the Waveform
graph on the JEITA screen.
The maximum value must be greater than the minimum value, but less than
99999.0.
This is an error in the manual setting of the vertical axis scale in the Waveform
graph on the JEITA screen.
The minimum value must be smaller than the maximum value, but no lower
than -5.0.
Reconnection to the device failed. The device must be reconnected.
Reconnection of the device failed. Check the device connection status.
Measurement failed.
Follow the instructions in the error message.
Reconnection of the device failed. Check the device connection status.
CH00 cannot be calibrated. Specify a dierent channel for calibration.
True value input of Lv for one of the RGBW colors is out of range.
Input a value between 0.0001 and 99999.0.
True value input of x for one of the RGBW colors is out of range.
Input a value between 0.0001 and 0.9999.
True value input of y for one of the RGBW colors is out of range.
Input a value between 0.0001 and 0.9999.
True value input of x or y for one of the RGBW colors is out of range.
Check whether x + y is 1 or less.
Memory ID input character count exceeds character limit.
Comment input character count exceeds character limit.
Loaded calibration data file has incorrect data format.
CA-S40Appendix
71
Page 74
Caution
MessageDetails
[Times] has input value error.
Please input a valid number.
[Times] has input value error.
Please input 0 - 40000 measurements.
[sec] has input value error.
Please input a valid number.
[sec] has input value error.
Please input 0 - 7259 sec.
[sec] has input value error.
Please input 1 - 7259 sec.
[Wait] has input value error.
Please input a valid number.
[Wait] has input value error.
Please input 0 - 36000 x 100 msec.
INT Sync Mode setting is invalid.
Please go to Settings and set INT
value to 0.50 - 130.00 Hz.
MANUAL Sync Mode setting is invalid.
Please go to Settings and set MANUAL
value to 7.7 - 4000.0 msec.
[INT] has input value error.
Please input a valid number.
[INT] has input value error.
Please input 0.50 - 240.00 Hz.
[INT] has input value error.
Please input 0.50 - 130.00 Hz.
[MANUAL] has input value error.
Please input a valid number.
[MANUAL] has input value error.
Please input 4.0 - 4000.0 msec.
[MANUAL] has input value error.
Please input 7.7 - 4000.0 msec.
[Averaging Times] has input value error.
Please input a valid number.
[Averaging Times] has input value error.
Please input 1 - 255.
File has incorrect format.
Please input a valid file.
Cannot get the saved weighting factor
data.
Filepath does not exist.
Please input a valid path.
User-defined weighting factor data was not
loaded. The previous setting will be
retained.
Do you want to proceed?
The weighting factor setting will be
reverted to the previous setting
Do you want to proceed?
A non-numeric value is entered for the number of measurements for interval
measurement. Use only numeric values.
An out-of-range value is entered for the number of measurements for interval
measurement. Use a value within the specified range.
A non-numeric value is entered for the measurement interval time for interval
measurement. Use only numeric values.
The measurement time interval input is out of range when the interval count
is specified within 1 to 4000. Enter a value within 0 to 7259 seconds.
The measurement time interval input is out of range when the interval count
is specified within 4001 to 40000. Enter a value within 1 to 7259 seconds.
A non-numeric value is entered for the inter-tone wait time for gamma
measurement. Use only numeric values.
An out-of-range value is entered for the inter-tone wait time for gamma
measurement. Enter a value within 0 to 36000.
Sync Mode INT out of range is entered for FMA measurement. Enter a value
within 0.50 Hz to 130.00 Hz.
Sync Mode INT out of range is entered for FMA measurement. Enter a value
within 7.7 to 4000.0 msec.
A non-numeric value is entered for Sync Mode INT. Use only numeric values.
An out-of-range value is entered for Sync Mode INT for colorimetry. Use a
value within the specified range.
An out-of-range value is entered for Sync Mode INT for FMA measurement.
Use a value within the specified range.
A non-numeric value is entered for Sync Mode MANUAL. Use only numeric
values.
An out-of-range value is entered for Sync Mode MANUAL for colorimetry. Use
a value within the specified range.
An out-of-range value is entered for Sync Mode MANUAL for FMA
measurement. Use a value within the specified range.
A non-numeric value is entered for averaging times. Use only numeric values.
An out-of-range value is entered for the number of averaging measurements.
Use a value within the specified range.
Weighting factor data for JEITA measurement could not be read. Check the
file format.
Weighting coecient data for flicker that has been read cannot be acquired.
Load the weighting coecient data file again.
File path is incorrect. Specify the correct path.
File cannot be loaded because of an error in the file format. Use the current
settings or try applying the settings again.
Settings reverted to the original setting.
72
Page 75
MessageDetails
Cannot overwrite existing file.
Check that it is not in use.
--no sync_signal
--input an external synchronization signal
--data value is too high
--set a color within CA's measuring range
Data cannot be saved to file due to reasons such as the save file is currently in
used. Check the file.
Vertical synchronization signals are not input in external synchronization
mode. Check the signal transmitter and the connection status.
The brightness of the measurement target exceeds the upper limit. Either
adjust the brightness so that it is within the measurable range or use a highluminance probe.
--failed to calcurate color temp
Retry the measurement.
--Retry the measurement
--FMA Flicker Value is out of rangeRetry the measurement.
CA-S40Appendix
73
Page 76
Instruction
The flow of operation is displayed. Perform the operation according to the information displayed.
MessageDetails
Please wait for a moment.A task is running. Please wait until the task is completed.
Transferring to Datalist…Continuous/Interval measurement data is being transferred from the buer
to the list. Please wait until the task is completed.
Please perform 0-CAL.This message appears when zero calibration is required. Perform zero
calibration according to the information displayed.
Continuous measurement in progress.
Please stop measurement first.
Interval measurement in progress.
Please stop measurement first.
Closing the application will clear all
measurement data.
Do you want to proceed?
Device was changed.
Do you want to apply previous
probe settings to the new device?
Start 0-CAL?This message appears when the zero calibration button is pressed. Press [No]
Please input a valid number.This message appears when an invalid value (such as characters) is entered
Start [WHITE/RED/GREEN/BLUE]
measurement?
Start Target setting?This message prompts whether or not to set the target.
The work is processing.
Switch to other tab?
Delete CH data from all probes?This message asks for confirmation whether to delete the data in the selected
Delete all data?This message asks for confirmation whether to delete all data in the channels
Overwrite [file name]?This message asks for confirmation whether to overwrite the file for channels
Executing continuous measurement
will clear previous measurement data.
Do you want to proceed?
Executing interval measurement will clear
previous measurement data.
Do you want to proceed?
Executing package measurement will clear
previous measurement data.
Do you want to proceed?
Executing Gamma measurement will clear
previous measurement data.
Do you want to proceed?
Executing measurement will clear gamma
measurement data.
Do you want to proceed?
This message appears when you try to exit the application while Continuous
measurement is running.
Exit the application after stopping the measurement.
This message appears when you try to exit the application while Interval
measurement is running.
Exit the application after stopping the measurement.
This message appears when you try to exit the application when there is still
measurement data. If the measurement data needs to be saved, use the [Save
to CSV] button to save the data before exiting the application.
This message appears when a new device has been connected to the
application. If you want to use the settings for an existing device, click [Yes]. If
you want to use the settings in the new device, click [No].
if you are not running the zero calibration.
for true value input for calibration. Enter a numeric value.
A message is displayed prompting whether or not to perform monochromatic
measurement for calibration. While this message is displayed, you can
change the position of the Color Window for measurement.
Switching to another tab results in the loss of edits and measurement data.
If necessary, finish the tasks before switching tabs.
channel that saves measurement data.
that save measurement data.
that save measurement data.
This message appears when you try to run a continuous measurement while
there is measurement data. If the measurement data is necessary, save it
before running the measurement.
This message appears when you try to run an interval measurement while
there is measurement data. If the measurement data is necessary, save it
before running the measurement.
This message appears when you try to run a package measurement while
there is measurement data. If the measurement data is necessary, save it
before running the measurement.
This message appears when you try to run a gamma measurement while
there is measurement data. If the measurement data is necessary, save it
before running the measurement.
This message appears when you try to run another measurement while there
is gamma measurement data. If the measurement data is necessary, save it
before running the measurement.
74
Page 77
MessageDetails
Entering gamma measurement mode will
clear previous measurement data.
Do you want to proceed?
Closing the window will clear
Gamma measurement data and graphs.
Do you want to proceed?
Changing JEITA sampling condition will
clear previous JEITA measurement data.
Do you want to proceed?
Data are over capacity.
Max capacity = 40000
This message appears when you try to open the gamma measurement
window while there is measurement data. If the measurement data is
necessary, save it before running the measurement.
This message appears when you try to close the gamma window while there
is gamma measurement data. If the measurement data is necessary, save it
before running the measurement.
This message appears when you try to change the JEITA sampling conditions
while there is JEITA measurement data. If the measurement data is necessary,
save it before running the measurement.
This message appears when the number of measurements reaches the
maximum limit.
Save the data first, and then clear it.
Delete data?This message appears when deleting list data. If you need the data, save it
The settings from the previously connected device can be applied to the
newly connected device.
to the new device?
Please specify image formatSpecify the image format.
Cannot save current data in file **Measurement values cannot be saved.
CA-S40Appendix
75
Page 78
Information
Updated information is provided.
MessageDetails
New CA device is connected.This message appears when a dierent device is connected after the USB
connection has been disconnected or when there is a device switchover from
the Settings screen.
New probe is connected.This message appears when a separate probe is selected while a data
processor is connected.
Another PC Software for Color Analyzer
instance is already running.
Re-connection to device in progress…Re-connection to the device is in progress. Wait without unplugging the USB
0-CAL in progress. . .Zero calibration is in progress.
Measurement is cancelled.The measurement has been canceled.
RGB+W Matrix is successfully applied.RGB+W matrix calibration has been successfully completed.
Single-point is successfully applied.Single-point calibration has been successfully completed.
Set Target is successfully applied.Color dierence target has been successfully set.
ID and Comment are successfully applied.ID and comment have been successfully set.
Calibration measurement failed.Measurement for calibration failed.
Target measurement failed.Measurement for setting the target failed.
Read data in progress. . .Data processor data is being read.
Saving data for Log No. [No.]…Log data is being saved.
No data selected.Data has not been selected.
No available data.There is no operable data.
Cannot delete measurement data.Measurement data cannot be deleted.
No data to save.There is no data to be saved.
Gamma measurement cancelled.Gamma measurement has been canceled.
Interval measurement in progress…Interval measurement is in progress.
Package measurement in progress…Package measurement is in progress.
JEITA flicker measurement in progress…JEITA measurement is in progress.
Reading memory…Memory is being read.
Please check the boxes corresponding to
the colors to be measured.
This message indicates a repeated launch attempt. Multiple copies of the
software cannot run simultaneously.
flash drive.
No color has been selected to be measured. Check the check boxes of the
colors to be measured.
List display color, error code: List color displayed on data processor, error code
List display colorError CodeMeaning
White——
Blue1:Dierent from cal
Probe
Blue2:∆t out of rangeThe temperature inside the probe has changed more than the specified
Blue3:∆t out of rangeError code 1 and error code 2 occurred at the same time.
Blue4:Below
measurement range
Blue5:Out of
measurement range
The probe(s) being used are dierent from when user calibration / target
setting and measurement was performed.
amount since zero calibration.
Measurement is out of measurement range.
Error code 1 and error code 4 occurred at the same time.
76
Page 79
List display colorError CodeMeaning
Blue6: ∆ t out of rangeError code 2 and error code 4 occurred at the same time.
Blue7: ∆ t out of rangeError code 1, error code 2, and error code 4 occurred at the same time.
Yellow10:Problem with
There is a problem with CA-410.
CA-410 device
Yellow20:Out of EXT range External synchronization signal is invalid. (External synchronization signal is
set to less than 0.5 Hz or over 240 Hz.)
Yellow50:FMA value over
The measured value in flicker mode exceeds 999.9%.
999.9%
Yellow51:Out of EXT range
for FMA
Synchronization signal is invalid in FMA flicker mode. (The synchronization
signal is set at less than 0.5 Hz or over 130 Hz.)
Yellow53:Not flicker probe CA-VP 410 or CA-VP 427 is connected in flicker mode.
White64:Low BatteryBattery level is low.
Blue65:Low Battery +
Error code 1 and error code 64 occurred at the same time.
other errors
Blue66:Low Battery +
Error code 2 and error code 64 occurred at the same time.
other errors
Blue67:Low Battery +
Error code 1, error code 2, and error code 64 occurred at the same time.
other errors
Blue68:Low Battery +
Error code 4 and error code 64 occurred at the same time.
other errors
Blue69:Low Battery +
Error code 1, error code 4, and error code 64 occurred at the same time.
other errors
CA-S40Appendix
Blue70:Low Battery +
Error code 2, error code 4, and error code 64 occurred at the same time.
other errors
Blue71:Low Battery +
other errors
Error code 1, error code 2, error code 4, and error code 64 occurred at the
same time.
* If correlated color temperature or dominant wavelength cannot be calculated, "----" is displayed as measured value.
* If there is a colorimetric value that has not been measured when reading the data of the data processor, its colorimetric
value will be displayed as "-9999", "- 99999999", etc.
77
Page 80
Information
Error code
For the error code displayed on the List in yellow, the error message shown in the table below is displayed
after measurement is completed.
Error messageMeaning
CA410 device has problem.
Please check configurations or device
conditions
An invalid external synchronization signal
(set an external synchronization signal
when less than 0.5 Hz, over 240 Hz)
Over the measurement range.The tristimulus value of the measurement subject is too large.
Measurement value is over 999.9% in the
flicker mode.
External synchronizing signal is over 130
Hz in the FMA flicker mode.
Flicker measurement is being attempted
using a Universal Measuring Probe CAVP410.
There is a problem with CA-410. Check the setting and CA-410 status.
External synchronization signal is invalid. (External synchronization signal is
set to less than 0.5 Hz or over 240 Hz.)
The measured value exceeds 999.9% in FMA measurement.
Frequency input exceeded 130 Hz in FMA measurement.
CA-VP 410 is connected in flicker mode.
78
Page 81
< CAUTION >
KONICA MINOLTA WILL NOT BE LIABLE FOR ANY DAMAGES RESULTING FROM THE MISUSE, MISHANDLING,
UNAUTHORIZED MODIFICATION, ETC. OF THIS PRODUCT, OR FOR ANY INDIRECT OR INCIDENTAL DAMAGES (INCLUDING
BUT NOT LIMITED TO LOSS OF BUSINESS PROFITS, INTERRUPTION OF BUSINESS, ETC.) DUE TO THE USE OF OR INABILITY
TO USE THIS PRODUCT.