Disinfecting the Sensor and Maintaining Hygiene
4–2 Disinfecting and Maintaining the Sensor
Disinfecting the Sensor and
Maintaining Hygiene
Carefully follow the procedure detailed earlier in this
manual on how to prepare the sensor to ensure
maximum hygienic safety for the patient.
To disinfect the sensor and maintain proper hygiene,
follow these guidelines:
• Always use the sensor with a hygienic barrier.
Change the barrier between each patient.
• When selecting a disinfectant product, check the
list with the product manufacturer's information.
• Thoroughly disinfect the sensor after each patient.
Remove the hygienic protective sheath and
thoroughly clean the sensor with a disinfecting
wipe.
• Use only cold disinfecting products that are
authorized by local dental regulatory agencies.
• Follow the manufacturer's recommendations for
safety precautions when using the disinfectant
product.
• Between each patient, clean the sensor and the
first centimeters of the cable using a disinfecting
cloth. Wipe down the sensor with a sterile
solution. Keep the sensor and the remote control
off of the floor at all times.
Note
Our sensors are
supplied non-sterile.
Caution!
Never place the
sensor or remote
control in an
autoclave. This can
damage the system.
Caution!
Never immerse the
connector located on
the other end of the
cable, nor the remote
control.
Note
Immerse part of the
cable to guarantee a
good disinfection. Do
not immerse the remote
control.