Agilent 89410A Fail radiation emissions requirement

89410A-06
SUPERSEDES: None
DATE: 15 June 1994
89410A Vector Signal Analyzer
Serial Numbers: 0000A00000 / 3346A00437
Analyzers may fail radiated emissions requirement due to missing gaskets
To Be Performed By: Agilent-Qualified Personnel
8160-0360 3 Rope Gasket
Situation:
The 3 rope gaskets which should be installed around the front bezel were found to be missing. The purpose of the gasket is to improve the contact between the front frame cast­ing and the covers. The analyzer may fail to meet the radiated emissions requirement with­out these gaskets.
1 of 2
AUTHOR: CMG
ENTITY: A100
© 1994 AGILENT TECHNOLOGIES PRINTED IN U.S.A.
SERVICE NOTE
ADMINISTRATIVE INFORMATION
MODIFICATION RECOMMENDED
SERVICE NOTE CLASSIFICATION:
ACTION CATEGORY:
LOCATION CATEGORY:
CUSTOMER INSTALLABLE ON-SITE SERVICE CENTER
AGILENT RESPONSIBLE UNTIL: 15 June 1996
ADDITIONAL INFORMATION:
IMMEDIATELY ON SPECIFIED FAILURE AGREEABLE TIME
AVAILABILITY:
PRODUCT’S SUPPORT LIFE
RETURN SCRAP SEE TEXT
SERVICE INVENTORY:
RETURN SCRAP SEE TEXT
USED PARTS:
STANDARDS:
Labor 1.0 Hour
Continued
Page 2 Service Note 89410A-06
Solution/Action:
Install gaskets.
STEP 1. Remove top, side, and bottom covers.
STEP 2. Add rope gaskets around the front frame.
1. Press 1 rope gasket into the bottom of the front frame (MP402).
2. Press 1 rope gasket into the top of the front frame.
3. Cut 1 rope gasket in half.
4. Press 1/2 rope gasket into each side of the front frame.
STEP 3. Replace covers.