Agilent 8594EM Data Sheet

specifications
8590EM
EMC Analyzers
Frequency specifications
Frequency range
HP 8591EM 9 kHz to 1.8 GHz HP 8593EM 9 kHz to 22 GHz
Option 027 9 kHz to 26.5 GHz (Type N) HP 8594EM 9 kHz to 2.9 GHz HP 8595EM 9 kHz to 6.5 GHz HP 8596EM 9 kHz to 12.8 GHz
Frequency reference
Standard Option 004
Aging rate ±2 x 10
–6
/year ±1 x 10–7/year
Temperature stability ±5 x 10
–6
±1 x 10
–8
Initial achievable accuracy±5 x 10
–6
±2.2 x 10
–8
Frequency readout accuracy
(start, stop, center, marker)
±(Frequency readout x frequency reference error + span accuracy + 1% of span + 20% RBW +100 Hz x N*)
* N = the LO harmonic number
Marker counter accuracy
Frequency span
10 MHz x N* ±(Marker frequency x frequency
reference error + counter
resolution + 100 Hz x N*) Frequency span > 10 x N* ±(Marker frequency x frequency
reference error + counter
resolution + 1 kHz x N*)
Frequency sweep time
Span 0 Hz, >1 kHz 20 µs to 100 s Span 0 Hz (Option 101) 20 µs to 100 s
Resolution bandwidths
30 Hz to 3 MHz (1, 3, 10
sequence)
200 Hz, 9 kHz, 120 kHz and
1 MHz (6dB) EMI bandwidths Accuracy ±10 %
Video bandwidth range
1 Hz to 1 MHz in 1, 3, 10
sequence
Comb generator
HP 8593EM, 8596EM 100 MHz fundamental frequency Accuracy ± .007 %
Span accuracy
Span 10 MHz ±2 % of span
Span > 10 MHz ±3 % of span
HP 8591EM, HP 8593EM, HP 8594EM, HP 8595EM, HP 8596EM, HP 84100EM, and HP 84110EM
HP 8590EM Series EMC Analyzers and Precompliance Systems
Amplitude Specifications
Amplitude range
Displayed average noise level to + 137 dBµV
Maximum safe input level
(input attenuator set at 10 dB)
Average continuous power +30 dBm Peak pulse power
HP 8591EM +30 dBm HP 8593EM, 8594EM, +50 dBm <10 µs 1 %
8595EM and 8596EM duty cycle 30 dB
input attenuation
DC
HP 8591EM 25 Vdc HP 8593EM 0 Vdc HP 8594EM, 8595EM 0 Vdc (dc coupled) 8596EM 50 Vdc (ac coupled)
Displayed average noise level
400 kHz to 1 GHz 30 Hz RBW 1 kHz RBW HP 8591EM –23 dBµV –8 dBµV HP 8593EM –20 dBµV –5 dBµV HP 8594EM 400 kHz to 5 MHz –15 dBµV 0 dBµV 5 MHz to 1 GHz –20 dBµV –5 dBµV HP 8595EM –18 dBµV –3 dBµV HP 8596EM –18 dBµV –3 dBµV
Frequency response
9 kHz to 1 GHz HP 8591EM, 8593EM, ±1.5 dB 8594EM, 8595EM and 8596EM
Calibrator output
300 MHz –20 dBm ±.4 dB
Option specifications
Option 010 tracking generator
Frequency range
HP 8591EM 100 kHz to 1.8 GHz HP 8593EM, 8594EM, 9 kHz to 2.9 GHz 8595EM and 8596EM
Output level
HP 8591EM 0 to –70 dBm HP 8593EM, 8594EM, –1 to –66 dBm 8595EM and 8596EM
Output flatness
HP 8591EM ±1.75 dB HP 8593EM, 8594EM, ±2.0 dB 8595EM and 8596EM
Weight
HP 8591EM 14.5 kg (32 lbs) HP 8593EM, 8594EM, 16.4 kg (36 lbs) 8595EM and 8596EM
Input/Output
Front panel connectors
Input 50 type N
Option 026 APC 3.5 mm male
Option 027 50 type N female
Cal output 50 BNC –20 dBm 300 MHz
Comb out 100 MHz ±0.007 %, SMA Probe power +15 Vdc, –12.6 Vdc and Gnd
(150 mA max each)
Rear panel connectors
Earphone 1/8 inch monaural jack Interface HP-IB and parallel Keyboard 5 pin DIN, HP C1405 Option ABA
and Option 003 IBM AT keyboard compatible
High sweep in/out BNC, high TTL = sweep
Aux video out 50 BNC, 0-1 V Aux IF output 50 BNC, –10 tp 60 DBM, 21.4 MHz
Ext. trigger input BNC TTL levels, positive edge trigger
Ext. ref. in 50 BNC, 10 MHz, –2 to + 10 dBm 10 MHz ref. output 50 BNC, 10 MHz, 0 dBm
Aux interface 9 pin “D” subminiature
Monitor out 50 BNC, NTSC, 15.75 kHz,
60 Hz or PAL, 15625 kHz, 50 Hz