Agilent 4155C Technical Data

Agilent 4155C Semiconductor Parameter Analyzer Agilent 4156C Precision Semiconductor Parameter Analyzer
Technical Data

Introduction

Basic Functions
Agilent 4155C and 4156C functions:
Set measurement and/or stress
conditions Control measurement and/or stress
execution Perform arithmetic calculations
 
Display measured and calculated results on the LCD display
Perform graphical analysis
Store and recall measurement setups,
and measurement and graphical display data
Dump to printers or plotters for hard-
copy output
Perform measurement and analysis
with the built-in instrument BASIC
Self test, Auto calibration
Configuration
4155C
4xMPSMU 2xVMU 2x
VSU
41501B
GNDU
(Optional)
4156C
4xHRSMU 2xVMU 2x
VSU
2xPGU(Option)
HPSMU(Option) or 2xMPSMU (Option)
SMU: Source Monitor Unit Display resolution: 6 digits at each current range (0.01fA display resolution at 10pA range) HRSMU: High Resolution SMU (1fA/2µV to 100mA/100V) MPSMU: Medium Power SMU
(
10fA/2µV to 100mA/100V)
HPSMU:High Power SMU
(
10fA/2µV to 100mA/200V)
VMU: Voltage Monitor Unit
(
0.2µV resolution in differential mode)
VSU: Voltage Source Unit PGU: Pulse Generator Unit (1 channel)
GNDU: Ground Unit
*1
*1: Minimum number of installable MPSMU *2: Accuracy not guaranteed. Minimum guaranteed
*1
resolution is 1fA at 10pA range

Hardware

Specification Condition
The “supplemental” information and “typical” entries in the following specifi­cations are not warranted, but provide useful information about the functions and performance of the instruments.
The measurement and output accuracy
are specified at the rear panel connector terminals when referenced to the Zero Check terminal under the following conditions:
1. 23°C ± 5°C (double between 5°C to 18°C, and 28°C to 40°C if not noted otherwise)
2. After 40 minutes warm-up
3. Ambient temperature change less than ± 1°C after auto calibration execution.
4. Integration time: medium or long
5. Filter: ON (for SMUs)
6. Kelvin connection (for HRSMU, HPSMU, and GNDU)
7. Calibration period: 1 year

Agilent 4156C Precision Semiconductor Parameter Analyzer

HRSMU (High Resolution SMU) Specifications
Voltage Range, Resolution, and Accuracy (HRSMU)
Voltage Range
± 20V ± 40V ± 100V
*1: 100mA (Vout 20V), 50mA (20V<Vout≤40V) *2: 100mA (Vout 20V), 50mA (20V<Vout≤40V), 20mA (40V<Vout≤100V)
Set. Reso.
1mV
5mV
Set.
Accuracy
±(0.025%+6mV)2mV
Meas. Reso.
Current Range, Resolution, and Accuracy (HRSMU)
Current Range
±A ±10µA ±100µA ±1mA
*1: The accuracy is applicable when offset cancellation has been performed. *2: The offset current specification is multiplied by one of the following factors depending upon the ambient temperature and humidity (RH = Relative Humidity):
*3: 100V (Iout 20mA) 40V (20mA<Iout50mA) 20V (50mA<Iout100mA)
Vout is the output voltage in volts. Iout is the output current in amps. For example, accuracy specifications are given as ±% of set/measured value (0.04%)
plus offset value (200nA+1nA×Vout) for the 1mA range. The offset value consists of a fixed part determined by the set/measuremet range and a proportional part that is multiplied by Vout or Vout/100.
Set. Reso.
10fA±10pA 10fA±100pA 100fA±1nA
±(0.5%+0.7pA+1fA×Vout)
1pA±10nA 10pA±100nA 100pA 1nA 10nA 100nA
±(0.5%+4pA+10fA×Vout)
±(0.12%+40pA+100fA×Vout)
±(0.12%+400pA+1pA×Vout)
±(0.07%+4nA+10pA×Vout)
±(0.07%+40nA+100pA×Vout)
±(0.06%+400nA+1nA×Vout)
±(0.06%+4µA+10nA×Vout)A±10mA
±(0.12%+40µA+100nA×Vout)10µA±100mΑ
5 - 60Temperature
Set.
Accuracy
RHHumidity % 60 - 80
×2×25°C to 18°C ×2×118°C to 28°C ×5×228°C to 40°C
*1*2
*1*2
Meas. Reso.
1fA±(4%+400fA)
±(4%+20fA+1fA×Vout/100)
1fA±(4%+400fA)
*2
±(4%+40fA+10fA×Vout/100)
10fA 10fA 100fA 1pA 10pA 100pA 1nA 10nA 100nA
±(0.5%+0.4pA+1fA×Vout)
±(0.1%+20pA+100fA×Vout)
±(0.1%+200pA+1pA×Vout) ±(0.05%+2nA+10pA×Vout)
±(0.05%+20nA+100pA×Vout)
±(0.04%+200nA+1nA×Vout)
±(0.04%+2µA+10nA×Vout)
±(0.1%+20µA+100nA×Vout)
Output terminal/connection:
Dual triaxial connectors, Kelvin (remote sensing)
HRSMU Measurement and Output Range
Meas.
Accuracy
±(0.01%+200µV)V±(0.02%+400µV)100µV± 2V
±(0.01%+1mV)20µV±(0.02%+3mV)
±(0.015%+2mV)40µV
±(0.02%+5mV)100µV±(0.03%+15mV)
Meas.
Accuracy
±(0.5%+2pA+10fA×Vout)
-100
Max. Current
100mA 100mA
*1 *2
*1*2
*1*2
*2
Voltage/Current Compliance (Limiting):
The SMU can limit output voltage or current to prevent damaging the device under test.
Voltage: 0V to ±100V Current: ±100fA to ±100mA Compliance Accuracy: Same as the
current (voltage) settling accuracy.
HRSMU Supplemental Informa-
tion :
Maximum allowable cable resistance
Max.
when using Kelvin connection (Force,
V
Sense): 10
100V
Typical voltage source output
100V
resistance (Force line/non-Kelvin
100V
connection): 0.2
100V
Voltage measurement input resistance/
100V
current source output resistance:
100V 100V 100V 100V 100V
*3
100
-20-40 20 4
-20
-50
-100
15
10
(10pA range)
Current compliance setting accuracy for opposite polarity: 10pA to 10nA range: V/I setting accuracy ± 12% of range
100nA to 100mA range: V/I setting accuracy ± 2.5% of range
Current (mA)
50
20
0
100
Voltage (V)
2
VSU (Voltage Source Unit)
Specifications
VSU Output Range:
Meas.
Voltage Range
Reso.
1mV
±20V *1: Specification is applicable under no load current. Max. Output Current: 100mA
±(0.05% of setting+10mV)
VSU Supplemental Information:
Output resistance: 0.2 Maximum load capacitance: 10µF Maximum slew rate: 0.2V/µs Current limit: 120mA (typical)
Output Noise: 1mV rms (typical)
Meas.
Accuracy
(Voltage Monitor
VMU
Specifications
VMU Measurement Range, Resolution, and Accuracy:
Voltage
Range
VMU Differential Mode Range Resolution, and Accuracy:
Diff V Range
Max. Common Mode Voltage: ± 20V Note: Vi is the input voltage of VMU2 in volts.
Meas. Reso.
Meas. Reso.
±(0.02%+200µV)2µV±2V
±(0.02%+1mV)20µV±20V
Meas.
Accuracy
±(0.03% + 10µV+0.3µV×Vi)µV±0.2V 0.2
±(0.02%+100µV+3 µ V × Vi)2µV±2V
Unit)
Meas.
Accuracy
For example, accuracy specifications are given as ±% of set/measured value (0.02%) plus offset value (1mV+13µV×Vi) for the 2V range. The differential mode offset value consists of a fixed part determined by the measurement range and a proportional part that is multiplied by Vi.
VMU Supplemental Information:
Input Impedance: 1G Input leakage current (@0V): 500pA (Typical)
Measurement noise: 0.01% of range (p-p) (Typical) Differential mode measurement noise:
0.005% of range (p-p) (Typical)

Agilent 4155C Semiconductor Parameter Analyzer

MPSMU
(Medium Power SMU) Specifications
Voltage Range, Resolution, and Accuracy (MPSMU)
Voltage Range
± 20V ± 40V ± 100V
*1: 100mA (Vout 20V), 50mA (20V<Vout≤40V) *2: 100mA (Vout 20V), 50mA (20V<Vout≤40V), 20mA (40V<Vout≤100V)
Set. Reso.
1mV 2mV 5mV
Set.
Accuracy
Meas. Reso.
±(0.02%+700µV+0.3×Iout) V±(0.03%+900µV+0.3×Iout)100µV± 2V
±(0.02%+2mV+0.3×Iout)20µV±(0.03%+4mV+0.3×Iout) ±(0.02%+3mV+0.3×Iout)40µV±(0.03%+7mV)+0.3×Iout) ±(0.03%+5mV+0.3×Iout)100µV±(0.04%+15mV)+0.3×Iout)
Meas.
Accuracy
Current Range, Resolution, and Accuracy (MPSMU)
Current Range
±A ±10µA ±100µA ±1mA
*1: 100V (Iout 20mA), 40V (20mA<Iout≤50mA), 20V (50mA<Iout≤100mA)
Set. Reso.
100fA±1nA 1pA±10nA 10pA±100nA 100pA 1nA 10nA 100nA
±(0.5%+3pA+2fA×Vout)
±(0.5%+7pA+20fA×Vout)
±(0.12%+50pA+200fA×Vout)
±(0.12%+400pA+2pA×Vout)
±(0.12%+5nA+20pA×Vout)
±(0.12%+40nA+200pA×Vout)
±(0.12%+500nA+2nA×Vout)
±(0.12%+4µA+20nA×Vout)A±10mA
±(0.12%+50µA+200nA×Vout)10µA±100
Set.
Accuracy
Meas. Reso.
10fA 10fA 100fA 1pA 10pA 100pA 1nA 10nA 100nA
±(0.5%+5pA+20fA×Vout)
±(0.1%+30pA+200fA×Vout)
±(0.1%+200pA+2pA×Vout)
±(0.1%+3nA+20pA×Vout)
±(0.1%+20nA+200pA×Vout)
±(0.1%+300nA+2nA×Vout)
±(0.1%+2µA+20nA×Vout)
±(0.1%+30µA+200nA×Vout)
Meas.
Accuracy
±(0.5%+3pA+2fA×Vout)
Max. Current
100mA 100mA
Max. V
100V 100V 100V 100V 100V 100V 100V 100V
Output terminal/connection:
Single triaxial connector, non-Kelvin (no remote sensing)
Voltage/Current Compliance (Limiting):
The SMU can limit output voltage or current to prevent damaging the device under test. Voltage: 0V to ±100V
Current: ±1pA to ±100mA Compliance Accuracy: Same as the current (voltage) settling accuracy.
MPSMU Supplemental Information:
Typical voltage source output resistance: 0.3 Voltage measurement input resistance/ current source output resistance: 10 Current compliance setting accuracy for opposite polarity: 1nA to 10nA range: V/I setting accuracy ± 12% of range
100nA to 100mA range: V/I setting accuracy ± 2.5% of range
13
(1nA range)
Vout is the output voltage in volts. Iout is the output current in amps. For example, accuracy specifications are given as ±% of set/measured value (0.1%)
plus offset value (30pA+200fA×Vout) for the 100nA range. The offset value consists of a fixed part determined by the set/measuremet range and a proportional part that is multiplied by Vout.
VSU Specifications
Same as 4156C VSU.
VMU
Specifications
Same as 4156C VMU.
MPSMU Measurem ent and Output Range
-100
Current (mA)
100
50
20
-20-40 20 40
-20
-50
-100
100
Voltage (V)
3