No part of this manual may be
reproduced in any form or by any
means (including electronic storage
and retrieval or translation into a
foreign language) without prior
agreement and written consent from
Keysight Technologies, Inc. as
governed by United States and
international copyright laws.
Trademark Acknowledgements
Manual Part Number
N9042-90001
Publication Date
February 2021
Supersedes: December 2020
Published in USA
Keysight Technologies Inc.
1400 Fountaingrove Parkway
Santa Rosa, CA 95403
Warranty
THE MATERIAL CONTAINED IN THIS
DOCUMENT IS PROVIDED “AS IS,”
AND IS SUBJECT TO BEING
CHANGED, WITHOUT NOTICE, IN
FUTURE EDITIONS. FURTHER, TO
THE MAXIMUM EXTENT PERMITTED
BY APPLICABLE LAW, KEYSIGHT
DISCLAIMS ALL WARRANTIES,
EITHER EXPRESS OR IMPLIED WITH
REGARD TO THIS MANUAL AND
ANY INFORMATION CONTAINED
HEREIN, INCLUDING BUT NOT
LIMITED TO THE IMPLIED
WARRANTIES OF
MERCHANTABILITY AND FITNESS
FOR A PARTICULAR PURPOSE.
KEYSIGHT SHALL NOT BE LIABLE
FOR ERRORS OR FOR INCIDENTAL
OR CONSEQUENTIAL DAMAGES IN
CONNECTION WITH THE
FURNISHING, USE, OR
PERFORMANCE OF THIS
DOCUMENT OR ANY INFORMATION
CONTAINED HEREIN. SHOULD
KEYSIGHT AND THE USER HAVE A
SEPARATE WRITTEN AGREEMENT
WITH WARRANTY TERMS
COVERING THE MATERIAL IN THIS
DOCUMENT THAT CONFLICT WITH
THESE TERMS, THE WARRANTY
TERMS IN THE SEPARATE
AGREEMENT WILL CONTROL.
Technology Licenses
The hardware and/or software
described in this document are
furnished under a license and may be
used or copied only in accordance
with the terms of such license.
U.S. Government Rights
The Software is “commercial
computer software,” as defined
by Federal Acquisition Regulation
(“FAR”) 2.101. Pursuant to FAR
12.212 and 27.405-3 and
Department of Defense FAR
Supplement (“DFARS”) 227.7202,
the U.S. government acquires
commercial computer software
under the same terms by which
the software is customarily
provided to the public.
Accordingly, Keysight provides
the Software to U.S. government
customers under its standard
commercial license, which is
embodied in its End User License
Agreement (EULA), a copy of
which can be found at
http://www.keysight.com/find/sweula
The license set forth in the EULA
represents the exclusive authority
by which the U.S. government
may use, modify, distribute, or
disclose the Software. The EULA
and the license set forth therein,
does not require or permit,
among other things, that
Keysight: (1) Furnish technical
information related to
commercial computer software
or commercial computer
software documentation that is
not customarily provided to the
public; or (2) Relinquish to, or
otherwise provide, the
government rights in excess of
these rights customarily provided
to the public to use, modify,
reproduce, release, perform,
display, or disclose commercial
computer software or
commercial computer software
documentation. No additional
government requirements
beyond those set forth in the
EULA shall apply, except to the
extent that those terms, rights, or
licenses are explicitly required
from all providers of commercial
computer software pursuant to
the FAR and the DFARS and are
set forth specifically in writing
elsewhere in the EULA. Keysight
shall be under no obligation to
update, revise or otherwise
modify the Software. With
respect to any technical data as
defined by FAR 2.101, pursuant
to FAR 12.211 and 27.404.2 and
DFARS 227.7102, the U.S.
government acquires no greater
than Limited Rights as defined in
FAR 27.401 or DFAR 227.7103-5
(c), as applicable in any technical
data.
Safety Notices
A CAUTION notice denotes a hazard. It
calls attention to an operating
procedure, practice, or the like that,
if not correctly performed or adhered
to, could result in damage to the
product or loss of important data. Do
not proceed beyond a CAUTION
notice until the indicated conditions
are fully understood and met.
A WARNING notice denotes a hazard.
It calls attention to an operating
procedure, practice, or the like that,
if not correctly performed or adhered
to, could result in personal injury or
death. Do not proceed beyond a
WARNING notice until the indicated
conditions are fully understood and
met.
In This Guide…
This guide contains the following information:
1 Quick-Start
This chapter explains how to initialize the signal analyzer and view a signal.
2 Front and Rear Panel Features
Refer to this chapter for information on front- and rear-panel key functionality, and display annotations.
3 Instrument Operating System
This chapter describes the Microsoft Windows 10 configuration and the settings used with the Keysight
instrument software.
4 Using Windows Tools
The information in this chapter provides some guidelines for using the Microsoft Windows 10 feature capabilities
with the signal analyzer.
5 Troubleshooting
This chapter details some basic steps that may solve any problems you are experiencing with either the signal
analyzer or Microsoft Windows 10.
3
Where to Find the Latest Information
Documentation is updated periodically. For the latest information about these products, including
instrument software upgrades, application information, and product information, browse to one of the
following URLs, according to the name of your product:
http://www.keysight.com/find/N9042B
To receive the latest updates by email, subscribe to Keysight Email Updates at the following URL:
http://www.keysight.com/find/MyKeysight
Information on preventing analyzer damage can be found at:
Periodically, Keysight releases software updates to fix known defects and incorporate product
enhancements. To search for software updates for your product, go to the Keysight Technical Support
website at:
http://www.keysight.com/find/sa_software
4
Contents
1 Quick Start
Initial Inspection 10
Verify the Contents 10
Shipping Problems? 10
Instrument Location and Rack Mounting Requirements 11
Locating the Analyzer 11
Cooling and Rack Mounting 11
Turning on the Analyzer the First Time 12
Initializing the Analyzer 12
Instrument Power Down Process 16
Contents
Recommended Instrument Shut Down 16
Anti-Virus Software and Firewalls 17
Instrument Information 18
Power Requirements 18
Protecting Against Overpowering 19
Environmental and Regulatory Information 20
Instrument Maintenance 22
Protecting against electrostatic discharge 24
2 Front and Rear Panel Features
Front-Panel Features 26
Display Features 29
Invalid Data Indicator 29
Screen Tabs 30
Meas Bar 31
Measurement Display 32
Menu Panel 33
Bottom Bar 35
Rear-Panel Features 36
5
Contents
Front and Rear Panel Symbols 38
3 Instrument Operating System
Microsoft Windows 40
Installed Software 40
Signal Analyzer Software 40
Vector Signal Analyzer Software 40
Customer installation of software 41
3rd Party Software Verified by Keysight 41
Installation of Other 3rd Party Software 41
User Accounts 42
Administrator Account 42
Instrument Accounts 42
Keysight Only User Account 42
Service User Accounts 42
Customer Creation of Accounts 42
Keysight X-Series Analyzer Licensing Options 44
Fixed Licenses 44
Transportable Licenses 44
Network Licenses 45
USB Portable Licenses 45
Configuring Network and USB Licenses 46
Licensing New Measurement Application Software -
After Initial Purchase 47
Installation Procedure Over USB for Fixed Licenses 47
Windows Configuration 49
Settings that Can Be Changed 49
Settings that Must Not Be Changed 50
Autoplay/Autorun 51
Configuring Printers 52
Configuring LAN 53
Hostname 53
6
IP Address & Gateway 53
Connecting To A Network Shared Folder 53
Windows Security 55
Windows Firewall 55
Automatic Updates 55
Virus Protection 56
Spyware Protection 56
System Maintenance 57
Backup 57
System Restore 57
Disk Defragmenting 57
USB Connections 58
Disk Drive Partitioning and Use 60
Contents
Backing-up Factory Calibration Data Using Alignment Data Wizard 60
Disk Drive Recovery Process 64
Updating the Software (required after a recovery) 66
4 Using Windows Tools
Navigating Windows Without a Mouse 70
Remote Desktop: Using the X-Series Signal Analyzers Remotely 71
Overview of Remote Desktop Operation 71
Setting Up Remote Desktop Operation 72
How to Locate the Computer Name of the Instrument 73
Running a Remote Desktop Session 74
Embedded Web Server: Using the X-Series Signal Analyzers Remotely 80
Browsers that support the X-Series Signal Analyzer Embedded Web Server
(these versions or later) 80
Accessing the instrument through the Internet 80
Controlling the Instrument 82
Enabling SCPI Control of the Instrument 83
Selecting the Get Image Tab 83
Selecting the Transfer Data Tab 84
7
Contents
Selecting the Save/Recall Tab 84
Selecting the Docs/Links Tab 85
Windows Shortcuts and Miscellaneous Tasks 86
Windows Shortcuts (key combinations) 86
Windows Taskbar Auto-hide Feature 87
5 Troubleshooting
Check the Basics 90
Problems with Microsoft Windows 10 92
Returning an Analyzer for Service 93
Calling Keysight Technologies 93
Locations for Keysight Technologies 94
Read the Warranty 94
Service Options 95
Packaging the Instrument 96
8
Keysight X-Series Signal Analyzers
N9042B
Getting Started and Troubleshooting Guide
1Quick Start
This section explains how to initialize the Signal Analyzer and view a signal.
The following topics can be found in this section:
“Initial Inspection” on page 10
“Instrument Location and Rack Mounting Requirements” on page 11
“Turning on the Analyzer the First Time” on page 12
“Anti-Virus Software and Firewalls” on page 17
“Instrument Information” on page 18
9
Quick Start
Initial Inspection
Initial Inspection
Inspect the shipping container and the cushioning material for signs of stress.
Retain the shipping materials for future use, as you may wish to ship the
analyzer to another location or to Keysight Technologies for service.
Verify the Contents
Verify the shipping container contents using the box contents list.
Shipping Problems?
If the shipping materials are damaged or the contents of the container are
incomplete:
— Contact the nearest Keysight Technologies office.
— Keep the shipping materials for the carrier’s inspection.
— If you must return an analyzer to Keysight Technologies, use the original (or
comparable) shipping materials. See “Returning an Analyzer for Service”
on page 93.
10 N9042B Signal Analyzer Getting Started and Troubleshooting Guide
Quick Start
Instrument Location and Rack Mounting Requirements
Instrument Location and Rack Mounting Requirements
Locating the Analyzer
Make sure that the fan inlet and exhaust vent areas on the sides of the analyzer
are not obstructed. The minimal required clearance is 2 inches. Airflow
restrictions cause additional airflow noise and cause the fans to speed up so
they can draw in enough air for the required cooling. This results in excessive
audible noise.
Cooling and Rack Mounting
Do not rack mount the analyzer side-by-side with any other instrument with
side-by-side ventilation. Make sure the exhaust air from the first instrument is
directed away from the inlet of the second unit. If the pre-heated air from the
first instrument is directed into the second instrument, it can cause excessive
operating temperatures in the second unit and can cause instrument failures.
When facing the rear panel, the analyzer draws air in from the left side and
exhausts air from the right side.
More than one person is required to lift or carry this instrument.
Alternately a mechanical lift can be used to eliminate the risk of
personal injury.
Safety of any system incorporating the equipment is the responsibility
olf the assembler of the system.
When installing the product in a cabinet, the convection into and out of
the product must not be restricted. The ambient temperature (outside the
cabinet) must be less than the maximum operating temperature of the
product by 4° C for every 100 watts dissipated in the cabinet. If the total
power dissipated in the cabinet is greater than 800 watts, then forced
convection must be used.
This product has been designed and tested in accordance with accepted industry
standards and has been supplied in safe condition. The documentation contains
information and warnings that must be followed by the user to ensure safe operation and
to maintain the product in a safe condition.
Only Keysight approved accessories shall be used.
N9042B Signal Analyzer Getting Started and Troubleshooting Guide 11
Quick Start
Turning on the Analyzer the First Time
Turning on the Analyzer the First Time
Initial power-on of the analyzer can be accomplished using the following
methods:
Initializing the Analyzer
Proper Ergonomics should be considered when using accessories such as a keyboard or a
mouse.
StepsActionsNotes
1. Power on the analyzera. Position the analyzer so
you have easy access to
the power cord and plug
it in.
b. Press the power switch
(located in the lower left
corner of the analyzer’s
front panel) to turn the
analyzer on.
2. Viewing the End-User
License Agreement
After a brief startup, the following window appears giving you information about the
End-User License Agreement.
See
“Instrument Location and Rack
Mounting Requirements
Requirements” on page 18
The analyzer can require more than 5 minutes to
power-on.
The Keysight Technologies screen appears followed by a
screen that allows you to select Windows10 or the
Recovery option.
and “Power
for more details.
At this time, it is safe to turn off the instrument before initializing the software. See
“Instrument Power Down Process” on page 16 for more details.
12 N9042B Signal Analyzer Getting Started and Troubleshooting Guide
Quick Start
Turning on the Analyzer the First Time
StepsActionsNotes
3. Reboot and log on
If you do not check the “Do not show this message again” check box, this message will be displayed
each time the analyzer is turned on. No application will start while this message is displayed. Before
continuing, make sure that you carefully read the Anti-Virus message and determine what action is
appropriate.
4. Disable the Anti-Virus
message
select
OK.
Continue.
— Select
— Select the check box and
When the instrument restarts, the following message
window appears:
This window
appears and covers
the Launch window.
Messages similar to the following continue to appear:
5. When the installation
is complete, you
should see a display
like this:
Several required processes continue.
The application initializes
N9042B Signal Analyzer Getting Started and Troubleshooting Guide 13
Quick Start
Turning on the Analyzer the First Time
StepsActionsNotes
6. Set user interface
language
a. On the instrument,
System, User
select
Interface tab
b. Choose the desired
language from the
Language drop-down
menu.
.
The System key is the in the upper right corner of
the screen.
7. Verify the installationa. On the instrument,
select
System, Show
System.
b. Verify that the
purchased
application(s) appear in
the list or have an
entitlement certificate.
If you require further assistance, contact the Keysight
support team.
Online assistance:
http://www.keysight.com/find/assist
14 N9042B Signal Analyzer Getting Started and Troubleshooting Guide
Quick Start
Turning on the Analyzer the First Time
StepsActionsNotes
8. View a signal
a. Select
RF Calibrator,
50 MHz.
b. Select
select
Frequency
c. Select
The 50 MHz reference signal appears on the
display.
Input/Output,
FREQ, then
Center
, 50MHz.
Span, 50 MHz.
This routes the internal 50 MHz signal to the analyzer
input.
N9042B Signal Analyzer Getting Started and Troubleshooting Guide 15
Quick Start
Instrument Power Down Process
Instrument Power Down Process
The following steps ensure that your data is saved before the instrument shuts
down.
Recommended Instrument Shut Down
Step
1. Briefly press the front panel power button.
Do not hold the button down (since this signals the processor
to immediately shut down before saving data, which could
potentially cause disk corruption).
The instrument will begin the power down sequence where the instrument
2.
performs clean-up activities such as closing applications and writing data to
disk. The “Shutting Down” Message will appear on screen.
Do not disconnect the AC power at this time. It is important
that the instrument be able to complete all power down
activities before power is interrupted.
3.
When the display goes blank, the fans stop, and the front panel yellow standby
light turns on,
4. You may safely remove the power cable at this time.
the instrument is completely shut down.
16 N9042B Signal Analyzer Getting Started and Troubleshooting Guide
Quick Start
Anti-Virus Software and Firewalls
Anti-Virus Software and Firewalls
No third-party anti-virus software is shipped with the analyzer. It is
recommended that you install anti-virus software if your analyzer is connected
to the LAN. Check with your IT department to see what they recommend.
Do not modify the default network settings as this may cause problems with
the operating system of the analyzer.
The analyzer is shipped with the Windows 10 firewall enabled and Windows
Defender.
To adjust Windows Defender settings you must be logged in as an "administrator" (default password:"Keysight4u!"). Minimize the X-Series Application, click
the Start button and type: defender. Then click on Windows Defender from the
Best match column.
N9042B Signal Analyzer Getting Started and Troubleshooting Guide 17
Quick Start
Instrument Information
Instrument Information
Power Requirements
The only physical installation of your Keysight signal analyzer is a connection
to a power source. Line voltage does not need to be selected.
This analyzer does not contain customer serviceable fuses.
The instruments can operate with mains supply voltage fluctuations up to ±10%
of the nominal voltage.
This is a Safety Class 1 Product (provided with a protective earthing ground
incorporated in the power cord). The mains plug shall only be inserted in a
socket outlet provided with a protective earth contact. Any interruption of the
protective conductor inside or outside of the product is likely to make the
product dangerous. Intentional interruption is prohibited.
Failure to ground the analyzer properly can result in personal injury. Before
turning on the analyzer, you must connect its protective earth terminals to the
protective conductor of the main power cable. Insert the main power cable plug
into a socket outlet that has a protective earth contact only. DO NOT defeat the
earth-grounding protection by using an extension cable, power cable, or
autotransformer without a protective ground conductor.
This product is designed for use in Installation Category II and Pollution Degree 2
and Measurement Category None.
This instrument has autoranging line voltage input. Be sure the supply voltage is
within the specified range.
The Mains wiring and connectors shall be compatible with the connector used in the premise
electrical system. Failure to ensure adequate earth grounding by not using the correct
components may cause product damage, and serious injury.
AC Power Cord
The analyzer is equipped with a three-wire power cord, in accordance with
international safety standards. This cable grounds the analyzer cabinet when
connected to an appropriate power line outlet. The cable appropriate to the
original shipping location is included with the analyzer.
Always use the three-prong AC power cord supplied with this product. Failure to
ensure adequate earth grounding by not using this cord can cause product
damage.
18 N9042B Signal Analyzer Getting Started and Troubleshooting Guide
Quick Start
Instrument Information
If this product is not used as specified, the protection provided by the
equipment could be impaired. This product must be used in a normal condition
(in which all means for protection are intact) only.
Install the instrument so that the detachable power cord is readily identifiable
and easily reached by the operator. The detachable power cord is the
instrument disconnecting device. It disconnects the mains circuits from the
mains supply before other parts of the instrument. The front panel switch is only
a standby switch and is not a LINE switch. Alternatively, an externally installed
switch or circuit breaker (which is readily identifiable and is easily reached by
the operator) may be used as a disconnecting device.
Protecting Against Overpowering
The input circuitry of the analyzer can be damaged by applying signals that
exceed the maximum safe input level of +30 dBm average total power or +/-
0.2 Vdc (when DC coupled) or +/- 70 Vdc (when AC coupled). If the instrument
signal path is set to Full Bypass, the maximum safe input level is reduced.
Refer to the analyzer's specification guide for more details regarding the
Maximum Safe Input Level. Repairing damage to the input circuitry can be
expensive.
If the analyzer will be used to measure signals which might be near the
maximum safe input level, use external attenuators and/or limiters to help
protect the analyzer input. The External Gain, amplitude Corrections, and/or
Ref Lvl Offset features may be used to compensate for the gains and losses of
external devices. External Gain and Corrections are under the Input/Output
menu and Ref Lvl Offset is under the AMPTD Y-Scale menu.
N9042B Signal Analyzer Getting Started and Troubleshooting Guide 19
Quick Start
Instrument Information
Environmental and Regulatory Information
Environmental Information
Samples of this product have been type tested in accordance with the Keysight
Environmental Test Manual and verified to be robust against the environmental
stresses of Storage, Transportation and End-use; those stresses include but
are not limited to temperature, humidity, shock, vibration, altitude and power
line conditions. Test Methods are aligned with IEC 60068-2 and levels are
similar to MIL-PRF-28800F Class 3.
Regulatory Information
This product has been designed and tested in accordance with accepted
industry standards, and has been supplied in a safe condition. The
documentation contains information and warnings that must be followed by
the user to ensure safe operation and to maintain the product in a safe
condition.
The V3050A SA Frequency Extender complies with the following
Electromagnetic Compatibility (EMC) compliances:
EMC Compliance
This product complies with the essential requirements of the European EMC
Directive as well as current editions of the following standards (dates and
editions are cited in the Declaration of Conformity):
—IEC/EN 61326-1
—CISPR Pub 11 Group 1, Class A
—AS/NZS CISPR 11
—ICES/NMB-001
This ISM device complies with Canadian ICES-001. Cet appareil ISM est
conforme a la norme NMB-001 du Canada.
This is a sensitive measurement apparatus by design and may have some performance
loss (up to 25 dBm above the Spurious Responses, Residual specification of –100 dBm)
when exposed to 3V/m ambient continuous electromagnetic phenomenon in the range
of 80 MHz to 6 GHz (similar to those used in testing per IEC 61000-4-3).
20 N9042B Signal Analyzer Getting Started and Troubleshooting Guide
Quick Start
Instrument Information
South Korean Class A EMC declaration
This equipment has been conformity assessed for use in business
environments. In a residential environment this equipment may cause radio
interference.
Regulatory Markings
The table below lists the definitions of markings that may be on or with the
product.
ccr.keysight@keysight.com
The CE marking is a registered trademark of the European Community. It indicates that
the product complies with all relevant directives.
ICES/NMB-001 indicates that this ISM device complies with the Canadian ICES-001.
Cet appareil ISM est conforme a la norme NMB-001 du Canada.
ISM GRP 1-A Class A indicates that this is an Industrial Scientific and Medical Group 1
Class A product.
The Keysight email address is required by EU directives applicable to our product.
China Restricted Substance Product Label. The EPUP (environmental protection use
period) number in the center indicates the time period during which no hazardous or
toxic substances or elements are expected to leak or deteriorate during normal use and
generally reflects the expected useful life of the product.
The RCM mark is a registered trademark of the Australian Communications and Media
Authority.
This WEEE symbol indicates separate collection for electrical and electronic equipment,
mandated under EU law as of August 13, 2005. All electric and electronic equipment are
required to be separated from normal waste for disposal.
South Korean Certification (KC) mark. It includes the marking’s identifier code in the
format shown.
The instruction documentation symbol. The product is marked with this symbol when it is
necessary for the user to refer to the instruction in the documentation.
N9042B Signal Analyzer Getting Started and Troubleshooting Guide 21
Quick Start
Instrument Information
Waste Electrical and Electronic Equipment (WEEE) Directive
2002/96/EC
This instrument complies with the WEEE Directive (2002/96/EC) marking
requirement. This affixed product label indicates that you must not discard this
electrical or electronic product in domestic household waste.
Product Category
With reference to the equipment types in the WEEE directive Annex 1, this
instrument is classified as a “Monitoring and Control Instrument” product.
The affixed product label is as shown below.
Do not dispose in domestic household waste.
To return this unwanted instrument, contact your nearest Keysight Service
Center, or visit
To prevent electrical shock, disconnect the signal analyzer from mains before
cleaning. Use a dry cloth or one slightly dampened with water to clean the
external case parts. Do not attempt to clean internally.
Cleaning Connectors
Cleaning connectors with alcohol shall only be done with the instrument power
cord removed, and in a well-ventilated area. Allow all residual alcohol moisture
to evaporate, and the fumes to dissipate prior to energizing the instrument.
Keep isopropyl alcohol away from heat, sparks, and flame. Store in a tightly
closed container. It is extremely flammable. In case of fire, use alcohol foam, dry
chemical, or carbon dioxide; water may be ineffective.
Use isopropyl alcohol with adequate ventilation and avoid contact with eyes,
skin, and clothing. It causes skin irritation, may cause eye damage, and is harmful
if swallowed or inhaled. It may be harmful if absorbed through the skin. Wash
thoroughly after handling.
In case of spill, soak up with sand or earth. Flush spill area with water.
Dispose of isopropyl alcohol in accordance with all applicable federal, state, and
local environmental regulations.
22 N9042B Signal Analyzer Getting Started and Troubleshooting Guide
Quick Start
Instrument Information
Battery Information
The analyzer uses a lithium battery located on the CPU board. This is not an
operator replaceable part. See “Returning an Analyzer for Service” on
page 93. Replaceable parts must be approved or supplied by Keysight
Technologies.
Danger of explosion if battery is incorrectly replaced. Replace only with the same
or equivalent type recommended. Discard used batteries according to the
manufacturer’s instructions.
Do not throw batteries away but collect as small chemical waste.
N9042B Signal Analyzer Getting Started and Troubleshooting Guide 23
Quick Start
Instrument Information
Protecting against electrostatic discharge
Electrostatic discharge (ESD) can damage or destroy electronic components
(the possibility of unseen damage caused by ESD is present whenever
components are transported, stored, or used).
Test Equipment and ESD
To help reduce ESD damage that can occur while using test equipment:
Do not use these first three techniques when working on circuitry with a voltage
potential greater than 500 volts.
— Before connecting any coaxial cable to an analyzer connector for the first
time each day, momentarily short the center and outer conductors of the
cable together.
— Personnel should be grounded with a 1 MΩ resistor-isolated wrist-strap
before touching the center pin of any connector and before removing any
assembly from the analyzer.
— Be sure that all instruments are properly earth-grounded to prevent
build-up of static charge.
— Perform work on all components or assemblies at a static-safe workstation.
— Keep static-generating materials at least one meter away from all
components.
— Store or transport components in static-shielding containers.
— Always handle printed circuit board assemblies by the edges. This reduces
the possibility of ESD damage to components and prevent contamination of
exposed plating.
Additional Information About ESD
For more information about ESD and how to prevent ESD damage, contact the
Electrostatic Discharge Association (http://www.esda.org). The ESD standards
developed by this agency are sanctioned by the American National Standards
Institute (ANSI).
24 N9042B Signal Analyzer Getting Started and Troubleshooting Guide
Keysight X-Series Signal Analyzers
N9042B
Getting Started and Troubleshooting Guide
2Front and Rear Panel Features
This section describes the following features:
“Front-Panel Features” on page 26
“Display Features” on page 29
“Rear-Panel Features” on page 36
“Front and Rear Panel Symbols” on page 38
25
Front and Rear Panel Features
Front-Panel Features
Front-Panel Features
Item
Description
#Name
1Measurement KeysThese keys (in the shaded area) enable you to set the parameters used for making
measurements in the current Mode and Measurement.
2Preset KeysMode Preset - local to the current mode, global to all measurements in the mode,
affects most but not all parameters in the mode, does not affect Input/Output or
System variables
User Preset - local to the current mode, global to all measurements in the mode,
affects all parameters in the mode as well as the Input/Output variables. Does not
affect System variables.
3Save/Recall KeysSave - enables you to save states, traces, screen images and other items from the
analyzer to files on the analyzer’s internal storage, to removable devices, and to
directories on the network.
Quick Save - enables you to repeat the previous Save. Whatever you saved before
gets saved again to the same directory.
Recall - enables you to recall previously saved states, traces and other items to the
analyzer from files on the analyzer’s internal storage, from removable devices, and
from directories on the network.
26 N9042B Signal Analyzer Getting Started and Troubleshooting Guide
Front and Rear Panel Features
Front-Panel Features
Item
Description
#Name
4Sweep KeysSingle/Cont - toggles between single and continuous measurement sweeps.
Restart - restarts the measurement.
5Mode/Measurement KeyThis key enables you to select the desired Mode (measurement application),
Measurement, and/or View.
6Enter and Arrow KeysThe Enter key terminates data entry when either no unit of measure is needed, or you
want to use the current unit.
The arrow keys:
— Increment and decrement the value of the current measurement
selection. (up/down = large increment, left/right = small increment)
— Navigate within tables.
7KnobIncrements and decrements the value of the current active function.
8Utility KeysThe following keys are available in the Utility section:
Numeric key pad
System
Help
Local/Cancel/(Esc)
Back-space
Delete
Control
Alt
Undo/Redo
Touch On/Off
Onscreen Keyboard
Tab
9RF InputConnector for inputting an external signal. Make sure that the total power of all
signals at the analyzer input does not exceed +30 dBm (1 watt).
10USB 3.0, Type A Connectors Standard USB 3.0 ports, Type A Connect to external peripherals such as a mouse,
keyboard, DVD drive, or hard drive. USB 3.0 connectors are capable of higher data
throughput that is required for certain peripherals. If your peripheral device has a
"blue" USB connector, you should connect it to a USB connector with a blue colored
insulator to ensure that it works as advertised.
11USB 2.0, Type A Connectors
The two USB connectors in the
middle)
12USB 3.0, Type C ConnectorUSB 3.0 Type C. Provides charging power for other devices.
High IF In (Option EXW) Connection to V3050A frequency extender.
13
14High LO Out(Option EXW) Connection to V3050A frequency extender.
15Ext MixerProvides LO output signal to and receives IF input signals from an external mixer. See
16Cal OutInternal calibrator output.
Standard USB 2.0 ports, Type A Connect to external peripherals such as a mouse,
keyboard, DVD drive, or hard drive.
the Specifications Guide for details on signal levels.
N9042B Signal Analyzer Getting Started and Troubleshooting Guide 27
Front and Rear Panel Features
Front-Panel Features
Item
Description
#Name
17Power Standby/ OnTurns the analyzer on. A green light indicates power on. A yellow light indicates
standby mode.
The front-panel switch is a standby switch, not a LINE switch
(disconnecting device). The analyzer continues to draw
power even when the front-panel switch is in standby.
The main power cord can be used as the system
disconnecting device. It disconnects the mains circuits from
the mains supply.
18Wide IF Out(Option CRW) Wide bandwidth IF output.
28 N9042B Signal Analyzer Getting Started and Troubleshooting Guide
Front and Rear Panel Features
Display Features
Display Features
This section describes the regions of the display.
Invalid Data Indicator
The invalid data indicator is displayed whenever the data on the display
does not match the settings of the analyzer. The most common example
of this is when instrument settings have changed in the time since the data in
the traces on the display was taken. This means that the screen annotation
cannot be guaranteed to match the trace data. For example, if you change
Center Frequency, the invalid data indicator will display until a new sweep has
completed.
If any Trace is in View mode (displaying but not updating) and analyzer settings
are changed, the invalid data indicator will display as long as that trace
remains in View. Traces that are blanked do not turn on the invalid data
indicator.
Not all analyzer settings require display of the invalid data indicator when they
change, only changes that require a new acquisition will cause it to display.
For example, changing the Y-Axis scale of the analyzer does not cause the
invalid data indicator to display, unless the attenuation changes.
Also, the invalid data indicator is turned on:
— When the counter is turned on, until the completion of the first count
N9042B Signal Analyzer Getting Started and Troubleshooting Guide 29
Front and Rear Panel Features
Display Features
— When a trace is imported from mass storage and the trace’s parameters do
not match the current analyzer settings
— When a trace is sent to the analyzer from a remote interface (since there is
no way to know if its settings match)
The Data Invalid Indicator has an associated status bit that can be checked at any time
to see if it is on.
Screen Tabs
Along the top of the display are tabs, one for each measurement screen you
have defined. Tap the icon to create a new tab as a "clone" of the current
measurement, which can be changed once it is created:
Tap the current screen tab (or press the Mode/Meas hardkey) to display the
following dialog:
This dialog allows you to choose a Mode, Measurement and View.
When you select a mode, the measurements that are available in the mode are
displayed in the Measurement column.
When you select the desired measurement, the views available for the
measurement are displayed under the View column.
30 N9042B Signal Analyzer Getting Started and Troubleshooting Guide
Loading...
+ 69 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.