KERN & SOHN TPE User Manual

Sauter GmbH
Ziegelei 1
TPE
D-72336 Balingen E-Mail: info@sauter.eu
Instruction Manual
MANUAL TEST STAND FOR PEEL TESTS
Thank you for buying this SAUTER Test Stand for peel tests. This test stand is very robust and will serve you many years if you are going to operate it correctly and maintain with care. We hope you are pleased with your high quality product with and its functions. If you have any queries, wishes or helpful s uggestions, do not hesitate to call our service number.
Introduction
This manual test stand was specially designed for 90° peel tests in connection with our SAUTER forc e gauges up to 500N. Hereby, typically, an adherent layer of material is torn off from the base material, shown in the illustration below:
The decisive factor is the force, which is necessary to peel off the adherent, above located material. The SAUTER TPE is designed in the way that the force gauge which is peeling off the layer, moves laterally and upwards simultaneously. With this, the peeling off movement is generated and the interference of falsified transverse forces can be avoided. The crank handle operation enables a very reliable and easy handling.
Tel: +49-[0]7433- 9933-199 Fax: +49-[0]7433-9933-149 Internet: www.kern-sohn.com
Operation
The test stand has to be placed onto a stable support plate. A s uitable clamp, for example AC 14 (as shown in the picture on the left) has to be screwed onto the force gauge. Subsequently, the force gauge has to be fixed by means of an adapter plate, exactly in vertical position, on the test stand. A piece of the layer, which has to be peeled off, has to be fixed in the clamp. It has to be made sure that the material for the peel test has been fixed correctly and tightly enough in the clamp, in the way that it can’t slide off. Now, by means of the crank handle movement in clockwise direction, the force gauge m oves up, vertically fixed and in a tilted way, as the movem ent of the test stand is effected in a 45° angle to the horizontal. The applied force can be read off on the display of the force gauge (here FH) soon afterwards.
Technical data
Distance of one stroke (travel): 3.1 mm per turn
Maximum peel off length: 105 mm
Dimensions: L 420 x B 215 x H 480 mm
Net weight: 22 kg
Maintenance
To avoid rusting, the test stand has to be cleaned with a smooth cloth after each use. There may not be applied any abrasive detergents.
TPE-BA-e-1310 1