KERN GAB-P Operating Manual

KERN & Sohn GmbH
Ziegelei 1
D-72336 Balingen
E-Mail: info@kern-sohn.com
Fax: +49-[0]7433-9933-149
Operating manual Table scales
KERN GAB-P
Version 1.1
03/2014
GB
GAB-P-BA-e-1411
GB
KERN GAB-P
Version 1.1 03/2014
Operating manual
Table scale
Contents
1 TECHNICAL DATA ............................................................................................. 4
2 APPLIANCE OVERVIEW ................................................................................... 6
2.1 Overview of display .................................................................................................................. 8
2.2 Keyboard overview ................................................................................................................... 9
2.3 Dimensions ............................................................................................................................. 11
3 BASIC INFORMATION (GENERAL) ................................................................ 11
3.1 Proper use ............................................................................................................................... 11
3.2 Improper Use........................................................................................................................... 11
3.3 Warranty .................................................................................................................................. 12
3.4 Monitoring of Test Resources ............................................................................................... 12
4 BASIC SAFETY PRECAUTIONS ..................................................................... 13
4.1 Pay attention to the instructions in the Operation Manual ................................................ 13
4.2 Personnel training .................................................................................................................. 13
5 TRANSPORT AND STORAGE ......................................................................... 13
5.1 Testing upon acceptance ...................................................................................................... 13
5.2 Packaging / return transport ................................................................................................. 13
6 UNPACKING, SETUP AND COMMISSIONING ............................................... 14
6.1 Installation Site, Location of Use .......................................................................................... 14
6.2 Unpacking / installation ......................................................................................................... 15
6.3 Scope of delivery .................................................................................................................... 16
6.4 Mains connection ................................................................................................................... 16
6.5 Rechargeable battery operation ............................................................................................ 16
6.6 Initial Commissioning ............................................................................................................ 16
6.7 Adjustment .............................................................................................................................. 17
6.8 Non-verifiable models: ........................................................................................................... 18
6.9 Verifiable models: ................................................................................................................... 18
6.10 Verification .............................................................................................................................. 19
6.10.1 Position of seals and adjusting switch ............................................................................. 20
6.11 Checking the balance verification settings ......................................................................... 20
2 GAB-P-BA-e-1411
7 OPERATION ..................................................................................................... 21
7.1 Switching on/off ...................................................................................................................... 21
7.2 Zeroing..................................................................................................................................... 21
7.3 Taring ....................................................................................................................................... 22
7.4 Overload warning ................................................................................................................... 23
7.5 Weighing units switch-over ................................................................................................... 23
7.6 Percent weighing .................................................................................................................... 24
7.7 Parts counting ........................................................................................................................ 25
7.8 Weighing with tolerance range ............................................................................................. 26
7.9 Manual totalizing .................................................................................................................... 28
7.10 Automatic adding-up .............................................................................................................. 30
8 THE MENU ........................................................................................................ 32
8.1 Navigation in the menu: ......................................................................................................... 32
8.2 Menu overview ........................................................................................................................ 33
8.2.1 Non-verifiable models:.......................................................................................................... 33
8.2.2 Verifiable models: ................................................................................................................. 34
9 PRINTER ........................................................................................................... 36
9.1 Change paper roll ................................................................................................................... 36
10 DATA OUTPUT RS 232C .............................................................................. 37
10.1 Technical data ......................................................................................................................... 37
10.2 Pin allocation of the bala nce output socket (front view) .................................................... 37
10.3 Remote control instructions .................................................................................................. 37
11 MAINTENANCE, SERVICING, REMOVAL, ERROR MESSAGES ............... 38
11.1 Clean ........................................................................................................................................ 38
11.2 Servicing, maintenance ......................................................................................................... 38
11.3 Disposal ................................................................................................................................... 38
11.4 Error messages ....................................................................................................................... 38
12 DECLARATION OF CONFORMITY .............................................................. 39
GAB-P-BA-e-1411 3
1 Technical data
KERN GAB 6K-5P GAB 10K-4P GAB 30K-4P
Weighing range (max) 6 kg 12 kg 30 kg Readability (d) 0.05 g 0.1 g 0.2 g Reproducibility 0.05 g 0.1 g 0.2 g Linearity ± 0.15 g ± 0.3 g ± 0.6 g Stabilization time 2 s 2 s 2 s Recommended adjusting
weight (not supplied)
6 kg (F2) 12 kg (F2) 30 kg (M1)
Heating time (operating temperature)
120 min
Net weight (kg) 6 kg Dimensions of housing
(l x L x h) [mm]
371 x 366 x 131
Weighing plate, stainless
steel
365 x 240 x 11 mm
Weighing Units kg, t, lb, oz Permissible
ambient condition
-10° C to 40° C Allowable air humidity 0 % - 80 % (non-condensing) Electric Supply
Mains adapter 220 V - 240 V AC, 50 Hz
balance 12 V, 500 mA
Rechargeable battery
6 V, 10 Ah
Operating period without printer – back ground lighting on 80 h. Operating period without printer – back ground lighting off 150 h if the printer is used frequently, the operating time will be halved
Loading time approx. 12 hrs.
Data interface RS 232C
4 GAB-P-BA-e-1411
KERN GAB 6K-3PM GAB 10K-3PM GAB 30K-3PM
Weighing range (max) 3 kg / 6 kg 6 kg / 15 kg 15 kg / 30 kg Readability (d)
1 g / 2 g
2 g / 5 g
5 g / 10 g
Minimum load 20 g 40 g 100 g Verification value 1 g / 2 g 2 g / 5 g 5 g / 10 g Verification class III III III Reproducibility
1 g /2 g
2 g / 5 g
5 g / 10 g
Linearity ± 1 g / 2 g ± 2 g / 5 g ± 5 g / 10 g Stabilization time 2 s 2 s 2 s Recommended adjusting
weight (not supplied)
6 kg (M1) 15 kg (F2) 30 kg (M1)
Heating time (operating temperature)
10 min / 10 min
Net weight (kg) 6 kg Dimensions caisse
(l x L x h) [mm]
371 x 366 x 131
Weighing plate, stainless
steel
365 x 240 x 11 mm
Weighing Units g, kg Permissible
ambient condition
-10° C to 40° C Allowable air humidity 0 % - 80 % (non-condensing) Electric Supply
Mains adapter 110 – 230 V AC
Balance 12 V DC, 500mA
Rechargeable battery
6 V, 10 Ah
Operating period without printer – back ground lighting on 80 h. Operating period without printer – background lighting off 150 h if the printer is used frequently, the operating time will be halved
Loading time approx. 12 hrs.
Data interface RS 232C
GAB-P-BA-e-1411 5
2 Appliance overview
1. Weighing pan
2. Printer
3. Buttons
4. Bubble level
5. RS 232 interface
6. Battery compartment
6 GAB-P-BA-e-1411
7. Transport guard screw
8. Levelling screw
9. Connection of mains adapter
10. ON/OFF switch
11. Seal of verifiable models
12. Seal with an underneath verification switch (verifiable models)
GAB-P-BA-e-1411 7
2.1 Overview of display
Display Description Description
1
The battery capacity display
appears when the capacity of the battery is getting low
2
Stability display Scales are in a steady state
3
weighing units selectable Non-verifiable models: kg, t, lb, oz
Verifiable models: g, kg
4
GROSS
Gross weight display Showing gross weight
NET
Net weight display Displays the net weight
ZERO
Zeroing display Should the balance not display exactly zero
despite empty scale pan, press the
0
button. The balance will be set to zero after a short standby time.
TARE
Taring Scales have been tared
AUTO
Automatic adding-up Is displayed if in t he m enu „ Aut om atic
totalizing“ has been selected
5
LO
OK
HI
Tolerance check Check weighing
Load above (HI), below (LO) or within (OK) tolerance range
8 GAB-P-BA-e-1411
1 2
2.2 Keyboard overview
1. Bubble level
Level balance with foot screws until the air bubble of the water balance is in the prescribed circle.
2. Storage battery status display
Power supply connected Icon is glowing for power supply via power pack
If the LED lights green, the rechargeable battery is fully charged.
When the battery is being recharged, the LED lights red
GAB-P-BA-e-1411 9
Selection Function
F
In piece counting mode change between refer ence weight - item number - total weight
UNIT In weighing mode: Weighing units switch-over
PRINT
Data transfer via interface
Save value to memory if memory
function “automatic“, is disabled
In menu: Select main menu items
Call up menu: Press both buttons at the same time
ESC Return to weighing mode
Changes between weighing mode and quantity counting mode
C
Delete displayed value
In menu: Return to weighing mode
%
Call up percent weighing Move to the left by one decimal point
TOL Enter upper, lower or both limits
Move to the right by one decimal point
TARE
Tare balance and enter pre-tare value
In menu: Select submenus
Increase displayed value
Reset scales to zero
In menu: Save selected setting
Confirm entered value or select funct ion
10 GAB-P-BA-e-1411
2.3 Dimensions
3 Basic Information (General)
3.1 Proper use
The balance acquired by you serves to determine weight. It is intended to be used as a “non-automatic balance”, i.e. the material to be weighed is manually and carefully placed in the centre of the weighing pan. As soon as the goods to be weighed are placed on the balance and the weight has been displayed.
3.2 Improper Use
Do not use balance for dynamic weighing. In the event that small quantities are removed or added to the material to be weighed, incorrect weighing results can be displayed due to the “stability compensation“. (Example: Slowly draining fluids from a container on the balance.) Do not leave permanent load on the weighing pan. This may damage the measuring system. Impacts and overloading exceeding the stated maximum load (max) of the balance, minus a possibly existing tare load, must be strictly avoided. Balance may be damage by this. Never operate balance in explosive environment. The serial version is not explosion protected. The structure of the balance may not be modified. This may lead to incorrect weighing results, safety-related faults and destruction of the balance. The balance may only be used according to the described conditions. Other areas of use must be released by KERN in writing.
GAB-P-BA-e-1411 11
3.3 Warranty
Warranty claims shall be voided in case
Our conditions in the operation manual are ignored
The appliance is used outside the described uses
The appliance is modified or opened
Mechanical damage or damage by media, liquids, natural wear and tear
The appliance is improperly set up or incorrectly electrically connected
The measuring system is overloaded
3.4 Monitoring of Test Resources
In the framework of quality assurance the measuring-related properties of the balance and, if applicable, the testing weight, must be checked regularly. The responsible user must define a suitable interval as well as type and scope of this test. Information is available on KERN’s home page (www.kern-sohn.com with regard to the monitoring of balance test substances and the test weights required for this.(www.kern-sohn.com). In KERN's accredited DKD calibration laboratory test weights and balances may be calibrated (return to the national standard) fast and at moderate cost.
12 GAB-P-BA-e-1411
4 Basic Safety Precautions
4.1 Pay attention to the instructions i n the Operation Manual
Carefully read this operation manual before setup and commissioning, even if you are already familiar with KERN balances.
All language versions contain a non-binding translation. The German original document is binding.
4.2 Personnel training
The appliance may only be operated and maintained by trained personnel.
5 Transport and st orage
5.1 Testing upon acceptance
When receiving the appliance, please check packaging immediately, and the appliance itself when unpacking for possible visible damage.
5.2 Packaging / return transport
Keep all parts of the original packaging for a possibly required
return.
Only use original packaging for returning. Prior to dispatch disconnect all cables and remove loose/mobile
parts.
Reattach transport securing devices. (see chapter 6.2) Secure all parts such as the glass wind screen, the weighing
platform, power unit etc. against shifting and damage.
GAB-P-BA-e-1411 13
6 Unpacking, Setup and Commissioning
6.1 Installation Site, Location of Use
The balances are designed in a way that reliable weighing results are achieved in common conditions of use. You will work accurately and fast, if you select the right location for your balance.
On the installation site observe the following:
Place the balance on a firm, level surface;
Avoid extreme heat as well as temperature fluctuation caused by installing
next to a radiator or in the direct sunlight;
Protect the balance against direct draughts due to open windows and doors;
Avoid jarring during weighing;
Protect the balance against high humidity, vapours and dust;
Do not expose the device to extreme dampness for longer periods of time.
Non-permitted condensation (condensation of air humidity on the appliance) may occur if a cold appliance is taken to a considerably warmer environment. In this case, acclimatize the disconnected appliance for ca. 2 hours at room temperature.
Avoid static charge of goods to be weighed or weighing container.
If electro-magnetic fields or static charge occur, or if the power supply is unstable major deviations on the display (incorrect weighing results) are possible. In that case, the location must be changed.
14 GAB-P-BA-e-1411
6.2 Unpacking / installation
Carefully remove the balance from the packaging, remove plastic cover and setup balance at the intended workstation.
Make sure that all transport locking devices are removed
To release the transport lock unscrew the transport screw
1 – non verifiable models or  2 –verifiable models
counter-clockwise.
For transportation carefully screw-in transport screw clockwise till to the stopper and then fix it using locknut.
GAB-P-BA-e-1411 15
6.3 Scope of delivery
Serial accessories:
Balance, see chap. 2
Mains adapter
Rechargeable battery
Operating manual
6.4 Mains connection
Power is supplied via the external mains adapter. The stated voltage value must be the same as the local voltage. Only use KERN original mains adapter. Using other makes requires consent by KERN.
6.5 Rechargeable battery operation The rechargeable battery is charged via the delivered power supply.
Before the first use, the battery should be charged by connecting it to the mains power supply for at least 12 hours.
The appearance of the battery symbol in the weight display indicates that the battery is almost exhausted. To charge the chargeable battery, connect the mains cable. The LED display informs you during loading about the loading status of the rechargeable battery.
If the LED lights green, the rechargeable battery is fully charged.
When the battery is being recharged, the LED lights red
6.6 Initial Commissioning
In order to obtain exact results with the electronic balances, your balance must have reached the operating temperature (see warming up time chap. During this warming up time the balance must be connected to the power supply (mains, accumulator or battery).
The accuracy of the balance depends on the local acceleration of gravity. Strictly observe hints in chapter Adjustment.
16 GAB-P-BA-e-1411
6.7 Adjustment
As the acceleration value due to gravity is not the same at every location on earth, each balance must be coordinated - in compliance with the underlying physical weighing principle - to the existing acceleration due to gravity at its place of location (only if the balance has not already been adjusted to the location in the factory). This adjustment process must be carried out for the first commissioning, after each change of location as well as in case of fluctuating environment temperature. To receive accurate measuring values it is also recommended to adjust the balance periodically in weighing operation.
The adjustment is locked for verified balances. To disable the access lock,
requires that the seal is destroyed and the adjusting switch is actuated in step 3 when turning on the scale. For position of adjusting switch, see chap.
6.10.1. Attention: After destruction of the seal the balance must be re-verified by an authorised agency and a new verification wire/seal mark fitted before it can
be reused for applications subject to verification.
Procedure when adjusting:
Observe stable environmental conditions. A warming up time (see chapter 1) is required for stabilization. Ensure that there are no objects on the weighing pan.
GAB-P-BA-e-1411 17
6.8 Non-verifiable models:
Switch on the balance and during the selftest press the
and the button at the same time. „UnLoad“ is displayed shortly.
Ensure that there are no objects on the weighing pan.
„LoAd“ is displayed
Place adjustment weight, the stability display appears After the adjustment the balance will carry out a self-test.
„PASS“ will be displayed shortly. Remove adjusting weight during selftest, the appliance will return into weighing mode automatically. An adjusting error or incorrect adjusting weight will be indicated by the error message; repeat adjustment procedure.
6.9 Verifiable models:
Switch on balance and press the adjustment switch at the
same time.
After that, during the balance’s selftest, press
and at the same time.
„UnLoad“ is displayed.
Acknowledge with . Ensure that there are no objects on
the weighing pan.
The currently set adjustment weight will be displayed.
To change by using the navigation buttons (see chap. 8)
select the desired setting, the active digit is flashing.
Acknowledge with . „LoAd“ is displayed.
18 GAB-P-BA-e-1411
Carefully place adjusting weight in the centre of the weighing
plate. Wait for stability display , then press . „PASS“ will be displayed shortly.
After the adjustment the balance will carry out a self-test.
Remove adjusting weight during selftest, the appliance will return into weighing mode automatically. An adjusting error or incorrect adjusting weight will be indicated by the error message; repeat adjustment procedure.
6.10 Verification
UUGeneral hints:
According to EU directive 90/384/EEC balances must be officially verified if they are used as follows (legally controlled area):
a) For commercial transactions if the price of goods is determined by weighing. b) For the production of medicines in pharmacies as well as for analyses in the
medical and pharmaceutical laboratory. c) For official purpose. d) For manufacturing final packages.
In cases of doubt, please contact your local trade in standard. After verification the balance is sealed at the indicated positions.
Verification of the balance is invalid without the "seal".
Verification notes An EU type approval exists for balances described in their technical data as
verifiable. If a balance is used where obligation to verify exists as described above, it must be verified and re-verified at regular intervals. Re-verification of a balance is carried out according to the respective national regulations. The validity for verification of balances in Germany is e.g. 2 years. The legal regulation of the country where the balance is used must be observed!
Balances with obligation to verify must be taken out of operation if:
The weighing res ult of the balance is outside the error limit. Therefore, in
regular intervals load balance with known test weight (ca. 1/3 of the max. load)
and compare with displayed value.
The reverification deadline has been exceeded.
GAB-P-BA-e-1411 19
6.10.1 Position of seals and adjusting switch
1. Seal of approval
2. Cover
3. Housing screw
4. Adjustment switch
6.11 Checking the balance verification settings
For the adjustment, the balance must be switched over to service mode.
In the service mode the parameters of the balance can be modified.
The service parameters may not be modified, as this could damage the
balance settings.
In calibrated scales the service mode is locked individually for each switch. To disable the access lock, destroy the seal and actuate the switch.
Attention: After destruction of the seal the balance must be re-verified by an authorised agency and a new verification wire/seal mark fitted before it can be reused for applications subject to verification.
20 GAB-P-BA-e-1411
7 Operation
7.1 Switching on/off
To turn on/off confirm forward at the underneath on the right of the scale and
keep pressed for a little while. The balance will carry out a self-test. As soon as the weight display appears, the balance is ready for weighing.
To switch-off push backward the switch-on/switch-out on the right lower side of
the balance.
Printout example:
No.1
2013/08/10
18:39
N:
5.000kg
G
5.000kg
T:
0.000kg
7.2 Zeroing
Resetting to zero corrects the influence of light soiling on the weighing plate.
Unload the balance
Press , the balance starts resetting to zero.
GAB-P-BA-e-1411 21
7.3 Taring
Deposit weighing receptacles
(example)
After successful standstill control press the button.
The zero display and „NET“ will appear. The weight of the weighing container is now internally saved.
Place material to be weighed and read weight value
(example)
The taring process can be repeated any number of times,
e.g. when adding several components for a mixture (adding). The limit is reached when the whole weighing range is exhausted.
To delete the tare value, remove load from weighing plate
and press .
Printout example:
No.1
1. Weighing
2013/08/10
18:39
Date and time
N:
2.500kg
Net weight
G
3.200kg
Gross weight
T:
0.700kg
Tare value
22 GAB-P-BA-e-1411
7.4 Overload warning
Overloading exceeding the stated maximum load (max) of the balance, minus a possibly existing tare load, must be strictly avoided. This could cause damage to the balance. Exceeding maximum load is indicated by the display of „----“ and an audio sound. Unload balance or reduce preload.
7.5 Weighing units switch-over
Use the button to switch over between the individual weighing units.
Only in non-verifiable models: How to enable weighing units:
In menu select the menu item „Ut on“ or „-off“
(see chap.8.2), the first unit will be displayed
(example)
To enable [on] / disable [off] the displayed weighing unit,
press
Acknowledge with . The next unit with the current
setting will be displayed.
(example)
To enable [on] / disable [off] the displayed weighing unit,
press
Repeat sequence for each weighing unit.
Return to weighing mode using
GAB-P-BA-e-1411 23
7.6 Percent weighing
Percent weighing allows to display weight in percent, in relation to a reference weight.
Switch on balance, wait for zero display
Put a reference weight on the weighing plate, which
corresponds to 100 %.
(example)
Press , in the display 100,00 % appears Remove the reference weight
The display will return to 0.00 %
Place a sample
The display will show the percentage value of the sample in relation to the reference value.
(example)
 To return to gram / kilo mode, press
Printout example:
G
200,00%
Weight value in percent
24 GAB-P-BA-e-1411
7.7 Parts counting
Before the balance can count parts, it must know the average part weight (i.e. reference). Proceed by putting on a certain number of the parts to be counted. The balance determines the total weight and divides it by the number of parts (the so-called reference quantity). Counting is then carried out on the basis of the calculated average piece weight.
The higher the reference quantity the higher the counting exactness.
Place reference quantity in weighing mode.
Press and the scales will change to quantity counting
mode; reference quantity of 10, „P 10“ will appear.
Use to set the desired reference quantity (e.g. 100),
selectable 10, 20, 50, 100 and 200
Acknowledge by . An audio signal will sound shortly and a dashed line will be
displayed before the respective item number appears.
(example)
Remove the items from the weighing plate. The balance is
from now in parts counting mode counting all units on the weighing plate.
To change between reference weight, total weight and
quantity, press the key.
 Back to W eighing mode by .
Printout example:
2013/08/10
18:39
Date/time
G
1.000kg
Reference weight
10g/pcs
Piece weight
PCS:
100pcs
Number of items
GAB-P-BA-e-1411 25
7.8 Weighing with toleranc e range
You can set an upper or lower limit when weighing with tolerance range and thus ensure that the material to be weighed remains exactly within the set limits. Exceeding or dropping below the tolerance range is indicated by an audio sound or optical signal.
Audio signal:
The audio sound depends on the setting in menu block “bP“ (see chap.8.2.) Options:
0
No audio sound during tolerance weighing
1
An acoustic signal sounds when material to be weighed is within tolerance limits
2
An acoustic signal sounds when material to be weighed is beyond tolerance limits
Optical signal:
The arrow symbols indicate whether load is within the two set limits. The icons provide the following information:
1. Material to be weighed above tolerance limit
2. Material to be weighed within tolerance range
3. Material to be weighed below tolerance limit
26 GAB-P-BA-e-1411
Settings:
In weighing mode press , the left number flashes and
the symbol „HI“ will be displayed.
Enter upper limit by pressing and .
(example)
Acknowledge with .
The weighing scales will change to the display of the lower limit.
Enter the lower limit value and confirm with
(example)
The scales will change to zero display
Weighing with toleranc e range
Tare when using a weighing container. Put on material to be weighed, tolerance control is started.
The tolerance control is not active when the weight is under
20d.
To finish tolerance weighing, set both limit values to zero.
GAB-P-BA-e-1411 27
7.9 Manual totalizing
This function is used to add the individual weighing values to the summation memory and to provide a printout via the installed printer, when the stability display appears
and the button is pressed. (in order to set this function see chap. 8.2 „ACC on“)
Place material to be weighed A
Wait until stability display appears
(example)
Press
„ACC 1“ followed by the weight value will appear. The weight value is added into the summation memory and will be
printed out when the button is pressed
Remove material to be weighed A
Further material to be weighed can only be added after the balance has returned to zero.
Place material to be weighed B
Wait until stability display appears
(example)
Press
„ACC 2“ followed by the weight value will appear. The weight value is added into the summation memory and will be
printed out when the button is pressed
Remove material to be weighed B
Further material to be weighed can only be added after the display has returned to zero.
Add more material to be weighed as described before.
Please note that the balance must be unloaded between the individual weighing procedures.
You can repeat this process until the capacity of the scales
is exhausted.
28 GAB-P-BA-e-1411
Display of the saved weighing data:
With the weighing platform unloaded, press and the number of weighing
processes, followed by the total weight will be displayed for 2 sec and afterwards printed.
Delete weighing data:
 With the weighing platform unloaded, press 2x.
The data in the summation memory are deleted.
Printout example:
1. Weighing
No.1
1. Weighing
2013/08/10
18:39
Date/time
N:
1.000kg
Net weight
G
1.000kg
Gross weight
T:
0.000kg
Tare value
2. Weighing
No.2
2. Weighing
2013/08/10
18:39
Date/time
N:
2.000kg
Net weight
G
2.000kg
Gross weight
T:
0.000kg
Tare value
Number of weighings / total
No.2
Number weighing processes
2013/08/10
18:39
Date/time
C
3.000kg
Total
GAB-P-BA-e-1411 29
7.10 Automatic adding-up
This function is used to add the individual weighing values automatically into the summation memory and to provide a printout via the installed printer. To set this function, see chap. 8.2 „AU on“
Place material to be weighed A
After stabilisation display has taken place, you will hear an audio sound. The symbol for „AUTO“ appears.
The weighing value will be saved to the summation memory.
(example)
Remove material to be weighed A
„ACC 1“followed by the weight value will be displayed shortly
and printed.
Afterwards the weighing scales change to zero display.
Place material to be weighed B
After stabilisation display has taken place, you will hear an audio sound.
The weighing value will be saved to the summation memory.
Remove material to be weighed B
„ACC 2“followed by the total weight of both weighings will be displayed shortly. The weight of the items to be weighed B will be printed.
After that the balance changes to the zero display.
Add more material to be weighed as described before.
Please note that the balance must be unloaded between the individual weighing procedures.
You can repeat this process until the capacity of the scales
is exhausted.
30 GAB-P-BA-e-1411
Display of the saved weighing data:
With the weighing platform unloaded, press and the number of weighing
processes, followed by the total weight will be displayed for 2 sec and afterwards printed.
Delete weighing data:
 With the weighing platform unloaded, press 2x.
The data in the summation memory are deleted.
Printout example:
1. Weighing
No.1
1. Weighing
2013/08/10
18:39
Date/time
N:
1.000kg
Net weight
G
1.000kg
Gross weight
T:
0.000kg
Tare value
2. Weighing
No.2
2. Weighing
2013/08/10
18:39
Date/time
N:
2.000kg
Net weight
G
2.000kg
Gross weight
T:
0.000kg
Tare value
Number of weighing processes / total
No.2
Number weighing processes
2013/08/10
18:39
Date/time
C
3.000kg
Total
GAB-P-BA-e-1411 31
8 The menu
8.1 Navigation in the menu:
Call up menu
Press and at the same time in weighing mode
Menu block Select main menu
Using the individual menu items of the main menu can
be selected subsequently
Menu item Select submenu
Switch over into the available menu items of the submenu
using .
Confirm setting
Use to save the selected setting
Modify/change the active digit
Change to the left
Change to the right
Increase number
Return to weighing mode
Press to exit menu.
32 GAB-P-BA-e-1411
8.2 Menu overview
8.2.1 Non-verifiable models:
Menu block Main menu
Menu item Submenu
Available settings / explanation
r dUAL
not documented r duAL, 12000, 15000, 3000d, 6000d, 15000d, 30000d, 60000d
EL AU*
Background
illumination
EL Au
Background light automatic off
EL on
Background lighting on
EL off
Background lighting of f
P Prn*
Data output
P Prn
Printout by pressing the print button
P Prt
not documented
P Cont
Continuous data output
b 4800*
Baud rate
Setting options for baud rate: 600/1200/2400/4800/9600
Lab 1*
Printout
Lab 1
not documented
Lab 2
Lab 3
AU off*
Add-up mode
(automatically)
AU on
Automatic totalizing function on
AU off
Automatic totalizing function of f
ACC off*
Add-up mode
ACC off
Totalizing function off
ACC on
Totalizing function on
A2 2d
A2 2d
Zero point correction
A4 4d
A2 off
A2 0.5d
A2 1d
Ut on*
Weighing units
Ut on
Switch on or off weighing units Ut off
GAB-P-BA-e-1411 33
bP0*
Acoustic signal
bP 0
No audio sound during tolerance weighing
bP 1
Audio sound when weight is within tolerance range
bP 2
Audio sound when weight is beyond tolerance range
SPd 15*
Display speed
SPd 15
not documented
SPd 30
SPd 7.5
oF 0*
Auto off
Auto off after 0, 3, 5, 15, or 30 minutes
return
Return to weighing mode
8.2.2 Verifiable models:
Menu block Main menu
Menu item Submenu
Available settings / explanation
r dUAL
Resolution r duAL, 12000, 15000, 3000d, 6000d, 15000d, 30000d, 60000d
EL AU*
Background
illumination
EL Au
Background light automatic off
EL on
Background lighting on
EL off
Background lighting of f
P Prn*
Printout
P Prn
Printout by pressing the print button
P Prt
not documented
P Cont
Continuous data output
AU off*
Add-up mode
(automatically)
AU on
Automatic totalizing function on
AU off
Automatic totalizing function of f
E Pn E Pn
Standard printer
LP 50
not documented
ACC off*
Add-up mode
ACC off
Totalizing function off
ACC on
Totalizing function on
Ut on*
Weighing units
Ut on
Switch on or off weighing units Ut off
34 GAB-P-BA-e-1411
beeP 0*
Acoustic signal
beeP 0
No audio sound during tolerance weighing
beeP 11
Audio sound when weight is within tolerance range
beeP 2
Audio sound when weight is beyond tolerance range
SPd 15*
Display speed
SPd 15
not documented
SPd 30
SPd 7.5
oF 0*
Auto off
Auto off after 0, 3, 5, 15, or 30 minutes
return
Return to weighing mode
GAB-P-BA-e-1411 35
9 Printer
9.1 Change paper roll
To open, lift green cover and pull upwards. Insert new paper roll as shown on diagram.
Pull a short length of the end of the paper roll towards the front and close cover.
Press evenly from both sides on the cover (See arrows).
Press and the receipt will be moved on.
36 GAB-P-BA-e-1411
10 Data output RS 232C
The balance is typically equipped with a RS 232C interface.
10.1 Technical data
ASCII code 8 data bits no parity bit Baud rate 9600 baud
10.2 Pin allocation of the balance output socket (front view)
Pin 2: Output Pin 3: Input Pin 5: Signal ground
10.3 Remote control instructions
Com mand
Function
T No data are sent, the balance carries out the
tare function.
Z No data are sent, the zero-display appears.
5 4 3 2 1
9 8 7
6
GAB-P-BA-e-1411 37
11 Maintenance, servicing, removal, error messages
11.1 Clean
Before cleaning, disconnect the appliance from the operating voltage. Please do not use aggressive cleaning agents (solvents or similar agents), but a
cloth dampened with mild soap suds. Take care that the device is not penetrated by fluids and polish it with a dry soft cloth. Loose residue sample/powder can be removed carefully with a brush or manual vacuum cleaner.
Spilled weighing goods m ust be removed immediately.
11.2 Servicing, maintenance
The appliance may only be opened by trained service technicians who are authorized by KERN. Before opening, disconnect from power supply.
11.3 Disposal
Disposal of packaging and appliance must be carried out by operator according to valid national or regional law of the location where the appliance is used.
11.4 Error messages
ERR 4 Zero range exceeded
ERR 5
Invalid entry
ERR 6 Damaged electronics ERR 8 Adjustment error: wrong adjustment weight ERR 9 Instability when the balance is switched on
38 GAB-P-BA-e-1411
12 Declaration of c onf ormity
KERN & Sohn GmbH
D-72322 Balingen-Frommern
Postbox 4052
E-Mail: info@kern-sohn.de
Phone: 0049-[0]7433- 9933-0
Fax: 0049-[0]7433-9933-149
Internet: www.kern-sohn.de
Declaration of conformity
EC Declaration of Conformity EC-Declaration of -Conformity EC- Déclaration de conformité EC-Declaración de Conformidad EC-Dichiarazione di conformità EC-Conformiteitverklaring EC- Declaração de conformidade EC- Prohlášení o shode EC-Deklaracja zgodności EC-Заявление о соответствии
D
Konformitäts­erklärung
Wir erk lären hiermit, dass das Produkt, auf das sich diese Erk lärung bezieht, mit den nachstehenden Normen übereinstimmt.
EN
Declaration of conformity
We hereby declare that the product to which this declaration refers conf orms to the following standards.
CZ
Prohlášení o shode
Tímto prohlašujem e, že výrobek, kterého se toto prohlášení týká, je v souladu s níže uvedenými normami.
E
Declaración de conformidad
Manifestamos en la presente que el producto al que se refiere esta declaración está de acuerdo con las normas siguientes
F
Déclaration de conformité
Nous déclarons avec cela responsabilité que le produit, auquel se rapporte la présente déclaration, est conforme aux normes citées ci-après.
I
Dichiarazione di conformitá
Dichiariamo con ciò che il prodotto al quale la presente dichiarazione si riferisce è conforme alle norme di seguito citate.
NL
Conformiteit­verklaring
Wij verklaren hiermede dat het product, waarop deze verklaring betrekking heeft, met de hierna vermelde normen overeenstemt.
P
Declaração de conformidade
Declaramos por meio da presente que o produto no qual se refere esta declaração, corresponde às normas seguintes.
PL
Deklaracja
zgodności
Niniejszym oświadczamy, że produkt, którego niniejsze oświadczenie dotyczy,
jest zgodny z poniższymi normami.
RUS
Заявление о
соответствии
Мы заявляем, что продукт, к которому относится данная декларация,
соответствует перечисленным ниже нормам.
Electronic Balance: KERN GAB-P
EU Directive
Standards
2004/108/EC
EN 55022 : 2010
EN 61000-3-3 : 2008 EN 55024 : 2010 (EN61000-4-2: 2009, EN61000-4-3: 2006/A2 : 2010, EN61000-4-4: 2006/A1: 2010, EN61000-4-5: 2006, EN61000-4-6: 2009, EN61000-4-11: 2004)
2006/95/EC
EN 60950-1 : 2006
Signatur
Signature
Datum
Date
11.09.2013
Ort der Auss tellung
Place of issue
72336 Balingen
Albert Sauter KERN & Sohn GmbH
Manager
Managing director
KERN & Sohn GmbH, Ziegelei 1, D-72336 Balingen, Tel. +49-[0]7433/9933-0 Fax +49-[0]7433/9933-149, E-Mail: info@kern-sohn.com, Internet: www.kern-sohn.com
GAB-P-BA-e-1411 39