KERN FFN-N, FFN 3K0.5IPN, FFN 1K-4N, FFN 6K1IPN, FFN 15K2IPN Operating Instructions Manual

...
KERN & Sohn GmbH
Ziegelei 1 D-72336 Balingen email: info@kern-sohn.com
Phone: +49-[0]7433- 9933-0 Fax: +49-[0]7433-9933-149 Internet: www.kern-sohn.com
Operating instructions Compact balance
KERN FFN-N
Version 2.3
08/2013 GB
FFN-N-BA-e-1323
FFN-N-BA-e-1323 2
GB
KERN FFN-N
Version 2.3 08/2013
Operating instructions Compact balance
Table of Contents
1 Technical Data ................................................................................................ 4
1.1 Dimensions .................................................................................................................................. 6
2 Appliance Overview ........................................................................................ 7
2.1 Overview of displays .................................................................................................................... 7
2.1.1 Not verifiable models ............................................................................................................... 7
2.1.2 Verifiable models ..................................................................................................................... 7
2.2 Keyboard Overview ..................................................................................................................... 9
3 Basic Information ......................................................................................... 10
3.1 Proper use ................................................................................................................................. 10
3.2 Improper use.............................................................................................................................. 10
3.3 Warranty .................................................................................................................................... 10
3.4 Monitoring of Test Resources .................................................................................................... 10
4 Basic Safety Precautions ............................................................................. 11
4.1 Pay attention to the instructions oft the Operation Manual ....................................................... 11
4.2 Personnel training ...................................................................................................................... 11
5 Transportation & Storage............................................................................. 11
5.1 Testing upon acceptance .......................................................................................................... 11
5.2 Packaging / return transport ...................................................................................................... 11
6 Unpacking, Setup and Commissioning ...................................................... 12
6.1 Installation Site, Location of Use ............................................................................................... 12
6.2 Unpacking and implantation ...................................................................................................... 13
6.2.1 Scope of supply ..................................................................................................................... 15
6.3 Battery power supply ................................................................................................................. 15
6.3.1 Mains connection during rechargeable battery operation ..................................................... 17
6.4 Initial Commissioning ................................................................................................................. 18
6.5 Protection type IP65 .................................................................................................................. 18
7 Adjustment .................................................................................................... 18
7.1 Adjustment ................................................................................................................................. 18
7.1.1 Adjusting non-verifiable models............................................................................................. 19
7.1.2 Adjusting verifiable models .................................................................................................... 20
7.2 Verification ................................................................................................................................. 21
7.2.1 Adjustment switch and seals ................................................................................................. 22
3 FFN-N-BA-e-1323
8 Operation ....................................................................................................... 23
8.1 Weighing .................................................................................................................................... 23
8.2 Taring ......................................................................................................................................... 23
8.3 Weighing units switch-over ........................................................................................................ 24
8.4 Background illumination ............................................................................................................ 24
9 The menu ....................................................................................................... 25
9.1 Navigation in the menu .............................................................................................................. 25
9.2 Menu overview........................................................................................................................... 26
9.3 Set auto-OFF function ............................................................................................................... 27
9.4 Setting the Multi-Tare function................................................................................................... 28
10 Error messages ............................................................................................. 29
11 Instant Help ................................................................................................... 30
12 Service, maintenance, disposal ................................................................... 31
12.1 Cleaning ..................................................................................................................................... 31
12.2 Service, maintenance ................................................................................................................ 31
12.3 Disposal ..................................................................................................................................... 31
13 Declaration of –Conformity ................................ ................................ .......... 32
FFN-N-BA-e-1323 4
1 Technical Data
KERN
FFN
1K-4N
FFN
3K0.5IPN
FFN
6K1IPN
FFN
15K2IPN
FFN
25K5IPN
Weighing range (max)
1,5 kg
3 kg
6 kg
15 kg
25 kg
Readability (d)
0,2 g
0,5 g
1 g
2 g
5 g
Reproducibility
0,2 g
0,5 g
1 g
2 g
5 g
Linearity
0,4 g
1 g
1 g
4 g
10 g
Recommended adjustment weight, not added (class)
1,5 kg
(M3)
3 kg
(M3)
6 kg
(M3)
15 kg
(M3)
25 kg
(M3)
Stabilization time (typical)
2 sec.
Units
kg, lb
Warm-up time
30 min.
Electric Supply
Supply voltage: 110V-230V AC
Mains adapter: 12 V, 500 mA
Battery power supply
Operating period: Background lighting on: 30 h
Background lighting off: 50 h
Loading time: 12 h
Auto-off (Rechargeable battery)
15 min., 5 min., 3 min., off
Display mode
LCD, digit size 25 mm
Operating temperature
0° C …. + 40° C
Humidity of air
25 % - 95 % (non-condensing)
Weighing plate dimensions (stainless steel) (mm)
230 x 190
Casing dimensions
230 x 300 x 130
Total dimensions mounted (mm)
230 x 300 x 130
Weight kg (net)
3,2
IP protection
IP65
(Only during operation on battery power)
5 FFN-N-BA-e-1323
KERN
FFN
1K-4NM
FFN
3K1IPM
FFN
6K2IPM
FFN
15K5IPM
FFN
25K10IPM
Weighing range (max)
1,5 kg
3 kg
6 kg
15 kg
25 kg
Readability (d)
0,5 g
1 g
2 g
5 g
10 g
Minimum weight
10 g
20 g
40 g
100 g
200 g
Verification values (e)
1 g
1 g
2 g
5 g
10 g
Accuracy class
III
III
III
III
III
Reproducibility
1 g
1 g
2 g
5 g
10 g
Linearity
1 g
1 g
2 g
5 g
10 g
Recommended adjustment weight, not added (class)
1,5 kg
(M1)
3 kg
(M3)
6 kg
(M3)
15 kg
(M3)
25 kg
(M3)
Stabilization time (typical)
2 sec.
Units
kg, g
Anwärmzeit
10 min.
Electric Supply
Supply voltage: 110V-230V AC
Mains adapter: 12 V, 500 mA
Battery power supply
Operating period: Background lighting on: 30 h
Background lighting off: 50 h
Loading time: 12 h
Auto-off (Rechargeable battery)
15 min., 5 min., 3 min., off
Display mode
LCD, digit size 25 mm
Operating temperature
-10° C …. + 40° C
Humidity of air
25 % - 95 % (non-condensing)
Weighing plate dimensions (stainless steel) (mm)
230 x 190
Casing dimensions
230 x 300 x 130
Total dimensions mounted (mm)
230 x 300 x 130
Weight kg (net)
3,2
IP protection
IP65
(Only during operation on battery power)
FFN-N-BA-e-1323 6
1.1 Dimensions
7 FFN-N-BA-e-1323
2 Appliance Overview
2.1 Overview of displays
2.1.1 Not verifiable models
2.1.2 Verifiable models
FFN-N-BA-e-1323 8
Display
Designation
Description
Zeroing display
Should the balance not display exactly zero despite empty scale pan, press the
button. Will be reset to zero after short waiting time.
Stability display
Scales are in a steady state
Net weight display
Net weight will be displayed
Capacity display Rechargeable battery
Rechargeable battery almost empty, please connect power pack
Power supply connected
Icon is glowing for power supply via power pack
Display weighing unit kg
Displayed weight in kg
(only in not verifiable units)
Display weighing unit lb
Displayed weight in lb (only in
verifiable units)
Display weighing unit g
Displayed weight in g
9 FFN-N-BA-e-1323
2.2 Keyboard Overview
Button
Designation
Function
in the menu
ON/OFF-switch
Turn on/off
TARE button
Tare balance
Call up menu Change to the next
menu item /parameter
Zero setting key
Balance set at zero
Select menu item /
parameter
+
TARE + ZERO-key
Change units
FFN-N-BA-e-1323 10
3 Basic Information
3.1 Proper use
The balance you purchased is intended to determine the weighing value of material to be weighed. It is intended to be used as a “non-automatic“ balance, i.e. the material to be weighed is manually and carefully placed in the centre of the weighing plate. As soon as a stable weighing value is reached the weighing value can be read.
3.2 Improper use
Do not use balance for dynamic weighing. In the event that small quantities are removed or added to the material to be weighed, incorrect weighing results can be displayed due to the “stability compensation“ in the balance. (Example: Slowly draining fluids from a container on the balance.) Do not leave permanent load on the weighing plate. This may damage the measuring system. Impacts and overloading exceeding the stated maximum load (max) of the balance, minus a possibly existing tare load, must be strictly avoided. Balance may be damaged by this. Never operate balance in explosive environment. The serial version is not explosion protected. The structure of the balance may not be modified. This may lead to incorrect weighing results, safety-related faults and destruction of the balance. The balance may only be used according to the described conditions. Other areas of use must be released by KERN in writing.
3.3 Warranty
Warranty claims shall be voided in case
Our conditions in the operation manual are ignored The appliance is used outside the described uses The appliance is modified or opened Mechanical damage or damage by media, liquids, natural wear and tear The appliance is improperly set up or incorrectly electrically connected The measuring system is overloaded
3.4 Monitoring of Test Resources
In the framework of quality assurance the measuring-related properties of the balance and, if applicable, the testing weight, must be checked regularly. The responsible user must define a suitable interval as well as type and scope of this test. Information is available on KERN’s home page (www.kern-sohn.com with regard to the monitoring of balance test substances and the test weights required for this. In
KERN’s accredited DKD calibration laboratory test weights and balances may be
calibrated (return to the national standard) fast and at moderate cost.
11 FFN-N-BA-e-1323
4 Basic Safety Precautions
4.1 Pay attention to the instructions oft the Operation Manual
Carefully read this operation manual before setup and commissioning, even if you are already familiar with KERN balances.
Versions in other languages are non-binding translations. The only binding version is the original document in German.
4.2 Personnel training
The appliance may only be operated and maintained by trained personnel.
5 Transportation & Storage
5.1 Testing upon acceptance
When receiving the appliance, please check packaging immediately, and the appliance itself when unpacking for possible visible damage.
5.2 Packaging / return transport
Keep all parts of the original packaging for a possibly required return. Only use original packaging for returning. Prior to dispatch disconnect all cables and remove loose/mobile
parts.
Reattach possibly supplied transport securing devices. Secure all parts such as glass wind screen, weighing platform, power
unit etc. against shifting and damage.
FFN-N-BA-e-1323 12
6 Unpacking, Setup and Commissioning
6.1 Installation Site, Location of Use
The balances are designed in a way that reliable weighing results are achieved in common conditions of use. You will work accurately and fast, if you select the right location for your balance.
Therefore, observe the following for the installation site:
Place scales on a stable, even surface; Avoid extreme heat as well as temperature fluctuation caused by installing
next to a radiator or in the direct sunlight;
Protect the balance against direct draughts due to open windows and
doors;
Avoid jarring during weighing; Protect the balance against high humidity, vapors and dust; Do not expose the device to extreme dampness for longer periods of time.
Non-permitted condensation (condensation of air humidity on the appliance) may occur if a cold appliance is taken to a considerably warmer environment. In this case, acclimatize the disconnected appliance for ca. 2 hours at room temperature.
Avoid static charge of goods to be weighed or weighing container.
Major display deviations (incorrect weighing results) may be experienced should electromagnetic fields (e.g. due to mobile phones or radio equipment), static electricity accumulations or instable power supply occur. Change location or remove source of interference.
13 FFN-N-BA-e-1323
6.2 Unpacking and implantation
Carefully remove the balance from the packaging, remove plastic cover and setup balance at the intended workstation.
2
1
3
5
4
6
7
8
9
1
Weighing plate
2
Locking screw
3
Carrier weighing plate
4
Display
5
Bubble level
6
Keyboard
7
Footscrews
8
Rechargeable battery compartment cover
9
Screws of battery compartment cover
FFN-N-BA-e-1323 14
Level balance with foot screws until the air bubble of the water balance is in the prescribed circle.
Pressure compensation screw:
[1]
On the lower side of the balance there is a pressure compensation screw [1] which has to be opened for the weighing process. Fasten the screw before cleaning the balance.
Mains connection
Battery compartment
Pressure compensation screw
15 FFN-N-BA-e-1323
6.2.1 Scope of supply Serial accessories:
Balance Weighing plate Carrier weighing plate Fastening screw of the carrier weighing plate Operating instructions Rechargeable battery Hexagonal socket wrench
6.3 Battery power supply The rechargeable battery is charged via the delivered power supply.
The operating time of the rechargeable battery with background illumination is 30h, without that it is 50 hrs. Charging time until complete recharging 12h. In the menu you can activate the AUTO-OFF function time off], see chap. 9.3. According to menu settings, the balance switches automatically off in order to spare the battery.
Insert rechargeable battery:
Remove any adherent moisture from
the balance
Move both levers 90° to the left
Membrane for degree of protection IP65
FFN-N-BA-e-1323 16
Remove the rechargeable battery
cover
Connecting the rechargeable battery
Watch the colours: Red on red ! Black on black!
Insert the rechargeable battery in a
manner that it cannot slip (fasten with a rubber mat)
Ensure that the cables are not squeezed.
17 FFN-N-BA-e-1323
Close the rechargeable battery cover Move both levers 90° to the right
If in the display the triangle appears above the capacity display , the rechargeable battery is nearly exhausted. Connect power pack, the rechargeable battery is loaded.
Avoid excess pressure on the balance in order to avoid damaging them, particularly as they are supported by the weighing plate.
6.3.1 Mains connection during rechargeable battery operation
During the rechargeable battery operation ensure that the mains connection is covered with a rubber cap.
Only by this the type of protection IP65 is ensured.
FFN-N-BA-e-1323 18
6.4 Initial Commissioning
In order to obtain exact results with the electronic balances, your balance must have reached the operating temperature (see warming up time chap. 1). The weighing scale must be connected to the power supply (batteries) during this warm-up time. The accuracy of the balance depends on the local acceleration of gravity. Strictly observe hints in chapter Adjustment.
6.5 Protection type IP65
Designed for temporary contact with liquids. Use a damp cloth for cleaning. Dustproof.
IP65 protection is only ensured during operation on battery power
7 Adjustment
As the acceleration value due to gravity is not the same at every location on earth, each balance must be coordinated - in compliance with the underlying physical weighing principle - to the existing acceleration due to gravity at its place of location (only if the balance has not already been adjusted to the location in the factory). This adjustment process must be carried out for the first commissioning, after each change of location as well as in case of fluctuating environment temperature. To receive accurate measuring values it is also recommended to adjust the balance periodically in weighing operation.
7.1 Adjustment
Carry out adjustment as near as possible to the maximum load of the scale (See chpt. 1 “Techn. data”). The accuracy of the weight used for adjustment has to correspond approximately to readability d of the scale, or rather closer than that. Info about test weights can be found on the Internet at: http://www.kern-sohn.com
Procedure when adjusting:
Observe stable environmental conditions. A warming up time (see chapter 1) is required for stabilization.
19 FFN-N-BA-e-1323
7.1.1 Adjusting non-verifiable models
Display
Operation
Start balance by pressing Whilst the weighing scale is carrying out a self-test,
(00…- 99…) press and hold until the message “F1 CAL“ appears.
Press ,
in the display appears „UnLod“.
(Example)
Press again
the weight value will be displayed.
Enter value for adjustment weight (See chpt 1) by Position weight
Press
Remove weight whilst scale is carrying out self-test Weighing scale changes to zero display. This completes the adjustment process.
FFN-N-BA-e-1323 20
7.1.2 Adjusting verifiable models
The adjustment is locked for verified balances.
with adjustment switch
To disable the access lock, destroy the seal and actuate the adjustment switch. Position of the adjustment switch see chapter 7.2.1
Attention: After destruction of the seal the balance must be re-verified by an authorised agency and a new verification wire/seal mark fitted before it can be reused for applications subject to verification.
Display
Operation
 Start balance by pressing While the balance carries out a self test (00…- 99…), press
and keep pressed until „F1 CAL“ appears on the
display.
Actuate the adjustment switch on the lower side of the
balance
Press
in the display appears „UnLod“.
(Beispiel)
Press anew
The weighing value is displayed.
Press to enter the value of the adjustment weight
(see chapter 1)
Place the adjustment weight Press
While the balance carries out a self-test, take away the adjustment weight
The balance changes over to zero display. Now the process of adjustment is completed.
21 FFN-N-BA-e-1323
7.2 Verification
General introduction:
According to EU directive 90/384/EEC or 2009/23EG balances must be officially verified if they are used as follows (legally controlled area):
a) For commercial transactions if the price of goods is determined by weighing. b) For the production of medicines in pharmacies as well as for analyses in the
medical and pharmaceutical laboratory. c) For official purposes d) For manufacturing final packages
In cases of doubt, please contact your local trade in standard. Verification notes:
An EU type approval exists for balances described in their technical data as verifiable. If a balance is used where obligation to verify exists as described above, it must be verified and re-verified at regular intervals. Re-verification of a balance is carried out according to the respective national regulations. The validity for verification of balances in Germany is e.g. 2 years. The legal regulation of the country where the balance is used must be observed!
Verification of the balance is invalid without the seal.
The seal marks attached on verified balances point out that the balance may only be opened and serviced by trained and authorised specialist staff. If the seal mark is destroyed, verification looses its validity. Please observe all national laws and legal regulations. In Germany a re-verification will be necessary.
FFN-N-BA-e-1323 22
7.2.1 Adjustment switch and seals
After a verification the balance is sealed at the indicated positions.
Position of the official seals:
1. Verification wire fastening
2. Verification wire fastening
3. Self-destroying seal mark
4. Cover of adjustment switch
5. Adjustment switch
6. Self-destroying seal mark
23 FFN-N-BA-e-1323
8 Operation
8.1 Weighing
Start balance by pressing
Program version is displayed. Afterwards the balance will carry out a self-test. The balance will be ready for operation as soon as weight reading “0.0” and the triangular icon above the stability display appear.
The key can be used to set the balance to zero at any
time.
Turn off scale by pressing .
The “0.0” display disappears and the weighing scale is now switched off.
8.2 Taring
The tare weight of any preloads can be deducted by pressing a button so that the actual weight of the baby is displayed in subsequent weighings.
Put on weighing receptacles and press .
The zero display appears as well as the triangle above the zero setting icon as well as the stability icon and the net weight icon . The weight of the container is now internally saved.
Place goods to be weighed in the weighing container.
The net weight of the goods to be weighed is displayed.
The weight of the weighing container will be displayed as a minus number after removing the weighing container (= gross weight).
FFN-N-BA-e-1323 24
The tare weight is saved until it is deleted. Remove the load from the balance and press . The zero display appears
and the triangle above the net weight icon disappears.
Gross weight:
Press as long as the weighing tray and the load are
present on the weighing platform.
Remove weighing goods and receptacle
Gross weight will be shown as a negative value.
8.3 Weighing units switch-over
Press and at the same time.
The weighing scale switches between the kg and lb units. (for not verifiable devices) or kg and g (in verifiable devices).
8.4 Background illumination
Press for approx. 3 seconds, „bK-AU“ will
be displayed.
Press to select background illumination
automatically off („bK-AU“) and background illumination off („bK-oF“).
Press to confirm selected setting.
The balance changes into weighing mode
25 FFN-N-BA-e-1323
9 The menu
9.1 Navigation in the menu
Start balance by pressing
While the balance carries out a self test, press and
keep pressed until „F1 CAL“ appears
Use to change to the next menu item
Select menu item using
Use to change between the individual parameters
Select the desired parameter using
FFN-N-BA-e-1323 26
9.2 Menu overview
Menu item
Function
Calibrate
not documented
not documented
not documented
Auto-OFF function, adjustable between off, 3, 5 and 15 minutes
not documented
Multi-Tare function: 0 tare Multi-Tare function disabled P tare Multi-Tare function enabled
not documented
Return to weighing mode
27 FFN-N-BA-e-1323
9.3 Set auto-OFF function
Start balance by pressing
While the balance carries out a self test, press and
keep pressed until „F1 CAL“ appears
Press repeatedly until „F5 oFF“ appears
Select function using
Use to change between the individual parameters
Select the desired parameter using
“F5 oFF“ appears, now the function is enabled
Press repeatedly until “bACK“ appears
Press
The balance carries out a restart and changes to weighing mode
FFN-N-BA-e-1323 28
9.4 Setting the Multi-Tare function
Start balance by pressing
While the balance carries out a self test, press and
keep pressed until „F1 CAL“ appears
Press repeatedly until „F8 tn“ appears For verifiable devices press the adjustment switch
Select function using , the function adjusted as last will
be displayed (PtArE or 0tArE)
Use to change between the individual parameters
Use to select „PtAre“
“F8 tn“ appears, now the function is enabled
Press repeatedly until “bACK“ appears
Press
The balance carries out a restart and changes to weighing mode
Now you have the ability to tare several times in a row.
29 FFN-N-BA-e-1323
10 Error messages
Display
Description
Remedy
Zero range exceeded
Unload the balance
A/D converter outside range
Unload the balance; check if the weighing is correctly placed and screwed on
Displayed weight is permanently changing
Avoid air draughts / air movement as well as vibrations of the table and of the floor
Overload
Unload the balance and adjust again
FFN-N-BA-e-1323 30
11 Instant Help
In case of an error in the program process, briefly turn off the balance and disconnect from power supply. The weighing process must then be restarted from the beginning.
Fault
Possible cause
The displayed weight does not glow.
The balance is not switched on.
Batteries are inserted incorrectly or empty
No batteries inserted.
The displayed weight is permanently changing
Draught/air movement Table/floor vibrations
Weighing plate has contact with other
objects.
Electromagnetic fields / static charging
(choose different location/switch off interfering device if possible)
The weighing value is obviously
The display of the balance is not at zero
wrong
Adjustment is no longer correct.
Great fluctuations in temperature. The balance is on an uneven surface. Electromagnetic fields / static charging
(choose different location/switch off interfering device if possible)
Should other error messages occur, switch balance off and then on again. If the error message remains inform manufacturer.
31 FFN-N-BA-e-1323
12 Service, maintenance, disposal
12.1 Cleaning
Remove batteries from instrument before cleaning.
Please do not use aggressive cleaning agents (solvents or similar agents), but a cloth dampened with mild soap suds. Ensure that no liquid penetrates into the device and wipe with a dry soft cloth. Loose residue sample/powder can be removed carefully with a brush or manual vacuum cleaner.
Spilled weighing goods must be removed immediately.
12.2 Service, maintenance
The appliance may only be opened by trained service technicians who are authorized by KERN. Before opening, disconnect from power supply.
12.3 Disposal
Disposal of packaging and appliance must be carried out by operator according to valid national or regional law of the location where the appliance is used.
FFN-N-BA-e-1323 32
13 Declaration of –Conformity
KERN & Sohn GmbH
D-72322 Balingen-Frommern Postfach 4052 E-Mail: info@kern-sohn.de
Tel: 0049-[0]7433- 9933-0 Fax: 0049-[0]7433-9933-149 Internet: www.kern-sohn.de
Declaration of -Conformity
EG-Konformitätserklärung EC-Declaration of -Conformity EC- Déclaration de conformité EC-Declaración de Conformidad EC-Dichiarazione di conformità EC-Conformiteitverklaring EC- Declaração de conformidade EC- Prohlášení o shode EC-Deklaracja zgodności EC-Заявление о соответствии
D
Konformitäts­erklärung
Wir erklären hiermit, dass das Produkt, auf das sich diese Erklärung bezieht, mit den nachstehenden Normen übereinstimmt.
GB
Declaration of conformity
We hereby declare that the product to which this declaration refers conforms with the following standards.
CZ
Prohlášení o
shode
Tímto prohlašujeme, že výrobek, kterého se toto prohlášení týká, je v souladu s níže uvedenými normami.
E
Declaración de conformidad
Manifestamos en la presente que el producto al que se refiere esta declaración está de acuerdo con las normas siguientes
F
Déclaration de conformité
Nous déclarons avec cela responsabilité que le produit, auquel se rapporte la présente déclaration, est conforme aux normes citées ci-après.
I
Dichiarazione di conformitá
Dichiariamo con ciò che il prodotto al quale la presente dichiarazione si riferisce è conforme alle norme di seguito citate.
NL
Conformiteit­verklaring
Wij verklaren hiermede dat het product, waarop deze verklaring betrekking heeft, met de hierna vermelde normen overeenstemt.
P
Declaração de conformidade
Declaramos por meio da presente que o produto no qual se refere esta declaração, corresponde às normas seguintes.
PL
Deklaracja
zgodności
Niniejszym oświadczamy, że produkt, którego niniejsze oświadczenie dotyczy, jest zgodny z poniższymi normami.
RUS
Заявление о
соответствии
Мы заявляем, что продукт, к которому относится данная декларация,
соответствует перечисленным ниже нормам.
Electronic Balances: KERN FFN_N, FFN_M
EU Directive
Standards
2004/108/EC
EN55022: 2006 A1:2007 EN61000-3-3:1995+A1:2001+A2:2005 EN55024: 1998+A1:2001+A2:2003
2006/95/EC
EN 60950-1:2006 EN 60065:2002+A1:2006
Signatur
Signature
Datum
Date
08.04.2013
Ort der Ausstellung
Place of issue
72336 Balingen
Albert Sauter KERN & Sohn GmbH
Geschäftsführer
Managing director
KERN & Sohn GmbH, Ziegelei 1, D-72336 Balingen, Tel. +49-[0]7433/9933-0 Fax +49-[0]7433/9933-149, E-Mail: info@kern-sohn.com, Internet: www.kern-sohn.com
Loading...