KERN ALT Operating Instructions Manual

ALT-BA-defsi-0410
69
GB
KERN ALT
Version 1.0 12/2004
Table of contents
1 Technical data _________________________________________________ 71
2 Declaration of conformity _______________________________________ 72
3 Fundamental information (general)________________________________ 74
3.1 Designated use __________________________________________________ 74
3.2 Inappropriate use ________________________________________________ 74
3.3 Guarantee ______________________________________________________ 74
3.4 Monitoring the test substances_____________________________________ 75
4 Fundamental safety instructions__________________________________ 75
4.1 Observe notices in the operating manual ____________________________ 75
4.2 Training of personnel_____________________________________________ 75
5 Transport and storage __________________________________________ 75
5.1 Acceptance check _______________________________________________ 75
5.2 Packing ________________________________________________________ 75
6 Unpacking, installation and commissioning ________________________ 76
6.1 Installation Site, Location of Use ___________________________________ 76
6.2 Unpack_________________________________________________________ 76
6.2.1 Assembly ___________________________________________________________ 77
6.2.2 Scope of delivery _____________________________________________________ 77
6.3 Mains supply____________________________________________________ 78
6.4 Connection of peripheral equipment ________________________________ 78
6.5 Initial Commissioning ____________________________________________ 78
7 User menu ____________________________________________________ 79
7.1 User principle of the menu control __________________________________ 80
7.2 Operating the balance with PS/2 keyboard (see chapter 6.3)_____________ 81
7.3 Menu selection __________________________________________________ 82
7.3.1 P1 adjustment________________________________________________________ 82
7.3.2 P2 GLP (Good laboratory practice) _______________________________________ 90
7.3.3 P3 Date/time_________________________________________________________ 92
7.3.4 P4 Readout__________________________________________________________ 94
7.3.5 P5 RS-232 __________________________________________________________ 96
7.3.6 P6 Printouts _________________________________________________________ 96
7.3.7 P7 Weighing units____________________________________________________ 101
7.3.8 P8 operating modes __________________________________________________ 101
7.3.9 P9 General functions _________________________________________________ 102
ALT-BA-defsi-0410
70
8 Operation ____________________________________________________ 103
8.1 Operating modes (P8 modes) _____________________________________ 105
8.1.1 Simple weighing _____________________________________________________ 107
8.1.2 Parts counting_______________________________________________________ 109
8.1.3 Weighing with tolerance range __________________________________________ 117
8.1.4 Filling _____________________________________________________________ 119
8.1.5 Percent determination ________________________________________________ 120
8.1.6 Animal weighing (Dynamic weighing)_____________________________________ 122
8.1.7 Density determination_________________________________________________ 124
8.1.8 Formulation_________________________________________________________ 125
8.1.9 Statistics ___________________________________________________________ 129
9 Data output __________________________________________________ 131
9.1 Tecnical data___________________________________________________ 131
9.2 Pin allocation of the balance exit plug (front view) ____________________ 131
9.3 Remote commands _____________________________________________ 132
10 Maintenance, upkeep, disposal ________________________________ 133
10.1 Cleaning ______________________________________________________ 133
10.2 Maintenance, upkeep ____________________________________________ 133
10.3 Disposal_______________________________________________________ 133
11 Troubleshooting ____________________________________________ 134
ALT-BA-defsi-0410
71
1 Technical data
KERN ALT 100-4M ALT 160-5D ALT 220-4M ALT 310-4
Readout (d) 0.1 mg 0.1mg/0.01mg 0.1 mg 0.1 mg
Weighing range (Max) 100 g 160 g/60 g 220 g 310 g
Taring range (subtractive) 100 g 160 g/60 g 220 g 310 g
Reproductability 0.2 mg 0.2mg/0.1mg 0.2 mg 0.2 mg
Linearity ± 0.2 mg ± 0,2mg/0,1mg ± 0.2 mg ± 0.3 mg
Minimal piece weight for counting function
0,2 mg 0,2 mg 0,2 mg 0,2 mg
Adjustment weight internal internal internal internal
Verifiable yes no yes no
Verification value (e) 1 mg - 1 mg -
Accuracy category I - I -
Reference piece numbers for parts counting
10, 20, 50,100, freely selectable
Weighing Units mg, g, ct,
mg, g, ct, lb, oz, ozt, dwt,
t {h}, t {S}, t {t}, mom,
GN, N, t,
freely pro-
grammable
Stabilization time (typically) 4 sec.
Permissible ambient temperature
+ 15° C .... +30° C
Humidity of air: max. 80 % (not condensing)
Under floor weighing device On-hook eyelets, serial
Weighing plate mm 100 80 100
Housing (W x D x H) mm 205 x 500 x 290
Weight kg (net) 8,9
ALT-BA-defsi-0410
72
2 Declaration of conformity
KERN & Sohn GmbH
D-72322 Balingen-Frommern Postfach 4052 E-Mail: info@kern-sohn.de
Tel: 0049-[0]7433- 9933-0 Fax: 0049-[0]7433-9933-149 Internet: www.kern-sohn.de
Konformitätserklärung
Declaration of conformity for apparatus with CE mark
Konformitätserklärung für Geräte mit CE-Zeichen
Déclaration de conformité pour appareils portant la marque CE
Declaración de conformidad para aparatos con marca CE
Dichiarazione di conformitá per apparecchi contrassegnati con la marcatura CE
English
We hereby declare that the product to which this declaration refers conforms with the following standards.
Deutsch
Wir erklären hiermit, dass das Produkt, auf das sich diese Erklärung bezieht, mit den nachstehenden Normen übereinstimmt.
Français
Nous déclarons avec cela responsabilité que le produit, auquel se rapporte la présente déclaration, est conforme aux normes citées ci-après.
Español
Manifestamos en la presente que el producto al que se refiere esta declaración est´´a de acuerdo con las normas siguientes
Italiano
Dichiariamo con ciò che il prodotto al quale la presente dichiarazione si riferisce è conforme alle norme di seguito citate.
Electronic Balance: KERN ALT
Mark applied EU Directive Standards
89/336EEC EMC EN 61000-4-2 :1999
EN 61000-4-3 :1996 EN 61000-4-4 : 1999 EN 61000-4-5 : 1998 EN 61000-4-6 : 1999 EN 61000-4-11 : 1997 EN 55022 :2000
Date: 18.01.2005 Signature:
Gottl. KERN & Sohn GmbH Management
Gottl. KERN & Sohn GmbH, Ziegelei 1, D-72336 Balingen, Tel. +49-[0]7433/9933-0,Fax +49-[0]7433/9933-149
ALT-BA-defsi-0410
73
KERN & Sohn GmbH
D-72322 Balingen-Frommern Postfach 4052 E-Mail: info@kern-sohn.de
Tel: 0049-[0]7433- 9933-0 Fax: 0049-[0]7433-9933-149 Internet: www.kern-sohn.de
Konformitätserklärung
Declaration of conformity for apparatus with CE mark
Konformitätserklärung für Geräte mit CE-Zeichen
Déclaration de conformité pour appareils portant la marque CE
Declaración de conformidad para aparatos con marca CE
Dichiarazione di conformitá per apparecchi contrassegnati con la marcatura CE
English
We hereby declare that the product to which this declaration refers conforms with the following standards.
This declaration is only valid with the certificate of conformity by a notified body.
Deutsch
Wir erklären hiermit, dass das Produkt, auf das sich diese Erklärung bezieht, mit den nachstehenden Normen übereinstimmt.
Diese Erklärung gilt nur in Verbindung mit der Konformitätsbescheinigung einer benannten Stelle.
Français
Nous déclarons avec cela responsabilité que le produit, auquel se rapporte la présente déclaration, est conforme aux normes citées ci-après.
Cette déclaration est valide seulement avec un certificat de conformité d’un organisme notifié.
Español
Manifestamos en la presente que el producto al que se refiere esta declaración está de acuerdo con las normas siguientes.
Esta declaración solo será válida acompañada del certificado de conformidad de conformidad de la parte nominal.
Italiano
Dichiariamo con ciò che il prodotto al quale la presente dichiarazione si riferisce è conforme alle norme di seguito citate.
Questa dichiarazione sarà valida solo se accompagnata dal certificato di conformità della parte nominale.
Model: KERN ALT 100-4M
KERN ALT 220-4M
EU Directive Standards
EC-type-approval certificate no.
Issued by
90/384/EEC EN 45501 T6655
NMI
Date: 18.01.2005 Signature:
Gottl. KERN & Sohn GmbH Management
Gottl. KERN & Sohn GmbH, Ziegelei 1, D-72336 Balingen, Tel. +49-07433/9933-0,Fax +49-074433/9933-149
ALT-BA-defsi-0410
74
3 Fundamental information (general)
3.1 Designated use
The balance you have acquired serves to determine the weighing value of the material to be weighed. It is intended to be used as a “non-automatic“ balance, i.e. the material to be weighed is manually and carefully placed in the centre of the weighing plate. The weighing value can be read off after a stable weighing value has been obtained.
3.2 Inappropriate use
Do not use balance for dynamic add-on weighings, if small amounts of goods to be weighed are removed or added. The “stability compensation“ installed in the balance may result in displaying an incorrect measuring value! (Example: Slowly draining fluids from a container on the balance.) Do not leave a permanent load on the weighing plate. This can damage the measuring system.
Be sure to avoid impact shock and overloading the balance in excess of the prescribed maximum load rating (max.), minus any possible tare weight that is already present. This could cause damage to the balance. Never operate the balance in hazardous locations. The series design is not explosion-proof. Structural alterations may not be made to the balance. This can lead to incorrect weighing results, faults concerning safety regulations as well as to destruction of the balance. The balance may only be used in compliance with the described guidelines. Varying areas of application/planned use must be approved by KERN in writing.
3.3 Guarantee
The guarantee is not valid in the following cases:
non-observation of our guidelines in the operating instructions
use outside the described applications
alteration to or opening of the device
mechanical damage and damage caused by media, liquids,
natural wear and tear
inappropriate erection or electric installation
overloading of the measuring equipment
ALT-BA-defsi-0410
75
3.4 Monitoring the test substances
The metrology features of the balance and any possible available adjusting weight must be checked at regular intervals within the scope of quality assurance. For this purpose, the answerable user must define a suitable interval as well as the nature and scope of this check. Information is available on KERN’s home page (www.kern-
sohn.com) with regard to the monitoring of balance test substances and the test
weights required for this. Test weights and balances can be adjusted quickly and at a reasonable price in KERN’s accredited DKD calibration laboratory (return to national normal).
4 Fundamental safety instructions
4.1 Observe notices in the operating manual
Please read the operating instructions carefully before erecting and commissioning, even if you already have experience with KERN balances.
4.2 Training of personnel
The device may only be operated and looked after by trained members of staff.
5 Transport and storage
5.1 Acceptance check
Please check the packaging immediately upon delivery and the device during unpacking for any visible signs of external damage.
5.2 Packing
Please retain all parts of the original packaging in case it should be necessary to return items at any time. Only the original packaging should be used for return consignments. Before despatch, disconnect all attached cables and loose/movable parts. Apply any intended transport security devices. Secure all parts e.g. weighing plate, mains power supply etc. against slipping and damage.
ALT-BA-defsi-0410
76
6 Unpacking, installation and commissioning
6.1 Installation Site, Location of Use
The balance is constructed in such a way that reliable weighing results can be achieved under normal application conditions. By selecting the correct location for your balance, you will be able to work quickly and precisely.
Therefore please observe the following at the place of installation:
Place the balance on a firm, level surface;
Avoid extreme heat as well as temperature fluctuation caused by installing
next to a radiator or in the direct sunlight;
Protect the balance against direct draughts due to open windows and doors;
Avoid jarring during weighing;
Protect the balance against high humidity, vapours and dust;
Do not expose the device to extreme dampness for longer periods of time.
Inadmissible bedewing (condensation of air moisture on the device) can occur if a cold device is taken into a significantly warmer environment. In this case, please acclimatise the device for approx. 2 hours at room temperature after it has been disconnected from the mains.
Avoid static charging of items to be weighed, or weighing container.
Major display deviations (incorrect weighing results) are possible if electromagnetic fields occur as well as due to static charging, currents and instable power supply. It is then necessary to change the location.
6.2 Unpack
Carefully remove the balance from its packaging, remove the plastic wrapping and position the balance in its intended working location.
ALT-BA-defsi-0410
77
6.2.1 Assembly
Structure of the balance:
Use the foot screws to level the balance until the air bubble in the bubble level is in the prescribed circle.
6.2.2 Scope of delivery
Standard accessories:
Scales
Weighing plate
Network appliance
Instruction manual
Wind protection
(1) Base plate
(2) Screen ring
(3) Weighing plate
ALT-BA-defsi-0410
78
6.3 Mains supply
Electric power supply is by means of the external mains supply circuit. The printed voltage level must comply with the local voltage. Only use original KERN mains supply circuits. The use of other makes is subject to approval by Kern.
1. Mains supply
2. PS/2 keyboard
3. RS232 interface
6.4 Connection of peripheral equipment
The balance must be disconnected from the mains before connecting or disconnecting additional equipment (printer, PC) to or from the data interface. Only use KERN accessories and peripheral equipment with your balance. These have been ideally coordinated to your balance.
6.5 Initial Commissioning
A warm-up time of 1 hour stabilises the measured values after switching on. The accuracy of the balance depends on the local acceleration of the fall. Instructions in chapter Adjustment must be observed.
ALT-BA-defsi-0410
79
7 User menu
The user menu has nine main menus with the following sub-menus:
P1 Calibration
P2 GLP P3 Date/Time P4 Readout P5 RS-232 P6 Printouts P7 Units P8 Modes P9 Globals
Int. calibr. Ext. calibr User calibr. Calibr. Test: Weight.corr. Aut. calibr. Print report
User No. Prj. No. Time print Date print UserNo. print Prj. No. print Id print Diff. print
Date format Time format Time Date Disp. time Disp. date
Filter Filter range Disp refresh Stable range Stable speed Autozero Last digit
Boud rate Parity Data bits Stop bits Handshake Auto print Interval Print on stab
Printout No. Pr. 1 start Pr. 1 stop Pr. 2 start Pr. 2 stop Pr. 3 start Pr. 3 stop Pr. 4 start Pr. 4 stop String 1 String 2 String 3 . . . String 78 String 79 String 80
Grams
Miligrams Kilograms Pounds Ounces, Ounces troy Carats Dwt Taels. Hk Taels. S Taels. T Mommes Granis Newtons Tical Custom Custom Factor
Parts Count Checkweigh ing Filling Percont Anima l Formulation Statistics Density
Beep Language Backlight Contrast Screensaver Temperature Balance Id Software rev. Par. Printout Par. Receive. Factory deff.
Main Menu
1. Menu number (P1 – P9)
2. Menu designation
3. Cursor () indicates the current menu item
Submenu:
1. Menu number (P1 – P9)
2. Parameter number and designation
3. Parameter value or ****** i.e. this parameter is available as a function
4. Functional Description
5. Cursor () indicates the current menu item (e.g. P1).
6. Cursor () indicates the current parameter (e.g. 06).
7. ”Setup“i.e. the user is in Setup Mode.
8. Use the arrow key to modify the parameter value, the active parameter blinks.
ALT-BA-defsi-0410
80
7.1 User principle of the menu control
In the menu the settings of the balance can be modified and functions can be activated. This way, the balance can be adjusted to individual weighing requirements. To do so proceed as follows:
Ö Use the -key to turn on the balance.
Ö After actuating the -key the display shows the main menus that can be set
P1 Calibration P2 GLP P3 Date/Time P4 Readout P5 RS-232 P6 Printouts P7 Units P8 Modes P9 Global
s
Setup2.02.05
Ö Select your setting using the cursor (►).
Use the -key to move the cursor () down, the -key to move it up
Ö The -key confirms your selected setting, the display shows the sub-menu or
the active menu item blinks:
Setup
P1
01 Int. calibr. 02 Ext. calib r. 03 User calibr. 04 Calib r. test 05 Weig ht. co rr. 06 Aut calibr. 07
Print report
o
Auto cal. time
08
******** ******** ******** ********
0.0 0
1
function
function function
off
on
function
31.01.05
1
1 hour
Ö Use the arrow keys to move within and enter into the menu:
Cursor control up
Parameter selection within a function
Increasing the value for the selected numeral or number
Access to the main
menu
Access to the sub-menu
In the menu, one menu
item to the right
Selecting the numeral to be modified to the right (the respective active place blinks)
Return to main menu
Selecting the numeral to
be modified to the left (the respective active place blinks)
Cursor control down
Parameter selection within a function
Decreasing the value for the selected numeral or number
ALT-BA-defsi-0410
81
Ö Use the -key to confirm or the -key to cancel
Ö Use the -key to return to the menu
Ö Actuate -key again, a message will inquire whether you want to save the
changes you made.
Ö Actuating the -key will save the selected setting. Ö If you do not want to save the selected setting press
Ö The balance will automatically return to weighing mode
Weighing
0.0000
g
0% 100%
0
27.01.05
7.2 Operating the balance with PS/2 keyboard (see chapter 6.3)
Using a PS/2 keyboard will help you to enter numbers and texts faster and easier.
Key allocation:
Balance keyboard
PS keyboard
ALT-BA-defsi-0410
82
7.3 Menu selection
7.3.1 P1 Calibration (Adjustment)
As the acceleration value due to gravity is not the same at every location on earth, each balance must be coordinated - in compliance with the underlying physical weighing principle - to the existing acceleration due to gravity at its place of location ( only if the balance has not already been adjusted to the location in the factory). This adjustment process must be carried out for the first commissioning, after each change of location as well as in case of fluctuating environment temperature. It is also recommendable to adjust the balance periodically during weighing operation in order to obtain exact measured values.
Using the inbuilt adjusting weight the balance automatically carries out an adjustment after any temperature change of 0.8 °C or any time it was disconnected from the power supply. Of course it is also possible to check the accuracy of the balance manually any time. You can choose between two options:
1. Adjustment is started by pressing the -key three times. After successful adjustment the balance automatically returns to weighing mode.
2. Under menu item “P1 01 Calibration with internal weight” (see table below)
Procedure for adjustment:
Observe stable environment conditions. A warming-up time of ca. 1 hour for stabilisation is necessary.
Important! During the adjustment process no objects should be on the weighing plate.
Operator Indication
Ö Call up menu item “P1 Calibration “
(see chapter 7.1)
Ö
Press -key
P1 Calibration P2 GLP P3 Date/Time P4 Readout P5 RS-232 P6 Printouts P7 Units P8 Modes P9 Globals
Setup2.02.05
01 Int. calibr. - Adjustment with internal weight
Ö Use the cursor () to select “01 Int. calibr. “
Setup
P1
01 Int. calib r. 02 Ext. calibr. 03 User calib r. 04 Calibr. test 05 Weigh t. corr. 06 Aut calibr. 07
Print report
o
Auto cal. time
08
******** ******** ******** ********
0.0 0
1
function
function function
off
on
function
31.01.05
1
1 hour
Ö Press the -key; the adjustment with internal
weight is carried out.
Calibration
Internal calibration
ALT-BA-defsi-0410
83
Ö After successful adjustment the balance
automatically returns to weighing mode. In case of an adjustment error (e.g. objects on the weighing plate) the display will show an error message, repeat adjustment.
Weighing
0.0000
g
0% 100%
0
27.01.05
02 Ext. Cal - Adjustment with an external weight
Locked for verifiable appliances
03 User cal. – locked for verifiable appliances
04 Cal. test – not documented
05 Weight corr. – locked for verifiable appliances
06 Autom. cal. – locked for verifiable appliances
07 Auto. Cal.time – locked for verifiable appliances
08 Print report – printout of the adjustment data
Ö Use the cursor () to select “08 “print report“
Ö Press the –key; the current menu item blinks.
Use the arrow keys (see chapter 7.1) to save your settings
0 Print report deactivated 1 Print report activated
Setup
P1
01 Int. calib r. 02 Ext. calib r. 03 User ca libr. 04 Calibr. test 05 Weig ht. cor r. 06 Aut calibr. 07
Print report
o
Auto cal. time
08
******** ******** ******** ********
0.0 0
1
function
function function
off
on
function
31.01.05
1
1 hour
Ö Confirm with -key
Ö When print report is activated you will receive a printout of your adjustment data
after each adjustment (Ex.):
*** Internal calibration report ***
Calibr. : internal
Diff.. : 0.0038 g
Signature………………………….
ALT-BA-defsi-0410
84
7.3.1.1 Calibration (adjustment) only for model KERN ALT 310
Operator indication
Ö Call up menu item “P1 Calibration“
(see chapter 7.1)
Ö Press -key
P1 Calibration P2 GLP P3 Date/Time P4 Readout P5 RS-232 P6 Printouts P7 Units P8 Modes P9 Globals
Setup2.02.05
01 Int. cal. - Adjustment with internal weight
Ö Use the cursor () to select “01 Int. cal.“
Setup
P1
01 Int. calib r. 02 Ext. calib r. 03 User ca libr. 04 Calibr. test 05 Weig ht. cor r. 06 Aut calibr. 07
Print report
o
Auto cal. time
08
******** ******** ******** ********
0.0 0
1
function
function function
off
on
function
31.01.05
1
1 hour
Ö Press the key; the adjustment with internal
weight is carried out.
Calibration
Internal calibration
Ö After successful adjustment the balance
automatically returns to weighing mode. In case of an adjustment error (e.g. objects on the weighing plate) the display will show an error message, repeat adjustment.
Weighing
0.0000
g
0% 100%
0
27.01.05
02 Ext. Cal - Adjustment with an external weight
Ö Use the cursor () to select “02 Ext. cal.“
Setup
P1
01 Int. calib r. 02 Ext. calib r. 03 User ca libr. 04 Calibr. test 05 Weig ht. cor r. 06 Aut calibr. 07
Print report
o
Auto cal. time
08
******** ******** ******** ********
0.0 0
1
function
function function
off
on
function
31.01.05
1
1 hour
Ö Press -key
Ensure that there are no objects on the weighing plate.
Ö Press -key; the display shows the measuring
value of the adjusting weight.
ALT-BA-defsi-0410
85
Ö Now place the adjusting weight in the center of
the weighing plate, press the -key. After successful adjustment the system will automatically return to the menu.
The -key can be used to cancel the adjustment. In case of an adjustment error or incorrect adjusting weight an error message is displayed. Repeat adjustment.
Setup
P1
01 Int. calib r. 02 Ext. calib r. 03 User ca libr. 04 Calibr. test 05 Weig ht. cor r. 06 Aut calibr. 07
Print report
o
Auto cal. time
08
******** ******** ******** ********
0.0 0
1
function
function function
off
on
function
31.01.05
1
1 hour
03 User calibr. – user-defined adjusting weight
Ö Use the cursor () to select “03 User. cal.“
Setup
P1
01 Int. calib r. 02 Ext. calib r. 03 User ca libr. 04 Calibr. test 05 Weig ht. cor r. 06 Aut calibr. 07
Print report
o
Auto cal. time
08
******** ******** ******** ********
0.0 0
1
function
function function
off
on
function
31.01.05
1
1 hour
Ö Press -key; the display shows the measuring
value of the adjusting weight. The first numeral of the displayed value blinks. You can modify it using the arrow keys. After setting your new adjusting weight use the
-key to confirm.
Ö Ensure that there are no objects on the weighing
plate. Press -key
Ö Press the key; the display shows the
measuring value you selected for the adjusting weight
ALT-BA-defsi-0410
86
Ö Now place the adjusting weight in the center of
the weighing plate, press the -key. After successful adjustment the system will automatically return to the menu.
The -key can be used to cancel the adjustment. In case of an adjustment error or incorrect adjusting weight an error message is displayed. Repeat adjustment.
Setup
P1
01 Int. calib r. 02 Ext. calib r. 03 User ca libr. 04 Calibr. test 05 Weig ht. cor r. 06 Aut calibr. 07
Print report
o
Auto cal. time
08
******** ******** ******** ********
0.0 0
1
function
function function
off
on
function
31.01.05
1
1 hour
We recommend: Carry out adjustment as close to the maximum load of the balance as possible. Information concerning the test weights is available at: http://www.kern-sohn.com
04 Calibration test
Here, deviation from the last adjustment is determined. This is only a check, i.e. no values are changed.
Ö Use the cursor () to select “04 Cal. test“
Setup
P1
01 Int. calib r. 02 Ext. calib r. 03 User ca libr. 04 Calibr. test 05 Weig ht. cor r. 06 Aut calibr. 07
Print report
o
Auto cal. time
08
******** ******** ******** ********
0.0 0
1
function
function function
off
on
function
31.01.05
1
1 hour
Ö Press the -key, the calibration test is started
Calibration
Calibration check
Ö The result is displayed
Calibration
Check results Cal.: 180.1354 Act.: 180.1438 Diff.: - 0.0084
05 Weight corr. – not documented
ALT-BA-defsi-0410
87
06 Automatic calibration
Ö Use the cursor () to select “06 Autom. cal.“
Ö Press the -key; the current menu item blinks.
Ö Use the - or - key to choose between the
following settings:
0 no Autom. calibration locked
1 Temp. Autom. Calibration for a Temperature change of 0.8 °C
2 Time Autom. Calibration after a user-defined interval
3 both Autom. Calibration for a Temperature change of
0.8 °C or after the defined interval
Setup
P1
01 Int. calib r. 02 Ext. calib r. 03 User ca libr. 04 Calibr. test 05 Weig ht. cor r. 06 Aut calibr. 07
Print report
o
Auto cal. time
08
******** ******** ******** ********
0.0 0
1
function
function function
off
on
function
31.01.05
1
1 hour
Ö Confirm with -key
07 Auto cal. time - Automatic calibration after a user-defined interval
Ö Use the cursor () to select “07 Auto.cal.time“
Ö Press the -key; the current menu item blinks.
Ö Use the arrow keys to enter a value between 1
and 12 hours.
Setup
P1
01 Int. calib r. 02 Ext. calib r. 03 User ca libr. 04 Calibr. test 05 Weig ht. cor r. 06 Aut calib r. 07
Print report
o
Auto cal. time
08
******** ******** ******** ********
0.0 0
1
function
function function
off
on
function
31.01.05
1
1 hour
Ö Confirm with -key
ALT-BA-defsi-0410
88
08 Print report
In this menu item you can activate the function for printing your calibrating data
Ö Use the cursor () to select “08 “print report“
Ö Press the key; the current menu item blinks.
Use the arrow keys (see chapter 7.1) to save your settings
0 Print report deactivated 1 Print report activated
Setup
P1
01 Int. calib r. 02 Ext. calib r. 03 User ca libr. 04 Calibr. test 05 Weig ht. cor r. 06 Aut calibr. 07
Print report
o
Auto cal. time
08
******** ******** ******** ********
0.0
0
1
function
function function
off
on
function
31.01.05
1
1 hour
Ö Confirm with -key
Ö When print report is activated you will receive a printout of your adjustment data
after each adjustment (Ex.):
*** Internal calibration report***
Calibr. : internal
diff.. : 0,0038 g
Name………………………….
*** External calibration report ***
Calibr. : external
diff.. : -0,0624 g
Name ………………………….
*** User calibration report***
Calibr. : User calibration
diff.. : -0,0003 g
Name ………………………….