KERN AEJ_N, AES_N Operating Instructions Manual

KERN & Sohn GmbH
Ziegelei 1
D-72336 Balingen
Tel: +49-[0]7433- 9933-0
Fax: +49-[0]7433-9933-149
Internet: www.kern-sohn.com
Operating instruct ions Analytical balance
KERN AEJ_N/AES_N
Version 2.1
03/2013
GB
AEJ_N/AES_N-BA-e-1321
2 AEJ_N/AES_N-BA-e-1321
GB
KERN AEJ_N/AES_N
Version 2.1 03/2013
Operating instructions
Electronic analytical balance
Contents
1 TECHNICAL DATA ............................................................................................. 5
1.1 Dimensions ............................................................................................................................... 7
2 APPLIANCE OVERVIEW ................................................................................... 8
2.1 Display and Keyboard Summary ............................................................................................ 9
2.1.1 Numeric entry ....................................................................................................................... 10
3 BASIC INSTRUCTIONS ................................................................................... 11
3.1 Proper use ............................................................................................................................... 11
3.2 Improper Use .......................................................................................................................... 11
3.3 Warranty .................................................................................................................................. 11
3.4 Monitoring of Test Resources ............................................................................................... 11
4 BASIC SAFETY PRECAUTIONS ..................................................................... 12
4.1 Pay attention to the instructions in the Operation Manual ................................................ 12
4.2 Personnel training .................................................................................................................. 12
5 TR ANSPORT AND STORAGE ......................................................................... 12
5.1 Testing upon acceptance ...................................................................................................... 12
5.2 Packaging / return transport ................................................................................................. 12
6 UNPACKING, SETUP AND COMMISSIONING ............................................... 15
6.1 Installation Site, Location of Us e .......................................................................................... 15
6.2 Unpacking and checking ....................................................................................................... 15
6.2.1 Scope of delivery / serial accessories .................................................................................. 15
6.2.2 Placing .................................................................................................................................. 16
6.3 Mains connection ................................................................................................................... 16
6.3.1 Turning On the Power .......................................................................................................... 17
6.4 Connection of peripheral devices ......................................................................................... 17
6.5 Initial Commissioning ............................................................................................................ 17
AEJ_N/AES_N-BA-e-1321 3
7 ADJUSTMENT .................................................................................................. 18
7.1 Automatic adjustment using internal weight (AEJ only) .................................................... 19
7.2 Menu settings ......................................................................................................................... 21
7.2.1 P1.1 iCAL ............................................................................................................................. 21
7.2.2 P1.2 ECAL - External adjustment (AES models) ................................................................. 22
7.2.3 P1.3 tCAL – adjustment test ................................................................................................ 23
7.2.4 P1.4 ACAL – Start criteria of the automatic internal adjustment (AEJ only) ........................ 25
7.2.5 P1.5 CALt– time controlled automatic adjustment (AEJ only) ............................................. 26
7.2.6 P1.5 CALr– Enable/disable issue of adjustment protocol .................................................... 27
7.3 Verification .............................................................................................................................. 28
8 MENU ................................................................................................................ 30
8.1 Navigation in the menu .......................................................................................................... 32
8.2 Exit menu / back to weighing mode. .................................................................................... 33
9 WEIGHING FUNCTIONS P6 FUNC ................................................................. 34
9.1 Turn on/off balance ................................................................................................................ 34
9.2 Weighing ................................................................................................................................. 34
9.3 Standard weighing unit .......................................................................................................... 35
9.4 Weighing units switch-over ................................................................................................... 36
9.5 Taring ....................................................................................................................................... 36
9.6 Underfloor weighing ............................................................................................................... 37
9.7 Selecting the operating mode ............................................................................................... 38
9.7.1 Selecting accessible operating modes ................................................................................. 40
10 BALANCE SETTINGS “P3 READ” ............................................................. 41
10.1 Filter - adaptation to ambient conditions and weighing options ...................................... 41
10.2 Rest position control display – Adaptation of weighing speed ........................................ 42
10.3 Auto Zero ................................................................................................................................. 43
10.4 Hide last decimal place .......................................................................................................... 44
10.5 Di spensing mode “RAPID” .................................................................................................... 45
11 FURTHER USEFUL FUNCTIONS “P7 OTHER“ .......................................... 46
11.1 Di splay background illumination .......................................................................................... 46
11.2 Acoustic signal for key operation ......................................................................................... 47
11.3 P rint weighing settings .......................................................................................................... 48
4 AEJ_N/AES_N-BA-e-1321
12 GLP/ISO LOG “P2 GLP“ .............................................................................. 49
13 OPERATION MODES ................................................................................... 53
13.1 P arts counting ........................................................................................................................ 54
13.2 Weig h ing with tolerance range ............................................................................................. 58
13.3 P ercent determination............................................................................................................ 61
13.3.1 Determining the reference weight by weighing (function F4 PrcA) ................................. 61
13.3.2 Numeric entry of reference weight (function F4 Prcb) ..................................................... 62
13.4 Dosing...................................................................................................................................... 64
13.5 Determining the density of solid matter and fluids ............................................................ 65
13.6 P ipette calibration .................................................................................................................. 66
13.6.1 Entering parameters ........................................................................................................ 66
13.6.2 Carry out series of measurements for test volume V0 ..................................................... 67
13.6.3 Carry out series of measurements for test volume V
1/2
................................................... 68
13.6.4 Carry out series of measurements for test volume V1 ..................................................... 69
14 DATA OUTPUT RS 232 “P4 PRINT“ ............................................................ 71
14.1 Technical data ......................................................................................................................... 71
14.2 Interface cable ........................................................................................................................ 72
14.3 Menu settin g s “P4 PRINT“ .................................................................................................... 73
14.4 P rintout examples (KERN YKB-01N) .................................................................................... 78
14.5 Output lo g ............................................................................................................................... 79
14.5.1 Issue after pressing the PRINT ´button ........................................................................... 79
14.5.2 PC controlled output ........................................................................................................ 80
14.5.3 Output of date/time .......................................................................................................... 80
14.6 Remote control instructions ................................................................................................. 81
15 SERVICE, MAINTENANCE, DISPOSAL ...................................................... 82
15.1 Cleaning .................................................................................................................................. 82
15.2 Service, maintenance ............................................................................................................. 82
15.3 Disposal ................................................................................................................................... 82
16 INSTANT HELP ............................................................................................. 83
16.1 Error messages ...................................................................................................................... 84
17 DECLARATION OF CONFORMITY .............................................................. 85
AEJ_N/AES_N-BA-e-1321 5
1 Technical data
KERN AEJ 100-4NM AEJ 200-4NM AEJ 200-5NM
Readability (d) 0.1 mg 0.1 mg 0.01 mg / 0.1 mg Weighing range (max) 160 g 220 g 60 g – 220 g Minimum load (Min) 10 mg 10 mg 1 mg Verification value (e) 1 mg 1 mg 1 mg Verification class I I I Reproducibility 0.2 mg 0.2 mg 0.04 mg / 0.2 mg Linearity ± 0.3 mg ± 0.3 mg ± 0.1 mg / ± 0.3 mg Stabilization time 4 sec. Adjustment weight internal Warm-up time 8 h Weighing Units mg, g, ct Smallest part weight for
piece counting
10 mg 10 mg 1 mg
Reference quantities at piece counting
10, 20, 50, freely selectable
Weighing plate, stainless steel
ø 85mm ø 70mm
Net weight (kg) 5.4 Permissible
ambient condition
+18° C to +30° C
Humidity of air 30 ~ 80 % relative (not condensing) Input voltage 110 - 230 V - 240 V AC‚50 Hz Power pack
secondary voltage
13.8 V
Interface RS 232C Underfloor weighing
device
Hooks
6 AEJ_N/AES_N-BA-e-1321
KERN AES 100-4N AES 200-4N
Readability (d) 0.1 mg 0.1 mg Weighing range (max) 160 g 220 g Reproducibility 0.2 mg 0.2 mg Linearity ± 0.3 mg ± 0.3 mg Recommended adjusting
weight not supplied (class)
150 g (E2) 200 g (E2)
Stabilization time 4 sec. Warm-up time 8 h Weighing Units g, mg, ct, oz, ozt, GN, mom, dwt, teal Smallest part weight for
piece counting
10 mg
Reference quantities at piece counting
10, 20, 50, freely selectable
Weighing plate, stainless steel
ø 85mm
Net weight (kg) 5.4 Permissible
ambient condition
+18° C to +30° C
Humidity of air 30 ~ 80 % relative (not condensing) Input voltage 110 - 230 V - 240 V AC‚50 Hz Power pack
secondary voltage
13.8 V
Interface RS 232C Underfloor weighing
device
Hooks
AEJ_N/AES_N-BA-e-1321 7
1.1 Dimensions
8 AEJ_N/AES_N-BA-e-1321
2 Appliance overview
1. Windshield
2. Weighing pan
3. Display
4. Keyboard
5. Levelling screw
6. Bubble level
7. Terminal power supply unit
8. RS232 interface
9. Not documented
AEJ_N/AES_N-BA-e-1321 9
2.1 Display and Keyboard Summ ary
1. Stability display
2. Zero indicator
3. The non-calibrated value is given in brackets in calibrated scales.
4. Weighing unit “g“
Key
Function
Turn on/off
Call up menu
Activate operating mode
Scroll backwards in menu
Switch-over weighing unit
Scroll forward in menu
Calculate weighing data via interface
Confirm / save
Taring
Zeroing
Exit menu / Return to weighing mode
10 AEJ_N/AES_N-BA-e-1321
2.1.1 Numeric entry
Key Description Function
Navigation button
Increase flashing dig it
Navigation button
Decrease flashing digi t
Navigation button
Digit selection to the right
Navigation button
Digit selection as per
Navigation button
Confirm entry
ESC Cancel input
AEJ_N/AES_N-BA-e-1321 11
3 Basic instructions
3.1 Proper use
The balance you purchased is intended to determine the weighing value of material to be weighed. It is intended to be used as a “non-automatic balcance”, i.e. the material to be weighed is manually and carefully placed in the centre of the weighing pan. As soon as a stable weighing value is reached the weighing value can be read.
3.2 Improper Use
Do not use balance for dynamic add-on weighing procedures, if small amounts of goods to be weighed are removed or added. The “stability compensation“ installed in the balance may result in displaying an incorrect measuring value! (Example: Slowly draining fluids from a container on the balance.) Do not leave permanent load on the weighing pan. This may damage the measuring system. Impacts and overloading exceeding the stated maximum load (m ax) of the balance, minus a possibly existing tare load, must be strictly avoided. Balance may be damage by this. Never operate balance in explosive environment. The serial version is not explosion protected. The structure of the balance may not be modified. This may lead to incorrect weighing results, safety-related faul ts and dest ruc ti on o f the bala nce. The balance may only be used according to the described conditions. Other areas of use must be released by KERN in writing.
3.3 Warranty
Warranty claims shall be voided in case
Our conditions in the operation manual are ignored
The appliance is used outside the described uses
The appliance is modified or opened
Mechanical damage or damage by media, liquids, natural wear and tear
The appliance is improperly set up or incorrectly electrically connected
The measuring system is overloaded
3.4 Monitoring of Test Resources
In the framework of quality assurance the measuring-related properties of the balance and, if applicable, the testing weight, must be checked regularly. The responsible user must define a suitable interval as well as type and scope of this test. Information is available on KERN’s home page (www.kern-sohn.com with regard to the monitoring of balance test substances and the test weights required for this. In KERN's accredited DKD calibration laboratory test weights and balances may be calibrated (return to the national standard) fast and at moderate cost.
12 AEJ_N/AES_N-BA-e-1321
4 Basic Safety Precautions
4.1 Pay attention to the instruc t ions in the Operation Manual
Carefully read this operation manual before setup and
commissioning, even if you are already familiar with KERN balances.
All language versions contain a non-binding translation.
The original German is binding.
4.2 Personnel training
The appliance may only be operated and maintained by trained personnel.
5 Transport and storage
5.1 Testing upon acceptance
When receiving the appliance, please check packaging immediately, and the appliance itself when unpacking for possible visible damage.
5.2 Packaging / return transport
Keep all parts of the original packaging for a possibly required
return.
Only use original packaging for returning.
AEJ_N/AES_N-BA-e-1321 13
Prior to dispatch disconnect all cables and remove loose/mobile
parts.
Reattach possibly supplied transport securing devices.
Secure all parts such as the glass wind screen, the weighing
platform, power unit etc. against shifting and damage.
14 AEJ_N/AES_N-BA-e-1321
AEJ_N/AES_N-BA-e-1321 15
6 Unpacking, Setup and Commissioning
6.1 Installation Site, Locati on of Us e
The balances are designed in a way that reliable weighing results are achieved in common conditions of use. You will work accurately and fast, if you select the right location for your balance.
Therefore, observe the following for the installation site:
Place the balance on a firm, level surface;
Avoid extreme heat as well as temperature fluctuation caused by installing
next to a radiator or in the direct sunlight;
Protect the balance against direct draughts due to open windows and doors;
Avoid jarring during weighing;
Protect the balance against high humidity, vapours and dust;
Do not expose the device to extreme dampness for longer periods of time.
Non-permitted condensation (co nd ens a ti on o f air humidi ty on the appliance) may occur if a cold appliance is taken to a considerably warmer environment. In this case, acclimatize the disconnected appliance for ca. 2 hours at room temperature.
Avoid static charge of goods to be weighed or weighing container.
If electro-magnetic fields or static charge occur, or if the power supply is unstable major deviations on the display (incorrect weighing results) are possible. In that case, the location must be chang ed.
6.2 Unpacking and checking
Open package, take out the appliance and accessories. Verify that there has been no damage and that all packing items are present.
6.2.1 Scope of delivery / serial accessories
Balance, see chap. 2
Mains adapter
Operating instructions
Protective cover
16 AEJ_N/AES_N-BA-e-1321
6.2.2 Placing
Installation of weighing plate
Models with weighing pan ø 85mm Models with weighing pan ø 70mm
Assemble all parts in the right order. Close gates. For instructions on how to connect the weighing scale to the
power supply see chap. 6.3.
Level balance with foot screws until the air bubble of the water balance is in the
prescribed circle
Check levelling regularly
6.3 Mains connection
Power is supplied via the external mains adapter. The stated voltage value must be the same as the local voltage. Only use original KERN mains adapters. Using other makes requires consent by KERN.
AEJ_N/AES_N-BA-e-1321 17
6.3.1 Turning On the Power
Supply balance with power via the mains adapter. The display unit lights up and shown the software version. The weighing scale carries out a self­test and automatically moves into weighing mode. In the AEJ models an automatic adjustment takes place.
6.4 Connection of periphera l devi c e s
Before connecting or disconnecting of additional devices (printer, PC) to the data interface, always disconnect the balance from the power supply.
With your balance, only use accessories and peripheral devices by KERN, as they are ideally tuned to your balance.
6.5 Initial Commissioning
In order to obtain exact results with the electronic balances, your balance must have reached the operating temperature (see warming up time chap. 1). During this warming up time the balance must be connected to the power supply (mains, accumulator or battery). To adapt the balance to ambient conditions, open wind screen doors. The accuracy of the balance depends on the local acceleration of gravity. Strictly observe hints in chapter Adjustment.
18 AEJ_N/AES_N-BA-e-1321
7 Adjustment
As the acceleration value due to gravity is not the same at every location on earth, each balance must be coordina ted - in compliance with the underlying physical weighing principle - to the existing acceleration due to gravity at its place of location (only if the balance has not already been adjusted to the location in the factory). This adjustment process must be carried out for the first commissioning, after each change of location as well as in case of fluctuating environment temperature. To receive accurate measur i ng values it is also recommended to adjust the balance periodically in weighing operation.
Possible adjustments include:
Automatic adjustment using internal weights (AEJ models only) For time and temperature controlled see chap. 7.1.
Manual adjustment (AEJ models only) Activated by pressing the F-key three times, see chap. 7.2.1
Adjustment with external weight, see chap. 7.2.2
Adjustment can only be carried out when the weighing pan is free of objects.
Any loads remaining on the weighing pan will be indicated by error message Er1Hi.
You can cancel the adjustment by pressing the TARE-key.
AEJ_N/AES_N-BA-e-1321 19
7.1 Automatic adjustment usi ng int e rnal weight (AEJ only)
With the internal adjustment weight, the weighing accuracy can be checked and re­adjusted at any time.
The automatic adjustment function is always enabled. You can start adjustment at any time by pressing the F-key 3 times.
Temperature / time controlled adjustment is taking place:
Automatic adjustment will be indicated 5 minutes in advance by “°C” (change of temperature t 0.8°C) or the “►”symbol on the display (after a period of 3 hours has passed or a defined time interval).
The user must complete his/her weighing process within this time. After 5 minutes [ CAL 30 ] appears on the display.
This starts a “count down” of 30 seconds [ CAL 30 ] [ CAL 0 ]. You may cancel adjustment during these 30 seconds by pressing the TARE-key. This makes the balance return to weighing mode in order to e. g. complete an unfinished measurement.
After a further 5 minutes the automatic adjustment will restart and [ CAL 30 ] appears. You may interrupt the adjustment process several times. However, keep in mind that an interruption of the adjustment may lead to incorrect weighing results.
20 AEJ_N/AES_N-BA-e-1321
Automatic adjustment is carried out:
When balance was disconnected from the mains.
When the temperature is changing
The moment a change in temperature of 0.8°C is detected by the temperature sensor, the weighing scale will carry out a fully automatic adjustment. The °C symbol indicates the pending adjustment.
at the end of a time interval After the time interval (selectable from 1-12 hours) set in the menu (P1.5 CALt)
has passed, the weighing scale will indicate the pending adjustment by displaying (indicator ►):
In weighing scales with type approval certificate
menu item P1.5 CALt is locked.
Adjustment takes place every 3 hours
AEJ_N/AES_N-BA-e-1321 21
7.2 Menu settings
Menu item “P1 CAL“ [adjustment]
AEJ AES
P1.1 iCAL Internal adjustment AEJ only P1.2 ECAL P1.1 ECAL External adjustment * P1.3 tCAL P1.2 tCAL Adjustment test P1.4 ACAL Automatic adjustment AEJ * only P1.5 CALt
Time controlled automatic adjustment AEJ *
only
P1.6 CALr P1.3 CALr Printout adjustment log
* = Locked in weighing scales with type approval certificate [Er 9 lock].
7.2.1 P1.1 iCAL
For automatic internal adjustment see chap. 7.1 Starting interna l adjustment manually
Press F-key three times
Ensure that there are no objects on the weighing pan.
You can cancel adjusting by pressing the TARE-key.
After successful adjustment the balance automatically returns
to weighing mode.
22 AEJ_N/AES_N-BA-e-1321
7.2.2 P1.2 ECAL - External adjustment (AES models)
Carry out adjustment with the help of the recommended adjustment weight (see chapter 1 “Technical Specifications”). Observe stable environmental conditions. A warming up time (see chapter 1) is required for stabiliz at ion.
Return to weighing mode, see chap. 8.2.
This function is locked in weighing scales with type approval certificate.
Ensure that there are no objects on the
weighing pan. Press PRINT-key, CAL will be shown. Wait until the weighed value for the required adjustment weight appears. Place the required calibration weight carefully in the centre of the weighing plate and press the PRINT key. The display is showing CAL and adjustment is starting. Adjustment will be complete when the display is showing UNLOAD; remove the adjustment weight. After successful adjustment the balance automatically returns to weighing mode.
AEJ_N/AES_N-BA-e-1321 23
7.2.3 P1.3 tCAL – adjustment test
AEJ models
This function is applied to compare the internal adjustment weight with the saved value. The test is carried out automatically. The result is shown on the display and, if an optional printer is connected, a printout will be issued via the RS 232 interface.
+ Return to weighing mode, see chap. 8.2.
d
24 AEJ_N/AES_N-BA-e-1321
AES models
Here, deviation from the last adjustm ent is determined. This is only a check, i.e. no values are changed.
+ Return to weighing mode, see chap. 8.2.
AEJ_N/AES_N-BA-e-1321 25
7.2.4 P 1 . 4 ACAL – Start criteria of the automatic internal adjustment
(AEJ only)
This function is applied to defin e the criteria required for s tarting up automatic adjustment.
Settings:
nonE Automatic adjus tm ent disabled tEmp Start automatic Adjustment after change in temperature timE The start of ti me controlled automatic adjustment depends on t he set ti ng
“P1.5 CALt”
both Start automatic Adjustment is temperature and time dependent
+ Return to weighing mode, see chap. 8.2.
This function is locked in weighing scales with type approval certificate.
26 AEJ_N/AES_N-BA-e-1321
7.2.5 P1.5 CALt– time controlled automatic adjustment (AEJ only)
This function is applied to define the time interval (selectable from 1-12 hours) for automatic adjustment.
+ Return to weighing mode, see chap. 8.2.
This function is locked in weighing scales with type approval certificate.