Kern 434 Service manual

Service Manual for Electronic
Precision Balances series
434
Version 1.0 7/97
434-SH-e-9710
Table of Contents
1 Specifications __________________________________________________________4 2 Introduction ___________________________________________________________5 3 Principles of operation ___________________________________________________5 4 Block Diagram _________________________________________________________6 5 Best Conditions for Weighing _____________________________________________7 6 Balance Functions ______________________________________________________8 7 Description of Function Menu_____________________________________________9 8 Description of Functions ________________________________________________10 9 Weighing Unit Initialisation _____________________________________________11
9.1 Keyboard Method_________________________________________________________ 11
9.2 Parameter Number Chart __________________________________________________ 12
10 Piece Counting Mode ___________________________________________________13
10.1 Removing establishment of the standard unit weight for piece counting__________ 14
11 Percent Mode _________________________________________________________15
11.1 Weighing Mixture Mode _________________________________________________ 15
12 Adjusting Introduction__________________________________________________17 13 Inner A /D Counter Check Mode__________________________________________19 14 Disassembly / Assembly _________________________________________________21
14.1 Removing the Top Case__________________________________________________ 21
14.2 Removing the Mechanical Unit____________________________________________ 22
14.3 Mechanical Unit Disassembly_____________________________________________ 23
14.4 Assembly ______________________________________________________________ 24
15 Jig Set _______________________________________________________________25 16 Additional Disassembly / Assembly ________________________________________26
16.1 Replacement of the Force Coil Bobbin _____________________________________ 26
16.2 Cleaning Inside the Magnet ______________________________________________ 26
16.3 Positioning Edge of the Beam Inside the Photo Sensor ________________________ 27
17 Primary Checking Procedure_____________________________________________28 18 Visual Check__________________________________________________________28 19 Mechanical Fault Finding_______________________________________________29
19.1 "+ / - .", "All Segment Lines" or No Display ________________________________ 29
19.2 Unstable Weighing Results Adjustment ____________________________________ 30
20 Mechanical Fault Finding Chart _________________________________________32 21 Short Wire Check ______________________________________________________33 22 Connecting Cable Check ________________________________________________34
434-SH-e-97102
23 Plate Bearing / Coupling Link Check ______________________________________35 24 Beam Stopper Adjustment _______________________________________________36 25 A / D Converter Check __________________________________________________38 26 Initialisation of EEPROM _______________________________________________39 27 Linearity Adjusting_____________________________________________________40
27.1 Linearity Adjusting Procedure____________________________________________ 40
27.2 Adjusting Masses Chart _________________________________________________ 41
28 Span Adjusting ________________________________________________________42
28.1 Adjusting Masses Chart _________________________________________________ 42
28.2 Adjusting Mass Tolerance Collection ______________________________________ 44
28.3 Disable the Span Adjusting Function_______________________________________ 46
29 Cornerload Adjustment _________________________________________________46
29.1 Cornerload Adjustment Masses ___________________________________________ 46
30 Electronic Fault Finding ________________________________________________48
30.1 Power Check___________________________________________________________ 48
30.2 EEPROM Check _______________________________________________________ 48
30.3 Voltage Check__________________________________________________________ 49
31 Electronic Fault Finding Charts __________________________________________50 32 Wave Form Check _____________________________________________________55 33 RS-232 C Interface_____________________________________________________57
33.1 Interface Specifications __________________________________________________ 57
33.2 Output Format _________________________________________________________ 59
33.3 Output Data Mode ______________________________________________________ 60
33.4 External Control Commands _____________________________________________ 61
33.5 Cabling Diagrams_______________________________________________________ 62
34 Installation of the Air Shield Case (OMJ-2 Option)___________________________63 35 Troubleshooting _______________________________________________________64 36 KERN 434 Series Parts List ______________________________________________68 37 Balance drawings ____________________________ Fehler! Textmarke nicht definiert.
434-SH-e-9710 3
1 Specifications
Model KERN 434-23 KERN 434-33 KERN 434-37 Gram Ounce Pound Carat Pennyweight Troy Ounce Grain Momme Tael (Hong Kong) Tael (Singapore) Tael (Taiwan) Linearity Readability Stabilization time Span drift (15° C - 30°C)
310 x 0.001 510 x 0.001 3100 x 0.001 10 x 0.0005 18 x 0.0005 109 x 0.0005
0.6 x 0.000005 1.1 x 0.000005 6.8 x 0.000005 1550 x 0.005 2550 x 0.005 15500 x 0.005
199 x 0.001 327 x 0.001 1993 x 0.001
9.9 x 0.00005 16 x 0.00005 99 x 0.00005
4784. x 0.02 7870. x 0.02 47840. x 0.02 82 x 0.0005 136 x 0.0005 826 x 0.0005 8 x 0.00005 13 x 0.00005 82 x 0.00005 8 x 0.00005 13 x 0.00005 82 x 0.00005 8 x 0.00005 13 x 0.00005 82 x 0.00005
±
0.002 g
0.001 g 0.01 g
2.3 sec
±
5 ppm/° C
±
0.02 g
±
3 ppm/° C
Operation temperature Pan size Total min. weight of
percent display Min. percent display Total min. weight of
piece counting Max. piece counting
display Sampling number of
piece counting Data output
Power supply Dimensions
Options: Auxiliary display (OMK-0), Underweighing hook (OMJ-1), Air shield case (OMJ-2) Animal weighing pan (OMJ-3), Small animal weighing pan (OMJ-4), Auxiliary display base stand (OMJ-10), Dust cover (O MJ- 11), Auxiliary display stand pole (OMJ-12)
31000 pcs 51000 pcs 31000 pcs
RS-232C bi-direction (standard accessory)
AC Adapter 220 V, 110 V or other (+10 V, 015 V)
135 mm
0.01 g 0.1 g
50/60 Hz Output AC19V 200 mA
194 (W) x 334 (D) x 81 (H) mm
10° C - 40° C
160 x 160 mm
1 g 10 g
0.01 %
5, 20, 100 or 100 pcs
434-SH-e-97104
2 Introduction
This Maintenance Manual covers three models from the KERN 434 series, the KERN 434-23, KERN 434-33 and KERN 434-37. Please read this Maintenance Manual and the owner’s Instruction Manual fully before beginning any maintenance work.
The KERN 434 series precision balances are the products of years of research, development, design and in-field testing. They incorporate the latest advances in mechanical and electronic engineering and offer the highest standard of reliability, easy to use functions and rugged durability.
3 Principles of operation
The KERN 434 series precision balances work on the principle of "Force Compensation". Any change in the load on the weighing pan causes the Beam to pivot on fulcrum Plate Bearings (see Block Diagram page 4). Attached to this Beam is a coil wound with fine wire, called the "Force Coil Bobbin", which floats in a permanent magnet. At the end of the Beam there is a small notch which allows light from a Light-Emitting Diode (LED) to pass through to a Photo Sensor (Light Measuring Diode). At zero weight, the light detected by the Photo Sensor is exactly equal to the light emitted by the LED.
When the end of the beam is forced up by the leverage exerted from a mass placed on the weighing pan, the Photo Sensor detects a change in the position of the beam and the attached Force Coil Bobbin, because the light reaching the Photo Sensor has become less than that emitted by the LED. The balance then feeds the Force Coil Bobbin with more voltage, which increases the magnetic power and pulls the Bobbin downward until the light reaching the Photo Sensor is once again equal to the light emitted by the LED. This is accomplished by a Differential Amplifier, a filter and an Analogue/Digital (A/D) Converter receiving photo current from the Photo Sensor, converting it to voltage and boosting it back to the Force Coil Bobbin.
The electrical current flowing through the Force Coil Bobbin generates a voltage proportional to the load weight on the pan. This is read back through the Differential Amplifier and filter, then the A/D Converter digitalises this voltage. The resulting value is then counted and fed to the microprocessor (CPU). The CPU performs a multitude of commands and mathematical operations in conjunction with parameter and adjusting information stored in Random Access Memory (RAM). Finally, the results are displayed on the Liquid Crystal Display (LCD) or sent to the RS-232 C Interface.
434-SH-e-9710 5
4 Block Diagram
434-SH-e-97106
5 Best Conditions for Weighing
1. Never turn off the power switch or disconnect the AC adapter when in use.
2. When making a measurement, always place the sample in the centre of the weighing pan. Slight errors may result if the sample is not near the centre of the pan.
3. Make sure the balance is level by using the level vial and the adjustable feet on the bottom.
4. Install the balance in a controlled environment. A) The weighing room should be kept clean, dry and free of cigarette
smoke.
B) Protect the balance from drafts (air currents). Use a draft shield if
necessary. C) Maintain the ambient temperature to D) Maintain the ambient humidity to E) If larger changes in temperature or humidity occur, re-calibrate the
balance. F) The balance table should be level and free from excessive vibration.
Corners of rooms are less prone to vibration. G) Do not expose the balance to direct sunlight or radiated heat. Keep away
from windows, heaters, hot plates, flames, fans, air conditioners, etc. H) Allow hot or cold sample containers to come to ambient temperature
before weighing. I) Discharge any statically charged sample before weighing. J) Do not expose the balance to corrosive gases.
±±±±
10 %.
±±±±
3° C.
5. Magnetised samples cannot be weighed accurately on an electronic balance. Keep equipment containing magnets away from the balance.
6. Make certain that the AC power supply is free from electrical disturbances.
7. Clean the balance with a damp cloth only (no solvents).
8. Always warm-up the balance before use or leave on Stand-by (display off) overnight.
9. Always handle the balance with care during use or when moving or storing.
434-SH-e-9710 7
6 Balance Functions
Sampling Time Function (SAP):
The
the surrounding environmental conditions. The
function allows the balance to adapt to
SAP
should be adjusted to a small
SAP
value for corresponds to the "integration time" in seconds that the balance is using to display readings.
Stability Indicator Function (STb):
All KERN 434 series balances use the star (Η) symbol to indicate stability. When the Η appears on the display, the balance has not yet stabilised. When the Η disappears from the display, the balance is stable. The weight reading should be made only after the Η has disappeared from the display and the indicator on the display for different sample types and weighing conditions. Small values fo r
require very stable conditions before the Η disappears from the
STb
display and large values allow the Η to disappear under more reliable conditions such as animal weighing.
Adjusting (Cal):
The
function is used to perform a Span Adjusting on the
Cal
balance. See Span Adjusting page 35 for the steps required for this procedure.
Weighing Unit Selection (Uni):
different weighing units. between units.
Uni-2
allows the user to switch from grams to another unit by pushing
Uni-1
locks in grams or another unit and prevents switching
The
function is used to select and lock in
Uni
the U key. See Weighing Unit Initialisation page 9 to select or change the available weighing units.
Output Data Mode (oUT):
The
function is used to change the output data
oUT
mode when using the RS-232 C interface. See RS-232 C Interface page 54.
RS-232 C Mode (rS):
The rS function allows you to switch between 8 bit non-parity (8N1) and 7 bit even-parity (7E1)when using the RS-232 C Interface. See RS-232 C Interface page 58 - 63
Baud Rate Speed (rSS):
The
function changes the baud rate speed when using
rSS
the RS-232 C Interface. See RS-232 C Interface page 8 for the steps required for this procedure.
Auto-Start Mode (AST-ON):
If
auto-start mode
position immediately without pressing the
ON/S
is on, the balance will be weighing
key.
TBY
434-SH-e-97108
7 Description of Function Menu
0.000g
⇓→
Press the F key.
Press the
!
Press the U key.
T
ARE
key.
SAP 1
Stb 1
CAL
Un i 1
OUT 1
!
!
!
!
"
!
!
2
"
!
2
!
3
3
!
A
H1
!
H2
!
H3
!
0
Full Point Value
Second Point Value
2
!
2
3
"
!
!
4
5
rS 8 n 1
!
"
7 E 1
r S S 1 . 2
!
2 . 4
"
!
4 . 8
AST ON
The display returns to the weighing mode if no operation is made for 3 sec. See the appropriate section for further explanation of individual function
!
"
OFF
434-SH-e-9710 9
8 Description of Functions
Display Command Selectable Setting Factory Set ting Remarks
1 Very steady
Range of 2 Normal
SAP
STb
CAL
Uni
OUT
rS
rSS
AST
Vibration 3 Unstable 2 Note 1
Adapter A Extremely unstable
1 Limited
Range of 2 Standard
Stability 3 Extensive 2 Note 1
H1 HOLD - Limited H2 HOLD - Standard H3 HOLD - Extensive
Adjusting See the adjusting section (page 11). Note 2
Unit Locking the key.
Selection Release the lock. 2 Note 3
Output upon print
1 com m and m ade
only after the balance has stabilised. After print command
2 made, output as
soon as balance has stabilised.
Data Output upon print
Transmission 3 command regard- Note 4
Mode less of stability.
Automatic output as
4 soon as balance
has stabilised. Automatic output
5 regardless of
stability.
RS-232 C 8N1 8 bit Non-parity
Interface 7E1 7 bit Even-parity 8N1 Note 5
Mode
RS-232 C 1.2 1200 BPS
Interface 2.4 2400 BPS 1.2 Note 5
Speed Mode 4.8 4800 BPS
Auto-Start ON Auto-start is on.
Mode OFF Auto-start is off. OFF Note 6
434-SH-e-971010
9 Weighing Unit Initialisation
9.1 Keyboard Method
Step 1:
Make certain that AST function
mode (Automatic start) is off.
Step 2:
To have the count-down display, once disconnect the AC adapter cord and reconnect it again.
Step 3:
keys at the same time when
F
111111111
Step 4:
Press and hold the
“ appears.
Release the keys when the „All
T
ARE
, U and
Segment“ line appears.
Step 5:
Parameter number will appear after „All Segment“ line has stopped flashing.
Step 6:
Two units in addition to grams can be selected and programmed into memory. Grams will automatically be in memory and does not need to be selected in this procedure. Unit 1 is selected by pressing the U key until the unit code place. Unit 2 is selected by pressing the
key until the unit does appear on the display in
T
ARE
the Unit 2 digit place. For example, if you select Units F and E as shown below the balance will show weighing results in Piece Counting and Percent.
Step 7:
Press the F key to store your selections. Check to make certain the desired units have been stored by pressing the U key. When the unit selection (Uni) is set at Uni-2 you can switch between units by pressing the U key. The Uni-1 setting locks in one unit and switching to another unit is not possible.
434-SH-e-9710 11
Step 8:
Press the F key to complete the Unit Selection procedure.
Display Circular
(# Press the U key)
Gram display
$%
9.2 Parameter Number Chart
0 Gram g 9 Carat C.M 1 Cancel last digit g A Momme mom 2 Kilogram kg B Taiwan Tael tl 3 Pound lb C Singapore Tael tl 4 Ounce oz D Hongkong Tael tl 5 Troy ounce ozt E
6Grain gr F
#
Unit 1
#
Unit 2
Η Η
#
Percent % Piece counting pcs
Gram display
7 Pennyweight dwt 8 Carat ct
Factory setting
Η
434-SH-e-971012
10 Piece Counting Mode
Step 1:
Press the U key to switch the display to Piece Counting Mode (see Weighing Unit Initialisation page 11).
Step 2:
Press the F key until the desired count value is displayed, 5, 20, 100 or 1000 pcs.
Step 3:
Count out the same number of sample pieces as you selected in Step 2, and place them on the weighing pan.
Step 4:
Press the
key to memorise
T
ARE
the number of sample pieces and their unit weight. The display will flash until the balance has memorised this data.
Step 5:
Proceed with piece counting.
Step 6:
message when pressing the
If the display shows a
T
ARE
"Lo"
key, the unit weight of the sample pieces is less than the balance can detect.
The minimum weight of each sample piece required to establish piece counting is as follows:
KERN 434-23 KERN 434-33 KERN 434-37
⇑ ⇑ ⇑
0.01 g
0.01 g
0.1 g
434-SH-e-9710 13
10.1 Removing establishment of the standard unit weight for piece counting
This unique function of the KERN 434 series is used to easily increase the number of pieces of sample used as your piece counting standard without taking the time to count out each standard sample separately. This is important because the greater the number of sample pieces used to set your standard, the more accurate an precise your piece counting will be.
Step 1:
sample pieces and unit weight at 5 PCS, f or example, by following the procedure on the previous page.
Step 2:
weighing pan. The display will show that there are 20 pieces on the pan.
Step 3:
pan and press the F key until 20 PCS appears on the display.
Step 4:
new number of sample pieces and the new unit weight.
Step 5:
increase the standard to 100 or 1000 pieces.
Establish the standard number of
Place 15 more sample pieces on the
Leave the 20 pieces on the weighing
Press the
Follow the same procedure to
key to memorise the
Tare
Note: The key is to let the balance do the counting for you when establishing the standard. For example, count out 20 pieces by yourself and set the standard at 20 pieces. Use the piece counting mode to count up to 100 pieces and set the new standard at 100 pieces. Then use the piece counting mode again to count up to 1000 pieces and set the new standard again at 1000 pieces.
434-SH-e-971014
11 Percent Mode
Target weight:
Step 1:
"%" appears.
Step 2:
weight on the weighing pan.
Step 3:
100% weight. Make certain that """ sign appears when entering a sample. Press the
T
ARE
Step 4:
when "100.00%" appears. Proceed with percent weighing by placing a new sample on the weighing pan. Percentage display is
100.00& for KERN 434-23, KERN 434-33 and KERN 434-37.
Continuously press the U key until
Place a sample of the ideal 100%
Press the F key to memorise the
key if the """ sign does not appear.
Percent mode procedure is complete
The minimum weight of the 100% sample required to establish percent weighing is as follows:
KERN 434-23 KERN 434-33 KERN 434-37
⇑ ⇑ ⇑
11.1 Weighing Mixture Mode
This mode is used when you want to mix two sample together. For instance when you add sample B to sample A to make a 100% mixture sample.
1 g 1 g 10 g
Sample B 40 % Sample A 60 %
434-SH-e-9710 15
Step 1:
the 100 % weight by using the percent mode procedure above.
First, memorise sample A to be
Step 2:
by pressing the U and P keys.
Step 3:
the 100 % weight is reached.
Reset sample A at 60 % weight
Add sample B to sample A until
434-SH-e-971016
12 Adjusting Introduction
Adjusting of KERN 434 series precision balances is required at initial installation, any time the balance is moved or bumped, whenever the ambient temperature changes by more than 3° C, and additionally every 30 days or so. Adjusting is necessary for two main reasons. First, with time and use, mechanical deviations can occur. Secondly, the weight of a mass in one location or under a certain set of conditions will not always be the same at a different location or under a different set of conditions.
There are a number of adjusting procedures that will need to be done during the life of a KERN 434 series balance. They are briefly explained below. The actual adjusting procedures are contained throughout this manual (see Table of Contents page 2).
Linearity Adjusting: The purpose of Linearity Adjusting is to create a straight line from Zero to the Full Capacity Scale Value to that the balance will display accurate weighing results at all values in between. On the graph below, the Y-axis represents the "true weight" of a mass on the weighing pan and the X-axis represents the weight displayed by the balance. An accurate mass of any weight value will be plotted on a linear (straight line) path from zero to the maximum capacity. However, although a straight line will be generated by Linear Adjusting, this line must be correctly placed by the balance during Span Adjusting (see Linearity Adjusting page
41).
Span Adjusting: With Span Adjusting, we are shifting the straight line generated during Linearity Adjusting by giving it the Zero Load Value and the full capacity adjusting mass value as reference points (see Span Adjusting page 43).
434-SH-e-9710 17
Temperature Compensation Adjusting: The KERN 434 Mechanical Unit operates by a force coil moving inside a permanent magnet. A change in ambient temperature causes a change in the temperature of the Mechanical Unit, in turn altering the characteristics of the magnet. Unless this is compensated for, it will cause sensitivity drift problems. KERN 434 balances use a transistor temperature sensor in the Mechanical Unit to detect changes in temperature. The temperature compensation settings, which match the temperature characteristics of the Mechanical Unit sensor, are stored in Electronic Erasable Programmable Read Only Memory (EEPROM).
All KERN 434 series balances undergo a Temperature Compensation Adjusting at the factory before shipment. At that time all temperature compensations settings are programmed into EEPROM. Therefore, it is not necessary to perform this adjusting regularly to avoid sensitivity drift problems.
EEPROM: If the EEPROM is lost due to a component replacement or short circuit, all adjusting data is cleared. If this happens, the EEPROM must be reinitialised. Reinitialisations of the EEPROM must always be followed by Linearity and Span Adjustings. Depending the user’s needs, a Temperature Compensation Adjusting may also be necessary (see Initialisation of EEPROM, page 39).
434-SH-e-971018
13 Inner A /D Counter Check Mode
Many of the maintenance procedures contained in this manual require adjustments to the Inner A / D Counter Dip Switch, hereby referred to as Dip Switch, located on the Display P.C.B. (MJ-1A).
The Check Modes needed to perform the repair procedures contained in this manual are found below. When a Dip Switch adjustment is required, simply flip the switches so that they correspond to the appropriate Modes below.
1234
Check 0 Mode
(Normal Weighing)
νννν
Off On
1234
Check 1 Mode
(A/D Count Data) (A/D Counter Range 0000000 - 3884160)
ννν
ν
Off On
434-SH-e-9710 19
1234
Check 2 Mode
(Temperature Compensation Data) (A/D Counter Range 180000 - 220000)
Check 3 Mode
(Linearity Data) (A/D Counter Range 0000000 - 3884160)
νν ν
ν
12
34
νν
νν
Off On
Off On
434-SH-e-971020
14 Disassembly / Assembly
14.1 Removing the Top Case
Step 1:
Step 2:
Step 3:
Step 4:
Step 5:
Unplug the AC adapter from the balance.
Remove the weighing pan and the pan support.
Loosen a forehead screw toward the display keyboard and screws on rear side of the balance.
Gently lift off the top cover, flipping it to the right.
Do not jerk the connecting cable.
Disconnect all cables inside the balance.
434-SH-e-9710 21
14.2 Removing the Mechanical Unit
Step 1:
Step 2:
Step 3:
Step 4:
Loosen 3 screws on the bottom of the balance.
Gently lift the Mechanical Unit box from the bottom case.
Slowly place this Mechanical Unit box a flat surface and loosen 4 screws on top of this box.
Open this box, then the Mechanical Unit will be exposed.
434-SH-e-971022
14.3 Mechanical Unit Disassembly
Step 1:
Jig Set (the beam Jig and the beam­suspension Jig) - see the Jig Set page
21.
Step 2:
plate bearing assembly and remove it.
Step 3:
plate bearing assembly and remove it.
Step 4:
vertical plate springs.
Step 5:
the balance weight and remove it.
Step 6:
coupling link.
Attach the appropriate KERN 434
Loosen 4 screws on the upper
Loosen 4 screws on the lower
Loosen 4 screws for removing
Loosen 2 screws on bottom of
Loosen 4 screws for removing a
Step 7:
beam positioning and remove it.
Step 8:
assembly, the suspension assembly and the mechanical unit base will be disassembled separately.
Step 9:
Step 10:
attached on the beam
Step 11:
from the beam. We recommend that you should wear gloves when handling the Force Coil Bobbin.
Loosen a screw on top of the
Make certain that the beam
Remove the magnet lid.
Unsolder the short wire
Remove the Force Coil Bobbin
434-SH-e-9710 23
14.4 Assembly
Step 1:
force coil bobbin attached to the beam .
Step 2:
force coil bobbin is attached to the beam.
Step 3:
positioning jigs in the mechanical unit base.
Step 4:
the right and left posit ioning jigs.
Step 5:
the mechanical unit.
Step 6:
force coil bobbin.
Step 7:
bobbin is completely positioned to have an even space between outer rim of the bobbin and the magnet inside. Then tighten the screws for fixing in between the force coil bobbin and the beam And an edge of the beam will be positioned in right space inside the photo sensor.
Note that the beam assembly is the
Slightly loosen the screws where the
Be laid and insert the right and left
Mount the suspension assembly on
Gently mount the beam assembly on
Insert the force coil bobbin j ig into the
Make certain that the force coil
Step 8:
suspension jigs by tightening the screws.
Step 9:
tighten the beam stopper with the beam stopper jig.
Step 10:
that the beam stopper will be installed at an proper height with the beam stopper jig.
Step 11:
then removing the mechanical unit and then removing the top case in reverse order to be complete.
Reattach the beam and the
Resolder the short wire and insert to
Make certain to tighten the screws
Proceed steps of the disassembly,
434-SH-e-971024
15 Jig Set
You should never attempt to replace Plate Bearings or disassemble the Mechanical Unit without first attaching the appropriate KERN 434 Jig Set. Even slight movements or shifts in the Mechanical Unit components can seriously hamper the performance and accuracy of the balance. The Jig Set’s primary purpose, therefore, is to secure the Mechanical Unit to prevent such movements.
434-SH-e-9710 25
16 Additional Disassembly / Assembly
16.1 Replacement of the Force Coil Bobbin
Step 1:
orders of the Disassembly / Assembly section page 21).
Step 2:
toward the force coil bobbin on the MJ-6 P.C.B. (Allow this small P.C.B. was attached on center of the beam).
Step 3:
from the beam by loosening the screws. Allow counter weight(s) is also mounted on one or two of the same screw holes.
Step 4:
were always isolated and attached in order on a new force coil bobbin.
Step 5:
reattaching the new force coil bobbin below the beam.
Completely follow all procedure
Solder to release 4 lead wires
Remove the force coil bobbin
Make certain that 4 lead wires
Slightly tighten the screws when
16.2 Cleaning Inside the Magnet
Step 6:
have to be resoldered exactly in order to the MJ-6 P.C.B.
Step 1:
attaching inside the magnet.
Step 2:
the magnet by air or adhesive tape.
make certain that 4 lead wires
Make certain that any dust is
Always blow out or clean inside
434-SH-e-971026
16.3 Positioning Edge of the Beam Inside the Photo Sensor
Edge of the have has to be positioned 1 or 1.5 mm space inside the photo sensor.
434-SH-e-9710 27
17 Primary Checking Procedure
When checking a damaged balance, there are primarily 3 major checking procedures that should be followed in order until the damage is detected and corrected.
1. Visual Check (page 28)
2. Mechanical Fault Finding (page 29)
3. Electronic Fault Finding (page 49)
Most repair problems that you will encounter will be a result of damage to the Mechanical unit during transport or rough usage. Occasionally you will also encounter problems resulting from damage to the Electronic Unit. Please carefully read the Disassembly / Assembly section on page 21 - 24 before going through the repair procedure.
18 Visual Check
A careful Visual Check should always be performed before attempting further repairs. If damage is detected during the Visual Check, replace the damaged unit.
Step 1:
Step 2:
Step 3:
Step 4:
Balance Case and Chassis - There should be no cracks or gaps in the balance housing. Check that all housing screws are in place.
Keyboard - They keyboard pad should be free of cracks or tears.
Weighing Pan - The weighing pan and pan support should be free of dents or bends and should be parallel to the balance case.
Level Vial - The level should be filled with fluid and the leveling bubble should be within the circle at the top. Use the adjustable feet to level the balance.
Step 5:
RS-232C Interface - Check for any damage, including dirt in the connectors.
434-SH-e-971028
19 Mechanical Fault Finding
This section provides a simple fault finding method when checking a balance for Mechanical Unit damage. When mechanical damage is present, there are two major problems that can occur when the power switch is turned on. First, the display may show only a decimal point with a polarity sign (+ / -), "all segment lines" message or no display at all. The second problem is when the display shows and unstable weighing result.
19.1 "+ / - .", "All Segment Lines" or No Display
The procedure below is designed to fix a balance that shows the following displays when the power switch is turned on.
(Only a decimal point and + or - sign)
Step 1:
Step 2:
Step 3:
Step 4:
(All Segment Lines)
(No Display)
Make certain that the Short Wires are correctly soldered (page 33) .
Make certain that all Connecting Cabl es are completel y connected ( page 21) .
Make certain that all Plate Bearings and the Coupling Link are not damaged, twisted or broken (page 35) .
Reprogram the Linearity into EEPROM by doing a Linearity Adjusting (page
41).
434-SH-e-9710 29
Step 5:
Return to the normal weighi ng mode by setting Dip Swit ch 2 to Check 0 Mode, and make certain that the three error messages above are gone and that the balance shows a correct weig hing di splay . If not, check all circui t boards.
Step 6:
Step 7:
19.2 Unstable Weighing Results Adjustment
The procedure below is designed to fix a balance that shows unstable weighing results after the power switch has been turned on. Fox example, when a 100 g sample is placed on the KERN 434-23 and the display shows a reading that rapidly shifts from one reading to another the following adjustment must be made to bring the display back to a stable reading of
100.00 g.
If the display shows an unstable weighing result, proceed to the Unstable Weighing Results Adjustment procedure on page 30.
If the display shows a stable weighing reading, the problem has been corrected. Do a Span Adjusting (page 43) before beginni ng to w eig h.
Step 1:
Step 2:
Step 3:
Step 4:
Step 5:
Make certain that all Plate Bearings and the Coupling Link are not damaged, twisted or broken (page 35).
Adjust the Inner A / D Counter (page 19).
Adjust the corner error by doing a Cornerload Adjustment (page 47 - 48).
Reprogram the Linearity into EEPROM by doing a Li nearity Adj usting (pag e 41).
Do a Span Adjusting (page 43).
434-SH-e-971030
Step 6:
If the problem has been corrected, the balance will display a "
Pass
"
message. If a "
" message does not appear, check all circuit boards.
Pass
434-SH-e-9710 31
20 Mechanical Fault Finding Chart
434-SH-e-971032
21 Short Wire Check
The Short Wire (two wires) is used to conduct electronic current from the Photo Sensor to the Force Coil Bobbin. The voltage that reaches the Bobbin via the Short Wires provides the magnetic power required for the Beam to reach equilibrium when a sample is placed on the Weighing Pan.
Step 1:
Step 2:
Step 3:
Remove the top cover (see Disassembly / Assembly page 21).
Check the Short Wire to make sure that it is securely soldered.
If the Short Wire is disconnected, then resolder. When resoldering, maintain contact with the soldering iron for only 1 second to avoid melting the wire.
434-SH-e-9710 33
22 Connecting Cable Check
Step 1:
Step 2:
Step 3:
Remove the top cover (see Disassembly / Assembly page 21).
Check all Cables for damage. Take special notice of all bending parts and Cable Connectors.
Make certain that all Connecting Cables are properly connected. Cables should be checked in the order given in the illustration below.
434-SH-e-971034
23 Plate Bearing / Coupling Link Check
To replace Plate Bearings or Coupling Links you must have the Jig Set used exclusively on KERN 434 series balances. Do not attempt to disassemble the Mechanical Unit or replace Plate Bearings without the appropriate Jig Set. Please read the Disassembly / Assembly and Jig Set sections (page 25) before attempting the steps below.
Step 1:
Step 2:
Step 3:
Step 4:
Step 5:
Step 6:
Remove the Mechanical Unit from the base unit.
Check all Plate Bearings and the Coupling Link for bending, breakage or other damage.
If replacement is required, install the appropriate KERN 434 Jig Set on the Mechanical Unit. Make certain that the Jig Set has been properly attached.
Remove the damaged part from the Mechanical Unit.
Replace the Plate Bearing or Coupling Link. Visually check to be sure the new part is perfectly straight and even.
Remove Jig Set and return Mechanical Unit to the balance housing.
434-SH-e-9710 35
Remark: When reattaching plate bearing(s), then your tightening torgue has to be exactly the same compared to other tightening torgues.
KERN 434-23, 434-33 KERN 434-37
1 Nm (10 kgfcm)
1.8 Nm (18 kgfcm)
24 Beam Stopper Adjustment
This procedure is required to adjust the up-down movement range of the Beam in order to get a proper analogue count. The Photo Sensor will catch the up­down movement of the Beam and determine the proper supply of electronic current sent to the magnet.
Step 1:
Step 2:
Step 3:
Step 4:
Step 5:
Remove the Balance Case (see Assembly / Disassembly page 21 - 24-).
Make certain that the beam stopper has to be set at exact height by using the beam stopper jig.
Reassemble the scale as sh own in the illustration. And make certain that all connecting cables have to be set completely.
Re-plug AC adapter cord to the balance.
Contact a Digital Tester to Ground and Contact Point TP 2 on the MJ-1 P.C.B.
434-SH-e-971036
Step 6:
Allow an adjustment hole right behind the pan suspension.
Step 7:
Step 8:
&
&
Insert a screw driver in the hole and screw up or down the extension piece space until the acceptable voltage + 3 V appears on the Tester.
Absolute voltage has to be ± 3 V even comparing No Load and Full Load.
- Contact Pin to GND + Contact Point to TP 2
Acceptable Voltage Range
No Load Full Load
+ 3 V ± 1 V
- 3 V ± 1 V
434-SH-e-9710 37
25 A / D Converter Check
When checking the A / D Converter you will be looking at the inner A / C count from the Mechanical Unit. Three weighing conditions will be tested and at the appropriate step you will be reading a number on the display and comparing it to Table A below.
Table A
Model Adjusting Masses A / D Count Value
No Load A B
138720 - 971040
KERN 434-23
KERN 434-33
KERN 434-37
Step 1:
Set the balance to Check 3 Mode (see Inner A / D Counter Check Modes page 19).
Step 2:
Turn the power switch
0 g 250 g 300 g 1720128 - 2164032
1803360 - 2635680 138720 - 971040
0 g 250 g 500 g 1720128 - 2164032
2913120 - 3745440 138720 - 971040
0 g 1500 g 3000 g 1720128 - 2164032
2913120 - 3745440
.
O
N
Step 3:
Step 4:
Make certain the balance has no load on the pan. Check to make certain that A / D count value will be appeared within the correct range.
Continuously place different adjusting masses A & B in accordance with Table A. And make certain that A / D count values are displayed as same as the Table A shows. If wrong A / D count value appears, then you must readjust to replace counter weight(s) on the force coil bobbin.
434-SH-e-971038
Step 5:
Remove to readjust the counter weight if attached for having the correct value. If not attached, then reattach the weight.
26 Initialisation of EEPROM
This procedure is necessary only when a complete loss of memory has occurred. A loss of memory can occur from a component replacement, short circuit or other electronic problem. To successfully complete this procedure you will need to initialise the RAM, do a Linearity Adjusting and a Span Adjusting. You may also need to do a Temperature Compensation Adjusting.
Step 1:
Step 2:
Readjust the Linearity Adjusting (see Linearity Adjusting page 41).
Readjust the Span Adjusting (see Span Adjusting page 43).
434-SH-e-9710 39
27 Linearity Adjusting
The purpose of Linearity Adjusting is to create a straight line from Zero to the Full Capacity Scale Value so that the balance will display accurate weighing results at all values in between. For this adjusting procedure, please use a high quality, non-magnetic-metric, stainless steel „Standard Mass“ that is OIML Class F2 or better. Whenever Linearity Adjusting is performed on the balance, it must be followed by a Span Adjusting.
27.1 Linearity Adjusting Procedure
Step 1:
The balance should be fully warmed-up (plugged in for 30 minutes) and all of the Best Conditions for Weighing (page 7) should be met.
Step 2:
With the power switch
O
, reset
FF
Dip-Switch 2 & 3 to the Check 3 Mode.
Step 3:
Turn the power switch
O
while
N
pressing the F key. The current A / D convert value will appear on the display.
Step 4:
Press the
T
ARE
key for proceeding the procedure steps. Allow that "
Lin CAL
Step 5:
Press the
" appears on the display.
key. Allow that
T
ARE
"
Step 6:
"
Step 7:
" appears on the display.
ZEro
Press the
HALF
" appears on the display.
Carefully place an adequate size
key. Allow that
T
ARE
of the adjusting weight A on the weighing pan. See the Adjusting Masses Chart.
434-SH-e-971040
27.2 Adjusting Masses Chart
Model Adjusting Mass Size
A - Half Size B - Full Size KERN 434-23 KERN 434-33 KERN 434-37
150 g 300 g 250 g 500 g
1500 g 3000 g
Step 8:
"
FULL
" appears on the display.
Step 9:
Press the
Carefully place an adequate size
key. Allow that
T
ARE
of the adjusting mass B on the weighing pan. See the Adjusting Masses Chart.
Step 10:
"
PASS
" appears on the display.
Press the
key. Allow that
T
ARE
The above procedure is successfully completed.
If "
CAL Err
" appears, then check to make certain the procedure was followed correctly including proper size adjusting mass.
434-SH-e-9710 41
28 Span Adjusting
The balance should be adjusted for span when it is first installed, any time it is moved or bumped, whenever the ambient temperature changes by more than 3° C, and additionally every 30 days or so. For this adjusting procedure, please use a high quality, non-magnetic, metric, stainless steel "Standard Mass" that is OIML Class F2 or better.
28.1 Adjusting Masses Chart
Model Adjusting Masses
A B KERN 434-23 KERN 434-33 KERN 434-37
300 g 200 g 500 g 300 g
3000 g 2000 g
Step 1:
The balance should be fully warmed-up (plugged in for 30 minutes) and all of the Best Conditions for Weighing (page 7) should be met.
Step 2:
Reset Dip Switch 2 to the Check
0 Mode (normal weighing).
Step 3:
until
Step 4:
Continuously press the F key
"CAL"
appears on the display.
Press the
key. Allow that
T
ARE
"CAL 0"
Step 5:
appears on the display.
Press the
key. The balance
T
ARE
automatically sets Zero. Do not disturb the balance when pressing
T
ARE
key.
the
434-SH-e-971042
(In case of KERN 434-37)
Step 6:
the appropriate size adjusting masses. Carefully place the adjusting masses on the weighing pan.
Step 7:
Step 8:
adjusting mass A size as same as a full capacity value. However, if unfortunately you have limited numbers of the adjusting mass, then choose this secondary procedure as follows the adjusting mass B size.
The balanced will then request
Press the U key.
It is better to place an adequate
Step 9: "PASS"
when adjusting is complete.
appears on the display
434-SH-e-9710 43
28.2 Adjusting Mass Tolerance Collection
If users desire to request for you to readjust KERN 434 balances by standard masses, it is common to know that Standard Masses have always been certified with an exact tolerance by your local Measurement Authority, then readjust the span of the KERN 434 balances with this exact tolerance as follows the below additional procedures.
Step 1:
until
Step 2:
Continuously press the F key
"CAL"
appears on the display.
Press the U key. Allow the deviation number appear on the display. Proceed to readjust a deviation value as calculation of programmed adjusting value - true value of the standard mass.
For instance, in case of KERN 434-37, programmed adjusting value is 2000.00 g and true value of the standard mass is
1999.85 g (you have to round off the numeric figures below than 1 mg readings). The deviation is + 0.15 g equal to programmed deviation 15.
Step 3:
Press continuously the U key for resetting the deviation value. Maximum setting value is up to ± 25 deviation value.
Step 4:
If setting value is - 15, then press the P key. Minus polarity will appear on the display for resetting negative values.
Step 5:
Press the
setting. Allow
"CAL 0"
key for this
T
ARE
to appear on the display. Then this procedure is complete. Follow the same steps of the span adjusting procedure.
434-SH-e-971044
28.3 Disable the Span Adjusting Function
If the user does not require to have the adjusting function on KERN 434 balances, then simply reset a combination of the dip switches to disable this function.
Dip Switch
1234
Step 1:
Step 2:
ννν
Off On
ν
Open the top case, and allow A / D counter has been set to be Check 0 Mode (see Inner A / C Counter Check Mode, page 19).
Reset No. 4 of the dip switch has to be
O
.
N
434-SH-e-9710 45
29 Cornerload Adjustment
Since the weighing pan is connected to the balance through one central point, as you move away from the centre toward the outer rim of the pan, mechanical distortions can occur, thus reducing the balance’s accuracy. Cornerload Adjustment is performed to compensate for the problem.
Step 1:
Step 2:
Step 3:
Step 4:
Step 5:
Step 6:
Step 7:
The balance must be fully warmed-up (plugged in for 30 minutes) before starting.
Remove the top case (see Disassembly / Assembly page 21 - 24).
Replace the weighing pan.
Make certain that the balance is level.
Plug the AC adapter into the power source.
Press the
ON/O
key to get a normal weighing display (Check 0 Mode).
FF
Place the appropriate Cornerload Adjustment Mass in the centre of the weighing pan (Point '). Place it around the pan. When using the draft shield, make certain that it does not touch the pan.
29.1 Cornerload Adjustment Masses
Model Masses KERN 434-23 KERN 434-33 KERN 434-37
Step 8:
Step 9:
Press the
Move the Cornerload Mass to Point ( and make a note of the display
T
ARE
key.
150 g 250 g
1500 g
reading.
434-SH-e-971046
Step 11:
Move the Cornerload Mass to Point ) and make a note of the display reading.
Step 12:
Compare the absolute values (ignore "+" or "-" signs) of the display readings at Point ( and Point ) and determine which is the greater deviation from zero. The side that is greater must be adjusted first. For example, if you get results similar to those left, then Point
(
must be adjusted.
Step 13:
Make the adjustment by slightly turning the appropriate Adjustment Screw shown left with a box driver (Screw B to adjust Point ) result and Screw A to adjust Point ( result). When the result is positive, "+", like the Point ( above, turn the screw counter-clockwise. When negative, "-", turn clockwise.
Step 14:
Repeat Steps 7 to 13 until both Point ( and ) display zero ± 3 deviations.
Step 15:
push the
Place the mass at Point ' and
key. Check Points ( to
T
ARE
*
and ) to + as diagonal to be certain that they are all at Zero ± 3 deviations.
434-SH-e-9710 47
30 Electronic Fault Finding
The following procedures are used to detect and repair damage to the Electronic Unit. Before proceeding with any of the repair procedures below, it is advisable to make a thorough visual check of the Electronic Unit. Look for dirt or other foreign objects throughout the balance. Check for objects touching the circuitry, broken circuit paths, grounding, solder dry joints or any other damage to the P.C. Boards.
30.1 Power Check
Fuse:
for touching objects. Also check the power supply electronics.
AD Adapter:
before shipment. If you find that the AC adapter is providing too little voltage or that it generates irregular heat, check the attached voltage label to make sure the proper voltage for your local power.
30.2 EEPROM Check
If the EEPROM (Electronic Erasable Programmable Read Only Memory) has been lost, the balance will not show the normal weighing display when the power switch is turned on. One of the following three messages will appear:
If the fuse (0.5 AMP) keeps blowing, there is a short circuit. Check the circuitry
The proper voltage AC adapter for your local power supply is provided
434-SH-e-971048
Step 1:
Attempt to restore the linearity program into EEPROM by using the procedures given in the Linearity Adjusting section on page 41.
Step 2:
30.3 Voltage Check
There are a number of voltage checks that can be carried out on the P.C. Boards for pinpointing faulty electronics. The main checking points (1A, 2A, 3A, 4A, 5A) are on the power circuit MJ-3. Always check the voltage range first. Other points are followed by listed numbers on the Electronic Fault Finding Chart (page 47) and attached Circuit Diagrams (hereafter).
If the above procedure does not solve the problem, then replace the EEPROM IC chip (IC 8).
434-SH-e-9710 49
31 Electronic Fault Finding Charts
434-SH-e-971050
434-SH-e-9710 51
434-SH-e-971052
434-SH-e-9710 53
434-SH-e-971054
32 Wave Form Check
If everything is correct in the Electronic Fault Finding Chart but the displayed value is still incorrect, then either the CPU, the EEPROM or the SCA is at fault. We will be able to tell by the wave form that is generated during testing. See the Wave Form Table for the testing points. If an incorrect wave form is found, then replace the unit that is the generator of the signal.
We recommend to use an oscilloscope (measuring capacity more than DC 100 MHz) and a digital tester (readability is 0.1
µµµµ
A).
P.C.B. No. Location Pinpoint No. Wave Form
1D 2D 3D 4D 5D
1C 2C 3C 9C
10C
MJ-1
MJ-1
IC 10 pin 3 IC 10 pin 5
CN 4 pin 12
CN 4 pin13
CN 4 pin 11
IC 11 pin 4 IC 11 pin 3
IC 11 pin 2 Resistor 37 Resistor 38
MJ-1 IC 13 pin 2 4C
MJ-1 Transistor 14 11C
434-SH-e-9710 55
P.C.B. No. Location Pinpoint No. Wave Form
IC 13 pin 3
MJ-1
MJ-3 IC 1 1A
IC 12 pin 7
IC 12 pin 14
IC 12 pin 8
IC 11 pin 1 12C
Condenser 6 2A
5C 6C 7C 8C
Condenser
73A
Diode 1 Diode 2
4A 5A
434-SH-e-971056
33 RS-232 C Interface
The data transmission RS-232 C Interface allows the user to use KERN 434 balances in conjunction with a computer, printer or other peripheral device. The Interface circuit is already mounted in the balance as a standard accessory.
Please read the following section and the instruction manuals of related equipment before using this device. The RS-232 C Interface is designed for exclusive use on KERN Balances and peripheral equipment designed by or recommended by KERN & Sohn GmbH. Use of this Interface with any other type of equipment is solely the responsibility of the user.
33.1 Interface Specifications
Type Method Speed System
Signal
RS-232 C Half Duplex Transmission 1200, 2400, 4800 BPS Asynchronous system 8 bit length, non-parity, 1 Start bit, 1 Stop bit 7 bit length, even-parity, 1 Start bit, 1 Stop bit
434-SH-e-9710 57
Configuration of Receptacle pins
Plug: Circle Type
Miniature connector TCP8580 Type
Receptacle: Circle Type
Miniature connector TCS75850-01-101
434-SH-e-971058
33.2 Output Format
1. Method
±±±±
9 characters decimal point inclusive
Polarity Unit
Total:
Unit I: Unit II: g:
14 characters
Space or unit Non-stable data Gram & other units
IIICRLF
2. Example, data output
Gram weighing
+ 3000.00 gCRLF
Percent weighing
+ 100.00 %CRLF
Piece counting
+100PCCRLF
434-SH-e-9710 59
3. Example, data output
Gram weighing
+.CRLF
The other weighing units
+CRLF
4. There will be no output when the balance is under following conditions:
The function data is being displayed.
Π
"All segments" display is on.
Π
The adjusting mark is being displayed.
Π
The stand-by mark is being displayed.
Π
The piece counting & percent weighing marks are being displayed.
Π
33.3 Output Data Mode
(Out 1)
(Out 2)
Transmits one data upon the print command when the balance is stabilised.
Step 1:
Step 2:
Step 3:
Transmits one data upon the print command as soon as the balance has stabilised.
Step 1:
Step 2:
Place a sample on the weighing pan.
Allow reading to stabilise.
Press the P key. The balance will output one data.
Place a sample on the weighing pan.
Press the P key. The balance will output one data as soon as the reading has stabilised.
434-SH-e-971060
(Out 3)
Transmits one data upon the print command even if the balance has not stabilised.
(Out 4)
(Out 5)
Step 1:
Step 2:
Place a sample on the weighing pan.
Press the P key. The balance will output one data even if the reading has not stabilised.
Automatically transmits one data when the balance has stabilised.
Step 1:
Step 2:
Place a sample on the weighing pan.
When the reading stabilises the balance will automatically output one data without pressing the P key.
Step 3:
Step 4:
Remove the sample and place another sample on the pan.
The balance will automatically output another data after the
reading has stabilised. Automatically transmits all data regardless of whether the balance has stabilised or not.
Step 1:
Place a sample on the weighing pan. The balance will
automatically output all data.
33.4 External Control Commands
The command (ASCII) will be transmitted into the balance through the miniature connector No.3 pin when pressing the appropriate code on the peripheral keyboard.
Command (ASCII) Functions
"D"
"Z"
"R"
"F"
Functions the same as pressing the P key. Functions the same as pressing the T
ARE
Functions the same as pressing the U key. Functions the same as pressing the F key.
key.
434-SH-e-9710 61
33.5 Cabling Diagrams
For connecting a computer or a peripheral device to the balance.
434-SH-e-971062
34 Installation of the Air Shield Case (OMJ-2 Option)
We recommend the installation of an air shield case whenever your operation will be held under unstable weighing conditions.
Remove the weighing pan and the pan support from the balance.
Loosen the 4 screws on top of the balance case.
Remove the dust ring from the top case.
Install the Air Shield Case by placing it squarely on the top of the balance.
Replace the dust ring on top of the air shield under panel.
Replace only the 2 outside screws and tighten them snugly into place.
Replace the weighing pan and the pan support.
434-SH-e-9710 63
35 Troubleshooting
Problem Research Cause Possible Solution
A) No display when
plug AC adapter into power.
B) No display after
count display when plug AC adapter into power.
C) Error message
display "Error 3" when plug AC adapter into power.
(1) Problem with check
point 4A, 5A voltage on MJ-3 P.C.B. Before checkin g mu st disconnect CN 1 connector.
(2) Problem with check
point 1A, 2A, 3A when correct voltage on check point 4A, 5A on MJ-3 P.C.B. Before checking must disconnect CN 1 connector.
(3) No display when
check mode 1.
(4) Problem with check
point 1D on MJ-1 P.C.B.
(5) Problem with check
point 2D on MJ-1 P.C.B.
(6) No signal at check
point 3D, 4D on MJ-1 P.C.B.
(7) No signal at check
point 5D on MJ-1 P.C.B.
(1) No display of co rrect
data when check mode 1.
(1) Correct display when
re-plug into power.
(2) Always error message
display "Erro r 3".
AC adapter malfunction. Blown fuse. Power Jack (JK) malfunction. MJ-3 P.C.B. malfunction.
IC 1, IC 2, IC 3 malfunction. C 5, C6, C7 malfunction or D3, D4, D5 malfunction on MJ-3 P.C.B.
Power cable malfunction. Disconnect power cable. Disconnect display cable.
Quarts malfunction. Replace the defective
IC 10 malfunction. Replace the defective
IC 5 malfunction. Replace the defective
IC 6 malfunction. Replace the defective
IC 6 malfunction on MJ-1 P.C.B.
IC 9 malfunction on MJ-1 P.C.B.
IC 8 malfunction on MJ-1 P.C.B.
Replace the defective parts.
Replace the defective parts.
Replace the new parts. Re-plug the power cable. Re-plug the display cable.
parts.
parts.
parts.
parts.
Replace the defective parts.
Replace the defective parts.
Replace the defective parts.
434-SH-e-971064
Problem Research Cause Possible Solution
D) Continuous flashing
display after all segment line when plug into power
E) "+." or "-." appears
after all segment line when AC adapter was plugged.
(3) No continuous error
message.
(1) Check whether zero
display is unsteady after pushing down the weighing pan to end.
(2) Problem with check
point 2A, 3A on MJ-3 P.C.B.
(3) No signal at check
point 5B.
(4) No triangle wave at
check point 4B.
(1) Problem with voltage
at check point 2A, 3A on MJ-3 P.C.B.
(2) Cable disconnection. Disconnect Photo sensor
(3) "CAL Err" appears A / D count value is
(4) Visual check. Object attaching weighing
(5) Problem with voltage
at check point 1B, 2B or 3B on MJ-1 P.C.B.
Caused by electronic noise.
Program data is unstable by draft, vibration, etc.
Dust cover attach ing weighing pan.
Power supply is fluctuated.
AC adapter malfunction. IC 2, IC 3 malfunction on
MJ-3 P.C.B. D1, D2 malfunction on
MJ-3 P.C.B. C3, C4 malfunction on
MJ-3 P.C.B. IC 6 (SCA) malfunction
on MJ-1 P.C.B. Tr 7, Tr8, C 13
malfunction on MJ-1. Same as D) 2. Same as D) 2.
cable.
wrong. Linearity program is
wrong.
pan. Defective plate bearing or
coupling link. ZD, IC 1, IC 2, Tr 1, Tr 2
malfunction.
Attach a win d shi el d case on the balance.
Install the balance on an anti-vibration table.
Set up the dust cover again.
Replace AC adapter.
Replace the defective parts.
Replace the defective parts.
Replace the defective parts.
Reconnect the cable again.
Recheck the A / D count value.
Restore the correct linearity program.
Clean around the weighing pan.
Replace the defective parts.
Replace the defective parts.
434-SH-e-9710 65
Problem Research Cause Possible Solution
F) Irregular span
adjusting.
(6) A / D count value at
Check 3 Mode is "0000000" or "3884160"
(7) Problem with TP 1,
TP 2.
(1) Problem with A / D
count value at Check 2 Mode.
(2) No signal at check
point 11C.
(3) Problem with signal at
check point 1C, 2C, 3C.
Defective plate bearings or coupling link.
Broken wire of force coil. Defective photo sensor. IC 3 malfunction on MJ-1
P.C.B. IC 4 malfunction on MJ-1
P.C.B. Broken wire of
temperature sensor or disconnecting the connector.
Problem with IC 10 on MJ-1 P.C.B.
Problem with IC 11, IC
13.
Replace the defective parts.
Replace the defective parts.
Replace the defective parts.
Resolder the wire or reconnecting the connector.
G) When increasing
load on the weighing pan, display does
not allow the corresponding results in half way.
H) No corresponding
results when loading on the weighing pan.
(4) Problem with voltage
at check point 5C, 6C, 7C, 8C.
(5) Problem with voltage
at check point 4C.
(6) Recheck plate
bearings or coupling link.
(1) No electricity into forth
coil wire.
(1) Problem with AC
adapter.
Problem with IC 12, Tr 11, Tr 12.
Problem with Tr 13, Tr 14, Tr 15.
Problem with plate bearings or coupling link.
Broken wire of the forth coil.
Problem with Tr 3, Tr 4, Tr 5, Tr 6.
Misuse of the wrong voltage AC adapter.
Replace the defective parts.
Replace the defective parts.
Replace the defective parts.
Replace a new AC adapter.
434-SH-e-971066
Problem Research Cause Possible Solution
I) Very poor
reproducibility.
(1) Recheck weighing
condition for effection of any vibration or air draft.
(2) Unstable installation
of the balance.
Air draft (by air condition or by vibration which caused by somebody walking near the balance).
Vibration (caused by machine operation or car moving outside).
Surface of the balance table is not completely flat.
Attach Air Shield Case on the balance.
Replace the balance somewhere in stable condition and on anti­vibration table.
434-SH-e-9710 67
36 KERN 434 Series Parts List
No. Description Qty 434-23 434-33 434-37
101 Pan holder (type 1) 1 325013 325013 - ­102 Packing cushion (left) 1 325010 325010 325010 103 Outer packing case 1 325009 325009 325009 104 Packing cushion (right) 1 325011 325011 325011 105 Shipping lock spacer 1 211013 211013 - ­106 Shipping lock 1 240232 240232 240232 108 Pan holder (type 2) 1 - - - - 325014 109 AC adapter box 1 325012 325012 325012 111 AC adapter (220 – 240 V) 1 122107 122107 122107
201 Weighing pan 1 230205 230205 - ­203 Pan support 1 240210 240210 - ­204 Pan support peg 1 240405 240405 - ­205 Weighing pan 1 - - - - 210202
207 Pan support 1 - - - - 240246 208 Pan support peg 1 - - - - 240406 211 Dust ring 1 240238 240238 240238 212 Dust cover 1 210906 210906 210906 213 Hook connection 1 210712 210712 210712 214 Underweighing hook 1 210712 210712 210712
301 Panel keyboard 1 121945 121945 121945 302 Temperature sensor cable 1 123015 123015 123015 303 Power cable 1 123009 123009 123009 304 Display board 1 130109-2 130110-2 130111-2 305 Display board guide 2 210815 210815 210815 306 Display board cable 1 123010 123010 123010 307 Photo sensor cable 1 123011 123011 123011 308 Main circuit board 1 130109-1 130110-1 130111-1 310 Case shaft 1 210463 210463 210463 311 Bed 1 240103 240103 240103 312 Adjustable legs 2 210809 210809 210809
434-SH-e-971068
No. Description Qty 434-23 434-33 434-37
315 Underweighing cap 2 210813 210813 210813 317 Circuit board stand (A) 2 210464 210464 210464 318 Circuit board stand (B) 2 210465 210465 210465 319 Spirit level washer 1 211106 211106 211106 320 Spirit level 1 211121 211121 211121 322 Data output board 1 130109-4 130110-4 130111-4 323 Rear socket plate (right) 1 210329 210329 210329 325 Data output cable 1 123009 123009 123009 326 Power circuit board 1 130109-3 130110-3 130111-3 327 Rear socket plate (left) 1 210330 210330 210330 329 Top case 1 210812 210812 210812 330 Spirit level window 1 211010 211010 211010
402 Transducer shield (upper) 1 210328 210328 210328 403 Magnet cover 1 240332 240332 240332 404 Upper parallel guide 1 210506 210506 210508 405 Parallel plate bearing 8 211214 211214 211215 407 Force coil board 1 120114 120114 120114 409 Beam plate bearing 2 2112166 2112166 211214 414 Beam extension piece 1 210239 210239 210239 415 Extension piece spacer 1 210470 210470 210470 419 Short wire (silver) 3 113107 113107 113107 420 Photo sensor board 1 120113 120113 120113 429 Suspension spring 1 210717 210717 210718 431 Coupling link 1 211204 211204 211205 432 Lower parallel guide 1 210507 210507 210509 434 Corner adjustment spring 2 210715 210715 210715 435 Corner adjust screw (B) 2 210467 210467 210467 436 Corner adjust screw (A) 2 210466 210466 210466 437 Beam stopper spring 1 210716 210716 210716 438 Beam stopper 1 241011 241011 241011 440 Force coil bobbin 1 251506 251506 251506 441 Beam counter weight 0-3 210473 210473 210473 444 Plate bearing spacer 16 - - - - 210472
434-SH-e-9710 69
No. Description Qty 434-23 434-33 434-37
447 Temperature sensor cable 1 123015 123015 123015 448 Balance weight 1 211634 211634 211634 451 Coupling link spacer 2 210398 210398 210398 453 Transducer shield (lower) 1 240397 240397 240397
500 434 series complete jig set 1 310102 310102 310102 501 Force co il jig 1 - - - - - ­502 Beam jig 2 - - - - - ­503 Suspension jig 2 - - - - - ­505 Left positioning jig 1 - - - - - ­506 Right positioning jig 1 - - - - - ­507 Beam stopper jig 1 - - - - - -
434-SH-e-971070
37 Balance drawings
434-SH-e-9710 71
434-SH-e-971072
434-SH-e-9710 73
434-SH-e-971074
434-SH-e-9710 75
Loading...