Kern 434 Service manual

Service Manual for Electronic
Precision Balances series
434
Version 1.0 7/97
434-SH-e-9710
Table of Contents
1 Specifications __________________________________________________________4 2 Introduction ___________________________________________________________5 3 Principles of operation ___________________________________________________5 4 Block Diagram _________________________________________________________6 5 Best Conditions for Weighing _____________________________________________7 6 Balance Functions ______________________________________________________8 7 Description of Function Menu_____________________________________________9 8 Description of Functions ________________________________________________10 9 Weighing Unit Initialisation _____________________________________________11
9.1 Keyboard Method_________________________________________________________ 11
9.2 Parameter Number Chart __________________________________________________ 12
10 Piece Counting Mode ___________________________________________________13
10.1 Removing establishment of the standard unit weight for piece counting__________ 14
11 Percent Mode _________________________________________________________15
11.1 Weighing Mixture Mode _________________________________________________ 15
12 Adjusting Introduction__________________________________________________17 13 Inner A /D Counter Check Mode__________________________________________19 14 Disassembly / Assembly _________________________________________________21
14.1 Removing the Top Case__________________________________________________ 21
14.2 Removing the Mechanical Unit____________________________________________ 22
14.3 Mechanical Unit Disassembly_____________________________________________ 23
14.4 Assembly ______________________________________________________________ 24
15 Jig Set _______________________________________________________________25 16 Additional Disassembly / Assembly ________________________________________26
16.1 Replacement of the Force Coil Bobbin _____________________________________ 26
16.2 Cleaning Inside the Magnet ______________________________________________ 26
16.3 Positioning Edge of the Beam Inside the Photo Sensor ________________________ 27
17 Primary Checking Procedure_____________________________________________28 18 Visual Check__________________________________________________________28 19 Mechanical Fault Finding_______________________________________________29
19.1 "+ / - .", "All Segment Lines" or No Display ________________________________ 29
19.2 Unstable Weighing Results Adjustment ____________________________________ 30
20 Mechanical Fault Finding Chart _________________________________________32 21 Short Wire Check ______________________________________________________33 22 Connecting Cable Check ________________________________________________34
434-SH-e-97102
23 Plate Bearing / Coupling Link Check ______________________________________35 24 Beam Stopper Adjustment _______________________________________________36 25 A / D Converter Check __________________________________________________38 26 Initialisation of EEPROM _______________________________________________39 27 Linearity Adjusting_____________________________________________________40
27.1 Linearity Adjusting Procedure____________________________________________ 40
27.2 Adjusting Masses Chart _________________________________________________ 41
28 Span Adjusting ________________________________________________________42
28.1 Adjusting Masses Chart _________________________________________________ 42
28.2 Adjusting Mass Tolerance Collection ______________________________________ 44
28.3 Disable the Span Adjusting Function_______________________________________ 46
29 Cornerload Adjustment _________________________________________________46
29.1 Cornerload Adjustment Masses ___________________________________________ 46
30 Electronic Fault Finding ________________________________________________48
30.1 Power Check___________________________________________________________ 48
30.2 EEPROM Check _______________________________________________________ 48
30.3 Voltage Check__________________________________________________________ 49
31 Electronic Fault Finding Charts __________________________________________50 32 Wave Form Check _____________________________________________________55 33 RS-232 C Interface_____________________________________________________57
33.1 Interface Specifications __________________________________________________ 57
33.2 Output Format _________________________________________________________ 59
33.3 Output Data Mode ______________________________________________________ 60
33.4 External Control Commands _____________________________________________ 61
33.5 Cabling Diagrams_______________________________________________________ 62
34 Installation of the Air Shield Case (OMJ-2 Option)___________________________63 35 Troubleshooting _______________________________________________________64 36 KERN 434 Series Parts List ______________________________________________68 37 Balance drawings ____________________________ Fehler! Textmarke nicht definiert.
434-SH-e-9710 3
1 Specifications
Model KERN 434-23 KERN 434-33 KERN 434-37 Gram Ounce Pound Carat Pennyweight Troy Ounce Grain Momme Tael (Hong Kong) Tael (Singapore) Tael (Taiwan) Linearity Readability Stabilization time Span drift (15° C - 30°C)
310 x 0.001 510 x 0.001 3100 x 0.001 10 x 0.0005 18 x 0.0005 109 x 0.0005
0.6 x 0.000005 1.1 x 0.000005 6.8 x 0.000005 1550 x 0.005 2550 x 0.005 15500 x 0.005
199 x 0.001 327 x 0.001 1993 x 0.001
9.9 x 0.00005 16 x 0.00005 99 x 0.00005
4784. x 0.02 7870. x 0.02 47840. x 0.02 82 x 0.0005 136 x 0.0005 826 x 0.0005 8 x 0.00005 13 x 0.00005 82 x 0.00005 8 x 0.00005 13 x 0.00005 82 x 0.00005 8 x 0.00005 13 x 0.00005 82 x 0.00005
±
0.002 g
0.001 g 0.01 g
2.3 sec
±
5 ppm/° C
±
0.02 g
±
3 ppm/° C
Operation temperature Pan size Total min. weight of
percent display Min. percent display Total min. weight of
piece counting Max. piece counting
display Sampling number of
piece counting Data output
Power supply Dimensions
Options: Auxiliary display (OMK-0), Underweighing hook (OMJ-1), Air shield case (OMJ-2) Animal weighing pan (OMJ-3), Small animal weighing pan (OMJ-4), Auxiliary display base stand (OMJ-10), Dust cover (O MJ- 11), Auxiliary display stand pole (OMJ-12)
31000 pcs 51000 pcs 31000 pcs
RS-232C bi-direction (standard accessory)
AC Adapter 220 V, 110 V or other (+10 V, 015 V)
135 mm
0.01 g 0.1 g
50/60 Hz Output AC19V 200 mA
194 (W) x 334 (D) x 81 (H) mm
10° C - 40° C
160 x 160 mm
1 g 10 g
0.01 %
5, 20, 100 or 100 pcs
434-SH-e-97104
2 Introduction
This Maintenance Manual covers three models from the KERN 434 series, the KERN 434-23, KERN 434-33 and KERN 434-37. Please read this Maintenance Manual and the owner’s Instruction Manual fully before beginning any maintenance work.
The KERN 434 series precision balances are the products of years of research, development, design and in-field testing. They incorporate the latest advances in mechanical and electronic engineering and offer the highest standard of reliability, easy to use functions and rugged durability.
3 Principles of operation
The KERN 434 series precision balances work on the principle of "Force Compensation". Any change in the load on the weighing pan causes the Beam to pivot on fulcrum Plate Bearings (see Block Diagram page 4). Attached to this Beam is a coil wound with fine wire, called the "Force Coil Bobbin", which floats in a permanent magnet. At the end of the Beam there is a small notch which allows light from a Light-Emitting Diode (LED) to pass through to a Photo Sensor (Light Measuring Diode). At zero weight, the light detected by the Photo Sensor is exactly equal to the light emitted by the LED.
When the end of the beam is forced up by the leverage exerted from a mass placed on the weighing pan, the Photo Sensor detects a change in the position of the beam and the attached Force Coil Bobbin, because the light reaching the Photo Sensor has become less than that emitted by the LED. The balance then feeds the Force Coil Bobbin with more voltage, which increases the magnetic power and pulls the Bobbin downward until the light reaching the Photo Sensor is once again equal to the light emitted by the LED. This is accomplished by a Differential Amplifier, a filter and an Analogue/Digital (A/D) Converter receiving photo current from the Photo Sensor, converting it to voltage and boosting it back to the Force Coil Bobbin.
The electrical current flowing through the Force Coil Bobbin generates a voltage proportional to the load weight on the pan. This is read back through the Differential Amplifier and filter, then the A/D Converter digitalises this voltage. The resulting value is then counted and fed to the microprocessor (CPU). The CPU performs a multitude of commands and mathematical operations in conjunction with parameter and adjusting information stored in Random Access Memory (RAM). Finally, the results are displayed on the Liquid Crystal Display (LCD) or sent to the RS-232 C Interface.
434-SH-e-9710 5
4 Block Diagram
434-SH-e-97106
5 Best Conditions for Weighing
1. Never turn off the power switch or disconnect the AC adapter when in use.
2. When making a measurement, always place the sample in the centre of the weighing pan. Slight errors may result if the sample is not near the centre of the pan.
3. Make sure the balance is level by using the level vial and the adjustable feet on the bottom.
4. Install the balance in a controlled environment. A) The weighing room should be kept clean, dry and free of cigarette
smoke.
B) Protect the balance from drafts (air currents). Use a draft shield if
necessary. C) Maintain the ambient temperature to D) Maintain the ambient humidity to E) If larger changes in temperature or humidity occur, re-calibrate the
balance. F) The balance table should be level and free from excessive vibration.
Corners of rooms are less prone to vibration. G) Do not expose the balance to direct sunlight or radiated heat. Keep away
from windows, heaters, hot plates, flames, fans, air conditioners, etc. H) Allow hot or cold sample containers to come to ambient temperature
before weighing. I) Discharge any statically charged sample before weighing. J) Do not expose the balance to corrosive gases.
±±±±
10 %.
±±±±
3° C.
5. Magnetised samples cannot be weighed accurately on an electronic balance. Keep equipment containing magnets away from the balance.
6. Make certain that the AC power supply is free from electrical disturbances.
7. Clean the balance with a damp cloth only (no solvents).
8. Always warm-up the balance before use or leave on Stand-by (display off) overnight.
9. Always handle the balance with care during use or when moving or storing.
434-SH-e-9710 7
6 Balance Functions
Sampling Time Function (SAP):
The
the surrounding environmental conditions. The
function allows the balance to adapt to
SAP
should be adjusted to a small
SAP
value for corresponds to the "integration time" in seconds that the balance is using to display readings.
Stability Indicator Function (STb):
All KERN 434 series balances use the star (Η) symbol to indicate stability. When the Η appears on the display, the balance has not yet stabilised. When the Η disappears from the display, the balance is stable. The weight reading should be made only after the Η has disappeared from the display and the indicator on the display for different sample types and weighing conditions. Small values fo r
require very stable conditions before the Η disappears from the
STb
display and large values allow the Η to disappear under more reliable conditions such as animal weighing.
Adjusting (Cal):
The
function is used to perform a Span Adjusting on the
Cal
balance. See Span Adjusting page 35 for the steps required for this procedure.
Weighing Unit Selection (Uni):
different weighing units. between units.
Uni-2
allows the user to switch from grams to another unit by pushing
Uni-1
locks in grams or another unit and prevents switching
The
function is used to select and lock in
Uni
the U key. See Weighing Unit Initialisation page 9 to select or change the available weighing units.
Output Data Mode (oUT):
The
function is used to change the output data
oUT
mode when using the RS-232 C interface. See RS-232 C Interface page 54.
RS-232 C Mode (rS):
The rS function allows you to switch between 8 bit non-parity (8N1) and 7 bit even-parity (7E1)when using the RS-232 C Interface. See RS-232 C Interface page 58 - 63
Baud Rate Speed (rSS):
The
function changes the baud rate speed when using
rSS
the RS-232 C Interface. See RS-232 C Interface page 8 for the steps required for this procedure.
Auto-Start Mode (AST-ON):
If
auto-start mode
position immediately without pressing the
ON/S
is on, the balance will be weighing
key.
TBY
434-SH-e-97108
7 Description of Function Menu
0.000g
⇓→
Press the F key.
Press the
!
Press the U key.
T
ARE
key.
SAP 1
Stb 1
CAL
Un i 1
OUT 1
!
!
!
!
"
!
!
2
"
!
2
!
3
3
!
A
H1
!
H2
!
H3
!
0
Full Point Value
Second Point Value
2
!
2
3
"
!
!
4
5
rS 8 n 1
!
"
7 E 1
r S S 1 . 2
!
2 . 4
"
!
4 . 8
AST ON
The display returns to the weighing mode if no operation is made for 3 sec. See the appropriate section for further explanation of individual function
!
"
OFF
434-SH-e-9710 9
8 Description of Functions
Display Command Selectable Setting Factory Set ting Remarks
1 Very steady
Range of 2 Normal
SAP
STb
CAL
Uni
OUT
rS
rSS
AST
Vibration 3 Unstable 2 Note 1
Adapter A Extremely unstable
1 Limited
Range of 2 Standard
Stability 3 Extensive 2 Note 1
H1 HOLD - Limited H2 HOLD - Standard H3 HOLD - Extensive
Adjusting See the adjusting section (page 11). Note 2
Unit Locking the key.
Selection Release the lock. 2 Note 3
Output upon print
1 com m and m ade
only after the balance has stabilised. After print command
2 made, output as
soon as balance has stabilised.
Data Output upon print
Transmission 3 command regard- Note 4
Mode less of stability.
Automatic output as
4 soon as balance
has stabilised. Automatic output
5 regardless of
stability.
RS-232 C 8N1 8 bit Non-parity
Interface 7E1 7 bit Even-parity 8N1 Note 5
Mode
RS-232 C 1.2 1200 BPS
Interface 2.4 2400 BPS 1.2 Note 5
Speed Mode 4.8 4800 BPS
Auto-Start ON Auto-start is on.
Mode OFF Auto-start is off. OFF Note 6
434-SH-e-971010
9 Weighing Unit Initialisation
9.1 Keyboard Method
Step 1:
Make certain that AST function
mode (Automatic start) is off.
Step 2:
To have the count-down display, once disconnect the AC adapter cord and reconnect it again.
Step 3:
keys at the same time when
F
111111111
Step 4:
Press and hold the
“ appears.
Release the keys when the „All
T
ARE
, U and
Segment“ line appears.
Step 5:
Parameter number will appear after „All Segment“ line has stopped flashing.
Step 6:
Two units in addition to grams can be selected and programmed into memory. Grams will automatically be in memory and does not need to be selected in this procedure. Unit 1 is selected by pressing the U key until the unit code place. Unit 2 is selected by pressing the
key until the unit does appear on the display in
T
ARE
the Unit 2 digit place. For example, if you select Units F and E as shown below the balance will show weighing results in Piece Counting and Percent.
Step 7:
Press the F key to store your selections. Check to make certain the desired units have been stored by pressing the U key. When the unit selection (Uni) is set at Uni-2 you can switch between units by pressing the U key. The Uni-1 setting locks in one unit and switching to another unit is not possible.
434-SH-e-9710 11
Step 8:
Press the F key to complete the Unit Selection procedure.
Display Circular
(# Press the U key)
Gram display
$%
9.2 Parameter Number Chart
0 Gram g 9 Carat C.M 1 Cancel last digit g A Momme mom 2 Kilogram kg B Taiwan Tael tl 3 Pound lb C Singapore Tael tl 4 Ounce oz D Hongkong Tael tl 5 Troy ounce ozt E
6Grain gr F
#
Unit 1
#
Unit 2
Η Η
#
Percent % Piece counting pcs
Gram display
7 Pennyweight dwt 8 Carat ct
Factory setting
Η
434-SH-e-971012
10 Piece Counting Mode
Step 1:
Press the U key to switch the display to Piece Counting Mode (see Weighing Unit Initialisation page 11).
Step 2:
Press the F key until the desired count value is displayed, 5, 20, 100 or 1000 pcs.
Step 3:
Count out the same number of sample pieces as you selected in Step 2, and place them on the weighing pan.
Step 4:
Press the
key to memorise
T
ARE
the number of sample pieces and their unit weight. The display will flash until the balance has memorised this data.
Step 5:
Proceed with piece counting.
Step 6:
message when pressing the
If the display shows a
T
ARE
"Lo"
key, the unit weight of the sample pieces is less than the balance can detect.
The minimum weight of each sample piece required to establish piece counting is as follows:
KERN 434-23 KERN 434-33 KERN 434-37
⇑ ⇑ ⇑
0.01 g
0.01 g
0.1 g
434-SH-e-9710 13
10.1 Removing establishment of the standard unit weight for piece counting
This unique function of the KERN 434 series is used to easily increase the number of pieces of sample used as your piece counting standard without taking the time to count out each standard sample separately. This is important because the greater the number of sample pieces used to set your standard, the more accurate an precise your piece counting will be.
Step 1:
sample pieces and unit weight at 5 PCS, f or example, by following the procedure on the previous page.
Step 2:
weighing pan. The display will show that there are 20 pieces on the pan.
Step 3:
pan and press the F key until 20 PCS appears on the display.
Step 4:
new number of sample pieces and the new unit weight.
Step 5:
increase the standard to 100 or 1000 pieces.
Establish the standard number of
Place 15 more sample pieces on the
Leave the 20 pieces on the weighing
Press the
Follow the same procedure to
key to memorise the
Tare
Note: The key is to let the balance do the counting for you when establishing the standard. For example, count out 20 pieces by yourself and set the standard at 20 pieces. Use the piece counting mode to count up to 100 pieces and set the new standard at 100 pieces. Then use the piece counting mode again to count up to 1000 pieces and set the new standard again at 1000 pieces.
434-SH-e-971014
11 Percent Mode
Target weight:
Step 1:
"%" appears.
Step 2:
weight on the weighing pan.
Step 3:
100% weight. Make certain that """ sign appears when entering a sample. Press the
T
ARE
Step 4:
when "100.00%" appears. Proceed with percent weighing by placing a new sample on the weighing pan. Percentage display is
100.00& for KERN 434-23, KERN 434-33 and KERN 434-37.
Continuously press the U key until
Place a sample of the ideal 100%
Press the F key to memorise the
key if the """ sign does not appear.
Percent mode procedure is complete
The minimum weight of the 100% sample required to establish percent weighing is as follows:
KERN 434-23 KERN 434-33 KERN 434-37
⇑ ⇑ ⇑
11.1 Weighing Mixture Mode
This mode is used when you want to mix two sample together. For instance when you add sample B to sample A to make a 100% mixture sample.
1 g 1 g 10 g
Sample B 40 % Sample A 60 %
434-SH-e-9710 15
Step 1:
the 100 % weight by using the percent mode procedure above.
First, memorise sample A to be
Step 2:
by pressing the U and P keys.
Step 3:
the 100 % weight is reached.
Reset sample A at 60 % weight
Add sample B to sample A until
434-SH-e-971016
12 Adjusting Introduction
Adjusting of KERN 434 series precision balances is required at initial installation, any time the balance is moved or bumped, whenever the ambient temperature changes by more than 3° C, and additionally every 30 days or so. Adjusting is necessary for two main reasons. First, with time and use, mechanical deviations can occur. Secondly, the weight of a mass in one location or under a certain set of conditions will not always be the same at a different location or under a different set of conditions.
There are a number of adjusting procedures that will need to be done during the life of a KERN 434 series balance. They are briefly explained below. The actual adjusting procedures are contained throughout this manual (see Table of Contents page 2).
Linearity Adjusting: The purpose of Linearity Adjusting is to create a straight line from Zero to the Full Capacity Scale Value to that the balance will display accurate weighing results at all values in between. On the graph below, the Y-axis represents the "true weight" of a mass on the weighing pan and the X-axis represents the weight displayed by the balance. An accurate mass of any weight value will be plotted on a linear (straight line) path from zero to the maximum capacity. However, although a straight line will be generated by Linear Adjusting, this line must be correctly placed by the balance during Span Adjusting (see Linearity Adjusting page
41).
Span Adjusting: With Span Adjusting, we are shifting the straight line generated during Linearity Adjusting by giving it the Zero Load Value and the full capacity adjusting mass value as reference points (see Span Adjusting page 43).
434-SH-e-9710 17
Temperature Compensation Adjusting: The KERN 434 Mechanical Unit operates by a force coil moving inside a permanent magnet. A change in ambient temperature causes a change in the temperature of the Mechanical Unit, in turn altering the characteristics of the magnet. Unless this is compensated for, it will cause sensitivity drift problems. KERN 434 balances use a transistor temperature sensor in the Mechanical Unit to detect changes in temperature. The temperature compensation settings, which match the temperature characteristics of the Mechanical Unit sensor, are stored in Electronic Erasable Programmable Read Only Memory (EEPROM).
All KERN 434 series balances undergo a Temperature Compensation Adjusting at the factory before shipment. At that time all temperature compensations settings are programmed into EEPROM. Therefore, it is not necessary to perform this adjusting regularly to avoid sensitivity drift problems.
EEPROM: If the EEPROM is lost due to a component replacement or short circuit, all adjusting data is cleared. If this happens, the EEPROM must be reinitialised. Reinitialisations of the EEPROM must always be followed by Linearity and Span Adjustings. Depending the user’s needs, a Temperature Compensation Adjusting may also be necessary (see Initialisation of EEPROM, page 39).
434-SH-e-971018
13 Inner A /D Counter Check Mode
Many of the maintenance procedures contained in this manual require adjustments to the Inner A / D Counter Dip Switch, hereby referred to as Dip Switch, located on the Display P.C.B. (MJ-1A).
The Check Modes needed to perform the repair procedures contained in this manual are found below. When a Dip Switch adjustment is required, simply flip the switches so that they correspond to the appropriate Modes below.
1234
Check 0 Mode
(Normal Weighing)
νννν
Off On
1234
Check 1 Mode
(A/D Count Data) (A/D Counter Range 0000000 - 3884160)
ννν
ν
Off On
434-SH-e-9710 19
1234
Check 2 Mode
(Temperature Compensation Data) (A/D Counter Range 180000 - 220000)
Check 3 Mode
(Linearity Data) (A/D Counter Range 0000000 - 3884160)
νν ν
ν
12
34
νν
νν
Off On
Off On
434-SH-e-971020
14 Disassembly / Assembly
14.1 Removing the Top Case
Step 1:
Step 2:
Step 3:
Step 4:
Step 5:
Unplug the AC adapter from the balance.
Remove the weighing pan and the pan support.
Loosen a forehead screw toward the display keyboard and screws on rear side of the balance.
Gently lift off the top cover, flipping it to the right.
Do not jerk the connecting cable.
Disconnect all cables inside the balance.
434-SH-e-9710 21
14.2 Removing the Mechanical Unit
Step 1:
Step 2:
Step 3:
Step 4:
Loosen 3 screws on the bottom of the balance.
Gently lift the Mechanical Unit box from the bottom case.
Slowly place this Mechanical Unit box a flat surface and loosen 4 screws on top of this box.
Open this box, then the Mechanical Unit will be exposed.
434-SH-e-971022
14.3 Mechanical Unit Disassembly
Step 1:
Jig Set (the beam Jig and the beam­suspension Jig) - see the Jig Set page
21.
Step 2:
plate bearing assembly and remove it.
Step 3:
plate bearing assembly and remove it.
Step 4:
vertical plate springs.
Step 5:
the balance weight and remove it.
Step 6:
coupling link.
Attach the appropriate KERN 434
Loosen 4 screws on the upper
Loosen 4 screws on the lower
Loosen 4 screws for removing
Loosen 2 screws on bottom of
Loosen 4 screws for removing a
Step 7:
beam positioning and remove it.
Step 8:
assembly, the suspension assembly and the mechanical unit base will be disassembled separately.
Step 9:
Step 10:
attached on the beam
Step 11:
from the beam. We recommend that you should wear gloves when handling the Force Coil Bobbin.
Loosen a screw on top of the
Make certain that the beam
Remove the magnet lid.
Unsolder the short wire
Remove the Force Coil Bobbin
434-SH-e-9710 23
Loading...
+ 52 hidden pages