![](/html/98/984c/984c678fd0e7b9ee6f25c17db83cfb450f7bf583150746a3ff01a71697aab0e3/bg1.png)
KS-DO-10 |Pocket Dissolved Oxygen Tester
Kensington Scientific
22a Cavan Rd, Dry Creek
South Australia, 5094
helpme@kensingtonscientific.com.au
www.kensingtonscientific.com.au
OVERVIEW
Below are the steps you need to follow to begin using your Kensington KS-DO-10 pocket dissolved oxygen tester. This
manual will cover how insert the batteries, using the function keys, making measurements, and sensor replacement.
PACKAGE CONTENTS
1. Dissolved Oxygen Probe
2. DO Electrolyte solution
3. AAA batteries
4. DO Membrane Cap
INSERTING THE BATTERIES
1. Your Kensington Pocket KS-DO-10 pocket salinity tester uses two AAA batteries.
2. Remove the button cap found on the bottom of the tester as shown in the above image.
3. Place the battery into its designated slot. Take note of the proper position of the battery to ensure the polarities
placed correctly.
4. Replace the battery cap.
KEYPAD INDEX
1. Serves as the power button of the tester
2. Initiates Data Hold. Press once to hold the current reading, press again to resume making
measurements.
3. Exit button during calibration
1. Press once to conduct calibration of the instrument
2. Press and hold to enter setup menu
3. In calibration mode, press to set calibration values.
![](/html/98/984c/984c678fd0e7b9ee6f25c17db83cfb450f7bf583150746a3ff01a71697aab0e3/bg2.png)
KS-DO-10 |Pocket Dissolved Oxygen Tester
Kensington Scientific
22a Cavan Rd, Dry Creek
South Australia, 5094
helpme@kensingtonscientific.com.au
www.kensingtonscientific.com.au
4. In setup mode, press to select default option
1. Press to confirm calibration or chosen option
2. Toggles between conductivity, TDS and salinity measurement modes.
DISPLAY
The KS-SAL-20 is built with a clear LCD display that shows the measured value, various mode indicators and units.
Below is a summary to describe the meaning behind each notification on the display.
Indicates tester is in the measurement mode
Indicates tester is in calibration mode
Indicates tester is in SETUP mode
Indicates the HOLD function is active and the current displayed value is on hold.
Indicates temperature compensation is enabled
REFILLING ELECTROLYTE SOLUTION FOR DO PROBE
1. Remove the packaging of the DO probe. Unscrew its cap from the probe as shown from the image below:
2. Fill the membrane cap halfway with the electrolyte solution
3. Screw back the membrane cap on the probe.
4. Make sure the cathode of the probe makes contact with the membrane cap. Make sure there are no bubbles found
on the electrolyte solution.
![](/html/98/984c/984c678fd0e7b9ee6f25c17db83cfb450f7bf583150746a3ff01a71697aab0e3/bg3.png)
KS-DO-10 |Pocket Dissolved Oxygen Tester
Kensington Scientific
22a Cavan Rd, Dry Creek
South Australia, 5094
helpme@kensingtonscientific.com.au
www.kensingtonscientific.com.au
SETUP MENU
The KS-SAL-20 contains integrated setup menu that allows you to customize your tester’s features to fit your
measurement requirements.
Setting range: 0 to 35 ppt
Atmospheric Pressure
Coefficient
Setting range: 450 to 850
mmHg
Calibration Points
1 (Default)
Temperature calibration
mode
Automatically freezes
stable readings
Automatically turns off
tester after 8 seconds of
idle
Restores factory settings
SETTING OF DEFAULT PARAMETERS
1. Press the CAL button to enter setup menu. The first display should be the current salinity coefficient.
2. Press CAL again to set the desired value. Press ENTER to confirm.
3. The tester should then display the atmospheric pressure setting. The display should show ‘760mmHg’.