Kensington KS-DO-10 User Manual

User Manual
KS-DO-10 |Pocket Dissolved Oxygen Tester
OVERVIEW
Below are the steps you need to follow to begin using your Kensington KS-DO-10 pocket dissolved oxygen tester. This manual will cover how insert the batteries, using the function keys, making measurements, and sensor replacement.
PACKAGE CONTENTS
1. Dissolved Oxygen Probe
2. DO Electrolyte solution
3. AAA batteries
4. DO Membrane Cap
INSERTING THE BATTERIES
1. Your Kensington Pocket KS-DO-10 pocket salinity tester uses two AAA batteries.
2. Remove the button cap found on the bottom of the tester as shown in the above image.
3. Place the battery into its designated slot. Take note of the proper position of the battery to ensure the polarities placed correctly.
4. Replace the battery cap.
KEYPAD INDEX
Key
Function
1. Serves as the power button of the tester
2. Initiates Data Hold. Press once to hold the current reading, press again to resume making measurements.
3. Exit button during calibration
1. Press once to conduct calibration of the instrument
2. Press and hold to enter setup menu
3. In calibration mode, press to set calibration values.
User Manual
KS-DO-10 |Pocket Dissolved Oxygen Tester
4. In setup mode, press to select default option
1. Press to confirm calibration or chosen option
2. Toggles between conductivity, TDS and salinity measurement modes.
DISPLAY
The KS-SAL-20 is built with a clear LCD display that shows the measured value, various mode indicators and units. Below is a summary to describe the meaning behind each notification on the display.
Indicator
Description
MEAS
Indicates tester is in the measurement mode
CAL
Indicates tester is in calibration mode
SETUP
Indicates tester is in SETUP mode
HOLD
Indicates the HOLD function is active and the current displayed value is on hold.
ATC
Indicates temperature compensation is enabled
REFILLING ELECTROLYTE SOLUTION FOR DO PROBE
1. Remove the packaging of the DO probe. Unscrew its cap from the probe as shown from the image below:
2. Fill the membrane cap halfway with the electrolyte solution
3. Screw back the membrane cap on the probe.
4. Make sure the cathode of the probe makes contact with the membrane cap. Make sure there are no bubbles found on the electrolyte solution.
User Manual
KS-DO-10 |Pocket Dissolved Oxygen Tester
SETUP MENU
The KS-SAL-20 contains integrated setup menu that allows you to customize your tester’s features to fit your measurement requirements.
Parameter
Description
Options
Description
SALT
Salinity Coefficient
0.0 (Default)
Setting range: 0 to 35 ppt
PRES
Atmospheric Pressure Coefficient
760 (Default)
Setting range: 450 to 850 mmHg
CAL
Calibration Points 1 (Default)
1 point calibration
2
2 points calibration
UNIT
Measurement Unit
mg/L (Default)
Concentration unit
ppm
Concentration unit
%
% saturation
°C (Default)
Celsius
°F
Fahrenheit
°C
Temperature calibration
CAL
Temperature calibration mode
HOLD
Auto-Hold YES
Automatically freezes stable readings
NO (Default)
Disable
OFF
Auto-off YES (Default)
Automatically turns off tester after 8 seconds of idle
NO
Disable
RST
Reset
YES
Restores factory settings
NO (Default)
Disable
SETTING OF DEFAULT PARAMETERS
1. Press the CAL button to enter setup menu. The first display should be the current salinity coefficient.
2. Press CAL again to set the desired value. Press ENTER to confirm.
3. The tester should then display the atmospheric pressure setting. The display should show ‘760mmHg’.