7076 Two-Pole Matrix Card
Dual 4×12
The Model 7076 is a general purpose matrix switching card that consists of two independent 4×12 switch- ing matrices. Each matrix has two switched circuits (HI and GUARD). The four row signal paths are con­nected through jumpers to the general purpose analog backplane in the Model 707A switching main­frame. This provides the interconnect between cards for column expansion (4×24, 4×36, etc.). The jumpers can be removed to isolate any matrix. The columns of the two matrices on each card can be tied together with on-card jumpers for expansion to an 8×12 matrix. Connections to the matrix are through standard 25-pin D connectors for mass termination. There are two column connectors, one for each bank, and one row connector.
MATRIX CONFIGURATION: Dual 4 rows by 12 columns. Also
configurable as 8 rows by 12 columns. Jumpers can be
Low cost
<5µV voltage offset
<100pA offset current
15MHz bandwidth
110V, 1A signal levels
Uses standard 25-pin
D connectors
Ordering Information
7076 Dual 4x12 Two-Pole
Extended warranty, service, and calibration contracts are available
Use with Model 707A and 708A switching matrix mainframes
Accessories Supplied
Matrix Card
.
removed to isolate any row from the backplane.
CROSSPOINT CONFIGURATION: 2-pole Form A (HI, GUARD).
CONNECTOR TYPE: 25-pin subminiature D connector, two for
column connection, one for row connection.
MAXIMUM SIGNAL LEVEL:
DC Signals: 110V DC, 1A switched, 30VA (resistive load).
AC Signals: 175V AC peak, 1A switched, 62.5VA (resistive
load).
COMMON MODE VOLTAGE: 110V DC, 175V AC peak pin-to-pin
or pin-to-chassis.
CONTACT LIFE: Cold Switching: 10
At Maximum Signal Level: 10
PATH RESISTANCE (per conductor): <0.50initial, <1.5at
end of contact life.
CONTACT POTENTIAL: <5µV per crosspoint (HI to GUARD).
OFFSET CURRENT: <100pA.
CROSSTALK (1MHz, 50Ω load) <–50dB.
INSERTION LOSS (1MHz, 50source, 50load): 0.05dB
typical.
8
5
closures.
closures.
Jumpers for multiplexer expansion.
123456789101112
A
ISOLATION:
Path: >10 Differential: >10 Common Mode: >10
3dB BANDWIDTH (50Ω load): 15MHz typical.
RELAY DRIVE CURRENT (per crosspoint): 28mA.
RELAY SETTLING TIME: <3ms.
EMC: Conforms to European Union Directive 89/336/EEC.
SAFETY: Conforms to European Union Directive 73/23/EEC
(meets EN61010-1/IEC 1010).
ENVIRONMENT:
Operating: 0° to 50°C, up to 35°C at 70% R.H. Storage: –25° to 65°C.
7076-RMTC High Isolation Row Cable Assembly, 3m (10 ft) 7076-CMTC High Isolation Column/Bank Cable Assembly,
7075-MTC Row/Column/Bank Standard Cable Assembly,
COLUMNS
(To rear panel connector)
10
, <5pF.
9
, 120pF nominal.
9
, 200pF nominal.
ACCESSORIES AVAILABLE
3m (10 ft)
3m (10 ft)
1.888.KEITHLEY (U.S. only)
SWITCHING & CONTROL
www.keithley.com
ROWS
(To rear
panel
connectors)
B
C
D
E
F
G
H
123456789101112
A GREATER MEASURE OF CONFIDENCE
COLUMNS
(To rear panel connectors)
Column
Jumpers
(User
installable;
to configure
as 8 x 12)
Backplane
Jumpers
(factory
installed)