7059 Low Voltage Scanner Card
10-Channel, 2-Pole
For low-level transducer switching, the Model 7059
is designed to reduce errors due to thermocouple
effects in the contacts (<1µV maximum) and
ensure that high isolation (10
between input signals. It will switch any one of ten
signals to one output or switch one signal to any
one of ten outputs. Switch settling time is less than
2 milliseconds.
• <1µV contact potential
• 200V switching
• Screw terminal connections
Ordering Information
7059 Low Voltage Scanner
Use with 7001 and 7002 Switch Mainframes
Card
7061 Universal Adapter Card
12
Ω) is maintained
HI
Ch. 1
LO
HI
Ch. 2
LO
Ch. 3 – 9
HI
Ch. 10 Output
LO
HI
LO
CHANNELS PER CARD: 10.
CONTACT CONFIGURATION: 2-pole Form A, common guard
connection.
CONNECTOR TYPE: Screw terminal, #18AWG maximum wire
size.
RELAY DRIVE CURRENT: 13mA per relay typical.
MAXIMUM SIGNAL LEVEL: 200V, 100mA, or 2VA (resistive load
only).
CONTACT LIFE: >10
maximum signal levels.
CONTACT RESISTANCE: <2Ω to rated life.
CONTACT POTENTIAL: <1µV input to output with copper leads
(<200nV typical within 1 minute of actuation).
WARM-UP: 1 hour in mainframe for thermal stability.
ACTUATION TIME: <2ms, exclusive of mainframe.
CHANNEL ISOLATION: >10
INPUT ISOLATION, DIFFERENTIAL: >10
INPUT ISOLATION, COMMON MODE: >10
COMMON MODE VOLTAGE: <200V peak.
OPERATING ENVIRONMENT: 0° to 50°C, up to 35°C at 70% RH.
STORAGE ENVIRONMENT: –25°C to 65°C.
8
closures cold switching; >106closures at
12
Ω, <10pF.
9
Ω, <50pF.
9
Ω, <150pF.
• Circuit prototyping
• Access to relay drive lines
• Extender cable
Ordering Information
7061 Universal Adapter Card
1.888.KEITHLEY (U.S. only)
SWITCHING & CONTROL
www.keithley.com
The Model 7061 was designed with a 6-volt supply,
ground trace, and a grid pattern of plated-through
holes on 0.1 inch centers for mounting special
relays or circuits. Relay drive lines from the scanner mainframe can be used for control. A supplied
ribbon cable assembly extends the relay drive lines
beyond the enclosure for easy access to relay cards
for troubleshooting.
RELAY DRIVE LINES: 350mA sink.
HARDWARE SPACE: 15/16 inch, maximum height a component
can protrude from the universal adapter card and fit in the
7001 or 7002 mainframe.
HOLE SIZE: 0.062 inch diameter.
HOLE SPACING: 0.1 inch centers
RIBBON CABLE: 2 ft length with connectors preassembled.
STRAIN RELIEF CLAMP: Assembly provided.
OPERATING CODES: Provided in manual.
A GREATER MEASURE OF CONFIDENCE