Keithley 7054 Service manual

Model 7054 High Voltage
Scanner Card
Instruction Manual
Contains Operating and Servicing Information
WARRANTY
Keithky Instruments, Inc. warrants this product to be free from defects in material and workmanship for a period of 3 years fmm date of shipment.
Keithky Inshuments, Inc. warrants the following items for 90 days from the date of shipment: probes, cables, rechargeable batter­ies, diskettes, and documentation.
During the warranty period, we will, at our option, either repair or replace any product that proves to be defective.
To exercise this warranty, write or call your local Keithley representative. or contact Keitbley headquarters in Cleveland, Ohio. You will be given prompt assistance and return instructions. Send the product transportation prepaid, to the indicated service facility. Repairs will be made and the pmduct retomed, transportation prepaid. Repaired or replaced products are warranted for the balance of the orighd warranty period, or at least 90 days.
LIMITATION OF WARRANTY
This warranty does not apply to defects resulting from product modification without Keithley’s express written consent, or misuse of any product or part. Tbis warranty also does not apply to fuses, software. non-rechargeable batteries, damage from battery leak­age., or problems arising from normal wear or failurc to follow instructions.
THISWARRANTYISINLJELJOFALLOTHEXW-, WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR USE THE REMEDIES PROVIDED HEREIN ARE BUYER’S SOLE AND EXCLUSIVE REMEDIES.
m KEZTHLEY INSTRUMENTS, INC. NOR ANY OF ITS EMPLOYEES SHALL BE LIABLE FOR ANY DIRECT, INDIRECT, SPECIAL, INCIDENTAL OR CONS!ZQJENTlAL DAMAGES ARISING OUT OF THE USE OF ITS INSTRU­MENTS AND SOFTWARE EVBN IF POSSIBILITY OF SUCH DAMAGES. SUCH EXCLUDED DAMAGES SHALL INCLUDE, BUT ARE NOT LIMITED To: COSTS OF REMOVAL AND INSTALLATION, LOSSES SUSTAINEZI AS THE RESULT OF INJURY TO ANY PERSON, OR DAMAGE TO PROPERTY.
KFJTHLBY INSTRUMENTS, INC., HAS BEEN ADVISED IN ADVANCE OF THE
EXPRJZXED OR IMPLIED, INCLUDING ANY IMPLIED
Model 7054 High Voltage Scannr Card
Instruction Manual
Q 1987, Keithley Instruments, Inc.
Test Instrumentation Group
All rights reserved.
Cleveland, Ohio, U.S.A.
February 1987, Third Printing
Document Number: 7054-901-01 Rec. C

SAFETY PRECAUTIONS

The following safety precautions should be observed before using this product and any associated instrumentation. Although some instruments and accessories would normally be used with non-hazardous voltages, there are situations where hazardous conditions may be present.
This product is intended for use by qualified personnel who recognize shock hazards and are familiar with the safety precautions required to avoid possible injury. Read the operating information carefully before us­ing the product.
Exercise extreme caution when a shock hazard is present. Lethal voltage may be present on cable connector jacks or test fixtures. The American National Standards Institute (ANSI) states that a shock hazard exists when voltage levels greater than 30V RMS, 42.4V peak, or 60VDC are present. A good safety practice is to expect that hazardous voltage is present in any unknown circuit before measuring.
Before operating an instrument, make sure the line cord is connected to a properly grounded power receptacle. Inspect the connecting cables, test leads, and jumpers for possible wear, cracks, or breaks before each use.
For maximum safety, do not touch the product, test cables, or any other instruments while power is applied to the circuit under test. ALWAYS re­move power from the entire test system and discharge any capacitors be­fore: connecting or disconnecting cables or jumpers, installing or removing switching cards, or making internal changes, such as installing or removing jumpers.
Do not touch any object that could provide a current path to the common side of the circuit under test or power line (earth) ground. Always make measurements with dry hands while standing on a dry, insulated surface capable of withstanding the voltage being measured.
Do not exceed the maximum signal levels of the instruments and acces­sories, as defined in the specifications and operating information, and as shown on the instrument or test fixture rear panel, or switching card.
Do not connect switching cards directly to unlimited power circuits. They are intended to be used with impedance limited sources. NEVER connect switching cards directly to AC main. When connecting sources to switch­ing cards, install protective devices to limit fault current and voltage to the card.
When fuses are used in a product, replace with same type and rating for
continued protection against fire hazard.
Chassis connections must only be used as shield connections for measur-
ing circuits, NOT as safety earth ground connections.
If you are using a test fixture, keep the lid closed while power is applied
to the device under test. Safe operation requires the use of a Bd interlock.
If a @ screw is present on the test fixture, connect it to safety earth
ground using #18 AWG or Iarger wire.
The t symbol on an instrument or accessory indicates that IOOOV or
more may be present on the terminals. Refer to the product manual for de-
tailed operating information.
Instrumentation and accessories should not be connected to humans,
Maintenance should be performed by qualified service personnel. Before
performing any maintenance, disconnect the line cord and all test cables.

SPECIFICATIONS

MODEL 7054 HIGH VOLTAGE SCANNER CARD
CHANNELS PER CARD: 10. CONTACT CONFIGURATION: l-pole Form A, common guard connection.
CONNECTOR TYPE: Quick disconnect screw terminal, No. 18 AWG maximum
wire size.
RELAY DRIVE CURRENT: 50mA (per relay). MAXIMUM SIGNAL LEVEL: lOOOV, 0.5A, 1OVA peak (resistive load). CONTACT LIFE: 10” closures (cold switching); 10 6 closures (at maximum signal
level).
CONTACT RESISTANCE: c 1l-l. ACTUATION TIME: <2Oms, exclusive of mainframe. CHANNEL ISOLATION: > 10’9 and c 10pF. INPUT ISOLATION: > 109R and < 150pF. GUARD TO CHASSIS VOLTAGE: 200V maximum. ENVIRONMENT:
Operating: 0” to SOT, up to 35°C at 70% non-condensing relative humidity. Storage: -25” to t7O”C.
DIMENSIONS, WEIGHT: 32mm high x l&run wide x 272mm long (l*/(” x 44’2”
x 10%“). Net weight 0.28kg (10 oz.).
Specifications subject to change without notice.

TABLE OF CONTENTS

SEcmON i-
1.1
1.2
1.3
1.4
1.5
1.6
SECTION 2 - OPERATION
2.1
2.2
2.3
2.4
2.5
2.6
2.7
SECTION 3 - SERVICING INFORMATION
3.1
3.2
3.3
3.4
3.5
3.5.1
3.5.2
Introduction. Warranty Information Manual Addenda Safety Symbols and Terms Unpacking and Inspection. Specifications
Zntroduction..............................~
Safety Precautions Wiring.
Installation and Removal Operating Considerations
Applications.
Switch Terminology
Introduction
Handling and Cleaning
Recommended Test Equipment.
Contact Resistance Test
Isolation Tests
GENERAL INFORMATION
.........................................
..................................... 2
........................................
....................................
..............................................
........................................
..................................
.........................................
.......................................
Channel isolation Input Isolation,
..................................
.................................... 19
.................................
............................
............................ 2
..............................
.............................
..............................
....................... 16
............................... 16
...........
1 1
2
3
5 5 5 6
6 11 13
l5 15
‘18 18
SECTION 4 -
Introduction
ii
4.3
4.4
4.5
Replaceable Parts.. Ordering Information Factory Service Component Layout and Schematic Diagram.
REPLACEABLE PARTS
.........................................
..................................
................................
......................................
...........
21
22
22
22
22

LIST OF TABLES

SECTION 3 -
3-1
SECTION 4 -
4-1
Recommended Test Equipment. . . . . . . . . . . . . . . . . . . . . . . . 16
Model 7054 Replaceable Parts . , . . . . . . . . , . . . . . . . . . . . . . . 21
SERVICING INFORMATION
REPLACEABLEPARTS

LIST OF ILLUSTRATIONS

SECTION 2 - OPERATION
2-1 2-2 2-3 2-4
SECTION 3 -
3-l
3-2 3-3
Limiting Ca Limiting In Measuring 3-Phase Voltages Switch Terminology
Contact Resistance
Channel Isolation Input Isolation
acitive
cp
uctive Reaction Voltage.
SERVICING INFORMATION
.................................... 19
......................................
Reaction Current
..........................
..................................
...................................
...................
...................
9 10 12 13
17
20
SECTION 4 -
4-1
4-2
ii
Model 7054 Component Layout . . . . . . . . . . . . . . . . . . . . . . . 23
Model 7054 Schematic Diagram . . . . . . . . . , . . . . . . . . . . . . . 25
REPLACEABLE PARTS
SECTION 1
GENERAL INFORMATION

1.1 INTRODUCTION

The Model 7054 is a high voltage scanner card that switches signal levels
up to lOOOV, 0.5A or 1OVA peak (lOmA at 1OOOV) for resistive loads only. Each of the 10 channels on the card conSists of a single-pole, Form A (normally open) relay with a common guard connection and a fused input. The output connection and the input connections have quick disconnect screw terminals with a maximum allowable wire size of No.
18 AWG.
Single-pole switching is done in less than 20msec exclusive of the main-
frame, and the specified contact life is 10B closures with cold switching and lo6 closures at maximum signal levels. Contact resistance is less than 10. The relay drive current is 50mAlrelay.
The Model 7054 is field installable in an appropriate Keithley scanner mainframe (e.g. Model 705 or Model 706). With the isolation between each channel being greater than 1O”Q and less than lOpF, each chan­nel is well insulated from cross channel noise and interference.

1.2 WARRANTY INFORMATION

Warranty information is stated on the inside front cover of this manual. If there is a need for service, contact the Keithley representative or authorized repair facility in your area. Check the back cover for ad­dresses. The service form supplied at the end of the manual should be used to provide the service facility with information concerning any difficulty.

1.3 MANUAL ADDENDA

Product improvements or changes to this manual will be explained on an addendum included with the manual. It is recommended that this information be incorporated immediately into the appropriate places in the manual.
If an additional instruction manual is required, order the manual package (Keithley Part Number 7054-901-00). The manual package in­cludes an instruction manual and all pertinent addenda.

1.4 SAFETY SYMBOLS AND TERMS

The symbol ! operating instructions.
The symbol terminals.
The WARNING used in this manual explains dangers that could result in personal injury or death.
The CAUTION used in this manual explains hazards that could damage the instrument.
A on the card denotes that the user should refer to the
denotes that a high voltage may be present on the
AIM

1.5 UNPACKING AND INSPECTION

The Model 7054 was inspected both electrically and mechanically before shipment. Upon receiving the Model 7054, unpack all items from the shipping carton and check for any obvious damage that may have oc­curred during transit. Report any damage to the shipping agent. Re­tain and use the original packaging materials in case reshipment is necessary. The following items are shipped with every Model 7054:
Model 7054 High Voltage Scanner Card Model 7054 Instruction Manual

1.6 SPECIFICATIONS

Detailed specifications of the Model 7054 precede the Table of Contents of this manual.
314
SECTION 2
OPERATION

2.1 INTRODUCTION

This section provides information needed to use the Model 7054 High
Voltage Card with the Model 705 and 706 scanner mainframes. Once
the card is installed in the mainframe, refer to the mainframe instruc-
tion manual for complete operating instructions.

2.2 SAFETY PRECAUTIONS

WARNING User supplied lethal voltages may be present on the PC board or the connections. Turn off all power and discharge stored energy in external circuitry before mak­ing or breaking connections.
1. Do not exceed the Model 7054 maximum voltages of ltlOOOV peak terminal to terminal and &l35OV peak terminal to chassis.
2. Make sure the scanner mainframe is grounded through an earth grounded receptacle before operation.
3. Ins
4. Use appropriately rated cables when switching high voltages.
ect alI test lead connections for wear and defects such as cracks
an f
exposed wires.

2.3 WIRING

Each channel on the Model 7054 card consists of a single-pole, Form A (normally open) switching relay. The Model 7054 will switch any one of the 10 signals (inputs) to one output, or switch one signal to any one of 10 outputs.
NOTE
Because of the high impedance of the board, take special care when handling and using to prevent degradation of performance. Handle the board by the edges to avoid con­taminating it with dirt, body oil, etc. For cleaning instruc­tions, see paragraph 3.2.
1. Wiring is accomplished by means of terminal strips as shown on the component layout. See Figure 4-1. Each channel has a HI connec­tion and a LO (GUARD) connection. Lb (Guard) is common to all channels.
2. Resistance of the relay contacts path is less than 10.
3. A common guard surrounds all analog signal paths.
4. Use wires or cables that are rated for greater than 1OOOV. The max­imum allowable wire size is No. 18 AWG. The minimum wire size is No. 26 AWG.
5. Route the wires through the rubber clamp pads located at the rear of the card.

2.4 INSTALLATION AND REMOVAL

Once the card is wired, insert it card edge first into the scanner main­frame by aligning it with the grooves in the appropriate slot. Make sure it is properly seated into the mainframe connector. Push the locking tabs forward to the center of the card to lock it in.
To remove a card, first turn off the mainframe and all other equipment connected to the card. Unfasten the locking tabs on the card by pull­ing the tabs outward. Grasp the end of the card and carefully pull it out of the mainframe.

2.5 OPERATING CONSIDERATIONS

Since the Model 7054 is a lo-channel card, set the scanner to the 2-pole mode when using the Model 7054 by itself or when intermixing with
other lo-channel cards (such as Models 7056, 7058, 7059, and 7066). In the 2-pole scanner mode, each scanner channel controls one channel on one lo-channel card.
Signal Levels-The maximum signal levels for the Model 7054 High Voltage Card are 1OOOV maximum, 0.5A or 1OVA l eak (1OmA at 1OOOV) for resistive loads only. The specified contact 1’ e is lO* closures with cold switching and lo6 closures at maximum signal levels.
WARNING When switching signals greater than 30V RMS or 42.2V peak take care to prevent contact with live circuits which could cause electrical shock resulting in injury or death.
Cables-Shielded cables should be used with the Model 7054 card when switching above 50V. The shield should be connected to the circuit LO
(GUARD). This prevents excessive radiation from the cables from in-
terfering with any equipment. The cables must be rated for greater than
1ooOV.
Fusible Links-The Model 7054 scanner card has a fusible link in series with the high input side of each channel relay. A link opens when the current level on that channel becomes excessive because of an external circuit misapplication or a relay failure. The links protect the PC board
from damage that is difficult to repair (e.g. lifted or burned traces on the board).
The links are covered with Plexiglas@ shields to prevent mechanical damage and still provide a visual check. A repair procedure follows:
1. Unscrew the Plexiglas@ shield. Clean out the PC board feedthru holes for the fusible link with a desoldering tool.
2. Insert a jumper of No. 38 AWG magnet wire into the holes and solder in place.
CAUTION When replacin the fusible links, use only No. 38 AWG magnet wire. 18 se of any other wire may cause severe damage to the PC board.
3. Check the same channel’s relay contacts for a short or insufficient input isolation resistance and replace the relay if necessary.
4. Clean the PC board and Plexiglas
@ shield according to the instruc-
tions given in paragraph 3.2, then remount the shield.
Depending on your application, a current limiting resistor can be used in place of a fusible link. Each resistor will limit the current flow through one channel. The resistors must be of sufficient power rating to safely
dissipate any overload.
Reactive Loads-The Model 7054 is specified for resistive loads. Since reactive loads can cause excessive currents and voltages, current surge limiting (for capacitive loads) and voltage clamping (for inductive loads) are required to prevent damaging the relays and external circuitry.
1. Ca acitive Loads - The surge current from a ca acitive load must be ess than 0.5A to protect the relays. Figure 2- P
P
shows typical cir­cuits to limit current surges. Also, consider the maximum load of 1OVA when determining the current limit. For example, when switch­ing 5OOV, the current must be limited to 2OmA (I = lOVA/5OOV) and the limiting resistor calculation of Figure 2-lA would be:
R = VII = 5OWl2OmA = 25k!J
2. Inductive Loads - Inductive reaction voltage, L(di/dt), must be less than 1OOUV. Typical clampin circuits are shown in Figure 2-2. Also,
consider the maximum loa B of 1OVA when determining the voltage limit. For example, when switching 4OmA, the voltage must be limited to 250V (V = lOVA14OmA) and the clamping resistor calcula­tion of Figure 2-2A would be:
R = VII = 25OVl4OmA = 625On
A. RESISTOR LIMITED
IN
0
L
l HIGH RESISTANCE COLD.
LOW RESISTANCE HOT. FAST THERMAL RECOVERY.
B. THERMISTOR LIMITED
Figure 2-l. Limiting Capacitive Reaction Current
LOAD
9
LOAD
A. RESISTOR CLAMPED (AC OR DC VOLTAGES)
B. DIODE CLAMPED (DC VOLTAGES)
C. ZENER CLAMPED (AC VOLTAGES)
LOAD
D. RESISTOR-CAPACITOR CLAMPED (AC VOLTAGES)
Figure 2-2. Limiting Inductance Reaction Voltage
10

2.6 APPLICATIONS

The Model 7054 can monitor high voltages from ten different circuits or switch a high voltage source to ten separate circuits.
CAUTION Because of the high voltage applications of the Model 7054, only one channel relay should be energized at a
time. If the break before make operation of the scanner channels is hindered by the application, the card could be damaged.
As an example of measuring high voltages, consider the sensing applica­tion shown in Figure 2-3. Three-phase line or load voltages are switched by the Model 7054 for monitoring. Since LO is common to all channels
and the output of the card, only one of the channel LOS is connected.
To prevent card damage, voltage clamping is necessary when monitor-
ing line voltages. Current limiting or voltage cIamping are necessary
when measuring capacitive or inductive loads. Put the protection cir-
cuitry between the source of the transient and the card.
An example of switching a high voltage source is given in Application Note #700, “Switching in Multipoint Testing”. A Model 7054 and a divider network are used to incrementally increase the output of a Model 247 High Voltage Supply.
This application is suitable for tests such as insulation resistance of cables, harnesses, and PC boards, high-pot testing, and breakdown testing on semiconductors.
11
MODEL 7054 IN SCANNER MAINFRAME
CHANNEL
HI LO HI LO HI LO HI LO
L
CHANNEL
CHANNEL 0uTPLrr
Figure 2-3. Measuring 3-Phase Voltages

2.7 SWITCH TERMINOLOGY

Terms used to describe switch configurations (see Figure 2-4) are defined as follows:
1. Form A is a single-pole, normally open (SPNO) switch. A Zpole, normally open switch is called 2 Form A.
2. Form B is similar to krm A except that its contacts are normally closed (SPNC). A 2-pole, normally closed switch is called 2 Form B.
3. Form C is a sin le-
figuration is c
ole, double-throw switch (SPILT). A 2-pole con­2 Form C.
z!iB e
FORM A (SPNO)
FORM B (SPNC) -
SINGLE-POLE. NORMALLY CLOSED
FORM C (SPDT)
--
SINGLE-POLE, NORMALLY OPEN
0
SINGLE-POLE, DOUBLE-THROW
Figure 2-4. Switch Terminology
SECTION 3
SERVICING INFORMATION

3.1 INTRODUCTION

This section describes tests for verifying the performance of the Model
7054. The tests include:
l Functionality test
or shorted relays or bad connections.
l Specification tests - Two isolation tests measuring leakage currents
to check for a contaminated board and possibly bad relays.
Perform these tests in an environment within the specified operating temperature range (0” to 35°C up to 70% RH) of the scanner card.
Do not perform the procedures in this section unless you are a qualified service person. Some of the pro­cedures may expose you to potentially lethal voltages (> 30V RMS) that could result In personal Injury or death if normal safety precautions are not observed.
Recommended maintenance includes inspection of the scanner card
and the card edge connector to ensure good electrical contact. The
Model 7054 does not require calibration.
- Measuring contact resistance to check for open
WARNING

3.2 HANDLING AND CLEANING

Because of the high impedance of the board, take special care when handling and using to prevent degradation of performance. Handle the board by the edges to avoid contaminating it with dirt, body oil, etc.
To clean the board, spray on an uncontaminated solvent, such as Freon@ TMS or TE and clean with cotton swabs or a soft brush. After the solvent has been applied and is stiIl liquid, blow-dry the board with dry nitrogen gas.

3.3 RECOMMENDED TEST EQUIPMENT

Table 3-l lists recommended test equipment for performance verifica­tion. Other test equipment may be substituted if specifications equal or exceed those stated.
NOTE
Since the Model 617 electrometer has an internal 1OOV source, it can be used in the V/I mode instead of the Model 614 electrometer and Model 230 voltage source.
Table 3-1. Recommended Test Equipment
Description Specification Mfr.
Scanner Mainframe ­Extender Card ­DMM Voltage Source l.OOVDC Electrometer lo-‘OA
<IQ sensitivity
Keithley 705 or 706 Keithley Keithley Keithley 230 Keithley 614 or 617
Model
7061 196

3.4 CONTACT RESISTANCE TEST

This test measures the resistance of each relay contact in an opened and closed state.
1. Set up the equipment as shown in Figure 3-l. Short the HI terminal to the LO terminal of each channel. Attach two copper wires to the output HI and Lo terminals.
2. Zero the Model 196 with the Kelvin test leads shorted together.
3. Set the channel mainframe to the channel mode, channel 1, and the step mode. Close channel 1 from the scanner front panel. The reading
on the Model 196 should indicate a short circuit (less than 1Q).
4. Open channel 1 from the scanner front panel. The reading on the Model 196 should indicate an open circuit.
5. Repeat steps 3 and 4 for the remaining channels on the card.
r
MODEL 7054 IN SCANNER MAINFRAME
CHANNEL
UNDER TEST
t
HI LO
KELVIN TEST LEADS ___)
OuTPur HI LO
Figure 3-1. Contact Resistance

3.5 ISOLATiON TESTS

The two isolation tests measure leakage currents for calculating leakage
resistances on the Model 7054. With the Model 7061 Universal Adpater Card, these tests can be performed faster.
3.5.1 Channel Isolation
This test measures the leakage current between two channels.
1. Set up the equipment as shown in Figure 3-2.
2. Set the scanner to channel 1 and the step mode. Program channel 1 open and the other channels as closed.
3. Set the electrometer to amps and zero check. Program the Model 230 to output 1OOV. Take the electrometer out of zero check.
4. The electrometer should read less than 10~‘OA. (Due to the capacitance of the circuit, the offset current may be high until the circuit capacitance is charged up. Wait until the readings settle out.)
5. Using Ohms law, calculate the channel isolation (the leakage resistance between two channels). For example:
R = E/I = 100VllO-‘“A = XP2n
NOTE
With the Model 617 electrometer in V/I mode, the resistance
is calculated automatically.
6. Put the Model 230 in standby mode and set the electrometer to zero.
7. Repeat steps 2 through 6 for the remaining channels.
la
h4ODEL 7054 IN SCANNER MAINFRAME
WANNEL
UNDER TEST
Figure 3-2. Channel Isolation
3.52 input Isolation
HI
10
OUTPUT
HI
!
This test measures the leakage current between a channel’s HI and Lo
terminals.
1. Set up the equipment as shown in Figure 3-3.
2. Set the scanner to channel 1 and the step mode. Program all chan­nels as open,
3. Set the electrometer to amps and zero check. Program the Model 230 to output 1OOV. Take the electrometer out of zero check.
4. The electrometer should read less than 10-‘A. (Due to the capacitance of the circuit, the offset current may be high until the circuit capacitance is charged up. Wait until the readings settle aut.)
19
5. Using Ohm’s law, calculate the input isolation (leakage resistance) for this channel. For example:
R = E/I = lOOV/lO-‘A = 10gQ.
NOTE
With the Model 677 electrometer in VII mode, the resistance is calculated automatically.
6. Put the Model 230 in standby mode and set the electrometer to zero.
7. Repeat steps 2 through 6 for the remaining channels.
MODEL 7054 IN SCANNER MAINFRAME
CHANNEL
UNDER TEST
HI
lo
OUTPUT
HI
20
Figure 3-3. input Isolation
SECTION 4
REPLACEABLE PARTS

4.1 INTRODUCTION

This section contains replacement parts information, a component layout, and a schematic diagram for the Model 7054.

4.2 REPLACEABLE PARTS

Table 41 Lists parts alphanumerically in the order of their circuit designa-
tions.
Table 4-1. Model 7054 Replaceable Parts
Circuit Desig.
Cl01
JlOOl J1002
J1003
KlOl­KllO
SQlOl­SQllO
Keithley
Description
Capacitor, lOpF, 25V, Aluminum Electrolytic C-314-10
Connector cs-464-1
Connector Connector
Relay, Single-Pole Form A
Fusible Link, No. 38 AWG Magnet Wire
Plexiglas* Shield 7054-302
Upper Clamp Assembly Lower Clamp Assembly 7055-308
Part No.
CS-464-2 cs-464-2
RL-76
-
7055-303-05
To place an order or to obtain information about replacement parts, con­tact your Keith@ representative or the factory. See the back cover for addresses. When ordering, include the following information:
l Model number
l Serial number l Part description l Circuit description (if applicable) l Keithley part number

4.4 FACTORY SERVICE

If fault can be isolated to a particular card, then it is sufficient to return just the card(s). Otherwise, send back both the card(s) and the scan­ner mainframe. For service, photocopy and complete the service form
which follows this section and return it with the equipment.

4.5 COMPONENT LAYOUT AND SCHEMATIC DIAGRAM

Figure 41 shows a component layout of the Model 7054. Figure 42 shows the schematic diagram of the Model 7054.
22
-.-.-
r-------7 L------A
b
% ;z
1
EO
-
-
N
I
(1
I
*
t
n
I
Figure 4-1. Model 7054 Component Layout
D
I
23124
\
5
W-t-F.S*
J
--
.ALL CAPACITOR VALUES AFIE v( WCROFAFWS
I
UNLESS OTHERWISE MARKED (p~-P~~OF~R~DS.).
DENOTES DIGITAL COMMON.
!.v
6
YATES WITH JIOIZ OR JIOU ON MAlNFRUlE OF HCiXLi’C6
Figure 4-2. Model 7054 Schematic Diagram
COMMON
/
25126
Service Form
Model No.
Serial No. i Date Name and Telephone No. Company
List all control settings, de&be problem and dwk boxes tkt apply to problem.
cl Jntermittent
P IElzfaillue 0 Obvious problem on power-up a Fmnt panel operational
Display or output (check one) LIDrift.
0 Unstable D Wtinot read applied input cl Overload
cl calibration only cl Certificate of aliiration required P Datarequired
(attach any additicmal sheets as necessary)
show a block diagram of your me Also, desaibe signal sowx.
cl An&g output fOllOWJ display
0 All ranges or functions are bad
p Unable to zero
ammnent system includiig all instruments mnnected (whether power is hmwd on or not).
0 l’artlcular range or function bad; specify 0 Batteries and fuses are OK
..-_.
0 Checked all cables
_-__ - -.l^_lll. --
where is the measurement b&g perfomxd? (factory, mnbolled laboratory, out-of-dwm, etc.)
what power Line voltage is used?
Any additional information (Jf special modifications have been made by the user, please desaibe.)
Ambient temperahue?
- ~
OF
Keithley Instruments, Inc.
Test Instrumentation Group
28775 Aurora Road Cleveland, Ohio 44139
Printed in the U.S.A.
Loading...