Keithley 7012-S, 7012-C Service manual

I
nstruction Manua
l
s
Models 7012-S and 7012-C
Contains Operating and Servicing Information
7012-901-01 Rev. B / 11-93
WARRANTY
Keithley Instruments, Inc. warrants this product to be free from defects in material and workmanship for a period of 1 year from date of shipment.
Keithley Instruments, Inc. warrants the following items for 90 days from the date of shipment: probes, cables, rechargeable batteries, diskettes, and documentation.
During the warranty period, we will, at our option, either repair or replace any product that proves to be defective.
To exercise this warranty, write or call your local Keithley representative, or contact Keithley headquarters in Cleveland, Ohio. You will be given prompt assistance and return instructions. Send the product, transportation prepaid, to the indicated service facility . Repairs will be made and the product returned, transportation prepaid. Repaired or replaced products are warranted fo r the balance of the original warranty period, or at least 90 days.
LIMITATION OF WARRANTY
This warranty does not apply to defects resulting from product modification without Keithley’s express written consent, or misuse of any product or part. This warranty also does not apply to fuses, software, non-rechargeable batteries, damage from battery leakage, or problems arising from normal wear or failure to follow instructions.
THIS WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR USE. THE REMEDIES PRO­VIDED HEREIN ARE BUYER’S SOLE AND EXCLUSIVE REMEDIES.
NEITHER KEITHLEY INSTRUMENTS, INC. NOR ANY OF ITS EMPLOYEES SHALL BE LIABLE FOR ANY DIRECT , INDIRECT, SPECIAL, INCIDENTAL OR CONSEQUENTIAL DAMAGES ARISING OUT OF THE USE OF ITS INSTRUMENTS AND SOFTWARE EVEN IF KEITHLEY INSTRUMENTS, INC., HAS BEEN ADVISED IN ADVANCE OF THE POSSIBILITY OF SUCH DAMAGES. SUCH EXCLUDED DAMA GES SHALL INCLUDE, BUT ARE NOT LIM­ITED TO: COSTS OF REMOVAL AND INSTALLATION, LOSSES SUSTAINED AS THE RESULT OF INJURY TO ANY PERSON, OR DAMAGE TO PROPERTY.
Keithley Instruments, Inc. • 28775 Aurora Road • Cle v eland, OH 44139 • 440-248-0400 • Fax: 440-248-6168 • http://www.k eithley.com
BELGIUM: Keithley Instruments B.V. Bergensesteenweg 709 • B-1600 Sint-Pieters-Leeuw • 02/363 00 40 • Fax: 02/363 00 64 CHINA: Keithley Instruments China FRANCE: Keithley Instruments Sarl B.P. 60 • 3, allée des Garays • 91122 Palaiseau Cédex • 01 64 53 20 20 • Fax: 01 60 11 77 26 GERMANY: Keithley Instruments GmbH Landsberger Strasse 65 • D-82110 Germering • 089/84 93 07-40 • Fax: 089/84 93 07-34 GREAT BRITAIN: Keithley Instruments Ltd INDIA: Keithley Instruments GmbH Flat 2B, WILOCRISSA • 14, Rest House Crescent • Bangalore 560 001 • 91-80-509-1320/21 • Fax: 91-80-509-1322 ITALY: Keithley Instruments s.r.l. Viale S. Gimignano, 38 • 20146 Milano • 02/48 30 30 08 • Fax: 02/48 30 22 74 NETHERLANDS: Keithley Instruments B.V. Postbus 559 • 4200 AN Gorinchem • 0183-635333 • Fax: 0183-630821 SWITZERLAND: Keithley Instruments SA Kriesbachstrasse 4 • 8600 Dübendorf • 01-821 94 44 • Fax: 01-820 30 81 TAIWAN: Keithley Instruments Taiwan 1 Fl. 85 Po Ai Street • Hsinchu, Taiwan, R.O.C. • 886-3572-9077• Fax: 886-3572-903
Y uan Chen Xin Building, Room 705 • 12 Y umin Road, De wai, Madian • Beijing 100029 • 8610-62022886 • F ax: 8610-62022892
The Minster • 58 Portman Road • Reading, Berkshire RG30 1EA • 0118-9 57 56 66 • Fax: 0118-9 59 64 69
10/99
Models 7012-S and 7012-C Instruction Manual
©1991, Keithley Instruments, Inc.
All Rights Reserved
Cleveland, Ohio, U. S. A.
New Contact Information
Keithley Instruments, Inc.
28775 Aurora Road
Cleveland, OH 44139
Technical Support: 1-888-KEITHLEY
Monday – Friday 8:00 a.m. to 5:00 p.m (EST)
Fax: (440) 248-6168
Visit our website at http://www.keithley.com
Manual Print History
The print history shown below lists the printing dates of all Revisions and Addenda created for this manual. The Revision Level letter increases alphabetically as the manual undergoes subsequent updates. Addenda, which are released between Revisions, contain important change information that the user should incorporate immediately into the manual. Addenda are numbered sequentially. When a new Revision is created, all Addenda associated with the previous Revision of the manual are incorporated into the new Revision of the manual. Each new Revi­sion includes a revised copy of this print history page.
Many product updates and revisions do not require manual changes and, conversely, manual corrections may be done without accompanying product changes. Therefore, it is recommended that you review the Manual Update History.
Revision A (Document Number 7012-901-01) ........................................................................November 1991
Addendum A (Document Number 7012-901-02) ........................................................................January 1992
Revision B (Document Number 7012-901-01).........................................................................November 1993
All Keithley product names are trademarks or registered trademarks of Keithley Instruments, Inc. Other brand and product names are trademarks or registered trademarks of their respective holders

Safety Precautions

The following safety precautions should be observed before using this product and any associated instrumentation. Although some in­struments and accessories would normally be used with non-haz­ardous voltages, there are situations where hazardous conditions may be present.
This product is intended for use by qualified personnel who recog­nize shock hazards and are familiar with the safety precautions re­quired to avoid possible injury. Read the operating information carefully before using the product.
The types of product users are:
Responsible body is the individual or group responsible for the use
and maintenance of equipment, for ensuring that the equipment is operated within its specifications and operating limits, and for en­suring that operators are adequately trained.
Operators use the product for its intended function. They must be
trained in electrical safety procedures and proper use of the instru­ment. They must be protected from electric shock and contact with hazardous live circuits.
Maintenance personnel perform routine procedures on the product
to keep it operating, for example, setting the line voltage or replac­ing consumable materials. Maintenance procedures are described in the manual. The procedures explicitly state if the operator may per­form them. Otherwise, they should be performed only by service personnel.
Service personnel are trained to work on live circuits, and perform
safe installations and repairs of products. Only properly trained ser­vice personnel may perform installation and service procedures.
Users of this product must be protected from electric shock at all times. The responsible body must ensure that users are prevented access and/or insulated from every connection point. In some cases, connections must be exposed to potential human contact. Product users in these circumstances must be trained to protect themselves from the risk of electric shock. If the circuit is capable of operating at or above 1000 volts, no conductive part of the circuit may be
exposed.
As described in the International Electrotechnical Commission (IEC) Standard IEC 664, digital multimeter measuring circuits (e.g., Keithley Models 175A, 199, 2000, 2001, 2002, and 2010) are Installation Category II. All other instruments’ signal terminals are Installation Category I and must not be connected to mains.
Do not connect switching cards directly to unlimited power circuits. They are intended to be used with impedance limited sources. NEV ­ER connect switching cards directly to A C mains. When connecting sources to switching cards, install protective devices to limit fault current and voltage to the card.
Before operating an instrument, make sure the line cord is connect­ed to a properly grounded power receptacle. Inspect the connecting cables, test leads, and jumpers for possible wear, cracks, or breaks before each use.
For maximum safety, do not touch the product, test cables, or any other instruments while power is applied to the circuit under test. ALWAYS remove power from the entire test system and discharge any capacitors before: connecting or disconnecting cables or jump­ers, installing or removing switching cards, or making internal changes, such as installing or removing jumpers.
Exercise extreme caution when a shock hazard is present. Lethal voltage may be present on cable connector jacks or test fixtures. The American National Standards Institute (ANSI) states that a shock hazard exists when voltage levels greater than 30V RMS, 42.4V peak, or 60VDC are present. A good safety practice is to expect
that hazardous voltage is present in any unknown circuit bef ore measuring.
Do not touch any object that could provide a current path to the common side of the circuit under test or power line (earth) ground. Always make measurements with dry hands while standing on a dry, insulated surface capable of withstanding the voltage being measured.
The instrument and accessories must be used in accordance with its specifications and operating instructions or the safety of the equip­ment may be impaired.
Do not exceed the maximum signal levels of the instruments and ac­cessories, as defined in the specifications and operating informa­tion, and as shown on the instrument or test fixture panels, or switching card.
When fuses are used in a product, replace with same type and rating for continued protection against fire hazard.
Chassis connections must only be used as shield connections for measuring circuits, NOT as safety earth ground connections.
If you are using a test fixture, keep the lid closed while power is ap­plied to the device under test. Safe operation requires the use of a lid interlock.
If a screw is present, connect it to safety earth ground using the wire recommended in the user documentation.
!
The symbol on an instrument indicates that the user should re­fer to the operating instructions located in the manual.
The symbol on an instrument shows that it can source or mea­sure 1000 volts or more, including the combined effect of normal and common mode voltages. Use standard safety precautions to avoid personal contact with these voltages.
The WARNING heading in a manual explains dangers that might result in personal injury or death. Alw ays read the associated infor­mation very carefully before performing the indicated procedure.
The CAUTION heading in a manual explains hazards that could damage the instrument. Such damage may invalidate the warranty.
Instrumentation and accessories shall not be connected to humans.
Before performing any maintenance, disconnect the line cord and all test cables.
To maintain protection from electric shock and fire, replacement components in mains circuits, including the power transformer, test leads, and input jacks, must be purchased from Keithley Instru­ments. Standard fuses, with applicable national safety approvals, may be used if the rating and type are the same. Other components that are not safety related may be purchased from other suppliers as long as they are equivalent to the original component. (Note that se­lected parts should be purchased only through Keithley Instruments to maintain accuracy and functionality of the product.) If you are unsure about the applicability of a replacement component, call a Keithley Instruments office for information.
T o clean an instrument, use a damp cloth or mild, water based clean­er. Clean the exterior of the instrument only. Do not apply cleaner directly to the instrument or allow liquids to enter or spill on the in­strument. Products that consist of a circuit board with no case or chassis (e.g., data acquisition board for installation into a computer) should never require cleaning if handled according to instructions. If the board becomes contaminated and operation is affected, the board should be returned to the factory for proper cleaning/servic­ing.
Rev.10/99

7012 SPECIFICATIONS

MODEL 7012-S 4×10 Matrix with Screw Terminals. MODEL 7012-C 4×10 Matrix with Mass Terminated Connector. MATRIX CONFIGURATION: 4 rows by 10 columns. Jumpers can be
removed to isolate any row from the backplane.
CONTACT CONFIGURATION: 2-pole Form A (Hi, Lo). CONNECTOR TYPE:
7012-S: Screw terminal, #16AWG maximum wire size, with .092 inch
O.D. 28 Conductors maximum. #22AWG typical wire size with .062 inch O.D. 88 Conductor maximum.
7012-C: 96-Pin male Euro card connector. Mates to female twisted wire
cable, crimp or solder connector.
MAXIMUM SIGNAL LEVEL:
DC Signals: 110V DC between any two pins, 1A switched. 30VA
(resistive load).
AC Signals: 125V RMS or 175V AC peak, between any two pins,
1A switched, 60VA (resistive load).
COMMON MODE VOLTAGE: 175V peak, any pin to chassis. CONTACT LIFE:
Cold Switching: 108 closures. At Maximum Signal Levels: 105 closures.
7012 4x10 Matrix
7012 4x10 Matrix
CHANNEL RESISTANCE (per conductor): < 1.
CONTACT POTENTIAL:
7012-S: < 500 nV per contact pair (Hi, Lo).
< 1.5 µV per single contact.
7012-C: < 1 µV per contact pair (Hi, Lo).
< 3 µV per single contact.
OFFSET CURRENT: < 100 pA. ACTUATION TIME: 3 ms. ISOLATION:
Path: >10 Differential: >10 Common Mode: >10
9
, < 50 pF.
9
, < 200 pF.
9
, < 400 pF.
CROSS TALK (1MHz, 50 Load): < -40 dB. INSERTION LOSS (50 Source, 50 Load): < 0.1dB below 1MHz, < 3 dB
below 2MHz.
RELAY DRIVE CURRENT (per relay): 16mA. ENVIRONMENT:
Operation:0°C to 50°C, up to 35°C < 80% RH. Storage: -25°C to 65°C.
Specifications subject to change without notice.
12345678910
>>>>>>>>>>>>>>>>>>>
>
1
>
>
2
>
>
3
>
>
4
>
HI
Matrix Crosspoint
LO
Matrix Crosspoint
>
J J
>
Backplane
>
J J
>
Backplane
>
J J
>
Backplane
>
J J
>
Backplane
>

Table of Contents

1.1 Introduction ..................................................................................................................................................... 1-1
1.2 Features............................................................................................................................................................. 1-1
1.3 Warranty information..................................................................................................................................... 1-2
1.4 Manual addenda ............................................................................................................................................. 1-2
1.5 Safety symbols and terms .............................................................................................................................. 1-2
1.6 Specifications ................................................................................................................................................... 1-2
1.7 Unpacking and inspection ............................................................................................................................. 1-2
1.7.1 Inspection for damage............................................................................................................................ 1-2
1.7.2 Shipping contents.................................................................................................................................... 1-2
1.7.3 Instruction manual.................................................................................................................................. 1-3
1.8 Repacking for shipment ................................................................................................................................. 1-3
1.9 Optional accessories........................................................................................................................................ 1-3
2.1 Introduction ..................................................................................................................................................... 2-1
2.2 Basic matrix configuration (4
2.3 Typical matrix switching schemes................................................................................................................ 2-3
2.3.1 Single-ended switching.......................................................................................................................... 2-3
2.3.2 Differential switching............................................................................................................................. 2-4
2.3.3 Sensing...................................................................................................................................................... 2-4
2.3.4 SMU connections..................................................................................................................................... 2-5
2.4 Matrix expansion............................................................................................................................................. 2-5
2.4.1 Two-card switching systems................................................................................................................. 2-5
2.4.2 Mainframe matrix expansion ................................................................................................................ 2-9
3.1 Introduction ..................................................................................................................................................... 3-1
3.2 Handling precautions..................................................................................................................................... 3-1
3.3 Connections...................................................................................................................................................... 3-1
3.3.1 Backplane row jumpers.......................................................................................................................... 3-2
3.3.2 Screw terminal connector card.............................................................................................................. 3-2
3.3.3 Multi-pin (mass termination) connector card..................................................................................... 3-4
3.4 Typical connection schemes .......................................................................................................................... 3-9
3.4.1 Single card system................................................................................................................................... 3-9
3.4.2 Two-card system..................................................................................................................................... 3-9
3.4.3 Two-mainframe system........................................................................................................................ 3-14
3.5 Model 7012 installation and removal......................................................................................................... 3-14
×
10) ............................................................................................................... 2-1
i
4.1 Introduction...................................................................................................................................................... 4-1
4.2 Power limits...................................................................................................................................................... 4-1
4.3 Mainframe control of matrix card................................................................................................................. 4-1
4.3.1 Channel assignments .............................................................................................................................. 4-2
4.3.2 Front panel control.................................................................................................................................. 4-4
4.3.3 IEEE-488 bus operation........................................................................................................................... 4-4
4.4 Matrix switching examples ............................................................................................................................ 4-5
4.4.1 Thick film resistor network testing....................................................................................................... 4-5
4.4.2 Transistor testing................................................................................................................................... 4-10
4.5 Measurement considerations....................................................................................................................... 4-13
4.5.1 Path isolation.......................................................................................................................................... 4-13
4.5.2 Magnetic fields....................................................................................................................................... 4-14
4.5.3 Radio frequency interference............................................................................................................... 4-14
4.5.4 Ground loops ......................................................................................................................................... 4-14
4.5.5 Keeping connectors clean..................................................................................................................... 4-15
4.5.6 AC frequency response......................................................................................................................... 4-15
5.1 Introduction..................................................................................................................................................... 5-1
5.2 Handling and cleaning precautions.............................................................................................................. 5-1
5.3 Performance verification ................................................................................................................................ 5-2
5.3.1 Environmental conditions...................................................................................................................... 5-2
5.3.2 Recommended equipment..................................................................................................................... 5-2
5.3.3 Matrix card connections ......................................................................................................................... 5-3
5.3.4 Channel resistance tests.......................................................................................................................... 5-3
5.3.5 Offset current tests................................................................................................................................... 5-4
5.3.6 Contact potential tests............................................................................................................................. 5-6
5.3.7 Path isolation tests................................................................................................................................... 5-7
5.3.8 Differential and common-mode isolation tests................................................................................... 5-9
5.4 Special handling of static-sensitive devices............................................................................................... 5-11
5.5 Principles of operation.................................................................................................................................. 5-12
5.5.1 Block diagram ........................................................................................................................................ 5-12
5.5.2 ID data circuits....................................................................................................................................... 5-13
5.5.3 Relay control........................................................................................................................................... 5-13
5.5.4 Relay power control.............................................................................................................................. 5-14
5.5.5 Power-on safeguard.............................................................................................................................. 5-14
5.6 Troubleshooting............................................................................................................................................. 5-15
5.6.1 Troubleshooting equipment................................................................................................................. 5-15
5.6.2 Troubleshooting access......................................................................................................................... 5-15
5.6.3 Troubleshooting procedure.................................................................................................................. 5-15
6.1 Introduction...................................................................................................................................................... 6-1
6.2 Parts lists ............................................................................................................................................................6-1
6.3 Ordering information...................................................................................................................................... 6-1
6.4 Factory service.................................................................................................................................................. 6-1
6.5 Component layouts and schematic diagrams ............................................................................................. 6-2
ii

List of Illustrations

Figure 2-1 Model 7012 simplified schematic.......................................................................................................... 2-1
Figure 2-2 Model 7001 analog backplane ............................................................................................................... 2-2
Figure 2-4 Single-ended switching example .......................................................................................................... 2-3
Figure 2-3 Matrix row connections to backplane .................................................................................................. 2-3
Figure 2-5 Differential switching example ............................................................................................................. 2-4
Figure 2-6 Sensing example...................................................................................................................................... 2-4
Figure 2-7 SMU connections..................................................................................................................................... 2-5
Figure 2-8 Two separate 4 Figure 2-9 Narrow matrix example (4 Figure 2-10 Wide matrix example (8
Figure 2-11 Mixed card type example....................................................................................................................... 2-9
Figure 2-12 Partial matrix expansion (8
×
10 matrices................................................................................................................. 2-6
×
20) ........................................................................................................... 2-7
×
10)................................................................................................................ 2-8
×
20) ........................................................................................................ 2-10
Figure 3-1 Backplane row jumpers.......................................................................................................................... 3-2
Figure 3-2 Screw terminal connector card.............................................................................................................. 3-3
Figure 3-3 Typical terminal block connections...................................................................................................... 3-3
Figure 3-4 Cable clamp for screw terminal connector card ................................................................................. 3-4
Figure 3-5 Multi-pin connector card terminal identification............................................................................... 3-5
Figure 3-6 Typical round cable connection techniques ........................................................................................ 3-7
Figure 3-7 Model 7011-MTR connector pinout...................................................................................................... 3-8
Figure 3-8 Model 7011-KIT-R (with cable) assembly............................................................................................ 3-8
Figure 3-9 Single card system example (multi-pin connector card) ................................................................. 3-10
Figure 3-10 Single card system example (screw terminal connector card)........................................................ 3-11
Figure 3-11 Two-card system example (multi-pin connector card).................................................................... 3-12
Figure 3-12 Two-card system example (screw terminal connector card) .......................................................... 3-13
Figure 3-13 Two-mainframe system example (multi-pin connector card) ........................................................ 3-15
Figure 3-14 Two-mainframe system example (screw terminal connector card)............................................... 3-16
Figure 3-15 Model 7012-S card installation in Model 7001 .................................................................................. 3-18
Figure 3-16 Model 7012-C card installation in Model 7001 ................................................................................. 3-19
Figure 4-1 Channel status display ........................................................................................................................... 4-2
Figure 4-2 Display organization for multiplexer channels .................................................................................. 4-3
Figure 4-3 Model 7012 programming channel assignments................................................................................ 4-3
Figure 4-4 Thick film resistor network testing....................................................................................................... 4-6
Figure 4-5 Four-terminal ohms measurements ..................................................................................................... 4-7
Figure 4-6 Voltage divider checks ........................................................................................................................... 4-9
Figure 4-7 Transistor testing................................................................................................................................... 4-10
Figure 4-8 DC parameter checks............................................................................................................................ 4-12
Figure 4-9 Common-emitter characteristics of an NPN silicon Transistor ...................................................... 4-13
Figure 4-10 Path isolation resistance........................................................................................................................ 4-13
Figure 4-11 Voltage attenuation by path isolation resistance .............................................................................. 4-13
Figure 4-12 Power line ground loops ...................................................................................................................... 4-15
Figure 4-13 Eliminating ground loops .................................................................................................................... 4-15
Figure 5-1 Path resistance testing............................................................................................................................. 5-3
Figure 5-2 Common-mode offset current testing................................................................................................... 5-5
Figure 5-3 Differential offset current testing .......................................................................................................... 5-5
Figure 5-4 Contact potential testing......................................................................................................................... 5-6
Figure 5-5 Path isolation testing (guarded) ............................................................................................................ 5-8
Figure 5-6 Differential isolation testing................................................................................................................. 5-10
Figure 5-7 Common-mode isolation testing......................................................................................................... 5-10
Figure 5-8 Model 7012 block diagram ................................................................................................................... 5-12
Figure 5-9 Start and stop sequences....................................................................................................................... 5-13
Figure 5-10 Transmit and acknowledge sequence................................................................................................. 5-14

List of Tables

Table 3-1 Mass termination accessories ................................................................................................................ 3-4
Table 5-1 Verification equipment........................................................................................................................... 5-2
Table 5-2 Path isolation tests................................................................................................................................... 5-8
Table 5-3 Differential and common-mode isolation testing............................................................................. 5-11
Table 5-4 Recommended Troubleshooting Equipment.................................................................................... 5-15
Table 5-5 Troubleshooting procedure ................................................................................................................. 5-16
v/vi
1

General Information

1.1 Introduction

This section contains general information about the Model 7012 4
There are two basic versions of this matrix card; the Model 7012-S and the Model 7012-C. The Model 7012­S assembly consists of a screw terminal connector card and the relay card. External test circuits are wired directly to the screw terminals of the connector card. Also available from Keithley is the Model 7012-ST. This is an extra screw terminal connector card. With an extra connector card, you can wire a second test system without disturbing the wiring conÞguration of the Þrst test system.
The Model 7012-C assembly consists of a multi-pin (mass termination) connector card and the relay card. External test circuit connections to the matrix are made via the 96-pin male DIN connector on the connector card. Keithley offers a variety of optional accessories that can be used to make connections to the connector card (see paragraph 1.9).
The rest of Section 1 is arranged in the following man­ner:
×
10 Matrix card.
1.4 Manual addenda
1.5 Safety symbols and terms
1.6 SpeciÞcations
1.7 Unpacking and inspection
1.8 Repacking for shipment
1.9 Optional accessories

1.2 Features

The Model 7012 is a two-pole, dual, 4 by 10 columns) matrix card. Some of the key features include:
¥ Low contact potential and offset current for mini-
mal effects on low-level signals.
¥ The connector board detaches from the relay board
allowing easy access to the screw terminals (Model 7012-S) and backplane row jumpers.
¥ Backplane row jumpers. Cutting jumpers discon-
nects matrix rows from the Model 7001 analog backplane.
×
10 (four rows
1.2 Features
1.3 Warranty information
1-1
General Information

1.3 Warranty information

Warranty information is located on the inside front cover of this instruction manual. Should your Model 7012 require warranty service, contact the Keithley rep­resentative or authorized repair facility in your area for further information. When returning the matrix card for repair, be sure to Þll out and include the service form at the back of this manual in order to provide the repair facility with the necessary information.

1.4 Manual addenda

Any improvements or changes concerning the matrix card or manual will be explained in an addendum in­cluded with the card. Addenda are provided in a page replacement format. Simply replace the obsolete pages with the new pages.

1.5 Safety symbols and terms

The following symbols and terms may be found on an instrument or used in this manual.
1.6 Specifications
Model 7012 speciÞcations are found at the front of this manual. These speciÞcations are exclusive of the ma­trix mainframe speciÞcations.

1.7 Unpacking and inspection

1.7.1 Inspection for damage
The Model 7012 is packaged in a re-sealable, anti-static bag to protect it from damage due to static discharge and from contamination that could degrade its perfor­mance. Before removing the card from the bag, observe the following precautions on handling.
Handling precautions:
1. Always grasp the card by the side edges and shields. Do not touch the board surfaces or components.
2. When not installed in a Model 7001 mainframe, keep the card in the anti-static bag and store it in the original packing carton.
The symbol on an instrument indicates that the user should refer to the operating instructions located in the instruction manual.
The symbol on an instrument shows that high volt­age may be present on the terminal(s). Use standard safety precautions to avoid personal contact with these voltages.
The WARNING heading used in this manual explains dangers that might result in personal injury or death. Always read the associated information very carefully before performing the indicated procedure.
The CAUTION heading used in this manual explains hazards that could damage the matrix card. Such dam­age may invalidate the warranty.
!
After removing the card from its anti-static bag, inspect it for any obvious signs of physical damage. Report any such damage to the shipping agent immediately.
1.7.2 Shipping contents
The following items are included with every Model 7012 order:
¥ Model 7012 4
¥ Model 7012 Instruction Manual
¥ Additional accessories as ordered
×
10 Matrix Card
1-2
General Information
1.7.3 Instruction manual
The Model 7012 Instruction Manual is three-hole drilled so that it can be added to the three-ring binder of the Model 7001 Instruction Manual. After removing the plastic wrapping, place the manual in the binder following the mainframe instruction manual. Note that a manual identiÞcation tab is included and should pre­cede the matrix card instruction manual.
If an additional instruction manual is required, order the manual package, Keithley part number 7012-901-
00. The manual package includes an instruction manu­al and any pertinent addenda.

1.8 Repacking for shipment

Should it become necessary to return the Model 7012 for repair, carefully pack the unit in its original packing carton or the equivalent, and include the following in­formation:
¥ Advise as to the warranty status of the matrix card.

1.9 Optional accessories

The following accessories are available for use with the Model 7012:
Model 7012-ST
is identical to the one provided with the Model 7012-S assembly. An extra screw terminal connector card al­lows you to wire a second test system without disturb­ing the wiring conÞguration of the Þrst connector card.
Model 7011-KIT-R
96-pin female DIN connector that will mate directly to the connector on the Model 7012-C or to a standard 96­pin male DIN bulkhead connector (see Model 7011­MTR). This connector uses solder cups for connections to external circuitry. It includes an adapter for a round cable and the housing.
Model 7011-MTC-2
sembly is terminated with a 96-pin female DIN connec­tor on each end. It will mate directly to the connector on the Model 7012-C and to a standard 96-pin male DIN bulkhead connector (see Model 7011-MTR).
This screw terminal connector card
This connection kit includes a
This two-meter round cable as-
¥ Write ATTENTION REPAIR DEPARTMENT on
the shipping label.
¥ Fill out and include the service form located at the
back of this manual.
Model 7011-MTR
connector uses solder cups for connections to external circuitry. It will mate to the Model 7011-KIT-R connec­tor and Model 7011-MTC-2 cable assembly.
This 96-pin male DIN bulkhead
1-3
2

Matrix Switching Basics

2.1 Introduction

This section covers the basics for matrix switching and is arranged as follows:
2.2 Basic matrix conÞguration: Covers the basic 4
10 matrix conÞguration. The signiÞcance of the backplane jumpers is also covered here.
2.3 Typical matrix switching schemes: Explains
some of the basic ways a matrix can be used to source or measure. Covers single-ended switch­ing, differential (ßoating) switching, and sensing.
2.4 Matrix expansion: Discusses the various matrix
conÞgurations that are possible by using multi­ple cards.
2.2 Basic matrix configuration (4
A simpliÞed schematic of the Model 7012 matrix card is shown in Figure 2-1. The card is conÞgured as a 4 10 matrix. Each of the 40 crosspoints is made up of a two-pole switch. By closing the appropriate crosspoint switch, any matrix row can be connected to any col­umn in the matrix.
×
10)
×
Column
110
23456789
1 2
Rows
3 4
Crosspoint (1 of 40)
HI
LO
Figure 2-1
Model 7012 simplified schematic
Backplane jumpers
Notice in Figure 2-1 there are four pairs of backplane jumpers located on the relay card. With the jumpers in­stalled, the matrix card is connected to the analog back-
×
plane of the Model 7001 allowing matrix expansion with a second 7001 card installed in the mainframe. With the jumpers removed (cut), the matrix card is iso­lated from another card installed in the mainframe.
Backplane
Jumpers (4 pairs)
To 7001
Analog
Backplane
2-1
Matrix Switching Basics
The three-pole analog backplane of the mainframe is shown in Figure 2-2. It is through this analog back­plane where the rows of a Model 7012 matrix card in­stalled in one slot can be connected to the rows (or banks) of a compatible card installed in the other slot of the mainframe.
Figure 2-3 shows how each row of the Model 7012 is connected to the backplane. Notice that, since the Mod­el 7012 is a two-pole card, there is no connection made to the Guard terminal of the backplane. The Model 7012 is shipped from the factory with the backplane row jumpers installed.
Model 7001
Card 1 Card 2
Row 1 or Bank A
H
Removing (cutting) the backplane jumpers isolates the card from the backplane, and subsequently, any card installed in the other slot. For information on removing the jumpers, refer to paragraph 3.3.1.
NOTE
The Model 7001 does not provide an analog backplane for the non-701X se­ries cards. As a result, anyone of these cards installed in one slot in the main­frame is electrically isolated from any card installed in the other slot. The only way to connect a Model 7012 to one of these cards is to wire them together.
Analog
Backplane
H
L
G
H
L
G
H
L
G
H
L
G
H = High
L = Low
G = Guard
Row 2 or Bank B
Row 3 or Bank C
Row 4 or Bank D
Row = Matrix Card (7012) Bank = Mux Card (7011)
L
G
H
L
G
H
L
G
H
L
G
Figure 2-2
Model 7001 analog backplane
2-2
Matrix Switching Basics
7012
Matrix Row
(1 of 4)
H
L
H = High L = Low G = Guard
Backplane
Jumpers
Figure 2-3
Matrix row connections to backplane
Row Columns
HI
LO
7001
Analog
Backplane
H
L
G

2.3 Typical matrix switching schemes

The following paragraphs describe some basic switch­ing schemes that are possible with a two-pole switch­ing matrix. These switching schemes include some various shielding conÞgurations to help minimize noise pick up in sensitive measurement applications. These shields are shown connected to chassis ground. For some test conÞgurations, shielding may prove to be more effective connected to circuit common. Chassis ground is accessible at the rear panel of the Model 7001.
2.3.1 Single-ended switching
In the single-ended switching conÞguration, the source or measure instrument is connected to the DUT through a single pathway as shown in Figure 2-4.
H
DUT
L
Optional
Shield
Source or
Measure
Figure 2-4
Single-ended switching example
7012
2-3
Matrix Switching Basics
2.3.2 Differential switching
The differential or ßoating switching conÞguration is shown in Figure 2-5. The advantage of using this con­Þguration is that the terminals of the source or measure instrument are not conÞned to the same matrix cross­point. Each terminal of the instrument can be connect­ed to any matrix crosspoint.
Rows Columns
HI
LO
Source or
Measure
2.3.3 Sensing
Figure 2-6 shows how the matrix card can be conÞg- ured to use instruments that have sensing capability. The main advantage of using sensing is to cancel the ef­fects of matrix card path resistance (<1 tance of external cabling. Whenever path resistance is a consideration, sensing should be used.
H
L
DUT
H
L
7012
) and the resis-
Figure 2-5
Differential switching example
Source HI Sense HI
Sense LO Source LO
Source or
Measure
Figure 2-6
Sensing example
Rows Columns
H L
DUT H L
7012
2-4
Matrix Switching Basics
2.3.4 SMU connections
Figure 2-7 shows how a Keithley Model 236, 237 or 238 Source Measure Unit could be connected to the matrix card. By using triax cables that are unterminated at one end, the driven guard and chassis ground are physical­ly extended all the way to the card.

2.4 Matrix expansion

With the use of additional matrix cards and main­frames, larger matrices can be conÞgured. Each Model
Rows Columns
Output HI
Guard
7001 Switch System mainframe will accommodate up to two cards, and up to six mainframes can be connect­ed together. Thus, a switch system using as many as 12 cards in six mainframes can be conÞgured.
2.4.1 Two-card switching systems
Each Model 7001 Switch System mainframe can ac­commodate two cards to allow the following switching conÞgurations.
H
L
Figure 2-7
SMU connections
Sense HI
Guard
Sense LO
Output LO
Output LO
Triax
Cables (3)
236/237/238
WARNING : Hazardous voltages may be present on GUARD. Make sure all cable shields are properly insulated before applying power.
H
L
7012
DUT
2-5
Matrix Switching Basics
Separate switching systems
Two single-card systems can be conÞgured by remov­ing the backplane jumpers from one of the cards. The two cards will be controlled by the same mainframe, but they will be electrically isolated from each other. Figure 2-8 shows an example using two Model 7012 matrix cards.
Card 1
7012
110
1
2
Rows
3
Columns
Narrow matrix expansion (4
A narrow 4 row
×
20 column matrix is conÞgured by
×
20 matrix)
simply installing two “as shipped” Model 7012s in the Model 7001 mainframe. By leaving the backplane jumpers installed, the rows of the matrix card installed in slot 1 (CARD 1) are automatically connected to the rows of the matrix card installed in slot 2 (CARD 2) through the analog backplane. The 4
×
shown in Figure 2-9.
Card 2
7001
Analog
Backplane
110
7012
Columns
20 matrix is
1
2
Rows
3
4
4 x 10 Matrix
Figure 2-8
Two separate 4 × 10 matrices
4
4 x 10 Matrix
Jumpers
Removed
2-6
Matrix Switching Basics
Card 1
7012
110
1
2
Rows
3
4
Notes : Backplane jumpers on both cards must be installed.
Columns
4 x 20 Matrix
Figure 2-9
Narrow matrix example (4 × 20)
Wide matrix expansion (8
A wide 8 row
×
10 column matrix is shown in Figure 2-
×
10 matrix)
10. For this conÞguration, the 10 columns of the two matrix cards must be physically hard-wired together. Also note that the backplane jumpers on one of the cards must be removed in order to isolate the rows of the two cards from each other.
Mixing card types
Different types of cards can be used together to create some unique switching systems. For example, you
Card 2
7001
Analog
Backplane
11 20
7012
Columns
could have a Model 7012 matrix card installed in one slot and a Model 7011 multiplexer card installed in the other slot.
Figure 2-11 shows a possible switching system using a matrix card and a multiplexer card. The backplane jumpers for both the matrix and multiplexer cards must be installed. This allows matrix rows to be con­nected to multiplexer banks. On the multiplexer card, the bank-to-bank jumpers must be removed to main­tain isolation between matrix rows. See the instruction manual for the Model 7011 for complete information on the multiplexer card.
2-7
Matrix Switching Basics
External
Column
Jumpers
Card 1
7012
110
1
2
Rows
3
4
Columns
Jumpers
Removed
7001 Analog
Backplane
1
2
Rows
3
4
Figure 2-10
Wide matrix example (8 × 10)
7012
Card 2
8 x 10 Matrix
2-8
Matrix Switching Basics
110
1
2
Rows
3
4
Notes : 1. Models 7011 and 7012 backplane jumpers must be installed.
2. Model 7011 bank-to-bank jumpers must be removed.
Figure 2-11
Mixed card type example
Card 1
7012
Columns
4 x 10 Matrix
7001
Backplane
Card 2
7011
110
110
110
110
Inputs
Bank A
Bank B
Bank C
Bank D
Quad 1 x 10 Mux
2.4.2 Mainframe matrix expansion
Matrices using up to 12 matrix cards are possible by us­ing six Model 7001 mainframes together. Using 12 Model 7012 matrix cards provides 480 cross-points.
In general, connecting the rows of a card in one main­frame to the rows of a card in a second mainframe in­creases the column numbers of the matrix. For example, if the rows of a 4 frame are connected to the rows of a 4 second mainframe, the resulting matrix would be 4
40. Paragraph 3.4.3 explains how to connect a test sys­tem using two mainframes.
Partial matrix implementation
A fully implemented matrix provides a relay at each potential crosspoint. For example, a fully implemented 8
×
20 matrix utilizing four 4 160 crosspoints. A partially implemented 8 would contain fewer crosspoints.
×
20 matrix in one main-
×
20 matrix in a
×
10 matrix cards contains
×
20 matrix
An example of a partially implemented 8 shown in Figure 2-12. The partial matrix is still consid­ered 8
×
20, but contains only 120 crosspoints using three Model 7012 matrix cards installed in two Model 7001 mainframes.
Matrix card #1 (7012 #1) installed in one of the slots of the Þrst mainframe (7001 #1) provides a 4 The other slot of the Þrst mainframe should be left empty. If another switching card is left in that slot, make sure it is isolated from the analog backplane (i.e. backplane jumpers removed). The two matrix cards (7012 #2 and #3) installed in the second mainframe
×
(7001 #2) are conÞgured as a an 8
×
10 matrix as ex­plained in paragraph 2.4.2 (Wide Matrix Expansion). Keep in mind that the rows of card #2 must be isolated from the rows of card #3. This is accomplished by re­moving the jumpers on one of the two cards. Finally, the partially implemented 8
×
20 matrix is realized by externally hard-wiring the rows of card #1 to the rows of card #2.
×
20 matrix is
×
10 matrix.
2-9
Matrix Switching Basics
An obvious advantage of a partial matrix is that fewer matrix cards are needed. Another reason to use a par­tial matrix is to keep speciÞc devices from being con­nected directly to other devices. For example, a source connected to rows 5, 6, 7 or 8 (Figure 2-12) cannot be
7001 #1
7012 #1
Columns
10
Rows
Rows
1
1 2 3 4
5 6 7 8
connected to a column of Model 7012 #1 with one “ac­cidental” crosspoint closure. Three speciÞc crosspoints must be closed in order to route the source signal to a column of card #1.
7001 #2
External
Row
Jumpers
11
7012 #2
Columns
20
Figure 2-12
Partial matrix expansion (8 × 20)
7012 #3
2-10
3

Card Connections & Installation

3.1 Introduction

WARNING
The procedures in this section are in­tended only for qualiÞed service per­sonnel. Do not perform these procedures unless qualiÞed to do so. Failure to recognize and observe nor­mal safety precautions could result in personal injury or death.
The information in this section is arranged as follows:
3.2 Handling precaution: Explains precautions that
must be followed to prevent contamination to the matrix card assembly. Contamination could de­grade the performance of the matrix card.
3.3 Connection:
ternal circuitry to the two available connector cards for the matrix; the screw terminal connec­tor card and the multi-pin connector card.
Covers the basics for connecting ex-

3.2 Handling precautions

To maintain high impedance isolation, care should be taken when handling the relay card to avoid contami­nation from such foreign materials as body oils. Such contamination can substantially lower leakage resis­tances, thus degrading performance.
To avoid possible contamination, always grasp the re­lay and connector cards by the side edges or shields. Do not touch the board surfaces or components. On connectors, do not touch areas adjacent to the electrical contacts. Dirt build-up over a period of time is another possible source of contamination. To avoid this prob­lem, operate the mainframe and matrix card in a clean environment.
If a card becomes contaminated, it should be thorough­ly cleaned as explained in paragraph 5.2.
3.4 Typical connection scheme: Provides some typi-
cal connection schemes for single card, two-card and two-mainframe system conÞgurations.
3.5 Model 7012 installation: Provides a procedure to
install the matrix card assembly in the Model 7001 mainframe.

3.3 Connections

This paragraph provides the basic information needed to connect your external test circuitry to the matrix. It includes the removal/installation of the backplane row jumpers on the relay card, and detailed information on the two available connector cards.
3-1
Card Connections & Installation
WARNING
The following connection informa­tion is intended to be used by quali­Þed service personnel. Failure to recognize and observe standard safe­ty precautions could result in person­al injury or death.
3.3.1 Backplane row jumpers
The Model 7001 mainframe has an analog backplane that allows the rows of a Model 7012 matrix to be inter­nally connected to a compatible switching card in­stalled in the other slot (see paragraph 2.4.1 for details).
The backplane row jumpers for the matrix card assem­bly are located on the relay card as shown in Figure 3-
1. The card is shipped from the factory with the jump­ers installed.
2. Physically remove a cut jumper by unsoldering it from the PC board.
3. Install a new #22 AWG jumper wire (Keithley P/N J-15) and solder it to the PC board.
4. Remove the solder ßux from the PC board. The cleaning procedure is explained in paragraph 5.2.
7012 Relay Card
Row 1
Row 2
Row 3
Row 4
H
L
H
L
H
L
H L
Jumper removal
Perform the following steps to remove backplane row jumpers:
1. If mated together, separate the relay card from the connector card by removing the mounting screw and then pulling the two cards away from each other. Remember to only handle the cards by the edges and shields to avoid contamination.
2. Use Figure 3-1 to locate the jumper(s) that are to be removed.
3. It is not necessary to physically remove the jump­ers from the PC board. Using a pair of wire cutters, cut one lead of each jumper.
Jumper installation
Referring to Figure 3-1 for jumper locations, perform the following steps to install backplane row jumpers:
1. If mated together, separate the relay card from the connector card by removing the mounting screw and then pulling the two cards away from each other. Remember to only handle the cards by the edges and shields to avoid contamination.
Jumpers
Figure 3-1
Backplane row jumpers
3.3.2 Screw terminal connector card
The screw terminal connector card is shown in Figure 3-2. Connections are made directly to the screw termi­nals of the four terminal blocks. Each screw terminal will accommodate #16-22 AWG wire.
Wiring procedure
Perform the following procedure to wire circuitry to the screw terminal connector card:
WARNING
Make sure all power is off and any stored energy in external circuitry is discharged.
3-2
Figure 3-3
Typical terminal block connections
H
L
H
L
H
L
COL 1
COL 2
COL 3
DUT
#16 - 22 AWG Wires
Card Connections & Installation
1. If mated together, separate the connector card from the relay card by removing the mounting screw and then pulling the two cards away from each other. Remember to only handle the cards by the edges and shields to avoid contamination.
2. Using an insulated screwdriver, connect the cir­cuitry to the appropriate terminals. Figure 3-3 shows how Column 1 of the matrix would be con­nected to a DUT.
3. Referring to Figure 3-4, remove the top half of the cable clamp as follows:
A. Loosen the cable clamp screw enough to disen-
gage it from the bottom half of the cable clamp.
B. Using your thumb and foreÞnger, press the re-
taining clips inward and, with your other hand, remove the top half of the clamp.
4. Route wires under wire guide/connector shim.
5. Route the wires through the bottom half of the ca­ble clamp.
6. Replace the top half of the clamp. It simply snaps onto the bottom half of the clamp. Tighten the ca­ble clamp screw. The clamp serves as a strain relief for terminal block wires.
7. Mate the connector card to the relay card. The Model 7012 is now ready to be installed in the Model 7001 mainframe. See paragraph 3.5 for de­tails.
Figure 3-2
Screw terminal connector card
3-3
Card Connections & Installation
Screw
Clips
Figure 3-4
Cable clamp for screw terminal connector card
3.3.3 Multi-pin (mass termination) connec­tor card
Since connections to external circuitry are made at the 96-pin male DIN bulkhead connector, there is no need to separate the connector card from the relay card. If the connector card is separated from the relay card, carefully mate them together. Make sure to handle the cards by the edges and shields to avoid contamination.
Terminal identiÞcation for the DIN connector of the multi-pin connector card is provided by Figure 3-5. This connector will mate to a 96-pin female DIN con­nector.
Keithley has a variety of cable and connector accesso­ries available to accommodate connections from the connector card to test instrumentation and DUT (de­vices under test). In general, these accessories, which are summarized in Table 3-1, utilize a round cable as­sembly for connections.
Table 3-1
Mass termination accessories
Model Description
7011-KIT-R 96-pin female DIN connector and
housing for round cable.
7011-MTC-2 Two-meter round cable assembly
terminated with a 96-pin female DIN connector on each end.
7011-MTR 96-pin male DIN bulkhead connec-
tor.
3-4
Pins of the Model 7012-C mass termination connector can be identified in one of three ways:
1. Matrix row (1-4) or column (1-10).
2. Connector designation, consisting of rows a-c and columns 1-32.
3. Schematic and component layout designation (1-96).
The following pinout diagrams show the correspondence between these arrangements:
3231302928272625242322212019181716151413121110987654321
c
b a
Card Connections & Installation
View from pin side of connector
Matrix
Terminal
Row 1 HI
LO
Row 2 HI
LO
Row 3 HI
LO
Row 4 HI
LO
Connector
Desig.
1a-32c
8c 8b
6c 6b
4c 4b
2c 2b
Schematic
Desig.
1-96
72 40
70 38
68 36
66 34
Matrix
Terminal
Col 1 HI
LO
Col 2 HI
LO
Col 3 HI
LO
Col 4 HI
LO
Col 5 HI
LO
Col 6 HI
LO
Col 7 HI
LO
Col 8 HI
LO
Col 9 HI
LO
Col 10 HI
LO
Connector
Desig. 1a-32c
32c 32b
30c 30b
28c 28b
26c 26b
24c 24b
22c 22b
20c 20b
18c 18b
16c 16b
14c 14b
Schematic
Desig.
1-96
96 64
94 62
92 60
90 58
88 56
86 54
84 52
82 50
80 48
78 46
Notes:
1. Refer to the schematic for shield pins.
2. Short pins 1a to 1b on the mating connector (pins 1 and 33 on schematic) to allow the output relays on the con­nector card to close.
Figure 3-5
Multi-pin connector card terminal identification
3-5
Card Connections & Installation
Typical connection techniques
All external circuitry, such as instrumentation and DUTs, that you wish to connect to the matrix card must be terminated with a single 96-pin female DIN connec­tor. The following connection techniques provide some guidelines and suggestions for wiring your circuitry.
WARNING
Before beginning any wiring proce­dures, make sure all power is off and any stored energy in external circuit­ry is discharged.
NOTE
It is recommended that external cir­cuitry be connected (plugged in) after the Model 7012 assembly is installed in the Model 7001 mainframe and with the 7001 power off. Installation is covered in paragraph 3.5.
Round cable assemblies
Figure 3-6 shows typical round cable connection techniques using accessories available from Keithley.
In Figure 3-6A, connections are accomplished using a Model 7011-MTC-2 cable and a Model 7011-MTR bulk­head connector. The two-meter round cable is termi­nated with a 96-pin female DIN connector at each end. This cable mates directly to the multi-pin connector card and to the bulkhead connector. The bulkhead con­nector has solder cups to allow direct connection to in­strumentation and DUT. Figure 3-7 provides pinout for the bulkhead connector. The view shown is from the solder cup end of the connector.
In Figure 3-6B, connections are accomplished using a Model 7011-MTC-2 cable assembly that is cut in half. The 96-pin female DIN connector on one end of the ca­ble mates directly to the multi-pin connector card. The unterminated end of the cable is wired directly to in­strumentation and DUT. The other half of the cable as­sembly could be used for a second switching card.
Output relays
The multi-pin connector card uses a relay for each of the four output banks. These output relays are normally open to prevent any hazardous voltages (via the mainframe backplane) from appear­ing on the pins of the male DIN connector. The output relays will only close when the Model 7011-MTC-2 ca­ble assembly is connected to card. If building you own cable assembly, you must make sure that it shorts pins 1a to 1b of the card connector (Figure 3-8) when it is mated to the card. Shorting pins 1a to 1b allows the output relays to close.
In Figure 3-6C, connections are accomplished using a custom-built cable assembly that consists of a Model 7011-KIT-R connector and a suitable round cable. Hita­chi cable p/n N2807-P/D-50TAB is a 50-conductor round cable. Two of these cables can be used to provide 100 conductors. The connector has solder cups to ac­commodate the individual wires of the unterminated cable. Figure 3-8 provides an exploded view of the con­nector assembly and shows how the cable is connected. The connector end of the resultant cable assembly mates directly to the multi-pin connector card. The un­terminated end of the cable assembly is wired directly to instrumentation and DUT.
3-6
T
A)
Multi-Pin
Connector
Card
Card Connections & Installation
Wire instrumentation and DUT to bulkhead connector (See Figures 3-5 and 3-7 for terminal identification)
7011-MTC-2
cable assembly
7011-MTR
bulkhead connector
B)
Connector
Card
Multi-Pin
Multi-Pin
C)
Connector
Card
Figure 3-6
ypical round cable connection techniques
Wire directly to
instrumentation
and DUT
7011-MTC-2
(Cut in Half)
Wire directly to
instrumentation
and DUT
Cable
7011-Kit-R
Connector Kit
Notes : Figure 3-8 provides an exploded view showing how the connector (with cable) is assembled.
Cable Hitachi p/n N2807-P/D-50TAB is a 50-conductor cable. Two of these cables can be used to supply 100 conductors.
3-7
Card Connections & Installation
Note : See Figure 3-5 for terminal identification.
Figure 3-7
Model 7011-MTR connector pinout
3231302928272625242322212019181716151413121110987654321
c b a
View from solder cup side of connector
Figure 3-8
Model 7011-KIT-R (with cable) assembly
3-8
Card Connections & Installation

3.4 Typical connection schemes

The following information provides some typical con­nection schemes for single card, two-card, and two­mainframe system conÞgurations. Connection schemes for the multi-pin connector card use some of the techniques presented in paragraph 3.3.2. Keep in mind that these are only examples to demonstrate var­ious ways to wire a test system. Connection details for both connector cards (multi-pin and screw terminal connector cards) are provided in paragraph 3.3.
3.4.1 Single card system
Figure 3-9 shows how external connections can be made to a single card system that uses the multi-pin connector card. Instrumentation and DUT are hard­wired to the Model 7011-MTR male bulkhead connec­tor. This connector has solder cups that will accept wire size up to #24 AWG. The test system is connected to the matrix using the Model 7011-MTC-2 round cable as­sembly. This cable mates directly to both the external bulkhead connector and the Model 7012-C matrix card assembly. Notice that the bulkhead connector is shown mounted to a Þxture to help keep the cabling stable during the test.
The single card system in Figure 3-10 is identical to the system in the previous illustration, except for the con­nection scheme. The system in Figure 3-10 uses the ter­minal block connector card. With this card, single conductor connections are made directly from the ter­minal blocks of the connector card to instrumentation and DUT.
When using a single card system, make sure that the card remains electrically isolated from any other
switching cards. There are several ways to ensure iso­lation for a single card in the Model 7001 mainframe:
1. Vacate the other mainframe slot. If there is a Model 701X card installed in the other slot, remove it.
2. Remove the backplane jumpers on the matrix card. This will disconnect the card from the analog back­plane of the mainframe.
3. Remove the backplane jumpers from the switching card installed in the other slot.
3.4.2 Two-card system
Both Figure 3-11 and Figure 3-12 show a system using two matrix cards installed in one Model 7001 main­frame to conÞgure a 4 connection schemes, row connections of the two matrix cards are accomplished internally through the back­plane of the Model 7001 mainframe. To connect rows internally, the backplane row jumpers of both matrix cards must be installed.
Figure 3-11 shows how external connections can be made for the multi-pin connector cards. In this exam­ple, a single Model 7011-MTC-2 round cable assembly is cut in half to provide two cables, each of which is un­terminated at one end. The unterminated ends of the two cables are hard-wired to instrumentation and DUT as shown in the drawing. The other ends of these cables mate directly to the Model 7012-C matrix card assem­blies.
Figure 3-12 shows how external connections can be made for the screw terminal connector card. Single conductor connections are made directly from the screw terminals of the connector card to instrumenta­tion and DUT.
20 test matrix. In both these
×
3-9
Card Connections & Installation
Instrument
Row 1
Fixture for
Bulkhead
Connector
7012-C
Instrument
Instrument
Instrument
Row 2
28 Individual Conductors
Row 3
Row 4
12345678910
DUT Test Fixture
Instruments
7011-MTR
Bulkhead
Connector
DUT
12 345678910
12 345678910
1
2
3
4
DUT
Columns
Rows
7011-MTC-2
Cable Assembly
Figure 3-9
Single card system example (multi-pin connector card)
3-10
Equivalent Circuit
Instrument
Instrument
Instrument
Instrument
Row 1
Row 2
Row 3
Row 4
Card Connections & Installation
7012-S
12345678910
DUT Test Fixture
12345678910
12345678910
1
Instruments
2
3
4
Equivalent Circuit
Figure 3-10
Single card system example (screw terminal connector card)
DUT
DUT
Columns
Rows
3-11
Card Connections & Installation
Instrument
Row 1
Row 2
Instrument
Row 3
Instrument
Row 4
Instrument
12345678910
DUT Test Fixture
1234 5678910
1234 5678910
1
Instruments
2
3
4
Note : Backplane row jumpers for both matrix cards must be installed.
7011-MTC-2 Cable Assembly (Cut in half to provide two cables)
7001
7012-C
7012-C
12345678910
DUT Test Fixture
7001
Backplane
DUT
DUT
DUT
1234 5678910
1234 5678910
Column Column
CARD 1 CARD 2
Backplane Row
Jumpers installed
DUT
C A R D 1
C A R D 2
Row
Figure 3-11
Two-card system example (multi-pin connector card)
3-12
Equivalent Circuit
T
Card Connections & Installation
Instrument
Instrument
Instrument
Instrument
Instruments
Row 1
Row 2
Row 3
Row 4
12345678910
DUT Test Fixture
DUT
DUT
1234 5678910
1234 5678910
1
2
3
4
Column Column
Note : Backplane row jumpers for both cards must be installed.
7012-S
7012-S
12345678910
7001
Backplane
7001
C A R D 1
C A R D 2
DUT Test Fixture
DUT
DUT
1234 5678910
1234 5678910
Row
CARD 1 CARD 2
Figure 3-12
wo-card system example (screw terminal connector card)
Backplane Row
Jumpers installed
Equivalent Circuit
3-13
Card Connections & Installation
3.4.3 Two-mainframe system
Figure 3-13 and Figure 3-14 show a system using three matrix cards in two Model 7001 mainframes to conÞg­ure a 4 Two-card System (see previous paragraph), except that a third matrix card (installed in a second mainframe) is added.
Figure 3-13 shows the connection scheme for the multi­pin connector cards. External circuit connections to the Model 7001 #1 mainframe are identical to the ones used for the Two-card System. The third matrix card (installed in Model 7001 #2 mainframe) shows how a custom-built cable can be used to make connections to external circuitry. A suitable round cable can be termi­nated with a 96-pin female DIN connector (Model 7011-KIT-R) that will mate to the Model 7012-C matrix card assembly. The unterminated end of the cable is connected directly to instrumentation and DUT. Notice that the row connections for the third matrix card are made at the instruments.
Figure 3-14 shows connections for the screw terminal connector card. Single conductor connections are made directly from the screw terminals of the connector card to instrumentation and DUT
×
30 test matrix. This system is similar to the
3.5 Model 7012 installation and re­moval
This paragraph explains how to install and remove the Model 7012 matrix card assembly from the Model 7001 mainframe.
WARNING
Installation or removal of the Model 7012 is to be performed by qualiÞed service personnel. Failure to recog­nize and observe standard safety pre­cautions could result in personal injury or death.
NOTE
If using the screw terminal connector card, make sure your external circuit­ry is wired to the card (as explained in paragraph 3.3.1) before installing the card assembly in the Model 7001 mainframe.
CAUTION
To prevent contamination to the ma­trix card that could degrade perfor­mance, only handle the card assembly by the edges and shields.
3-14
Card Connections & Installation
DUT Test Fixture
12345678910
Instrument
Instrument
Instrument
Instrument
12345678910
DUT Test Fixture
7011-MTC-2 Cable Assembly (Cut in half to provide two cables)
7011-Kit-R
Connector Kit
Cable
7001 #2
C A
7012-C
7012-C
7001 #1
7012-C
7012-C
R D 1
C A R D 2
C A R D 1
C A R D 2
Trigger Link
I
N O
U T
Trigger Link
I
N O
U T
12345678910
DUT Test Fixture
Trigger
Link
Cable
Note : Backplane row jumpers for both cards in 7001 #1 must be installed.
7001 #1 7001 #2
7001
Backplane
DUT
Instruments
1234 5678910
DUT
1234 5678910
1
2
3
4
1234 5678910
1234 5678910
Column Column Column
CARD 1 CARD 2 CARD 3
Backplane Row
Jumpers installed
Equivalent Circuit
Figure 3-13
Two-mainframe system example (multi-pin connector card)
DUT
DUT
DUT
1234 5678910
1234 5678910
DUT
External Row Jumpers
Row
3-15
Card Connections & Installation
12345678910
Instrument
DUT Test Fixture
7012-S
Not Used
7001 #2
C A
Trigger Link
R D 1
I
N
C A
O
R D
U
2
T
Trigger
Link
Cable
Instruments
Instrument
Instrument
Instrument
12345678910
DUT Test Fixture
DUT
1234 5678910
1234 5678910
1
2
3
4
DUT
Column Column Column
CARD 1 CARD 2 CARD 3
7001 #1
C A
7012-S
7012-S
Trigger Link
R D
1
I N
C A
O
R D
U
2
T
Note : Backplane row jumpers for both cards in 7001 #1 must be installed.
12345678910
DUT Test Fixture
7001 #1 7001 #2
7001
Backplane
DUT
DUT
1234 5678910
1234 5678910
Backplane Row
Jumpers installed
1234 5678910
1234 5678910
DUT
DUT
Row
Equivalent Circuit
Figure 3-14
Two-mainframe system example (screw terminal connector card)
3-16
External Row Jumpers
Card Connections & Installation
Matrix card installation
Perform the following steps to install the matrix card assembly in the Model 7001 mainframe:
WARNING
Turn off power from all instrumenta­tion (including the Model 7001 main­frame) and disconnect their line cords. Make sure all power is re­moved and stored energy in external circuitry is discharged.
1. Mate the connector card to the relay card if they are separated. Install the supplied 4-40 screw at the end of the card to secure the assembly. Make sure to handle the cards by the edges and shields to pre­vent contamination.
2. Facing the rear panel of the Model 7001, select the slot (CARD 1 or CARD 2) that you wish to install the card in.
3. Referring to Figure 3-15 for Model 7012-S installa­tion, or Figure 3-16 for Model 7012-C installation, feed the matrix card assembly into the desired slot such that the edges of the relay card ride in the rails.
4. With the ejector arms in the unlocked position, push the card assembly all the way into the main­frame until the arms engage into the ejector cups. Then push both arms inward to lock the card into the mainframe.
5. For the 7012-C, also install the screw shown in Fig­ure 3-16.
Matrix card removal
To remove the matrix card assembly, Þrst unlock it by pulling the latches outward, then pull the card assem­bly out of the mainframe. Remember to handle the card assembly by the edges and shields to avoid contamina­tion that could degrade performance.
3-17
Card Connections & Installation
Unlock car d
Unlock card
Ejector Arms (2)
Ejector Arms (2)
Figure 3-15
Model 7012-S card installation in Model 7001
3-18
Lock card
Lock car d
1
Screw
Screw
Unlock card
2
Unlock car d
Card Connections & Installation
Ejector Arms (2)
Ejector Arms (2)
Screw
2
Figure 3-16
Model 7012-C card installation in Model 7001
1
Lock card
Lock car dScrew
3-19
4

Operation

4.1 Introduction

The information in this section is arranged as follows:
4.2 Power limits: Summarizes the maximum power
limits of the Model 7012 matrix card assembly.
4.3 Mainframe control of matrix card: Summarizes
programming steps to control the matrix card from the Model 7001 Switch System mainframe.
4.4 Multiplexer switching examples: Provides some
typical applications for using the Model 7012.
4.5 Measurement considerations: Reviews a num-
ber of considerations when using the Model 7012 to make measurements.

4.2 Power limits

CAUTION
To prevent damage to the card, do not exceed the maximum signal level speciÞcations of the card.
Maximum signal levels
To prevent overheating or damage to the relays, never exceed the following maximum signal levels:
DC signals: 110V between any two pins (termi-
nals), 1A switched, 30VA (resistive load)
AC signals: 125V rms or 175V AC peak be-
tween any two pins (terminals), 1A switched, 60VA (resistive load)

4.3 Mainframe control of matrix card

The following information pertains to the Model 7012 matrix card. It assumes that you are familiar with the operation of the Model 7001 mainframe.
If you are not familiar with the operation of the main­frame, it is recommended that you proceed to Getting Started (Section 3) in the Model 7001 Instruction Man­ual after reading the following information.
4-1
Operation
4.3.1 Channel assignments
The Model 7001 has a channel status display (Figure 4-
1) that provides the real-time state of each available channel. The left portion of the display is for slot 1 (Card 1), and the right portion is for slot 2 (Card 2).
Notice that the matrix organization of the channel sta­tus display corresponds directly to the 4 tion of the matrix card. With a matrix card installed, the top row of the display corresponds to Row 1 of the ma­trix card. The 10 columns of the matrix are labeled 1 through 10 on the display. The next rows down corre­spond to Rows 2, 3, and 4 respectively.
Matrix organization of the channel status display cor­responds to the 4
×
10 organization of the matrix card
×
10 organiza-
7001 Display
as shown in Figure 4-2. Each channel is designated as a row/column crosspoint.
To control the matrix card from the mainframe, each matrix crosspoint must have a unique CHANNEL as­signment. The CHANNEL assignments for the matrix card are provided in Figure 4-3. Each CHANNEL as­signment is made up of the slot designator (1 or 2) and the matrix crosspoint. To be consistent with Model 7001 operation, the slot designator, row and column are separated by exclamation points (!). Some exam­ples of CHANNEL assignments:
CHANNEL 1!1!1 = Slot 1, Row 1, Column 1 CHANNEL 1!4!10 = Slot 1, Row 4, Column 10 CHANNEL 2!2!9 = Slot 2, Row 2, Column 9 CHANNEL 2!3!4 = Slot 2, Row 3, Column 4
1 234567891012345678910
= Open Channel = Closed Channel
Figure 4-1
Channel status display
CARD 1 CARD 2
4-2
Operation
Column
123456 78910
R1,C1 R1,C3 R1,C4 R1,C5 R1,C6 R1,C7 R1,C8 R1,C9 R1,C10
1
R1,C2
2
R2,C1 R2,C2 R2,C3 R2,C4 R2,C5 R2,C6 R2,C7
Row
3
R3,C1 R3,C2 R3,C3 R3,C4 R3,C5 R3,C6 R3,C7 R3,C8 R3,C9 R3,C10
4
R4,C1 R4,C2 R4,C3 R4,C4 R4,C5 R4,C6 R4,C7 R4,C8 R4,C9 R4,C10
R = Row C = Column
Figure 4-2
Display organization for multiplexer channels
1 2 3 4 5 6 7 8 9 10
1!1!1
1!2!1
1!1!2
1!2!2
1!1!3
1!2!3
1!1!4
1!2!4
1!1!5
1!2!5
1!1!6
1!2!6
1!1!7
1!2!7
R2,C8
R2,C9 R1,C10
1!1!8
1!2!8
1!1!9
1!2!9
1!1!10
1!2!10
1!3!1
1!3!2
1!3!3
A. Slot 1 (Card 1)
1!4!1
1!4!2
1!4!3
1 2 3 4 5 6 7 8 9 10
2!1!1
2!2!1
2!3!1
2!1!2
2!2!2
2!3!2
2!1!3
2!2!3
2!3!3
B. Slot 2 (Card 2)
2!4!1
2!4!2
2!4!3
Examples : 1!2!4 = Slot 1, Row 2, Column 4 2!3!6 = Slot 2, Row 3, Column 6
Figure 4-3
Model 7012 programming channel assignments
1!3!4
1!4!4
2!1!4
2!2!4
2!3!4
2!4!4
1!3!5
1!4!5
2!1!5
2!2!5
2!3!5
2!4!5
1!3!6
1!4!6
2!1!6
2!2!6
2!3!6
2!4!6
1!3!7
1!4!7
2!1!7
2!2!7
2!3!7
2!4!7
1!3!8
1!4!8
2!1!8
2!2!8
2!3!8
2!4!8
1!3!9
1!4!9
2!1!9
2!2!9
2!3!9
2!4!9
1!3!10
1!4!10
2!1!10
2!2!10
2!3!10
2!4!10
4-3
Operation
4.3.2 Front panel control
Closing and opening channels
A matrix crosspoint is closed form the front panel by simply keying in the channel assignment and pressing CLOSE. For example, to close Row 3, Column 4 cross­point of a matrix card installed in slot 2, key in the fol­lowing channel list and press CLOSE:
SELECT CHANNELS 2!3!4
The above closed channel can be opened by pressing OPEN or OPEN ALL. The OPEN key opens only the channels speciÞed in the channel list, and OPEN ALL opens all channels.
The following display is an example of a channel list that consists of several channels:
SELECT CHANNELS 2!1!1, 2!1!3, 2!2!1-2!2!5
Notice that channel entries are separated by commas (,). A comma is inserted by pressing ENTER or the right cursor key ( ). The channel range is speciÞed by us­ing the hyphen (-) key to separate the range limits. Pressing CLOSE will close all the channels speciÞed in the channel list. Pressing OPEN (or OPEN ALL) will open the channels.
A manual scan can be performed by using the RESET default conditions of the Model 7001. RESET is selected from the SAVESETUP menu of the main MENU. When RESET is performed, the mainframe is conÞgured for an inÞnite number of manual scans. The Þrst press of STEP takes the mainframe out of the idle state. The next press of STEP will close the Þrst channel speciÞed in the scan list. Each subsequent press of STEP will se­lect the next channel in the scan list.
4.3.3 IEEE-488 bus operation
Bus operation is demonstrated using HP BASIC 4.0. The programming statements assume that the primary address of the mainframe is 07.
Closing and opening channels
The following SCPI commands are used to close and open channels:
:CLOSe <list>
:OPEN <list>|ALL
The following statement closes channels 1!1!1, and 1!1!3 through 1!1!6:
Scanning channels
Matrix crosspoints are scanned by creating a scan list and conÞguring the Model 7001 to perform a scan. The scan list is created in the same manner as a channel list (see Closing and Opening Channels). However, the scan list is speciÞed from the “SCAN CHANNEL” dis­play mode. (The SCAN LIST key toggles between the channel list and scan list.) The following shows an ex­ample of a scan list:
SCAN CHANNELS 2!1!1, 2!1!3, 2!2!1-2!2!5
When a scan is performed, the channels speciÞed in the scan list will be scanned in the order that they are pre­sented in the scan list.
OUTPUT 707; “:clos (@ 1!1!1, 1!1!3:1!1!6)”
Notice that the colon (:) is used to separate the range limits.
Either of the following statements will open channels 1!1!1, and 1!1!3 through 1!1!6:
OUTPUT 707; “:open (@ 1!1!1, 1!1!3:1!1!6)”
OUTPUT 707; “:open all”
Scanning channels
There are many commands associated with scanning. However, it is possible to conÞgure a scan using as lit­tle as four commands. These commands are listed as follows:
4-4
Operation
*RST :TRIGger:SEQuence:COUNt:AUTO ON” :SCAN <list> :INIT
The Þrst command resets the mainframe to a default scan conÞguration. The second command automatical­ly sets the channel count to the number of channels in the scan list, the third command deÞnes the scan list, and the fourth command takes the Model 7001 out of the idle state.
The following program will perform a single scan through all 40 channels of a multiplexer card installed in slot 1:
10 OUTPUT 707; “*RST” 20 OUTPUT 707; “:trig:seq:coun:auto on” 30 OUTPUT 707; “:scan (@ 1!1!1:1!4!10)” 40 OUTPUT 707; “:init” 50 END
5 OUTPUT 707; “:open all” 10 OUTPUT 707; “*RST” 20 OUTPUT 707; “:trig:seq:coun:auto on” 25 OUTPUT 707; “:trig:del 0.25” 30 OUTPUT 707; “:scan (@ 1!1!1:1!4!10)” 40 OUTPUT 707; “:INIT” 50 END
Line 5 Opens all channels. Line 25 Sets a 1/4 second delay after each channel
closes.

4.4 Matrix switching examples

Some applications to test thick Þlm resistor networks and transistors are provided in the following para­graphs. These applications are intended to demon­strate the versatility of using the matrix card in test systems.
Line 10 Selects a default conÞguration for the scan. Line 20 Sets channel count to the scan-list-length. Line 30 DeÞnes the scan list. Line 40 Take the Model 7001 out of the idle state. The
scan is conÞgured to start as soon as this command is executed.
When the above program is run, the scan will be com­pleted in approximately 240msec (3msec delay for each relay close and a 3msec delay for each open), which is too fast to view from the front panel. An additional re­lay delay can be added to the program to slow down the scan for viewing. The program is modiÞed by add­ing line 25 to slow down the scan. Also, Line 5 is added to the beginning of the program to ensure that all chan­nels are open before the scan is started.
4.4.1 Thick film resistor network testing
A dedicated matrix system for testing thick Þlm resis­tor networks is shown in Figure 4-4. This particular system provides two different methods to check thick Þlms; four-wire resistance measurements, and voltage measurements using an applied voltage.
The system shown in Figure 4-4 tests two 4-element thick Þlms, but can be expanded to test more by simply using additional Model 7012 matrix cards. The Model 7001 will accommodate two matrix cards. Daisy-chain­in six Model 7001s expands the system to 12 matrix cards allowing 24 four-element thick Þlms to be tested.
4-5
Operation
Measure V or 4-terminal
Source V
Model 196
Model 230
Ohms Sense Volts Ohms
Volts/Ohms HI
Ohms Sense HI
Volts/Ohms LO
Ohms Sense LO Output
Sense Output
Common
Sense Common
TF-1 TF-2
R1R2R3R
1 345 6789102
1
2
Rows
3
4
LHLHLHLHL HLHLHLHLHL
H
4
Cols
7012
R1R
2
4
3
H L
H L
H L
H
L
R
R
Figure 4-4
Thick film resistor network testing
Four-terminal ohms measurements
For general purpose testing, the Keithley Model 196 can be used to make four-terminal resistance measure­ments of each thick Þlm. As shown in Figure 4-5, OHMS HI and OHMS SENSE HI are connected to one matrix row, and OHMS LO and OHMS SENSE LO are connected to another matrix row. With this conÞgura- tion, the resistance of each resistor element and/or combined elements can be measured by closing the ap­propriate crosspoints. In Figure 4-5, crosspoints 1!1 (Row 1, Column 1) and 2!3 (Row 2, Column 3) are closed to measure the combined resistance of R1 and R2.
The effects of thermal EMFs generated by relay con­tacts and connections can be cancelled by using the off­set compensated ohms feature of the Model 196. (The Model 7012 has been designed to keep relay EMF at a minimal level.) To compensate for thermal EMFs, close two crosspoints (such as 1!1!1 and 1!2!1), this will short the input of the Model 196, enable zero to cancel inter­nal offset, and then enable offset compensated ohms.
4-6
Operation
Thick Film
R
R
1
R
2
R
3
4
Model 196
Volts/Ohms HI Ohms Sense HI
Volts/Ohms LO Ohms Sense LO
X = Closed Crosspoints
R
1
HL LH HLHL
196
Rows
R
2
Equivalent Circuit
HL1HL2HL3HL
H
1
X
L
H
2
L
R
R
3
4
HL
X
4
HL
5
Cols
Figure 4-5
Four-terminal ohms measurements
4-7
Operation
Voltage divider checks
For thick Þlm resistor networks that are to be used as voltage dividers, it may be desirable to test them using voltages that simulate actual operating conditions. This is a particularly useful test for resistor networks that have a voltage coefÞcient speciÞcation. The test system in Figure 4-4 uses a Keithley Model 230 to source voltage and the Model 196 to measure voltage.
A consideration in these checks is the effect of the Mod­el 196 input impedance on voltage measurements. The input impedance is shunted across the resistor being measured. The resultant divider resistance is the paral­lel combination of the resistor under test and the input impedance. As long as the input impedance is much larger than the resistor being tested, the error intro­duced into the measurement will be minimal. Mini­mum input impedance requirements are, of course, determined by the accuracy needed in the measure­ment. The input impedances of the Model 196 are as follows: 300mV and 3V ranges, 1G 300V range, 10.1M
. For better input impedance re-
; 30V range, 11M
quirements, the Keithley Model 617 Electrometer can be incorporated into the test system to measure volt­age.
Another factor to be considered when checking low voltage dividers is thermal EMFs generated by the ma­trix card. (The Model 7012 has been designed to keep relay EMF at a minimal level.) A matrix card crosspoint can generate up to 5
µ
V of thermal EMF. Thus, when making low voltage measurements be sure to account for this additional error.
Even though four-terminal connections are made at the Model 196 and the resistor networks, the sense leads are internally disconnected from the input of the DMM when the volts function is selected. The simpliÞed test system is shown in Figure 4-6.
The thick Þlm is tested by applying a voltage across the resistor network and measuring the voltage across each resistor element and/or across combined ele­ments. In Figure 4-6, crosspoints 1!3!1 and 1!4!4 are closed to apply voltage across the network, and cross­points 1!1!3 and 1!2!4 are closed to measure the voltage drop across R3.
;
4-8
Operation
Thick Film
R
1
R
2
R
3
R
4
Model 196
Measure V
Model 230
Source V
Output
Sense Output
Common
Sense Common
HLHLHLHL
1234
HI
H
1
L
LO
H
2
L
Rows
H
3
X
L
H
4
L
X = Closed Crosspoints
R
1
HL
5
Cols
X
X
X
Model 7012
R
2
R
3
R
4
Figure 4-6
Voltage divider checks
H H H H
V
196
+/-
230
Equivalent Circuit
H
4-9
Operation
4.4.2 Transistor testing
A matrix system for testing DC parameters of transis­tors is shown in Figure 4-7. This system uses two Source Measure Units (SMU). There are three SMUs available from Keithley; the Model 236 Source Measure Unit, Model 237 High Voltage Source Measure Unit and Model 238 High Current Source Measure Unit. Keep in mind that if using the Models 237 (high voltage capability) or 238 (high current capability), do not ex­ceed the maximum signal levels of the matrix card. Maximum allowable DC signals are 110V and 1A, 30W with resistive load.
This system tests three transistors, but can be expand­ed to test more by simply using additional Model 7012 matrix cards. The Model 7001 will accommodate two matrix cards. Daisy-chaining six Model 7001s expands the system to 12 matrix cards allowing 36 or more tran­sistors to be tested.
NOTE
The Model 7012 is a general purpose matrix card and cannot be used to check FETs or transistors that have high gain or low power. To test these devices, a matrix card with low offset current and high isolation characteris­tics must be used.
SMU #1
Output HI
Sense HI
Sense LO
Output LO
SMU #2
Output HI
Sense HI
Sense LO
Output LO
Figure 4-7
Transistor testing
123
1
2
Rows
3
4
HL HL HL
456
Columns
HL HL HL
7012
789
HL HL HL
10
HL
H L
H L
H L
H L
4-10
Operation
DC parameter checks
With a transistor conÞgured as a common-emitter am­pliÞer, the test system shown in Figure 4-8 can be used to determine the following DC parameters: Collector current (I rent (I
), base current (I
C
) and base-to-emitter voltage (V
E
), current gain, emitter cur-
B
BE
).
Figure 4-8 shows which crosspoints to close to conÞg- ure the ampliÞer circuit. SMU #1 is conÞgured to source voltage and measure current. It is used to power the collector circuit (V current (I
). SMU #2 is conÞgured to source current
C
) and measure the collector
CE
and measure voltage. It is used to provide the base cur­rent (I base-to-emitter voltage (V (I
) for the transistor, and will also measure the
B
) and base current (I
C
). With collector current
BE
) known, the current gain can
B
be calculated as follows:
I
Gain
C
-----=
I
B
Common-emitter characteristic curves
A proÞle of the transistor operating characteristics can be obtained by measuring the collector current over a speciÞed voltage range (V currents (I
). For example, Figure 4-9 shows the charac-
B
) for different base bias
CE
teristics of a typical NPN silicon transistor at base bias currents (I
) of 20
B
µ
A, 40
µ
A, 60
µ
A, and 80
µ
A.
Extensive trigger capabilities facilitate synchronization of the Keithley Source Measure Unit operations. By performing a subordinate sweep, SMU #1 will perform a staircase sweep at every base bias current level set by SMU #2. On every step of each staircase sweep, SMU #1 will source a voltage level (V subsequent collector current (I
) and measure the
CE
). For the characteris-
C
tics shown in Figure 4-9, four staircase sweeps were performed; one staircase sweep at each base bias level.
Refer to a Keithley Source Measure Unit instruction manual for details on performing sweeps.
The emitter current (I
) can be determined by using
E
Kirchoff’s Current Law as follows:
I
+=
EICIB
4-11
Operation
SMU #1
A
±
Source V
Measure I
I
C
V
BE
I
E
Equivalent Circuit
SMU #1
Output HI
Sense HI
Sense LO Output LO
I
B
SMU #2
A = Measure I ± = Source V V = Measure V
V
= Source I
X
1
2
X
Figure 4-8
DC parameter checks
Source I
Measure V
SMU #2
Output HI
Sense HI
Sense LO Output LO
Rows
3
4
HL HL HL
X = Closed Crosspoint
X
X
7012
4-12
Operation
10
8
6
, ma
c
I
4
2
012345
V , volts
CE
+80 µa
+60 µa
+40µa
+20 µa
I
B
Figure 4-9
Common-emitter characteristics of an NPN silicon Transistor

4.5 Measurement considerations

= 0
sideration; however, it can seriously degrade measurement accuracy when testing high-impedance devices. The voltage measured across such a device, for example, can be substantially attenuated by the volt­age divider action of the device source resistance and path isolation resistance, as shown in Figure 4-11. Also, leakage currents can be generated through these resis­tances by voltage sources in the system.
R
DUT
R
PATH
E
DUT
DUT
= Source Resistance of DUT
R
DUT
E
= Source EMF of DUT
DUT
R
= Path Isolation Resistance
PATH
R
= Input Resistance of Measuring Instrument
IN
Matrix
Card
R
IN
Measure Instrument
V
Many measurements made with the Model 7012 are subject to various effects that can seriously affect low­level measurement accuracy. The following para­graphs discuss these effects and ways to minimize them.
4.5.1 Path isolation
The path isolation is simply the equivalent impedance between any two test paths in a measurement system. Ideally, the path isolation should be inÞnite, but the ac­tual resistance and distributed capacitance of cables and connectors results in less than inÞnite path isola­tion values for these devices.
Path isolation resistance forms a signal path that is in parallel with the equivalent resistance of the DUT, as shown in Figure 4-10. For low-to-medium device resis­tance values, path isolation resistance is seldom a con-
Figure 4-10
Path isolation resistance
R
DUT
E
DUT
DUT
R
PATH
R
PATH
R
+
PATH
E
DUT
=
E
OUT
R
Figure 4-11
Voltage attenuation by path isolation resistance
4-13
Operation
Any differential isolation capacitance affects DC mea­surement settling time as well as AC measurement ac­curacy. Thus, it is often important that such capacitance be kept as low as possible. Although the distributed ca­pacitance of the matrix card is generally Þxed by de­sign, there is one area where you do have control over the capacitance in your system; the connecting cables. To minimize capacitance, keep all cables as short as possible.
4.5.2 Magnetic fields
When a conductor cuts through magnetic lines of force, a very small current is generated. This phenomenon will frequently cause unwanted signals to occur in the test leads of a switching matrix system. If the conduc­tor has sufÞcient length, even weak magnetic Þelds like those of the earth can create sufÞcient signals to affect low-level measurements.
Two ways to reduce these effects are: (1) reduce the lengths of the test leads, and (2) minimize the exposed circuit area. In extreme cases, magnetic shielding may be required. Special metal with high permeability at low ßux densities (such as mu metal) is effective at re­ducing these effects.
Even when the conductor is stationary, magnetically­induced signals may still be a problem. Fields can be produced by various signals such as the AC power line voltage. Large inductors such as power transformers can generate substantial magnetic Þelds, so care must be taken to keep the switching and measuring circuits a good distance away from these potential noise sourc­es.
At high current levels, even a single conductor can gen­erate signiÞcant Þelds. These effects can be minimized by using twisted pairs, which will cancel out most of the resulting Þelds.
4.5.3 Radio frequency interference
RFI (Radio Frequency Interference) is a general term used to describe electromagnetic interference over a wide range of frequencies across the spectrum. Such RFI can be particularly troublesome at low signal lev­els, but is can also affect measurements at high levels if the problem is of sufÞcient severity.
RFI can be caused by steady-state sources such as radio or TV signals, or some types of electronic equipment (microprocessors, high speed digital circuits, etc.), or it can result from impulse sources, as in the case of arcing in high-voltage environments. In either case, the effect on the measurement can be considerable if enough of the unwanted signal is present.
RFI can be minimized in several ways. The most obvi­ous method is to keep the equipment and signal leads as far away from the RFI source as possible. Shielding the matrix switching card, signal leads, sources, and measuring instruments will often reduce RFI to an ac­ceptable level. In extreme cases, a specially-construct­ed screen room may be required to sufÞciently attenuate the troublesome signal.
Many instruments incorporate internal Þltering that may help to reduce RFI effects in some situations. In some cases, additional external Þltering may also be re­quired. Keep in mind, however, that Þltering may have detrimental effects on the desired signal.
4.5.4 Ground loops
When two or more instruments are connected together, care must be taken to avoid unwanted signals caused by ground loops. Ground loops usually occur when sensitive instrumentation is connected to other instru­mentation with more than one signal return path such as power line ground. As shown in Figure 4-12, the re­sulting ground loop causes current to ßow through the instrument LO signal leads and then back through power line ground. This circulating current develops a small but undesirable voltage between the LO termi­nals of the two instruments. This voltage will be added to the source voltage, affecting the accuracy of the mea­surement.
4-14
Operation
Signal Leads
Instrument 1 Instrument 2 Instrument 3
Ground Loop
Current
Power Line Ground
Figure 4-12
Power line ground loops
Figure 4-13 shows how to connect several instruments together to eliminate this type of ground loop problem. Here, only one instrument is connected to power line ground.
Instrument 1 Instrument 2 Instrument 3
manner. When in doubt, consult the manual for all in­strumentation in the test setup.
4.5.5 Keeping connectors clean
As is the case with any high-resistance device, the in­tegrity of connectors can be damaged if they are not handled properly. If connector insulation becomes con­taminated, the insulation resistance will be substantial­ly reduced, affecting high-impedance measurement paths.
Oils and salts from the skin can contaminate connector insulators, reducing their resistance. Also, contami­nants present in the air can be deposited on the insula­tor surface. To avoid these problems, never touch the connector insulating material. In addition, the multi­plexer card should be used only in clean, dry environ­ments to avoid contamination.
Power Line Ground
Figure 4-13
Eliminating ground loops
Ground loops are not normally a problem with instru­ments having isolated LO terminals. However, all in­struments in the test setup may not be designed in this
If the connector insulators should become contaminat­ed, either by inadvertent touching, or from air-borne deposits, they can be cleaned with a cotton swab dipped in clean methanol. After thoroughly cleaning, they should be allowed to dry for several hours in a low-humidity environment before use, or they can be dried more quickly using dry nitrogen.
4.5.6 AC frequency response
The AC frequency response of the Model 7012 is im­portant in test systems that switch AC signals. Refer to the speciÞcations at the front of this manual.
4-15
Operation
4-16
5

Service Information

WARNING
The information in this section is in­tended only for qualiÞed service per­sonnel. Some of the procedures may expose you to hazardous voltages that could result in personal injury or death. Do not attempt to perform these procedures unless you are qualiÞed to do so.

5.1 Introduction

This section contains information necessary to service the Model 7012 matrix card and is arranged as follows:
5.2 Handling and cleaning precautions: Discusses
handling procedures and cleaning methods for the matrix card.
5.3 Performance veriÞcation: Covers the procedures
necessary to determine if the card is operating properly.
5.4 Special handling of static-sensitive devices:
Reviews precautions necessary when handling static-sensitive devices.

5.2 Handling and cleaning precautions

Because of the high impedance circuits on the Model 7012, care should be taken when handling or servicing the card to prevent possible contamination, which could degrade performance. The following precau­tions should be taken when handling the matrix card.
Do not store or operate the card in an environment where dust could settle on the circuit board. Use dry ni­trogen gas to clean dust off the card if necessary.
Handle the card only by the side edges. Do not touch any board surfaces, components, or connectors. Do not touch areas adjacent to electrical contacts. When servic­ing the card, wear clean cotton gloves.
If making solder repairs on the circuit board, use an OA-based (organic activated) ßux. Remove the ßux from these areas when the repair is complete. Use pure water along with plenty of clean cotton swabs to re­move the ßux. Take care not to spread the ßux to other areas of the circuit board. Once the ßux has been re­moved, swab only the repaired area with methanol, then blow dry the board with dry nitrogen gas.
5.5 Principles of operation: Brießy discusses circuit
operation.
5.6 Troubleshooting: Presents some troubleshooting
tips for the matrix card.
After cleaning, the card should be placed in a 50 humidity environment for several hours.
C low
°
5-1
Service Information
5.3 Performance verification
The following paragraphs discuss performance veriÞ- cation procedures for the Model 7012, including path resistance, offset current, contact potential, and isola­tion.
With the Model 7012’s backplane jumpers installed, the performance veriÞcation procedures must be per­formed with only one matrix card (the one being checked) installed in the Model 7001 mainframe. These conditions do not apply if the backplane jumpers are removed.
CAUTION
Contamination will degrade the per­formance of the matrix card. To avoid contamination, always grasp the card by the side edges. Do not touch the connectors, and do not touch the board surfaces or components. On plugs and receptacles, do not touch areas adjacent to the electrical con­tacts.
NOTE
Failure of any performance veriÞca- tion test may indicate that the matrix card is contaminated. See paragraph
5.2 to clean the card.
5.3.1 Environmental conditions
All veriÞcation measurements should be made at an ambient temperature between 18 relative humidity of less than 70%.
°
and 28
°
C, and at a
5.3.2 Recommended equipment
Table 5-1 summarizes the equipment necessary for per­formance veriÞcation, along with an application for each unit.
Table 5-1
Verification equipment
Description Model or part SpeciÞcations Applications
DMM Keithley Model 196 300
Electrometer w/voltage source Keithley Model 617 20pA; 1.6%
Sensitive DVM Keithley Model 182 3mV; 0.015% Contact potential
Triax cable (unterminated) Keithley Model 7025
Low thermal cable (unterminated)
Keithley Model 1484
; 0.01% Path resistance
Offset current, isolation 20nA, 200nA; 0.25% 100V source; 0.2%
 
Offset current
Contact potential
5-2
Service Information
5.3.3 Matrix card connections
The following information summarizes methods that can be used to connect test instrumentation to the two connector cards. Detailed connection information is provided in Section 3.
Model 7012-S
wired directly to the screw terminals of the connector card. Jumper wires should be kept as short as possible.
Model 7012-C
nections to the matrix card is by hard-wiring a 96-pin female DIN connector and then mating it to the con­nector on the Model 7012-C. Row and column shorting connections can also be done at the connector. The con­nector in the Model 7011-KIT-R connection kit (see Ta­ble 3-1) can be used for this purpose. Pin identiÞcation for the multi-pin connector for the matrix card is pro­vided by Figure 3-5.
Instrumentation can simply be hard-
One method to make instrument con-
CAUTION
After making solder connections to a connector, remove solder ßux as ex­plained in paragraph 5.2. Failure to clean the solder connections could result in degraded performance pre­venting the card from passing veriÞ- cation tests.
Before pre-wiring any connectors plugs, study the fol­lowing test procedures to fully understand the connec­tion requirements.
5.3.4 Channel resistance tests
Referring to Figure 5-1, perform the following steps to verify that each contact of every relay is closing prop­erly and that the resistance is within speciÞcation.
Model 196
(Measure 4-Wire Ohms)
Ohms Sense LO
Note : Set up shown is configured to test the high (H) terminal of row 1 through crosspoints 1!1 through 1!12.
Figure 5-1
Path resistance testing
Ohms Sense HI
Ohms HI Ohms LO
1 34567 89
1
2
Rows
3
4
Jumpers
Columns
2
Model 7012
10
H L
H L
H L
H L
5-3
Service Information
1. Turn the Model 7001 off if it is on.
2. As shown in Figure 5-1, connect all terminals of matrix columns 1-10 together to form one common terminal.
3. Set the Model 196 to the 300 four test leads to the OHMS and OHMS SENSE in­put.
4. Short the four test leads together and zero the Model 196. Leave zero enabled for the entire test.
5. Connect OHMS HI and OHMS SENSE HI of the Model 196 to the common terminal. It is recom­mended that the physical connections be made at columns 1 and 10 as shown in the illustration.
6. Connect OHMS LO and OHMS SENSE LO to the high (H) terminal of Row 1.
7. Install the Model 7012 in slot 1 (CARD 1) of the Model 7001.
8. Turn on the Model 7001 and program it to close Channel 1!1!1 (Slot 1, Row 1, Column 1). Verify that the resistance of this channel is <1
9. Open Channel 1!1!1 and close 1!1!2. Verify that the resistance of this channel is <1
10. Open Channel 1!1!2 and close 1!1!3. Verify that the resistance of this channel is <1
11. Repeat the basic procedure of opening and closing channels to check the resistance of Row 1 high (H) terminal paths for Columns 4 through 10 (Chan­nels 1!1!4 through 1!1!10).
12. Turn off the Model 7001 and connect the OHMS LO and OHMS SENSE LO test leads of the Model 196 DMM to the low (L) terminal of Row 1.
13. Repeat steps 8 through 11 to check the low (L) channel paths of Row 1.
14. Turn off the Model 7001 and repeat the basic proce­dure in steps 7 through 13 for Rows 2, 3 and 4.
range and connect
.
.
.
5.3.5 Offset current tests
These tests check leakage current from high (H) to low (L) (differential), and from high (H) and low (L) to
chassis (common mode) for each pathway. In general, these tests are performed by simply measuring the leakage current with an electrometer. In the following procedure, the Model 617 is used to measure leakage current.
Referring to Figure 5-2, perform the following proce­dure to check offset current:
1. Turn the Model 7001 off if it is on.
2. Connect the Model 617 electrometer to Row 1 of the matrix card as shown in Figure 5-2. Note that electrometer HI is connected to both high (H) and low (L) of Row 1. Electrometer LO is connected to chassis ground, which is accessible at the rear pan­el of the mainframe.
3. Install the matrix card in slot 1 (CARD 1) of the Model 7001.
4. On the Model 617, select the 200pA range, and en­able zero check and zero correct in that order. Leave zero correct enabled for the entire proce­dure.
5. Turn on the Model 7001
6. Program the Model 7001 to close Channel 1!1!1.
7. On the Model 617, disable zero check and verify that it is <100pA. This measurement is the leakage current of the pathway.
8. On the Model 617, enable zero check and on the Model 7001, open Channel 1!1!1.
9. Repeat the basic procedure in steps 6 through 8 to check the rest of the pathways (Channels 1!1!2 through 1!1!10) of the row.
10. Turn off the Model 7001 and connect the Model 617 to Row 2. Repeat the basic procedure in steps 6 through 9 to check Channels 1!2!1 through 1!2!10.
11. Repeat the basic procedure in step 10 to check Rows 3 and 4.
12. Turn off the Model 7001.
13. To check differential offset current, connect the Model 617 to Row 1 as shown in Figure 5-3, and re­peat steps 4 through 12.
5-4
Model 7025 Unterminated Triax Cable
Service Information
INPUT
MODEL 617
(Measure Current)
Note : Setup shown is configured to test Row 1 pathways for offset current.
Figure 5-2
Common-mode offset current testing
Model 7025 Unterminated Triax Cable
INPUT
MODEL 617
(Measure Current)
Note : Setup shown is configured to test Row 1 pathways for offset current.
HI
LO
LO
1
2
Rows
3
4
HL HL HL HL HL HL HL HL HL HL
Chassis ground is accessible at rear panel of the 7001.
HI
1
2
Rows
3
4
HL HL HL HL HL HL HL HL HL HL
1 3456789102
Columns
H
L
H
L
H
L
H
L
Model 7012
Columns
1 3456789102
H L
H L
H L
H L
Figure 5-3
Differential offset current testing
Model 7012
5-5
Service Information
5.3.6 Contact potential tests
These tests check the EMF generated by each relay con­tact pair (H and L) for each pathway. The tests simply consist of using a sensitive digital voltmeter (Model
182) to measure the contact potential.
Perform the following procedure to check contact po­tential of each path:
1. Turn the Model 7001 off if it is on.
2. Place a short between HI to LO on each input col­umn 1-10.
3. Connect all row HI together on the common bus.
4. Connect all row LO together on the common bus.
5. Place a short between HI to LO on the rows.
6. Connect the Model 182 input leads to HI and LO of the rows.
7. Install the Model 7012 in the Model 7001 slot 1 and turn the Model 7001 on.
8. Allow the Models 7012, 7001, and 182 to warm up for two hours.
9. Select the 3mV range on the Model 182.
10. Press REL READING on the Model 182 to null out internal offsets. Leave REL READING enabled for the entire procedure.
11. Turn the Model 7001 off. Remove the Model 7012 front slot 1. Cut the short from HI to LO on the rows.
12. Install the Model 7012 in the Model 7001 slot 1 and turn power on.
13. Wait 15 minutes.
14. Program the Model 7001 to close Channel 1!1!1.
15. After settling, verify that the reading on the Model 182 is <500mV (7012-S). This measurement repre­sents the contact potential of the pathway.
16. From the Model 7001, open Channel 1!1!1.
17. Repeat steps 14 through 16 for all 40 crosspoints.
Model 1484 Low Thermal Cable (Unterminated)
KEITHLEY
182 SENSITIVE DIGITAL VOLTMETER
Model 182
Note : Setup shown is configured to test Row 1 crosspoints for contact potential.
TRG SRQ REM TALK LSTN
Figure 5-4
Contact potential testing
Low Thermal short­clean high purity copper (1 of 10)
Columns
HI LO
1 3456789102
1
2
Rows
3
4
HL HL HL HL HL HL HL HL HL HL
Model 7012
H L
H L
H L
H L
5-6
Service Information
5.3.7 Path isolation tests
These tests check the leakage resistance (isolation) be­tween adjacent paths. A path is deÞned as the high (H) and low (L) circuit from a row to a column that results by closing a particular crosspoint. In general, the test is performed by applying a voltage (+100V) across two adjacent paths and then measuring the leakage current across the paths. The isolation resistance is then calcu­lated as R = V/I. In the following procedure, the Model 617 functions as both a voltage source and an ammeter. In the V/I function, the Model 617 internally calculates the resistance from the known voltage and current lev­els, and displays the resistance value.
1. Turn the Model 7001 off if it is on.
2. Jumper the high (H) terminal to the low (L) termi­nal for each row (see Figure 5-5).
3. Connect the Model 617 to Rows 1 and 2 as shown in Figure 5-5. Make sure the voltage source is in standby. Also, make sure there are no other con­nections to the card.
4. Install the Model 7012 in slot 1 of the Model 7001.
WARNING
The following steps use high voltage (100V). Be sure to remove power from the circuit before making con­nection changes.
5. On the Model 617, select the 2pA range, and enable zero check and zero correct in that order. Leave zero correct enabled for the entire procedure.
6. On the Model 617, select the 20pA range and re­lease zero check.
7. On the Model 617, press suppress to cancel offset current and then enable zero check.
8. On the Model 617, set the voltage source for +100V and select the 200nA current range. Make sure the voltage source is in standby.
9. Place the Model 617 in the V/I measurement func­tion by pressing SHIFT OHMS.
10. Turn on the Model 7001 and program it to close Channels 1!1!1 (Row 1, Column 1) and 1!2!2 (Row 2, Column 2).
11. On the Model 617, disable zero check and press OPERATE to source +100V.
12. After allowing the reading on the Model 617 to set­tle, verify that it is >1G leakage resistance (isolation) between Row 1, Col­umn 1 and Row 2, Column 2.
13. Place the Model 617 in standby and enable zero check.
14. Turn off the Model 7001.
15. Disconnect the Model 617 from Rows 1 and 2, and in a similar manner, reconnect it to Rows 2 and 3 (picoammeter high to Row 2 and voltage source high to Row 3).
16. Turn on the Model 7001 and program it to close Channels 1!2!2 and 1!3!3.
17. On the Model 617, disable zero check and press OPERATE to source +100V.
18. After allowing the reading on the Model 617 to set­tle, verify that it is >1G
19. Using Table 5-2 as a guide, repeat the basic proce­dure in steps 13 through 18 for the rest of the path pairs (starting with test #3).
. This measurement is the
.
5-7
Service Information
Banana to Banana Cable Ground Link
Removed
INPUT
Source V and Measure V/I
Note : Setup shown is configured to test isolation between row 1 column 1, and row 2 column 2.
Model 617
Unterminated Banana Cables
Model 7025 Unterminated Triax Cable
HI
(Red)
HI
Jumpers
(1 of 4)
Columns
1 3456789102
1
2
Rows
3
4
HL HL HL HL HL HL HL HL HL HL
Model 7012
H L
H L
H L
H L
Figure 5-5
Path isolation testing (guarded)
Table 5-2
Path isolation tests
Test
no. Path isolation Test equipment locations Channels closed
1 Row 1, Col 1 to Row 2, Col 2 Row 1 and Row 2 1!1!1 and 1!2!2
2 Row 2, Col 2 to Row 3, Col 3 Row 2 and Row 3 1!2!2 and 1!3!3
3 Row 3, Col 3 to Row 4, Col 4 Row 3 and Row 4 1!3!3 and 1!4!4
4 Row 3, Col 4 to Row 4, Col 5 Row 3 and Row 4 1!3!4 and 1!4!5
5 Row 3, Col 5 to Row 4, Col 6 Row 3 and Row 4 1!3!5 and 1!4!6
6 Row 3, Col 6 to Row 4, Col 7 Row 3 and Row 4 1!3!6 and 1!4!7
7 Row 3, Col 7 to Row 4, Col 8 Row 3 and Row 4 1!3!7 and 1!4!8
8 Row 3, Col 8 to Row 4, Col 9 Row 3 and Row 4 1!3!8 and 1!4!9
9 Row 3, Col 9 to Row 4, Col 10 Row 3 and Row 4 1!3!9 and 1!4!10
5-8
Service Information
5.3.8 Differential and common-mode isola­tion tests
These tests check the leakage resistance (isolation) be­tween high (H) and low (L) (differential), and from high and low to chassis (common-mode) of every row and column. In general, the test is performed by apply­ing a voltage (100V) across the terminals and then mea­suring the leakage current. The isolation resistance is then calculated as R = V/I. In the following procedure, the Model 617 functions as a voltage source and an am­meter. In the V/I function, the Model 617 internally cal­culates the resistance from the known voltage and current levels, and displays the resistance value.
1. Turn the Model 7001 off if it is on.
2. Connect the Model 617 to Row 1 as shown in Fig­ure 5-6. Make sure the voltage source is in standby. Also, make sure there are no other connections to the card.
3. Install the Model 7012 in slot 1 of the Model 7001.
WARNING
The following steps use high voltage (100V). Be sure to remove power from the circuit before making con­nection changes.
4. On the Model 617, select the 2pA range, and enable zero check and zero correct in that order. Leave zero correct enabled for the entire procedure.
5. On the Model 617, set the voltage source for +100V, and select the 200nA current range. Make sure the voltage source is still in standby.
6. Place the Model 617 in the V/I measurement func­tion by pressing SHIFT OHMS.
7. Turn on the Model 7001, but do not program any channels to close. All channel crosspoints must be open.
8. On the Model 617, disable zero check and press OPERATE to source 100V.
9. After allowing the reading on the Model 617 to set­tle, verify that it is >1G leakage resistance (isolation) of Row 1.
10. Place the Model 617 in standby and enable zero check.
11. Program the Model 7001 to close Channel 1!1!1.
12. On the Model 617, disable zero check and press OPERATE to source +100V.
13. After allowing the reading on the Model 617 to set­tle, verify that it is also >1G checks the isolation of Column 1.
14. Using Table 5-3 as a guide, repeat the basic proce­dure in steps 10 through 13 for the rest of the col­umns and rows (test numbers 3 through 14 of the table).
15. Place the Model 617 in standby and turn the Model 7001 off.
16. For each matrix row, jumper the high (H) terminal to the low (L) terminal as shown in Figure 5-7.
17. Connect the Model 617 to Row 1 as shown in Fig­ure 5-7, and repeat steps 7 through 15 to check common-mode isolation.
. This measurement is the
. This measurement
5-9
Service Information
Banana to Banana Cable
Ground Link Removed
INPUT
Source V and Measure V/I
Model 617
Unterminated Banana Cable
Figure 5-6
Differential isolation testing
Model 7025 Unterminated Triax Cable
HI
(Red)
HI
Columns
1 3456789102
1
2
Rows
3
4
HL HL HL HL HL HL HL HL HL HL
Model 7012
H L
H L
H L
H L
Banana to Banana Cable
Ground Link Removed
INPUT
Source V and Measure V/I
Model 617
Unterminated Banana Cable
Figure 5-7
Common-mode isolation testing
(Red)
Jumpers (1 of 4)
Model 7025 Unterminated Triax Cable
HI
HI
Rows
Chassis Ground is accessible at 7001 rear panel
Columns
1 3456789102
1
2
3
4
HL HL HL HL HL HL HL HL HL HL
Model 7012
H L
H L
H L
H L
5-10
Service Information
Table 5-3
Differential and common-mode isolation testing
Differential or
Test
no.
1 Row 1 None
2 Column 1 1!1!1
3 Column 2 1!1!2
4 Column 3 1!1!3
5 Column 4 1!1!4
6 Column 5 1!1!5
7 Column 6 1!1!6
8 Column 7 1!1!7
9 Column 8 1!1!8
10 Column 9 1!1!9
11 Column 10 1!1!10
12 Row 2 1!1!1 and 1!2!1
13 Row 3 1!1!1 and 1!3!1
14 Row 4 1!1!1 and 1!4!1
common-mode
test Channels closed
5.4 Special handling of static-sensitive
devices
CMOS and other high-impedance devices are subject to possible static discharge damage because of the high-impedance levels involved. The following pre­cautions pertain speciÞcally to static-sensitive devices. However, since many devices in the Model 7012 are static-sensitive, it is recommended that they all be treated as static-sensitive.
1. Such devices should be transported and handled only in containers specially designed to prevent or dissipate static build-up. Typically, these devices will be received in anti-static containers made of plastic or foam. Keep these parts in their original containers until ready for installation.
2. Remove the devices from their protective contain­ers only at a properly grounded work station. Also, ground yourself with a suitable wrist strap while working with these devices.
3. Handle the devices only by the body; do not touch the pins.
4. Any printed circuit board into which the device is to be inserted must Þrst be grounded to the bench or table.
5. Use only anti-static type de-soldering tools and grounded-tip soldering irons.
5-11
Service Information

5.5 Principles of operation

The following paragraphs discuss the basic operating principles for the Model 7012, and can be used as an aid in troubleshooting the matrix card. The schematic drawing of the matrix card is shown on drawing num­ber 7012-106, located at the end of Section 6.
CLK
Relay
Drivers
U100­U104
To Mainframe
Data Strobe Enable
5.5.1 Block diagram
Figure 5-8 shows a simpliÞed block diagram of the Model 7012. Key elements include the relay drivers and relays, as well as the ROM, which contains card ID and conÞguration information. These various elements are discussed in the following paragraphs.
Relays
User connections
+3.5V (Steady State) +5.7 (100 msec during
relay actuation)
To Mainframe
Figure 5-8
Model 7012 block diagram
ID CLK
ID DATA
+6V, +14.6V
ROM
U105
Relay
Power
Control
Q100, Q101 U106, U107
5-12
Service Information
5.5.2 ID data circuits
Upon power-up, card identiÞcation information from each card is read by the mainframe. This ID data in­cludes such information as card ID, hardware settling time, and relay conÞguration information.
ID data is contained within an on-card EEPROM (U105). In order to read this information, the sequence described below is performed on power-up.
1. The IDDATA line (pin 6 of U105) is set from high to low while the IDCLK line (pin 5 of U105) is held high. This action initiates a start command to the ROM to transmit data serially to the mainframe (Figure 5-9).
2. The mainframe sends the ROM address location to be read over the IDDATA line. The ROM then transmits an acknowledge signal back to the main­frame, and it then transmits data at that location back to the mainframe (Figure 5-10).
3. The mainframe then transmits an acknowledge signal, indicating that it requires more data. The ROM will then sequentially transmit data after each acknowledge signal it receives.
4. Once all data is received, the mainframe sends a stop command, which is a low-to-high transition of the IDDATA line with the IDCLK line held high (see Figure 5-9).
5.5.3 Relay control
Card relays are controlled by serial data transmitted via the relay DATA line. A total of Þve bytes for each card are shifted in serial fashion into latches located in the card relay driver ICs. The serial data is clocked in by the CLK line. As data overßows one register, it is fed out the Q’S line of the register down the chain.
Once all Þve bytes have shifted into the card, the STROBE line is set high to latch the relay information into the Q outputs of the relay drivers, and the appro­priate relays are energized (assuming the driver out­puts are enabled, as discussed below). Note that a relay driver output goes low to energize the corresponding relay.
ID CLK
ID DATA
Figure 5-9
Start and stop sequences
Start Bit Stop Bit
5-13
Service Information
ID CLK
IDDATA
(Data output from mainframe or ROM)
IDDATA
(Data output from mainframe or ROM)
Start
Figure 5-10
Transmit and acknowledge sequence
189
Acknowledge
5.5.4 Relay power control
A relay power control circuit, made up of U106, U107, Q100, Q101, and associated components, keeps power dissipated in relay coils at a minimum, thus reducing possible problems caused by thermal EMFs.
During steady-state operation, the relay supply volt­age, +V, is regulated to +3.5V to minimize coil power dissipation. When a relay is Þrst closed, the STROBE pulse applied to U106 changes the parameters of the re­lay supply voltage regulator, Q100, allowing the relay supply voltage, +V, to rise to +5.7V for about 100msec. This brief voltage rise ensures that relays close as quickly as possible. After the 100msec period has elapsed, the relay supply voltage (+V) drops back down to its nominal steady-state value of +3.5V.
5.5.5 Power-on safeguard
NOTE
The power-on safeguard circuit dis­cussed below is actually located on the digital board in the Model 7001 main­frame.
A power-on safeguard circuit, made up of U114 (a D­type ßip-ßop) and associated components, ensures that relays do not randomly energize on power-up and power-down. This circuit disables all relays (all relays are open) during power-up and power-down periods.
The PRESET line on the D-type ßip-ßop is controlled by the 68302 microprocessor, while the CLK line of the D-type ßip-ßop is controlled by a VIA port line on the 68302 processor. The Q output of the ßip-ßop drives each switch card relay driver IC enable pin (U100­U104, pin 8).
When the 68302 microprocessor is in the reset mode, the ßip-ßop PRESET line is held low, and Q out imme­diately goes high, disabling all relays (relay driver IC enable pins are high, disabling the relays.) After the re­set condition elapses (
200msec), PRESET goes high while Q out stays high. When the Þrst valid STROBE pulse occurs, a low logic level is clocked into the D­type ßip-ßop, setting Q out low and enabling all relay drivers simultaneously. Note that Q out stays low, (en­abling relay drivers) until the 68302 processor goes into a reset condition.
5-14
Service Information

5.6 Troubleshooting

5.6.1 Troubleshooting equipment
Table 5-4 summarizes recommended equipment for troubleshooting the Model 7012.
Table 5-4
Recommended Troubleshooting Equipment
Manufacturer
Description
Multimeter Keithley 196 Measure DC voltages
Oscilloscope TEK 2243 View logic waveforms
5.6.2 Troubleshooting access
In order to gain access to the relay card top surface to measure voltages under actual operation conditions, perform the following steps:
and model Application
WARNING
Lethal voltages are present within the Model 7001 mainframe. Some of the procedures may expose you to hazardous voltages. Observe stan­dard safety precautions for dealing with live circuits. Failure to do so could result in personal injury or death.
CAUTION
Observe the following precautions when troubleshooting or repairing the switch card:
To avoid contamination, which could degrade card performance, always handle the card only by the handle and side edges. Do not touch edge connectors, board surfaces, or com­ponents on the card. Also, do not touch areas adjacent to electrical con­tacts on connectors.
1. Disconnect the connector card from the relay card.
2. Remove the Model 7001 cover.
3. Install the relay card in the CARD 1 slot location.
4. Turn on Model 7001 power to measure voltages (see following paragraph).
5.6.3 Troubleshooting procedure
Table 5-5 summarizes switch card troubleshooting.
Use care when removing relays from the PC board to avoid pulling traces away from the circuit board. Before attempting to remove a relay, use an appropriate de-soldering tool, such as a solder sucker, to clear each mounting hole completely free of solder. Each relay pin must be free to move in its mounting hole before re­moval. Also, make certain that no burrs are present on the ends of the relay pins.
5-15
Service Information
Table 5-5
Troubleshooting procedure
Step Item/Component Required Condition Comments
1 GND pad All voltages referenced to digital ground
(GND pad).
2 +6V pad +6VDC Relay voltage.
3 +5V pad +5VDC Logic voltage.
4 +14.6V pad +14.6VDC Relay bias voltage.
5 +V pad +3.5VDC* Regulated relay voltage.
6 U105, pin 5 ID CLK pulses During power-up only.
7 U105, pin 6 ID DATA pulses During power-up only.
8 U100, pin 7 STROBE pulse End of relay update sequence.
9 U100, pin 2 CLK pulses During relay update sequence only.
10 U100, pin 3 DATA pulses During relay update sequence only.
11 U100-U104, pins 10-18 Low with relay energized;
Relay driver outputs.
high with relay de-energized.
*+3.5VDC present at +V pad under steady-state conditions. This voltage rises to +5.7VDC for about 100msec when relay conÞguration is changed.
5-16
6

Replaceable Parts

6.1 Introduction

This section contains replacement parts information, schematic diagrams, and component layout drawings for the Model 7012-S and 7012-C.

6.2 Parts lists

Parts lists for the various circuit boards are included in tables integrated with schematic diagrams and compo­nent layout drawings for the boards. Parts are listed al­phabetically in order of circuit designation.

6.3 Ordering information

To place an order, or to obtain information concerning replacement parts, contact your Keithley representa­tive or the factory (see inside front cover for addresses).
When ordering parts, be sure to include the following information:
1. Card model number (7012-S or 7012-C)
2. Card serial number
3. Part description
4. Circuit description, if applicable
5. Keithley part number

6.4 Factory service

If the card is to be returned to Keithley Instruments for repair, perform the following:
1. Complete the service form at the back of this man­ual and include it with the card.
2. Carefully pack the card in the original packing car­ton.
3. Write ATTENTION REPAIR DEPT on the shipping label.
Note: It is not necessary to return the mainframe with the card.
6-1
Replaceable Parts

6.5 Component layouts and schematic diagrams

Component layout drawings and schematic diagrams are included on the following pages integrated with the parts lists:
Table 1 Parts List, Relay Card for 7012-S and 7012-C. 7011-100 Component Layout, Relay Card for 7012-S
and 7012-C.
7011-106 Schematic, Relay Card for 7012-S and 7012-C.
NOTE
The Model 7011 and 7012 use the same relay card, only the connector cards are different.
Table 2 Parts List, Screw Terminated Connector
Card for 7012-S.
7012-160 Component Layout, Screw Terminated Con-
nector Card for 7012-S.
7012-166 Schematic, Screw Terminated Connector
Card for 7012-S.
Table 3 Parts List, Mass Terminated Connector Card
for 7012-C.
7012-170 Component Layout, Mass Terminated Con-
nector Card for 7012-C.
7012-176 Schematic, Mass Terminated Connector
Card for 7012-C.
6-2
Replaceable Parts
Table 1. Relay Board for Model 7012-S and 7012-C, Parts List
Circuit Desig. Description Keithley Part No.
EJECTOR ARM ROLL PIN (FOR EJECTOR ARMS) SHIELD SOCKET (FOR U105) 2-56X1/4 PHILLIPS PAN HD (FOR SCANNER SHIELD) 2-56X3/8 PHILLIPS PAN HEAD (P2001 TO STANDOFF) 4-40X1/4 PHILLIPS PAN HD SEMS SCREW
7011-301 DP-6-1 7011-305 SO-72 2-56X1/4PPH 2-56X3/8PPH 4-40X1/4PPHSEM
(SCANNER BOARD TO TERMINAL BOARD)
C100-105,107-109,118,119 C106 C110,111 C112 C113,114 C115-117
4-40X3/16 PHIL. PAN HD SEMS (FOR Q100)
CAP,.1UF,20%,50V,CERAMIC CAP,.1UF,20%,50V,CERAMIC CAP,1UF,20%,50V, CERAMIC CAP, 0.001uF, 20%, 500V, CERAMIC CAP, 10UF,-20+100%,25V,ALUM ELEC CAP,150PF,10%,1000V,CERAMIC
4-40X3/16PPHSEM
C-365-.1 C-365-.1 C-237-1 C-22-.001 C-314-10 C-64-150P
J1002,1003 CONNECTOR, MALE CS-736-2
K100-139 RELAY, ULTRA-SMALL POLARIZED TF2E-5V RL-149
P2001 CONNECTOR, RIGHT ANGLE MALE CS-775-1
Q100 Q101
R100 R101 R102,103 R104 R105 R106
TRANS, NPN PWR, TIP31, (TO-220AB) TRANS,N CHAN MOSPOW FET,V11713 (TO-92)
RES, 2.49K, 1%, 1/8W, METAL FILM RES, 1.15K, 1%, 1/8W, METAL FILM RES, 560, 10%, 1/2W, COMPOSITION RES,1K,1%,1/8W,METAL FILM RES,220K,5%,1/4W,COMPOSITION OR FILM RES,10K,5%,1/4W,COMPOSITION OR FILM
TG-253 TG-195
R-88-2.49K R-88-1.15K R-1-560 R-88-1K R-76-220K R-76-10K
U100-104 U105 U106 U107
IC, 8-BIT SERIAL-IN LATCH DRIVER, 5841A EPROM PROGRAM IC,RETRIG MONO MULTIVIB,74HC123 IC,AJD SHUNT REGULATOR,TL431CLP
IC-536 7012-800-*** IC-492 IC-677
W100-107 JUMPER J-15
*** Order current Þrmware revision level.
6-3
E2
DELETED SO-72.
CHG’D U105 FROM IC-737 TO TC17-100.
25917
KK
6/18/01
TC17-100 BOARD ASS’Y. ORIENT ARROW TOWARDS PIN 1 OF DEVICE.
25917
2
TC17-100
25917
2
Replaceable Parts
Table 2. Screw Terminal Board for Model 7012-S, Parts List
Circuit Desig. Description Keithley Part No.
CABLE CLAMP CAPTIVE SCREW (FOR TOP CLAMP) CONNECTOR SHIM (FOR P1002, 1003) SHIELD STRIP, POLYURETHANE (FOR BOTTOM CLAMP) TOP CLAMP 2-56X7/16 PHILLIPS PAN HEAD
7011-304-2 FA-243-1 7011-309 7011-305 2001-345-1 7011-302 2-56X7/16PPH
(FOR TERMINAL SHIELD BOARD)
J1004,1006 J1005,1007
CONN, 6 PIN CONN, 8 PIN
TE-115-6 TE-115-8
P1002,1003 CONNECTOR, FEMALE CS-748-3
6-5
Replaceable Parts
Table 3. Mass Terminated Connector Card for 7012-C, Parts List
Circuit Desig. Description Keithley Part No.
BRACKET CONNECTOR SHIM SHIELD STANDOFF
7011-307 7011-309 7011-311 ST-203-1
C101 CAP,.1UF,20%,50V,CERAMIC C-365-.1
CR101-104 DIODE,SILICON,IN4148 (DO-35) RF-28
E101,102 BEAD, FERRITE CT-9
J1004 CONN, 96-PIN, 3 ROWS CS-514
K101-104 RELAY, ULTRA-SMALL POLARIZED TF2E-4.5V RL-162
P1002,1003 CONNECTOR, FEMALE CS-748-3
Q101 TRANS,NPN SILICON,BUK456-1000B TG-47
R101,102 R103
RES, 220,10%, 1/2W, COMPOSITION RES,100K,5%,1/4W,COMPOSITION OR FILM
R-1-220 R-76-100K
6-7
Service Form
Model No. Serial No. Date
Name and Telephone No.
Company
List all control settings, describe problem and check boxes that apply to problem.
Intermittent
❏ ❏
IEEE failure Front panel operational
Display or output (check one)
Drifts
Unstable
❏ ❏
Overload
Calibration only
Data required
(attach any additional sheets as necessary)
Show a block diagram of your measurement system including all instruments connected (whether power is turned on or not). Also, describe signal source.
Analog output follows display
❏ ❏
Obvious problem on power-up All ranges or functions are bad
Unable to zero
Will not read applied input
CertiÞcate of calibration required
Particular range or function bad; specify
❏ ❏
Batteries and fuses are OK Checked all cables
Where is the measurement being performed? (factory, controlled laboratory, out-of-doors, etc.)
What power line voltage is used? Ambient temperature? ¡F
Relative humidity? Other?
Any additional information. (If special modiÞcations have been made by the user, please describe.)
Be sure to include your name and phone number on this service form
.
Keithley Instruments, Inc.