Number 1138
Switching in
Multipoint Testing
Switching in Multipoint Testing
Selection of suitable switching is an extremely important but
sometimes under-emphasized function in the process of measurement system design. Switch selection is, in fact, no less critical
than the selection of system instrumentation or the design of the
test interface.
The objective of any system design is to transfer the accuracies and specified performance of system instrumentation to
the Unit Under Test (UUT). These specifications are guaranteed
only at the instrument terminals. To complete the test procedure,
it may be necessary to insert meters of cable, complex switching,
and a unique test interface between the system instrumentation
and the UUT. The system designer must deal with errors created
by cabling, connections, measurement techniques, and system
switching. Add the need to accommodate the designer’s worstcase measurement situation, and achievement of the test objective becomes extremely difficult.
This application note describes the use of currently available switching equipment in typical system measurement situations—accommodating or compensating as much as possible for
error from various sources. Covered in some detail are system
temperature measurements, resistance measurements, high voltage switching, and matrix switching.
System Temperature Measurements
Temperature is often measured in test systems, environmental
chambers, automated processes, etc. The fundamental properties
being sensed are electrical characteristics, which vary predictably with temperature. Resistance temperature detectors (RTDs),
thermistors, and thermocouples are all common sensors suitable
for such applications. Adetailed discussion of each sensor type
can be found in Application Note #401.
The most common temperature sensor in measurement systems is the thermocouple because it is durable, inexpensive, and
covers a broad measurement range from about –50°C to 1800°C.
In using thermocouples, however, the test engineer is faced with
unique problems in dealing with microvolt-level signals. It is
difficult to control these levels and to obtain repeatability and
accuracy, even in a simple bench setup.
Thermoelectric offset voltages are present in every connection and termination because of the thermoelectric properties of
the materials involved. Each connection and switch contact is
basically a thermocouple. Thermoelectric offset voltages can
create measurement errors of several degrees; steps must be
taken to recognize, control, and minimize these error sources.
In addition, a thermocouple produces an EMF proportional
to the difference in temperature from end to end. Therefore, the
temperature to be measured cannot be determined unless the
temperature is known at the cold junction. For this reason, a temperature reference must be provided.
The reference must be held at or as near as practical to the
temperature of the thermocouple cold junction. This is generally
accomplished by mounting the terminal connector and the reference device in one thermally conductive block.
The System Solution
There are at least three ways to use thermocouples for multipoint
temperature measurements. The method chosen depends on the
“acceptable measurement uncertainty.”
1. Direct without Compensation. The method shown in Figure
1 is simple and can provide adequate accuracy in monitoring
high temperatures (typically above 300°C). The computing
power of the controller is used to convert the voltage output
of the sensor to a temperature value, and to insert a constant
reference temperature into the conversion equation. This
assumed reference may be valid in well controlled operating
environments.
The major error is the difference between the constant
temperature assumed for the thermocouple reference and the
actual temperature at the terminal block (∆T
A
in Figure 1).
If the difference is no greater than ±6°C, the additional error
in monitoring temperatures above 300°C is less than 2%. An
error of this magnitude may be acceptable in many
applications.
Application Note
Series
Figure 1: Direct Multipoint Temperature Measurement without Compensation
.
.