Kanomax A031 Operating Manual

Page 1
CLIMOMASTER
ANEMOMASTER
MODEL A031/A041/A034/A044
Operation Manual
Read this manual carefully and understand the warnings
described in this manual before operating the product.
Keep this manual handy for future reference.
02001
06.07
No. 603101
Page 2
Thank you for purchasing a product of Kanomax, Inc.
Please read this operation manual carefully and operate the
instrument appropriately by following the instructions given in
this manual.
Page 3
List of Components
Standard
Item Model Qty. Feature and Functions
Straight probe.
A031 1
Main Body
A034 1
&
Probe
A041 1
A044 1
Operation Manual 1
RS232C Cable 1
Measures air velocity, air temperature and volumetric flow
rate.
Articulating probe.
Measures air velocity, air temperature and volumetric flow
rate.
Straight probe.
Measures air velocity, air temperature, volumetric flow
rate and differential pressure.
Articulating probe.
Measures air velocity, air temperature, volumetric flow
rate and differential pressure.
Software CD 1
Carrying Case 1
AC Adapter
Manganese AA
Batteries
STC9R0V0300-A
 6
1
Options
Item Model Function
Analog Output
Printer DPU-H245
Printer Cable 6000-03
Analog Output Terminal
For printing out stored data.
Printer cable for connecting the instrument with the printer.
Page 4
Important Safety Information
[Classifications]
Danger: To Prevent Serious Injury or Death
Warnings in this classification indicate danger that may result in serious injury or death if not
observed.
Caution: To Prevent Damage to the Product
Warnings in this classification indicate risks of damage to the product that may void the product
warranty if not observed.
[Description of Symbols]
This symbol indicates a condition (including danger) that requires caution.
The subject of each caution is illustrated inside the triangle (e.g., high temperature caution symbol shown
on the left).
This symbol indicates prohibition. Do not take a prohibited action shown inside or near this symbol (e.g.,
disassembly prohibition symbol shown on the left).
This symbol indicates a mandatory action. A specific action is given near the symbol.
i
Page 5
Danger
Never bring the probe close to a flammable gas
atmosphere.
Do not use near
flammable gas
--- The heated sensor may cause fire or explosion.
Never disassemble, modify or repair the product.
--- Failure to observe the above may cause short circuit and/or other failure that will affect the
Do not modify /
disassemble
performance.
Carefully follow the instructions provided in this Manual.
--- Failure to observe the instructions may lead to electrical shock, fire or damage to the
Handle Properly
instrument.
If abnormal noise, smell or smoke is observed, or if liquid has entered the instrument,
turn off the instrument immediately, and remove the batteries or pull out the plug.
--- There is possibility of malfunction, electric shock, and fire.
Please contact your local distributor or our service center for repair.
Do not use the instrument in a water vapor atmosphere.
--- Failure to observe the above may cause electrical shock, fire, or
Prohibition
damage to the sensor.
High Temperature
Warning
Never touch the sensor.
--- The sensor is heated during operation.
Touching the heated sensor may cause burns, and may also damage the
sensor itself.
SHAPECLEAR MODE
ENTER
CALC
HOLD
TIMECONS.
STORE
MR
UNITZERO
PRINT
Water drops
LI
LI
ii
Page 6
Handle Properly
Prohibited
Installation
Caution
Always unplug the instrument from the electrical outlet when the instrument is not in
use.
--- Failure to do so may cause electrical shock, fire or circuit damage.
Remove the batteries from the battery compartment when storing the instrument.
Do not leave exhausted batteries in the battery compartment.
--- Failure to do so may cause battery leakage.
Do not use or leave the instrument in a high temperature/
humidity environment, or in a dusty environment.
Do not leave the instrument under direct sunlight for a
prolonged period.
--- The instrument may not function properly out of the specified
operating conditions.
SHAPECLEAR MODE
ENTER
CALC
HOLD
TIMECONS.
STORE
MR
UNITZERO
PRINT
Do not wipe the instrument with a volatile solvent.
--- The body may deform or deteriorate. Use soft dry cloth to remove stains.
If stains persist, soak the cloth in a neutral detergent and wipe the instrument with the soft
Prohibition
cloth. Never use volatile solvents such as thinner or benzine.
Do not apply strong shock or place/drop anything heavy
on the instrument.
--- Failure to observe the above may cause damage or
Prohibition
malfunction to the instrument
Discharge any built-up static electricity from your body before touching the
instrument.
--- The built-up static electricity may influence the readings and cause damage to the circuit.
iii
Page 7
Table of Contents
1. Part Names and Functions ......................................................................................... 1
1.1 Overview..............................................................................................................................1
1.2 Main Body ........................................................................................................................... 2
1.3 Touch Panel.......................................................................................................................... 3
1.4 Probe ....................................................................................................................................4
2. Getting Started ............................................................................................................ 5
2.1 Installing Batteries ...............................................................................................................5
2.2 Turning ON/OFF the Power ................................................................................................6
2.3 Precautions for Measurements............................................................................................. 8
2.3.1 Air Velocity Measurement Precautions ....................................................................................................8
2.3.2 Air Temperature Measurement Precautions..............................................................................................8
2.3.3 Differential Pressure Measurement Precautions (A041, A044)................................................................8
3. Duct Shape and Size .................................................................................................. 10
4. Measurement.............................................................................................................. 11
4.1 Changing the Measurement Mode..................................................................................... 11
4.2 Hold the Reading ...............................................................................................................12
4.3 Setting the Time Constant.................................................................................................. 13
5. Data Storage and Statistical Calculation ................................................................ 14
5.1 Storing Measurement Data ................................................................................................14
5.2 Data Storage and Statistical Calculation Procedure ..........................................................15
5.3 Viewing and Deleting Stored Data .................................................................................... 17
6. Setting the Measurement Unit and Baud Rate....................................................... 18
7. Data Output ...............................................................................................................19
7.1 Printing Out the Measurement Data ..................................................................................19
7.1.1 Preparation for Print Out ........................................................................................................................19
7.1.2 Printing Directly from the Measurement Mode......................................................................................20
7.1.3 Printout Examples (Measurement Mode) ...............................................................................................20
7.1.4 Printing Out Stored Data ........................................................................................................................21
7.1.5 Printout Examples (Stored Data) ............................................................................................................22
7.2 Analog Output (Optional) .................................................................................................. 22
7.3 S232C Serial Communication ...........................................................................................23
8. Cleaning the Probe .................................................................................................... 24
9. Specifications.............................................................................................................. 25
10. Measurement Principles ......................................................................................... 26
11. Air Velocity Compensation .................................................................................... 28
11.1 Influence of Air Temperature...........................................................................................28
11.2 Influence of Atmospheric Pressure.................................................................................. 28
11.3 Influence of Air Composition ..........................................................................................28
12. Probe Directivity (Air Velocity) ............................................................................. 29
12.1 Horizontal Directivity ...................................................................................................... 29
12.2 Vertical Directivity........................................................................................................... 29
Page 8
13. Troubleshooting....................................................................................................... 30
14. Warranty and After Service................................................................................... 32
15. Contact Information................................................................................................ 34
Page 9
)
1. Part Names and Functions
1.1 Overview
Main Body
Probe
Approx. φ6mm
CLEA
HOLD
CAL
ENTER
TIME
STOR
CONS.
ZER UNIT
MOD
SHAPE
MR
PRINT
Probe Cable
(Approx. 2m
Approx. 284~854mm
pprox.φ13mm
1
Page 10
1.2 Main Body
Approx. 188mm
Approx. 51mm
Analog Output Ter minal (Optional)
RS232C Terminal
DC Input Terminal
Power Switch
I : ON
: OFF
Touch Panel
Battery Compartment
Approx. 88mm
Pressure Port
(Optional)
2
Probe Socket
CLEAR
CALC
STORE
ZERO
MODE
ENTER
HOLD
TIME CONS.
UNIT
Approx. 66mm
LCD Display
SHAPE
MR
PRINT
Page 11
t
1.3 Touch Panel
STORE Key:
ZERO Key: (A041, A044)
To provide zero adjustment in pressure measurement mode.
CLEAR Key:
To clear stored data.
CALC Key:
To obtain average, max and min.
To store current measurement data.
ENTER/HOLD Key:
1. To put the current display on hold during a measurement. Press the key again to release hold.
2. To confirm current setting.
MODE Key:
To change measurement mode.
CLEAR
MODE
ENTER
CALC
HOLD
TIME
STORE
CONS.
ZERO UNIT
UNIT Key:
To select the unit of each parameter and to set the communication baud rate.
SHAPE
MR
PRINT
Air Velocity
Air Temp
Flow Volume
Air Temp.
Differential
Pressure
(A041/A044)
SHAPE Key:
To select duct type and size for measuring volumetric flow rate.
▲,▼ Key:
1. To move the cursor for selecting an item and setting values.
2. To scroll through the stored data.
MR/PRINT Key:
1. To view stored data.
2. To printout stored data.
TIME CONS. Key:
To select the time constan for measurements.
3
Page 12
1.4 Probe
Straight Probe (ModelA031/A041)
Articulating Probe (ModelA034/A044)
L
I
Direction Mark
Approx. φ6mm
L
L I
I
How to extend the Articulating Probe:
1. Hold the upper part  of the probe, and unscrew .
2. Pull out the flex-neck .
Fix the probe in its extended position by holding  and turning .
3. Slowly bend the flex-neck. (&)
Caution: Do not excessively bend the flex-neck or apply excessive force. When the instrument is not in use, the probe must be returned to its original position (the flex-neck stored) to avoid possible damage.
Air Velocity Sensor
Temp. Sensor and Temp. Compensation Sensor
L
I
4
Page 13
t
r
2. Getting Started
2.1 Installing Batteries
Backside of the body
1. Press down the battery cover with your
finger as shown left.
2. Slide the cover toward the bottom of the
instrument until it stops.
3. Lift the cover away from the body.
Types of Batteries that can be Used
- Manganese (R6), AA batteries
-
Alkaline (LR6), AA batteries
- Ni-Cd, AA batteries
4. Insert the batteries by observing the polarity. The instrument requires
six (6) AA size batteries. Types of batteries that can be used are:
Manganese (R6), Alkaline (LR6) or Ni-Cd batteries.
The six (6) batteries must be of the same type. Do NOT mix differen
types of batteries. Failure to observe this may cause battery leakage o
damage to the instrument.
* Batteries CANNOT be recharged by the AC adapter.
5. Put the cover back on by reversing the above procedures.
5
Page 14
2.2 Turning ON/OFF the Power
The power switch for turning ON/OFF the instrument is located at the side of the instrument.
When powered up, the LCD test screen will be displayed, which will switch to the Air Velocity / Air Temperature
measurement screen in approx. 2 seconds.
Power Switch
| : Power ON
: Power OFF
Power ON
LCD Test Screen
Air Velocity / Air Temperature Measurement Screen
6
Page 15
r
Battery Level Indicator
The indicator changes as below according to the remaining battery level:
Check the “battery level indicator” to confirm the
remaining battery level. The battery consumption rate
largely depends on the measured air velocity.
When the battery drops to a level requiring
replacement, the indicator will start blinking.
The screen may freeze if high velocity is measured afte
the battery level indicator starts blinking.
Blinks
Batteries must be replaced.
7
Page 16
2.3 Precautions for Measurements
Direction Mark
2.3.1 Air Velocity Measurement Precautions
Probe has its own directivity characteristics. Make sure that the direction mark is
facing against the airflow (for details of the directivity characteristics, refer to Section
12 “Probe Directivity”). If you are not sure of the airflow direction, slowly rotate the
probe and select the point where you get the maximum velocity reading,
The probe compensates air velocity change due to temperature change by using the air velocity
sensor with the temperature compensation sensor. In order to obtain this compensation effect, it
is required that both sensors are evenly exposed to the airflow under the same temperature
condition.
For measurements in an environment with rapid temperature change, you must wait for the
reading to become stable after measuring for more than 20 seconds before taking the data.
Airflow
2.3.2 Air Temperature Measurement Precautions
The response time for temperature measurement improves as the air velocity increases. The normal response
time is approximately 5 seconds when the air velocity is 1m/s. You must wait for the reading to become stable
before taking the data.
When a measurement is performed in a no-airflow condition, the air temperature reading may become higher
than actual due to the heat generated by the air velocity sensor. It is recommended that the measurement is
performed in an environment with at least 0.1m/s airflow to obtain accurate readings.
2.3.3 Differential Pressure Measurement Precautions (A041, A044)
Do not apply pressure more than 75kPa to the pressure sensor. Excess pressure may cause permanent damage to
the sensor.
O
The operating temperature is 5 - 40
operate properly out of this temperature range.
Make sure to perform zero adjustment before measuring the pressure. When performing the zero adjustment,
leave both pressure ports (+) and (-) open.
C (or 41 - 104OF) when measuring the pressure. The instrument may not
8
Page 17
t
r
<Zero Adjustment Procedure>
Display Description
Press
Press and hold the
The pressure value will be 0.00 kpa.
<Connecting the Pressure Tube>
Connect the pressure tube to the (+) or (-) pressure port as shown
right.
Connect the other end of the tube to the port (e.g. duct) where the
pressure is to be measured.
When the pressure to be measured is positive, connect the tube to
the (+) port, and when it is negative, connect the tube to the (-)
port.
In order to take an accurate measurement, make sure tha
the tube is properly secured without any leakage, o
being bent.
key to enter the pressure measurement screen.
key for more than 2 seconds.
9
Page 18
3. Duct Shape and Size
Before measuring the volumetric flow rate, duct shape and size setting must be made.
Duct Shape:
Size of duct: For a rectangular duct, set the width (W) and height (H).
Size Range:
There are two duct shapes - Rectangular and Circular, which are indicated as and at the lower
right corner of the LCD screen.
For a circular duct, set the diameter (D).
Maximum dimension of a side: 2550mm
Input Increment: Range of 0 to 1000mm: 1mm
Range of 1000 to 2550: 10mm
*When inch is selected for the input unit, maximum dimension will be 255 inch.
Input Increment: Range of 0 to 100 inch: 0.1 inch
Range of 100 to 255 inch: 1 inch
Display Description
Press
<Setting of a Rectangular Duct>
Select with the
<Setting of a Circular Duct>
Select with the keys, and press the
<Setting the Dimension for a Rectangular Duct>
Set the height (H) value with the keys, and
press the key.
Set the width (W) value with the keys,
and press the key.
Note: Press and hold or key over 2 seconds, and the increment
speed will accelerate.
key to enter the setting screen.
keys, and press the
key.
key.
H
W
<Setting the Dimension for a Circular Duct>
Set the diameter (D) with the keys,
and press the key.
10
Page 19
N
w
4. Measurement
When the instrument is turned on, the LCD test screen will be displayed for approx. 2 seconds. The test screen will then
switch to the Air Velocity / Air Temperature measurement screen. The reading will be updated each second.
4.1 Changing the Measurement Mode
To change the measurement mode, press the key while each measurement screen is displayed.
The screen will switch in the following sequence.
*Note: If the duct dimension is not set, the screen
Air Velocity / Temperature
umber of Data Records
will switch to the duct dimension setting screen.
Volumetric flow / Temperature*
Air velocity Reading
Air temperature
Duct shape
Volumetric flow rate
Only For A041, A044
Displayed when volumetric flo rate exceeds 99999.
Volumetric Flow rate = reading × 100
Pressure reading
Differential Pressure / Temperature
11
Page 20
4.2 Hold the Reading
Display Description
While measuring, press the key to hold the current reading.
HOLD
mark will be displayed at the upper right of the reading while the
reading is held.
To recover from the hold mode, press the key again.
By pressing the key while the reading is held, the displayed
reading will be saved temporarily in the memory, and hold mode will be
released to resume measurement.
For how to store the data, refer to Section5.2 “Data Storage and
Statistical Calculation Procedure” (P.15).
12
Page 21
4.3 Setting the Time Constant
When there is rapid change in the measurement data, the readings may become difficult to read. In such case, the speed
of updating the readings can be reduced by changing the time constant setting.
Time constant determines the time span of the moving average. When a larger (longer) time constant is selected, the
readings will be rather stable, and when a smaller (shorter) time constant is selected, the readings will be more
responsive and sensitive to the change.
The time constant can be selected from 1, 5, 10 or 20 seconds by pressing the key.
Display Description
< Time Constant Setting Screen >
By pressing the key in a measurement mode, the display will
switch to the time constant setting screen.
Time constant
Time cons. unit
EXAMPLE when setting the time constant to 20 seconds:
Select the time constant “20.0” by pressing the
the key .
Time constant can be set only for the Air Velocity and Volumetric Flow
measurements.
Note: The time constant selected from 1, 5, 10 and 20 seconds will be
initialized once the power is turned off, and will return to the
default setting of 1 second.
keys, and press
13
Page 22
5. Data Storage and Statistical Calculation
5.1 Storing Measurement Data
Measurement data can be stored in the built-in memory of the instrument. The instrument can hold up to 800 data
records. When storing the measurement data, average, maximum and minimum values will be calculated for the data
DNo
group to be stored. Each data group is stored with a storage number (shown as
mode. Contents of the data stored in each measurement mode are shown in the following table.
Measurement Mode Stored Data
Air Velocity /
Air Temperature
Volumetric Flow Rate /
Air Temperature
Pressure /
Air Temperature
Air Velocity, Air Temperature and Data Storage No.
Volumetric Flow Rate, Air Temperature, Data Storage No., and Duct Shape/Size
Pressure, Air Temperature, and Data storage No.
xxx) which starts from 001 for each
14
Page 23
N
5.2 Data Storage and Statistical Calculation Procedure
Data storage procedure will be described below by taking an example of an operation in the “Air Velocity / Air
Temperature Measurement Mode”. Data storage procedure is same in other modes as well.
Display Description
umber of stored data records
< Temporary Data Storage>
By pressing the key while measuring, the current reading will be
stored in the temporary memory.
NOTE:
1. Temporary Memory -- Data will be cleared once measurement
mode is changed or power is turned off.
2. Number of Stored Data Records -- Indicates the number of data
records stored in the temporary memory.
The displayed reading will be stored in the temporary memory each time
the key is pressed, and the number of stored data records will
increase by one. A maximum of 800 data records can be stored.
Storage Number of
the data group
< Data Storage and Statistical Calculation >
Press the key to execute the statistical calculation (Average,
Maximum and Minimum) for the group of data records stored in the
temporary memory.
Average
value will be displayed with the
Storage Number
of the data
group.
Display will switch in the sequence shown left (AVG→MAX→MIN)
as you press the key.
15
Page 24
Display Description
< Storing the Calculation Result >
To store the calculation result, press the key while either average,
maximum or minimum value is displayed. The calculation data will be
stored as a data group.
The display will return to the original measurement mode and the
temporary memory will be cleared.
< To Clear the Data Group >
To clear the data group without saving, press and hold the key
while either average, maximum or minimum value is displayed.
NOTE: Data records in the temporary memory and the calculation
result will be cleared.
will be displayed, and the screen will return to the original
measurement mode.
16
Page 25
5.3 Viewing and Deleting Stored Data
The average, maximum and minimum value of the stored data can be viewed on the display and deleted. Only the
calculation result can be viewed from the instrument. To view each data of the data group, the data must be printed out
from an optional printer (please refer to section 7 “Data Output” P.19).
When deleting the stored data, the calculation result and each data of the selected data group will be deleted.
Data viewing and deleting procedure will be described below by taking an example of an operation in the “Air Velocity
/ Air Temperature Measurement Mode”.
Display Description
< Viewing the Stored Data >
Press the
which the data is stored.
The most recently stored average value (AVG) will be displayed. (Figure
)
< Viewing Other Data >
By pressing the key, maximum, minimum and average value of
the data group can be viewed.
To move to the previous/next data group, press the keys.
key when the instrument is in the measurement mode in
< Exit from Data Viewing >
Press the key to exit (cancel) data viewing.
When cancelled, the screen will return to the measurement mode.
(Figure )
< Deleting the Stored Data Group >
Select the data group (Figure) to be deleted, and press the
key. The air velocity value will start blinking (Figure).
- To delete the selected data, press the
will be displayed (Figure ).
- To delete
press and hold the key for approx. 5 seconds until is
displayed. All Data Groups stored under this mode will be cleared.
After is displayed, the screen will return to the measurement
mode.
data groups stored in the relevant measurement mode,
All
key again. When pressed,
17
Page 26
t
t
6. Setting the Measurement Unit and Baud Rate
List of Measurement Units:
- Air velocity: m/s, FPM
- Air temperature:
- Volumetric flow rate: m
o
C, oF
3
/h, m3/min, ft3/min
- Length: mm, inch
RS232C Communication Baud Rates:
4800bps, 9600bps, 19200bps, 38400bps
Procedure for setting the Units and Baud Rate.
Display Description
Measurement Mode
Length Unit
Air Velocity Uni
Air Velocity: 1m/s=196 FPM
Air Temperature: T(
Volumetric Flow: 1m
Length: 1inch=25.4mm
Press the key to enter the unit setting mode.
The unit to be set will blink.
Use the keys to select the unit. Press the key to save the
setting and to proceed to the next unit setting.
Repeat the above procedure for each unit.
< Unit Conversion Table >
o
F)=1.8×T(oC)+32
3
/h=35.32ft3/h
Volumetric Flow Uni
NOTE:
Air Temperature Unit
- The saved setting will be kept even
when the instrument is turned off.
Baud Rate
- To exit (cancel) the unit/baud rate
setting, press the “MODE” key. Any
setting saved before then is valid.
18
Page 27
f
7. Data Output
7.1 Printing Out the Measurement Data
To print out the stored measurement data, an optional printer and printer cable
are required. The printer cable must be connected to the RS232C terminal located
at the side of the instrument.
RS232C Ter min al
7.1.1 Preparation for Print Out
Equipment (Optional )
- Printer
- Printer Cable
< Communication Protocol >
The baud rate setting of the instrument and printer must be consistent.
List of Communication Protocol:
Data Bit Length 8 bit
Parity None
Stop Bit 1
Delimiter CRLF
Baud Rate Set Value*
< Connecting the Printer and the Instrument >
1. Connect the printer and the instrument with the printer cable by inserting the printer cable in the RS-232C terminal located at the side of the instrument.
2. Turn on the power of the instrument first, and then turn on the printer.
3. Confirm that the instrument is in the measurement mode.
* For how to set the baud rate of the instrument, refer to Section 6
“Setting the Measurement Unit and Baud Rate”.
* For setting the printer, please refer to the operational manual o
your printer.
19
Page 28
7.1.2 Printing Directly from the Measurement Mode
Procedure for printing directly from the measurement mode will be described below taking an example of an operation
in the “Air Velocity / Air Temperature Measurement Mode”.
Display Description
Hold the reading by pressing the key.
sign will be displayed at the upper right part of the LCD screen.
Press the key to execute printing.
While printing, will be displayed on the screen.
When printing is complete, the screen will return to the measurement
screen with the reading held. Measurement can be resumed by releasing
the hold mode.
7.1.3 Printout Examples (Measurement Mode)
Measurement Modes
Air Velocity /
Air Temperature
EXAMPLE:
Vel 0.15m/s Temp 21.8
EXAMPLE:
(When duct shape is Rectanglular.)
o
C
Duct Shape R Size 2550*2550mm
FlowRate 407.3 m3/h Temp 22.5
EXAMPLE:
(When duct shape is Circular.)
Duct Shape C Size 2550mm
FlowRate 407.3m3/h Temp 23.6
Volumetric Flow /
Air Temperature
Air Temperature
EXAMPLE:
Press 1.03 KPa Temp 20.1
o
C
o
C
Pressure /
o
C
20
Page 29
7.1.4 Printing Out Stored Data
Procedure of printing the stored data will be described below by taking an example of printing the sixth data group
stored in the “Air Velocity / Air Temperature Measurement Mode”.
Display Description
Press the key in the measurement mode. The most recently
stored average value will be displayed (Figure ).
Select the Data group to be printed by using the keys.
When selected, press the
(Figure ).
Connect the printer.
key and the reading will start blinking
< Error Message >
Press the
On the screen, will be displayed (Figure
To print
key for approx. 5 seconds until appears on the
screen. All data stored under the relevant mode will be printed out
continuously.
When printing is complete, the display will return to the measurement
mode (Figure ).
If the printer is not connected when executing print, an error message
will be displayed for approx. 2 sec, and the screen will return to the
measurement mode (Figure ).
key again to execute printing.
data group stored in this mode, press and hold the
all
).
21
Page 30
7.1.5 Printout Examples (Stored Data)
Measurement Modes
Air Velocity /
Air Temperature
EXAMPLE for group DNo006
DNo.006
001 Vel 0.15m/s Temp 21.8
002 Vel 0.16m/s Temp 21.8
003 Vel 0.14m/s Temp 21.8
004 Vel 0.13m/s Temp 21.8
AVG Vel 0.15m/s Temp 21.8
EXAMPLE for group DNo006
(When duct shape is Rectanglular.)
o
DNo.006 Duct Shape R Size 2550*2550mm
C
o
001 FlowRate 407.3 m3/h Temp 22.5
C
o
C
002 FlowRate 405.6 m3/h Temp 22.5
o
003 FlowRate 400.9 m3/h Temp 22.5
C
o
004 FlowRate 401.4 m3/h Temp 22.5
C
AVG FlowRate 403.8 m3/h Temp 22.5
EXAMPLE for group DNo005
(When duct shape is Circular.)
DNo.005 Duct Shape C Size 2550mm
001 FlowRate 407.3m3/h Temp 23.6
002 FlowRate 405.6m3/h Temp 23.6
003 FlowRate 400.9m3/h Temp 23.6
004 FlowRate 401.4m3/h Temp 23.6
AVG FlowRate 403.8m3/h Temp 23.6
Volumetric Flow /
Air Temperature
Pressure /
Air Temperature
EXAMPLE for group DNo006
DNo.006
001 Press 1.03 KPa Temp 20.1
o
002 Press 1.02 KPa Temp 20.1
C
o
C
003 Press 1.01 KPa Temp 20.1
o
004 Press 1.00 KPa Temp 20.1
C
o
AVG Press 1.02 KPa Temp 20.1
C
o
C
o
C
o
C
o
C
o
C
o
C
o
C
o
C
o
C
o
C
o
C
7.2 Analog Output (Optional)
Analog output is limited to the output of air velocity in the Air Velocity / Air Temperature Mode.
1. Data update interval……………… 0.1sec
2. Load impedance……………………5KΩ and above
3. Output voltage…………………… DC 0~3V (0~30m/s)
4. Polarity of output voltage…………
22
Analog output terminal
Page 31
7.3 S232C Serial Communication
For serial communication, the RS232C cable must be connected
to the RS232C terminal located at the side of the instrument.
<Equipment >
- Computer
- RS232C Communication Cable
- Communication Software
PC-LINK Software: For transferring stored data to the PC.
Data Acquisition Software: For transferring real time data to the PC.
<Baud Rate>
The baud rate will be set automatically by the communication software.
<Connecting the Instrument with a Computer>
1. Turn off the instrument.
2. Connect the instrument to a computer with the RS232C communication cable.
3. Turn on the instrument.
4. Confirm that the instrument is in normal measurement mode.
RS232C Cable Wiring Diagram
Computer (D-Sub9 pin) ANEMOMASTER (Model A031/A034/A041/A044)
Signal Pin No.
Connection
Pin No. Signal Description of Signal Signal Direction
RS232C terminal
NC 1 1 GND Signal Ground
RXD 2 2 TXD Transmit Data Output
TXD 3 3 RXD Receive Data Input
NC 4 4 CTS Clear to Send Input
GND 5 5 RTS Request to Send Output
NC 6 6 NC
RTS 7
CTS 8
NC 9
* Refer to the operation manual provided with the software for the operation procedure.
23
Page 32
y
8. Cleaning the Probe
When the sensor is contaminated with impurities such as dust, particles, soot or machine oil, the heat dissipation rate
will change. In most cases, heat dissipation will decrease, resulting in lower air velocity readings.
This is same for the probes which are equipped with a mesh cover. The same problem will occur if the mesh is deformed,
or clogged with impurities.
If impurities are attached to the sensor or mesh from using the instrument in an unclean environment, it is recommended
that the sensor is cleaned right after use.
Clean the sensor of the probe in an ultrasonic cleaner for approx. 10-20 sec.
Do not clean the sensor longer than required as excess cleaning will lead to sensor coating damage.
Use water for cleaning.
The sensor can also be cleaned in a vessel filled with neutral detergent diluted with water, and by gentl
stirring it in the vessel.
! Caution !
!) Make sure to TURN OFF the power before cleaning.
Dry the probe completely after cleaning. Do not turn on the power before completely dried.
!)
Method of Cleaning
24
Page 33
9. Specifications
Product Anemomaster
Model A031, A041, A034, A044
Measuring Object Clean Air Flow
Measuring
Range
(Resolution)
Accuracy
Response
Time
Temperature Compensation
Accuracy (Velocity)
Air Velocity 0.10 to 30.0 m/s (0.00 to 9.99m/s: 0.01m/s, 10.0 to 30.0m/s: 0.1m/s)
Air Temperature -20.0 to 60.0 °C (0.1°C)
1
Pressure*
-5.00 to +5.00 kPa (0.01kPa)
Duct Size Range 0 to 2550mm (0 to 255inch)
Air Velocity ± (3% of the reading +0.1 ) m/s
Air Temperature ± 0.5 °C
Pressure*
1
± (3% of the reading +0.01) kPa
Air Velocity Approx.1sec. (at Air Velocity: 1 m/s, 90% Response)
Air Temperature Approx 30sec. (at Air Velocity: 1 m/s, 90% Response)
Pressure*
1
Approx 1 sec.
± (5% of the reading +0.1)m/s in temperature range of 5 to 60°C
Sampling Functions
Output
Power
Battery Life
Operating
Environment
Weigh t
Standard Accessories
Main Unit
Probe
Storage
Temperature
- Hold the reading
- Statistical calculation (Average, Maximum, Minimum)
- Time constant setting (1, 5, 10, 20 sec)
- Remaining battery level indicator (4 levels)
- Selection of sampling units (Air velocity: m/s or ft/min; Flow rate: m
3
or ft
/min; Air temperature: °C or °F )
3
/h, m3/min
- Data storage: Max. 800 data records.
- Duct shape setting : Rectangular or Circular / Duct size unit: mm or inch
Digital output: RS232C (Baud rate: 4800, 9600, 19200, 38400bps) for outputting
to printer and PC.
Analog output*
2
: DC 0 to 3V (Only for air velocity output)
Manganese AA Batteries × 6 (Alkaline or Ni-Cd batteries can be used as well) AC Adapter: AC 100 to 240V (50/60Hz)
Approx. 10 hours (when, Air velocity: 5m/s, Air temp: 20°C, and using Alkaline batteries)
5 to 40°C
-20 to 60°C
5 to 40°C
Approx. 500g (Including Batteries)
Carrying case, Operation manual, Manganese AA Batteries × 6, RS232C cable, Software (for Windows), and AC adapter
Optional Accessories
Analog output, Printer, AC adapter for the printer
*1: Pressure measurement is only available for A041 and A044
*2: Optional
25
Page 34
H
=
p
p
10. Measurement Principles
Hot-wire Anemometer Principle
When the heated air velocity sensor is exposed to airflow, the sensor
temperature will change by the heat drawn by the airflow. Accordingly, the
sensor resistance value will change. This change in the resistance value will
vary largely as the air velocity increases. Therefore, if the relationship
between the air velocity and the resistance value is known, the air velocity can
Airflow
be obtained by measuring the resistance value (or current).
Air velocity sensor (platinum coil)
The Anemomaster anemometer is based on the above principle.
Generally, a hot-wire anemometer employs a feedback circuit to control the sensor to
maintain constant temperature. (Constant Temperature Type)
When there is a change in the air velocity, the heat drawn from the sensor (heat dissipation)
will change accordingly. In order to maintain constant temperature, current is applied to the
Current i
Ve lo c i ty Sensor
sensor to compensate this change. Thus, the air velocity value can be obtained from the
amount of the applied current (i).
The amount of heat [H] drawn from the air velocity sensor can be expressed by the following
Heat Dissi
ation
Ta1 < Ta2
[H]
Ta1
formula.
Ta
2
Where;
H: Heat Dissipation T: Sensor Temperature Ta: Air Temperature
U: Air Velocity a,b: Constant
Air Velocity
[U]
The Heat Dissipation [H] can also be expressed by the following formula from the sensor resistance (R) and current (i).
Heat Dissi
Temperature
Compensation
RI
2
(R is kept constant regardless of the air velocity change)
ation
[H]
Thus:
As shown by this formula, the change in the air velocity “U” can be interpreted as the change
))(( TaTUba +
in the current “i”.
Air Velocity [U]
2
UbaRI +
Temperature Compensation
When the air temperature changes, the amount of heat dissipation will change accordingly even when the air velocity is
constant. Thus, Anemomaster employs a temperature compensation circuit to enable accurate air velocity measurement
by eliminating the influence of the temperature change. For this purpose, a temperature measurement sensor Rc having
the same temperature coefficient as the air velocity is provided at the opposite side of the bridge, and the bridge is
adjusted so that the difference with the air temperature (T-Ta) is kept constant.
Current
Cooling
King’s Formula
Formula sKing'))(( += @TaTUbaH
mount of Current
Air Velocity m/s
26
Page 35
Air Temperature Measurement
An air temperature element (platinum thin film element) which its resistance value changes by the air temperature is
incorporated in one side of the bridge. The air temperature can be obtained by measuring the variance in the resistance
value.
Constant Voltage
Output
Air temperature element
Pressure Measurement (A041, A044)
A diffusion-type semiconductor pressure sensor is employed to measure the pressure. This diffusion-type
semiconductor pressure sensor is based on the Piezoreistance Effect, in which the resistance value changes when the
pressure is applied. It is configured with four (4) diffusion resistances (sensor chips) located on the thin silicon
diaphragm (Fig. 1).
The pressure is applied from above the pressure sensors, and when the diaphragm is deflected as shown in Fig. 2,
compressive stress is applied to R3 and R4 which are located at the center of the diaphragm, and tensile stress is applied
to R1 and R2. The resistance value of the diffusion resistance changes in accordance with the strength of the stress
applied.
Diaphragm
Constant Current (i)
R1-R4: Diffusion Resistance
Fig.1 Pressure Sensor
R3
R1
R2
R4
Pressure
Fig.2
R1
R4
Output
Fig.3 Detection Circuit
R3
R2
By configuring the bridge of the detection circuit with these diffusion resistances, voltage that is proportional to the
pressure can be obtained. In addition, since the diffusion resistance is dependent on temperature, temperature
compensation resistance is employed for the resistance.
27
Page 36
11. Air Velocity Compensation
When the heated air velocity sensor of the instrument is exposed to airflow, the heat is drawn from the sensor. The
instrument obtains air velocity readings by using this relationship between the amount of heat removed (heat
dissipation) and air velocity.
Since the instrument is calibrated with clean airflow with normal temperature and pressure, when the condition of air to
be measured is different from that of the air used for calibration, the heat dissipation amount will differ even when the
velocity is consistent (i.e. velocity reading is influenced by the condition of air).
11.1 Influence of Air Temperature
The instrument is a hot-wire anemometer, which measures the air velocity by using the heat dissipation amount. Thus,
if temperature compensation is not provided, air velocity readings will by affected by the ambient air temperature
change even when the air velocity is consistent. In order to prevent such influence, the instrument is equipped with a
temperature compensation circuit for measuring and compensating the air temperature in the range of 5°C to 60°C.
11.2 Influence of Atmospheric Pressure
The instrument is calibrated under atmospheric pressure of 1013hPa. Since change in the atmospheric pressure will
influence the heat dissipation amount, compensation of the atmospheric pressure is required. Compensation can be
provided by using the following formula.
Where, Um: Actual Air Velocity [m/s]
Um ×=
1013
Uc
Pm
Uc: Air Velocity Reading
Pm: Atmospheric Pressure at the Time of Sampling [hPa]
11.3 Influence of Air Composition
Compensation is required if the measurement is to be performed in an environment including any gas other than air.
Compensation shall be performed by calculating the heat dissipation amount from the physical properties of the gas,
and comparing it with the heat dissipation amount of the air.
28
Page 37
12. Probe Directivity (Air Velocity)
12.1 Horizontal Directivity
-180°
-90°
Direction Mark
90°
+90°
120%
110%
100%
90%
80%
60%
40%
20%
-90°
180°
When Air Velocity is 5m/s
12.2 Vertical Directivity
Direction Mark
90°
-90°
-180°
+90°
120%
110%
100%
90%
80%
60%
40%
20%
-90°
180°
29
When Air Velocity is 5m/s
Page 38
13. Troubleshooting
13.1 Troubleshooting
Symptom Possible Cause / Solution
Display does not appear when power
is turned ON.
”blinks
Reading is displayed as “-----”.
Incorrect air velocity reading.
High air temperature readings.
Air velocity reading does not change
from “0.0” when sampling in
Volumetric Flow / Air Temperature
measurement mode.
Printing Failure
Refer To
(Page No.)
Battery is inserted in wrong polarity.
Turn off the power and insert the battery correctly.
Battery is drained.
Turn off the power and replace the batteries.
Battery is drained.
Turn off the power and replace the batteries.
Measurable range is exceeded.
The instrument must be used in the specified measurement
range.
Probe wire disconnection or sensor damage.
Contact your local distributor for repair
Confirm that the direction mark of the probe is facing against
the airflow direction.
Sensor of the probe is dirty.
Turn off the power, and clean the sensor.
Correct reading cannot be obtained when there is no airflow.
Minimum 0.1m/s velocity is required for measurement.
Duct shape and dimension settings are not made properly
(setting is zero).
Provide correct duct shape/dimension settings.
Confirm that the printer cable is connected properly. 19
Printer is not connected in the right order.
After connecting the printer, turn on the instrument first,
and then turn on the printer.
5
5, 6
5, 6
25
33, 34
8
24
8
10
19
Baud rate is not set properly.
Confirm the instrument and printer settings.
Data Transfer Failure
Analogue Output Failure
Incorrect Output Value Load impedance is set lower than the specified value.
Confirm that the RS232C cable is connected properly. Make
sure that it is not confused with the printer cable.
Confirm that the polarity of the output terminal is correct. 22
The reading is in “hold” mode.
Press the key to release the hold mode.
Load impedance must be set to 5K and over.
30
18
23
12
22
Page 39
13.2 Various Status Displays
Display
The number of stored data records has exceeded the maximum limit of 800 data
records. Stored data must be deleted to enable further data storage.
Printing failure.
Confirm the communication setting and cable connection.
Also confirm that the printer is in normal operating condition.
Communication buffer overflow during data transfer (to printer or PC).
Normal operation will be possible once communication is completed.
Description
Displayed during printing.
When you want to cancel printing, press and hold the key.
31
Page 40
14. Warranty and After Service
14.1 Kanomax Limited Warranty
The limited warranty set forth below is given by KANOMAX JAPAN, Inc. (hereafter referred to as “KJI”) with
respect to the KANOMAX brand anemometer, and its attachment parts including probe and other accessories
(hereafter referred to as “PRODUCT”) purchased directly from KJI or from an authorized distributor.
Your PRODUCT, when delivered to you in new condition in its original container, is warranted against defects in
materials or workmanship as follows: for a period of one (1) year from the date of original purchase, defective parts or
a defective PRODUCT returned to KJI, as applicable, and proven to be defective upon inspection, will be exchanged
for a new or comparable rebuilt parts, or a refurbished PRODUCT as determined by KJI. Warranty for such
replacements shall not extend the original warranty period of the defective PRODUCT.
This limited warranty covers all defects encountered in normal use of the PRODUCT, and does not apply in the
following cases:
(1) Use of parts or supplies other than the PRODUCT sold by KJI, which cause damage to the PRODUCT or
cause abnormally frequent service calls or service problems.
(2) If any PRODUCT has its serial number or date altered or removed.
(3) Loss of damage to the PRODUCT due to abuse, mishandling, alternation, improper packaging by the owner,
accident, natural disaster, electrical current fluctuations, failure to follow operation, maintenance or
environmental instructions prescribed in the PRODUCT's operation manual provided by KJI, or service
performed by other than KJI.
NO IMPLIED WARRANTY, INCLUDING ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS
FOR A PARTICULAR PURPOSE, APPLIES TO THE PRODUCT AFTER THE APPLICABLE PERIOD OF THE
EXPRESS LIMITED WARRANTY STATED ABOVE, AND NO OTHER EXPRESS WARRANTY OR
GUARANTY, EXCEPT AS MENTIONED ABOVE, GIVEN BY ANY PERSON OR ENTITY WITH RESPECT TO
THE PRODUCT SHALL BIND KJI. KJI SHALL NOT BE LIABLE FOR LOSS OF STORAGE CHARGES, LOSS
OR CORRUPTION OF DATA, OR ANY OTHER SPECIAL, INCIDENTAL OR CONSEQUENTIAL DAMAGES
CAUSED BY THE USE OR MISUSE OF, OR INABILITY TO USE, THE PRODUCT, REGARDLESS OF THE
LEGAL THEORY ON WHICH THE CLAIM IS BASED, AND EVEN IF KJI HAS BEEN ADVISED OF THE
POSSIBILITY OF SUCH DAMAGES. IN NO EVENT SHALL RECOVERY OF ANY KIND AGAINST KJI BE
GREATER IN AMOUNT THAN THE PURCHASE PRICE OF THE PRODUCT SOLD BY KJI AND CAUSING
THE ALLEGED DAMAGE. WITHOUT LIMITING THE FOREGOING, THE OWNER ASSUMES ALL RISK
AND LIABILITY FOR LOSS, DAMAGE OF, OR INJURY TO THE OWNER AND THE OWNER'S PROPERTY
AND TO OTHERS AND THEIR PROPERTY ARISING OUT OF USE OR MISUSE OF, OR INABILITY TO USE,
THE PRODUCT NOT CAUSED DIRECTLY BY THE NEGLIGENCE OF KJI. THIS LIMITED WARRANTY
SHALL NOT EXTEND TO ANYONE OTHER THAN THE ORIGINAL PURCHASER OF THE PRODUCT, OR
THE PERSON FOR WHOM IT WAS PURCHASED AS A GIFT, AND STATES THE PURCHASER'S EXCLUSIVE
REMEDY.
32
Page 41
14.1 After Service
When you have a problem with your instrument, please check out the “Troubleshooting” section first.
If that does not help, please contact your local distributor, or call our service center (See last page for contact
information).
During the warranty period, we will repair at no charge a product that proves to be defective due to material or
workmanship under normal use. (See Section 14.1 Kanomax Limited Warranty)
All return shipping charges are the responsibility of the customer.
Repair after warranty expiration:
Upon request, we will repair the instrument at the customer’s expense, if the instrument’s performance is found to
be recoverable by providing the repair.
Replacement parts are available for a minimum period of five (5) years after termination of production. This
storage period of replacement parts is considered as the period during which we can provide repair service.
For further information, please contact your local distributor or our service center.
When making an inquiry, please provide the following information.
(1) PRODUCT name: Anemomaster
(2) Model Number: A031, A034, A041 or A044 (3) Serial number: xxxxxx (4) Probe number: xxxxxx
(5) Description of Symptom in detail:
(6) Data of Purchase: Day, Month and Year
33
Page 42
15. Contact Information
U.S.A.
KANOMAX USA, INC.
PO Box 372, 219 Route 206, Andover, NJ 07821 U.S.A.
Te l: (800)-247-8887 / (973)-786-6386 FAX : (973)-786-7586
URL:
E-Mail: info@kanomax-usa.com
http://www.kanomax-usa.com/
JAPAN
KANOMAX JAPAN, INC.
2-1 Shimizu Suita City, Osaka 565-0805, Japan
TEL: 81-6-6877-0183 FAX: 81-6-6879-2080
URL: http://www.kanomax.co.jp/
E-Mail: sales@kanomax.co.jp
CHINA
Shenyang Kano Scientific Instrument Co., Ltd
No. 12, 4 Jia Wencui Road Heping District
Shenyang City PRC
TEL: 86-24-23845309 FAX : 86-24-23898417
URL: http://www.kanomax.com.cn/
E-Mail: sales@kanomax.com.cn
34
Loading...