4.1 Changing the Measurement Mode..................................................................................... 11
4.2 Hold the Reading ...............................................................................................................12
4.3 Setting the Time Constant.................................................................................................. 13
5. Data Storage and Statistical Calculation ................................................................ 14
5.1 Storing Measurement Data ................................................................................................14
5.2 Data Storage and Statistical Calculation Procedure ..........................................................15
5.3 Viewing and Deleting Stored Data .................................................................................... 17
6. Setting the Measurement Unit and Baud Rate....................................................... 18
7. Data Output ...............................................................................................................19
7.1 Printing Out the Measurement Data ..................................................................................19
7.1.1 Preparation for Print Out ........................................................................................................................19
7.1.2 Printing Directly from the Measurement Mode......................................................................................20
7.1.4 Printing Out Stored Data ........................................................................................................................21
To provide zero adjustment
in pressure measurement
mode.
CLEAR Key:
To clear stored data.
CALC Key:
To obtain average,
max and min.
To store current
measurement data.
ENTER/HOLD Key:
1. To put the current display on
hold during a measurement.
Press the key again to release
hold.
2. To confirm current setting.
MODE Key:
To change
measurement mode.
CLEAR
MODE
ENTER
CALC
HOLD
TIME
STORE
CONS.
ZEROUNIT
UNIT Key:
To select the unit of each
parameter and to set the
communication baud rate.
SHAPE
MR
PRINT
Air Velocity
Air Temp
Flow Volume
Air Temp.
Differential
Pressure
(A041/A044)
SHAPE Key:
To select duct type
and size for measuring
volumetric flow rate.
▲,▼ Key:
1. To move the cursor for
selecting an item and
setting values.
2. To scroll through the
stored data.
MR/PRINT Key:
1. To view stored data.
2. To printout stored data.
TIME CONS. Key:
To select the time constan
for measurements.
3
Page 12
1.4 Probe
Straight Probe (ModelA031/A041)
Articulating Probe (ModelA034/A044)
L
I
Direction Mark
Approx. φ6mm
L
L I
I
How to extend the Articulating Probe:
1. Hold the upper part of the probe, and unscrew .
2. Pull out the flex-neck .
Fix the probe in its extended position by holding and turning .
3. Slowly bend the flex-neck. (&)
Caution:
Do not excessively bend the flex-neck or apply excessive force.
When the instrument is not in use, the probe must be returned to its
original position (the flex-neck stored) to avoid possible damage.
Air Velocity Sensor
Temp. Sensor and
Temp. Compensation Sensor
L
I
4
Page 13
t
r
2. Getting Started
2.1 Installing Batteries
Backside of the body
1. Press down the battery cover with your
finger as shown left.
2. Slide the cover toward the bottom of the
instrument until it stops.
3. Lift the cover away from the body.
Types of Batteries that can be Used
- Manganese (R6), AA batteries
-
Alkaline (LR6), AA batteries
- Ni-Cd, AA batteries
4. Insert the batteries by observing the polarity. The instrument requires
six (6) AA size batteries. Types of batteries that can be used are:
Manganese (R6), Alkaline (LR6) or Ni-Cd batteries.
The six (6) batteries must be of the same type. Do NOT mix differen
types of batteries. Failure to observe this may cause battery leakage o
damage to the instrument.
* Batteries CANNOT be recharged by the AC adapter.
5. Put the cover back on by reversing the above procedures.
5
Page 14
2.2 Turning ON/OFF the Power
The power switch for turning ON/OFF the instrument is located at the side of the instrument.
When powered up, the LCD test screen will be displayed, which will switch to the Air Velocity / Air Temperature
measurement screen in approx. 2 seconds.
Power Switch
| : Power ON
○: Power OFF
Power ON
LCD Test Screen
Air Velocity / Air Temperature
Measurement Screen
6
Page 15
r
■ Battery Level Indicator
The indicator changes as below according to the remaining battery level:
Check the “battery level indicator” to confirm the
remaining battery level. The battery consumption rate
largely depends on the measured air velocity.
When the battery drops to a level requiring
replacement, the indicator will start blinking.
The screen may freeze if high velocity is measured afte
the battery level indicator starts blinking.
Blinks
Batteries must be replaced.
7
Page 16
2.3 Precautions for Measurements
Direction Mark
2.3.1 Air Velocity Measurement Precautions
Probe has its own directivity characteristics. Make sure that the direction mark is
facing against the airflow (for details of the directivity characteristics, refer to Section
12 “Probe Directivity”). If you are not sure of the airflow direction, slowly rotate the
probe and select the point where you get the maximum velocity reading,
The probe compensates air velocity change due to temperature change by using the air velocity
sensor with the temperature compensation sensor. In order to obtain this compensation effect, it
is required that both sensors are evenly exposed to the airflow under the same temperature
condition.
For measurements in an environment with rapid temperature change, you must wait for the
reading to become stable after measuring for more than 20 seconds before taking the data.
Airflow
2.3.2 Air Temperature Measurement Precautions
The response time for temperature measurement improves as the air velocity increases. The normal response
time is approximately 5 seconds when the air velocity is 1m/s. You must wait for the reading to become stable
before taking the data.
When a measurement is performed in a no-airflow condition, the air temperature reading may become higher
than actual due to the heat generated by the air velocity sensor. It is recommended that the measurement is
performed in an environment with at least 0.1m/s airflow to obtain accurate readings.
Do not apply pressure more than 75kPa to the pressure sensor. Excess pressure may cause permanent damage to
the sensor.
O
The operating temperature is 5 - 40
operate properly out of this temperature range.
Make sure to perform zero adjustment before measuring the pressure. When performing the zero adjustment,
leave both pressure ports (+) and (-) open.
C (or 41 - 104OF) when measuring the pressure. The instrument may not
8
Page 17
t
r
<Zero Adjustment Procedure>
Display Description
Press
Press and hold the
The pressure value will be 0.00 kpa.
<Connecting the Pressure Tube>
Connect the pressure tube to the (+) or (-) pressure port as shown
right.
Connect the other end of the tube to the port (e.g. duct) where the
pressure is to be measured.
When the pressure to be measured is positive, connect the tube to
the (+) port, and when it is negative, connect the tube to the (-)
port.
In order to take an accurate measurement, make sure tha
the tube is properly secured without any leakage, o
being bent.
key to enter the pressure measurement screen.
key for more than 2 seconds.
9
Page 18
3. Duct Shape and Size
Before measuring the volumetric flow rate, duct shape and size setting must be made.
Duct Shape:
Size of duct: For a rectangular duct, set the width (W) and height (H).
Size Range:
There are two duct shapes - Rectangular and Circular, which are indicated as and at the lower
right corner of the LCD screen.
For a circular duct, set the diameter (D).
Maximum dimension of a side: 2550mm
Input Increment: Range of 0 to 1000mm: 1mm
Range of 1000 to 2550: 10mm
*When inch is selected for the input unit, maximum dimension will be 255 inch.
Input Increment: Range of 0 to 100 inch: 0.1 inch
Range of 100 to 255 inch: 1 inch
Display Description
Press
<Setting of a Rectangular Duct>
Select with the
<Setting of a Circular Duct>
Select with the keys, and press the
<Setting the Dimension for a Rectangular Duct>
Set the height (H) value with the keys, and
press the key.
Set the width (W) value with the keys,
and press the key.
Note: Press and hold or key over 2 seconds, and the increment
speed will accelerate.
key to enter the setting screen.
keys, and press the
▲
▼
▲
▼
▲
▼
▲
▼
▼
▲
key.
key.
H
W
<Setting the Dimension for a Circular Duct>
Set the diameter (D) with the keys,
and press the key.
10
▼
▲
Page 19
N
w
4. Measurement
When the instrument is turned on, the LCD test screen will be displayed for approx. 2 seconds. The test screen will then
switch to the Air Velocity / Air Temperature measurement screen. The reading will be updated each second.
4.1 Changing the Measurement Mode
To change the measurement mode, press the key while each measurement screen is displayed.
The screen will switch in the following sequence.
*Note: If the duct dimension is not set, the screen
Air Velocity / Temperature
umber of Data Records
will switch to the duct dimension setting screen.
Volumetric flow / Temperature*
Air velocity
Reading
Air temperature
Duct shape
Volumetric flow rate
Only For A041, A044
Displayed when volumetric flo
rate exceeds 99999.
Volumetric Flow rate = reading × 100
Pressure reading
Differential Pressure / Temperature
11
Page 20
4.2 Hold the Reading
Display Description
While measuring, press the key to hold the current reading.
HOLD
mark will be displayed at the upper right of the reading while the
reading is held.
To recover from the hold mode, press the key again.
By pressing the key while the reading is held, the displayed
reading will be saved temporarily in the memory, and hold mode will be
released to resume measurement.
For how to store the data, refer to Section5.2 “Data Storage and
Statistical Calculation Procedure” (P.15).
12
Page 21
4.3 Setting the Time Constant
When there is rapid change in the measurement data, the readings may become difficult to read. In such case, the speed
of updating the readings can be reduced by changing the time constant setting.
Time constant determines the time span of the moving average. When a larger (longer) time constant is selected, the
readings will be rather stable, and when a smaller (shorter) time constant is selected, the readings will be more
responsive and sensitive to the change.
The time constant can be selected from 1, 5, 10 or 20 seconds by pressing the key.
Display Description
< Time Constant Setting Screen >
By pressing the key in a measurement mode, the display will
switch to the time constant setting screen.
Time constant
Time cons. unit
EXAMPLE when setting the time constant to 20 seconds:
Select the time constant “20.0” by pressing the
the key .
Time constant can be set only for the Air Velocity and Volumetric Flow
measurements.
Note: The time constant selected from 1, 5, 10 and 20 seconds will be
initialized once the power is turned off, and will return to the
default setting of 1 second.
▼
▲
keys, and press
13
Page 22
5. Data Storage and Statistical Calculation
5.1 Storing Measurement Data
Measurement data can be stored in the built-in memory of the instrument. The instrument can hold up to 800 data
records. When storing the measurement data, average, maximum and minimum values will be calculated for the data
DNo
group to be stored. Each data group is stored with a storage number (shown as
mode. Contents of the data stored in each measurement mode are shown in the following table.
Measurement Mode Stored Data
Air Velocity /
Air Temperature
Volumetric Flow Rate /
Air Temperature
Pressure /
Air Temperature
Air Velocity, Air Temperature and Data Storage No.
Volumetric Flow Rate, Air Temperature, Data Storage No., and Duct Shape/Size
Pressure, Air Temperature, and Data storage No.
xxx) which starts from 001 for each
14
Page 23
N
5.2 Data Storage and Statistical Calculation Procedure
Data storage procedure will be described below by taking an example of an operation in the “Air Velocity / Air
Temperature Measurement Mode”. Data storage procedure is same in other modes as well.
Display Description
umber of stored data records
< Temporary Data Storage>
By pressing the key while measuring, the current reading will be
stored in the temporary memory.
NOTE:
1. Temporary Memory -- Data will be cleared once measurement
mode is changed or power is turned off.
2. Number of Stored Data Records -- Indicates the number of data
records stored in the temporary memory.
The displayed reading will be stored in the temporary memory each time
the key is pressed, and the number of stored data records will
increase by one. A maximum of 800 data records can be stored.
Storage Number of
the data group
< Data Storage and Statistical Calculation >
Press the key to execute the statistical calculation (Average,
Maximum and Minimum) for the group of data records stored in the
temporary memory.
Average
value will be displayed with the
Storage Number
of the data
group.
Display will switch in the sequence shown left (AVG→MAX→MIN)
as you press the key.
15
Page 24
Display Description
< Storing the Calculation Result >
To store the calculation result, press the key while either average,
maximum or minimum value is displayed. The calculation data will be
stored as a data group.
The display will return to the original measurement mode and the
temporary memory will be cleared.
< To Clear the Data Group >
To clear the data group without saving, press and hold the key
while either average, maximum or minimum value is displayed.
NOTE: Data records in the temporary memory and the calculation
result will be cleared.
will be displayed, and the screen will return to the original
measurement mode.
16
Page 25
5.3 Viewing and Deleting Stored Data
The average, maximum and minimum value of the stored data can be viewed on the display and deleted. Only the
calculation result can be viewed from the instrument. To view each data of the data group, the data must be printed out
from an optional printer (please refer to section 7 “Data Output” P.19).
When deleting the stored data, the calculation result and each data of the selected data group will be deleted.
Data viewing and deleting procedure will be described below by taking an example of an operation in the “Air Velocity
/ Air Temperature Measurement Mode”.
Display Description
< Viewing the Stored Data >
Press the
which the data is stored.
The most recently stored average value (AVG) will be displayed. (Figure
)
< Viewing Other Data >
By pressing the key, maximum, minimum and average value of
the data group can be viewed.
To move to the previous/next data group, press the keys.
key when the instrument is in the measurement mode in
▼
▲
< Exit from Data Viewing >
Press the key to exit (cancel) data viewing.
When cancelled, the screen will return to the measurement mode.
(Figure )
< Deleting the Stored Data Group >
Select the data group (Figure) to be deleted, and press the
key. The air velocity value will start blinking (Figure).
- To delete the selected data, press the
will be displayed (Figure ).
- To delete
press and hold the key for approx. 5 seconds until is
displayed. All Data Groups stored under this mode will be cleared.
After is displayed, the screen will return to the measurement
mode.
data groups stored in the relevant measurement mode,
All
key again. When pressed,
17
Page 26
t
t
6. Setting the Measurement Unit and Baud Rate
List of Measurement Units:
- Air velocity: m/s, FPM
- Air temperature:
- Volumetric flow rate: m
o
C, oF
3
/h, m3/min, ft3/min
- Length: mm, inch
RS232C Communication Baud Rates:
4800bps, 9600bps, 19200bps, 38400bps
Procedure for setting the Units and Baud Rate.
Display Description
Measurement Mode
Length Unit
Air Velocity Uni
Air Velocity: 1m/s=196 FPM
Air Temperature: T(
Volumetric Flow: 1m
Length: 1inch=25.4mm
Press the key to enter the unit setting mode.
The unit to be set will blink.
Use the keys to select the unit. Press the key to save the
▼
▲
setting and to proceed to the next unit setting.
Repeat the above procedure for each unit.
< Unit Conversion Table >
o
F)=1.8×T(oC)+32
3
/h=35.32ft3/h
Volumetric Flow Uni
NOTE:
Air Temperature Unit
- The saved setting will be kept even
when the instrument is turned off.
Baud Rate
- To exit (cancel) the unit/baud rate
setting, press the “MODE” key. Any
setting saved before then is valid.
18
Page 27
f
7. Data Output
7.1 Printing Out the Measurement Data
To print out the stored measurement data, an optional printer and printer cable
are required. The printer cable must be connected to the RS232C terminal located
at the side of the instrument.
RS232C
Ter min al
7.1.1 Preparation for Print Out
Equipment (Optional )
- Printer
- Printer Cable
< Communication Protocol >
The baud rate setting of the instrument and printer must be consistent.
List of Communication Protocol:
Data Bit Length 8 bit
Parity None
Stop Bit 1
Delimiter CRLF
Baud Rate Set Value*
< Connecting the Printer and the Instrument >
1. Connect the printer and the instrument with the printer cable by inserting the printer cable in the RS-232C terminal
located at the side of the instrument.
2. Turn on the power of the instrument first, and then turn on the printer.
3. Confirm that the instrument is in the measurement mode.
* For how to set the baud rate of the instrument, refer to Section 6
“Setting the Measurement Unit and Baud Rate”.
* For setting the printer, please refer to the operational manual o
your printer.
19
Page 28
7.1.2 Printing Directly from the Measurement Mode
Procedure for printing directly from the measurement mode will be described below taking an example of an operation
in the “Air Velocity / Air Temperature Measurement Mode”.
Display Description
Hold the reading by pressing the key.
sign will be displayed at the upper right part of the LCD screen.
Press the key to execute printing.
While printing, will be displayed on the screen.
When printing is complete, the screen will return to the measurement
screen with the reading held. Measurement can be resumed by releasing
the hold mode.
7.1.3 Printout Examples (Measurement Mode)
Measurement Modes
Air Velocity /
Air Temperature
EXAMPLE:
Vel 0.15m/s Temp 21.8
EXAMPLE:
(When duct shape is Rectanglular.)
o
C
Duct Shape R Size 2550*2550mm
FlowRate 407.3 m3/h Temp 22.5
EXAMPLE:
(When duct shape is Circular.)
Duct Shape C Size 2550mm
FlowRate 407.3m3/h Temp 23.6
Volumetric Flow /
Air Temperature
Air Temperature
EXAMPLE:
Press 1.03 KPa Temp 20.1
o
C
o
C
Pressure /
o
C
20
Page 29
7.1.4 Printing Out Stored Data
Procedure of printing the stored data will be described below by taking an example of printing the sixth data group
stored in the “Air Velocity / Air Temperature Measurement Mode”.
Display Description
Press the key in the measurement mode. The most recently
stored average value will be displayed (Figure ).
Select the Data group to be printed by using the keys.
When selected, press the
(Figure ).
Connect the printer.
key and the reading will start blinking
▼
▲
< Error Message >
Press the
On the screen, will be displayed (Figure
To print
key for approx. 5 seconds until appears on the
screen. All data stored under the relevant mode will be printed out
continuously.
When printing is complete, the display will return to the measurement
mode (Figure ).
If the printer is not connected when executing print, an error message
will be displayed for approx. 2 sec, and the screen will return to the
measurement mode (Figure ).
key again to execute printing.
data group stored in this mode, press and hold the
all
).
21
Page 30
7.1.5 Printout Examples (Stored Data)
Measurement Modes
Air Velocity /
Air Temperature
EXAMPLEfor group DNo006
DNo.006
001 Vel 0.15m/s Temp 21.8
002 Vel 0.16m/s Temp 21.8
003 Vel 0.14m/s Temp 21.8
004 Vel 0.13m/s Temp 21.8
AVG Vel 0.15m/s Temp 21.8
EXAMPLE for group DNo006
(When duct shape is Rectanglular.)
o
DNo.006 Duct Shape R Size 2550*2550mm
C
o
001 FlowRate 407.3 m3/h Temp 22.5
C
o
C
002 FlowRate 405.6 m3/h Temp 22.5
o
003 FlowRate 400.9 m3/h Temp 22.5
C
o
004 FlowRate 401.4 m3/h Temp 22.5
C
AVG FlowRate 403.8 m3/h Temp 22.5
EXAMPLE for group DNo005
(When duct shape is Circular.)
DNo.005 Duct Shape C Size 2550mm
001 FlowRate 407.3m3/h Temp 23.6
002 FlowRate 405.6m3/h Temp 23.6
003 FlowRate 400.9m3/h Temp 23.6
004 FlowRate 401.4m3/h Temp 23.6
AVG FlowRate 403.8m3/h Temp 23.6
Volumetric Flow /
Air Temperature
Pressure /
Air Temperature
EXAMPLE for group DNo006
DNo.006
001 Press 1.03 KPa Temp 20.1
o
002 Press 1.02 KPa Temp 20.1
C
o
C
003 Press 1.01 KPa Temp 20.1
o
004 Press 1.00 KPa Temp 20.1
C
o
AVG Press 1.02 KPa Temp 20.1
C
o
C
o
C
o
C
o
C
o
C
o
C
o
C
o
C
o
C
o
C
o
C
7.2 Analog Output (Optional)
Analog output is limited to the output of air velocity in the Air Velocity / Air Temperature Mode.
1. Data update interval……………… 0.1sec
2. Load impedance……………………5KΩ and above
3. Output voltage…………………… DC 0~3V (0~30m/s)
4. Polarity of output voltage…………
22
Analog output terminal
Page 31
7.3 S232C Serial Communication
For serial communication, the RS232C cable must be connected
to the RS232C terminal located at the side of the instrument.
<Equipment >
- Computer
- RS232C Communication Cable
- Communication Software
PC-LINK Software: For transferring stored data to the PC.
Data Acquisition Software: For transferring real time data to the PC.
<Baud Rate>
The baud rate will be set automatically by the communication software.
<Connecting the Instrument with a Computer>
1. Turn off the instrument.
2. Connect the instrument to a computer with the RS232C communication cable.
3. Turn on the instrument.
4. Confirm that the instrument is in normal measurement mode.