Push-in RTD temperature probes
with terminal head form B
■ For temperatures from -50 to +600 °C
■ with replaceable measuring insert
■ As single or double RTD temperature probes
■ Connection head form B, BUZ, BUZH, BBK
■ Available with transmitter
Push-in RTD temperature probes are preferentially used for measuring temperatures in liquids
and gasses. Application areas encompass, among others, the air conditioning technology and
refrigeration engineering, as well as the heating system, kiln and apparatus engineering sector,
and in process technology.
The connection head is suitable for ambient temperatures up to +100 C. The versions BUZ,
BUZH and BBK are also available in addition to the standard connection head, form B.
The measuring insert is normally fitted with a Pt100 temperature probe as per DIN EN 60751,
class B in 2-wire circuit, versions with Pt500 or Pt1000 are also possible. Connection is
available in 3-wire or 4-wire-circuitry as an option. A transmitter can be optionally integrated
into the connection head.
Data Sheet 90.2120 (90.2102)Page 1/5
Technical data
Terminal headForm B DIN 43729, die-cast aluminum, M 20x1.5; IP65, ambient temperature -40 to +100 °C
Form BUZ, die-cast aluminum, M 20x1.5; IP65, ambient temperature -40 to +100 °C
Form BUZH, die-cast aluminum, M 20x1.5; IP65, ambient temperature -40 to+ 100 °C
Form BBK, plastic, M 20x1.5; IP54, ambient temperature -30 to +130 °C
Caution: reduced ambient temperature range when using transmitters,
data sheets 70.7010, 70.7030 and 70.7060
Process connectionSheet steel flange Ø 9 mm, Ø 11 mm, galvanized steel
SheathStainless steel 1.4571, Ø 9 mm, Ø 11 mm, Ø 15 mm
Measuring insertReplaceable, Pt100 temperature probe, DIN EN 60751, class B, 2-wire circuit
Transmitteranalog transmitter, 4 to 20 mA output, data sheet 70.7030
Stop flange Ø 15 mm, DIN 43734, gray cast iron
Pipe screw-connection Ø 12 mm, Ø 15 mm, stainless steel
analog transmitter, 0 to 10 V output, data sheet 70.7030
programmable transmitter, 4 to 20 mA/20 to 4 mA output, data sheet 70.7010
Wtrans B programmable head transducer with radio transmission, data sheet 70.7060
(suitable Wtrans receivers, data sheet 90.2931)
Order specifications: Push-in RTD temperature probe with connection head form B
(1) Basic type
902120/10
Push-in RTD temperature probes
with continuous sheath
(2) Operating temperature in °C
x150-200 to +600 °C
x402-50 to +400 °C (standard)
x415-50 to +600 °C
(3) Measuring insert
x10011x Pt100 in 3-wire-circuit
x10031x Pt100 in 2-wire circuit
x10111x Pt100 in 4-wire circuit
x20012x Pt100 in 3-wire circuit
x20032x Pt100 in 2-wire circuit
(4) Tolerance class as per DIN EN 60751
x1Class B (standard)
x2Class A
(5) Sheath diameter D in mm
x9Ø 9mm
x11Ø 11mm
x15Ø 15mm
(6) Nominal length NL in mm (100≤ΝΛ£1000)
x180180 mm, fitting length (EL) 100 to 140 mm
x270270 mm, fitting length (EL) 100 to 230 mm
x290290 mm, fitting length (EL) 100 to 250 mm
x350350 mm, fitting length (EL) 100 to 310 mm
x410410 mm, fitting length (EL) 100 to 370 mm
x420420 mm, fitting length (EL) 100 to 400 mm
x500500 mm, fitting length (EL) 100 to 460 mm
x710710 mm, fitting length (EL) 100 to 670 mm
x10001000 mm, fitting length (EL) 100 to 960 mm
x...Specification in plain text (50 mm steps)
(7) Process connection
x000without process connection
x254Sheath screw connection G 1/2 (for a 15 mm diameter
only)
x663Sheet steel flange Ø 9 mm
x665Sheet steel flange Ø 11 mm
x668Stop flange Ø 15 mm, DIN 43734
(8) Extra codes
x000no extra code
x320terminal head Form BUZ
x321terminal head Form BUZH
x322Connection head form BBK
x3301x analog transmitter, 4 to 20 mA
x3311x transmitter programmable output 4 to 20 mA/20 to 4 mA
x3331x analog transmitter, 0 to 10 V
x3342x analog transmitters, 4 to 20 mA
x3352x transmitter programmable output 4 to 20 mA/20 to 4 mA
x3372x analog transmitters, 0 to 10V
2
output, data sheet 70.7030
2
output, data sheet 70.7030
2
output, data sheet 70.7030
2
output, data sheet 70.7030
3
, data sheet 70.7010
3
, data sheet 70.7010
x8591x Wtrans B programmable head transducer with radio transmission, data sheet 70.7060
Order code------/ , ...
(1)(2)(3)(4)(5)(6)(7)(8)
1
Order example902120/10-402-1001-1-9-410-663/000
1
List extra codes in sequence, separated by commas.
2
Specify measuring range in plain text.
3
Specify measuring range and output signal in plain text.
Order specifications: Push-in RTD temperature probe with connection head form B
(1) Basic type
Data Sheet 90.2120 (90.2102)Page 4/5
902120/11
Push-in RTD temperature probes
with stepped sheath
(2) Operating temperature in °C
x150-200 to +600 °C
x402-50 to +400 °C (standard)
x415-50 to +600 °C
(3) Measuring insert
x10011x Pt100 in 3-wire-circuit
x10031x Pt100 in 2-wire circuit
x10111x Pt100 in 4-wire circuit
x20012x Pt100 in 3-wire circuit
x20032x Pt100 in 2-wire circuit
(4) Tolerance class as per DIN EN 60751
x1Class B (standard)
x2Class A
(5) Sheath diameter D in mm
x12Ø 12 mm stepped down to 9 mm
(6) Nominal length NL in mm (100≤ΝΛ£700)
x180180 mm, fitting length (EL) 100 to 140 mm
x270270 mm, fitting length (EL) 100 to 230 mm
x290290 mm, fitting length (EL) 100 to 250 mm
x350350 mm, fitting length (EL) 100 to 310 mm
x410410 mm, fitting length (EL) 100 to 370 mm
x500500 mm, fitting length (EL) 100 to 460 mm
x...Specification in plain text (50 mm steps)
(7) Process connection
x000without process connection
x254Sheath screw-connection G 1/2
(8) Extra codes
x000no extra code
x320terminal head Form BUZ
x321terminal head Form BUZH
x322Connection head form BBK
x3301x analog transmitter, 4 to 20 mA
x3311x transmitter programmable output 4to20mA/20to4mA
x3331x analog transmitter, 0 to 10 V
x3342x analog transmitters, 4 to 20 mA
x3352x transmitter programmable output 4to20mA/20to4mA
x3372x analog transmitters, 0 to 10 V
2
output, data sheet 70.7030
2
output, data sheet 70.7030
2
output, data sheet 70.7030
2
output, data sheet 70.7030
3
, data sheet 70.7010
3
, data sheet 70.7010
x8591x Wtrans B programmable head transducer with radio transmission, data sheet 70.7060
Order code------/ ,...
(1)(2)(3)(4)(5)(6)(7)(8)
1
Order example902120/11-402-1001-1-12-410-000/000
1
List extra codes in sequence, separated by commas.
2
Specify measuring range in plain text.
3
Specify measuring range and output signal in plain text.
Accessories for programmable 2-wire transmitter type 707010
in the accessories data sheet 70.9770
PartSales No.:
PC interface with TTL/RS232 converter and adapter (socket connector)70/00350260
PC interface with USB/TTL converter, adapter (socket connector) and adapter (pins)70/00456352
Special programming (standard measuring range 0 to 100 °C)Setup program on CD-ROM, multilingual70/00378733
Accessories for Wtrans B, programmable head transmitter with radio transmission
Type 707060
PartSales No.:
Lithium battery 3.6 V, 2.2 Ah (size AA)70/00547559
PC interface with USB/TTL converter, adapter (socket) and adapter (pins)70/00456352
PC interface with TTL/RS232 converter and adapter (socket connector)70/00350260
Setup program on CD-ROM, multilingual70/00488887
Setup program including OnlineChart on CD-ROM, multilingual70/00549067
OnlineChart activation70/00549188
2011-03-17/00311498
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.