Screw-in RTD temperature probes
with terminal head form B
For temperatures from -50 to +600 °C
With replaceable measuring insert
As single or double RTD temperature probes
Connection head form B, BUZ, BUZH, BBK
Available with transmitter
Screw-in RTD temperature probes are mainly used for measuring temperatures in fluids and
gasses. A decisive selection criterion is the reliable sealing feature of this installation type with
vacuum and with overpressure. The application areas are, among others, in the air conditioning
technology and refrigeration engineering as well as the HVAC, kiln and apparatus engineering
sector.
The connection head is suitable for temperatures up to +100 °C. The versions BUZ, BUZH and
BBK are also available in addition to the standard connection head, form B.
A Pt100 temperature sensing element according to DIN EN 60751, class B in 2-wire circuit is
used as a measuring insert by default – versions with Pt500 or Pt1000 are also possible.
Connection is available in 3-wire or 4-wire circuit as an option. A transmitter can be optionally
integrated into the connection head.
Data Sheet 902020Page 1/6
Technical data
Terminal headForm B DIN 43729, die-cast aluminum, M20 × 1.5; IP65, ambient temperature -40 to +100 °C
Form BUZ, die-cast aluminum, M20 × 1.5; IP65, ambient temperature -40 to +100 °C
Form BUZH, die-cast aluminum, M20 × 1.5; IP65, ambient temperature -40 to +100 °C
Form BBK, plastic, M20 × 1.5; IP54, ambient temperature -30 to +130 °C
Caution: reduced ambient temperature range when using transmitters,
data sheets 707010, 707030 and 707060
Extension tubeStainless steel 1.4571, length 130 mm
Process connectionThread, stainless steel 1.4571
Thermowell, stainless steel 1.4571 or steel 1.7335
ThermowellStainless steel 1.4571, Ø 9 mm, Ø 11 mm, Ø 12 mm
Measuring insertReplaceable, Pt100 temperature probe, DIN EN 60751, class B, 2-wire circuit
Response timest
Transmitteranalog transmitter, 4 to 20 mA output, data sheet 707030
AccessoriesThermowell, data sheet 909710
approx. 50 s, in water 0.2 m/s, Ø 9 mm
0.9
analog transmitter, 0 to 10 V output, data sheet 707030
programmable transmitter, 4 to 20 mA/20 to 4 mA output, data sheet 707010
Wtrans B programmable head transducer with radio transmission, data sheet 707060
(suitable Wtrans receivers, data sheet 902931)
Order specifications: Screw-in RTD temperature probe with form B connection head
(1) Basic type
902020/10
902020/11
Screw-in RTD temperature probes
with extension tube and continuous thermowell
Screw-in RTD temperature probes
with extension tube and stepped thermowell
(2) Operating temperature in °C
x x402-50 to +400 °C (standard)
x x415-50 to +600 °C
(3) Measuring insert
x x10011× Pt100 in 3-wire-circuit
x x10031× Pt100 in 2-wire circuit
x x10111× Pt100 in 4-wire circuit
x x20012× Pt100 in 3-wire circuit
x x20032× Pt100 in 2-wire circuit
(4) Tolerance class according to DIN EN 60751
x x1Class B (standard)
x x2Class A
(5) Thermowell diameter D in mm
x9Ø9mm
x11Ø11mm
x12Ø 12 mm stepped down to 9 mm
(6) Fitting length EL in mm (EL 100 ... 1000 mm, EL ≤ 700 mm for basic type 902020/11)
x x160160 mm
x220220 mm
x x250250 mm
x280280 mm
x x400400 mm
x x...Specification in plain text (50 mm steps)
(7) Process connection
x x104Screw-connection G 1/2
x x105Screw-connection G 3/4
x x106Screw connection G 1
(8) Extra codes
x x000no extra code
x x306Extension tube 70 mm
x x320terminal head Form BUZ
x x321terminal head Form BUZH
x x322Connection head form BBK
x x3301× analog transmitter, 4 to 20 mA
x x3311× transmitter programmable output 4 to 20 mA/20 to 4 mA
x x3331× analog transmitter, 0 to 10V
x x3342× analog transmitters, 4 to 20 mA
x x3352× transmitter programmable output 4 to 20 mA/20 to 4 mA
x x3372× analog transmitters, 0 to 10 V
2
output, data sheet 707030
2
output, data sheet 707030
2
output, data sheet 707030
2
output, data sheet 707030
3
, data sheet 707010
3
, data sheet 707010
x x8591× Wtrans B programmable head transducer with radio transmission, data sheet 707060
(1)(2)(3)(4)(5)(6)(7)(8)
Order code
Order example
1
List extra codes in sequence, separated by commas.
2
Specify measuring range in plain text.
3
Specify measuring range and output signal in plain text.
Order specifications: Screw-in RTD temperature probe with connection head form B
(1) Basic type
902020/20
902020/21
Screw-in RTD temperature probes
without extension tube and continuous thermowell
Screw-in RTD temperature probes
without extension tube and stepped thermowell
(2) Operating temperature in °C
x x402-50 to +400 °C (standard)
x x415-50 to +600 °C
(3) Measuring insert
x x10011× Pt100 in 3-wire-circuit
x x10031× Pt100 in 2-wire circuit
x x10111× Pt100 in 4-wire circuit
x x20012× Pt100 in 3-wire circuit
x x20032× Pt100 in 2-wire circuit
(4) Tolerance class according to DIN EN 60751
x x1Class B (standard)
x x2Class A
(5) Thermowell diameter D in mm
x9Ø9mm
x11Ø11mm
x12Ø 12 mm stepped down to 9 mm
(6) Fitting length EL in mm (EL 100 ... 1000 mm, EL ≤ 700 mm for basic type 902020/21)
x x100100 mm
x x160160 mm
x x250250 mm
x x400400 mm
x x...Specification in plain text (50 mm steps)
(7) Process connection
x x104Screw-connection G 1/2
x x105Screw-connection G 3/4
x x106Screw connection G 1
(8) Extra codes
x x000no extra code
x x320terminal head Form BUZ
x x321terminal head Form BUZH
x x322Connection head form BBK
x x3301× analog transmitter, 4 to 20 mA
x x3311× transmitter programmable output 4 to 20 mA/20 to 4 mA
x x3331× analog transmitter, 0 to 10 V
x x3342× analog transmitters, 4 to 20 mA
x x3352× transmitter, programmable, output 4 to 20 mA/20 to 4 mA
x x3372× analog transmitters, 0 to 10 V
2
output, data sheet 707030
2
output, data sheet 707030
2
output, data sheet 707030
2
output, data sheet 707030
3
, data sheet 707010
3
, data sheet 707010
x x8591× Wtrans B programmable head transducer with radio transmission, data sheet 707060
Order code------/ ,...
(1)(2)(3)(4)(5)(6)(7)(8)
Order example902020/20-402-1001-1-9-160-104/000
1
List extra codes in sequence, separated by commas.
2
Specify measuring range in plain text.
3
Specify measuring range and output signal in plain text.
902020/10-402-1003-1-9-160-104/00000055562
902020/10-402-1003-1-9-250-104/00000055563
902020/10-402-1003-1-9-400-104/00000055564
902020/10-402-2003-1-9-160-104/00000055565
902020/10-402-2003-1-9-250-104/00000055566
902020/10-402-2003-1-9-400-104/00000055567
902020/10-402-1001-1-9-160-104/330 (0 to 100 °C)00054616
902020/10-402-1001-1-9-160-104/330 (0 to 200 °C)00087522
902020/10-402-1001-1-9-250-104/330 (0 to 100 °C)00054617
902020/10-402-1001-1-9-250-104/330 (0 to 200 °C)00087527
902020/10-402-1001-1-9-400-104/331 (0 to 100 °C)00335111
Accessories for programmable 2-wire transmitter type 707010
in the accessories data sheet 709770
PartPart no.
PC interface with TTL/RS232 converter and adapter (socket connector)00350260
PC interface with USB/TTL converter, adapter (socket connector) and adapter (pins)00456352
Special programming (standard measuring range 0 to 100 °C)Setup program on CD-ROM, multilingual00378733
Accessories for Wtrans B, programmable head transmitter with radio transmission
Type 707060
PartPart no.
Lithium battery 3.6 V, 2.2 Ah (size AA)00547559
PC interface with USB/TTL converter, adapter (socket) and adapter (pins)00456352
PC interface with TTL/RS232 converter and adapter (socket connector)00350260
Setup program on CD-ROM, multilingual00488887
Setup program including OnlineChart on CD-ROM, multilingual00549067
OnlineChart activation00549188
2012-08-16/00311468