Push-in thermocouples with terminal head form A
For temperatures from -200 to +1600 °C
With thermowells made of various materials
As single or double thermocouple
Available with transmitter
Push-in thermocouples are preferentially used for measuring temperatures in liquids and gases. The application areas are, among others, in the HVAC, kiln and apparatus engineering sector.
The connection head form A is suitable for ambient temperatures up to +100 °C. The support tube is made of steel. When using a counter flange, see data sheet 90.9750 (90.9725), the installation point can be sealed up to 1 bar.
Sheaths made of various ceramic materials and heat resistant steels protect the measuring insert against chemical influences and mechanical damage. The selection of a suitable sheath material depends on the conditions on site.
Thermocouples (elements) as per DIN EN 60584 and DIN 43710, class 2, are set into the measuring insert. Versions with two thermocouples are also possible.
Data Sheet 90.1110 (90.1101) Page 1/4
Technical data
Terminal head Form A DIN 43729, die-cast aluminum, M 20x1.5; IP54, ambient temperature -40 to +100 °C
Process connection Sheath screw-connection G 1
Sheath Steel 1.4749, Ø 22 mm, operating temperature up to +1150 °C (seamless)
Measuring insert Insulated structure:
Transmitter Programmable transmitter, 4 to 20 mA/20 to 4 mA output, data sheet 70.7010
Caution: reduced ambient temperature when using connection head transmitters, data sheets
70.7010 and 70.7060
Stop flange Ø 22 mm, DIN 43734 Stop flange Ø 32 mm, DIN 43734
Steel 1.4841, Ø 22 mm, operating temperature up to +1150 °C (seamless) Ceramic C 610 (KER 610), Ø 15 mm, Ø 24 mm, operating temperatures up to +1300 °C Ceramic C 799 (KER 710), Ø 15 mm, Ø 24 mm, operating temperatures up to +1600 °C
1x Fe-CuNi „J“, DIN EN 60584, cl. 2, operating temperature -200 to +600 °C 1x Fe-CuNi „L“, DIN 43710, cl. 2, operating temperature -200 to +600 °C 1x NiCr-Ni „K“, DIN EN 60584, cl. 2, operating temperature -200 to +1200 °C 1x Pt10Rh-Pt „S“, DIN EN 60584, cl. 2, operating temperature 0 to 1300 °C 1x Pt30Rh-Pt6Rh „B“, DIN EN 60584, cl. 2, operating temperature 600 to 1600 °C 2x Fe-CuNi „J“, DIN EN 60584, cl. 2, operating temperature -200 to +600 °C 2x Fe-CuNi „L“, DIN 43710, cl. 2, operating temperature -200 to +600 °C 2x NiCr-Ni „K“, DIN EN 60584, cl. 2, operating temperature -200 to +1200 °C 2x Pt10Rh-Pt „S“, DIN EN 60584, cl. 2, operating temperature 0 to 1300 °C 2x Pt30Rh-Pt6Rh „B“, DIN EN 60584, cl. 2, operating temperature 600 to 1600 °C
Wtrans B programmable head transducer with radio transmission, data sheet 70.7060 (suitable Wtrans receivers, data sheet 90.2931)
2011-03-17/00314543
JUMO GmbH & Co. KG
Delivery address:Mackenrodtstraße 14,
36039 Fulda, Germany
Postal address:
36035 Fulda, Germany Phone: +49 661 6003-0 Fax: +49 661 6003-607 e-mail: mail@jumo.net Internet: www.jumo.net
JUMO Instrument Co. Ltd.
JUMO House Temple Bank, Riverway Harlow, Essex CM20 2DY, UK Phone: +44 1279 635533 Fax: +44 1279 635262 e-mail: sales@jumo.co.uk Internet: www.jumo.co.uk
JUMO Process Control, Inc.
8 Technology Boulevard Canastota, NY 13032, USA Phone: 315-697-JUMO
1-800-554-JUMO Fax: 315-697-5867 e-mail: info@jumo.us Internet: www.jumo.us
Dimensions
Data Sheet 90.1110 (90.1101) Page 2/4
Basic type 901110/10 Basic type 901110/20
2011-03-17/00314543