Product data sheet 1/2
ME-16G Visco Bath
Visco Bath for internal temperature applications
JULABO Visco Baths are used for precise temperature applications with
capillary viscometers and density meters within the bath tank. The units offer
high temperature accuracy and optimal temperature distribution throughout
the bath. An integral cooling coil for cooling water is provided for temperature
applications below or near ambient temperatures. The units are equipped
with a holder for thermometers and therefore offer very good options for
temperature calibration procedures and comparisons with accredited
thermometers.
Your advantages
VFD COMFORT DISPLAY
Keypad for setpoints, warning/safety values and menu functions
PID3 cascade temperature control
ATC3 3-Point-Calibration
Pt100 External sensor connection for measurement and control
SMART PUMP, electronically adjustable pump stages
Early warning system for low liquid level (DBGM 203 06 059.8)
Adjustable high temperature cut-out, visible via display
RS232 interface for online communication
Integrated programmer for 10 program steps
ME-16G
Technical Data
Order No. 9162616
Model series TopTech
Category Visco Baths
Working temperature range (°C) +20 ... +100
Temperature stability (°C) ±0.01
Setting / display resolution 0.01 °C
Integrated programmer 1x10 steps
Temperature Display VFD
Heating capacity (kW) 2
Pump capacity flow rate (l/min) 11-16
Pump capacity flow pressure (psi) 3.3-6.5
Bath opening / bath depth (W x L / D inch) 3 x 3 / 0.8 / 12.2
Filling volume liters 16
External Pt100 sensor connection integrated
Digital interface RS232
Optional: Profibus
Ambient temperature 5...40 °C
Dimensions W x L x H (inch) Ø 11.4 x 18.9
Product data sheet 2/2
Weight (LBS) 20
Classification according to DIN12876-1 Classification III (FL)
Cooling coil integrated
Power requirement V / Hz / A 115/60/9
Number of viscometers 2
Available voltage versions 208 - 230 V / 50 - 60 Hz
100 - 115 V / 50 - 60 Hz
Characteristics
ME-16G
Display
A perfect view
Ample, easy to read VFD Comfort
display for simultaneous display of 3
values, warning functions, high
temperature cut-off, pump stages
(resolution 0.01 °C)
Pump stage and liquid level
Backlit indicator for selected pump
stages and filling volume on Presto®
PLUS, Magnum 91 & Forte HT
Operation
Comfortable and detailed
Comfortable keypad with additional
menu functions for pump stages,
calibration, control parameters,
programmer, warnings, etc.
Temperature Control
For higher demands
PID Temperature control with drift
compensation and adjustable
parameters, improved temperature
stability for external applications,
temperature stability ±0.01 °C
internal, <±0.1 °C external
100 % Cooling capacity
‘Active Cooling Control’ for cooling
available throughout the entire
working temperature range, fast
cool-down even at higher
temperatures
Technical Features
Clever pump system
Reliable and consistent pump
capacity, electronically adjustable
pump stages
Control from the external
application
External Pt100 sensor connection
for precise measurement and
control directly in the external
application
Serial connection
RS232 interface for PC connection,
e.g. for data communication and
recording of measured values
Easy program control
Integrated programmer for the
execution of time and temperature
dependant profiles, 1 temperature
profile with 10 steps max., with real
time clock
Enhanced protective functions
Maximum safety, adjustable high
temperature cut-off or dry-running
protection, additional display of
setpoints permits easy and precise
adjustments
For flammable bath fluids
Classification III (FL) according to
DIN 12876-1
Powered by TCPDF (www.tcpdf.org)
Highest measuring accuracy
‘Absolute Temperature Calibration’
for manual compensation of a
temperature difference, 3-point
calibration
Refrigeration Technology
Consistent cooling capacity
Easily removable venting grid for
quick and easy cleaning
Refer to www.julabo.us for more information regarding the entire JULABO product portfolio.
Technical changes without prior notification. Images may deviate from the original.
Warning & Safety Functions
Early warning system for low
liquid level
Maximum safety for applications,
optical and audible alarm, allows
user to refill bath fluid before the unit
shuts down
Early warning system for high/low
temperature limits
Maximum safety for applications,
optical and audible alarm,
convertible to automated cut-off
function