Temperature control devices for research and industry are developed,
9001
Congratulations!
You have made an excellent choice.
JULABO thanks you for the trust you have placed in us.
This operating manual has been designed to help you gain an understanding of the operation
and possible applications of our circulators. For optimal utilization of all functions, we
recommend that you thoroughly study this manual prior to beginning operation.
The JULABO Quality Management System
produced, and distributed according to the requirements of ISO
and ISO 14001. Certificate Registration No. 01 100044846
Unpacking and inspecting
Unpack the circulator and accessories and inspect them for possible transport damage.
Damage should be reported to the responsible carrier, railway, or postal authority, and a
damage report should be requested. These instructions must be followed fully for us to
guarantee our full support of your claim for protecting against loss from concealed damage.
The form required for filing such a claim will be provided by the carrier.
Printed in Germany Changes without prior notification reserved
Important: keep original operating manual for future use
13. JULABO Service – Online remote diagnosis ....................................................... 51
14. Cleaning / repairing the unit ................................................................................. 52
4
MA
JULABO circulators are not suitable for direct temperature control of foods,
bath medium (bath fluid).
PID2
ATC
3
RS232
PUMP
SMART
temperatures to ensure an accurate temperature pattern at the selected spot
independent from the control circuit. This protection can be
The pump capacity (electronically adjustable via the motor speed) enables to
Operating manual
1. Intended use
JULABO circulators have been designed to control the temperature of specific fluids in a bath
tank.
The units feature pump connections for temperature control of external systems (loop circuit).
semi-luxury foods and tobacco, or pharmaceutical and medical products.
Direct temperature control means unprotected contact of the object with the
1.1. Description
The circulators are operated via the splash-proof keypad. The implemented
microprocessor technology allows to set and to store different values that
can be indicated on the MULTI-DISPLAY (LED). Three menu keys facilitate
adjusting setpoints, warning and safety functions and menu functions.
The PID temperature control adapts the heat supplied to the thermal
requirements of the bath.
Absolute Temperature Calibration (ATC3) provides a high temperature
stability in the bath. With the 3-point calibration an offset is adjusted at three
in the bath over the full temperature range.
Electrical connections:
The serial interface RS232 allows modern process technology without
additional interface.
Alarm output for external alarm message or control of JULABO refrigerating
baths or solenoid valve (cooling water).
The excess temperature protection conforming to IEC 61010-2-010 is a
safety installation
indicated and set on the MULTI-DISPLAY (LED).
The early warning system for low level signals that bath fluid needs to be
refilled before the low level protection conforming to IEC 61010-2-010
causes a complete shut-down of the main functional elements.
adapt to varying conditions for internal and external temperature
applications.
The circulator conforms to the relevant requirements specified by European
The products of JULABO ensure safe operation when installed, operated, and maintained
according to common safety regulations. This section explains the potential dangers that may
arise when operating the circulator and also specifies the most important safety precautions to
preclude these dangers as far as possible.
The operator is responsible for the qualification of the personnel operating the units.
The personnel operating the units should be regularly instructed about the dangers
involved with their job activities as well as measures to avert these dangers.
Make sure all persons tasked with operating, installing, and maintaining the unit have read
and understand the safety information and operating instructions.
When using hazardous materials or materials that could become hazardous, the circulator
may be operated only by persons who are absolutely familiar with these materials and the
circulator. These persons must be fully aware of possible risks.
If you have any questions concerning the operation of your unit or the information in this
manual, please contact us!
You have received a product designed for industrial use. Nevertheless, avoid strikes to the
housing, vibrations, damage to the operating-element panel (keypad, display), and
contamination.
Make sure the product is checked for proper condition regularly (depending on the
conditions of use). Regularly check (at least every 2 years) the proper condition of the
mandatory, warning, prohibition and safety labels.
Make sure that the mains power supply has low impedance to avoid any negative effects
on instruments being operated on the same mains.
This unit is designed for operation in a controlled electromagnetic environment. This
means that transmitting devices (e.g., cellular phones) should not be used in the
immediate vicinity.
Magnetic radiation may affect other devices with components sensitive to magnetic fields
(e.g., monitors). We recommend maintaining a minimum distance of 1 m.
Permissible ambient temperature: max. 40 °C, min. 5 °C.
Permissible relative humidity: 50% (40 °C).
Do not store the unit in an aggressive atmosphere.
Protect the unit from contamination.
Do not expose the unit to sunlight.
6
MA
Appropriate operation
Only qualified personnel is authorized to perform configuration, installation, maintenance and
repairs of the circulator.
Routine operation can also be carried out by untrained personnel who should however be
instructed by trained personnel.
Use:
The bath can be filled with flammable materials. Fire hazard!
There might be chemical dangers depending on the bath medium used.
Observe all warnings for the used materials (bath fluids) and the respective instructions
(safety data sheets).
Insufficient ventilation may result in the formation of explosive mixtures. Only use the unit in
well ventilated areas.
Only use recommended materials (bath fluids). Only use non-acid and non corroding
materials.
When using hazardous materials or materials that could become hazardous, the operator
must affix the enclosed safety labels (1 + 2) to the front of the unit so they are highly visible:
1
2
or
2
Particular care and attention is necessary because of the wide operating range.
There are thermal dangers: Burn, scald, hot steam, hot parts and surfaces that can be
touched.
Observe the instructions in the manuals for instruments of a different make that you connect to
the circulator, particularly the respective safety recommendations. Also observe the pin
assignment of plugs and technical specifications of the products.
Warning label W00: Colors: yellow, black
Danger area. Attention! Observe instructions.
(operating manual, safety data sheet)
Mandatory label M018: Colors: blue, white
Carefully read the user information prior to beginning operation.
Scope: EU
Semi S1-0701 Table A1-2 #9
Carefully read the user information prior to beginning operation.
Scope: USA, NAFTA
Warning label W26: Colors: yellow, black
Hot surface warning.
(The label is put on by JULABO)
7
Operator responsibility – Safety recommendations
municipal waste is not permitted!
2.1. Disposal
The circulator contains a back-up battery that supplies voltage to memory chips when the unit
is switched off. Do not dispose of the battery with household waste!
Depending on battery regulations in your country, you might be obliged to give back used or
defect batteries to gathering places.
The product may be used with oil as bath fluid. These oils fully or partially consist of mineral oil
or synthetic oil. For disposal, observe the instructions in the safety data sheets.
Valid in EU countries
See the current official journal of the European Union – WEEE directive.
Directive of the European Parliament and of the Council on waste electrical
and electronic equipment (WEEE).
This directive requires electrical and electronic equipment marked with a
crossed-out trash can to be disposed of separately in an environmentally
friendly manner.
Contact an authorized waste management company in your country.
Disposal with household waste (unsorted waste) or similar collections of
2.2. Warranty conditions
JULABO GmbH warrants its products against defects in material or in workmanship, when
used under appropriate conditions and in accordance with appropriate operating instructions
for a period of ONE YEAR.
Extension of the warranty period – free of charge
With the ‘1PLUS warranty’ the user receives a free of charge extension to the warranty of up
to 24 months, limited to a maximum of 10 000 working hours.
To apply for this extended warranty the user must register the unit on the JULABO web site
www.julabo.com, indicating the serial no. The extended warranty will apply from the date of
JULABO GmbH’s original invoice.
JULABO GmbH reserves the right to decide the validity of any warranty claim. In case of faults
arising either due to faulty materials or workmanship, parts will be repaired or replaced free of
charge, or a new replacement unit will be supplied.
Any other compensation claims are excluded from this guarantee.
8
2.3. EC Conformity
MA
9
Operator responsibility – Safety recommendations
10
2.4. Technical specifications
Open Bath Circulators
MA-4
MA-6
MA-12
MA-26
Working temperature range
°C
20 ... 200
20 ... 200
20 ... 200
20 ... 200
Temperature stability
°C
±0,01
±0,01
±0,01
±0,01
Bath opening (WxL)
cm
13x15
13x15
22x15
22x30
Bath depth
cm
15
20
20
20
Filling volume from ... to
liters
3 ... 4,5
4,5 ... 6
7,5 ... 12
18 ... 26
Overall dimensions (WxDxH)
cm
21x42x38
21x43x42
30x43x45
36x61x45
Weight
kg
9,6
12,5
13,0
26,0
Mains power connection
V; Hz
208 - 230; 50/60 (perm. range)
Current draw (at 208V)
A 8 Current draw (at 230V)
A 9 Mains power connection
V; Hz
100 - 115; 50/60 (perm. range)
Current draw (at 100V)
A 8 Current draw (at 115V)
A
9 MA
Temperature selection
digital
via keypad indication on
MULTI-DISPLAY(LED)
remote control via personal computer
indication on monitor
Temperature indication
MULTI-DISPLAY (LED)
Resolution (-9.99 …. +99.99 = 0.01)
°C
0.01 / 0.1
Absolute Temperature Calibration
°C
±3 Temperature control
PID
Heater wattage (at 230 V)
kW
2,0
Electronically adj. pump capacity
stages
1 ... 4
Flow rate max.at 0 bar
l/min
11 ... 16
Pressure max. max. at 0 l
bar
0.23 ... 0.45
Electrical connections:
External alarm device
Vdc/mA
24-0 / max. 25
Computer interface
RS232
Ambient temperature
°C
5 ... 40
All measurements have been carried out at: rated voltage and frequency
operating temperature: 70 °C ambient temperature: 20 °C bath fluid: water
Technical changes without prior notification reserved.
MA
Heater wattage (at 115V) kW 1,0
11
Operator responsibility – Safety recommendations
Caution:
Safety installations according to IEC 61010-2-010:
Excess temperature protection adjustable from 0 °C ... 230 °C
Low liquid level protection float switch
Classification according to DIN 12876-1 class III
Supplementary safety installations
Early warning system for low level float switch
High temperature warning function optical + audible (in intervals)
Low temperature warning function optical + audible (in intervals)
Supervision of working sensor plausibility control
Reciprocal sensor monitoring between
working and safety sensors difference >35 K
Alarm message optical + audible (permanent)
Warning message optical + audible (in intervals)
Environmental conditions according to IEC 61 010-1:
Use only indoor.
Altitude up to 2000 m - normal zero.
Ambient temperature: +5 ... +40 °C
Air humidity:
Max. rel. humidity 80 % for temperatures up to +31 °C,
linear decrease down to 50 % relative humidity at a temperature of +40 °C
Max. mains fluctuations of ±10 % are permissible.
Protection class according to IEC 60 529 IP21
The unit corresponds to Class I
Overvoltage category II
Pollution degree 2
Standards for interference resistance according to EN 61326-1
This unit is an ISM device classified in Group 1 (using high frequency for internal purposes)
Class A (industrial and commercial range).
The unit is not for use in explosive environment
12
In addition to the safety warnings listed, warnings are posted throughout the
Read and follow these important instructions for averting dangers.
Warning:
Caution:
dangerous situation. If this is not avoided, slight or minor
Notice:
anything in its surroundings can be damaged.
Note!
Important!
This icon is used in the operating instructions to indicate flashing values or
Follow the safety instructions to avoid personal injury and property damage.
• Only connect the unit to a power socket with an earthing contact (PE –
material.
Operating instructions
3. Safety notes for the user
3.1. Explanation of safety notes
operating manual. These warnings are designated by an exclamation mark
inside an equilateral triangle. “Warning of a dangerous situation (Attention!
Please follow the documentation).”
The danger is classified using a signal word.
Describes a possibly highly dangerous situation. If these instructions are not
followed, serious injury and danger to life could result.
Describes a possibly
injuries could result. A warning of possible property damage may also be
contained in the text.
MA
Describes a possibly harmful situation. If this is not avoided, the product or
3.2. Explanation of other notes
Draws attention to something special.
3.3. Safety recommendations
Indicates usage tips and other useful information.
parameters which have to be set or confirmed.
Also, the valid safety instructions for workplaces must be followed.
protective earth)!
• The power supply plug serves as a safe disconnecting device from the line
and must always be easily accessible.
• Place the unit on an even surface on a base made of nonflammable
13
Safety notes for the user
• Do not stay in the area below the unit.
source before performing any service or maintenance procedures, or before
caution when touching these parts.
Caution:
• Make sure you read and understand all instructions and safety precautions
listed in this manual before installing or operating your unit.
• Set the excess temperature safety installation at least 25 °C below the fire
point of the bath fluid.
• Observe the limited working temperature range when using plastic bath
tanks.
• Never operate the unit without bath fluid in the bath.
• Pay attention to the thermal expansion of bath oil during heating to avoid
overflowing of the fluid.
• Prevent water from entering the hot bath oil.
• Do not drain the bath fluid while it is hot!
Check the temperature of the bath fluid prior to draining (e.g., by switching
the unit on for a short moment).
• Use suitable connecting tubing.
• Avoid sharp bends in the tubing, and maintain a sufficient distance from
surrounding walls.
• Make sure that the tubing is securely attached.
• Regularly check the tubing for material defects (e.g., for cracks).
• Never operate damaged or leaking units.
• Always turn off the unit and disconnect the mains cable from the power
moving the unit.
• Always turn off the unit and disconnect the mains cable from the power
source before cleaning the unit.
• Always empty the bath before moving the unit.
• Transport the unit with care.
• Sudden jolts or drops may cause damage in the interior of the unit.
• Observe all warning labels.
• Never remove warning labels.
• Never operate units with damaged mains power cables.
• Repairs are to be carried out only by qualified service personnel.
Some parts of the bath tank and the pump connections may become
extremely hot during continuous operation. Therefore, exercise particular
The temperature controlling i.e. of fluids in a reactor constitutes normal
circulator practice.
We do not know which substances are contained within these vessels.
Many substances are:
• inflammable, easily ignited or explosive
• hazardous to health
• environmentally unsafe
i.e.: dangerous
14
determine the method of decontamination.
screwdriver for example.
The user alone is responsible for the handling of these substances!
The following questions shall help to recognize possible dangers and to reduce
the risks to a minimum.
• Are all tubes and electrical cables connected and installed?
Note:
sharp edges, hot surfaces in operation, moving machine parts, etc.
• Do dangerous steams or gases arise when heating?
Is an exhaust needed when working?
• What to do when a dangerous substance was spilled on or in the unit?
Before starting to work, obtain information concerning the substance and
Notice: Check the safety installations at least twice a year!
• Excess temperature protection according to IEC 61010-2-010.
With a screwdriver turn back the adjustable excess temperature protection
until the shut-down point (actual temperature).
• Low level protection according to IEC 61010-2-010.
To check the function of the float, it can be manually lowered with a
MA
15
Operating controls and functional elements
4. Operating controls and functional elements
4.1. Circulator
Front view Rear view
1
Navigation keys
2
3
Mains power switch, illuminated
1. Key: >OK< Start / Stop (pump / heater )
2. >OK< in the menu Menu item / select submenu for setting
Save set value
Save selected parameter
A beep signals the end of setting
After the actions Start, Stop and change from VFD Display to
standard display the key is locked for a short time.
The above graph “front side” shows an example for standard display.
1. Key: >Return< Stop (pump / heater )
2. >Return< in the menu one menu level down
Correction function for parameters or values (prior to OK)
immediately back to standard display
16
- icon for „keep key pressed down“.
Loading...
+ 36 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.