Unpacking and inspecting
2
Unpack the recirculating cooler and accessories and check for damages incurred
during transit. These should be reported to the responsible carrier, railway, or postal
authority, and a request for a damage report should be made. These instructions
must be followed fully for us to guarantee our full support of your claim for protecting
against loss from concealed damage. The form required for filing such a claim will be
provided by the carrier.
EC Conformity
The products described in the operating instructions conform to the
requirements of the following European guidelines:
Directive of the European Parliament and of the Council on the approximation of the laws of
the Member States relating to machinery.
Low voltage regulations with respect to legal harmonization of the member countries
concerning electric devices for use within certain voltage limits.
EMC guideline with respect to legal harmonization of the member countries concerning
JULABO recirculating coolers have been designed for temperature application to specific fluids. The
pump connections can be used for cooling applications in an external circuit at a constant
temperature.
JULABO water baths are not suitable for direct temperature control of foods, semiluxury foods and tobacco, or pharmaceutical and medical products. Direct temperature
control means unprotected contact of the object with the bath medium (bath fluid).
1.1. Description
The recirculating coolers are operated via the splash-proof keypad. The
implemented microprocessor technology allows to set and to store the setpoint that
PID1
can be indicated on the LED temperature display.
The PID temperature regulation is used to withdraw heat from the bath fluid by
means of the cooling machine and to automatically regulate the required need.
Electrical connections:
1. The serial interface RS232 allows modern process technology without
additional interface.
2. Alarm output for external alarm message.
RS232
2. Operator responsibility – Safety instructions
The products of JULABO ensure safe operation when installed, operated, and maintained according
to common safety regulations. This section explains the potential dangers that may arise when
operating the circulator and also specifies the most important safety precautions to preclude these
dangers as far as possible.
6
The operator is responsible for the qualification of the personnel operating the units.
The personnel operating the units should be regularly instructed about the dangers involved with
their job activities as well as measures to avert these dangers.
Make sure all persons tasked with operating, installing, and maintaining the unit have read and
understand the safety information and operating instructions.
When using hazardous materials or materials that could become hazardous, the unit may be
operated only by persons who are absolutely familiar with these materials and the unit. These
persons must be fully aware of possible risks.
If you have any questions concerning the operation of your unit or the information in this manual,
please contact us!
Contact JULABO GmbH
Eisenbahnstraße 45
77960 Seelbach / Germany
Safety recommendations for the operator
You received a product conceived for industrial use. Nevertheless, avoid strikes to the housing,
vibrations, damages to the keypad foil (keys, display) or contamination.
Make sure the product is regularly checked for proper condition. Regularly check (at least every
2 years) the proper condition of the mandatory, warning, prohibition and safety labels.
Take care that the mains supply features a low impedance to avoid any negative affects on the
instrument being operated in the same mains.
This unit is designed for operation in a controlled electromagnetic environment. This means that
transmitting devices (e.g. cellular phones) should not be used in the immediate vicinity. Magnetic
radiation may influence other units with components susceptible to magnetic fields (e.g. a
monitor). We recommend to keep a minimum distance of 1 m.
Permissible ambient temperature: max. 40 °C, min. 5 °C.
Permissible relative air humidity: 50 % (40 °C).
Do not store in an aggressive atmosphere. Protect from contaminations.
Do not expose to sunlight.
Appropriate Operation
Only qualified personnel is authorized to perform configuration, installation, maintenance and repairs
of the water bath.
Routine operation can also be carried out by untrained personnel who should however be instructed
by trained personnel.
Use
For the use according to the intended purpose, special material requirements have to be respected
(bath fluids). Only use non-acid and non corroding materials.
Observe all warnings for the used materials (bath fluids) and the respective instructions (safety data
sheets). Only use the unit in well ventilated areas. (see page 20).
The recirculating coolers are not for use in explosive atmosphere
8
When using hazardous materials or materials that could become hazardous, the operator must affix
the enclosed safety labels (1 + 2) to the front of the unit so they are highly visible:
1
2
or
2
Observe the instructions in the manuals for instruments of a different make that you connect to the
circulator, particularly the corresponding safety instructions. Also observe the pin assignment of
plugs and technical specifications of the products.
Warning label W00: Colors: yellow, black
Danger area. Attention! Observe instructions.
(operating manual, safety data sheet)
Mandatory label M018: Colors: blue, white
Carefully read the user information prior to beginning operation.
Scope: EU
Semi S1-0701 Table A1-2 #9
Carefully read the user information prior to beginning operation.
Scope: USA, NAFTA
9
2.1. Disposal
This unit contains the refrigerants R134a or R404A, which at this time are not considered harmful to
the ozone layer. However, over the long operating period of the unit, disposal rules may change.
Therefore, only qualified personnel should handle the disposal.
Valid in EU countries
See the current official journal of the European Union – WEEE directive.
Directive of the European Parliament and of the Council on waste electrical and
electronic equipment (WEEE).
This directive requires electrical and electronic equipment marked with a crossedout trash can to be disposed of separately in an environmentally friendly manner.
Contact an authorized waste management company in your country.
Disposal with household waste (unsorted waste) or similar collections of municipal
waste is not permitted!
10
3. Technical specifications
Recirculating Cooler FL300
Working temperature range °C -20 ... +40
Temperature stability
Temperature selection: digital
via key pad indication on LED-DISPLAY
remote control via personal computer
Temperature indication: LED-DISPLAY
Resolution °C 0.1
Temperature control
Temperature sensor
Excess temperature protection
Low liquid level protection float switch
Electrical connections:
°C ±0.5
indication on monitor
PID 1
Pt 100
85 °C - fixed value
Computer interface RS232
Alarm output for external alarm signal
Circulating pump: discharge, max.at 0 bar l/min 15
pressure, max. at 0 Liters
bar 0,35
11
Recirculating Cooler FL300
bar
Filling level indicator sight glass
Filling volume from ... toLiters 3 ... 4.5
Dimensions (WxLxH) cm 25x50x60
Weight
Ambient temperature range°C 5 ... 40
Return flow temperature°C 80 max.
IP class according to IEC 60 529 IP 21
Cooling compressor
Refrigerant
Cooling capacity (at A, B, C) °C +20 +10 0 -10 -20
Medium: Mixture water-glycol kW 0.3 0.25 0.2 0.15 0.1
A Mains power connection 230 V/50 Hz V/ Hz 207-253 / 50
kg 39
1- stage / air cooled
R134a
Current input at 230 V A 3.0
B Mains power connection 208 - 220 V/60 Hz V/ Hz 197 - 242 / 60
Current input at 208V / 220 V A 3.0 / 3.0
C Mains power connection 115 V/60 Hz V/ Hz 103-127 / 60
Current input at 115 V A 6.0
D Mains power connection 200 V/50-60 Hz V/ Hz 190-254 / 50 190-244 / 60
Current input at 200 V A 3.0
All measurements have been carried out at: rated voltage and frequency,
ambient temperature: 20 °C
13
Recirculating Cooler FL601
Working temperature range °C -20 ... +40
Temperature stability
Temperature selection: digital
via key pad indication on LED-DISPLAY
remote control via personal computer
Temperature indication: LED-DISPLAY
Resolution °C 0.1
Temperature control
Temperature sensor
Excess temperature protection
Electrical connections:
Computer interface RS232
Alarm output for external alarm signal
14
Recirculating Cooler FL601
Circulating pump: discharge, max.at 0 bar l/min 23
pressure, max. at 0 Liters
bar -----
Filling level indicator sight glass
Filling volume from ... toLiters 5.5 ... 8
Dimensions (WxLxH) cm 32x50x60
Weight
Ambient temperature range°C 5 ... 40
Return flow temperature°C 80 max.
IP class according to IEC 60 529 IP 21
Mains power connection 230 V/50 Hz V/ Hz 207-253 / 50
Current input at 230 V A 4.6
Mains power connection 208 - 220 V/60 Hz V/ Hz 197 - 242 / 60
Current input at 208V / 220 V A 5.0
bar 1.0
kg 48
Mains power connection 115 V/60 Hz V/ Hz 103-127 / 60
Current input at 115 V A 9.1
All measurements have been carried out at: rated voltage and frequency,
ambient temperature: 20 °C
15