Julabo FL2503, FLW2503, FL2506, FLW2506, FL4003 Operating manual

...
English
2
A
Operating manual
Recirculating Coolers
FL2503 FLW2503 FL2506 FLW2506
FL4003 FLW4003 FL4006 FLW4006
FL2506
°C
LARM
JULABO USA, Inc. 884 Marcon Boulevard Allentown, PA 18109 Phone: +1(610) 231-0250 Fax: +1(610) 231-0260 info @ julabo.com
19534829-V1.doc Print date: 02.01.2013
www.julabo.com
Congratulations!
You have made an excellent choice. JULABO thanks you for the trust you have placed in us. This operating manual has been designed to help you gain an understanding of the operation and possible applications of our circulators. For optimal utilization of all functions, we recommend that you thoroughly study this manual prior to beginning operation.
The JULABO Quality Management System
Temperature control devices for research and industry are developed, produced, and distributed according to the requirements of
ISO 9001:2008. Certificate Registration No. 01 100044846
Unpacking and inspecting
1.953.4829-V1 01/13 Printed in Germany Changes without prior notification reserved
Unpack the recirculating cooler and accessories and check for damages incurred during transit. These should be reported to the responsible carrier, railway, or postal authority, and a request for a damage report should be made. These instructions must be followed fully for us to guarantee our full support of your claim for protecting against loss from concealed damage. The form required for filing such a claim will be provided by the carrier.
Lifting device for transportation by crane see page 18
Important: keep operating manual for future use
2
TABLE OF CONTENTS
1. Intended use ................................................................................................................................. 4
1.1. Description ........................................................................................................................ 4
2. Operator responsibility – Safety instructions ................................................................................ 4
2.1. Disposal ............................................................................................................................ 6
3. Technical specifications ................................................................................................................ 7
3.1. Cooling water connection ................................................................................................ 12
4. Safety notes for the user ............................................................................................................ 13
4.1. Explanation of safety notes ............................................................................................. 13
4.2. Explanation of other notes .............................................................................................. 13
4.3. Safety instructions ........................................................................................................... 13
5. Installation .................................................................................................................................. 15
5.1. Tubing ............................................................................................................................. 17
6. Operating controls and functional elements ............................................................................... 18
7. Operating procedures ................................................................................................................. 20
7.1. Bath fluids ....................................................................................................................... 20
7.2. Power connection ........................................................................................................... 20
7.3. Filling ............................................................................................................................... 21
7.4. Switching on / Start - Stop .............................................................................................. 21
7.5. Setting the feed pressure ................................................................................................ 22
7.6. Setting the temperatures ................................................................................................. 22
7.7.
7.8. Remote control: activate – deactivate ............................................................................. 23
8. Safety installations...................................................................................................................... 23
8.1. Excess temperature protection ....................................................................................... 23
8.2. Low level protection ........................................................................................................ 23
9. Troubleshooting guide / Error messages .................................................................................... 24
10. Electrical connections ....................................................................................................... 26
11. Remote control .................................................................................................................. 27
11.1. Setup for remote control ................................................................................................. 27
11.2. Communication with a PC or a superordinated data system .......................................... 27
11.3. List of commands ............................................................................................................ 28
11.4. Status messages ............................................................................................................ 28
11.5. Error messages ............................................................................................................... 29
12. Cleaning / repairing the unit .............................................................................................. 30
12.1. JULABO Service – Online remote diagnosis .................................................................. 32
12.2. Draining ........................................................................................................................... 33
13. Adequate storing of operating manual .............................................................................. 33
14. WARRANTY PROVISIONS .............................................................................................. 34
AUTOSTART ON / OFF .................................................................................................. 22
3
1. Intended use
JULABO recirculating coolers have been designed for temperature application to specific fluids. The pump connections can be used for cooling applications in an external circuit at a constant temperature.
JULABO water baths are not suitable for direct temperature control of foods, semi­luxury foods and tobacco, or pharmaceutical and medical products. Direct temperature
control means unprotected contact of the object with the bath medium (bath fluid).
1.1. Description
The recirculating coolers are operated via the splash-proof keypad. The
implemented microprocessor technology allows to set and to store the setpoint that
PID1
RS232
can be indicated on the LED temperature display.
The PID temperature regulation is used to withdraw heat from the bath fluid by
means of the cooling machine and to automatically regulate the required need.
Electrical connections:
1. The serial interface RS232 allows modern process technology without additional interface.
2. Alarm output for external alarm message.
Manually adjustable by-pass (handwheel) to reduce the pump capacity
(e. g. for glass equipment).
2. Operator responsibility – Safety instructions
The products of JULABO ensure safe operation when installed, operated, and maintained according to common safety regulations. This section explains the potential dangers that may arise when operating the circulator and also specifies the most important safety precautions to preclude these dangers as far as possible.
The operator is responsible for the qualification of the personnel operating the units. The personnel operating the units should be regularly instructed about the dangers involved with
their job activities as well as measures to avert these dangers.
Make sure all persons tasked with operating, installing, and maintaining the unit have read and
understand the safety information and operating instructions.
When using hazardous materials or materials that could become hazardous, the unit may be
operated only by persons who are absolutely familiar with these materials and the unit. These persons must be fully aware of possible risks.
4
If you have any questions concerning the operation of your unit or the information in this manual, please contact us!
Contact
Safety recommendations for the operator
You received a product conceived for industrial use. Nevertheless, avoid strikes to the housing,
vibrations, damages to the keypad foil (keys, display) or contamination.
Make sure the product is regularly checked for proper condition. Regularly check (at least every
2 years) the proper condition of the mandatory, warning, prohibition and safety labels.
Take care that the mains supply features a low impedance to avoid any negative affects on the
instrument being operated in the same mains.
This unit is designed for operation in a controlled electromagnetic environment. This means that
transmitting devices (e.g. cellular phones) should not be used in the immediate vicinity. Magnetic radiation may influence other units with components susceptible to magnetic fields (e.g. a monitor). We recommend to keep a minimum distance of 1 m.
Permissible ambient temperature: max. 40 °C, min. 5 °C. Permissible relative air humidity: 50 % (40 °C). Do not store in an aggressive atmosphere. Protect from contaminations. Do not expose to sunlight.
JULABO USA, Inc. 884 Marcon Boulevard Allentown, PA 18109
Phone: +1(610) 231-0250 Fax: +1(610) 231-0260 info @ julabo.com www.julabo.com
Appropriate Operation
Only qualified personnel is authorized to perform configuration, installation, maintenance and repairs of the water bath. Routine operation can also be carried out by untrained personnel who should however be instructed by trained personnel.
Use
For the use according to the intended purpose, special material requirements have to be respected (bath fluids). Only use non-acid and non corroding materials. Observe all warnings for the used materials (bath fluids) and the respective instructions (safety data sheets). Only use the unit in well ventilated areas. (see page 14). The recirculating coolers are not for use in explosive atmosphere
When using hazardous materials or materials that could become hazardous, the operator must affix
the enclosed safety labels to the front of the unit so they are highly visible:
If this unit is intended for use within the United States of America, all 3 warning labels must be affixed
to the housing of the unit prior to use. Directions for the positioning of the individual warning labels are enclosed with the warning labels included in the delivery. Warning labels must be easily visible to users.
5
1
Warning label W00: Colors: yellow, black Danger area. Attention! Observe instructions. (operating manual, safety data sheet)
2
or 2
3
Observe the instructions in the manuals for instruments of a different make that you connect to the circulator, particularly the corresponding safety instructions. Also observe the pin assignment of plugs and technical specifications of the products.
Mandatory label M018: Colors: blue, white Carefully read the user information prior to beginning operation.
Scope: EU
Semi S1-0701 Table A1-2 #9 Carefully read the user information prior to beginning operation.
Scope: USA, NAFTA
Warning label Proposition 65
2.1. Disposal
This unit contains the refrigerant R404A, which at this time is not considered harmful to the ozone layer. However, over the long operating period of the unit, disposal rules may change. Therefore, only qualified personnel should handle the disposal.
6
3. Technical specifications
Recirculating Cooler
Cooling Working temperature range °C -20 ... +40 -20 ... +40 Temperature stability °C ±0.5 Temperature selection: digital via key pad indication on LED-DISPLAY remote control via personal computer indication on monitor Temperature indication: LED-DISPLAY Resolution °C 0.1 Temperature control PID 1 Temperature sensor Pt 100 Excess temperature protection 85 °C - fixed value Low liquid level protection float switch Cooling capacity Medium: Mixture water-glycol Cooling compressor 1-stage Refrigerant R404A
Electrical connections: Computer interface RS232 Alarm output for external alarm signal
Circulating pump: discharge, max. at 0 bar l/min 40 40 pressure, adjustable at 0 Liter bar 0.5 ... 3.0 0.5 ... 3.0 Feed pressure indication bar Manometer Manometer Filling level indicator sight glass Filling volume from ... to liters 24 … 30 Dimensions (WxLxH) cm 60x76x115 Weight kg 146 143 Ambient temperature range °C 5 ... 40 Return flow temperature °C 80 max. Cooling water Flow rate at 20 °C inlet temperature IP class according to IEC 60 529 IP 21
Mains power connection 230 V/50 Hz V/ Hz 207-253 / 50 207-253 / 50
Current input at 230 V A 11 12
Mains power connection 208 - 220 V/60 Hz V/ Hz 197 - 242 / 60 197 - 242 / 60 Current input at 208V / 220 V A 13 ----
All measurements have been carried out at: rated voltage and frequency, ambient temperature: 20 °C
°C kW
l/min
FL2503 FLW2503
air cooled water cooled
+20 0 -10 +20 0 -10
2.5 1.5 1.2 2.7 1.7 1.0
----- 4.1
7
Recirculating Cooler
Cooling Working temperature range °C -20 ... +40 -20 ... +40 Temperature stability °C ±0.5 Temperature selection: digital via key pad indication on LED-DISPLAY remote control via personal computer indication on monitor Temperature indication: LED-DISPLAY Resolution °C 0.1 Temperature control PID 1 Temperature sensor Pt 100 Excess temperature protection 85 °C - fixed value Low liquid level protection float switch Cooling capacity Medium: Mixture water-glycol Cooling compressor 1- stage Refrigerant R404A
Electrical connections: Computer interface RS232 Alarm output for external alarm signal
Circulating pump: discharge, max. at 0 bar l/min 60 60 pressure, adjustable at 0 Liter bar 0.5 ... 6.0 0.5 ... 6.0 Feed pressure indication bar Manometer Manometer Filling level indicator sight glass Filling volume from ... to liters 24 … 30 Dimensions (WxLxH) cm 60x76x115 Weight kg 158 160 Ambient temperature range °C 5 ... 40 Return flow temperature °C 80 max. Cooling water Flow rate at 20 °C inlet temperature IP class according to IEC 60 529 IP 21
Mains power connection 230 V/50 Hz V/ Hz 207-253 / 50 207-253 / 50 Current input at 230 V A 14 15 Mains power connection 208 - 220 V/60 Hz V/ Hz 197 - 242 / 60 197 - 242 / 60 Current input at 208V / 220 V A 16 ------
All measurements have been carried out at: rated voltage and frequency, ambient temperature: 20 °C
°C kW
l/min
FL2506 FLW2506
air cooled water cooled
+20 0 -10 +20 0 -10
2.5 1.0 0.3 2.5 1.0 0.3
----- 4.1
8
Recirculating Cooler
Cooling
FL4003 FLW4003
air cooled water cooled Working temperature range °C -20 ... +40 -20 ... +40 Temperature stability °C ±0.5 Temperature selection: digital via key pad indication on LED-DISPLAY remote control via personal computer indication on monitor Temperature indication: LED-DISPLAY Resolution °C 0.1 Temperature control PID 1 Temperature sensor Pt 100 Excess temperature protection 85 °C - fixed value Low liquid level protection float switch Cooling capacity Medium: Mixture water-glycol
°C kW
+20 0 -10 +20 0 -10
4.0 2.4 1.5 4.3 2.2 1.3 Cooling compressor 1- stage Refrigerant R404A
Electrical connections: Computer interface RS232 Alarm output for external alarm signal
Circulating pump: discharge, max. at 0 bar l/min 40 40 pressure, adjustable at 0 Liter bar 0.5 ... 3.0 0.5 ... 3.0 Feed pressure indication bar Manometer Manometer Filling level indicator sight glass Filling volume from ... to liters 24 ... 30 Dimensions (WxLxH) cm 60x76x115 Weight kg 148 143 Ambient temperature range °C 5 ... 40 Return flow temperature °C 80 max. Cooling water Flow rate at 20 °C inlet temperature
l/min ----- 6.6
IP class according to IEC 60 529 IP 21
Mains power connection 400 V/3PNPE/50 Hz
V/ Hz 360-440/3PNPE/50 360-440/3PNPE/50
Current input at 230 V A 8 8 Mains power connection 230 V/3PPE/60 Hz
V/ Hz 197-242/3PPE/60 207-253/3PPE/60
Current input at 208V / 220 V A 11 ------
All measurements have been carried out at: rated voltage and frequency, ambient temperature: 20 °C
9
Recirculating Cooler
Cooling
FL4006 FLW4006
air cooled water cooled Working temperature range °C -20 ... +40 -20 ... +40 Temperature stability °C ±0.5 Temperature selection: digital via key pad indication on LED-DISPLAY remote control via personal computer indication on monitor Temperature indication: LED-DISPLAY Resolution °C 0.1 Temperature control PID 1 Temperature sensor Pt 100 Excess temperature protection 85 °C - fixed value Low liquid level protection float switch Cooling capacity Medium: Mixture water-glycol
°C kW
+20 0 -10 +20 0 -10
4.0 1.9 0.9 4.0 1.7 0.7 Cooling compressor 1- stage Refrigerant R404A
Electrical connections: Computer interface RS232 Alarm output for external alarm signal
Circulating pump: discharge, max. at 0 bar l/min 60 60 pressure, max. at 0 liter bar 0.5 ... 6.0 0.5 ... 6.0 pressure, adjustable at 0 Liter Feed pressure indication bar Manometer Manometer Filling level indicator sight glass Filling volume from ... to liters 24 ... 30 Dimensions (WxLxH) cm 60x76x115 Weight kg 157 160 Ambient temperature range °C 5 ... 40 Return flow temperature °C 80 max. Cooling water Flow rate at 20 °C inlet temperature
l/min ----- 6.6
IP class according to IEC 60 529 IP 21
Mains power connection 400 V/3PNPE/50 Hz
V/ Hz 360-440/3PNPE/50 360-440/3PNPE/50
Current input at 230 V A 12 13 Mains power connection 230 V/3PPE/60 Hz
V/ Hz 197-242/3PPE/60 207-253/3PPE/60
Current input at 208V / 220 V A 16 ------
All measurements have been carried out at: rated voltage and frequency, ambient temperature: 20 °C
10
Warning functions and safety installations
Excess temperature protection 85 °C - fixed value Low liquid level protection float switch Alarm message optical + audible (permanent) Excess temperature - Warning function 75 °C Overload protection for compressor and pump motor Classification according to DIN 12876-1 class I
Environmental conditions according to IEC 61 010-1:
Use only indoor. Altitude up to 2000 m - normal zero. Ambient temperature: +5 ... +40 °C Air humidity: Max. rel. humidity 80 % for temperatures up to +31 °C, linear decrease down to 50 % relative humidity at a temperature of +40 °C Max. mains fluctuations of ±10 % are permissible.
The unit corresponds to Class I
Overvoltage category II Pollution degree 2
Caution:
The unit is not for use in explosive atmosphere
Standards for interference resistance according to EN 61326-1 This unit is an ISM device classified in Group 1 (using high frequency for internal purposes) Class A (industrial and commercial range).
11