JULABO FK31-SL User Manual

Operating Manual
Calibration Baths
FK30-SL
English
FK31-SL
JULABO GmbH 77960 Seelbach / Germany Tel. +49 (0) 7823 / 51-0 Fax +49 (0) 7823 / 24 91
1.951.2829-V2 06/13
19512829-V2.doc 22.07.13
info@julabo.de www.julabo.de
Congratulations!
You have made an excellent choice. JULABO thanks you for the trust you have placed in us. This operating manual has been designed to help you gain an understanding of the operation and possible applications of our circulators. For optimal utilization of all functions, we recommend that you thoroughly study this manual prior to beginning operation.
The JULABO Quality Management System
Temperature control devices for research and industry are developed, produced, and distributed according to the requirements of ISO 9001 and ISO 14001. Certificate Registration No. 01 100044846
Unpacking and inspecting
Unpack the circulator and accessories and inspect them for possible transport damage. Damage should be reported to the responsible carrier, railway, or postal authority, and a damage report should be requested. These instructions must be followed fully for us to guarantee our full support of your claim for protecting against loss from concealed damage. The form required for filing such a claim will be provided by the carrier.
Printed in Germany Changes without prior notification reserved
Important: keep operation manual for future use
2
TABLE OF CONTENTS
Operating manual .................................................................................................................. 5
1. Intended use ................................................................................................................... 5
1.1. Description ................................................................................................................. 5
2. Operator responsibility – Safety recommendations ........................................................ 6
2.1. Disposal ..................................................................................................................... 8
2.2. EC Conformity ............................................................................................................ 9
2.3. Warranty conditions ................................................................................................... 9
2.4. Technical specifications ........................................................................................... 10
Operating instructions ......................................................................................................... 12
3. Safety notes for the user .............................................................................................. 12
3.1. Explanation of safety notes ...................................................................................... 12
3.2. Explanation of other notes ....................................................................................... 12
3.3. Safety recommendations ......................................................................................... 13
4. Operating controls and functional elements ................................................................. 15
5. Preparations ................................................................................................................. 19
5.1. Installation ................................................................................................................ 19
5.2. Bath fluids ................................................................................................................ 20
5.3. Filling / draining ........................................................................................................ 21
6. Operating procedures ................................................................................................... 22
6.1. Power connection .................................................................................................... 22
6.2. Switching on / Selecting the language ..................................................................... 22
6.2.1. Switching on the cooling machine ....................................................................... 23
6.3. Start - Stop ............................................................................................................... 23
6.4. Adjusting the temperature chamber ......................................................................... 23
SL
7.
8.
8.0.1. Early warning system, low level protection .......................................................... 28
9.
9.1. Start of a program .................................................................................................... 31
9.1.1. Status at the end of a program ............................................................................ 33
9.1.2. Indication after successful start ........................................................................... 33
9.1.3. Editing after Start ................................................................................................. 34
9.1.4. Interrupting a program ......................................................................................... 34
9.1.5. Interruption after a power failure .......................................................................... 35
9.1.6. Termination of a program .................................................................................... 35
9.2. Program administration, creation ............................................................................. 36
9.3. Setting the pump pressure ....................................................................................... 40
9.4. Configuration ............................................................................................................ 41
9.4.1. Remote control via the serial interface ................................................................ 44
9.4.2. Keypad control or setpoint setting via the analog input ....................................... 44
9.4.3. Autostart .............................................................................................................. 45
9.4.4. Off-Mode ............................................................................................................. 45
Setting of temperatures ................................................................................... 24
Safety installations, warning functions ............................................................ 24
Menu functions ............................................................................................... 29
3
9.4.5. ActVar - actuating variable ................................................................................... 46
9.4.6. Setting of clock and date ...................................................................................... 47
9.4.7. Language ............................................................................................................. 47
9.4.8. Unit ....................................................................................................................... 47
9.4.9. Program type ........................................................................................................ 47
9.4.10. Reset ............................................................................................................... 47
9.4.11. C-Sensor - Precision reference sensor ........................................................... 48
9.5. Control ...................................................................................................................... 49
9.5.1. Control internal / external .................................................................................... 51
9.5.2. Selftuning ............................................................................................................. 52
9.5.3. Dynamik intern ..................................................................................................... 52
9.5.4. CoSpeed - extern ................................................................................................. 53
9.5.5. Control parameters – internal/external ................................................................. 53
9.6. Serial Interface .......................................................................................................... 55
9.7. ATC Absolute Temperature Calibration, 3-point calibration ...................................... 56
9.7.1. ATC Fühler - intern / extern .................................................................................. 59
9.7.2. ATC Status - yes / no ........................................................................................... 59
9.7.3. Type 1 -/ 2 -/ 3 point ............................................................................................ 59
9.7.4. Example: 3-point calibration for internal control .................................................. 60
9.8. Limits ........................................................................................................................ 62
9.8.1. Limits for internal control and external control ...................................................... 63
9.8.2. Limits for external control ..................................................................................... 64
9.9. Analog inputs/outputs ............................................................................................... 65
9.9.1. Outputs of the connector - REG+E-PROG ........................................................... 69
9.9.2. Input of the connector - REG+E-PROG ............................................................... 70
9.9.3. External Stand-by input ........................................................................................ 72
9.9.4. Alarm-output ......................................................................................................... 73
10. Troubleshooting guide / Error messages....................................................................... 74
11. Electrical connections .................................................................................................... 77
12. Remote control .............................................................................................................. 80
12.1. Setup for remote control ........................................................................................... 80
12.2. Communication with a PC or a superordinated data system .................................... 80
12.3. List of commands ...................................................................................................... 81
12.4. Status messages ...................................................................................................... 85
12.5. Error messages......................................................................................................... 85
13. Installation of electronic module with analog connectors .............................................. 86
14. JULABO Service – Online remote diagnosis ................................................................. 87
15. Cleaning / repairing the unit .......................................................................................... 88
4
Operating manual
1. Intended use
JULABO circulators have been designed to control the temperature of specific fluids in a bath tank.
JULABO circulators are not suitable for direct temperature control of foods, semi­luxury foods and tobacco, or pharmaceutical and medical products. Direct temperature control means unprotected contact of the object with the bath medium (bath fluid).
1.1. Description
The circulators are operated via the splash-proof keypad. The implemented
microprocessor technology allows to set and to store different values that can be indicated on the VFD COMFORT-DISPLAY and LCD DIALOG-DISPLAY. Three menu keys facilitate adjusting setpoints, warning and safety functions and menu functions.
The integrated programmer allows storing and running temperature and time-
dependent processes.
SL
ICC
TCF
ATC
RS232 RS485
Pt100
3
The control electronics including “ICC - Intelligent Cascade Control“ automatically
adapts the heat supplied to the thermal requirements of the bath.
The TCF - Temperature Control Features allow the user to have access to all
important temperature control parameters. This means: Full control on the control mode and the chance to manually adjust or adapt control to the specific application.
Absolute Temperature Calibration (ATC3) provides high temperature stability in the
bath. With the 3-point calibration an offset is adjusted at three temperatures to ensure an accurate temperature pattern at the selected spot in the bath over the full temperature range.
Electrical connections:
The serial interface, switchable from RS232 to RS485, allows modern process technology without additional interface. Connection for Pt100 external sensor for external temperature measurement and control. Alarm output for external alarm message or control of JULABO refrigerating baths or solenoid valve (cooling water). The electronic module (option) provides 3 further analog connections (alarm input, standby input, recorder output, programmer input).
The excess temperature protection conforming to IEC 61010-2-010 is a safety
installation independent from the control circuit. This protection can be indicated and set on the LCD DIALOG-DISPLAY.
SMART
PUMP
The early warning system for low level signals that bath fluid needs to be refilled
before the low level protection conforming to IEC 61010-2-010 causes a complete shut-down of the main functional elements.
Intelligent pump system: The pump capacity (electronically adjustable via the motor
speed) enables to adapt to varying conditions for internal and external temperature applications.
5
Operator responsibility – Safety recommendations
2. Operator responsibility – Safety recommendations
The products of JULABO ensure safe operation when installed, operated, and maintained according to common safety regulations. This section explains the potential dangers that may arise when operating the circulator and also specifies the most important safety precautions to preclude these dangers as far as possible.
The operator is responsible for the qualification of the personnel operating the units. The personnel operating the units should be regularly instructed about the dangers involved with
their job activities as well as measures to avert these dangers.
Make sure all persons tasked with operating, installing, and maintaining the unit have read and
understand the safety information and operating instructions.
When using hazardous materials or materials that could become hazardous, the circulator may be
operated only by persons who are absolutely familiar with these materials and the circulator. These persons must be fully aware of possible risks.
If you have any questions concerning the operation of your unit or the information in this manual, please contact us!
Contact
Safety instructions for the operator:
You have received a product designed for industrial use. Nevertheless, avoid strikes to the
housing, vibrations, damage to the operating-element panel (keypad, display), and contamination.
Make sure the product is checked for proper condition regularly (depending on the conditions of
use). Regularly check (at least every 2 years) the proper condition of the mandatory, warning, prohibition and safety labels.
Make sure that the mains power supply has low impedance to avoid any negative effects on
instruments being operated on the same mains.
This unit is designed for operation in a controlled electromagnetic environment. This means that
transmitting devices (e.g., cellular phones) should not be used in the immediate vicinity. Magnetic radiation may affect other devices with components sensitive to magnetic fields (e.g., monitors). We recommend maintaining a minimum distance of 1 m.
Permissible ambient temperature: max. 40 °C, min. 5 °C. Permissible relative humidity: 50% (40 °C). Do not store the unit in an aggressive atmosphere. Protect the unit from contamination. Do not expose the unit to sunlight.
JULABO GmbH Eisenbahnstraße 45 77960 Seelbach / Germany
Tel. +49 (0) 7823 / 51-0 Fax +49 (0) 7823 / 24 91 info@julabo.de www.julabo.de
Appropriate operation
Only qualified personnel is authorized to perform configuration, installation, maintenance and repairs of the circulator. Routine operation can also be carried out by untrained personnel who should however be instructed by trained personnel.
6
SL
Use:
The bath can be filled with flammable materials. Fire hazard! There might be chemical dangers depending on the bath medium used. Observe all warnings for the used materials (bath fluids) and the respective instructions (safety data sheets). Insufficient ventilation may result in the formation of explosive mixtures. Only use the unit in well ventilated areas.
Only use recommended materials (bath fluids). Only use non-acid and non corroding materials. When using hazardous materials or materials that could become hazardous, the operator must affix
the enclosed safety labels to the front of the unit so they are highly When using hazardous materials or materials that could become hazardous, the operator must affix
the enclosed safety labels (1 + 2) to the front of the unit so they are highly visible:
visible:
1
2
or 2
Particular care and attention is necessary because of the wide operating range. There are thermal dangers: Burn, scald, hot steam, hot parts and surfaces that can be touched.
Observe the instructions in the manuals for instruments of a different make that you connect to the circulator, particularly the respective safety recommendations. Also observe the pin assignment of plugs and technical specifications of the products.
Warning label W00: Colors: yellow, black Danger area. Attention! Observe instructions. (operating manual, safety data sheet)
Mandatory label M018: Colors: blue, white Carefully read the user information prior to beginning operation. Scope: EU
Semi S1-0701 Table A1-2 #9 Carefully read the user information prior to beginning operation. Scope: USA, NAFTA
Warning label W26: Colors: yellow, black Hot surface warning.
(The label is put on by JULABO)
7
Operator responsibility – Safety recommendations
2.1. Disposal
The circulator contains a back-up battery that supplies voltage to memory chips when the unit is switched off. Do not dispose of the battery with household waste! Depending on battery regulations in your country, you might be obliged to give back used or defect batteries to gathering places. The product may be used with oil as bath fluid. These oils fully or partially consist of mineral oil or synthetic oil. For disposal, observe the instructions in the safety data sheets.
This unit contains refrigerants, which at this time are not considered harmful to the ozone layer. However, over the long operating period of the unit, disposal rules may change. Therefore, only qualified personnel should handle the disposal.
Valid in EU countries See the current official journal of the European Union – WEEE directive. Directive of the European Parliament and of the Council on waste electrical and electronic equipment (WEEE). This directive requires electrical and electronic equipment marked with a crossed­out trash can to be disposed of separately in an environmentally friendly manner. Contact an authorized waste management company in your country. Disposal with household waste (unsorted waste) or similar collections of municipal waste is not permitted!
8
2.2. EC Conformity
The products described in the operating instructions conform to the requirements of the following European guidelines:
Directive of the European Parliament and of the Council on the approximation of
the laws of the Member States relating to machinery.
EMC guideline with respect to legal harmonization of the member countries
concerning electromagnetic compatibility.
SL
JULABO GmbH Eisenbahnstr. 45 77960 Seelbach / Germany
2.3. Warranty conditions
JULABO GmbH warrants its products against defects in material or in workmanship, when used under appropriate conditions and in accordance with appropriate operating instructions
for a period of ONE YEAR.
Extension of the warranty period – free of charge
With the ‘1PLUS warranty’ the user receives a free of charge extension to the warranty of up to 24 months, limited to a maximum of 10 000 working hours.
To apply for this extended warranty the user must register the unit on the JULABO web site www.julabo.de, indicating the serial no. The extended warranty will apply from the date of JULABO GmbH’s original invoice.
JULABO GmbH reserves the right to decide the validity of any warranty claim. In case of faults arising either due to faulty materials or workmanship, parts will be repaired or replaced free of charge, or a new replacement unit will be supplied.
Any other compensation claims are excluded from this guarantee.
9
Operator responsibility – Safety recommendations
2.4. Technical specifications
Calibration Baths FK30-SL FK31-SL
Working temperature range °C -30 ... 200 -30 ... 200 Temperature stability Temperature uniformity °C <±0.01 <±0.01 Cooling capacity Medium: ethanol Refrigerant R134a R134a Overall dimensions (WxDxH) cm 32x45x79 32x45x91 Bath opening (WxL) cm Bath depth cm 17 31 Filling volume Liter 14 24 Weight kg 48 51 Mains power connection 230 V/50 Hz V/ Hz 207-253 / 50 207-253 / 50 Current draw (at 230 V)
Current draw (at 230 V) <GB> Mains power connection 115 V/60 Hz V/ Hz 103-127 / 60 103-127 / 60
Current draw (at 115 V) A
(1
±0.02 °C@+200 °C
Circulator (SL-C) SL
Temperature selection digital via keypad indication on LCD DIALOG-DISPLAY (°C/°F) remote control via PC indication on monitor Temperature indication VFD COMFORT-DISPLAY (°C/°F) Resolution °C 0.01 ATC3 INT / EXT °C ±3 / ±9 Temperature control ICC - Intelligent Cascade Control Heater wattage (at 230 V) kW 2.0 Heater wattage (at 115 V) kW 1.0
Electronically adj. pump capacity stages 1 ... 4 Flow rate at 0 bar l/min 22 ... 26 Max. pressure at 0 liters bar 0.7 Max. suction at 0 liters bar 0.4 Ambient temperature °C 5 ... 40
Mains power connection 230 V/50-60 Hz V/ Hz Current draw (at 208 V / 230 V) A Mains power connection 115 V/50-60 Hz V/ Hz Current draw (at 100 V / 115 V) A
All measurements have been carried out at: rated voltage and frequency ambient temperature: 20 °C Technical changes without prior notification reserved.
(1
°C ±0.005 ±0.005
°C kW
+20 0 -20
0.46 0.34 0.15
+20 0 -20
0.46 0.34 0.15
12 12
A A A
A
14 10 (Circulator) 4 (Cooling machine)
10 (Circulator) 6 (Cooling machine)
14 10 (Circulator) 4 (Cooling machine)
10 (Circulator) 6 (Cooling machine)
190 ... 253 / 50-60 12 / 13 90 ... 127 / 50-60 9/10
10
Electrical connections:
External alarm device 24-0 V DC / max. 25 mA Computer interface RS232 or RS485 External Pt100 sensor
Optional for HL, SL (Order No. 8900100 Electronic module with analog connections) Programmer input -100 °C to 400 °C = 0 - 10 V or 0 - 20 mA or 4 - 20 mA Input for the signal of a flow meter or external manipulated variable Temperature recorder outputs 0 - 10 V (0 V = -100 °C, 10 V = 400 °C) 0 - 20 mA (0 mA = -100 °C, 20 mA = 400 °C) 4 - 20 mA (4 mA = -100 °C, 20 mA = 400 °C) Standby input for external emergency switch-off Alarm output for external alarm signal
Safety installations according to IEC 61010-2-010: Excess temperature protection adjustable from 0 °C ... 320 °C Low liquid level protection float switch Classification according to DIN 12876-1 class III
Supplementary safety installations Early warning system for low level float switch High temperature warning function optical + audible (in intervals) Low temperature warning function optical + audible (in intervals) Supervision of working sensor plausibility control Reciprocal sensor monitoring between working and safety sensors difference >35 K Alarm message optical + audible (permanent) Warning message optical + audible (in intervals)
Environmental conditions according to IEC 61 010-1: Use indoors only. Altitude up to 2000 m - normal zero. Ambient temperature: see Technical specifications Humidity: Max. relative humidity 80% for temperatures up to +31 °C, linear decrease down to 50% relative humidity at a temperature of +40 °C Max. mains voltage fluctuations of ±10% are permissible. Protection class according to IEC 60 529 IP21 The unit corresponds to Class I Overvoltage category II Pollution degree 2
Caution:
The unit is not suitable for use in explosive environment
EMC requirements according to EN 61326-1 This unit is an ISM device classified in Group 1 (using high frequency for internal purposes), Class A (industrial and commercial range).
SL
11
Operating instructions
Operating instructions
3. Safety notes for the user
3.1. Explanation of safety notes
In addition to the safety warnings listed, warnings are posted throughout the operating manual. These warnings are designated by an exclamation mark inside an equilateral triangle. “Warning of a dangerous situation (Attention! Please follow the documentation).” The danger is classified using a signal word. Read and follow these important instructions for averting dangers.
Warning:
Describes a possibly highly dangerous situation. If these instructions are not followed, serious injury and danger to life could result.
Caution:
Describes a possibly dangerous situation. If this is not avoided, slight or minor injuries could result. A warning of possible property damage may also be contained in the text.
Notice:
Describes a possibly harmful situation. If this is not avoided, the product or anything in its surroundings can be damaged.
3.2. Explanation of other notes
Note! Draws attention to something special.
Important! Indicates usage tips and other useful information.
This icon is used in the operating instructions to indicate flashing values or parameters which have to be set or confirmed.
12
3.3. Safety recommendations
SL
Follow the safety instructions to avoid personal injury and property damage. Also, the valid safety instructions for workplaces must be followed.
Only connect the unit to a power socket with an earthing contact (PE – protective
earth)!
The power supply plug serves as a safe disconnecting device from the line and
must always be easily accessible.
Place the unit on an even surface on a base made of nonflammable material. Do not stay in the area below the unit. Make sure you read and understand all instructions and safety precautions listed in
this manual before installing or operating your unit.
Set the excess temperature safety installation at least 25 °C below the fire point of
the bath fluid.
Observe the limited working temperature range when using plastic bath tanks. Never operate the unit without bath fluid in the bath. Pay attention to the thermal expansion of bath oil during heating to avoid
overflowing of the fluid.
Prevent wat e r from entering the hot bath oil. Do not drain the bath fluid while it is hot!
Check the temperature of the bath fluid prior to draining (e.g., by switching the unit on for a short moment).
Use suitable connecting tubing. Avoid sharp bends in the tubing, and maintain a sufficient distance from
surrounding walls.
Make sure that the tubing is securely attached. Regularly check the tubing for material defects (e.g., for cracks). Never opera te damaged or leaking units. Always turn off the unit and disconnect the mains cable from the power source
before performing any service or maintenance procedures, or before moving the unit.
Always turn off the unit and disconnect the mains cable from the power source
before cleaning the unit.
Always empty the bath before moving the unit. Transport the unit with care. Sudden jolts or drops may cause damage in the interior of the unit. Observe all warning labels. Never remove warning labels. Never opera te units with damaged mains power cables. Repairs are to be carried out only by qualified service personnel.
Some parts of the bath tank and the pump connections may become extremely hot
during continuous operation. Therefore, exercise particular caution when touching these parts.
13
Safety notes for the user
Caution:
The temperature controlling i.e. of fluids in a reactor constitutes normal circulator practice.
We do not know which substances are contained within these vessels. Many substances are:
inflammable, easily ignited or explosive hazardous to health environmentally unsafe
i.e.: dangerous
The user alone is responsible for the handling of these substances!
The following questions shall help to recognize possible dangers and to reduce the risks to a minimum.
Are all tubes and electrical cables connected and installed?
Note: sharp edges, hot surfaces in operation, moving machine parts, etc.
Do dangerous steams or gases arise when heating?
Is an exhaust needed when working?
What to do when a dangerous substance was spilled on or in the unit?
Before starting to work, obtain information concerning the substance and determine the method of decontamination.
Notice: Check the safety installations at least twice a year!
Excess temperature protection according to IEC 61010-2-010.
With a screwdriver turn back the adjustable excess temperature protection until the shut-down point (actual temperature).
Low level protection according to IEC 61010-2-010.
To check the function of the float, it can be manually lowered with a screwdriver for example.
14
4. Operating controls and functional elements
Front view Rear view
SL
1a 1b
2
Mains power switch, illuminated for circulator Mains power switch, illuminated for cooling machine
VFD COMFORT-DISPLAY
Header: Control indicators Line 1: Actual value internal or external
The display is depending on the selected control mode in the menu > Control < (internal or external). or Actual value of the precision reference sensor „C“ xxx.xx
Line 2: Working temp. setpoint, constantly S xxx.xx Line 3: Actual value (E = external or I = internal or C= reference sensor)
Alternating with the display in line 1
Use the keys PI Capacity in % - with manipulated variable set to >control<*
to indicate further values in line 3
15
Operating controls and functional elements
PS Capacity in % - with manipulated variable set to >serial<* or >eprog<* H Heater capacity in Watts U Mains voltage Volts F Flow rate in liters/minute (providing EPROG input set to >Flow rate<)
*refer to >MENU / CONFIGIGURATIO> page 31
2.1
Control indicators in the header: Heating / Cooling / Alarm /
2.2
2.3
3
Remote control
Control indicators in the header: Temperature indication Internal or External actual value Temperature indication in °C or °F
Display for the adjusted pump pressure stage in the -OFF- mode. Display for the effective pump pressure stage (rotation speed) after start. Four stages, adjustable via the
LCD DIALOG-DISPLAY
Line 1: Setpoint and origin of setpoint programming
(Key / RS232 or RS485 / ext. Pt100 / EProg)
Line 2: Actual value - internal or external,
identical to line 1 of the VFD-COMFORT-DISPLAY
Line 3: Heating capacity in %.
Line 4: Control type: internal / external
Navigation aid in MENU - Window
Keys Selection of menu items / parameters
- Setting in line 3:
Actual value / parameter
Orientation aid in MENU - Window
Line 1 - name with allocation to key
2 Submenu Level 2 Example: CONFIGURATION
3 Submenu Level 3
/ :
Safety values
T – Setpoint
1 Main menu Level 1
button, in the menu >PUMP<.
16
SL
4 Keypad 1
4.1
4.2
Navigation keys
1. Key: >OK< Start / Stop (pump / heater )
2. >OK< in the menu Menu item / select submenu for setting Save set value Save selected parameter A beep signals the end of setting After the actions Start, Stop and change from VFD Display to standard display the key
The above graph “front side” shows an example for standard display.
1. Key: >Return< Stop (pump / heater) Special display on VFD Display leave line 3.
2. >Return< in the menu one menu level down Correction function for parameters or values (prior to OK)
immediately back to standard display
- icon for „keep key pressed down“.
is locked for a short time.
4.3
4.4
4.5
4.6
5 Keypad 2
6
..
1. Key: >Up / Down <temperature – increase/decrease setpoint Push key quickly for single steps, Keep key pressed for fast change.
2. >Up/Down< in the menu selection of menu items / parameters
Menu keys
Key: start the menu > warning and safety values<
Key: start the menu >temperature setpoints<
Key: display of MENU structure
Numeric keypad: numerals 0 to 9
minus / decimal point
Backspace key on LCD DIALOG-DISPLAY.
Adjustable excess temperature protection according to IEC 61010-2-010
17
Operating controls and functional elements
Rear view 7
ext Pt100
8
SERIAL
9
Option: Electronic module Order No. 8 900 100
Socket for external measurement and control sensor or external setpoint programming
Interface RS232 / RS485: remote control via personal computer or Socket for a precision reference sensor Socket: control cable of JULABO refrigerated circulator
or output for alarm messages
The circulator automatically recognizes the connected electronic module.
10
ALARM
11
STAND-BY
12
REG+E-PROG
16a 16b
17a 17b
18
19
20
Alarm output (for external alarm signal)
Standby input (for external emergency switch-off)
Programmer input and temperature recorder output
Mains fuses for Circulator: T16A Mains fuses for cooling machine: T 10,0 A Mains power cable with plug for circulator Mains power cable with plug cooling machine Built-in mains outlet for connection of circulator (230 V / 16 A)
Drain screw with drain connection
Venting grid, removable
18
5. Preparations
5.1. Installation
SL
Place the unit on an even surface on a pad made of non-flammable
material.
Cooling machine, pump motor and electronics produce intrinsic heat that
is dissipated via the venting openings! Never cover these openings!
Keep at least 20 cm of open space on the front and rear venting grids. The place of installation should be large enough and provide sufficient
air ventilation to ensure the room does not warm up excessively because of the heat the instrument rejects to the environment. (Max. permissible ambient temperature: 40 °C). For a fault (leakage) in the refrigeration system, the standard EN 378 prescribes a certain room space to be available for each kg of refrigerant. The refrigerant quantity is specified on the type plate. > > For 0.25 kg of refrigerant R134a, a room space of 1 m
3
is required.
Do not set up the unit in the immediate vicinity of heat sources and do
not expose to sun light
Before operating the unit after transport, wait about one hour after setting
it up. This will allow any oil that has accumulated laterally during transport to flow back down thus ensuring maximum cooling performance of the compressor.
The bath cover (1) is mounted on the bath tank with 4 captive screws (2).
Required tool: SW3 hexagonal socket wrench with ball-shaped head
The circulating pump transports the bath fluid to the temperature chamber (4) at the bottom (3). The overflow on top of the chamber ensures a constant fluid level during operation.
The bath cover (5) provides 6 measuring places (6) that can be fitted with different Viton
®
sleeves..
®
Order-No.: Viton
sleeves 8 930 602 2 mm inner dia. 8 930 603 3 mm inner dia. 8 930 604 4 mm inner dia. 8 930 605 5 mm inner dia. 8 930 605 6 mm inner dia. 8 930 608 8 mm inner dia.
Take off the complete insert for cleaning. The gasket below the bath
cover should be replaced in this case. Part No.: 7.901.1088 Gasket for bath cover FK30, FK31
19
Preparations
5.2. Bath fluids
Caution:
Carefully read the safety data sheet of the bath fluid used, particularly with regard to the fire point! If a bath fluid with a fire point of 65 °C is used, only supervised operation is possible.
Water: The quality of water depends on local conditions. Due to the high concentration of lime, hard water is not suitable for temperature
control because it leads to calcification in the bath.
Ferrous water can cause corrosion - even on stainless steel.  Chloric water can cause pitting corrosion.  Distilled and deionized water is unsuitable. Their special properties cause
corrosion in the bath, even in stainless steel.
Recommended bath fluids:
Bath fluid Temperature range soft/decalcified water 5 °C to 80 °C
mixture water/glycol, mixture 1:1
-20°C to 50°C
JULABO bath fluids JULABO
Description
Thermal
G
Thermal HY
Thermal
H5 Order Number 10 liters 8 940 124 8 940 104 8 940 106 5 liters 8 940 125 8 940 105 8 940 107 Temperature range °C -30 ... 80 -80 ... 55 -50 ...105 Flash point °C -- 78 124 Fire point °C -- 80 142 Color
light yellow clear clear
JULABO Description
Thermal
H10
Thermal
H20S Order Number 10 liters 8 940 114 8 940 108 5 liters 8 940 115 8 940 109 Temperature range °C -20 ... 180 0 ... 220 Flash point °C 190 230 Fire point °C 216 274 Color
clear light brown
See website for list of recommended bath fluids.
2
ATTENTION: The maximum permissible viscosity is 70 mm
/s.
20
Caution:
Fire or other dangers when using bath fluids that are not recommended:
Use only non-acidic and non-corroding bath fluids.
JULABO assumes no liability for damage caused by the selection of an unsuitable bath liquid. Unsuitable bath fluids are fluids which, e.g.,
are highly viscous
(much higher than recommended at the respective working temperature)
have a low viscosity and have creep characteristics
have corrosive characteristics or tend to crack.
No liability for use of other bath fluids!
5.3. Filling / draining
Notice:
Pay attention to the thermal expansion of bath oil during heating to avoid
overflowing of the liquid.
Do not drain the bath fluid while it is hot!
Recommendation: Temperature range 5 °C to 40 °C Check the temperature of the bath fluid prior to draining (by switching the unit on for a short moment, for example).
Store and dispose the used bath fluid according to the laws for environmental
protection.
SL
Filling
Take care that no liquid enters the interior of the circulator.
The recommended filling level is marked with an
angle plate (M) inside the bath.
When the filling level does not exceed the marker
(M) and providing a suitable bath fluid is used for the specified working temperature range of the unit, neither low liquid level alarm should be triggered nor overflowing of the bath fluid should be caused.
Draining:
Turn off the unit and disconnect the mains cable
from the power source.
Hold the venting grid, pull out and remove.
19
Slide a short piece of tube onto the drain
connection and hold it into a container.
Loosen the drain screw a few turns and drain the
unit completely.
Tighten the drain tap.
21
Operating procedures
6. Operating procedures
6.1. Power connection
Caution:
Only connect the unit to a power socket with earthing contact (PE – protective
Check to make sure that the line voltage matches the supply voltage
earth)!
The power supply plug serves as safe disconnecting device from the line and must
be always easily accessible.
Never operate equipment with damaged mains power cables. Regularly check the mains power cables for material defects (e.g. for cracks). We disclaim all liability for damage caused by incorrect line voltages!
specified on the identification plate.
Connect the circulator with mains power cable (17a) to the mains
outlet (18).
Connect the control cable (9) to the sockets (9a, 9b). Connect the refrigerated circulator with mains power cable (17b) to
the mains socket.
6.2. Switching on / Selecting the language
Switching on:
Turn on the mains power switch (1a).
The unit performs a self-test.
Then the software version (example: V 1.12-S008C) appears. The display „OFF“ or „R OFF“ indicates the unit is ready to operate.
The circulator enters the operating mode activated before switching
the circulator off:
keypad control mode (manual operation) or remote control mode (operation via personal computer).
Selecting the language:
There are two options for the language of the LCD DIALOG-DISPLAY: German or English. Select the desired language in the menu >Configuration< under the submenu >Language/Sprache<.
Press the respective keys in the following order:
22
3x up to menu Configuration 6x up to submenu >Language/Sprache< deutsch / english
6.2.1. Switching on the cooling machine
Switch on the cooling machine using the switch (1b) .
SL
6.3. Start - Stop
Control of the cooling machine:
With the mains switch (1b) turned on, the circulator automatically switches the cooling machine off and on.
To save energy, turn off the cooling machine with the mains switch (1b) whenever cooling is not required.
Start:
Press
Stop:
Press
The actual bath temperature is displayed on the VFD COMFORT­DISPLAY. The circulating pump starts with a slight delay.
or Keep The VFD COMFORT-DISPLAY indicates the message "OFF".
key.
key.
key pressed.
6.4. Adjusting the temperature chamber
Adjusting the temperature chamber after set up:
If bath fluid overflow in the temperature
chamber is not equal in every direction, an adjustment can be made with the 3 screws that are accessible from above.
Required tool: SW3 hexagonal socket wrench with ball-shaped head
23
Setting of temperatures
7. Setting of temperatures
Factory settings:
Press the 3 different working temperatures can be adjusted. Their values are freely
selectable within the operating temperature range.
key to call up the menu for temperature selection.
The temperatures can be set in start or stop mode.
Setting of working temperature in the
1. Press the key
2. Select SETPOINT 1 or 2 or 3 using the key
3. Confirm by pressing the
The circulator uses the new working temperature value for
temperature control.
Example: Adjustment/modification of the pre-setting of "SETPOINT 3"
. The value flashes
key.
menu
or .
If the active Setpoint is changed, the new value is immediately used
for the control of the working temperature. The heater control indicator flashes.
Setpoint3: Change the value from 70.00°C to 80.00°C. See standard display line 1
Use the numeric keypad to enter
Then press enter
Notice: Refer to SETPOINT MAX / MIN in chapter „LIMITS“
to store the value.
and .
8. Safety installations, warning functions
Check the safety installations at least twice a year! Refer to page 14.
Settings for the excess temperature protection > SafeTemp< and for the
warning functions > OverTemp< and > SubTemp< are made in a menu which is called up by pressing the key
.
24
Menu item > Type< allows choosing between a warning and an alarm cut-off for the menu items > OverTemp< and > SubTemp<.
SL
Warning:
The excess temperature protection must be set at least 25 °C below the fire point of the bath fluid used! In case of wrong setting there is a fire hazard! We disclaim all liability for damage caused by wrong settings!
SafeTemp:
Setting range: 20 °C ... 320 °C
Rough setting can be effected by using the
temperature scale.
SafeTemp:
Type:
OverTemp :
SubTemp:
Sensor:
Settings:
Press the key
menu item
For setting proceed as follows.
Set the new cut-out value within 30 seconds with a screw driver. Exact
display and setting on LCD display Recommendation: Set over temperature-protective setting >SafeTemp< 5 °C to 10 °C above the set point of the working temperature.
Switch-over from warning to shutdown function
Change the parameter by pressing the key
pressing the key or
pressing the key
Possible parameters: warning or alarm
Set value on numeric keypad and confirm by pressing
Example: 83.00 °C Press the keys
Indicated is the temperature value of the safety sensor.
SafeTemperature:
Setting range: 20 °C ... 320 °C Indicated is the cut-out value set with a screwdriver on the excess
temperature protection device. This safety installation is independent of the control circuit. When the
temperature of the bath fluid has reached the safety temperature, a
and confirm the by pressing the key .
and by pressing the key or select the
and confirm by
.
if the parameter is to be retained.
and and confirm by pressing
25
Safety installations, warning functions
O
S
complete shutdown of the heater and pump is effected. The alarm is indicated by optical and audible signals (continuous tone).
The following error message appears on the VFD COMFORT-DISPLAY:
Press
for help on the LCD
Depending on the setting of >ControlType - internal or external< the
actual temperature values for both sensors are shown/indicated on both displays.
Examples: I 94.06
Type: >Warning< or >Alarm<
For the two menu items >OverTemp< and >SubTemp< choose between a warning message being signaled or a complete shutdown of the main functional elements such as heater and circulating pump being effected.
and Ext. : xx.xx°C
OverTemperature: SubTemperature:
If for a sensitive temperature application task adherence to a working temperature value >Setpoint< is to be supervised, then set high and low temperature warning values. In the example below, the >Setpoint< of 85 °C is surrounded by the values >OverTemp< 87 °C and >SubTemp< 83 °C. The electronics immediately registers when the actual temperature attains a temperature out of the limits and it follows a reaction according to what is set in the menu item >-Type<.
Setpoint3 85.0°C
verTemp 87.0°C
ubTemp 83.0°C
The warning functions are only triggered when the actual bath
temperature, after start from the „OFF“ or „rOFF“ mode, lies within the set limits for 3 seconds.
26
SL
Setting >WARNING<
A mere warning function with optical and audible warning signal (interval tone) A message appears on the VFD COMFORT-DISPLAY:
or
OverTemp SubTemp The LCD –DISPLAY shows the message:
Setting >ALARM<
Temperature limit with shutdown of heater and circulating pump. An audible alarm sounds (continuous tone) and a message appears on the VFD COMFORT-DISPLAY:
or
OverTemp SubTemp The LCD -DISPLAY shows the message
Press
for help on the LCD
27
Safety installations, warning functions
8.0.1. Early warning system, low level protection
This low level protection is independent of the control circuit and is divided into two sections:
DBGM: 20306059.8
1. Switch in stage 1 recognizes a critical fluid level An audible warning (interval tone) sounds and a message
appears on the VFD COMFORT-DISPLAY.
Refill bath fluid!
.
2. Switch in stage 2 recognizes a low fluid level
If stage 2 of the low level protection device (according to IEC 61010-2-010) is triggered, a complete shutdown of the heater and circulating pump is effected. A continuous alarm tone sounds and a message >ALARM< >CODE 01< appears on the VFD COMFORT-DISPLAY.
Press
Turn off the unit with the mains switch, refill bath fluid and turn
Warning:
For refill always use the same bath fluid type that is already in the bath. Bath oils must not contain any water contaminants and should be pre-heated to the actual bath temperature! Explosion hazard at higher temperatures!
the unit on again!
3. Float
4. Circulating pump
5. Heater
for help on the LCD
.
28
9. Menu functions
SL
Menu level 1
The term „menu functions“ refers to adjustments such as
Start program Page 31 This menu will start a previously set program.
Administration and creation of programs Page 36
Electronically adjustable pump capacity Page 40 The capacity of the circulating pump is set by adjusting the motor speed.
Configuration of the unit Page 41
Remote – on / off (remote control via RS232) Setpoint ext - Switch over external setpoint setting Autostart – AUTOSTART on / off Off-Mode – Pump on / off ACTVAR - Switch over input variable Time / Date – setting time and date Language – english / deutsch Unit – Temperature values can be displayed in °C or °F. Program Type – Time or Gradient (grad) Reset – factory settings C-Sensor (Precision Pt100 reference sensor)
Control characteristics and parameters Page 49 ControlType - Control internal or external Selftuning Dynamic - internal CoSpeed – external Control parameters - XP-, TN-, TV- internal Control parameters - XP-, TN-, TV- XPU-, external
29
Menu functions Menu level 1
Adjustable interface parameters Page 55 BAUDRATE, PARITY, HANDSHAKE
ATC - Absolute Temperature Calibration, Page 56 Sensor calibration internal sensor, Sensor calibration external sensor 3-point calibration
Limitations of temperature and capacity Page 63 Setpoint Max / Min - Maximum and minimum setpoint Heating Max – Set maximum heating Cooling Max – Set maximum cooling Internal Max / Min – Limitation of the temperature range Upper band / Lower band – Band limit
Analog inputs/outputs Page 65 Recorder output – CHANNEL 1, 2, 3 EPROG – External programmer input Ext-StBy - Stand-By input ALARM - output
30
9.1. Start of a program
The start menu of the integrated programmer allows one of six previously stored temperature programs to be called up and started. The profiles are started manually or via the integrated timer.
Requirements:
SL
1. Create a program. (refer to next chapter)
2. Return to the Start-MENU and confirm the desired setting of each MENU item with the key
3. Set a start time (>TIME< >DATE< >YEAR<) if the program is to be started by the internal timer.
Level 2 Parameter /Value
>Program Number< Select the program that should be started.
0 ... 5
> at Step< Select the Step at which the program should be
started. 0 … 59 > Runs < Select the number of program repetitions. 1 ... 99 > Program End< Status at program end .
StdBy, PSetp, Setp 1, Setp 2, Setp 3 > RUN < Start time (now / time)
Press the key if a parameter is to be retained. (prior to OK)
Set a parameter using the numeric
keypad and confirm by pressing
Possible parameters: 0 …5
Set a parameter using the numeric
.
keypad and confirm by pressing
Possible parameters: 0 … 59
Set a parameter using the numeric
keypad and confirm by pressing
Possible parameters: 1 … 99
.
.
Set desired parameters with
and .
Possible parameters: StdBy , PSetp, Setp 1, Setp 2 , Setp 3
31
Menu functions Level 2 Parameter /Value
Set desired parameters with
and .
Submenu TIMER
or
If the parameter time is selected, a further
Confirm >now< with the program will start immediately or start at the set time under parameter (time ). Set time in the example below:
09. August 2009, 11:15 hrs
submenu opens for setting the start time.
Set the time for the start of the program in the submenu >3 TIMER<.
key and the
Set starting time using the numeric
keypad and confirm by pressing
Example: (hh:mm)
+
Set day and month using the numeric
.
keypad and confirm by pressing
Display of waiting time until start: The unit switches to waiting mode and a flashing line “wait” appears on the LCD DIALOG-DISPLAY. The start time and current time are permanently indicated on the display.
It might be necessary to check the correct setting of the internal real time clock (see MENU configuration)
Example: (TT/MM)
+
Set year using the numeric keypad and
confirm by pressing
Example: (JJJJ)
+
Confirm >Starten/Start?< by pressing and the program starts at the set time.
.
.
32
9.1.1. Status at the end of a program
SL
Program end StdBy , PSetp, Setp 1, Setp 2 , Setp 3
This parameter is used to determine if the unit switches to the -OFF­condition at the end of a profile or if temperature control is continued, and the working temperature setpoint that is to be used if temperature control is continued.
StdBy – If the end of the profile is reached, the circulator switches to the -OFF- condition. The VFD display indicates the working temperature setpoint that was displayed in line 2 before the program was started (example: S 25.00 °C ).
PSetp –(Setpoint of the ProGrammer) At the end of the program, the setpoint of the last program step is written to the position of setpoint 3 (example: 80.00 °C) and indicated in line 1 of the LCD display. The circulator operates and controls to this temperature until the key
is pressed or a new profile is started.
Setp 1, Setp 2, Setp 3 – (“Setp” short for Setpoint) Before starting the programmer, a temperature value is set for one of
the three setpoints in the
menu. (example: setpoint 2 58.00 °C) The circulator operates and controls to this temperature at the end of the profile until the key
is pressed or a new program is started.
9.1.2. Indication after successful start
Examples:
The started programmer displays the currently calculated setpoint in line
1. The value changes until the final temperature of the section is reached (temperature ramp).
st
1
line: Setpoint of the programmer
nd
line: Actual temperature value
2 for internal control = IntAct: xxx.xx for external control = ExtAct: xxx.xx
rd
line: Selected profile and the current section (step).
3
th
4
line: With the keys, the following information can be
displayed in line 4.
Remaining time of the current section (standard display)
Total remaining time: profile x number of repetitions
(All: Days/hours:minutes:seconds) Particularities :
1. If a total time of more than 999 days is calculated, only the hours are displayed by the unit. (All: hours:minutes:seconds)
2. If the time period of a section is edited after
the start, then the
remaining total time can no longer be displayed.
33
Menu functions
AK
S
9.1.3. Editing after Start
Particularity for changes in a section that is currently in process.
The programmer allows a running program to be edited. Initiation and exit of the editing function:
After the start enter the submenu „3 EDITOR“.by pressing the key
See chapter „program administration, creation“ page 36
Exit the editing modus anytime by pressing the key
.
If the target temperature is changed, the slope of the ramp is
automatically calculated based on the remaining time of the section.
If the time period is changed, the slope of the ramp is automatically
calculated based on the new total time of the section.
The “remaining total time” (profile x number of repetitions) as
described on page 33 can no longer be displayed.
9.1.4. Interrupting a program
2
1
3
Press the
The setpoint and time period set for the corresponding section are thus stopped at the values presently achieved. The instrument is put on hold and the message “pause” flashes on the LCD DIALOG DISPLAY.
key to interrupt or continue a profile.
A program can be interrupted or restarted by an external shut-off.
Important: For this, in the >Inputs/Outputs< menu, the menu item >Ext. StBy< must be set to >active< and in the >Configuration< menu, >Autostart< needs to be set
34
Setpoint control and timer are interrupted by breaking the contact
TAND-BY
Warning:
Following a power interruption, it would be possible in this condition for the instrument to restart automatically. The safety and warning functions of the instrument should always be used to their fullest capacity. Refer to Warning page 45.
“AK”. The instrument is put on hold and the message “Stand-By” flashes on the LCD DIALOG-DISPLAY.
Notice: This is not an actual emergency switch-off.
9.1.5. Interruption after a power failure
SL
9.1.6. Termination of a program
It is possible to operate the programmer safely in case of a power failure. Important: For this, >Autostart< needs to be set in the >Configuration< menu.
If the AUTOSTART function is activated, the programmer starts
again at a point approx. 20 seconds before the interruption took place. However, an uncontrolled change of the bath temperature has occurred.
Power failure when the instrument is on hold if start is effected via
the built-in timer: If the starting time is not yet reached, the programmer is put on hold again. If the starting time is exceeded, the programmer starts immediately.
A program can be terminated by pressing . The programmer
switches back to the Start menu. Further procedures are defined in the MENU >START PROGRAM<,
menu item >Program End <. Example: >Program End - StdBy<
The VFD COMFORT-DISPLAY displays “OFF”.
Interruption after a power failure
In case of a power failure, the reaction of the circulator is the same as when switched off and on again with the mains switch.
The VFD COMFORT-DISPLAY displays “OFF”.
35
Menu functions
9.2. Program administration, creation
The integrated programmer permits fast and easy programming of setpoint temperature sequences. This temperature sequence is called program. A program is composed of individual sections (Step). The sections are defined by duration (t) or Gradient (°/t) and target temperature. The target temperature is the setpoint (SP), which is achieved at the end of a section. The programmer calculates the temperature ramp from the difference in time and temperature.
Create programs
A flashing segment indicates that a number needs to be entered. Under the submenu “Edit”, enter a program number. Six programs may be stored (numbers 0 to 5).
When the program is running, only sections with complete information for the target temperature and time period are considered. It makes sense to skip section numbers in the profile in order to use them later for corrections to the profile.
Important: If a time of 00:00 is set for a profile, the profile is continued with the next section only after the setpoint temperature (±0.2 °C) is reached.
Example: Step (Nr.) 1 3 5 6 9 11 14 Setpoint (°C) 50 50 100 100 80 80 20 Time (h:m) 00:08 00.02 00:10 01:05 00:10 00:05 00:30
36
SL
Editing
PrgNr: Set the program you wish to create. 0 ... 5
Level 2
SP: Setpoint TI : Time [hh:mm] or
GRD: Gradient [°/t] = [°/min] Change type of program in MENU >configuration<, menu item >Programmtyp/program type< .
ST: Step 0 … 59
Press the key if a parameter is to be retained. (prior to OK)
Example:
Program (PrgNr.) 2 is to have a target temperature with a set point (SW) of 100.00 °C in section t (ABS) 3. Duration (ZI) 1 hour.
Select the program number.
Set the desired >PrgNr. < by pressing
and .
Level 3
or
For the selected section >ST< a setpoint >SP< and the
duration >TI< or the gradient >GRD< are entered in the right half of the input window.
Input window for >ST<, >SP< and >TI< / >GRD<
1. Set the desired Step >ST < by pressing
and .
2. Set set point >SP< by using the numeric keypad and
confirm by pressing
.
3. Set the time >TI< or a gradient >GRD< using the numeric
keypad and confirm by pressing
.
By confirming time or GRD the values on the left side of
the input window are moved.
The values for the next section are entered in the right half
of the input window. See above 1. / 2. / 3.
37
Menu functions
Delete It is possible to delete an entire program or consecutive sections
Delete PrgNr: Set the program you wish to delete
0 ... 5
Delete from Step X to Step X (0 … 59)
Level 2
Level 3
Press the
key if a parameter is to be retained. (prior to OK)
Submenu >Delete<
Set menu item >Delete PrgNr.< by pressing
and .
Set desired PrgNr with
Example: PrgNr. 0
Input window: delete sections.
Select menu items >
after the other by pressing
Select the desired section for every menu item by using the
numeric keypad and confirm by pressing Example: >from 0< >to 59<
from Step< and > to Step< one
and .
and .
.
Select menu item >delete< by pressing Confirm >Delete?< by pressing
.
and
The delete procedure is confirm by indication of the
message >delete< on the LCD display.
38
SL
Level 2
Level 3
Print
Every program can be printed via the serial interface for control and documentation purposes.
Press the key if a parameter is to be retained. (prior to OK)
Submenu >Print<
Set menu item >Print PrgNr.< by pressing
and .
Set desired >PrgNr.< by pressing
and .
Example: 4
Confirm
> Print< by pressing .
Confirm >Print?< by pressing
.
Printing and end of printing are confirmed by successive
messages on the LCD display as shown on the left.
39
Menu functions
9.3. Setting the pump pressure
Level 2
Example: Adjusted Effective
The capacity of the circulating pump is set by adjusting the motor speed
Press the key if a parameter is to be retained. (prior to OK)
Submenu >Pump<
Adjustable pump capacity stage 1 ... 4 Illuminated display:
The set pump stage is displayed. Example: 1
Confirm by pressing
Set required pump stage by pressing
and .
Example: 3
for pump pressure
.
Indication of the adjusted pump pressure stage in the –OFF- mode. Indication of the effective pump pressure stage (rotational speed)
after start.
To protect the pump motor, the rotational speed changes with the load applied, for example, depending on the viscosity of the bath fluid at different working temperatures.
Flow rate: 22 ... 26 l/min Pump capacity stage 1 2 3 4 Pump pressure [bar] 0.4 0.5 0.6 0.7 Suction pump [bar] 0.2 0.26 0.33 0.4 Total capacity [bar] 0.6 0.76 0.93 1.1
in a loop circuit
40
9.4. Configuration
SL
Level 2 Parameter level
Press the key if a parameter is to be retained. (prior to OK)
or
Switch over setpoint
setting by pressing
and
Remote control via serial interface. Possible parameters: off / on
Switch on and off remote control by
pressing
and
First set the >type< of interface (RS232
or RS485) in the MENU >INTERFACE SERIAL< and then switch >Remote< to >on<.
Control display in the top line
for Remote
Connect SERIAL with PC.
External setpoint setting Possible parameters: off / Eprog / PT100
Switch over setpoint setting by pressing
and
off – Setpoint setting with the keys or Setpoint setting via the analog socket „ext. Pt100 or analog socket >REG+EPROG<
Autostart Possible parameters: on / off
Switch on and off Autostart by pressing
and
41
Menu functions Level 2 Parameter level
Pump-Mode Possible parameters: P. on / P. off
Switch on and off >Off-Mode by
pressing
and
P.on continuous operation of circulating
pump
P.off circulating pump is linked to
Start/Stop
ActVar - actuating variable Possible parameters: control, Eprog, serial
Set desired parameters with
and .
Programming of variables for the
parameters > serial < or > EPROG < is only accepted, if the unit is in Start mode
Time, Date, Year see next page
Language Possible parameters: deutsch / english
Set desired parameters with
and .
Unit Possible parameters: °C / °F
Set desired parameters with
and .
Program type Possible parameters: time / grad
Set desired parameters with
and .
42
Level 2 Parameter level
SL
Reset
Return to factory settings by pressing >Reset?< returns all set values to the
factory setting except for date and time.
A RESET can be effected only in the –
OFF- mode.
During the message >Running < all
parameters are reset to factory settings.
Level 2 Level 3 Parameter / Value
C-Sensor (Precision Pt100 reference sensor)
Possible parameters: yes / no
Set desired parameters with
and .
Possible parameters: Time >hour:min Date >Day:Mon Year
Succesively set time, date and year
using the numeric keypad and confirm by pressing
.
Example:
The internal real time clock allows starting a program any time. The clock is set to the local mean time (MEZ) at the factory.
If the unit is operated in a different time zone, the clock can be
adjusted in this menu.
Only the actual time is displayed.
Change summer/ winter time in this menu
43
Menu functions
9.4.1. Remote control via the serial interface
Factory setting: off
SERIAL
Setpoint is set via the serial RS232/RS485 interface through a PC or superordinated data system.
In the header of the VFD COMFORT-DISPLAY, an “ indicates that remote control mode is active.
The selected setting is shown on the LCD DIALOG-DISPLAY.
IMPORTANT: additional measures for remote control
RS232 or RS485:
First set >type< of interface in the MAIN MENU > Interface serial < the switch >Remote< to >on<.
Connect the circulator to the PC using an interface cable. Check the interface parameters of both interfaces (circulator and PC)
and make sure they match. (See 12.1. Setup for remote control page 80)
9.4.2. Keypad control or setpoint setting via the analog input
R” illuminates. It
Factory setting: off
The selected mode is indicated on standard display.
In addition to the serial interface via remote control the circulator offers the possibility to adjust the setpoint via analog interface >ext. Pt100< or >REG+E-PROG<.
Possible parameters: off - Setpoint is set via the keypad or via the integrated programmer.
(factory setting
Pt100 - Setpoint setting via the analog socket „ext. Pt100“ using an
external temperature sensor or an appropriate voltage/current source.
Eprog - Can only be adjusted when an electronic module with analog
connections is used (option). Setpoint setting via the analog interface REG+E-PROG connection with an external voltage or current source or a programmer.
Important:
Connect the external voltage or current source or a programmer to
the circulator via the socket REG+E-PROG (see page 70). In the menu >MENU Inputs/Outputs< set the parameter >Erog­INPUT< and the input variables >Eprog-signal< (see page 70).
44
The E-Prog input can only be used either under menu item
>Setpoint ext < or under menu item >ActVar<
9.4.3. Autostart
SL
Possible parameters:
on - Autostart on off - Autostart off (Factory setting)
Note:
The circulator has been configured and delivered by JULABO according to N.A.M.U.R. recommendations. This means for the start mode that the unit must enter a safe operating state after a power failure (non­automatic start mode). This safe operating state is indicated by “OFF” or “R OFF” on the VFD-COMFORT-DISPLAY. A complete shutdown of the main functional elements such as the heater and circulating pump is effected simultaneously.
The values set on the circulator remain stored, and the unit is returned to operation by pressing the start/stop key (in manual control mode). In remote control mode, the values need to be resent by the PC via the interface.
Should such a safety standard not be required, the AUTOSTART function (automatic start mode) may be activated, thus allowing the unit to be started directly by pressing the mains power switch or using a timer.
Warning:
For supervised or unsupervised operation with the “AUTOSTART“ function avoid any hazardous situation to persons or property
Take care to fully observe the safety and warning functions of the circulator.
9.4.4. Off-Mode
The AUTOSTART function can be used only if the setpoint is set via
>keyboard<, >eprog<, or >ext. Pt100<.
OFF-Mode
Normally the circulating pump is switched via the start/stop signal. However, if circulation should be maintained even in the -off- condition, the parameter >pump on< needs to be set.
Possible parameters: P.on Circulating pump in continuous operation P.off Circulating pump linked to Start/Stop (factory setting)
45
Menu functions
9.4.5. ActVar - actuating variable
The variable (ACTuating VARiable) corresponds to the extent to which the heater or cooling unit of the circulator is controlled. Heat or cold is applied to the bath according to this variable. If this happens with the control electronics of the circulator, called > control < in this particular case, the bath temperature is exactly heated and maintained constant at the adjusted setpoint.
Programming of variables for the parameters > SERIAL < or >
EPROG < is only accepted, if the unit is in Start mode.
Possible parameters: control – The internal control electronics of the circulator controls the
heater and the connected cooling unit. Self-tuning is possible.. (factory setting)
serial – – The heater or the connected cooling unit receives the control
signal via the serial interface. Self-tuning is not possible.
Eprog - The heater or the connected cooling unit receives the control
signal via the E-Prog input. Self-tuning is not possible.
- Setting requires electronic module.
Important: Set the parameter >Eprog< and the input variables also in the menu >Inputs/Outputs< (see page 70).
Note: The E-Prog input can only be used either under menu item >Setpoint ext< or under menu item > ActVar <.
Warning:
The working temperature range of the circulator is determined during configuration. If set to >Control<, this range cannot be exceeded. If set to > Serial < and > Eprog <, heat or cold is applied to the bath without control. The permissible maximum temperature can be exceeded. The user has to take adequate precautions for temperature control. Materials, such as gaskets or insulations for example, may be damaged or destroyed, if the permissible maximum temperature is exceeded.
The safety and warning functions > < of the instrument must always be used to their fullest capacity.
46
9.4.6. Setting of clock and date
The internal real time clock allows starting a program any time. The clock is set to the local mean time (MEZ) at the factory.
If the unit is operated in a different time zone, the clock can be
adjusted in this menu.
Change summer/ winter time in this menu
9.4.7. Language
SL
9.4.8. Unit
9.4.9. Program type
There are two options for the language of the LCD DIALOG-DISPLAY: German and English.
Possible parameters:
deutsch / english
Temperature values can be displayed in °C or °F. With this change, all adjusted temperature values are converted and
displayed in the new unit. Possible parameters:
°C (factory setting) °F
A program can be created with the parameters time or gradient.
9.4.10. Reset
Possible parameters:
time (factory setting) Grd (Gradient) [°/t] = [°/min]
A Reset will return all values to factory setting except for date and time.
A Reset can be effected in the >OFF< mode only.
47
Menu functions
9
5
9.4.11. C-Sensor - Precision reference sensor
Factory setting: no
1
SERIAL
6
°C INT
Start
Temperature measurement with a precision reference sensor Precision reference sensor for applications in quality assurance,
calibration services and laboratories. Possible parameters:
yes – Temperature measurement with a precision reference sensor no
Connect the higly accurate temperature sensor to the RS232 socket
SERIAL) on the rear of the circulator.
(
Set menu item > C-Sensor < to > yes <
In the start status the value „C“ of the precision reference sensor is
permanently shown in line 1. The control indicator „°C“ in the header blinks.
In the stop status (OFF) it is in line 3.
Stop
CONTROL –INTERNAL / EXTERNAL Control
The precision reference sensor cannot be used for external control. Accesories: Highly accurate temperature probe Order-No. 8 981 002
Technical specifications Immersion depth minimum 90 mm maximum 140 mm Measurement range -40,00 °C ... +300,00 °C Measurement accuracy Range –40,00 °C .. . –0,001 °C ±(0,05 °C +0,05 % v. Mw.) Range 0,00 °C .. +100,00 °C ±0,05 °C Range +100,001 °C .. +300,00 °C ±(0,05 °C +0,05 % v. Mw.)
48
9.5. Control
SL
Level 2 Parameter-Level
The circulator is designed for internal and external temperature control. This can be switched in this submenu. Only the set of parameters corresponding to the particular setting is displayed.
Control internal Control external
For external temperature control and measurement connect a Pt100
external sensor to the socket at the rear of the circulator.
Press the key if a parameter is to be retained. (prior to OK)
Control Type internal Parameter-Level
or
Switch over control type by pressing
and
Possible parameters: intern / extern
The control type can be adjusted in the
-OFF- mode only.
The parameter flashes, switch by pressing
and
Possible parameters: off / once / always.
The parameter flashes, switch by pressing
and
Possible parameters: aperio / normal
This parameter affects the temperature
sequence in case of internal control.
49
Menu functions Control Type internal Parameter-Level
Control Type external Parameter-Level
The value flashes. Set by using the numeric
keypad and confirm by pressing
Example: Setting range: 0.1 … 99.9 K
.
The value flashes. Set by using the numeric
keypad and confirm by pressing
Setting range: 3 … 9999 s
.
The value flashes. Set by using the numeric
keypad and confirm by pressing
Setting range: 0 … 999 s
.
The value flashes. Set by using the numeric
keypad and confirm by pressing
Example:
Setting range: 0.0 … 5.0
.
The value flashes. Set by using the numeric
keypad and confirm by pressing
Setting range: 0.1 … 99.9 K
.
The value flashes. Set by using the numeric
keypad and confirm by pressing
Setting range: 3 … 9999 s
.
The value flashes. Set by using the numeric
keypad and confirm by pressing
Setting range: 0 … 999 s
.
The value flashes. Set by using the numeric
keypad and confirm by pressing
Setting range: 0.1 … 99.9 K
.
50
9.5.1. Control internal / external
The control type can be adjusted in the -OFF- mode only.
ext.
Pt100
Possible parameters:
intern internal temperature control extern external temperature control with external Pt100 sensor
IMPORTANT:
Additional measures for external temperature control.
Connect a Pt100 sensor to the socket on the rear of the circulator. Sensor calibration of the external Pt100 sensor is performed in the
>ATCalibration< menu.
Suggested adjustments for external temperature control:
Upper band / Lower band and IntMax / IntMin see Section >Limits<.
SL
Notice:
Place the external sensor into the temperature-controlled medium and securely fix the sensor.
Pt100 M+R
External temperature control does not allow the setpoint to be set via
the “ext. Pt100” socket.
Accessory: Pt100 external sensor
Order No. Description Material Cable 8981003 200x6 mm Ø, stainless steel 1.5 m
8981005 200x6 mm Ø, glass 1.5 m 8981006 20x2 mm Ø, stainless steel 1.5 m 8981010 300x6 mm Ø, stainless steel 1.5 m 8981015 300x6 mm Ø, stainless steel / PTFE coated 3 m 8981013 600x6 mm Ø, stainless steel / PTFE coated 3 m 8981016 900x6 mm Ø, stainless steel / PTFE coated 3 m 8981014 1200x6 mm Ø, stainless steel / PTFE coated 3 m 8981103 Extension cable for Pt100 sensor 3.5 m 8981020 M+R in-line Pt100 sensor
The M+R in-line Pt100 sensor is a flow sensor and can be installed loop circuit
51
Menu functions
9.5.2. Selftuning
When performing a selftuning for the controlled system (temperature application system), the control parameters Xp, Tn and Tv are automatically determined and stored.
Possible parameters:
off - no
once - single
always - continual
selftuning The control parameters ascertained during the last identification are used for control purposes.
selftuning (factory setting) The instrument performs a single selftuning of the controlled system after each start with the command via the interface.
The instrument performs a selftuning of the controlled system whenever a new setpoint is to be reached. Use this setting only when the temperature application system changes permanently.
key or after receiving a start
selftuning
9.5.3. Dynamik intern
°C
normal temp. stability
aperiodic
setpoint
°C
normal temp. stability
aperiodic temperature ramp
This parameter affects the temperature sequence only in case of internal control.
Factory setting: APER (aperiodic)
Possible parameters:
normal Allows for reaching the setpoint faster – with
t
aperio Ramp function: the increase of temperature
setpoint change or ramp function – but overshooting of up to 5 % is possible.
occurs temporally offset and achieves the target temperature without overshooting. Setpoint change: The temperature increases at the same rate, the target temperature is achieved without overshooting.
With both settings constant temperature is achieved
after approximately the same time.
t
52
9.5.4. CoSpeed - extern
SL
°C
CoSpeed:
This parameter affects the temperature pattern only in case of external control.
Possible parameters:
Int. temperature
Ext. temperature
5 3 0
Setpoint
t
9.5.5. Control parameters – internal/external
0 ... 5
During selftuning, the control parameters Xp, Tn, and Tv of a controlled system are automatically determined and stored. The time required for tuning may vary depending on the controlled system. This controller design allows protection of sensitive objects requiring temperature control. As soon as a co-speed factor is set, it is considered in calculating the control parameters. As shown in the diagram, tuning times are shorter the higher the co-speed factor is, but overshoot can occur in the internal system.
Setting range: internal / external
0.1 ... 99.9 K
Setting range: internal / external 3 ...9999 s
Setting range: internal / external 0 ... 999 s
Control parameters – internal/external
The control parameters preset at the factory are in most cases adequate to achieve an optimal temperature pattern in the samples requiring temperature control. Each parameter may be manually set via the keypad if necessary to allow optimal control performance.
Proportional range >Xp<
The proportional range is the temperature range below the setpoint temperature value in which the control circuit reduces the heating power from 100% to 0%.
Resetting time >Tn<
Compensation of the remaining control deviation due to proportional control. An insufficient resetting time may cause instabilities to occur. An excessive resetting time will unnecessarily prolong compensation of the control difference.
Lead time >Tv< (Differential component)
The differential component reduces the control settling time. An insufficient lead time will prolong the time required to compensate for disturbance effects and cause high overshoot during run-up. An excessive lead time could cause instabilities (oscillations) to occur.
(Integral component)
Setting range:
0.1 ... 99.9 K
Proportional range >Xpu<
The proportional range Xpu of the cascaded controller is only needed for external control.
53
Menu functions
°C
°C
Optimization instructions for the PID control parameters:
optimum setting
The heat-up curve reveals inappropriate control settings.
t
Inappropriate settings may produce the following heat-up curves:
Xp too low
Tv/Tn too low
°C
°C
Xp too high or Tv too high
t
Tv/Tn too high or Xp too high
t
°C
t
t
54
9.6. Serial Interface
SL
Factory settings: 4800 Baud even hardware handshake
Level 2 Parameter-level
For communication between the circulator and a PC or a superordinated process system, the interface parameters of both units must be identical. The adjustment is usually carried out one time only in the >Interface serial< menu.
RS232 RS485
Press the key if a parameter is to be
retained. (prior to OK)
The parameter flashes, switch by pressing
and
Possible parameters: RS232 / RS485 Address RS485: 0 to 127
The parameter flashes, switch by pressing
and
Possible parameters:
4800 / 9600 / 19200 / 38400
The parameter flashes, switch by pressing
and
Possible parameters: no/even/odd even: Data bits = 7; Stop bits = 1 odd: Data bits = 7; Stop bits = 1 no: Data bits = 8; Stop bits = 1
The parameter flashes, switch by pressing
and
Possible parameters: hardw / softw Xon/Xoff-protocol (Software handshake)
Protocol RTS/CTS (Hardware handshake)
55
Menu functions
°
9.7. ATC Absolute Temperature Calibration, 3-point calibration
ATC is used to compensate for a temperature difference between the
circulator and a defined measuring point in the bath tank that may develop for physical reasons. ATC internal external
Example: 1-point calibration
°C
Principle: For ATC calibration, in steady state the bath temperature at the location of the temperature sensor (CT) is determined at the respective adjusted working temperature. This value is then set on the circulator in the menu >ATCalibration< under menu item > CTEMP X <. This can be a 1-point, 2-point or 3-point calibration.
3-point calibration
°C
T
1 = Original curve
T
T 1
T
T 1
T
M B (INT) T
C
°C
TT
ATC (internal)
M = Temperature measuring instrument with temperature sensor B = Bath tank (INTernal or EXTernal) T = circulator
CT = Temperature on measuring point TT = Temperature on circulator
°C
56
SL
Press the key if a parameter is to be retained. (prior to OK)
Level 2 Parameter-Level
or
or
The parameter flashes, switch by pressing
and
The parameter flashes, switch by
pressing
and
>no< Carry out an ATC calibration >yes< return to standard operation after
calibration.
The parameter flashes, switch by pressing
A >1-point<, >2-point< or >3-point<
calibration can be carried out.
The value >TmpValue< is only indicated
and
In addition the measured temperature
value >CalValue X< is saved during the next step.
The value flashes. Set by using the
numeric keypad and confirm by pressing
.
If only a 1-point calibration is carried out, the following menu items are
not indicated anymore
57
Menu functions
The value >TmpValue< is only indicated
In addition the measured temperature
If only a 2-point calibration is carried out, the following menu items are
not indicated anymore
value >CalValue X< is saved during the next step.
The value flashes. Set by using the
numeric keypad and confirm by pressing
.
The value >TmpValue< is only indicated.
In addition the measured temperature
value >CalValue X< is saved during the next step.
The value flashes. Set by using the
numeric keypad and confirm by pressing
.
58
9.7.1. ATC Fühler - intern / extern
In the first submenu the ATC function is set for the >intern< internal or the >extern< external temperature sensor.
SL
9.7.2. ATC Status - yes / no
Calibration can be carried out for the internal temperature sensor and for the external temperature sensor connected to the socket „ext. Pt100“. The circulator is able to save both parameter sets. However only the one which has been set under menu item > sensor < is displayed.
In the second submenu the ATC function for the temperature sensor selected above is activated >yes< or deactivated >no<.
>yes< (factory setting) The controller of the circulator uses the original
>no< An ATC calibration is to be carried out.
In the > Status < >yes< the ATC calibration always affects the current
curve of the temperature sensor or the new curve measured during the ATC calibration.
Important: Set to >NO< during the calibration process
Important: Set to >yes< after calibration.
working temperature; also the one set via interface.
9.7.3. Type 1 -/ 2 -/ 3 point
A >1-point<, >2-point< or >3-point< calibration can be carried out. First geometrically define the location for calibration (measuring point
CT), then determine the temperature values of the calibration points. The type of calibrations also determines the number of the following pairs of values indicated on the LCD DIALOG-DISPLAY.
Pairs of values: TmpValue X: Circulator temperature 1 or 2 or 3 (actual value TT )
CalValue X: Calibration temperature 1 or 2 or 3 (actual value CT )
The „calibration value“ is determined with a temperature measuring device and saved under menu item >CalValue<.
(value flashes
The actual temperature of the bath is simultaneously saved with the “calibration value“ >CalValue< and can be indicated for control purposes (value does not flash).
)
59
Menu functions
9.7.4. Example: 3-point calibration for internal control
In the temperature range from 80 °C to 160 °C the calibration curve of the
temperature sensor (TT) is to be adjusted to the actual temperatures at measuring point (CT).
80.00 °C
120.00 °C
160.00 °C
1. Set circulator to internal control:
The type of control can be set only in the –OFF- mode.
2. Set working temperature setpoint See standard display line 1
Example 80.00 °C
Press the keys
The bath is heated up.
Wait for approx. 5 minutes until the temperature is constant.
3. Reading of temperature measuring device
Read the value of measuring point CT on the device and enter under menu item > CalValue X< by using the keypad.
CalValue 1< (79.70 °C) CalValue 2< (119.50 °C) CalValue 3< (159.30 °C)
4. Calibration
Press the key if a parameter is to be retained. (prior to OK) Setting is required only for the first calibration point.
Level 2 Parameter-Level
and and confirm by pressing .
Set sensor: >intern<.
The parameter flashes, switch by
pressing and
Set to >no<
The parameter flashes, switch by
pressing and
A >3-point< calibration is carried out.
60
The parameter flashes, switch by
pressing
and
SL
The value >TmpValue< is only indicated.
In addition the measured temperature
Return to 2. Set working temperature value: 120.00 °C
value >CalValue X< is saved during the next step.
The value flashes. Set by using the
numeric keypad and confirm by pressing
.
The value >TmpValue< is only indicated.
In addition the measured temperature
Return to 2. Set working temperature value: 160.00 °C
value >CalValue X< is saved during the next step.
The value flashes. Set by using the
numeric keypad and confirm by pressing
.
5. Return to standard operation
The value >TmpValue< is only indicated.
In addition the measured temperature
value >CalValue X< is saved during the next step.
The value flashes. Set by using the
numeric keypad and confirm by pressing
.
Set >YES< after calibration. (Standard operation)
The parameter flashes, switch by
pressing
and
61
Menu functions
9.8. Limits
The >Limits< menu allows the minimum and maximum values to be set for all important setting ranges and capacity variables.
Control internal
Press the key if a parameter is to be retained. (prior to OK).
Level 2 Parameter-level
Control external
The parameter flashes. Set by using the
numeric keypad and confirm by pressing
.
The parameter flashes. Set by using the
numeric keypad and confirm by pressing
.
The parameter flashes. Set by using the
numeric keypad and confirm by pressing
.
The parameter flashes. Set by using the
numeric keypad and confirm by pressing
.
62
SL
The parameter flashes. Set by using the
numeric keypad and confirm by pressing
.
The parameter flashes. Set by using the
numeric keypad and confirm by pressing
.
The parameter flashes. Set by using the
numeric keypad and confirm by pressing
.
The parameter flashes. Set by using the
9.8.1. Limits for internal control and external control
Set Max / Min– Maximum and minimum setpoint Restriction of the adjustable temperature range.
The limitation of the operating temperature range effects the temperature setting in the menu with the key Only setting of working temperatures which lie within the determined limits
is possible Existing settings for Setpoint 1, -2, -3, as well as those for >OverTemp< and > SubTemp <, are automatically deferred into the limit range.
Setting range:
-94,90 °C ... +300,0 °C
numeric keypad and confirm by pressing
.
.
Set maximum heating / cooling
The heating and cooling capacity of the unit are adjustable. 100 % corresponds to the technical specification of the equipment.
Setting range:
Heating Max – 0 to 100 % in 1 % steps Cooling Max – 0 to 100 % in 1 % steps
63
Menu functions
2
9.8.2. Limits for external control
> SAFETMP <
INTERN MAX / MIN Restriction for the temperature range of the internal bath.
Setting range: The limits IntMax and IntMin are only active in external control. IntMax
and IntMin determine fixed limits for the temperature within the internal bath. The temperature controller cannot exceed these limits even if it would be necessary for achieving the temperature in an external system. Therefore it is possible that the external setpoint cannot be achieved.
Sense of limit setting:
-94,9 °C ... +300,0 °C
Protects the bath fluid from overheating. Prevents an undesired alarm shutdown by the excess temperature
protection - >ALARM CODE 14<. Set the value of > Int Max at least 5 °C below the value of >SafeTemp<.
Protects the pump motor from high viscosity of the bath fluid at low
temperatures.
For refrigerated circulators. Freezing protection when using water as
bath fluid.
00
150
100
75 50
°C
BAND HIGH
Upper band: and Lower band: The band limitation is active during external control. Varied, practice-
oriented setting are feasible for heat-up and cool-down phases.
Internal bath external system
Setting range:
Upper band: and Lower band allow for the limitation of the difference between the temperatures in the internal bath and the external system to any maximum value for the heat-up and cool-down phase. During the heat-up phase this difference value is always added to the actual external temperature. During the cool-down phase the difference value is subtracted.
0 °C ... 200 °C
Sense of a band limitation:
Protection of objects and samples
by gentle temperature control
Protection of e.g. glass reactors
from thermal shock.
BAND LOW
25
t
10 20 30 40 50 60 10 20 30 40 50 60 10
64
[min]
9.9. Analog inputs/outputs
SL
ALARM
STAND-BY
REG+ E-PROG
If the electronic module is not fitted, the item >inputs/outputs< will be missing in the menu.
In order to use the analog inputs and outputs, the circulator must be
equipped with the electronic module, which is available as option. Order No. 8 900 100 Electronic module
This submenu enables the input and output values to be set for the programmer input and the temperature recorder outputs of the REG+E­PROG socket.
The >STAND-BY< input and the >ALARM< output are configurable.
Menu item >SV-OUT< is not used.
Channel 1 voltage output for recorder (V) Channel 2 voltage output for recorder (V)) Channel 3 current output for recorder (mA) EPROG external programmer input External standby Standby input
(for external switch-off)
Alarm output Output for external alarm signal
NOTE:
Channel 2 is pre-configured for the connection of an HST Booster Heater.
Press the key if a parameter is to be retained. (prior to OK)
Level 2 Parameter-Level
Level 2 Level 3 Parameter-level
Menu item >SV-OUT< is not used.
Define the output value for channel 1 Possible parameters:
ActInt, ActExt, Power, Setpoint
The parameter flashes, switch by pressing
and
65
Menu functions Level 2 Level 3 Parameter-level
Select the scale for channel 1.
The value flashes. Set by using the
numeric keypad and confirm by pressing
.
The value flashes. Set by using the
numeric keypad and confirm by pressing
.
Define the output value for channel 2. Possible parameters:
ActInt, ActExt, Power, Setpoint
The parameter flashes, switch by pressing
and
Select the scale for channel 2.
The value flashes. Set by using the
numeric keypad and confirm by pressing
.
The value flashes. Set by using the
numeric keypad and confirm by pressing
.
Define the output value for channel 3 Possible parameters:
ActInt, ActExt, Power, Setpoint
The parameter flashes, switch by pressing
and
66
Level 2 Level 3 Parameter-level
Select the scale for channel 3.
The value flashes. Set by using the
numeric keypad and confirm by pressing
The value flashes. Set by using the
Possible parameters: 0-20 mA, 4-20 mA
.
numeric keypad and confirm by pressing
.
The parameter flashes, switch by pressing
SL
and
Possible parameters: Setpoint, variable, flow rate, pressure
The parameter flashes, switch by pressing
and
Possible parameters: voltage, current
The parameter flashes, switch by pressing
and
67
Menu functions Level 2 Level 3 Parameter-level
The value flashes. Set by using the
numeric keypad and confirm by pressing
.
The value flashes Set by using the numeric
keypad and confirm by pressing
.
(Parameter Level)
The value >setpoint< is only indicated.
Possible parameters: inactive, active
The parameter flashes, switch by pressing
Possible parameters: Alarm, Stand By, Alarm+ Stand By
The parameter flashes, switch by pressing
and
and
68
Possible parameters: normal, inverse
The parameter flashes, switch by pressing
and
9.9.1. Outputs of the connector - REG+E-PROG
SL
Outputs of the connector
1. First define the desired output value for channels 1 to 3
:
ActInt internal actual temperature value (bath temperature) ActExt external actual temperature value (external sensor) Power periodic or intermittent heating or cooling Setpoint active setpoint temperature
(setpoint 1, 2, 3/integr. programmer/external programmer)
2. Then select the scale for channels 1 to 3: Voltage outputs channels 1 and 2
Assign the voltage values of 0 V to the lowest and 10 V to the highest necessary temperature or power rating required as an output value (°C/%). Current output channel 3 Assign the current values 0 mA or 4 mA to the lowest and 20 mA to the highest temperature or power rating required as an output value (°C/%).
3. The current output (channel 3) offers 2 ranges for selection: 0 to 20 mA or 4 to 20 mA
0 V
10 °C 210 °C
  200 K
197 °C 202 °C
 5 K
0 mA 4 mA
20 mA
Examples: lowest temperature value: 10 °C highest temperature value 210 °C Fig. shows 200 C scaled to paper width slope: 50 mV/°C
lowest temperature value: 197 °C highest temperature value: 202 °C Fig. shows 5 C scaled to paper width slope: 2000 mV/°C
69
Menu functions
9.9.2. Input of the connector - REG+E-PROG
Menu >Configuration<
1.
2. Menu >Inputs/Outputs<
3.
Example:
E-PROG - Input
Setting is necessary if
1. the Setpoint is to be set via an external voltage or current source or programmer For this, in the menu >Configuration<, first set the menu item >Setpoint< to >Eprog<.
2. the heater variable should be controlled via an external control pulse. For this, in the menu >Configuration<, set the menu item > ActVar.< to >Eprog<.
3. the signal of an external flow meter should be registered.
The E-Prog input can only be used either under menu item
>Setpoint< or under menu item > Actvar <. If the input is occupied neither by >Setpoint< or > ActVar <, the signal of a flow meter can be connected.
First set menu >Configuration<, then menu >Inputs/Outputs<.
Connect the external voltage or current source or programmer to
the REG+E-PROG socket (12) of the circulator.
Selecting the signal:
The programmer (E-PROG) input of the circulator can be matched to the output signal of the external voltage or current source.
Voltage voltage input Current current input
“L Value” - Setting the LOW value: (See below
First adjust and set the lowest voltage or current on the external voltage or current source (e.g., 0 V or 0 mA). Then after approx. 30 seconds, enter the corresponding temperature value (e.g., 20.00°C).on the circulator by pressing the
appropriate keys on the keypad and press
“H Value” - Setting the HIGH value: (See below
First adjust and set the highest voltage or current on the external voltage or current source (e.g., 10 V or 20 mA). Then after approx. 30 seconds, enter the corresponding temperature value (e.g., 300 °C).on the circulator by pressing the
appropriate keys on the keypad and press
Example:
)
to set.
)
to set.
Set the external voltage or current source output for the equivalent
of a 50 °C temperature setpoint. The value adjusted and set on the external programmer is displayed in line 4 of the LCD DIALOG-DISPLAY for inspection.
70
SL
C
A
After returning the LCD display to standard display by pressing the
key voltage or current source is displayed in line 1 (Example: Eprog 50.00 C).
This EPROG input enables the use of different voltage and current values as program parameters.
, the temperature value adjusted and set on the external
300
250
200
150
100
80 60 40 20
°
B
12345678910V
246810 14 18 20 m
“L Value” - Setting the Low value: (See below
)
1) Adjust and set the lowest desired value on the voltage or current source (Example A: 1 V). Wait approximately 30 seconds.
2) Assign a lower temperature threshold value to this adjusted voltage/current value by pressing the appropriate keys on the
A
keypad of the instrument (Example A: 20 C) and set by pressing
.
“H Value” - Setting the High value: (See below
)
1) Adjust and set the highest desired value on the voltage or current source (Example A: 10 V). Wait approximately 30 seconds.
2) Assign an upper temperature threshold value to this adjusted voltage/current value by pressing the appropriate keys on the keypad of the instrument (Example A: 200 C) and set by pressing .
Example B in the diagram illustrates that the end point values
are freely selectable (e.g., 8 mA and 16 mA).
Example out of diagram A:
Adjusting the voltage source for an output of 7.6 V!
Line 5 of the LCD DIALOG-DISPLAY shows the externally set setpoint value. The instrument calculates this value from the slope of the two specified end points (in example A, 7.6 V corresponds to an external
setpoint temperature of 152.0 C ). After returning the LCD display to standard display by pressing
, this value is displayed in line 1 (Example: EPROG 152.00 C).
Notice:
If this adjustment is not correctly performed at two different points, the setpoint setting will be incorrect.
Important:
The usable temperature range between > L Value < and > H Value < is limited to the configured working temperature range of the circulator or unit combination. For the working temperature range, see technical specifications.
71
Menu functions
AK
S
9.9.3. External Stand-by input
AK
2
1
STAND-BY
3
Ext. StBy (External STAND-BY input)
Stand-By input for external switch-off.
Possible parameters:
inactive - Stand-By input is ignored active - Stand-By input is active
Activate the standby input:
1. Under menu item >Ext. StBy<, set the parameter to >active<.
2. Connect an external contact “AK” (e.g., for external switch-off) or an alarm contact of the superordinated system.
If the connection between pin 2 and pin 3 is interrupted by opening the contact “AK”, a complete shutdown of the circulating pump and heater is effected, and the unit enters the condition “E OFF”. As long as the contact remains open, line 4 of the LCD DIALOG­DISPLAY flashes and displays the message “Stand-By”.
If the contact is reclosed, the instrument returns to the standby state and “ E OFF “ is displayed. Press
to start.
Additional tips for using the STAND-BY input:
The standby function can be used in conjunction with the AUTOSTART feature.
2
1
TAND-BY
3
1. If the Autostart function is NOT turned ON, the standby input is used as described above.
2. If the Autostart function is enabled, the instrument will revert back to the original method of entering the setpoint (i.e., keypad, RS232, analog input, etc.).
Entering the setpoint with the keypad. As described above, a complete shutdown is accompanied by the “Stand-By” display and the OFF state. The programmable controller starts again when the contact is reclosed. The temperature of the bath fluid changed during the Stand-By state.
Entering the setpoint with the programmer The display “StandBy” appears. The setpoint value and the time are both held at the current value. The temperature of the bath fluid will be held constant at this temperature. The programmer continues once the contact is reclosed.
Notice: this is not an actual emergency switch-off.
e.g.,
.
72
9.9.4. Alarm-output
ALARM
SL
Alarm output (10)
(for external alarm signal) This socket is a potential-free change-over contact. With the
adjustments in the menu >Inputs/Outputs<, all operating conditions can be signaled without having to change the pin assignments.
Meaning of the terms under menu item >Function< : The circulator is in condition
>StandBy< or >Alarm<
For >Type: normal<, pins 2 and 3 are connected in any case according to the selected >Function<.
For >Type: inverse<, pins 2 and 1 are connected in any case according to the selected >Function<.
Switching capacity max. 30 W/40 VA Switching voltage max. 125 V/ Switching current max. 1 A
73
Troubleshooting guide / Error messages
10. Troubleshooting guide / Error messages
Alarm with complete shutdown:
If one of the following failures occur a complete, all-pole shutdown of the heater and circulating pump is effected.
+
“ lights up and a continuous signal sounds.
„ The code for the cause of alarm is indicated on the VFD COMFORT­DISPLAY.
Alarm without shutdown:
The code for the cause of alarm is indicated on the VFD COMFORT­DISPLAY. The warning signal sounds in regular intervals. The messages appear every 10 seconds.
Press the key Press to indicate the help text in the LCD DIALOG-DISPLAY when
The circulator is operated without bath fluid, or the liquid level is insufficient. Replenish the bath tank with the bath fluid.
Tube breakage has occurred (insufficient filling level due to excessive bath fluid pumped out). Replace the tubing and replenish the bath tank with the bath fluid.
The float is defect (e. g., because damaged in transit). Repair by authorized JULABO service personnel.
During the self-test after switch-on a short-circuit is registered between pin 2 and pin 4 of the control cable or the control cable is interrupted during operation. Reconnect the cable or eliminate the short-circuit.
The cable of the working temperature sensor is disconnected or short­circuited.
to stop the signal
alarm messages appear.
Excess temperature warning
or Excess temperature alarm
Warn-Type: Set to >Warning< or >Alarm<
Low temperature warning
or Low temperature alarm.
Warn-Type: Set to >Warning< or >Alarm<
74
SL
Defect of the working or excess temperature sensor. The working temperature and excess temperature sensors report a temperature difference of more than 35 K.
Other errors (I
2
C-BUS errors)
Error in A/D converter
Excess temperature sensor is defective. The safety temperature is below the working temperature setpoint. Set the
safety temperature to a higher value.
External control selected, but external Pt100 sensor is not connected or is defective.
Cooling of the condenser is impaired. Clean the air-cooled condenser. For a water-cooled condenser, check the flow rate and temperature of the cooling water.
The first-stage compressor does not work. After a short cooling interval, the compressor motor will be automatically reconnected and the message “CODE 21” will no longer appear.
The second-stage compressor does not work. Cooling compressor overload protection
The motor of the cooling compressor is equipped with an overload protector, which will be activated in case of excessive temperature in the capsule or excessive current draw.
Causes of motor shut-off:
- poor air circulation
- short distance to walls
- dirt accumulated on condenser
- high ambient temperature
- compressor switched off and on at short intervals Excess temperature in the first-stage compressor.
Excess temperature in the second-stage compressor.
75
Troubleshooting guide / Error messages
Short-circuit in the control cable for the cooling machine during self-test.
Cable of the excess temperature sensor disconnected or short-circuited.
Ext. Pt100 sensor input without signal, but setpoint programming set to external Pt100.
The early warning system for low level signals a critical fluid level. Replenish the bath tank with bath fluid.
60
1545
4 s
30
After eliminating the malfunction, turn the mains power switch off and on again to cancel the alarm state. If the unit cannot be returned to operation, contact an authorized JULABO service center.
Special message “CONFIGURATION ERROR” The configuration of the circulator does not correspond to its present use.
Press configuration.
to automatically perform a one-time modification of the
Then contact an authorized JULABO service center.
Disturbances that are not indicated.
The electronic pump motor is protected against overload by an electronic current limiter. If the viscosity of the bath fluid is or becomes too high, the motor stops running.
Fuses: The mains fuses on the rear of the unit may easily be replaced as shown at
left. Circulator: Fine fuses (16a)– T 16 A, 250 V~ , D5 x 20 mm Cooling machine: Fine fuses (16b)– T 10 A, 250 V~ , D5 x 20 mm
Warning:
Before exchanging the fuses, turn off the mains power switch and disconnect the power plug from the mains socket! Only use fine fuses with a nominal value as specified.
Example:
76
Manufacturer Supplier Type Order-No. Schurter Schurter G-fuse link SPT T16A
No. 0001.2516
5x20mm
11. Electrical connections
S
g
9
5
Notice:
Use shielded cables only. The shield of the connecting cable is electrically connected to the plug housing.
The unit ensures safe operation if connecting cables with a maximum length of 3 m are used. The use of longer cables does not affect proper performance of the unit, however external interferences may have a negative impact on safe operation (e.g. cellular phones).
2
3
ext. Pt100
1
4
Socket for external Pt100 sensor
Pt100
hield Plug
SL
Pin assignment: Pin Signal 1 I+ 2 U+
2
1
3 U­ 4 I-
3
Look on
solderin
The shield of the connecting cable is electrically connected to the plug housing
and the sensor tube.
1 6
SERIAL
RS232/RS485 serial interface
This port can be used to connect a computer with an RS232 or RS485 cable for remote control of the circulator.
Pin assignments RS232:
Pin 2 RxD Receive Data Pin 3 TxD Transmit Data Pin 5 0 V Signal GND Pin 7 RTS Request to send Pin 8 CTS Clear to send
Pin 1; 4; 6, 9 Reserved - do not use!
Pin assignments RS485:
Pin 3 B Pin 5 0 V Signal GND Pin 6 +5 V (max. 50 mA) Pin 8 A
Pin 1; 2; 4; 7; 9 Reserved - do not use!
4
side.
Notice: RS485 serial interface
The max. load current between pin 5 and pin 6 is 50 mA.
77
Electrical connections RS232 interface cable
Circulator (9-pol) PC (9-pol) Pin 2 RxD Pin 3 TxD Pin 3 TxD Pin 2 RxD Pin 5 GND Pin 5 GND Pin 7 RTS Pin 8 CTS Pin 8 CTS Pin 7 RTS
Accessories: Order No. Description
8 980 073 RS232 interface cable 9-pol./9-pol. , 2,5 m 8 900 110 USB interface adapter cable
/ Control output
The connector may only be used for control of a JULABO refrigerated circulator or JULABO MVS Solenoid valve controller for cooling water.
Pin assignment: Pin Signal (only with attached JULABO equipment)
1 +24 V (I max. current 25 mA) 2 0 V 3 Alarm relay 4 Reserved - do not use! 5 Cooling pulse
Alarm output
(for external alarm signal)
ALARM
AK
2
1
STAND-BY
This potential-free change-over contact is activated in case of an alarm when pins 2 and 3 are connected.
Switching capacity max. 30 W / 40 VA Switching voltage max. 125 V/ Switching current max. 1 A
Stand-by input
(for external switch-off) Pin assignment: Pin Signal
1 not connected 2 5 V / DC 3 0 V
3
78
SL
3
6
2
1
REG+E-PROG
4
5
Programmer input / temperature recorder output
Pin Signal 1 Voltage output Channel 1 0 ... 10 V 2 Voltage output Channel 2 0 ... 10 V 3 GND for outputs 0 V 4 Programmer input EPROG 0 to 10 V / 0 to 20 mA 5 Current output Channel 3 0 to 20 mA / 4 to 20 mA 6 GND for Progammer 0 V
79
Remote control
9
6
5
12. Remote control
12.1. Setup for remote control
1
SERIAL
1. Check the interface parameters for both interfaces (on the circulator and PC) and make sure they match. In the >Interface< menu, set the menu item >Type< to >RS232< or >RS485<.
2. In the >Configuration< menu, set the menu item >Setpoint< to >RS232< or >RS485<.
3. Connect both units with an interface cable.
Like all parameters that can be entered through the keypad, interface parameters are stored in the memory even after the circulator is turned off.
12.2. Communication with a PC or a superordinated data system
If the circulator is put into remote control mode via the configuration level, the VFD COMFORT-DISPLAY will read “R -OFF-” = REMOTE STOP. The circulator is now operated via the computer. In general, the computer (master) sends commands to the circulator (slave). The circulator sends data (including error messages) only when the computer sends a query.
In remote control mode, the start command and all values to be set must be resent by the PC via the interface in case of a power interruption. AUTOSTART is not possible.
A transfer sequence consists of:
address (RS485 interface only) command space (; Hex: 20)
parameter (decimal separation with a period) end of file (; Hex: 0D)
The response (data string) after an in command is always followed by a
line feed (LF, Hex: 0A).
The commands are divided into in or out commands.
in commands: retrieve parameters out commands: set parameters
80
SL
Important times for a command transmission:
To ensure secure data transfer, the time gap between two commands should be at least 250 ms. The circulator automatically responds to an in command with a data string followed by an LF (Line Feed). The next command should be sent only after 10 ms.
The out commands are valid only in remote control mode. Examples:
When the RS485 interface is used, the three-digit instrument address precedes each command. (example: address Ad32 = A032)
Command to set the working temperature >Setpoint1< to 55.5 °C
out_sp_00  55.5 A032_out_sp_00  55.5
Command to retrieve the working temperature >Setpoint1<
in_sp_00 A032_in_sp_00
Response from the circulator:
55.5 LF A032_55.5 LF
Notice: out commands
Temperature values can be displayed on the circulator in °C or °F. The settings sent via the interface must also be made in the selected unit according to this setting.
12.3. List of commands
out commands: Setting parameters or temperature values. Command Parameter Response of circulator
out_mode_01 0 Use working temperature >Setpoint1< out_mode_01 1 Use working temperature >Setpoint2< out_mode_01 2 Use working temperature >Setpoint3< out_mode_02 0 Selftuning “off”.
Temperature control using the stored parameters.
out_mode_02 1 Selftuning “once”
Single selftuning of the controlled system after the next start.
out_mode_02 2 Selftuning “always”
Continual selftuning of the controlled system whenever a new setpoint is to be reached.
81
Remote control
Command Parameter Response of circulator
out_mode_03 0 Set external programmer input to voltage.
Voltage 0 V ... 10 V
out_mode_03 1 Set external programmer input to current.
Current 0 mA ... 20 mA
out_mode_04 0 Temperature control of internal bath. out_mode_04 1 External control with Pt100 sensor. out_mode_05 0 Stop the unit = R –OFF-. out_mode_05 1 Start the unit. out_mode_08 0 Set the control dynamics - aperiodic out_mode_08 1 Set the control dynamics - standard
out_sp_00 xxx.xx Set working temperature. “Setpoint 1” out_sp_01 xxx.xx Set working temperature. “Setpoint 2” out_sp_02 xxx.xx Set working temperature. “Setpoint 3”
out_sp_03 xxx.xx Set high temperature warning limit “OverTemp” out_sp_04 xxx.xx Set low temperature warning limit “SubTemp” out_sp_06 xxx.xx Set manipulated variable for heater via serial interface
-100 ... +100 [%] out_sp_07 x Set the pump pressure stage. (1 ... 4) out_par_04 x.x CoSpeed for external control 0 ... 5.0.
out_par_06 xxx Xp control parameter of the internal controller. 0.1 ... 99.9 out_par_07 xxx Tn control parameter of the internal controller. 3 ... 9999 out_par_08 xxx Tv control parameter of the internal controller. 0 ... 999 out_par_09 xxx Xp control parameter of the cascade controller. 0.1 ... 99.9 out_par_10 xxx Proportional share of the cascade controller. 1 ... 99.9 out_par_11 xxx Tn control parameter of the cascade controller. 3 ... 9999 out_par_12 xxx Tv control parameter of the cascade controller. 0 ... 999 out_par_13 xxx Maximum in ternal temperature of the cascade controller. out_par_14 xxx Minimum internal temperature of the cascade controller. out_par_15 xxx Band limit (upper) 0 ... 200 °C out_par_16 xxx Band limit (lower) 0 ... 200 °C out_hil_00 -xxx Set the desired maximum cooling power (0% to 100%).
Note: Enter the value with a preceding negative sign! This setting is meaningful only for FP cooling machines.
out_hil_01 xxx Set the desired maximum heating power (10% to 100%).
82
in commands: Asking for parameters or temperature values to be displayed.
Command Parameter Response of circulator
version None Version number of the software (V X.xx) status none Status message, error message (see page 85)
in_pv_00 none Current bath temperature. in_pv_01 none Heating power being used (%). in_pv_02 none Temperature value registered by the external Pt100 sensor. in_pv_03 none Temperature value registered by the safety sensor. in_pv_04 none Setpoint temperature (“SafeTemp”) of the excess
temperature protection
SL
in_sp_00 none Working temperature “Setpoint 1” in_sp_01 none Working temperature “Setpoint 2” in_sp_02 none Working temperature “Setpoint 3” in_sp_03 none High temperature warning limit “OverTemp” in_sp_04 none Low temperature warning limit “SubTemp” in_sp_05 none Setpoint temperature of the external programmer
(REG+E-PROG).
in_sp_06 none Temperature indication in °C or °F in_sp_07 none 1. Adjusted pump stage in the -OFF- condition.
2. Pump stage corresponding to the effective rotational speed after start.
in_sp_08 none Value of a flow rate sensor connected to the E-Prog input
in_sp_09 none Value of a pressure sensor connected to the E-Prog input
in_par_00 none Temperature difference between working sensor and safety
sensor
in_par_01 none Te - Time constant of the external bath. in_par_02 none Si - Internal slope in_par_03 none Ti - Time constant of the internal bath. in_par_04 none CoSpeed - Band limit (max. difference between the
temperatures in the internal bath and external system).
in_par_05 none Factor pk/ph0:
Ratio of max. cooling capacity to max. heating capacity
in_par_06 none Xp control parameter of the internal controller. in_par_07 none Tn control parameter of the internal controller.
83
Remote control
Command Parameter Response of circulator
in_par_08 none Tv control parameter of the internal controller. in_par_09 none Xp control parameter of the cascade controller. in_par_10 none Proportional share of the cascade controller. in_par_11 none Tn control parameter of the cascade controller. in_par_12 none Tv control parameter of the cascade controller. in_par_13 none Adjusted maximum internal temperature of the cascade
controller.
in_par_14 none Adjusted minimum internal temperature of the cascade
controller.
in_par_15 none Band limit (upper) in_par_16 none Band limit (lower) in_mode_01 none Selected setpoint:
0 = Setpoint 1 1 = Setpoint 2 2 = Setpoint 3
in_mode_02 none Selftuning type:
0 = Selftuning “off” 1 = Selftuning “once” 2 = Selftuning “always”
in_mode_03 none Type of external programmer input:
0 = Voltage 0 V to 10 V
1 = Current 0 mA to 20 mA
in_mode_04 none
Internal/external temperature control: 0 = Temperature control with internal sensor.
1 = Temperature control with external Pt100 sensor.
in_mode_05 none
Circulator in stop/start condition: 0 = stop
1 = start
in_mode_08 none
Adjusted control dynamics 0 = aperiodic
1 = standard
in_hil_00 in_hil_01
84
none Max. cooling power (%). none Max. heating power (%).
12.4. Status messages
Status messages Description
SL
00 MANUAL STOP 01 MANUAL START 02 REMOTE STOP 03 REMOTE START
12.5. Error messages
Error messages
-01 LOW LEVEL ALARM
-02 REFRIGERATOR ALARM
-03 EXCESS TEMPERATURE WARNING
Circulator in „OFF“ state. Circulator in keypad control mode. Circulator in „r OFF“ state. Circulator in remote control mode.
Description Low liquid level alarm Control cable of the refrigerated circulator or MVS
solenoid valve controller short-circuited or disconnected. High temperature warning
-04 LOW TEMPERATURE WARNING
-05 WORKING SENSOR ALARM
-06 SENSOR DIFFERENCE ALARM
-07 I2C-BUS ERROR
-08 INVALID COMMAND
-09 COMMAND NOT ALLOWED IN CURRENT OPERATING MODE
-10 VALUE TOO SMALL
-11 VALUE TOO LARGE
-12 TEMPERATURE MEASUREMENT ALARM
-13 WARNING : VALUE EXCEEDS TEMPERATURE LIMITS
-14 EXCESS TEMPERATURE PROTECTOR ALARM
Low temperature warning. Working temperature sensor short-circuited or
disconnected. Sensor difference alarm.
Working temperature and safety sensors report a temperature difference of more than 35 K.
Internal error when reading or writing the I Invalid command. Invalid command in current operating mode.
Entered value too small. Entered value too large. Error in A/D converter.
Value lies outside the adjusted range for the high and low temperature warning limits. But the value is stored.
Excess temperature protector alarm
2
C bus.
-15 EXTERNAL SENSOR ALARM
External control selected, but external Pt100 sensor not connected.
85
Installation of electronic module with analog connectors
A
A
Error messages
-20 WARNING: CLEAN CONDENSOR OR CHECK COOLING WATER CIRCUIT OF REFRIGERATOR
-21 WARNING: COMPRESSOR STAGE 1 DOES NOT WORK
-22 WARNING: COMPRESSOR STAGE 2 DOES NOT WORK
-23 WARNING: HIGH TEMPERATURE ON COMPRESSOR STAGE 1
-24 WARNING: HIGH TEMPERATURE ON COMPRESSOR STAGE 2
-25 REFRIGERATOR WARNING
-26 WARNING: STAND-BY PLUG IS MISSING
-30 CONFIGURATION ERROR: CONFIRM BY PRESSING <OK> ON CIRCULATOR
Description Cooling of the condenser is impaired.
Clean air-cooled condenser. Check the flow rate and cooling water temperature for a water-cooled condenser.
First-stage compressor does not work.
Second-stage compressor does not work.
Excess temperature at first-stage compressor.
Excess temperature at second-stage compressor.
Error in the cooling machine. External standby contact is open.
The configuration of the circulator does not correspond to its present use. Press OK to automatically perform a one-time modification of the configuration.
-33 SAFETY SENSOR ALARM
Excess temperature sensor short-circuited or disconnected.
-38 EXTERNAL SENSOR SETPOINT PROGRAMMING ALARM
-40 NIVEAU LEVEL WARNUNG
Ext. Pt100 sensor input without signal and setpoint programming set to external Pt100.
Low liquid level warning in the internal reservoir.
13. Installation of electronic module with analog connectors
Caution:
Always turn off the unit and disconnect the mains cable from the power source before performing any service or maintenance procedures, or before moving the unit. Configuration, installation, maintenance, and repairs on the circulator may be performed only by qualified personnel.
LARM
B
STAND-BY
REG+ E-PROG
Turn off the unit with the mains switch and disconnect the
power plug.
Remove the screws (A) and remove the dummy plate. Carefully insert the upper edge (B) into the rear panel from
below, align the electronic module at the top and bottom, and connect the 15-pin connector by applying light pressure.
Attach the electronic module using the screws (A). The circulator is ready for use. It automatically recognizes the
installed electronic module.
86
14. JULABO Service – Online remote diagnosis
SL
JULABO circulators of the HighTech series are equipped with a black box. This box is implemented in the controller and records all significant data for the last 30 minutes. In case of a failure, this data can be read out from the unit by using special software. This software is available as a free download from www.julabo.de EasyBlackBox.
Installation is easy and is performed step by
step. Please observe the instructions.
Data read-out is possible in the conditions
“OFF”, “R OFF” or “ALARM”.
Connect the circulator to the computer using an
interface cable.
Start the EasyBlackBox program.
The program asks for the port used (COM1, ......)
and the baud rate of the unit. You do not have this information on hand? Simply try it out! The program continues to send the request until the correct settings are made.
Data is read out and shown on the monitor
divided into the sections >Einstellungen/Settings<, >Alarmspeicher/Alarms stored<, >Blackbox<
see example
After pressing >Speichern/Save<, a text file is
created. The program suggests a filename ­>C:\model description and barcode no.<. Modifications are possible.
E-mail this file to service@julabo.de
service department. JULABO is thus able to provide rapid support.
, JULABO's
\
87
Cleaning / repairing the unit
15. Cleaning / repairing the unit
Caution:
Always turn off the unit and disconnect the mains cable from the power source
before cleaning the unit.
Prevent humidity from entering into the circulator. Electrical connections and any other work must be performed by qualified
personnel only.
In dusty environments, the refrigerating machine’s condenser should be cleaned by vacuuming every 1 to 2 months. The condenser is located behind the venting grid on the front of the unit.
Switch off the unit, disconnect mains power cable. Remove the venting grid. Clean the ribbed condenser with a vacuum cleaner. Replace the venting grid. Switch on the unit.
Cleaning:
For cleaning the bath tank and the immersed parts of the circulator, use low surface tension water (e.g., soap suds). Clean the outside of the unit using a wet cloth and low surface tension water.
The circulator is designed for continuous operation under normal conditions. Periodic maintenance is not required.
The tank should be filled only with a bath fluid recommended by JULABO. To avoid contamination, it is essential to change the bath fluid from time to time.
Repairs Before asking for a service technician or returning a JULABO
instrument for repair, please contact an authorized JULABO service station.
When returning the unit:
Clean the unit in order to avoid any harm to the service personnel. Attach a short fault description. During transport the unit has to stand upright. Mark the packing
correspondingly.
When returning a unit, take care of careful and adequate packing. JULABO is not responsible for damages that might occur from
insufficient packing.
JULABO reserves the right to carry out technical modifications with repairs for providing improved performance of a unit.