www.joy-it.net
Pascalstr. 8 47506 Neukirchen-Vluyn
LABORATORY POWER SUPPLY
JT-RD6006
1. GENERAL INFORMATION
Dear costumer,
thank you very much for choosing our product.
In following, we will introduce you to what to observe while starting up
and using this product.
Should you encounter any unexpected problems during use, please do
not hesitate to contact us.
The RD6006 is a constant voltage and constant current DC power supply
which can convert voltages down.
It is very easy to handle and the adjustment can either be made by keypad or by the encoder of the or even both.
It is easy and quick to store and recall data and the RD6006 is very con-
nectable with a PC or via WIFI with a handheld device in addition that you
can use an App for control. On the 2.4 " colour screen you can read out
many values and additional information.
www.joy-it.net
Pascalstr. 8 47506 Neukirchen-Vluyn
SAFETY INSTRUCTIONS
Our product is suitable for up to 70 Volt DC. Working with voltage over 60 V DC is reserved for trained electricians, because of the hazard of an electric shock which may
cause serious injuries or death. Non-electricians have to use a voltage source below
60 Volt DC.
Power source input interface must be connected to
6-70V constant DC power source. The external sensor cable (as shown on right) must be connected
to the external temperature sensor interface.
The fan interface cannot be connected to other fans. When the output current is
higher than 4A or the system temperature higher than 45 °C , the fan start
to work, when the temperature is less than 40°C and output current lower than 3.9A,
the fan will stop working.
When the system temperature is higher than 80°C , the output will be shut down
because of OTP. CR1220 is the clock battery. communication interface is a special
interface, please don’t connect to other modules or cables.
Note: This instruction is corresponding to firmware V1.25.
www.joy-it.net
Pascalstr. 8 47506 Neukirchen-Vluyn
2. TECHNICAL SPECIFICATIONS
3. SCREEN OVERVIEW
Display 2.4 " colour LCD
Input voltage range 6 - 70 V
Output voltage range 0 - 60 V
Output current range 0 - 6 A
Output power range 0 - 360 W
Input voltage accuracy ± 1 % + 5 digits
Output voltage accuracy ± 0.3 % + 3 digits
Output current accuracy ± 0.5% + 5 digits
Battery voltage measurement accuracy ± 0.5% + 3 digits
Input voltage measurement resolution 0.01 V
Output voltage measurement resolution 0.01 V
Current setting measurement resolution 0.001 A
Battery voltage measurement resolution 0.01 V
Constant voltage mode response time 2 ms (at 0.1 - 4 A load)
Constant voltage mode load regulation ± 0.1 % + 2 digits
Constant current mode load regulation ± 0.1 % + 3 digits
Capacity measurement range 0 - 9999.99 Ah
Energy measurement range 0 - 9999.99 Wh
Capacity and Energy statistical error ± 2 %
Output ripple typical 100mV VPP
Sensor temperature detection range -10 - 100°C / 0- 200 ° F
Sensor temperature detection accuracy ± 3° C / ± 6 °F
Working mode Buck mode
Voltage drop > 1V and > 10 %
Screen brightness setting 0 - 5 Level , total 6 level
Working temperature range - 10° - 40 ° C
Current data group
Constant voltage/Constant current status
Protection status indication
Button lock status
Communication Interface
Battery charging indication
Battery-related information
Overcurrent protection value
Overvoltage protection value
Output voltage preset value
Output current preset value
Button tune
Time
Input voltage
Current output voltage
Current output power
Current output current