Joy-it RD6006 User guide [ml]

www.joy-it.net
Pascalstr. 8 47506 Neukirchen-Vluyn
LABORATORY POWER SUPPLY
JT-RD6006
1. GENERAL INFORMATION
Dear costumer, thank you very much for choosing our product. In following, we will introduce you to what to observe while starting up and using this product. Should you encounter any unexpected problems during use, please do not hesitate to contact us.
The RD6006 is a constant voltage and constant current DC power supply which can convert voltages down. It is very easy to handle and the adjustment can either be made by key­pad or by the encoder of the or even both.
It is easy and quick to store and recall data and the RD6006 is very con-
nectable with a PC or via WIFI with a handheld device in addition that you can use an App for control. On the 2.4 " colour screen you can read out many values and additional information.
www.joy-it.net
Pascalstr. 8 47506 Neukirchen-Vluyn
SAFETY INSTRUCTIONS
Our product is suitable for up to 70 Volt DC. Working with voltage over 60 V DC is re­served for trained electricians, because of the hazard of an electric shock which may cause serious injuries or death. Non-electricians have to use a voltage source below 60 Volt DC.
Power source input interface must be connected to 6-70V constant DC power source. The external sen­sor cable (as shown on right) must be connected to the external temperature sensor interface.
The fan interface cannot be connected to other fans. When the output current is
higher than 4A or the system temperature higher than 45 °C , the fan start
to work, when the temperature is less than 40°C and output current lower than 3.9A, the fan will stop working.
When the system temperature is higher than 80°C , the output will be shut down because of OTP. CR1220 is the clock battery. communication interface is a special interface, please dont connect to other modules or cables.
Note: This instruction is corresponding to firmware V1.25.
www.joy-it.net
Pascalstr. 8 47506 Neukirchen-Vluyn
2. TECHNICAL SPECIFICATIONS
3. SCREEN OVERVIEW
Display 2.4 " colour LCD
Input voltage range 6 - 70 V Output voltage range 0 - 60 V Output current range 0 - 6 A Output power range 0 - 360 W Input voltage accuracy ± 1 % + 5 digits Output voltage accuracy ± 0.3 % + 3 digits Output current accuracy ± 0.5% + 5 digits Battery voltage measurement accuracy ± 0.5% + 3 digits Input voltage measurement resolution 0.01 V Output voltage measurement resolution 0.01 V Current setting measurement resolution 0.001 A Battery voltage measurement resolution 0.01 V Constant voltage mode response time 2 ms (at 0.1 - 4 A load) Constant voltage mode load regulation ± 0.1 % + 2 digits Constant current mode load regulation ± 0.1 % + 3 digits Capacity measurement range 0 - 9999.99 Ah Energy measurement range 0 - 9999.99 Wh Capacity and Energy statistical error ± 2 % Output ripple typical 100mV VPP Sensor temperature detection range -10 - 100°C / 0- 200 ° F
Sensor temperature detection accuracy ± 3° C / ± 6 °F
Working mode Buck mode Voltage drop > 1V and > 10 % Screen brightness setting 0 - 5 Level , total 6 level Working temperature range - 10° - 40 ° C
Current data group
Constant voltage/Constant current status
Protection status indication
Button lock status
Communication Interface
Battery charging indication
Battery-related information
Overcurrent protection value
Overvoltage protection value
Output voltage preset value Output current preset value
Button tune
Time
Input voltage
Current output voltage
Current output power
Current output current