JBL DC280, DSH115, DS280, DSX280, DSX280HD Installation Manual

...
Page 1
Installation Manual
JBL Professional Intellivox - DC/DS/DSH/DSX 115, 180, 280, 380, 430 & 500 models
(Part Nos. 585000, 585001, 585020, 585021, 585060, 585061, 585120, 585121, 585160, 585161, 585300, 585301, 587000, 587001, 587020, 587021, 587040, 587041, 587060, 587061, 587080, 587081, 587120, 587121, 587140, 587141, 587160, 587161, 587180, 587181, 587300, 587301, 587350, 587351, 587400, 587480, 587481)
Page 2
2
JBL Intellivox Installation Manual
This document confirms that products manufactured by
D
uran Audio bearing the CE label meet all the requirements
in the EMC directive 2004/108/EC and LV directive
2006/95/EC laid down by the Member States Council for
adjustment of legal requirements. Furthermore the products
comply with the rules and regulations from 30 August 1995
referring to the electromagnetic compatibility of devices.
Duran Audio products bearing the CE label comply with the
following harmonised or national standards:
EMC:
EN 55103-1 :1996 EN 55103-2 :1996
Safety:
IEC60065:2001 (ed7), A1:2005, A2:2010
CAN/CSA 60065-03 + Amd 1 and
UL 60065-07 (pollution degree 2)
Mains Harmonics:
EN 61000-3-2 :2001
Insulation:
Class1
Duran Audio BV
Koxkampseweg 10
5301 KK Zaltbommel
The Netherlands
Tel: +31 418 515583
Fax: +31 418 518077
Zaltbommel, January 2012.
USER’S NOTICE AND DISCLAIMER:
No part of this manual including the software described in
it may be reproduced, transmitted, transcribed, stored in a
database system or translated without the express written
permission of JBL Professional.
Documentation kept by the end user for back-up purposes
is excluded from the above.
All products and corporate names mentioned in this manual
may be registered trademarks or copyrights of their
respective companies. They are used here for indicative
purposes only.
The information contained in this manual has been carefully
checked for accuracy; however no guarantee is given with
respect to its correctness. JBL Professional accepts no
responsibility or liability for any errors or inaccuracies that
may appear in this manual or the products and software
described in it.
Specifications and information contained in this manual are
subject to change at any time without notice.
© 1998-2014 JBL Professional. All rights reserved.
REFERENCE TO EC STATEMENT OF CONFORMITY
Page 3
3
JBL Intellivox Installation Manual
Reference to Statements of Conformity . . . . . . . . . . . . . . . . . . . 2
User’s Notice and disclaimer . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
1. Important Safety Instructions. . . . . . . . . . . . . . . . . . . . . . . . . 4
2. Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Applicable models and variants . . . . . . . . . . . . . . . . . . . . 5
What’s in the packaging. . . . . . . . . . . . . . . . . . . . . . . . . . 6
3. Installation Guide . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Preparing for Installation . . . . . . . . . . . . . . . . . . . . . . . . . 7
Signal and Control cables . . . . . . . . . . . . . . . . . . . . . . . 7-8
Connector and wiring details . . . . . . . . . . . . . . . . . . . . 8-9
AC Mains . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
WAGO connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Audio inputs. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Network connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Fault monitoring/fan control connector. . . . . . . . . . . 13-14
External Ambient SPL/Temperature Sensor. . . . . . . . . . . 15
Mechanical Installation . . . . . . . . . . . . . . . . . . . . . . . . . 16
- Acoustic centre . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
- Mounting Options. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
- Mounting Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
4. Unit checks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Connecting the PC to the Intellivox . . . . . . . . . . . . . . . . 19
Settings file upload . . . . . . . . . . . . . . . . . . . . . . . . . . 20-21
5. Appendix. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Optional accessories . . . . . . . . . . . . . . . . . . . . . . . . . 22-23
Common analogue grounding issues . . . . . . . . . . . . . . . 24
Software and firmware updates . . . . . . . . . . . . . . . . . . . 24
Network cables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Failure monitoring on 70 V/100 V connection. . . . . . . . . 25
-
Configuring PCB960 . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
-
Configuring PCB124 . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Failure Relay Configuration. . . . . . . . . . . . . . . . . . . . . . . 27
TABLE OF CONTENTS
Page 4
4
JBL Intellivox Installation Manual
This symbol is intended to alert you to the
presence of uninsulated dangerous voltages
within the product's enclosure that may be of
sufficient magnitude to constitute a risk of
electric shock.
This symbol is used throughout this manual and
is intended to alert you to the presence of
important instructions.
1) Read these instructions.
2) Keep these instructions.
3) Heed all warnings.
4) Follow all instructions.
5) Do not use this apparatus near water.
6) Clean only with dry cloth.
7) Do not block any ventilation openings. Install in
accordance with the manufacturer's instructions.
8) Do not install near any heat sources such as radiators,
heat registers, stoves, or other apparatus (including
amplifiers) that produce heat.
9) Do not defeat the safety purpose of the polarised or
grounding-type plug. A grounding-type plug has two blades
and a third grounding prong. The wide blade or the third
prong are provided for your safety. If the provided plug does
not fit into your outlet, consult an electrician for
replacement of the obsolete outlet.
10) Protect the power cord from being walked on or
pinched particularly at plugs, convenience receptacles, and
the point where they exit from the apparatus.
11) Only use attachments/accessories specified by the
manufacturer.
12) Use only with the cart, stand, tripod, bracket or
table specified by the manufacturer, or sold with the
apparatus. When a cart is used, use caution when moving
the cart/apparatus combination to avoid injury from tip-over.
13) Unplug this apparatus during lightning storms or when
unused for long periods of time.
14) Refer servicing to qualified service personnel. Servicing
is required when the apparatus has been damaged in any
way, such as power-supply cord or plug is damaged, liquid
has been spilled or objects have fallen into the apparatus,
the apparatus has been exposed to rain or moisture, does
not operate normally, or has been dropped.
Warning -To reduce the risk of fire or electric
shock, do not expose this apparatus to rain or
moisture and objects filled with liquids, such as vases,
should not be placed on this apparatus.
Warning - In order to disconnect this apparatus
from the mains power supply, the connector
should be removed from the mains socket located on the
amplifier module.
Warning - Do not expose the internal backup battery to
excessive heat such as sunshine, fire or the like.
Caution - Danger of explosion if the internal backup
battery is incorrectly replaced. Replace only with the same
or equivalent type.
Warning - The installer should ensure that the
mechanical mounting method employed should
be capable of supporting four times the weight of the unit
(i.e., with a safety factor of 4). Always use all the brackets
provided with the unit. Only skilled persons may install
the units.
Warning - This apparatus is a Class I device and
must be connected to a mains socket outlet that
provides a safety ground connection.
1. IMPORTANT SAFETY INSTRUCTIONS
Page 5
5
JBL Intellivox Installation Manual
This manual describes the recommended installation
p
rocedure for the JBL Intellivox DC (DDC2.0) and DS (DDS)
range of loudspeakers.
The JBL Intellivox is a type of loudspeaker known as an
active DSP-controlled loudspeaker array, which utilises the
principle of multiple, individually-driven loudspeaker drive
units in the vertical plane at specific spacings. In addition to
the drive units, the Intellivox housing contains a mains-
powered electronics module containing multiple power
amplifier channels and a Digital Signal Processing (DSP)
section.
This manual describes the following aspects of an
installation:
• Necessary cabling
• Connector wiring
• Mechanical installation
After installation, the loudspeaker should be tested for basic
functionality using JBL WinControl software; the manual
also describes this procedure. (Note that full configuration
and control of the loudspeaker is beyond the scope of this
manual; this topic is fully documented in the WinControl
Help files.)
Please note that a full range of optional
mounting brackets and connection coverboxes
are available. Please refer to our website for
further details.
APPLICABLE MODELS, AND VARIANTS
The various models in the Intellivox range are fitted with
different numbers of drive units.
This manual applies to the following JBL
Intellivox models:
Intellivox-Intellivox-DC115, DS115 and DSH115* (fig.1a)
Intellivox-DC180, DS180 and DSX180 (fig.1b)
Intellivox-DC280, DS280, DSX280 and DSX280HD** (fig.1c)
Intellivox-DC380, DS380 and DSX380 (fig.1d)
Intellivox-DC430, DS430 and DSX430 (fig.1e)
Intellivox-DC500, DS500 and DSX500 (fig.1f)
The DS and DSX models use a different DSP algorithm from
the DC models, and the DSX models use two different types
of drive units in each loudspeaker instead of the DS’s single
type†. These model differences may be ignored for the
purposes of installation and commissioning.
2. INTRODUCTION
fig.1
The Intellivox models covered by this manual
abcdef
* The DSH models are 'high-frequency-only' and can be used to augment an existing DS model.
** Models with the HD suffix offer increased directivity control for high frequencies.
†The DC115 & DS115 models are an exception, also employing two different driver types.
Page 6
6
JBL Intellivox Installation Manual
A standard option applicable to all models is the position
within the housing of the electronics module, which may
e
ither be at the top or bottom of the housing. The position
is relevant to the installation procedure as the “acoustic
centre” (see fig 2) of the loudspeaker array will be different
in the two cases. It is not possible for the electronics
module’s position to be changed after production.
A further option available to special order is for the
electronics module to be mounted at the rear of the unit.
This results in a loudspeaker with different dimensions from
the standard models, requiring a different mounting
procedure. Note that this manual is NOT applicable to these
versions; please contact JBL for more information.
Reference point for
acoustical mounting
h
eight
DC115
a
mp at bottom
DC115
a
mp at top
Reference point for
acoustical mounting
height
WHAT’S IN THE PACKAGING
In addition to the loudspeaker itself, each unit is shipped
with the following items:
• Installation manual (this document)
• Datasheet specific to the model
• Box containing:
A
B
I
C
J
G
D
E
F
H
REF ITEM QTY USE
A Mounting brackets 2 or 3* Speaker mounting
B Set of 30 mm hex-head screws with washers and wall plugs 4 or 6* Mounting bracket fixings
(wall)
C Female IEC mains connector 1 AC input
D 6-pole female WAGO connector, with cover 1 Mating connector
for audio
E 5-pole female WAGO connector, with cover 1 Mating connector for
RS-485 control port
F, G Set of PVC inserts for WAGO connectors, with wire insertion tool 1 WAGO connector wiring
H Set of M4 x 10 round-head screws with star retaining washers 4 or 8* Grille retaining screws
I M5 x 12 pan-head screws with shakeproof washers 8 or 12* Mounting bracket fixings
(loudspeaker)
J Caution Label 1 Mains cable labeling
* depending on model
fig.2
fig.3 - Box contents
The mounting brackets, screws and wall plugs as well as the connector set are packed together in a
separate box enclosed in the main packaging.
Page 7
7
JBL Intellivox Installation Manual
PREPARING FOR INSTALLATION
Before starting to install the Intellivox loudspeaker, a number of points should be considered.
• It is assumed that the installation site has been the subject of an extensive DDA investigation or optimisation prior to the actual install. (DDA = Digital Directivity Analysis, the proprietary simulation and optimisation software for these products*). This procedure will have identified the optimum physical location for the Intellivox, to accurately achieve the desired acoustic coverage. In particular, the “acoustic centre” position will have been specified, in terms of height above the finished floor. The installer should confirm that he/she has this information.
• Unless specified otherwise by the sound system designer, the Intellivox column should be installed at a position exactly perpendicular to the listening plane (i.e. vertical if the floor is horizontal). Unlike conventional loudspeaker systems, minor mechanical misalignment may lead to degraded coverage and intelligibility. Ensure (by use of a levelling device such as a spirit level) that the surface to which the Intellivox is to be attached is absolutely vertical and that all mounting positions (2 or 3 depending on model) are in the same plane.
• The Intellivox is configured to operate from either 230 V or 115 V AC mains. A suitable AC mains outlet should be provided adjacent to the loudspeaker’s intended mounting position, taking the position of the electronics module into account. The outlet must be on a circuit able to handle the full-load power consumption as well as supply the maximum in-rush current drawn by the electronics module; these figures
vary with model, and can be found in the datasheet supplied
with the unit. The third terminal of the mains outlet must be connected to a proper safety ground. Local safety regulations may require a separate mains distribution network or an uninterruptible mains power supply (UPS).
SIGNAL AND CONTROL CABLES
• In addition to the mains supply, some or all of the following
cables (shown in fig 4) will need to be run to the loudspeaker
position. The necessity for each of these will have been
determined by the system designer; the installer should ensure
that he/she has these details to hand. Depending on the specific
access issues, it may be better to ensure that these cables are
in place and correctly terminated “ready-to-plug-in” before the
Intellivox is mounted in position. All wiring details can be found
on pages 9-15.
• Audio inputs: the Intellivox accepts line level audio
(0 dBV). Two separate transformer-balanced audio inputs
are provided; however, units are shipped with only input 1
active. Input 2 can be made active via WinControl.
Alternative input sections are available separately:
accepting either one line level input and one 70 V/100 V
input or two 70 V/100 V inputs. This option needs to be
specified at the time of ordering.
• Network connection: all Intellivox products are
configured, and may also be monitored, from a PC running
WinControl software. With all installations it will be
3. INSTALLATION GUIDE
*Authorised electro-acoustic consultants or 'build & design' sound contractors can apply for a DDA License free of charge.
This license can be obtained through our website at http://www.jblpro.com/ For further information please refer to the Help files and manual with the DDA software.
Audio
RS-485
AC Mains
supply
External
Ambient noise
+ Temperature
sensor
Failure Relay /
Fan Control
REQUIRED
CONNECTIONS
OPTIONAL
CONNECTIONS
fig.4 - Intellivox Connections
Page 8
8
JBL Intellivox Installation Manual
necessary to upload the correct settings for each unit to obtain the desired acoustic coverage. Once this has been done, there is no longer an operational need to have the Intellivox connected to a PC because all settings are stored locally within the unit’s memory and in the event of a power failure, will be automatically reloaded on power-up. The system designer will have determined whether a permanent network connection needs to be provided. This will be the case if the Intellivox is to be monitored remotely during normal operation. Otherwise, the network wiring can be omitted, but consideration should still be given to the unit’s location with regard to ease of both initial and future access for a PC connection. It may be appropriate to install network wiring in order to provide a network access point at a location more easily accessible than the rear of the Intellivox unit itself. A cable with two individually foil­screened twisted pairs should be used. Details of suitable cables can be found on page 24.
• Fault monitoring: if network monitoring is not to be used, a simple surveillance function may be realised by means of the Intellivox’s ‘failure detect’ volt-free relay. This connection requires a 2-core cable.
• Additional cooling: if the ambient temperature is expected to exceed 40 °C an external fan should be installed. This may be activated automatically via the Intellivox’s temperature-controlled fan output; 2-core cable will be required. The fan should be mounted in a position which ensures sufficient airflow around the heatsink. For outdoor applications a weather-proofing kit is available as an option (see accessories in the appendix).
• Ambient Noise Sensing: the Intellivox is equipped with a built-in microphone to provide a signal for its embedded autogain algorithm, but an external ambient microphone (available as an option*) should be connected if specified by the system designer. Standard screened two-core microphone cable should be used for this purpose.
• Frost protection: the Intellivox is equipped with a frost protection system which activates an internally generated pilot tone if the ambient temperature drops below a pre-
determined threshold. The frost protection thresholds can be set using WinControl. If this function is required, an external ambient temperature sensor (available as an option*) should be connected with screened two-core cable (or similar) for its correct operation.
CONNECTOR AND WIRING DETAILS
All connections to the Intellivox (except mains power) are
made via the connector block at one end of the electronics
module. The IEC mains socket is immediately adjacent to
the connector block.
*Note that the optional external ANS microphone is combined with a temperature sensor in a single unit (JBL Order Code
97661101). If both functions are required, the use of 2-pair screened cable is recommended.
USE ONLY 230V FUSES
Audio input
connector
Network
connector
fig.5.1 - Intellivox Connector block
Page 9
9
JBL Intellivox Installation Manual
Additional access to the connectors is available via the connector plate located between the amplifier and loudspeaker baffle. Removal of the front grille will be necessary to gain access.
AC MAINS
The Intellivox is equipped with a 3-pin IEC
mains connector with an integral fuse holder.
Only ever replace the fuse with one of the same type and
rating. Never attempt to replace a fuse without first
removing the IEC plug from the unit. Be sure that the mains
voltage stated on the unit complies with the local standard
before connecting the Intellivox to the mains supply. Be sure
to use the appropriate cable diameter for the power
consumption of your particular Intellivox model.
A pre-made IEC mains cable (with the appropriate plug for
the territory) may be used if one of a suitable length is
a
vailable. Otherwise, use the mating IEC cable socket
supplied, and wire to the intended mains outlet. Note that
this connector is only suitable for use with stranded
“appliance” cable, not rigid mains installation cable.
A stranded wire clamp is available from JBL to make the
transition from rigid mains installation wiring to stranded
appliance cable, should no suitable mains outlet or
connection plate be available. Refer to page 22 for ordering
information.
Warning – This apparatus is a Class I device
and must be connected to a mains socket
outlet that provides a safety ground connection.
TERM EURO USA L Brown Black E Green/yellow Green N Blue White
fig.5.2 - Additional access to connector block via
the front cover plate
fig.6.1 - IEC mains cable wiring
L
E
N
fig.6.2 IEC mains cable label
Warning – The caution label contained in the
Intellivox standard connector set must be
placed in close proximity to the IEC female
cable plug as indicated in fig.6.2.
Page 10
10
JBL Intellivox Installation Manual
WAGO CONNECTORS
A
ll other Intellivox connections are made via the four on-
board WAGO Type 231 multi-pin connectors. These are:
• Audio (6-way male)
• Network (5-way male)
• Ambient noise mic/Ext. temp. sensor (5-way female)
• Failure detection/Fan control (5-way female)
One 5-way and one 6-way mating female connector is
supplied with every unit for the network and audio
connections. 5-way mating male connectors for the other
two sockets are available separately from JBL if required
(see page 22), but will be shipped with the unit if specified
at the time of order.
To wire the WAGO Type 231 connectors, proceed as follows:
1) Prepare the cable; strip back approximately 8-9 mm
(0.35”) of the insulation of each core.
2) Refer to the table below: select the appropriate PVC
insert for the wire size. If the wire is too big for the largest
size insert, the inserts can be ignored and the wire
connected directly (next step). Insert the wire into the insert
so that the bare end goes through the hole but the
insulated portion does not.
3) Using the wire insertion tool supplied as shown below,
open the clamp contact in the connector shell by pressing
i
nwards towards the shell.
4) Insert the stripped wire, with or without a PVC insert,
into the hole as shown. If using an insert, ensure that it is
fully seated. Release the wire insertion tool, and check that
the wire is secured. Repeat from Step 2 for all other cores.
WIRE SIZE INSERT COLOUR
0.08 to 0.2 mm
2
White
0.25 to 0.5 mm
2
Light grey
0.75 to 1.0 mm
2
Dark grey
fig.7.1 - Inserts
fig.7.2 - Using the WAGO insertion tool
Page 11
11
JBL Intellivox Installation Manual
AUDIO INPUTS
T
he Intellivox is normally supplied with two transformer-
balanced line inputs. Selection of Input 1 or Input 2 as audio
source is made from the WinControl software via the
network, and for most installations only Input 1 need be
wired. However, Input 2 may be connected to a secondary
signal path (in case of a failure elsewhere in the system), or
to an emergency sound source.
For optimum results only use good quality balanced audio
cable consisting of a twisted pair and an overall screen. The
audio source(s) (e.g. the Octadrive) should have a low
impedance balanced output. The nominal line input level
is 0 dBV.
For applications requiring direct connection to a 70 V/100 V
distributed loudspeaker system the Intellivox may be
ordered with an alternative input module having either
one 70 V/100 V audio input and one 0 dBV line level input
or two 70 V/100 V inputs. Refer to Appendix for order codes.
The audio connector pinout is as follows:
If the source equipment has a well-designed balanced
output and the wiring strategy shown below is used, there
s
hould be no problems with hum or other extraneous noise.
Note that pins 3 and 6 of the audio connector are analogue
audio grounds, they should be used only for connecting the
screen of the audio cables, and for no other purpose.
Due to the complex architecture of some large audio
systems, unexpected hum or noise may occur due to ground
loops, or injected noise from other equipment elsewhere in
the system. If this occurs, please refer to the Appendix (p24)
and to our website, where additional guidance on grounding
strategies will be found, with general techniques that may
be applied in order to reduce ground-induced noise.
PIN NO. STANDARD VERSION
1 X 70 V/100 V
1 X LINE VERSION
DUAL 70 V/100 V
VERSION
1
Line level
audio input 1
(0 dBV)
Line 1 +
70 V/100 V
input*
70 V/100 V +
70 V/100 V
input*
70 V/100 V +
2 Line 1 - 70 V/100 V - 70 V/100 V -
3 Line 1 Screen GND (n/c) GND (n/c)
4
Line level
audio input 2
(0 dBV)
Line 2 +
Line level
audio input
(0 dBV)
Line +
70 V/100 V
input
70 V/100 V +
5 Line 2 - Line - 70 V/100 V -
6 Line 2 Screen Line Screen GND (n/c)
1
2
3
Balanced source output (e.g. XLR)
Intellivox line level
audio input
(Line 1 only shown)
LINE 1+ LINE 1 ­LINE 1 SCN
1 2 3 4 5 6
fig.8.1 - Audio input connections
fig.8.2 - Balanced audio
connections
1 2 3 4 5 6
6 5 4 3 2 1
or
*A simple failure monitoring scheme based on impedance switching can be implemented on this 70 V/100 V input.
Refer to the Appendix (page 25) for details.
Page 12
12
JBL Intellivox Installation Manual
NETWORK CONNECTION
The RS-485 network connection should be permanently wired in any of the following situations:
• The Intellivox is to be constantly monitored in operation.
• The physical location of the Intellivox makes access to the network connector problematical.
• It is not practical (e.g. on a large scale installation) to configure each device locally.
The RS-485 interface permits multiple Intellivox units to be “daisy-chained” in parallel, so that they are controlled by the same PC. In such a system, each Intellivox must have its own unique network address; this is determined via WinControl for each individual loudspeaker.
The network connection should be wired as shown in the table and diagram below:
Note that the table above contains information with respect to 5-pin XLR wiring. A 5-pin XLR connector is commonly used to connect the network to the USB to RS-485 converter.
Note that 2-pair, individually-screened cable should be used for RS-485 connection. The transmit (Tx) and receive (Rx) balanced data lines must be wired via their own twisted pairs.
CAT-5 type UTP or FTP cable is NOT suitable. Please refer to the A
ppendix section of this manual for cable specifications.
The maximum cable length over which the network connection will operate reliably depends on the cable type and the baud rate used. With good quality cable, a safe maximum figure (at 19.2 kbaud) is 2000 m. If the distance is significantly longer than this, a network repeater will be required.
Before connecting multiple devices to the same
network subnet, ensure that their network addresses do not overlap. Each device should be set to a unique network address (determined via WinControl).
In some installations, it is normal practice to
disconnect the RS-485 network from the USB to RS-485 converter at the host (computer) during normal operation. This removes the possibility of extraneous data being inadvertently transmitted over the network. In such a situation, the converter is only reconnected when system maintenance or checks are required. Disconnecting from the converter in this way leaves the Rx lines of the devices unterminated. In most situations this does not present any problems, but with a large network in a hostile EMC environment, it is recommended that the host Y and Z lines (host Tx/device Rx) are terminated in a 100 ohm resistor for normal operation. This can be easily achieved by plugging the XLR5 female connector into a spare male connector with the resistor on the appropriate pins. This termination can then be removed and the converter reconnected when necessary. (If there is any doubt about which pins should be terminated, shorting all five pins will generally be equally effective.)
Further information on the RS-485 network can be found in the RS-485 converter manual and in the WinControl Help files.
to other devices
unit 3 unit 2 unit 1
host
A B Y Z
DGND
A B Y Z
DGND
A B Y Z
DGND
Y Z A B D
GND
network cable twin twisted i
ndividually
shielded
A +
B -
Y +
Z -
D
GND
RS-485 port
fig.9.1 - Multiple units “daisy chained” in parallel
fig.9.2 - Network connections
fig.9.3 - Twisted pairs
1 2 3 4 5 5 4 3 2 1
or
WAGO PIN NO. XLR5 PIN NO. LABELLED FUNCTION
1 5 A Data Rx + 2 4 B Data Rx ­3 3 Z Data Tx ­4 2 Y Data Tx + 5 1 DGND
Page 13
13
JBL Intellivox Installation Manual
FAULT MONITORING/FAN CONTROL CONNECTOR
T
he pinout of this connector is as follows:
Although accessed via the same connector, the fan control
and failure relay functions are independent of each other
and either or both may be used, as required.
The diagram above illustrates how this connector may
typically be wired.
Please note that Failure Relay and SPL/Temp
sensor connectors are both of the same type,
do NOT interchange!
Fault Monitoring
T
he Intellivox includes a single-pole changeover relay whose contacts are accessible on the Failure Relay connector. The conditions under which the relay operates are defined as part of the device configuration in WinControl. The relay connections may be configured either as volt-free, or with additional internal resistors to allow them to be connected directly to impedance-sensing line monitoring equipment*. This selection is made by moving an internal PCB jumper, refer to the Appendix for details.
In volt-free operation, COM is connected to NC; a failure condition causes COM to be connected to NO.
When configured for impedance-sensing use, the impedance between COM and NC is 10k Ω and open-circuit between COM and NO in normal operation. When a fault condition occurs, the impedance between COM and NC rises to 20k Ω, and COM is short-circuited to NO. Impedance-sensing fault monitoring equipment may be connected to COM and NC.
The relay contacts have maximum current and voltage ratings of 100 mA and 24 V respectively.
PIN NO. LABELLED USE
1 COM Failure relay Common
2 NO Failure relay Normally Open
3 NC Failure relay Normally Closed
4 + Fan +
5 - Fan -
To external
fault reporting
system
Intellivox
Failure Relay
Fan control signal
24 V DC
FAN –
FAN + FAULT NC FAULT NO
FAULT COM
(OR)
24 V
Fan
fig.10.1 - Failure relay / fan control connector
fig.10.2 Fault monitoring / fan control connections
5 4 3 2 1
1 2 3 4 5
or
*Intellivox devices with early serial numbers do not have the option of re-configuring the volt-free relay contacts for impedance-sensing operation. Please contact the service department for more information.
Page 14
14
JBL Intellivox Installation Manual
Fan Control
A
n external cooling fan should be installed if the ambient temperature is expected to exceed 40 °C and/or the Intellivox will be exposed to direct sunlight. For outdoor applications JBL offers an off-the-shelf solution (see weather-proofing kit in the appendix). Dedicated forced cooling solutions however can be constructed by means of the information provided in this section.
An internal temperature-controlled output is available at two pins of the connector; these pins should be wired to an external 24 V DC fan as indicated in fig 10.2. One of the pins is connected to an internal positive supply rail (FAN+, approximately 24 V), the other one (FAN-) is driven by an internal transistor. Note that external protection diodes are normally not required, the FAN- signal is diode-clamped both to the internal positive supply rail as well as to the internal GND.
The fan should be a 24 V DC type consuming less than 3 W. Applicable fan types are:
EBM-Papst 8314. EBM-Papst 8414. Sunon KDE2408.
Please note that the FAN- signal is a PWM (Pulse Width Modulated) signal. This control r
esults in a temperature dependent fan speed. It implies that this signal cannot be used to directly drive an external relay (e.g. to control a mains-voltage fan). C
ontact JBL for details with respect to the interfacing to
mains-voltage fans.
Ensure that the airflow of the fan enhances the natural airflow (caused by convection).
Note that the supply available at the fan control connector has a maximum current
rating of 125 mA. Fans or any other loads consuming more than 3 W should not be connected. This might result in the activation of an internal protection*.
*For Intellivox devices with early serial numbers this protection is non-resetable. Please contact the service department for more information.
Page 15
15
JBL Intellivox Installation Manual
EXTERNAL AMBIENT SPL/TEMPERATURE SENSOR
T
he pinout of this connector is as follows:
Although accessed via the same connector, the external
ANS microphone and temperature sensor functions are
independent of each other and either or both may be used,
as required. The use of the optional JBL combined Ambient
SPL/Temperature sensor is recommended, although either
of the two functions may be used alone if wished.
The diagram above illustrates how this connector should
be wired to the optional combined sensor using twin
twisted-pair cable with an overall screen. Note that there
are two AGND terminals on both the Intellivox and the
sensor; care should be taken not to confuse these, and in
particular, not to connect AGND to GND at any time.
E
xternal Ambient SPL Sensor
T
he Intellivox incorporates an autogain algorithm which
intelligently alters the volume of the audio according to the
ambient noise level. An internal microphone is fitted to
monitor the ambient noise level in the gaps in programme
material. The autogain function is activated and adjusted
via WinControl. This automatic volume adaptation feature
will only give good results if the signal measured by the
microphone is truly representative of the diffuse ambient
noise level, and in some circumstances it may be preferable
to monitor the ambient noise at a position remote from the
Intellivox itself. Connecting the external ambient noise
sensor automatically disconnects the internal microphone.
Care should be taken in selecting a suitable location for the
external sensor; it should not be near direct sound sources
such as people talking, air conditioning system outlets, or
other constant noise sources such as electrical plant, etc.
The most favourable external sensor locations are walls or
ceilings (height > 3 m) within the coverage range of the
Intellivox.
Temperature sensor
The external temperature sensor is for activating the
Intellivox’s frost detection system, and should be
connected if it is anticipated that the ambient temperature
may drop below 0 °C. If it does, an internally-generated
pilot tone is applied to the drivers, warming the voice coils
slightly, and thus preventing any build-up of moisture in
the magnet gap. Note that this pilot tone is always
generated if the internal load monitoring surveillance
function is enabled. In that case the ambient temperature
sensor is not required.
PIN NO. LABELLED FUNCTION
1 MIC ANS mic +
2 AGND ANS mic -
3 NTC Temp Sensor +
4 AGND Temp Sensor -
5 GND Screen
NTC
AGND
Mic
AGND
Shield Shield
AMBIENT SPL / TEMPERATURE
SENSOR
5 4 3 2 1
Pair 1 Pair 1
Pair 2 Pair 2
INTELLIVOX
fig.11.1 - External ambient SPL/Temperature
sensor connector
fig.11.2 - External ambient SPL/Temperature
sensor connections
5 4 3 2 1
1 2 3 4 5
or
Page 16
16
JBL Intellivox Installation Manual
MECHANICAL INSTALLATION
A
coustic Centre
M
ounting an Intellivox loudspeaker is a straightforward
procedure, but it is essential to understand that the height
of the unit above floor level is extremely critical. The
“correct operational height” is defined as part of the
acoustical design process and the installer should make
sure that he/she has this information before commencing to
mount the Intellivox.
The correct operational height defined by the system
designer is the height of the Intellivox’s acoustic centre. This
point does not necessarily coincide with the position of any
particular physical component on the unit. The location of
the speaker’s acoustic centre is marked by the centre of the
yellow sticker on the front face of the unit at the time of
shipping, and it is this position that must correspond to the
operational height defined in the design process. The
“vertical” position of the acoustic centre relative to the
bottom of the unit varies from one Intellivox model to
another, though it will always be on the vertical centre line.
When calculating the positions for the mountings (see page
18), the distance from the acoustic centre to the top or
bottom of the loudspeaker (whichever is being used as the
reference datum) should be measured and then allowed for
when marking the mounting positions. The model-specific
specification sheet supplied with each loudspeaker includes
a mechanical drawing clearly showing the mounting
position locations.
Because of the extreme tight vertical radiation pattern of
the Intellivox, a small deviation from the calculated
acoustical mounting height may cause severe degradation
of the expected performance.
Note also that the yellow sticker indicates
which way is “up”; ensure that this orientation
is observed.
Ensure the protective blue foam behind the
grille is removed before operation.
fig.12 Sticker showing acoustic centre and unit orientation
Page 17
17
JBL Intellivox Installation Manual
Mounting Options
The Intellivox has two or three (depending on the model) attachment points at the rear of the enclosure; the device should be mounted using the appropriate brackets fixed at these points. Each point consists of four M5 threaded inserts and only the screws supplied with the original mounting hardware should be used.
The threaded inserts are M5 metric and the depth is 16 mm.
There are three options for mounting the Intellivox column:
1. The first (and most often used) method is to use the standard mounting brackets supplied with the unit. Two or three brackets will be supplied, according to the Intellivox model. Dimensions of the bracket are shown in the drawing below.
The standard brackets allow mounting of the Intellivox on a straight wall or on a curved surface (e.g., a pillar) with a gap of 25 mm (1.0”) between the rear of the enclosure and the wall. If the unit requires three mounting points (as on the larger Intellivox models) ensure that all three points are in the same plane, to avoid the unit being bent when the screws are tightened.
The mounting procedure is given on page 18 under “Mounting Procedure”.
2. A second method is to mount the Intellivox using (optional) swivel brackets which allow the plane of the column’s vertical radiating axis to be at an angle other than 90 degrees to the wall. In this case two or three (depending on the model) swivel brackets are needed. The horizontal angle of the swivel brackets can be secured by the hexagonal nut. The swivel brackets are mounted to the rear
of the enclosure by the four M5 pan head screws instead of the standard brackets. Two versions of swivel bracket are available, permitting the Intellivox to be rotated through 45 or 90 degrees either way respectively. Refer to page 22 for ordering information.
3. The third method is to mount the column within a recess in the wall or dedicated mounting panel so that the front of the unit is flush. This is a specialised mounting situation which will be encountered only rarely. In this case longer M5 screws than those supplied will be required, and it should be noted that the threaded inserts on the rear of the enclosure are 16 mm deep. Access to the rear of the unit will still be required for the cables. Because of the proximity of the unit to the walls, particular attention should be paid to the avoidance of acoustic resonances.
With all mounting methods, it is advised to keep the rear side of the Intellivox free from the wall in order to prevent unwanted resonances. If mounting the unit onto or into a cavity wall, it is advisable to insert rockwool or similar sound-absorbing material into the cavity in the vicinity of the loudspeaker.
An Intellivox Accessories brochure, which includes a full range of optional brackets for various situations can be downloaded from http://www.jblpro.com
fig.13.1 - Standard bracket dimensions in mm
fig.13.2 ­Direction of rotation when using swivel brackets
Page 18
18
JBL Intellivox Installation Manual
Mounting Procedure
The general mounting procedure using the standard
brackets is described below. It is assumed that all necessary
c
abling is already prepared with connectors fitted as
described in the previous chapters.
Proceed as follows:
1. Use of the wall plugs and hexagon head screws
supplied is recommended. If using other types, ensure
that the “across-flats” head size does not exceed
11.5 mm (0.45”).
2. Mark and drill the holes (8 mm) for the wall plugs. Use
the two holes with the slots running vertically (refer to
Fig 13.1).
3. Insert the plugs and screws into the holes. Tighten the
screws with a spanner, but stop with a few mm of screw
shaft visible between the head and the wall.
4. Check that it is possible to slide a bracket over the
heads.
5. Check that the chosen mounting points will position
the column in the correct vertical angle specified by the
acoustic design (which is usually exactly perpendicular
to the listening area). Use spacers on one or more of
the brackets if necessary.
6. Using a Philips H2 Screwdriver, attach the brackets to
the Intellivox enclosure with the M5 x 12 pan-head
s
crews and shakeproof washers supplied. The 10 mm
dia. holes allow screwdriver access to the screw heads.
7. Lift the Intellivox and slide the brackets over the heads
of the hex-head screws protruding from the wall.
8. Firmly fasten the screws securing the brackets to the
wall. Re-check the verticality (or other angle if specified)
with a spirit level or similar levelling device.
9. Remove the front grille(s) from the Intellivox by gently
lifting the grille at the locations of the ‘snap-on’
fittings which hold it in place. Remove the protective
foam. Unscrew the four self-tapping screws securing
the connector compartment cover and remove the
cover. Thread all cabling through the connector
compartment from the rear so that it is free and
accessible from the front.
10. Terminate and plug in all the connectors (see fig 5.1)
and replace the connector compartment cover and the
grille(s)*.
11. Secure grille(s) at each corner with supplied M4 round
head screws and star retaining washers.
*If it is intended to connect a PC directly to the Intellivox to carry out the settings file upload (see page 19), leave the network connector (if any) unplugged and do not replace the connector compartment cover until the upload and verification procedure has been carried out.
fig.14 - Bracket attachment
Page 19
19
JBL Intellivox Installation Manual
After installation and connection are complete, the installer
s
hould upload the settings file into the Intellivox and verify
correct operation with WinControl software.
This section of the manual describes only the file upload
and a simple verification procedure. It assumes that a
laptop (or other computer) with both WinControl software
and a driver for the RS-485 interface already installed is
available, with the RS-485 interface. It also assumes that
the installer has the settings files, prepared by the system
designer, for each Intellivox loudspeaker making up the
system. These may be elsewhere on the laptop’s hard drive,
or may be supplied separately on removable media such as
a memory stick or a CD-ROM. No previous knowledge of
WinControl is necessary, but basic familiarity with Windows
file handling is required.
A full description of WinControl is beyond the scope of this
manual and further information is available in the
application’s Help files. Instructions on how to install
WinControl on a PC are included with the Program Set.
Instructions on the use of the RS-485 interface are included
with the interface itself.
CONNECTING THE PC TO THE INTELLIVOX
If connecting directly to the Intellivox, use the 5-pin XLR to
WAGO cable (supplied with the WinControl Program Set)
and plug the WAGO connector into the loudspeaker’s
network socket. Connect the other end to the RS-485
output socket on the RS-485 interface. If the network
connection to the Intellivox has been made permanent and
network access is consequently in a different physical
location, connect the PC at this point, using a different
(installation-specific) cable if necessary.
The RS-485 interface should then be connected to the
PC’s USB port.
4. UNIT CHECKS
Page 20
20
JBL Intellivox Installation Manual
SETTINGS FILE UPLOAD
Apply mains power to the Intellivox, turn on the PC and
launch WinControl. Open Options > Communications
options, and confirm that the Off-line box is unchecked
and that the correct COM port is selected. Close the
Communication options dialogue box. If the PC is
connected to only one Intellivox, a screen similar to that
shown below should now be displayed.
The green rectangle in the main pane represents the
Intellivox. The model number is shown below and the
unit’s current network address to the right. If a
complete network comprising multiple Intellivoxes is
being accessed, additional similar rectangles will be
displayed.
• If the display does not look like that shown, click on
Command > Netscan to force the PC to re-scan the
network. (This will be the case if the Intellivox(es) were
unpowered when WinControl was launched.) If this fails
to produce the correct display, please check all network
connections, and that the correct port on the PC is
selected (open Options > Communications options
to do this).
• Select the Intellivox to have its settings file uploaded
by clicking on it once, highlighting its model number
and address, if it is not already highlighted.
Click File > Load settings…, which will open the two
dialogue boxes shown below,
Load control parameters
s
ettingsand Control parameters
.
fig.15.1 - WinControl connected to single unit
fig.15.2 - Uploading the settings
File Name field Browse button
Page 21
21
JBL Intellivox Installation Manual
• The File Name field (at the top of the
Load control
parameters settings
dialogue box) indicates the
l
ast-used settings file name. If this is the first time
that the installation of WinControl has been used, a
default settings file is selected. The default location
for this file is in [My Documents folder]\Duran
Audio BV\WinControl\Settings, though as
this is alterable, it is possible that a different location
may be displayed, depending on how the particular PC
being used has been set up. To navigate to the location
containing the settings file to be uploaded, click on the
Browse button (the folder symbol to the right of
the File Name field), which will open a further
Open
dialogue box. Navigation may be continued in the
normal Windows manner. Note that Intellivox settings
files always have an *.ini file extension.
• Select the correct settings file and click Open. Then in
the
Load control parameters settings
dialogue box,
e
nsure that the proper parameter groups are selected
and specify the preset(s) to be uploaded, click OK* to
continue. An
Info
dialogue box will then open showing
additional information, in this box click OK to start the
upload process. This will transfer the pre-determined
Intellivox parameters to the loudspeaker. The
Control
parameters
dialogue box can now be closed and the
upload is complete.
• If there are further loudspeakers on the network to have
their settings files uploaded at this time, the procedure
can be repeated by selecting the next Intellivox. It is
possible to load the same settings file into several
Intellivoxes simultaneously, should the installation
require this. See
Addresses and Groups
in the
WinControl Help file for more information.
* See WinControl Help files for information regarding further Load options.
Page 22
22
JBL Intellivox Installation Manual
OPTIONAL ACCESSORIES
Listed below are a number of additional components which may be required for an individual installation.
Mating connectors and components for external wiring:
Other installation-related accessories:
5. APPENDIX
CONNECTOR DESCRIPTION ORDER CODE
Mains power IEC socket 6A 17824103
Audio Input WAGO 6-pin female 17959146
Network WAGO 5-pin female 17959145
Failure/Fan or Ext SPL Mic/Temp. WAGO 5-pin male 17949145
DESCRIPTION ORDER CODE
WAGO stranded wire clamp 17969406
Intellivox wall bracket, 25 mm 802225 (2pcs); 802226 (3pcs)
Intellivox wall bracket, 35 mm 802235 (2pcs); 802236 (3pcs)
Intellivox wall bracket, 60 mm 802260 (2pcs); 802261 (3pcs)
Intellivox hinge bracket 90° 802000
Intellivox swivel bracket 90° 806608
Intellivox swivel bracket 45° 806618
Small Hinge 806602 (2pcs); 807402 (3pcs)
Cover plate with 2 x 13.5 mm cable gland holes 802110
Cover box, 42 mm, with 2 x 16 mm cable gland holes 802105
Cover box, 58 mm, with 2 x 25 mm cable gland holes 802100
Cover box, 42 mm, with 6 x XLR connectors 802120
Cover box, 42 mm, pre-punched for 6 x XLR connectors 191810043
External ambient mic/temperature sensor 97661101
WAGO wire inserts (white) for wire dias. 0.08 to 0.2 mm 17949605
WAGO wire inserts (light grey) for wire dias. 0.25 to 0.5 mm 17949615
WAGO wire inserts (dark grey) for wire dias. 0.75 to 1 mm 17949625
WAGO wire insertion tool 17949901
Page 23
23
JBL Intellivox Installation Manual
Available input modules:
Weather proofing kits:
Network accessories:
DESCRIPTION ORDER CODE
Weather-proofing kit for Models DC115/180, DS115/180, DSX/180 (amp at top only) 20010
Weather-proofing kit for Models DC280, DC380, DS280, DS380, DSX280, DSX380, DSX280HD (amp at top only)
20012
Weather-proofing kit for Models DC430, DS430, DSX430 (amp at top only) 20014
Weather-proofing kit for Models DC500, DS500, DSX500 (amp at top only) 20016
DESCRIPTION ORDER CODE
Serial interface converter, USB to RS-485 387802
Program Set (inc. WinControl CD, USB interface, cable) 386602
DESCRIPTION ORDER CODE
1 x Line Level and 1 x 70V/100V 381001
Dual 70V/100V 381003
Dual 70V/100V with DC blocking 381005
DM1-TP Optocore SANE input board 381006
Page 24
24
JBL Intellivox Installation Manual
COMMON ANALOGUE GROUNDING ISSUES
C
orrectly connecting the Intellivox to ground has several
benefits:
1. Safety. The GND terminal of the IEC mains connector
provides a direct low impedance path from the
metal parts of the chassis to ground. Always
connect this terminal.
2. Reduction of RF emission. Although the Intellivox
electronics module is well shielded and external
connections are decoupled to prevent RF emission from
the internal high speed digital circuits, this protection
will not work properly if the chassis is not connected to
ground.
3. RF Immunity. RF currents induced in the signal cables by
external RF fields are effectively shorted to chassis-
ground, provided that the cable screen (shield) is of
sufficiently low impedance.
In addition to the mains ground, the Intellivox connector
block has ground pins marked in 3 different ways. Each
serves a dedicated purpose and care should be taken that
they are not interchanged or connected to each other.
• Pins marked GND are connected internally directly to
the chassis of the Intellivox. There are GND pins on the
audio line input connector and the ambient SPL
mic/temperature sensor connector. These GND pins are
for connecting the screen (shield) of the audio and
sensor cables.
• Two pins marked AGND are fitted to the ambient SPL
mic and temperature sensor connector, and serve as a
signal reference only. AGND should not be connected
to any other ground pin (GND or DGND) on the chassis.
• The RS-485 connector has one pin marked DGND and is
intended as a terminal for the network cable screen
(shield). Since the RS-485 interface is optically isolated
(to prevent ground loops), DGND has no relation to the
chassis ground. DGND should not be connected to any
other ground pin (GND or AGND) on the chassis.
SOFTWARE AND FIRMWARE UPDATES
T
he WinControl application is freely available and can be
downloaded from the download area of our website;
http://www.jblpro.com. We advise installers, users and
engineers to check our site regularly for updates. Intellivox
firmware and DSP software upgrades will also be made
available through the website.
For further information about how to use WinControl please
refer to the WinControl Help files.
NETWORK CABLES
The type of cable necessary for correct operation of the
RS-485 network is twin twisted pair with each pair individually
shielded. Numerous cables of this type are readily available
and cables broadly meeting the specifications of the
example cable given below are likely to be suitable.
Example of a preferred cable type:
PARAMETER VALUE
Type
BELDEN ‘Datalene’ series No. 9729 2-pair
Characteristic impedance 100 ohms
Capacitance (core to core) 41 pF/m
Capacitance (core to screen) 72.5 pF/m
DC resistance (core) 78.7 ohms/km
DC resistance (screen) 59.1 ohms/km
Page 25
25
JBL Intellivox Installation Manual
FAILURE MONITORING ON 70 V/100 V CONNECTION.
If the Intellivox is equipped with an input module that supports direct interfacing to a 70 V/100 V signal, a simple failure monitoring scheme requiring no additional connections can be easily implemented.
This scheme is based on impedance switching of the first 70 V/100 V input. An amplifier that is driving this input should be capable of detecting changes in connected impedance.
Impedance switching is supported for input boards PCB960 (see fig 16) and PCB124 (see fig 17). Configuration details of the impedance switching are depending on the input board type as described hereafter.
Reconfiguring the failure monitoring requires access to an internal jumper.
This requires the electronics module to be carefully removed from the enclosure by a qualified engineer.
Ensure that all connectors are unplugged before removing the electronics module.
C
onfiguring PCB960
T
he impedance between pin 1 and 2 of the audio input
connector has two states:
• High impedance (approximately 100 kOhm @ DC).
• Low impedance (approximately 2.2 kOhm @ DC).
The states are switched by a separate change-over contact
of the Intellivox's internal Failure Relay.
An internal jumper (JMP1) on the input board determines
the impedance states:
• Low impedance if the device status is OK, High
impedance if the device has a failure or is not powered. This
is the factory default configuration, the jumper is placed on
the contact pins of JMP1 that are closest to CONN1
(furthest away from the edge of the board, see fig 16).
• High impedance if the device status is OK, Low
impedance if the device has a failure or is not powered. The
jumper should be placed on the contact pins of JMP1
closest to the edge of the board.
• Always High impedance, regardless of the device status.
The jumper should be removed.
CONFIGURATION JMP1 NOTES
High-impedance on failure Towards CONN1 Factory default
Low-impedance on failure Towards PCB edge -
Always High-impedance Removed -
fig.16 - Input board PCB960 with location of jumper JMP1
Page 26
26
JBL Intellivox Installation Manual
C
onfiguring PCB124
T
he impedance between pin 1 and 2 of the audio input
connector has two states:
• High impedance (approximately 140 kOhm @ DC).
• Low impedance (approximately 2.2 kOhm @ DC).
The states are switched by a separate change-over
contact of the Intellivox's internal Failure Relay.
An internal jumper (JMP2) on the input board determines
the impedance states:
• Low impedance if the device status is OK, High
impedance if the device has a failure or is not powered.
This is the factory default configuration, the jumper is
placed on both contact pins of JMP2 (see fig 17).
• Always High impedance, regardless of the device status.
The jumper should be removed.
Notes
• The fault conditions that result in impedance switching
are user definable (these are the same as for the internal
Failure Relay, see section Fault Monitoring on page 13 for
details).
• The High impedance state is affected by the option 'DC
blocking' (as for example offered by part # 381005). In
case of the 'DC blocking' option, series capacitors result
in a very high input impedance for DC. The Low
impedance state is not affected by the 'DC blocking'
option.
CONFIGURATION JMP2 NOTES
High-impedance on failure Present Factory default
Always High-impedance Removed -
fig.17 - Input board PCB124 with location of jumpers JMP1 and JMP2
Page 27
27
JBL Intellivox Installation Manual
FAILURE RELAY CONFIGURATION
Intellivox units equipped with input board PCB124 (see fig.17) support configuration of the internal Failure Relay for either volt-free or impedance-sensing (10k / 20k Ω) operation.
Reconfiguring the internal Failure Relay requires access to an internal jumper.
This requires the electronics module to be carefully removed from the enclosure by a qualified engineer.
Ensure that all connectors are unplugged before removing the electronics module.
An internal jumper on the input board (JMP1) defines the operation:
• Volt-free mode: In volt-free operation, COM is connected to NC; a failure condition causes COM to be connected to NO. This is the factory default configuration, the jumper is placed on the contact pins of JMP1 that are closest to R29 (furthest away from the edge of the board, see fig 17).
• Impedance sensing mode: When configured for impedance-sensing use, the impedance between COM and NC is 10k Ω and open­circuit between COM and NO in normal operation. When a fault condition occurs, the impedance between COM and NC rises to 20k Ω, and COM is short-circuited to NO. The jumper should be placed on the contact pins of JMP1 closest to the edge of the board.
N
ONCCOM
C
onnect to NO or
NC as required
I
NTERNAL RELAY
V
OLT-FREE
CONFIGURATION
NOTE: Relay shown in
energised state - i.e., ‘no
fault’.
fig.18 - Volt-free mode
2 x 10 k
Ω
I
NTERNAL RELAY
IMPEDANCE-SENSING
CONFIGURATION
NOTE: Relay shown in
energised state - i.e., ‘no
fault’.
NONCCOM
fig.19 - Impedance-sensing mode
CONFIGURATION JMP1 NOTES
Volt-free Towards R29 Factory default
Impedance-sensing Towards PCB edge -
Page 28
JBL Professional
8500 Balboa Boulevard
Northridge, CA 91329 U.S.A.
© Copyright 2014 JBL Professional
www.jblpro.com
Loading...