ION PhoCheck User Guide

ION PhoCheck User Guide

PhoCheck

TIGER SELECT

Instrument User Manual V1.1

For Sales & Service Contact

2650 E. 40th Ave. • Denver, CO 80205

Phone 303-320-4764 • Fax 303-322-7242

1-800-833-7958

www.geotechenv.com

Register your instrument online to receive your Extended 5 year

Warranty. See page 44 of the Standard TIGER Manual for details.

rt Number: 861489

PhoCheck TIGER Select

Ion Science Ltd

About this Manual

This manual describes the function and operation of the TIGER Select instrument.

It is a supplement to the standard Phocheck Tiger manual (Part number: 861265) and should be used together with the standard manual.

Please read and understand both user manuals completely before operating the TIGER Select instrument.

Page 1 of 19

PhoCheck TIGER Select Ion Science Ltd

Contents Page

About this Manual

1

Contents

2

Statements

3

Responsibility for use

3

Caution

3

Quality Assurance

3

Disposal

3

Calibration Facility

3

Ion Science Contact Details

3

Introduction to PhoCheck TIGER Select

4

Getting Started

5

Understanding the Display Screen

6

Start up screens

6

Soft Key areas

6

Using your TIGER Select

7

TIGER Select mode

8-9

TIGER PC Software

10

Maintenance

11-12

Calibration

11

Probes

12

TIGER SELECT Parts

13-14

Main Assembly

13

Front End Filter Assembly

14

Accessories

Please refer to Tiger Manual

Instrument Warranty and Service

Please refer to Tiger Manual

Technical Specification

15

Manual History

16

Page 2 of 19

PhoCheck TIGER Select

Ion Science Ltd

Statements

RESPONSIBILITY FOR USE

Inadequate performance of the gas detection equipment described in this manual may not necessarily be self-evident and consequently equipment must be regularly inspected and maintained. Ion Science recommends that personnel responsible for equipment use institute a regime of regular checks to ensure it performs within calibration limits, and that a record be maintained which logs calibration check data. The equipment should be used in accordance with this manual, and in compliance with local safety standards.

LEGAL NOTICE

Whilst every attempt is made to ensure the accuracy of the information contained in this manual, Ion Science accepts no liability for errors or omissions, or any consequences deriving from the use of information contained herein. It is provided "as is" and without any representation, term, condition or warranty of any kind, either express or implied. To the extent permitted by law, Ion Science shall not be liable to any person or entity for any loss or damage which may arise from the use of this manual. We reserve the right at any time and without any notice to remove, amend or vary any of the content which appears herein.

!CAUTION!

It is essential that the PhoCheck TIGER is always used with a supplied PTFE 0.5 micron filter fitted to the front of the instrument. Without a filter, particles of debris and dust can be drawn into the detector inhibiting the function of the instrument. These filters are consumable and should be changed after every 100 hours of use. The frequency of replacement should be increased for dusty or moisture laden environments. Filters are available from your distributor or at www.ionscience.com .

Quality Assurance

PhoCheck TIGER has been manufactured in compliance with ISO9001:2000, which ensures that the equipment supplied to our customers has been designed and assembled reproducibly, from traceable components, and leaves Ion Science calibrated to stated standards.

Disposal

Dispose of PhoCheck TIGER, its components and any used batteries in accordance with all local and national safety and environmental requirements. This includes the European WEEE (Waste Electrical and Electronic Equipment) directive. Ion Science Ltd offers a take back service. Please contact us for more information. The PhoCheck TIGER field case material is recyclable polypropylene.

Calibration Facility

Ion Science Ltd offers a calibration service including the issue of certification confirming calibration with equipment traceable to national standards. A PhoCheck TIGER calibration kit is available from your distributor or service centre or at www.ionscience.com . Ion Science recommends annual return of all instruments for yearly service and calibration.

Ion Science Contact Details

 

 

UK Head Office

USA Office

German Office

Ion Science Ltd

Ion Science Americas LLC

The Way,

33 Commercial Drive

Fowlmere

Waterbury

Cambridge

VT 05676

SG8 7UJ

USA

UK

 

 

Tel: +1 802 244 5153

Tel: +44 (0)1763 207206

Fax: +1 802 244 8942

Fax: +44 (0) 1763 208814

Email: info@ionscience-

Email: info@ionscience.com

americas.com

Web: www.ionscience.com

Web: www.ionscience.com

Ion Science Messtechnik GMBH

Laubach 30 Metmann-Neandertal 40822

GERMANY

Tel: +49 2104 14480 Fax: +49 2104 144825 Email: info@ism-d.de Web: www.ism-d.de

Page 3 of 19

PhoCheck TIGER Select

Ion Science Ltd

Introduction to PhoCheck TIGER Select

Benzene vapour is carcinogenic. Benzene is a constituent of certain types of fuel, and used as a solvent, and in the production of a wide range of chemicals and plastics.

PID technology can easily detect Benzene vapour as well as many of other volatile organic compounds.

The Tiger Select uses a combination of a ultraviolet and chemical filtering to deliver a response to Benzene free from interferents.

When the Select upgrade is activated, data logging, H&S mode and ppb automatically become available.

At start up a specific Select screen appears however the instrument initially runs like a standard Tiger. When in Select mode the instrument takes control over many of the standard features.

The Tiger Select uses a standard 10.6 eV lamp and a selective 10 eV filter incorporated in the pellet/stack.

The lower eV output prevents sensing many of the gases co-present in Benzene containing atmospheres.

10.0eV 10.6eV

The Tiger Select starts up in normal running mode.

Ion Science recommend operating the instrument in the working environment for 10 to 15 minutes before use. This allows the tubes and instrument to acclimatize and ensures the best results.

Ion Science suggest that the TIGER Select is calibrated before each use and a new PID-Pre –Filter Tube is used.

With no tube fitted, the customer can check the general VOC level, if there are no VOC's present then no tube test is required.

Page 4 of 19

PhoCheck TIGER Select

Ion Science Ltd

Getting started

The TIGER Select has all the features and functions of the standard TIGER instrument with the additional feature of Benzene detection using the PID-Pre-Filter Tube in TIGER Select mode..

This manual describes the operation of the TIGER in the TIGER Select mode indicated by the following symbol.

When first turned on, the TIGER Select will run through a start up routine as described below in

‘Understanding the display screen’. The TIGER Select will then run in standard mode with all the standard TIGER functions. All standard functions are described in the Standard TIGER manual (part number 861265).

When in TIGER Select mode the instrument will default to the Benzene calibration.

PID-Pre-Filter Tube for Benzene.

The PID-Pre-Filter Tube is used in a specifically designed way with the TIGER Select. The instructions supplied with the tubes do not affect the operation of the TIGER Select.

Please note:

1.If your TIGER Select has just been calibrated please ensure that a new PID-Pre-Filter Tube is used when in TIGER Select mode.

2.A 10eV filter Pellet is fitted to the TIGER Select is used for Benzene and Isobutylene detection. To revert to the full standard TIGER function with the whole range of gases the 10eV filter should be replaced with a standard 10.6eV pellet. Ensure that the 10.6eV lamp is selected from the Tiger PC Software. A custom calibration should then be carried out.

Page 5 of 19