Washer Replacement Kit for Reliant Scale RSC600
Assembly, Installation and Operating Instructions
SA VE THESE INSTRUCTIONS
NOTE: Check all parts for shipping damage before using. In case of damage, DO NOT use the equipment.
Contact the Dealer/Carrier for further instructions.
SAFETY SUMMARY
GENERAL WARNINGS
DO NOT install or use this equipment without first
reading and understanding these instructions. If
you are unable to understand the Warnings, Cautions or Instructions, contact a healthcare professional, dealer or technical personnel before attempting to install this equipment - otherwise, injury or damage may occur.
The weight capacity is limited to the lowest rated
capacity of any one (1) of the components in
use (i.e. Patient Lift, Sling or Scale). The patient's
weight MUST NOT exceed the lowest rated capacity of any component.
INST ALLATION W ARNINGS
After ANY adjustments, repair or service and
BEFORE use, make sure all attaching hardware
is tightened securely.
Patient and sling MUST be removed from the lift
during ALL installation procedures.
This kit contains the following:
DESCRIPTION QTY
Washers 2
CHECKING FOR PROPER SCALE
INSTALLATIONN (FIGURE 1)
NOTE: Refer to the INSTALLATION WARNINGS in
the SAFETY SUMMARY of this instruction sheet.
Proper Shoulder Screw Installation
1. Examine shoulder bolt and nut securing the load
cell assembly to the boom mounting bracket. Look
for the following conditions:
A. The shoulder bolt head should be flush against
the outside of the boom mounting bracket.
B. The nut should be tight against the shoulder bolt
shaft.
C. The shoulder bolt with the nut installed should turn
freely in the load cell assembly mounting holes.
2. Perform one (1) of the following:
ANY CONDITIONS FROM STEP 1 ARE NOT
A.
MET - Proceed to STEP 3.
ALL CONDITIONS FROM STEP 1 ARE MET
B.
- Proceed to
TION in this instruction sheet.
PROPER WASHER INSTALLA-
3. Remove the nut from the shoulder bolt.
4. Position the shoulder bolt so the head is flush
with the outside of the boom mounting bracket.
5. Install the nut until it is tight against the shaft of
the shoulder bolt.
6. Repeat STEPS 1-2.
Proper Washer Installation
1. Examine the washers on the inside of the boom
mounting bracket to see if the washers are binding
against the corners of the boom mounting bracket.
2. Spin the washers on the inside of the boom mounting bracket to see if the washers spin freely.
3. If the washers ARE binding against the corners of
the boom mounting bracket
freely, the washers MUST be replaced. Refer to
REPLACING WASHERS in this instruction sheet.
and/or DO NOT spin
REPLACING WASHERS (FIGURE 1)
NOTE: Refer to the INSTALLATION WARNINGS in
the SAFETY SUMMARY of this instruction sheet.
1. Remove the nut, shoulder bolt and two (2) existing large washers securing the load cell assembly to the boom mounting bracket.
2. Position the load cell assembly into the boom mounting bracket, with the scale facing the direction shown
in FIGURE 1.
NOTE: The end of the scale with the Invacare logo
should be towards the inside of the lift.
3. Perform one (1) of the following:
A. ALL RELIANT MODEL LIFTS - Position one
(1) NEW large washer on each side of the load
cell assembly, inside the boom mounting bracket.
B. MODEL 9805 LIFTS - Proceed to STEP 4.
4. Align the mounting holes in the load cell assembly,
NEW large washers (if applicable) and the boom
mounting bracket.
5. Secure the load cell assembly with washers (if applicable) to the boom mounting bracket with the shoulder bolt and nut. Perform the following steps:
A. Insert the shoulder bolt through the boom
mounting bracket, large washers (if applicable)
and load cell assembly so the head is flush
with the outside of the boom mounting bracket.
B. Install the nut until it is tight against the shaft
of the shoulder bolt.
1
Shoulder Bolt (Head)
Washer
Large Washers
(NOT present on model 9805.)
*Shoulder Bolt
(Shaft)
*NOTE: Ensure the shaft of the shoul-
der bolt passes through both sides of
the boom mounting bracket.
Boom
Boom Mounting Bracket
Nut
Logo
Mounting Hole
(NOT SHOWN)
Load Cell
Assembly
Scale
450 Lift Swivel
Bar (600 Lift
Swivel Bar NOT
SHOWN)
FIGURE 1 - CHECKING FOR PROPER SCALE INST ALLATION / INST ALLING WASHERS
Invacare Corporation www.invacare.com