Intek RheoVac DR User Manual

TABLE OF CONTENTS
SECTION 1 GENERAL INFORMATION ...................................... - 1 -
1.1 INTRODUCTION .................................................... - 1 -
1.2 PRINCIPLE OF OPERATION .......................................... - 2 -
1.3 TECHNICAL SPECIFICATIONS ........................................ - 2 -
1.4 PRECAUTIONS AND RECOMMENDATIONS ............................ - 3 -
SECTION 2 I NSTALLATION ................................................ - 4 -
2.1 INSTALLATION ..................................................... - 4 -
2.2 RheoVac DR SYSTEM INSTALLATION/SITE SELECTION ................. - 4 -
2.2.1 Probe Site Selection ............................................. - 4 -
2.2.2 Electronics Unit Site Selection .................................... - 6 -
2.3 PROBE INSTALLATION .............................................. - 6 -
2.4 ELECTRICAL CONNECTIONS ........................................ - 8 -
SECTION 3 OPERATI ON .................................................. - 12 -
3.1 GENERAL INFORMATION .......................................... - 12 -
3.2 SYSTEM START-UP ................................................ - 12 -
3.3 PORTABLE USB DATA STORAGE/Warranty Registration Instructions ....... - 12 -
3.4 DISPLAY .......................................................... - 12 -
3.5 COMMUNICATIONS ................................................ - 13 -
3.6 DATA PROCESSING ................................................ - 14 -
3.7 CUSTOM SOFTWARE .............................................. - 15 -
SECTION 4 MAINTENANCE ............................................... - 16 -
4.1 GENERAL MAINTENANCE .......................................... - 16 -
4.2 CALIBRATION ..................................................... - 16 -
4.3 SPARE PARTS ..................................................... - 16 -
4.4 TROUBLESHOOTING ............................................... - 17 -
4.5 HARDWARE AND SOFTWARE MODIFICATIONS AND UPGRADES ...... - 19 -
SECTION 5 CUSTOMER SERVICE ......................................... - 20 -
5.1 QUESTION ON EXISTING HARDWARE ............................... - 20 -
5.2 TROUBLESHOOTING ............................................... - 20 -
5.3 FACTORY AND FIELD SERVICE ..................................... - 20 -
5.4 NEW EQUIPMENT AND SERVICES ................................... - 20 -
SECTION 6 CUSTOM I NFORMATION ...................................... - 21 -
6.1 UNIT IDENTIFICATION ............................................. - 21 -
6.2 CONFIGURATION .................................................. - 21 -
6.3 SPECIAL INSTRUCTIONS ........................................... - 21 -
Appendix A - RheoVac Instrument User Interface Software Appendix B - RheoVac DR Networking
Manual no. RV DR Rev. B
Intek, Inc. 2011
WARRANTY
Intek, Inc. warrants each RheoVac DR product to be free from defects in material and workmanship under normal use and service, Intek's obligation under this warranty being limited to making good any part or parts thereof which shall, within one (1) year after delivery of such product to the original purchaser, be returned to Intek with transportation charges prepaid and which Intek's examination shall disclose to its satisfaction to have been thus defective; this warranty being expressly in lieu of all other warranties, express or implied and all other obligation or liabilities on Intek's part. The purchaser will assume all responsibility and expense for removal, decontamination and reinstallation of equipment.
RheoVac instruments are manufactured under United States patent numbers 4,255,968, 5,485,754, 5,752,411, 6,385,974 and 6,526,755. Inte k, Rheotherm and RheoVac are registered trademarks of Intek, Inc.
Intek, Inc.
751 Intek Way
Westerville, Ohio 43082-9057
Phone (614) 895-0301 Fax (614) 895-0319
web site www.intekflow.com
e-mail techsupp ort@intekflow.com
sales@intekflow.com
SECTION 1 — GENERAL INFORMATION
1.1 INTRODUCTION
For the first time, all necessary properties of the gases in the vacuum dryer exhaust line are directly measured to provide an accurate determination of water vapor removal and dryer performance related parameters. These properties are measured to provide the industry with the most advantageous and complete product for vacuum dryer system diagnostics:
®
the RheoVac
A RheoVac DR system consists of sensor probes reporting to a central signal conditioner and processor unit. The sensing probes, consisting of multiple sensors configured in two separate probes (the standard configuration is a main FTP probe and a RS probe), is installed in the vacuum line, generally between the dryer and the exhauster. The RheoVac DR system is superior to all other methods in that it makes no assumptions about the dynamics of the dryer and vacuum line environment. The main FTP probe sensor head employs the patented Rheotherm® technology to provide an accurate mass flow measurement. Additionally, temperature,pressure and water vapor relative saturation measurements are made using a high accuracy platinum resistance temperature detector (RTD), a strain gauge pressure sensor and a specially configured and calibrated water vapor saturation sensor.
System*
*USPNs 4,255,648; 5,485,754; 5,752,411; 6,385,974; 6,526,755
Figure 1 RheoVac DR
- 1 -
1.2 PRINCIPLE OF OPERATION
The principal features of the RheoVac DR system are shown in Figure 1. At the heart of the RheoVac DR system is the Rheotherm mass flow transducer, which uses the same patented thermal sensing technique employed in all precision flow instruments manufactured by Intek. Two temperature sensors are used one sensor is in thermal equilibrium with the flow medium and provides a temperature and flow signal reference, while the second sensor is located near a constant power heater so that its temperature is always above that of the fluid. The temperature of the heated sensor will vary with the stream velocity of the fluid. Hence, the measured temperature differential between the reference sensor and heated sensor is a function of flow rate, which is approximately proportional to the logarithm of mass flow rate (USPN 4,255,968).
The Rheotherm flow sensor is calibrated to measure the total mass flow of the water vapor/air mixture. From the other three measurements - temperature, pressure and relative saturation - the RheoVac DR electronics converts the total mass flow signal from the probe into two components, air mass flow rate and water vapor mass flow rate. This unique measurement method is disclosed in two separate patents (USPN 5,485,754 & 5,752,411).
The RheoVac DR system is fully calibrated in the factory under dynamic fluid conditions similar to those within the dryer vacuum line. Typically, field adjustments are not required.
1.3 TECHNICAL SPECIFICATIONS
1.3.1 Sensor (Probe) Specifications
Primary Cali bration Accuracy:
±5% of total mass flow
Repeatability:
±0.5% of reading
Operating Temperature:
Electronics: 40 to 120ºF (5 to 49ºC) Probe: 40 to 160ºF (5 to 71ºC)
Never subject probe to temperatures above 210ºF (99ºC)
Operating Pressure:
0.5 to 30 inches Hg absolute 15 psi maximum
Storage Temperature:
-20 to 210ºF (-29 to 99ºC)
Storage Pressure:
15 psig (maximum)
Process Connection:
NPT Hot tap assembly (optional, 1½” thread-o-let must be welded to pipe for hot tap installation)
Wetted Surface:
300 Series SS and engineered plastic
- 2 -
1.3.2 Main Electronics (Remote or PC) Signal & Data Access
Local Display:
Back-lit LCD Selectable display of air mass flow and 6 additional instrument output parameters Parameter scrolling Metric/English units
Input Power:
100-250 Vac, 50/60 Hz
Signal Output or Data Access:
RS-232/RS-422/Serial Modbus, Ethernet, TCP/IP Eight 4-20mA signals (optional) Wireless (Optional) OPC (Optional)
Temperature Environment:
Operating: 40 to 120
/F (5 to 49/C)
Storage: -20 to 210/F (-29 to 99/C)
1.4 PRECAUTIONS AND RECOMMENDATIONS
Read the entire manual before installing and operating the RheoVac DR system.
Carefully select the best location for installation of the probe(s). Access, clearances, freedom from standing water, and straight-run should all be considered when selecting probe locations.
Use reasonable care in handling the probes the sensing components are delica te. Do not bend the probes, damage the tips, or obstruct the sensing ports. If shipping the unit, make sure the probes are adequately protected from foreign objects and damage; save and reuse factory provided custom probe protector and shipping boxes.
Use proper input power it must be between 100 and 250 Vac (nominal 120/240 Vac) at 50/60 Hz (60 Hz nominal).
Confirm the line and environmental temperature is always below the probe and electronics ratings never operate a probe at or subject it to temperatures or pressures beyond its specified limits. (See SECTION 1.3)
.. WARNING - Never allow live high temperature steam to flow either direction in
the line where a probe is located.
Keep moisture out of the enclosures once all service connec tions are made, make sure the enclosure lids are tightly closed and all gaskets are in place. Seal all conduit lines.
The RJ45 is the recommended network connection for all data traffic (as opposed to serial and 4-20 mA communication).
Intek recommends a service contract to ensure probes are within calibration specifications and electronics are maintained with appropriate updates. Instrument probes should be returned to the factory for inspection and calibration service every two years.
- 3 -
SECTION 2 — INSTALLATION
2.1 INSTALLATION
These instructions are generalguidelines for the installation of RheoVac DR instruments in their standard configuration. Additional information pertaining to your unit is covered in SECTION 6 CUSTOM INFORMATION. Carefully read these instructions prior to installing the equipment. Also, see preceding SECTION 1.4 PRECAUTIONS AND RECOMMENDATIONS.
2.2 RheoVac DR SYSTEM INSTALLATION/SITE SELECTION
The RheoVac DR can be configured with one or two probes, with the two probe being the standard configuration. In the standard two-probe configuration, the main probe (FTP) is the larger probe, and the relative saturation (RS) sensor is located in a second probe, separate from the remaining primary sensors in the FTP probe. The RS probe is typically installed in the same line at a location slightly downstream of the FTP probe.
2.2.1 Probe Site Selection
The location for the probe pair should be selected so as to provide the probe sensing area with well-established smooth flow, uniform system temperature and pressure, and consistent non-liquid phase flow medium. Pipe sections ahead of a probe, in which water can accumulate, must be avoided. Refer to Figure 2 and select the most preferred location for the probe pair. Do not install the probes beyond any “trap” sections as shown in Figure 2, Configurations B and D. Special installation instructions unique to your unit, where applicable, will be noted in SECTION 6.3 SPECIAL INSTRUCTIONS. Refer to this section now to review special instructions.
Check installation clearance. The installation location should allow sufficient clearance for ease of probe removal (for servicing), including the required clearance space for the hot tap, where used. Be sure there are no obstructions around the vacuum line that will interfere with probe insertion. Figure 3 shows the proper insertion angle. THIS ORIENTATION IS IMPORTANT FOR PROPER OPERATION, particularly if liquid water is expected to wet the wall surfaces of the probe(s).
Observe the selected sites; check for ease of access. They should be convenient for the removal and replacing probes at any time for service; without requiring ladders, building scaffolding or waiting for plant shutdown.
Check operating conditions. The temperature and pressure limits (see TECHNICAL SPECIFICATIONS; SECTION 1.3) of the unit should be checked to ensure compatibility with your application.
- 4 -
Figure 2 RheoVac DR Probe Insertion Recommendation
Figure 3 Probe Installation Detail
- 5 -
2.2.2 Electronics Unit Site Selection
The RheoVac DR system may have up to three electronics enclosures; a main processor box, an optional distribution box, and an optional transmitter box. These enclosures should be installed in a convenient indoor location and should be kept away from direct sources of heat. The maximum temperature rating of electronics is 120°F; ensure that this temperature will not be exceeded inside the enclosure. Once the wiring connections are made, close and latch down the box lid to protect the contents from damage and debris. The enclosures should be located in a dry area and should be kept clamped shut during normal operation. Do not allow water to get into the enclosures. If installed outdoors, build a roof over all enclosures to prevent potential water infiltration, or direct sunlight to overheat contents.
A. Processor Enclosure: This 12x10 NEMA 4 enclosure houses the display and central
processing unit. Input power (100-250 Vac, 50/60 Hz) is connected inside this enclosure. This enclosure should be installed in a convenient location and should be kept away from direct sources of high heat, such as uninsulated steam lines.
B. Distribution Box (optional): This 8x6 NEMA 4 enclosure is typically located near the probe
that is closest to the processor enclosure unit. It is connected to the main processor with an RS-485 bus DeviceNet™ cable, which can be hundreds of feet long. The probe and transmitter RS-485 cable connections are also made within this box. The probe and
transmitter cable lengths should be 15 ft. or less.
C. Transmitter Box (optional): This 10x8 NEMA 4 enclosure contains terminals for accessing
the optional eight 4-20mA signals. It can be located in or near the control room so that the 4-20 wires do not have to be run from the plant floor. An RS-485 DeviceNet cable connects the distribution box to the transmitter box. (There are restrictions on how this can be done, so check manual SECTION 6, or contact the factory.)
2.3 PROBE INSTALLATION
!!WARNING!! The FTP probe and the RS probe(s) contain components and materials that have a temperature rating limit of 210°F. If line and fluid temperatures are anticipated to exceed this limit, the probe(s) must be removed from the vacuum pipe before and must not be reinstalled until the problem condition has passed. Failure to do this can result in severe damage to the sensor probe(s), especially when steam enters the vacuum pipe due to vacuum pump operating anomalies. If Steam Jet Air Ejectors are used, always pump start up and do not reinstall the RS probe until stable operating conditions are achieved.
A. NPT/Hottap Installation
1. Check hardware. Verify that the probe slides easily through the NPT fitting or hottap assembly and pipe penetration hole.
2. Verify there are the necessary minimum probe insertion clearance four (4) feet- between pipe surface and any obstruction.
3. For NPT fitting installation, ensure that the threaded mating fittings (not provided) for mounting the sensing probe(s) are installed at the proper locations in the process line. Typically, 1" threaded couplings are needed for the FTP and the RS probes.
remove the RS probe from the line before
- 6 -
4. If using a hottap: install the mounting hardware; drill a 1½” through-hole, center the thread-o-let over the hole and weld it onto the dryer vacuum pipe; thread the hottap assembly into the thread-o-let. Use thread tape or pipe dope to seal the connection. Alternate: weld thread-o-let to pipe wall, then drill a 1½" hole in pipe wall using a hottap drill.
5. Make sure the probes are parallel to the floor (see Figure 3). Be sure location is accessible for probe removal and maintenance.
B. Installing/Removing the Probe
1. Check proper installation direction and probe orientation. The FTP probe has a directional arrow on the junction box. Before installing the unit, note proper flow direction. This is important to instrument operation.
2. Check serial number (S/N). If more than one RheoVac DR system has been purchased, make sure the first five digits of the serial numbers of the probe(s) match the first five digits of the serial number of the main processor unit. The electronics and probe pairs are a matched set. Mismatched components will not work correctly. The dash number on the probe S/N is the probe number shown on the display. Record the probe pair number and installation location for future reference.
3. Verify stop clamp location (see Figure 4). A stop clamp is attached to each FTP probe as an indication of its insertion depth. It is important that the stop clamp is securely in place to position the sensors in the correct location and to ensure that the probes do not contact the opposite pipe wall. Contact with the pipe wall could damage the probe. The clamp’s location is determined based on your submitted pipe diameter, as shown in SECTION 6.2, and is marked with a groove on each probe shaft. Refer to this mark if a stop clamp is inadvertently moved. When installed in the line, the sensors in the probe should be positioned in the middle of the pipe.
4. Prior to inserting probe, loosen the compression nut on the thermocouple connector of the hot tap or the NPT fitting and clean the inner surface of the thermocouple connector to ensure it is free of particles that may cause probe damage.
5. Be sure to power up your RheoVac DR instrument system and probe pair for at least 30 minutes before inserting probes into the vent line. DO NOT leave probes in vent line without power.
6. Install probes. The probes should be mounted through the pipe wall using the hottap assembly or NPT fitting. a. The FTP probe installs so that the two sensor tips are
side-by-side across the gas stream. Each probe has a flow directional arrow on the junction box. Make sure the probe orientation is correct.
b. The RS probe does not have an orientation requirement.
The RS probe has a 4 foot (1.2 m) cable with a plug-in connector (with attached dust cap) for connecting to the FTP probe. Unscrew the dust caps on the RS cable and the FTP probe junction box, plug the cable connector in securely first, before inserting the RS probe in the line.
- 7 -
Figure 4 Probe Stop
Clamp
7. When installing under vacuum, do not allow the clamp to “slam” against the seal nut
upon opening the valve. Grasp each probe firmly, with hand against the seal nut, before opening the ball valve. Allow the probe shaft to slide slowly through the valve by controlling the amount of grip on the probe shaft. Do not let it slam against the seal nut. Once fully inserted, tighten the seal nut. Special installation instructions, if any, will be noted in SECTION 6.
8. It may be necessary to apply a force of about 23 lb (102 Newtons) to remove or
replace the probe under operating conditions. Firmly grip the probe shaft when removing a probe from the line.
2.4 ELECTRICAL CONNECTIONS
IMPORTANT Inspect and VERIFY these electrical connections carefully. Improper connection could damage electronic components and sensor function. If additional holes need to be drilled in the processor enclosure, remove the electronics subassembly (mounted on a mounting plate) and temporarily store inside an ESD bag in a safe, clean place. Do not drill with electronics boards inside the enclosure.
A. Main Processor Unit (see Figure 5)
1. Sensor Power and Communication Line
: Connect the distribution box to this main processor box using the RS-485 communications/power cable. Follow indicated connector color code. [communications: white (A), blue (B) and shield (SH); power: 24Vdc, red (+), and black (
2. Main Power
: Connect main power terminals to a dedicated 100-250Vac, single phase,
!)]
15-amp circuit. An external disconnect switch should be used for disconnecting power to the system during outages. Power connection wires should be at least 18 gauge and comply with accepted wiring codes. SW1 on the power input PWA (printed wiring board #01011-5) is used for cycling power to reset.
3. Network Connection (Recommended)
: The Ethernet connection at CN3 on the CPU interface PWA (printed wiring board #08017-1) is an RJ-45 style jack. A 10-foot Ethernet Cat5 crossover cable is supplied with the unit for laptop connections. Intek recommends using this connection for all data transmissions and RheoVac communications because: a. More measured data is accessible
through the network connection.
b. Software and calibration file updates
can be done remotely.
- 8 -
Figure 5 Electrical Connections and Set-up
4. Serial Communication: Connector JP3 on the CPU interface PWA (printed wiring board #08017-1) is the RS-232 serial communication interface. This interface should only be used for distances of 20 feet or less, such as to a laptop computer. A 20-foot serial cable with a DB-9 connector is available from Intekupon request (see Figure 6 and Table I). An RS-422 serial communication interface is present for long data communications when configured without 4-20 mA outputs. Note: Intek recommends using the
network connection for all data transmissions and RheoVac communications.
Figure 6 Serial Communication Interface
TABLE I. RJ-11 to DB Module Adapter
RS-232 CONFIGURATION RS-422 CONFIGURATION
RJ-11 Pin Out DB-9 Pin Out RJ-11 Pin Out DB-9 Pin Out
1 Tx (transmit) 1 N/C 1 Tx+ (transmit+) 1 Rx- (receive-)
2 N/C 2 Tx (transmit) 2 Tx- (transmit-) 2 Rx+ (receive+)
3 Rx (receive) 3 Rx (receive) 3 Rx+ (receive+) 3 Tx+ (transmit+)
4 N/C 4 N/C 4 Rx- (receive-) 4 N/C
5 Power (+5V) 5 Ground 5 Power (+5V) 5 Ground
6 Ground 6 Pulled high 6 Ground 6 Tx- (transmit-)
7 N/C 7 7 TBD
8 Pulled high 8 8 TBD
9 N/C 9 9
B. Distribution Box (optional) (see Figure 7)
1. Connect the RS-485 wires (blue, white and shield) and 24Vdc power (red and black)
from the main processor unit.
2. Install ½” liquid-tight conduit between the
distribution enclosure and the probes unless ½” rigid conduit is used for long distance runs. Use a minimum of 6 feet of liquid­tight at the probes.
3. The probe connector comes with an adapter
which allows the attachment of the ½” flexible conduit connector.
4. Connect the probes to the distribution
enclosure using the manufacturer supplied four conductor shielded cable to any of the screw terminals labeled JP3 to JP6. Probe cable connections are shown in Figure 7.
5. Optional - Connect the transmitter box to the
distribution box using the supplied RS-485 cable.
Figure 7 Distribution Box
- 9 -
Loading...
+ 23 hidden pages