Appendix A - RheoVac Model 950 User Interface Software
Appendix B - RheoVac Model 950 Networking
Intek, Inc. 2007
Manu al no. RV950 Rev. D
I:\OFFICE\WPMANU AL\RV950 Rev D.wpd
Page 2
WARRANTY
Intek, Inc. warrants each RheoVac product to be free from defects in
material and workmanship under normal use and service, Intek's
obligation under this warranty being limited to making good any part or
parts thereof which shall, within one (1) year after delivery of such
product to the original purchaser, be returned to Intek with transportation
charges prepaid and which Intek's examination shall disclose to its
satisfaction to have been thus defective; this warranty being expressly in
lieu of all other warranties, express or implied and all other obligation or
liabilities on Intek's part. The purchaser will assume all responsibility
and expense for removal, decontamination and reinstallation of
equipment.
RheoVac instruments are manufactured under United States patent numbers
4,255,968, 5,485,754, 5,752,411 and 6,526,755. Intek, Rheotherm, Rheovec,
Rheomax, RheoVac and RheoSmart are registered trademarks of Intek, Inc.
Windows is a registered trademark of Microsoft Corporation in the United States and other countries.
Intek, Inc.
751 Intek Way
Westerville, Ohio 43082-9057
Phone (614) 895-0301 Fax (614) 895-0319
web site www.intekflow.com
e-mail techsupport@intekflow.com
Page 3
SECTION 1 — GENERAL INFORMATION
1.1INTRODUCTION
For the first time, all necessary properties of the gases in the condenser exhauster line are directly
measured to provide an accurate determination of air in-leakage and condenser performance related
parameters. These properties are measured to provide the power industry with the most advantageous
and complete product for condenser system diagnostics:
the RheoVac®System*
*USPNs 4,255,648; 5,485,754; 5,752,411; 6,526,755
A model 950 RheoVac system consists of one, two or three probes reporting to a central signal
conditioner and processor unit. The sensing probe, consisting of multiple sensors configured in a single
probe head, is installed in the vacuum line between the condenser and the exhauster. The RheoVac
system is superior to all other methods in that it makes no assumptions about the dynamic condenser and
vacuum line environment. The sensor head employs the patented Rheotherm
an accurate mass flow measurement. Additionally, temperature, pressure and water vapor relative
saturation measurements are made using a high accuracy platinum resistance temperature detector
(RTD), a strain gauge pressure sensor and a specially configured and calibrated water vapor saturation
sensor.
®
technology to provide
Figure 1 RheoVac Model 950
- 1 -
Page 4
1.2PRINCIPLE OF OPERATION
The principal features of the RheoVac system are shown in Figure 1. At the heart of the RheoVac
system is the Rheotherm mass flow transducer, which uses the same patented thermal sensing technique
employed in all precision flow instruments manufactured by Intek. Two temperature sensors are used
one sensor is in thermal equilibrium with the flow medium and provides a temperature and flow signal
reference, while the second sensor is located near a constant power heater so that its temperature is
always above that of the fluid. The temperature of the heated sensor will vary with the stream velocity
of the fluid. Hence, the measured temperature differential between the reference sensor and heated
sensor is a function of flow rate, which is approximately proportional to the logarithm of mass flow rate
(USPN 4,255,968).
The Rheotherm flow sensor is calibrated to measure the total mass flow of the gaseous water vapor/air
mixture. From the other three measurements, the RheoVac electronics converts the total gas mass flow
signal from the probe into two components, air mass flow rate and water vapor mass flow rate. This
unique measurement method is disclosed in two separate patents (USPN 5,485,754 & 5,752,411).
The RheoVac system is fully calibrated in the factory under dynamic fluid conditions identical to those
within the power plant vacuum line. No field adjustments are required.
1.3TECHNICAL SPECIFICATIONS
Primary Calibration Accuracy:
±5% Total Mass Flow
Repeatability:
±0.5% of reading
Operating Temperature:
Electronics: 40 to 120°F (5 to 49°C)
Probe: 40 to 160°F (5 to 71°C)
Never subject probe to temperatures
above 210°F (99°C)
Operating Pressure:
0.5 to 30 inches Hg absolute
15 psi maximum
Process Connection:
Hot tap assembly
(1½” thread-o-let must be welded to
pipe for hot tap installation)
Wetted Surface:
300 Series SS and engineered plastic
1.4PRECAUTIONS
Read the entire manual before installing and operating the RheoVac system.
Local Display:
Back-lit LCD
Selectable display of air in-leakage and 6
of the other output parameters
Carefully select the best locations for installation of the probes. Access, installation clearances,
freedom from standing water, and straight-run should all be considered when selecting a probe
location.
- 2 -
Page 5
Use reasonable care in handling the probe the sensing components are delicate. Do not
bend the probes, damage the tips, or obstruct the sensing ports. If shipping the unit, make sure
the probe is adequately protected from foreign objects and damage.
Use proper input power it must be between 100 and 250 Vac (nominal 120/240 Vac) at
50/60 Hz (60 Hz nominal).
Confirm the line and environmental temperature is always below the probe and electronics
ratings never operate a probe at or subject it to temperatures or pressures beyond its
specified limits. (See Section 1.3)
.. WARNING - Never allow live high temperature steam to flow either direction in the
exhauster line where a probe is located. This can happen if steam jet ejectors are
operated incorrectly.
.. WARNING - Do not allow the instrument’s RS sensor (or separate RS probe) to come
into contact with liquid water, water mist, or fog – these conditions may be present in
your condenser vent line due to design configuration. The RS sensor measures relative
saturation up to 100% but may fail under the above wet condition. Intek provides design
services to improve condenser venting and reduce or eliminate entrained liquid water and
excess condenser back pressure.
Keep moisture out of the enclosures once all service connections are made, make sure the
enclosure lids are tightly closed and all gaskets are in place. Seal all conduit lines.
.. WARNING - Be sure to power up your RheoVac instrument system and probe(s) for
at least 30 minutes before inserting probes into the vent line hot taps. DO NOT leave
probe in vent line unpowered or when flooding the condenser.
- 3 -
Page 6
SECTION 2 — INSTALLATION
2.1INSTALLATION
These instructions cover installation of RheoVac instruments in their standard configuration. Additional
information pertaining to your unit is covered in SECTION 6 CUSTOM INFORMATION . Carefully
read these instructions prior to installing the equipment. Also, see preceding SECTION 1.4.
2.2RheoVac SYSTEM INSTALLATION/SITE SELECTION
The RheoVac 950 can be configured with one, two or three probes. A three probe system is intended
to be installed so that one probe is in each of the two exhaust lines coming out of the condenser (A side
and B side). The third probe should go in the header pipe that runs to the exhauster(s). Other
installation arrangements of multiple probes should be discussed with the factory.
2.2.1 Probe Site Selection
#The location for each probe should provide the probe sensing area with well-established
smooth flow, uniform system temperature and pressure, and consistent non-liquid phase flow
medium. Pipe sections ahead of a probe, in which water can accumulate, must be avoided.
Refer to Figure 2 and select the most preferred location for each probe. Do not install the
probes beyond any “trap” sections as shown in Figure 2, Configurations B and D. Special
installation instructions unique to your unit, where applicable, will be noted in SECTION
6.3 SPECIAL INSTRUCTIONS. Refer to this section now to review special instructions.
Figure 2 RheoVac Probe Insertion Recommendation
- 4 -
Page 7
#Check installation clearance. Each transducer probe is almost 3 feet (0.9m) long and the hot
tap assembly is approximately 13 inches (0.33m) long, therefore, allow minimum clearance
of 4 feet (1.3m) for probe installation. Be sure there are no obstructions around the vacuum
line that will interfere with probe insertion. Figure 3 shows the proper insertion angle. THIS
ORIENTATION IS IMPORTANT FOR PROPER OPERATION.
#Observe the selected sites. They should be convenient for the removal and replacement of
probes at any time for service without building scaffolding or waiting for plant shutdown.
#Check operating conditions. The temperature and pressure limits (see Technical
Specifications; Section 1.3) of the unit should be checked to ensure compatibility with your
application.
#Gases in the air removal line should be free of liquid water, mist or fog. Wetness in the line
will result in erroneous readings from the instrument. If wet conditions exist in the air
removal line, there is either air removal section damage or design flaws which adversely
affect condenser performance. Consult Intek for assistance in evaluating the severity of the
problem and possible remedies.
2.2.2 Electronics Unit Site Selection
The Model 950 typically has two or three electronics enclosures (see Figure 1). The processor
enclosure and other electronics boxes should be installed in a convenient location and should be
kept away from direct sources of heat, such as uninsulated steam lines, or moisture. The
maximum temperature rating of electronics is 120°F; ensure that this temperature will not be
exceeded. Once the wiring connections are made, close and latch down the box lid to protect the
contents from damage and debris. All these enclosures should be located in a dry area and should
be kept clamped shut during normal operation. Do not allow water to get into the enclosures. If
installed outdoors, build a roof over all enclosures to prevent eventual water infiltration.
A.Processor Enclosure: This 12x10 NEMA 4 enclosure houses the display and central
processing unit. Input power (100-250 Vac, 50/60 Hz) is connected inside this enclosure.
This enclosure should be installed in a convenient location and should be kept away from
direct sources of high heat, such as uninsulated steam lines.
B.Distribution Box: This 8x6 NEMA 4 enclosure is typically located near the probe that is
closest to the processor enclosure unit. It is connected to the main processor with an RS-485
bus DeviceNet™ cable, which can be hundreds of feet long. The probe and transmitter RS485 cable connections are also made within this box. The probe and transmitter cable
lengths should be 15 ft. or less, if possible.
C.Transmitter Box (optional): This 10x8 NEMA 4 enclosure contains terminals for accessing
the eight 4-20mA signals. It can be located in or near the control room so that the 4-20 wires
do not have to be run from the plant floor. An RS-485 DeviceNet cable runs from the
distribution box to the transmitter box. (There are restrictions on how this can be done, so
check manual Section 6, or contact the factory.)
- 5 -
Page 8
2.3MOUNTING HARDWARE INSTALLATION
ÎCheck hardware. Verify that the probe slides through the hot tap assembly.
ÏCheck installation configuration. Make sure the probe is parallel to the floor (see Figure 4).
Be sure location is accessible for probe removal and maintenance.
ÐCheck installation clearance. Verify there is a minimum probe insertion clearance of 4 feet
(1.3m) between pipe surface and any obstruction.
ÑInstall the mounting hardware. Drill a 1½” through-hole, center the thread-o-let over the
hole and weld it onto the condenser vacuum pipe (see Figure 4). Thread the hot-tap
assembly into the thread-o-let. Use thread tape or pipe dope to seal the connection.
(Alternate: weld thread-o-let to pipe wall, then drill a 1¼" hole in pipe wall using a hot-tap
drill.)
ÒIt may be necessary to apply a force of about 23 lb (102 Newtons) to remove or replace the
probe under plant operating conditions.
2.4PROBE INSTALLATION AND ORIENTATION
ÎCheck proper installation direction. Each probe has a directional
arrow on the junction box. Before installing the unit, note proper
flow direction. This is important to instrument operation.
ÏCheck serial number (S/N). If more than one RheoVac unit has been
purchased, make sure the first five digits of the serial numbers of the
probe(s) match the first five digits of the serial number of the main
processor unit. The electronics and up to three probes are a matched
set. Mismatched components will not work correctly. The dash
number on the probe S/N is the probe number shown on the display.
Record the probe number and installation location for future
reference.
ÐVerify stop clamp location (see Figure 3). A stop clamp is attached
to each probe as an indication of its insertion depth. It is important
that the stop clamp is securely in place to position the sensors in the
correct location and to ensure that the probes do not contact the
opposite pipe wall. Contact with the pipe wall could damage the
probe. The clamp’s location was determined based on your
submitted pipe diameter, as shown in SECTION 6.2, and is marked
with a groove on each probe shaft. Refer to this mark if a stop clamp
is inadvertently moved. The two metal probe tips should be in the
middle of the pipe.
Figure 3
Probe Stop Clamp
- 6 -
Page 9
Figure 4 Transducer Installation Detail
- 7 -
Page 10
ÑInspect the probe tips. Be sure wetted surfaces are clean before installing. If cleaning is
needed, use a damp cloth wetted with alcohol and wipe dry using a second dry, soft, lint-free
cloth. Do not immerse probes in alcohol or any other liquids. This may damage the probe
sensors.
ÒBe sure to power up your RheoVac instrument system and probe(s) for at least 30 minutes
before inserting probes into the vent line hot taps.DO NOT leave probe in vent line
unpowered or when flooding the condenser.
ÓInstall each probe. The probe should be mounted through the pipe wall using the hot-tap
assembly. The probe installs so that the two sensor tips are side-by-side across the gas stream.
The probes have a flow directional arrow on the junction box. When installing under vacuum,
do not allow the clamp to "slam" against the seal nut upon opening the valve. Grasp the probe
firmly, with hand against the seal nut, before opening the ball valve. Allow the probe shaft to
slide slowly through the valve by controlling the amount of grip on the probe shaft. Special
installation instructions, if any, will be noted in SECTION 6.
ÔPrior to inserting probe, loosen the compression nut on the thermocouple connector of the hot
tap and clean the inner surface of the thermocouple connector to ensure it is free of particles
that may cause probe damage.
2.5ELECTRICAL CONNECTIONS
. IMPORTANT . Inspect and VERIFY these electrical connections carefully. Improper
connection could damage electronic components and sensor function. If additional holes need to
be drilledin the processor enclosure, remove the electronicssubassembly (mounted ona mounting
plate) and temporarily store inside an ESD bag in a safe, clean place. Do not drill with electronics
boards inside the enclosure.
A.Main Processor Unit (see Figure 5)
1. Sensor Power and Communication Line
box using the RS-485 Communications/Power cable. Follow indicated connector color code.
[communications
2. Main Power
circuit. An external disconnect switch should be used for disconnecting power to the
: white (A), blue (B) and shield (SH); power: 24Vdc, red (+), and black (!)]
: Connect main power terminals to a dedicated 100-250Vac, single phase, 15 amp
: Connect the distribution box to this main processor
system during outages. Power connection wires should be at least 18 gauge and comply with
accepted wiring codes. SW1 is used for cycling power to reset.
3. Serial Communication
: Connector JP3 is the RS-232 serial communication interface. This
interface should only be used for distances of 20 feet or less, such as to a laptop computer. A
20 foot serial cable with a DB-9 connector can be shipped upon request (see Figure 9 and
Table I below).An RS-422 serial communication interface is present for long data
communications when configured without 4-20 mA outputs.
Table I RJ-11 to DB-9 Module Adapter
RS-232 CONFIGURATIONRS-422 CONFIGURATION
RJ-11 Pin OutDB-9 Pin OutRJ-11 Pin OutDB-9 Pin Out
4. Network Connection
Ethernet Cat 5 crossover cable is supplied with the unit.
B.Distribution Box (see Figure 6)
1. Connect the RS-485 wires (blue, white and shield) and 24Vdc power (red and black) from
the main processor unit.
2. Install ½” liquid-tight conduit between the distribution enclosure and the probes unless ½”
rigid conduit is used for long distance runs. Use a minimum of 6 feet of liquid tight at the
probes.
3. The probe connector comes with an adapter which allows the attachment of the ½” flexible
conduit connector.
4. Connect the probes to the distribution enclosure using the manufacturer supplied four
conductor shielded cable. Probe cable connections are shown in Figure 6.
5. Optional - Connect the transmitter box to the distribution box using a RS-485 cable.
: The Ethernet connection at JP1 is an RJ-45 style jack. A 10 foot
Figure 6 Distribution Box
- 10 -
Page 13
C.Probe - .. CAUTION - Do not cross thread connection. The probe is supplied with a
convenient plug-in connector. The male side of the connector comes installed in the probe
junction box. The female side is usually shipped loose and must be installed onto the RS485 cable once it is run from the Distribution Box to the probe. A separate instruction sheet
comes with the connector, showing the wiring connections.
D.Transmitter Box (optional) - For driving eight (8) remote 4-20 mA analog signals from one
RS-485 input port. (See Figure 8)
1. Connect the RS-485 communications/power cable from the distribution box (RS-485: white
and blue wires; 24Vdc power: red and black wires).
2. Connect up to eight (8) signal wire pairs to the indicated terminals for isolated 4-20mA
outputs.
3. Figure 8 provides the RheoVac wiring detail for
the 8 channel 4-20 mA outputs. Table II
provides the appropriate connection
identification. Model 950 transmitters are
configured as active (transmitter sources the
current) when shipped. To change to the passive
mode (receiver to source the current), extract
each small 4-20 board, find the JP1 pins, and
move the two jumpers from the "Act" pins to the
"Pass" pins (two positions to the right of factory
settings). Figure 7 shows the current output
TABLE II - Optional 4-20m A Configuration
Channel
1Actual volume flow
2Total mass flow
3Water vapor flow
4Pressure
5Water vapor/A ir mass ratio
6Relative saturation
7Air in-leak
8Temperature
Output Parameter
circuit. The figure also illustrates the active
mode and the passive mode configurations.
Figure 7 4-20 mA
Output Circuit
- 11 -
Page 14
Figure 8 Optional Transmitter Box
Figure 9 Serial Communications Interface
- 12 -
Page 15
E.Cable Connection from Probe to Distribution Box
The RheoVac probes require special connectors to be installed on the supplied DeviceNet
type 5711 cable sections that go from the Distribution Box(es) to the probes. The probes
have the male mating connectors already installed. These connectors will use either the
“backshell nut” or “conduit connector” (see Figure 10) depending on whether the cable is
installed in a liquid-tight conduit. When installing without conduit, use the backshell nut;
when using liquid-tight conduit, use the conduit connector with o-ring.
1. Slide all of the appropriate parts onto the cable as shown.
2. Strip the cable conductors as shown. The connector (Turck p/n B4151-0/9) has 5 retention
screws to hold the wires in place. The use of crimp pins on the wires will greatly increase
connection reliability.
3. Loosen all 5 retention screws (do not completely remove).
4. Insert the wires, in accordance with color-coding shown (see Figure 6, insert).
5. Tighten the retention screws on each wire.
6. Reassemble the connector parts (Figure 10).
Figure 10 Probe Connector Assembly
- 13 -
Page 16
SECTION 3 — OPERATION
3.1GENERAL INFORMATION
The RheoVac system is calibrated, compensated, and linearized for a wide range of flowing media
temperatures, pressures, and water vapor contents. However, abrupt changes in these parameters can
cause the instrument to temporarily read the flow rate improperly, which could lead to transient spikes
in the flow indication. In particular, if liquid (water) hits the probe tips, there will be high flow
indications until all the water vaporizes. If liquid (water) is present for extended periods, the
performance of the instrument will be compromised or the sensors can be damaged. See SECTION 2
INSTALLATION.
3.2SYSTEM START-UP
Verify wiring from SECTION 2.5 is correct. With all power off, place the power OFF/ON switch,
SW1, to “ON.” Close the main processor enclosure lid. While watching the main processor display,
apply power to the instrument. After about twenty seconds the display should initialize to “HealthCheck in Progress” followed by additional initialization status messages. Within two minutes the
display should indicate “Air In-Leak.”
Upon a "cold" initial start-up of the RheoVac system, the probe may take several minutes to give
accurate readings. During this time, the probe is transitioning to thermal equilibrium conditions under
vacuum.
3.3PORTABLE USB DATA STORAGE DEVICE/Warranty Registration Instructions
Intek’s warranty program is stated at the front of the manual. To validate warranty, return the portable
USB Data Storage Device after the RheoVac instrument has been in operation with the unit on-line for
one week.
To return the portable USB Data Storage Device to Intek:
ÎLocate portable USB Data Storage Device in the main electronics enclosure (see Figure 5).
ÏGrasp bracket and carefully remove portable USB Data Storage Device.
ÐPlace portable USB Data Storage Device in the pre-addressed, padded envelope provided
by Intek and mail.
The initial data stored in the device is used to verify proper operation of the RheoVac instrument and
is required to provide baseline data for future technical support.
.. CAUTION Do not attempt to use non-Intek approved USB flash drives with the RheoVac
instrument.
- 14 -
Page 17
3.4DISPLAY
The RheoVac 950 has a 2x20 alphanumeric scroll-through display that shows seven parameters for each
probe. The display shows Air In-leakage, W/A Mass Ratio, Total Mass Flow, Temperature, Pressure,
Relative Saturation, or Actual Volume Flow (as well as Time/Date, Software Version, or Serial
Number). To change the displayed parameter, press the Scroll Up or Down arrows. The display units
can be changed from English or Metric units by pressing the Units button.
If the Model 950 has two or three probes the display will show two or three serial numbers at a time,
one for each probe. The numbers will be in order, from the lowest serial number to the highest, reading
from left to right on the display. Example: 02611-1, 02611-2 and 02611-3.
3.5COMMUNICATIONS
Typical RheoVac 950 instruments have an Ethernet port, two serial communications choices (RS-232
or RS-422) and eight 4-20mA output signals for each probe. Each process variable is a linear, fully
temperature and pressure compensated value on any of these readable outputs. All 4-20 mA output
signals are scaled such that 4 mA represents 0% of the rated full scale value (except temperature, which
is 0°C) and 20 mA represents 100% of the rated full scale value (temperature is 100°C). The standard
full scale values and definitions of all process variables are listed in Table III.
Table III. Process Variable Definitions and 4-20 mA Range*
4-20 mA
PROCESS VARIABLE
FULL
SCALE
VALUE
ACTUAL VOLUME FLOW
ACFM [m
TOTAL MASS FLOW
lbs/hr [kg/hr]
WATER VAPOR MASS FLOW
lbs/hr [kg/hr]
RheoVac PRESSURE
" Hg [mm Hg]
WATER VAPOR SPECIFIC
VOLUME [cu. ft/lb]
WATER to AIR M AS S RATIO20Ratio of w ater va por flo w rate t o dry a ir flow r ate. D efines “ vacuum quality .”
RELATIVE SATURATION
PARTIAL PRESSURE, WATER
"Hg [mm Hg]
AIR IN-LEAK
SCFM [nm
RheoVac TEMPERATURE
[%]
°F [°C]
3
/hr]
3
/hr]
5000
[8,500]
10000
[4,536]
8000
[3,629]
30
[762]
N/AThe inv erse density of th e water v apor pres ent in the lin e. Not su pplied as 4-20m A output.
100The percent concentration of water vapor in the extraction line relative to saturation.
N/AThe pa rtial pressure of w ater vapo r in the vac uum line. Not su pplied as 4-20m A outpu t.
100
[169.9]
212
[100]
The actual volumetric flow rate of gases leaving the condenser. It is a measure of exhauster
capacity. Decreased capacity means pump degradation.
The total m ass flow rate of the flo wing gas. N ote: this va lue is n ot a me asure o f air in-le ak.
It is a measure of total vapor and air removal and exhauster operating capacity.
The water v apor comp onent of the flowing ga s being remo ved from the c ondenser.
Absolute pressure at the RheoVac probe head. Should be equal to or less than turbine back
pressure.
Actual meas ure of air volume flow rate passing the RheoVac sensor head, normalized to
standard conditions (70°F, 29.9" HgA).
4 mA = 0°C /32°F 20 m A = 100°C/21 2°F
Temperature of the flow med ia at the RheoVac pro be he ad.
* See Custom Information section (Section 6) for custom units of measure.
PROCESS VARIABLE DEFINITION
Communication with the RheoVac unit using Intek supplied software is discussed in Appendix A of this
manual.
- 15 -
Page 18
3.6DATA PROCESSING
TABL E IV. Column Headings for Data Downloads and Archived Data*
O
Time
Stamp
(Time of
day)
012345678910
Actual
Volume
Flow
(ACFM)
Total Mass
Flow
(lbs/hr)
Water
Vapor
Flow
(lbs/hr)
RheoVac
Pressure
("Hg abs)
Water
Vapor Sp.
Vol.
(Cu. ft/lb)
Water to
Air Mass
Ratio
(lb/lb)
Relative
Saturation
(%)
H
2
Partial
Pressure
("Hg abs)
Air Flow
In-Leak
(SCFM)
Probe
Temp. (°F)
* See Custom Information section (Section 6) for custom units of measure.
A very important part of understanding your condenser system comes from evaluating the data that is
created and stored by the RheoVac instrument. The ability to evaluate data has been greatly enhanced
with the RheoVac Model 950 by providing nearly unlimited data storage capacity. A data file is created
for each calender day that the monitor is operating. The ability to review stored data has been invaluable
when troubleshooting condenser problem s.
The importance of reviewing graphed data cannot be overstated. Reviewing data for normal operating
conditions will greatly decrease the time needed to troubleshoot problem conditions when they arise. To
simplify the graphing process, an Excel add-in has been created by the technical support team at Intek
and is included with your RheoVac user software. This Excel add-in graphing utility, called
RheoGraph, is an automated graphing macro that imports and graphs RheoVac data files.
The RheoGraph utility is contained in a self-extracting zip file called “InstallRheoGraph.exe” on the
User Interface software CD. Executing the file will begin the menu driven installation process. The
default installation path is to C:\RheoVac, do not change this path nam e during installation. The file
RheoGraph.xla will be copied to the installation folder along with help files, sample data, and a readme
text file. Microsoft Excel is required for the add-in to function.
Executing RheoGraph.xla will launch Excel and the graphing menu will appear. Selecting Auto will
prompt the user to enter Company, Plant, and Unit info, which will be placed in the graph titles. Then
the user is prompted to browse to the data file to be graphed. The data will be inserted into a spreadsheet
page and four graph pages will then be automatically added as separate tabs in the workbook. When the
graphs are complete, macro execution terminates.
Several advanced features are available by selecting Advanced on the Auto Start menu. For additional
help, use the keyboard arrow keys to set focus on any button, then press F1. The following is an example
of one of the graphs made by RheoGraph showing air in-leak and water vapor to air mass ratio.
- 16 -
Page 19
3.7CUSTOM SOFTWARE
Custom software may be developed by the user to receive and archive RheoVac data into a computer
system. The electronics has a serial data protocol of 9600 baud, no parity check, eight data bits and one
stop bit (i.e., 9600,N,8,1). Each transmitted group of data is sent in a standard ASCII coded format
representing each process variable value, instrument identification and status information.
The data stream consists of 13 fields, followed by a carriage return <RETURN>. The first ten fields,
nine bytes each, are the process variables. Following the process variables are the RheoVac serial
number, nine bytes, the process identification tag number, 15 bytes, and the RheoVac system status,
seven bytes. The data stream is then ended by a single <RETURN> byte (ASCII code 13). The total
number of bytes transmitted in each data stream is 122 bytes per probe including the trailing
<RETURN>. This data group is sent approximately once every three seconds. Table V shows the field
names and number of bytes in one data stream.
TABLE V. Serial Output Data Stream**
(°F)
Instr.
Serial
Numbe
9
bytes
ID
Status
Tag
No.
r
15
9
bytes
7
bytes
Term.
<CR>
1
byte
Actual
Volume
Flow
(ACFM)
9
bytes
Total
Mass
Flow
(lbs/hr)
9
bytes
Water
Vapor
Flow
(lbs/hr)
9
bytes
RheoVac
Pressure
("Hg abs)
9
bytes
Water
Vapor
Sp. V ol.
3
/16)
(ft
9
bytes
Water to
Air Mass
Ratio
(lb/lb)
9
bytes
Relative
Saturation
(%)
9
bytes
H2O
Partial
Pressure
("Hg abs)
9
bytes
Air Flow
** See Custom Information section (Section 6) for custom units of measure.
In-Leak
(SCFM)
9
bytes
Probe
Temp.
bytes
Each of the first ten process values are sent in the fixed decimal format of XXXX.XXXX with leading
and trailing zeros inserted to maintain the nine character length. The next three fields are ASCII text
strings followed by the <RETURN>. Example: The nine bytes for an air in-leak of 10.0 SCFM would
be: 0010.0000, or 48,48,49,48,46,48,48,48,48 ASCII.
- 17 -
Page 20
SECTION 4 — MAINTENANCE
4.1GENERAL MAINTENANCE
Precautions should be taken to insure proper performance of all sensors. Since the quantification
technique involves signal measurements, care should be exercised to prevent build-up of dirt and/or
corrosive layers on the various terminal strip connections. Periodic checks with necessary cleaning
should be performed to insure clean terminals. The joints of the sensor leads should occasionally be
inspected for corrosion or presence of moisture.
4.2CALIBRATION
The RheoVac instrument is calibrated at the factory in a calibration system which replicates the
condenser and vacuum line environment. The system is designed to simulate the temperature, pressure,
water vapor relative saturation and flow rate under the gaseous fluid conditions found within the power
plant vacuum line.
In general, calibrations should be valid over a two year period if the probe is well maintained. Should
the unit require re-ranging or recalibration, note the serial number of the RheoVac instrument and
contact the factory concerning recalibration cost and turn around times. Refer to SECTION 5
CUSTOMER SERVICE of this manual for additional information.
4.3SPARE PARTS
It is advantageous to have a spare probe. Should a spare probe be desired, it can be ordered and stored
for installation at any time. A spare probe will come with a parameter disk for uploading from a laptop.
Spare fuses should be available for replacement of blown fuses. Appropriate fuses to stock are:
500 mA fuse, slow blow, Wickmann part number 3720500041 or equivalent, for probes.
1 A fuse, slow blow, Wickmann part number 372110041 or equivalent, for the main electronics.
There are no other normally recommended spare components to stock.
To replace the fuse in the main electronics: locate the AC Power Input board (shown in Figure 5) in the
lower right hand corner of the main electronics box. Slide the power switch to the “OFF” position to
disable power to the electronics. The fuses are shown at locations F1 and F2 (spare). Gently pull the
fuse(s) from their sockets and gently insert the replacement 1 Amp fuse(s) as appropriate. Slide the
power switch back to the “ON” position to enable power to the electronics. Close hinged lid and verify
display backlight is on. Clamp close the enclosure lid, and tighten the latch screws.
4.4TROUBLESHOOTING
Table VI provides a guide for plant personnel to identify causes of problems and determine appropriate
actions to resolve problems observed. If problems are encountered and factory assistance is desired,
please contact the factory.
- 18 -
Page 21
MESSAGE CODE
Description/Symptom
TABLE VI. Troubleshooting Guide
PROBABLE CAU SE
ACTION
MSG-0
Communications not
being received from
probe(s)
MSG-1
Invalid data received
from probe(s)
MSG-2
Flow sensor heater
“OFF”
MSG-3
Relative saturation
sensor power “OFF”'
MSG-4
RS heater “FF”
MSG-5
Pressure sensor heater
“FF”
MSG-6
Temperature alarm
(above 210°F/99°C)
MSG-7
Wet probe
1. Improper cable hookup
2. Blown main fuse
3. Failed RS-485 circuit
4. Damaged flow sensor
1. Failed RS-485 communications
component
1. Blown heater fuse
2. Failed electronic component
1. Probe temperature too high
1. Com ponent failure1. Contact factory
1. Problem with pressure port
circuitry
1. Steam in exhaust pipe
1. This usually indicates a condenser
design or operation problem that
requires corrective action
2. Hogging mode during start-up
1. Verify plug-in connector properly mated
2. Check all cable connections
3. Check F1 fuse on probe board
4. Contact factory
1. Check wiring
2. Contact factory
1. Contact factory
1. Ch eck to be su re probe tem perature is
<160°F
2. Contact factory
1. Contact factory
1. Remove probe or cool line ASAP!
Once line h as cooled down and probe is
reinstalled, check unit for proper function
1. Contact factory
MSG-8
RS sensor problem
Mass flow output
saturates high, will not
respond to flow changes
Mass flow output
saturates low, will not
respond to flow changes
1. RS sensor has failed2. Contact factory
1. Flow rate is not within range of
calibration
2. Blown heater fuse
3. Failed electronic component
1. Flow rate is not within range of
calibration
2. Failed electronic component
1. Contact factory about re-ranging
instrument
2. Replace fuse
3. Contact factory
1. Contact factory about re-ranging
instrument
2. Contact factory
When using the RheoVac software on a PC, the software may give you a “Communication Error or
Message.” Use the Table VII to determine the source of this message and appropriate action.
Note: If the RheoVac instrument is operating without message indications but output readings are
questionable, please send to Intek by telefax or e-mail the following plant data
: turbine back pressure,
hot well temperature, load, and inlet and outlet circulation water temperatures, along with at least 24
hours of data from the RheoVac instrument/probe for the same time period.
- 19 -
Page 22
TABLE VII. Troubleshooting Guide (software communication errors)
OBSERVATIONPROBABLE CAU SEACTION
1. Instrument not connected to the software
Appears the first time the
application was executed
Appears intermittently
after application has
been running normally
Completely stops
working after application
was been running
normally
Cannot be made to work
at all with Com3 or
Com4
Note: Windows is a registered trademark of Microsoft Corporation in the United States and other countries.
defaulted serial port
2. Communication connections not made
or instrument is not powered
1. Electrical noise interfering with
communications
2. Too many applications running in
windows
3. Another application is conflicting with
this comm port or IRQ
1. Instrument has stopped communicating
2. Loose or damaged communication
connection
1. Works fine on Com1 or Com2 but does
not work on other port due to other
hardware conflicts such as a modem
1. Change serial comm port setting and hit
“Retry”
2. Check connections and instrument power
1. Change to RS-422 communications, reroute or shield cable
2. Close other applications until problem
self corrects
3. Change to a different comm port
1. Check instrument power or look at
instrument display for fault status
2. Check cable adapter at back of PC or at
any other splices or at instrument
1. Using Windows Control Panel - System
utility, check for IRQ or I/O hardware
conflict - ADVANCED USERS ONLY
4.5HARDWARE AND SOFTWARE MODIFICATIONS AND UPGRADES
Note: The workgroup setting of the identification tab in Network Neighborhood Properties must be set
to WORKGROUP which is the Windows 98 default setting. A Cat5 crossover network cable is required
for proper Ethernet network communication to the RheoVac instrument.
4.5.1UPDATING SOFTWARE
Normally, updating the main software residing on the RheoVac Model 950 should not need to be done
by the user. However, software improvements or other factors may necessitate a user installation or
upgrade of a new main program. When this is required, instructions will be supplied.
4.5.2REPLACING PROBES
If it is necessary to replace one of the probes, a new probe and diskette (with probe-specific calibration
data) can be ordered from Intek.
4.5.3ADDING ADDITIONAL PROBES (maximum total probes is 3)
Each RheoVac Model 950 can communicate with up to three probes simultaneously. Additional probes
can be easily added to systems that have fewer than three probes.
Each additional probe will be shipped with a set of calibration files that must be installed on the
RheoVac instrument. Refer to drawings for mechanical and electricalconnections. Specific instructions
will be supplied with the probe(s).
- 20 -
Page 23
SECTION 5 — CUSTOMER SERVICE
Intek's corporate philosophy is to help solve our customers’ difficult flow measurement problems.
When you purchase a RheoVac system you also receive Intek's outstanding customer service. For sales
or product service, call your local representative or Intek directly at (614) 895-0301 8AM to 5PM
EST/EDT weekdays or fax us anytime at (614) 895-0319. E-mail inquiries should be sent to
sales@Intekflow.com or techsupport@Intekflow.com. Our customer service staff will provide
assistance promptly.
5.1QUESTIONS ON EXISTING HARDWARE
To allow us to help you more quickly, please have the serial number of the equipment available before
you call.
5.2TROUBLESHOOTING
If you have reviewed SECTION 4.4 TROUBLESHOOTING and have questions, please call our
experienced engineers for assistance. In many cases we can solve a problem over the phone. Please
provide as complete a description as possible of the problems encountered.
5.3FACTORY AND FIELD SERVICE
If you request field service to help with condenser performance problems, Intek has, for a fee,
experienced engineers available to meet your needs. For RheoVac instrument related questions, if a
problem cannot be solved over the phone, with your help, we will determine if factory service or field
service will be the best solution. Be sure to carefully pack the instrument/electronics; extra care should
be taken to protect the probes from damage in shipment.
To request factory service on your instrument, a Return Material Authorization (RMA) and purchase
order is required. Our customer service staff will assist you with the required information to return
instruments for service.
5.4QUESTIONS ON NEW EQUIPMENT
For a new RheoVac system or any liquid or gas flow meter or flow switch need, contact the Intek
technical sales department at the above phone/fax numbers. Our staff will be pleased to answer all
questions and provide our recommended solutions or instruments.
5.5CONDENSER SERVICES
Intek has developed unique solutions for plants experiencing problems with dissolved oxygen, heat rate
and excessive back pressure. We provide comprehensive condenser diagnostic and report services.
Plant and RheoVac data is automatically transmitted to Intek and results are published via online
reporting and regular hard copy reports. For a quotation, please call us at the numbers listed above.
- 21 -
Page 24
SECTION 6 — CUSTOM INFORMATION
6.1UNIT IDENTIFICATION
Model no.:
Serial no.:
Customer identification:
6.2CONFIGURATION
The marked (X) items denote the configuration of this unit, as originally shipped from the factory.
Pipe Connection:: Hot tap with 1½" MNPT connection
Input Power:
Outputs:
Analog:9 8 outputs (4-20mA) for each probe
Digital:
: 100-250 Vac, 50/60 Hz
: Ethernet
9 RS-232/RS-422
Data Access:
Portable Data Storage:: Portable USB Data Storage Device
Enclosures shipped:Cables Shipped:
: Ethernet communication
: Processor Box:5711 interconnect cable (RS485) -
: Distribution Box:Cat5 Ethernet cable - 10 ft.
9 Transmitter Box9RS232 communication - 20 ft.
Software:Windows compatible RheoVac 950 software on CD - Version
Calibrated for customer line size of inches, schedule
6.3SPECIAL INSTRUCTIONS
- 22 -
Page 25
APPENDIX A - RheoVac® Instrument User Interface Software
A.1 INTRODUCTION
User Interface Software (UIS) is provided on CD-ROM with every RheoVac instrument. This
software is used to communicate with the RheoVac unit from a User Interface Computer (UIC).
Some of the functions of the software can be performed using a serial cable, while all functions are
available using an Ethernet connection. An Ethernet crossover cable is provided with each RheoVac
instrument for the purpose of connecting a UIC directly to the RheoVac unit. The network functions
are also capable of being used from any network computer if the RheoVac instrument is connected
directly to the same network. In this case, the supplied crossover cable is not used; rather a straight,
high noise-immune Ethernet cable with RJ-45 plug is used (supplied by customer).
Most details regarding use of the software can be found in the active help files from the pull down
Help menu when using the software.
A.2 SOFTWARE INSTALLATION
A User Interface Computer (UIC) is required to access stored data and real time data of all
parameters. A laptop is suggested for easy access to the main processor unit. Intek does not supply
UICs but recommends the following System Requirements (minimum recommended):
Minimum System Requirements:
Windows 98, Windows NT, Windows 2000 or Windows XP Operating Systems
Pentium III, 400 MHZ processor with 128 MB RAM
SVGA 800 x 600 display
One CD-ROM drive
One 10/100 Ethernet Port
Recommended (Optional):
One 3.5 inch floppy disk drive (probe recalibration data files can be provided on diskette)
One RS-232 serial port with DB-9 connector
The User Interface Software should be installed onto the User Interface Computer. Install the
software by inserting the CD-ROM supplied with the manual into the CD-ROM drive and follow
the on screen prompts. A folder, C:\RHEOVAC will be created and files will be copied to this
folder. The executable file is RheoVac User Interface.exe and the others are drivers and
configuration files that must remain in the RHEOVAC folder. If a previous version of User
nd
Interface Software is on the UIC, you will need to launch the installer a 2
time.
A.3 USER INTERFACE SOFTWARE (UIS) SELECTION
After installation, the user interface is initiated by selecting [START\PROGRAMS\RHEOVAC\
RHEOVAC USER INTERFACE]. The main menu appears, prompting for a selection of RS-232
Serial Communication or Ethernet Communication Utility. A list of functions for each utility
appears below the green selection buttons.
Make sure the proper cable connection has been made and select the desired communication type
by clicking on either green button.
Figure A - Initial UIS computer Screen
A.4 RS-232/422 COMMUNICATION
This function requires a serial cable and connector, available on request, to connect the UIC and the
RheoVac instrument electronics. This connection allows the user to view real time data of important
parameters in a convenient gauge screen layout. From this gauge screen, the user may choose to
graph a single parameter for all probes, graph all parameters for a single probe, or archive data
directly to the UIC. The archive rate can also be adjusted. These are the only functions available
using serial communication. No commands or updates can be sent to the RheoVac instrument using
this communication method.
A.5 ETHERNET COMMUNICATION
User interface using Ethernet connection allows the user to: synchronize the RheoVac clock with
the UIC clock, retrieve stored data
1
, update files1, change line size, and access data. These network
access utilities are performed from a user interface computer (UIC) connected directly (peer-to-peer)
to the RheoVac unit using a crossover network cable or from a network computer when the RheoVac
instrument is connected to a network (LAN/WAN)2 using a straight Ethernet cable.
Each RheoVac instrument is identified by a unique network ID. The ID begins with the letter “R”
followed by the first five digits of the serial number found on the unit. (RheoVac Model 950 devices
which have not been serviced by the factory since 2005 may use the serial numbers or “RheoVac
1
File transfer can also be accomplished using Windows Explorer or Network Neighborhood
950" network ID). The user must select the network ID from the pull down menu following the
selection of Ethernet Communication utility from the Main Menu. If the Network ID is not in the
list, select Add New ID and enter the correct six character Network ID. When a new ID is added
to the list, it will be saved as the default for later use. Selecting the correct ID will initiate
communication with the RheoVac instrument. Software features, in addition to those detailed above,
are described within the User Interface Software itself.
A.6 SOFTWARE OPERATION
1.Set Clock
(available with Ethernet connection only): Initiating this action will reset the
RheoVac time clock to agree with the time on your UIC. Make sure the UIC has the
correct time before proceeding.
2.Change Line Size (available with Ethernet connection only): Use this option to update
the instrument’s stored line size variable for proper volume and mass flow calculations.
You should only need this option if the probe is installed in a pipe size that is different
from the value set in the instrument at the factory. In Ethernet Communication, click on
the Line Size tab. On the pull down menu for pipe size, click on the new value, then click
on Save. Click Load to verify that the change was saved by the instrument.
3.Update Files
recalibration, or some other change in the RheoVac operating software is needed, this tab
will be used. Do not use it until then. Instructions will be supplied at that time.
4.Retrieve Data
accessed via Ethernet connection from the
The
RheoVac software creates a new data file each day. The file name is in the
(available with Ethernet connection only): If the probes are ever returned for
(available with Ethernet connection only): Stored data can be easily
RheoVac instrument to the UIC.
following format:
MMDDYY - P.dat
where MM = two digit month
DD = two digit day
YY = two digit year
P = probe designator (1, 2, or 3)
.dat = file extension
These data files are stored in tab delimited text format. See SECTION 3.5 to process data.
To copy a file for analysis, you must access the files on the RheoVac instrument. Open
Windows Explorer on the UIC. Type \\Network ID\RHEOVAC\DATA (Network ID is
generally “R” plus the first 5 digits of the serial number; see section A.5) on the address
line and press enter. Be careful not to alter any support files in this data folder. Use
the copy/paste functions (copy - Ctrl C; paste - Ctrl V) to put data files (.dat) on the
UIC.
5.Monitor Data
(available with Ethernet or RS-232 connection): Once RS-232 or Ethernet
is selected, the first screen, Figure B, shows the most recent value for each parameter.
These four parts work together in the following ways: As soon as the software is opened and valid
communication is established, the data collected for each probe is plotted on the chart(s). The
newest data points are displayed on the right of the chart and push older data points to the left, much
like many old-style strip chart recorders.
The horizontal scroll bar immediately below the chart(s) can be used to select which data points (i.e.,
data from a specific timestamp) from the chart are displayed on the legend below. Data points can
be selected either by clicking-and-dragging the scroll bar pointer or by simply single-clicking a spot
on the scroll bar.
The four buttons below and to the left of the scroll bar labeled 30 min, 1 hour, 2 hour, and 4 hour
can be used to quickly set the x-axis range of the chart(s) and the scroll bar. When using these
buttons keep in mind that the maximum x-axis value is always used as a reference and only the
minimum displayed x-axis value will be changed. You can also change both the minimum and
maximum x-axis value by double-clicking them with your mouse and entering a new value with
your keyboard. Changing the range of the chart (X or Y axis) changes only the data that is displayed
and does not delete or alter the data in any way.
The blue block (time/date) to the right of the 4 hour button displays the timestamp for the data point
currently being pointed to by the horizontal scroll bar. If the scroll bar is in the right-most position,
this value will automatically update when new data is received from the connected instrument.
The legend at the bottom of the screen displays the value of the data being pointed to by the scroll
bar and also is used to control where each data parameter is plotted. To access this feature singleclick on any parameter in the legend. Two things will happen. First, the background color of the
selected parameter will change in the legend, and second, the parameter will disappear from the
chart if you are in single-chart mode or jump from one chart to the other if you are in dual-chart
mode.
Single-chart and dual-chart modes are the first two selections available in the chart user menu. Also
available in the chart menu are selections to choose between manual and auto scaling for the y-axis
and to reset the maximum x-axis value to the most recently collected data point.
The legend menu is used to select which parameter or which probe is displayed. If a parameter is
chosen from this menu, that parameter will be displayed for all probes connected to the instrument.
If a probe is selected, then all parameters for that probe will be displayed.
The archive menu is used to set the rate that data will be archived to the hard disk of the computer
on which the user has installed the User Interface Software. These selections have no effect on the
data that is stored every minute on the hard disk of the RheoVac instrument itself. Data archived
to the user’s hard disk is stored in the “C:\RHEOVAC\DATA\” directory. The data is subdivided
into folders for each probe. A new file for each probe will be created each day that data is archived.
A.7 WARRANTY REGISTRATION/TECHNICAL SUPPORT
Sending Data to Intek: IMPORTANT
If the instrument is equipped with a portable USB data storage device as shown in Figure 5 (in the
main manual), remove this storage device after one week of operation and send to "Intek, Inc., 751
Intek Way, Westerville, OH 43082" to activate/secure warranty registration. This storage device
is programmed to capture the initial baseline data recorded by the RheoVac instrument.
The following paragraph is applicable to RheoVac 950 instrument users only:
The availability of plant data for periods corresponding to the RheoVac data are important to proper
diagnostic evaluation. Plant data of interest to the factory for the same time period are: turbine back
pressure, hotwell temperature, load, and inlet and outlet circulation water temperatures. The plant
data is necessary for correlating RheoVac measured data. The availability of this information at the
factory is very important to Intek’s ability to assist you in troubleshooting your instrument and/or
to respond to inquiries. This provides a baseline for the system and will help us support you should
future system problems arise. Our experienced support engineers can access the “before” and “after”
effects of an upset and work with you to identify and understand the problem, and develop solutions.
Additionally, most air removal piping conditions that are detrimental to probe operation can be
diagnosed before probe damage occurs. For this reason, we require a data download one week after
RheoVac system is placed into service. The download activates the one year warranty service.
the
Alternately, you may send to the factory an initial week or so of data (zipped format is preferred). The
simplest way is to copy all data files and the ‘event.log’ file to your UIC and send it via email to
techsupport@intekflow.com.
Note: Windows is a registered trademark of Microsoft Corporation in the United States and other countries.
Use these instructions when making a peer-to-peer Ethernet connection between the RheoVac 950 and
the User Interface Computer (UIC), or an Ethernet connection between the RheoVac 950 and a Local
Area Network (LAN). Locate the section that covers your operating software.
B.2WINDOWS 98 OR WINDOWS ME OPERATING SYSTEM
1. Ethernet Cable
:
a. UIC Connection: use CAT5 network crossover cable (supplied by Intek) to connect the
Ethernet port of the UIC to the Ethernet port on the RheoVac 950.
b. LAN Connection: use CAT5 straight high noise-immune Ethernet cable (not supplied)
with RJ45 plug to connect the Ethernet port of the LAN to the Ethernet port of the
RheoVac 950.
2. Workgroup Identification
: Unless otherwise requested, the Workgroup identification for the
RheoVac 950 will be WORKGROUP, which is the Windows 98 default setting. To check/set
the workgroup name: Right click on the Network Neighborhood icon on the UIC desktop.
Select Properties from the menu that appears. Select the Identification tab of the Network
dialog box that appears. Verify that Workgroup is WORKGROUP.
3. Check for Installed Network Components
: Right click on the Network Neighborhood icon
on the UIC or LAN computer desktop. Select Properties from the menu that appears. Select
the Configuration tab of the Network dialog box that appears. In the list under “The
following network components are installed” there should be all three of the following: Client
for Microsoft Networks, Ethernet adapter, and NetBEUI (with the Ethernet adapter identified
on the same line next to it). Each will have a different icon associated with it; the ‘Client’
icon looks like a computer screen, the ‘Adapter’ icon looks like an adapter card, and the
‘NetBEUI’ protocol icon looks like a cable in a “T” shape. If these components are not
installed, one or more of the following installations may be necessary. Note: The Windows
installation disk may be needed.
a. Installing Client for Microsoft Networks: If Client for Microsoft Networks is not in the
list of “The following network components are installed,” then do the following:
i.Click the Add button next to the list.
ii.The Select Network Component Type dialog box will appear.
iii.Select Client from the list then click the Add button.
iv.Select Microsoft from the Manufacturers list.
v.Select Client for Microsoft Networks in the Network Clients list.
b. Installing NetBEUI: If the NetBEUI (with the Ethernet adapter identified on the same line
next to it) is not in the list of “The following network components are installed,” then do
the following:
i.Search the installed network components list for the installed Ethernet adapter.
ii.Select the Ethernet adapter and click the Add button below the list.
iii.Select Protocol from the Select Network Component Type dialog box.
iv.Click the Add button next to the list.
v.The Select Network Protocol dialog box will appear.
vi.Select Microsoft from the Manufacturers list.
vii.Select NetBEUI from the Network Protocols list.
4. Networking with the RheoVac 950: If all of the items above are completed successfully, the
UIC or LAN computer should be able to network with the RheoVac 950. To search for the
RheoVac 950 from the UIC or LAN computer:
Method 1
a. Open Network Neighborhood.
b. Each RheoVac instrument is identified by a unique Network ID. The ID is “R” plus the
first five digits of the serial number found on the unit. (RheoVac Model 950 devices
which have not been serviced by the factory since August, 2003 may use the serial
numbers or “RheoVac 950” as the Network ID).
c. Look for the Network ID of your RheoVac 950. If the expected Network ID (see previous
paragraph) does not show up in the list, then type in the appropriate Address line (either
“\\RHEOVAC 950” or “\\R” and the first five digits of the serial number) and click the Go
button. This may take several seconds and then a RHEOVAC folder will appear, which
is the hard drive of the RheoVac 950.
d. Open the RHEOVAC folder then open the DATA folder. The \\Network ID
\RHEOVAC\DATA folder contains all the stored data. Use copy/paste functions (copy -
Ctrl C; paste - Ctrl V) to save the data files to the UIC or LAN computer.
Method 2
a. Open My Computer
b. Type in the Address line: \\Network ID (Network ID is defined above) and click the Go
button. This may take several seconds and then a RHEOVAC folder will appear, which
is the hard drive of the RheoVac 950.
c. Open the RHEOVAC folder then open the DATA folder. The \\Network ID\
RHEOVAC\DATA folder contains all the stored data. Use copy/paste functions to save
the data files to the UIC or LAN computer.
B.3WINDOWS NT OR WINDOWS 2000 OPERATING SYSTEM
1. Ethernet Cable
:
a. UIC Connection: use CAT5 network crossover cable (supplied by Intek) to connect the
Ethernet port of the UIC to the Ethernet port on the RheoVac 950.
b. LAN Connection: use CAT5 straight high noise-immune Ethernet cable (not supplied)
with RJ45 plug to connect the Ethernet port of the LAN to the Ethernet port of the
RheoVac 950.
2. Workgroup Identification
: Unless otherwise requested, the Workgroup identification for the
RheoVac 950 will be WORKGROUP. To check/set the workgroup name: Right click on the
My Computer icon on the UIC or LAN computer desktop. Select Properties from the menu
that appears. Select the Network Identification tab of the System Properties dialog box that
appears. Verify that Workgroup is WORKGROUP.
3. Check for Installed Network Components
: Click the Start button, select Settings, select
Network and Dial-up Connections. Or, right click My Network Places and select
Properties. Highlight the Local Area Connection that represents the Ethernet adapter. Right
click on Local Area Connection and select Properties. Under “Components checked are
used by this connection:” look for Client for Microsoft Networks and NetBEUI Protocol.
If these components are not installed, one or more of the following installations may be
necessary. Note: The Windows installation disk may be needed as well as administrator
security rights.
a. Installing Client for Microsoft Networks: If Client for Microsoft Networks is not in the
list of “The following network components are installed,” then do the following:
i.Click the Install button below the list.
ii.The Select Network Component Type dialog box will appear.
iii.Select Client from the list then click the Add button.
iv.Select Microsoft from the Manufacturers list.
v.Select Client for Microsoft Networks in the Network Client list.
vi.Click OK.
b. Installing NetBEUI: If the NetBEUI (with the Ethernet adapter identified on the same line
next to it) is not in the list of “The following network components are installed,” then do
the following:
i.Click the Install button below the list.
ii.The ‘Select Network Component Type’ dialog box will appear.
iii.Select Protocol and click the Add button.
iv.The ‘Select Network Protocol’ dialog box will appear.
v.Select NetBEUI from the Network Protocol list.
vi.Click OK.
4. Networking with the RheoVac 950
UIC or LAN computer should be able to network with the RheoVac 950. To search for the
RheoVac 950 from the UIC or LAN computer:
a. Open My Computer
b. Each RheoVac instrument is identified by a unique Network ID. The ID is “R” plus the
first five digits of the serial number found on the unit. (RheoVac Model 950 devices
purchased prior to August, 2003 may use the serial numbers or “RheoVac 950” as the
Network ID.)
c. Look for the Network ID of your RheoVac 950. If the expected Network ID (see previous
paragraph) does not show up in the list, then type in the appropriate Address line (either
‘\\RHEOVAC 950’ or ‘\\’ and the first five digits of the serial number) and click the Go
button. This may take several seconds and then a RHEOVAC folder will appear, which
is the hard drive of the RHEOVAC 950.
d. Open the RHEOVAC folder and then open the DATA folder. The \\Network ID\
RHEOVAC\DATA folder contains all the stored data. Use copy/paste functions (copy Ctrl C; paste - Ctrl V) to save the data files to the UIC or LAN computer.
B.4WINDOWS
1. Ethernet Cable
XP OPERATING SYSTEM
:
a. UIC Connection: use CAT5 network crossover cable (supplied by Intek) to connect the
Ethernet port of the UIC to the Ethernet port on the RheoVac 950.
b. LAN Connection: use CAT5 straight high noise-immune Ethernet cable (not supplied)
with RJ45 plug to connect the Ethernet port of the LAN to the Ethernet port of the
RheoVac 950.
2. Workgroup Identification
RheoVac 950 will be WORKGROUP. To check/set the workgroup name: Click the Start
button and select Control Panel. Right click on the System icon and select Open. Select
the Computer Name tab. Verify that Workgroup is WORKGROUP. Use the Change button
to type in a new Workgroup if needed.
: If all of the items above are completed successfully, the
: Unless otherwise requested, the Workgroup identification for the
3. Check for Installed Network Components: Click the Start button and select Control Panel.
Right click on the Network Connections icon and select Open. Or, click the Start button,
select Connect To, then select Show all connections. Highlight the Local Area Connection
that represents the Ethernet adapter. Right click the icon and select Properties. In the new
window that pops up, under “This connection uses the following items:,” look for Client for
Microsoft Networks, NWLink NetBIOS, and NWLink IPX/SPX/NetBios protocols. If these
components are not installed, one or more of the following installations may be necessary.
Note: The Windows installation disk may be needed as well as administrator security rights.
a. Installing Client for Microsoft Networks: If Client for Microsoft Networks is not in the
list of “This connection uses the following components:,” then do the following:
i.Click the Install button below the list.
ii.The Select Network Component Type dialog box will appear.
iii.Select Client from the list then click the Add button.
iv.Select Microsoft from the Manufacturers list.
v.Select Client for Microsoft Networks in the Network Client list.
vi.Click OK.
b. Installing NWLink NetBIOS: If NWLink NetBIOS is not in the list of “This connection
uses the following components:,” then do the following:
i.Click the Install button below the list.
ii.The Select Network Component Type dialog box will appear.
iii.Select Protocol and click the Add button.
iv.The Select Network Protocol dialog box will appear.
v.Select NWLink NetBIOS from the Network Protocol list.
vi.Click OK.
c. Installing NWLink IPX/SPX/NetBIOS: If NWLink IPX/SPX/NetBIOS is not in the list
of “This connection uses the following components:,” then do the following:
i.Click the Install button below the list.
ii.The Select Network Component Type dialog box will appear.
iii.Select Protocol and click the Add button.
iv.The Select Network Protocol dialog box will appear.
v.Select NWLink IPX/SPX/NetBIOS from the Network Protocol list.
vi.Click OK.
4. Networking with the RheoVac 950
: If all of the items above are completed successfully, the
UIC or LAN computer should be able to network with the RheoVac 950. To search for the
RheoVac 950 from the UIC or LAN computer:
a. Open My Computer.
b. Each RheoVac instrument is identified by a unique Network ID. The ID is “R” plus the
first five digits of the serial number found on the unit. (RheoVac Model 950 devices
purchased prior to August, 2003 will use the Network ID “RheoVac 950.”)
c. Look for the Network ID of your RheoVac 950. If the expected Network ID (see previous
paragraph) does not show up in the list, then type in the appropriate Address line (either
‘\\RHEOVAC 950’ or ‘\\’ and the first five digits of the serial number) and click the Go
button. This may take several seconds and then a RHEOVAC folder will appear, which
is the hard drive of the RheoVac 950.
d. Open the RHEOVAC folder and then open the DATA folder. The \\Network ID\
RHEOVAC\DATA folder contains all the stored data. Use copy/paste functions (copy Ctrl C; paste - Ctrl V) to save the data files to the UIC or LAN computer.