Instek GDS-2072E, GDS-2102E, GDS-2202E, GDS-2074E, GDS-2104E User's Manual

...
Digital Storage Oscilloscope
GDS-2000E Series
USER MANUAL
GW INSTEK PART NO. 82DS-2KE00E01
ISO-9001 CERTIFIED MANUFACTURER
This manual contains proprietary information, which is protected by copyright. All rights are reserved. No part of this manual may be photocopied, reproduced or translated to another language without prior written consent of Good Will company.
The information in this manual was correct at the time of printing. However, Good Will continues to improve products and reserves the rights to change specification, equipment, and maintenance procedures at any time without notice.
Good Will Instrument Co., Ltd. No. 7-1, Jhongsing Rd., Tucheng Dist., New Taipei City 236, Taiwan
TABLE OF CONTENTS
Table of Contents
SAFETY INSTRUCTIONS ................................................... 5
GETTING STARTED ......................................................... 10
GDS-2000E Series Overview ................. 11
Appearance .......................................... 15
Set Up .................................................. 27
Built-in Help ........................................ 36
MEASUREMENT .............................................................. 37
Basic Measurement ............................. 38
Automatic Measurement ...................... 45
Cursor Measurement ........................... 59
Math Operation ................................... 67
CONFIGURATION ........................................................... 77
Acquisition .......................................... 80
Segmented Memory Acquisition Overview
............................................................ 85
Display................................................. 98
Horizontal View ................................. 104
Vertical View (Channel) ...................... 113
Bus Key Configuration ........................ 121
Trigger ............................................... 144
Search ................................................ 176
System Settings and Miscellaneous
Settings ............................................. 185
APPS .............................................................................. 190
Applications ....................................... 191
SAVE/RECALL ................................................................ 213
3
GDS-2000E Series User Manual
File Format/Utility ............................. 214
Create/Edit Labels .............................. 219
Save ................................................... 222
Recall ................................................. 230
Reference Waveforms ......................... 236
FILE UTILITIES ............................................................... 238
HARDCOPY KEY ............................................................. 245
REMOTE CONTROL CONFIG ......................................... 249
Interface Configuration ...................... 250
MAINTENANCE ............................................................. 261
FAQ ............................................................................... 267
APPENDIX ..................................................................... 270
GDS-2000E Specifications .................. 270
Probe Specifications .......................... 274
GDS-2000E Dimensions ..................... 276
Declaration of Conformity .................. 277
INDEX ............................................................................ 278
4
SAFETY INSTRUCTIONS
WARNING
Warning: Identifies conditions or practices that could result in injury or loss of life.
CAUTION
Caution: Identifies conditions or practices that could result in damage to the GDS-2000E or to other properties.
DANGER High Voltage
Attention Refer to the Manual
Protective Conductor Terminal
Earth (ground) Terminal
SAFETY INSTRUCTIONS
This chapter contains important safety instructions that you must follow during operation and storage. Read the following before any operation to insure your safety and to keep the instrument in the best possible condition.
Safety Symbols
These safety symbols may appear in this manual or on the GDS­2000E.
5
GDS-2000E Series User Manual
Do not dispose electronic equipment as unsorted municipal waste. Please use a separate collection facility or contact the supplier from which this instrument was purchased.
General Guideline
CAUTION
Make sure the BNC input voltage does not
exceed 300Vpk.
Never connect a hazardous live voltage to the
ground side of the BNC connectors. It might lead to fire and electric shock.
Do not place any heavy object on the GDS-
2000E.
Avoid severe impact or rough handling that
leads to damaging the GDS-2000E.
Do not discharge static electricity to the GDS-
2000E.
Use only mating connectors, not bare wires, for
the terminals.
Do not block the cooling fan opening. Do not perform measurement at a power source
or building installation site (Note below).
Do not disassemble the GDS-2000E unless you
are qualified.
(Measurement categories) EN 61010-1:2010 specifies the measurement categories and their requirements as follows. The GDS-2000E falls under category I.
Measurement category IV is for measurement performed at the
source of low-voltage installation.
Measurement category III is for measurement performed in the
building installation.
Measurement category II is for measurement performed on the
circuits directly connected to the low voltage installation.
Measurement category I is for measurements performed on
circuits not directly connected to Mains.
Safety Guidelines
6
SAFETY INSTRUCTIONS
Power Supply
WARNING
AC Input voltage: 100 - 240V AC, 50 - 60Hz,
auto selection. Power consumption: 30 Watts.
Connect the protective grounding conductor of
the AC power cord to an earth ground, to avoid electrical shock.
Cleaning the GDS-2000E
Disconnect the power cord before cleaning. Use a soft cloth dampened in a solution of mild
detergent and water. Do not spray any liquid.
Do not use chemicals containing harsh materials
such as benzene, toluene, xylene, and acetone.
Operation Environment
Location: Indoor, no direct sunlight, dust free,
almost non-conductive pollution (Note below)
Relative Humidity: ≤80%, 40°C or below; 45%,
41°C ~ 50°C
Altitude: < 2000m Temperature: 0°C to 50°C
(Pollution Degree) EN 61010-1:2001 specifies the pollution degrees and their requirements as follows. The GDS-2000E falls under degree 2.
Pollution refers to “addition of foreign matter, solid, liquid, or
gaseous (ionized gases), that may produce a reduction of dielectric strength or surface resistivity”.
Pollution degree 1: No pollution or only dry, non-conductive
pollution occurs. The pollution has no influence.
Pollution degree 2: Normally only non-conductive pollution
occurs. Occasionally, however, a temporary conductivity caused by condensation must be expected.
Pollution degree 3: Conductive pollution occurs, or dry, non-
conductive pollution occurs which becomes conductive due to condensation which is expected. In such conditions, equipment is normally protected against exposure to direct sunlight, precipitation, and full wind pressure, but neither temperature nor humidity is controlled.
Storage environment
Location: Indoor Temperature: -10°C to 60°C
40°C /93% RH 41°C ~60°C /65% RH
7
GDS-2000E Series User Manual
Disposal
Do not dispose this instrument as unsorted municipal waste. Please use a separate collection facility or contact the supplier from which this instrument was purchased. Please make sure discarded electrical waste is properly recycled to reduce environmental impact.
8
SAFETY INSTRUCTIONS
Green/ Yellow:
Earth
Blue:
Neutral
Brown:
Live (Phase)
Power cord for the United Kingdom
When using the oscilloscope in the United Kingdom, make sure the power cord meets the following safety instructions.
NOTE: This lead/appliance must only be wired by competent persons
WARNING: THIS APPLIANCE MUST BE EARTHED IMPORTANT: The wires in this lead are coloured in accordance with the following code:
As the colours of the wires in main leads may not correspond with the coloured marking identified in your plug/appliance, proceed as follows:
The wire which is coloured Green & Yellow must be connected to the Earth terminal marked with either the letter E, the earth symbol
or coloured Green/Green & Yellow.
The wire which is coloured Blue must be connected to the terminal which is marked with the letter N or coloured Blue or Black.
The wire which is coloured Brown must be connected to the terminal marked with the letter L or P or coloured Brown or Red.
If in doubt, consult the instructions provided with the equipment or contact the supplier.
This cable/appliance should be protected by a suitably rated and approved HBC mains fuse: refer to the rating information on the equipment and/or user instructions for details. As a guide, a cable of 0.75mm2 should be protected by a 3A or 5A fuse. Larger conductors would normally require 13A types, depending on the connection method used.
Any exposed wiring from a cable, plug or connection that is engaged in a live socket is extremely hazardous. If a cable or plug is deemed hazardous, turn off the mains power and remove the cable, any fuses and fuse assemblies. All hazardous wiring must be immediately destroyed and replaced in accordance to the above standard.
9
GDS-2000E Series User Manual
GDS-2000E Series Overview ............................................ 11
Series lineup .......................................................................... 11
Main Features ....................................................................... 12
Accessories ............................................................................ 13
Appearance ..................................................................... 15
GDS-2074E/2104E/2204E Front Panel ......................... 15
GDS-2072E/2102E/2202E Front Panel ......................... 16
Rear Panel ............................................................................. 23
Display ................................................................................... 25
Set Up ............................................................................. 27
Tilt Stand ............................................................................... 27
Power Up .............................................................................. 28
First Time Use ...................................................................... 29
How to Use This Manual .................................................... 31
GETTING STARTED
This chapter describes the GDS-2000E in a nutshell, including its main features and front / rear panel introduction. After going through the overview, follow the Set Up section to properly set up the oscilloscope for first time use. The Set Up section also includes a starter on how to use this manual effectively.
10
GETTING STARTED
Model name
Frequency bandwidth
Input channels
Max. Real-time Sampling Rate
GDS-2072E
70MHz
2
1GSa/s
GDS-2102E
100MHz
2
1GSa/s
GDS-2202E
200MHz
2
1GSa/s
GDS-2074E
70MHz
4
1GSa/s
GDS-2104E
100MHz
4
1GSa/s
GDS-2204E
200MHz
4
1GSa/s
GDS-2000E Series Overview
Series lineup
The GDS-2000E series consists of 6 models, divided into 2-channel and 4-channel versions.
11
Main Features
Features
8 inch, 800 x 480, WVGA TFT display. Available from 70MHz to 200MHz. Real-time sampling rate of 1GSa/s (2 channel
models), 1GSa/s max. (4 channel models).
Deep memory: 10M points record length. Waveform capture rate of 120,000 waveforms
per second.
Vertical sensitivity: 1mV/div~10V/div. Segmented Memory: Optimizes the acquisition
memory to selectively capture only the important signal details. Up to 29000 successive waveform segments can be captured with a time-tag resolution of 4ns.
Waveform Search: Allows the scope to search
for a number of different signal events.
On-screen Help. 32 MB internal flash disk.
Interface
USB host port: front panel, for storage devices. USB device port: rear panel, for remote control
or printing.
Probe calibration output with selectable output
frequency (1kHz ~ 200kHz).
Ethernet port as standard. Calibration output.
GDS-2000E Series User Manual
12
Accessories
Standard Accessories
Part number
Description N/A
Quick Start Guide
N/A region dependent
Power cord
GTP-070A-4, for GDS-2072E/GDS-2074E
Passive probe; 70 MHz
GTP-150A-4, for GDS-2102E/GDS-2104E
Passive probe; 150 MHz
GTP-300A-4, for GDS-2202E/GDS-2204E
Passive probe; 300 MHz
Standard Apps
Name
Description
Go-NoGo
Go-NoGo testing app.
DataLog
Waveform or image data logging app.
DVM
Digital Volt Meter app.
Digital Filter
High or low pass digital filter for analog inputs.
Remote Disk
Allows the scope to mount a network share drive.
Demo Mode
Demonstration mode that is used with the GDB-03 demo board.
Optional Accessories
Part number
Description
GTC-001
Instrument cart, 470(W)x430(D)mm (U.S. type input socket)
GTC-002
Instrument cart, 330(W)x430(D)mm (U.S. type input socket)
GDB-03
Demo board
GTL-110
test lead, BNC to BNC heads
GETTING STARTED
13
GDS-2000E Series User Manual
GTL-242
USB cable, USB2.0A-B type cable 4P
GTP-070A-4
Passive probe; 70 MHz
GTP-150A-4
Passive probe; 150 MHz
GTP-300A-4
Passive probe; 300 MHz
Drivers
USB driver
LabVIEW driver
14
GETTING STARTED
VARIABLE
POSITION
HORIZONTAL
POSITION
POSITION
POSITION
POSITION
VERTICAL
MATH
REF
BUS
SCALE
TRIGGER
LEVEL
PUSH TO
ZERO
SCALESCALESCALE
PUSH TO
ZERO
PUSH TO
ZERO
PUSH TO
ZERO
PUSH TO
ZERO
PUSH TO ZERO
SCALE
POWER
CH1 CH2 CH3 CH4
2V
CAT I
1MW16pF 300Vpk MAX.
Digital Storage Oscilloscope
GDS-2204E
200 MHz 1 GS/s
Visual Persistence Oscilloscope
LCD
Variable knob and Select key
Autoset, Run/Stop, Single and Default settings
CH1~CH4
Trigger controls
Function keys
USB Host port, Probe Compensation terminals
Power button
Hardcopy key
Option key
Math, Reference and Bus keys
Bottom menu keys
Horizontal controls
Menu key
Vertical controls
Side menu keys
Appearance
GDS-2074E/2104E/2204E Front Panel
15
GDS-2000E Series User Manual
VARIABLE
POSITION
HORIZONTAL
TRIGGER
LEVEL
PUSH TO
ZERO
PUSH TO
ZERO
SCALE
Measure
Display
Cursor
Help
Test
Save/Recall
Acquire
Utility
Autoset
Run/Stop
Single
Default
Select
Search
Set/Clear
Zoom
Menu
50 %
Force-Trig
POSITION
POSITION
VERTICAL
MATH
REF
BUS
SCALESCALE
PUSH TO
ZERO
PUSH TO
ZERO
M
R
B
Digital Storage Oscilloscope
GDS-2202E
200 MHz 1 GS/s
Visual Persistence Oscilloscope
POWER
CH1 CH2 EXT TRIG
2V
CAT I
1MW16pF 300Vpk MAX.
CAT I
1MW16pF 300Vpk MAX.
LCD
Variable knob and Select key
Autoset, Run/Stop, Single and Default settings
CH1~CH2, EXT TRIG
Trigger controls
Function keys
USB Host port, Probe Compensation terminals
Power button
Hardcopy key
Option key
Math, Reference and Bus keys
Bottom menu keys
Horizontal controls
Menu key
Vertical controls
Side menu keys
LCD Display
8” WVGA TFT color LCD. 800 x 480 resolution, wide angle view display.
Menu Off Key
Use the Menu Off key to hide the onscreen menu system.
Option Key
The Option key is used to access future installed options.
GDS-2072E/2102E/2202E Front Panel
16
GETTING STARTED
Menu Keys
The side menu and bottom menu keys are used to make selections from the soft-menus on the LCD user interface.
To choose menu items, use the 7 Bottom menu keys located on the bottom of the display panel.
To select a variable or option from a menu, use the side menu keys on the side of the panel. See page 31 for details.
VARIABLE
POSITION
HORIZONTAL
TRIGGER
LEVEL
PUSH TO
ZERO
PUSH TO
ZERO
SCALE
Measure
Display
Cursor
Help
Test
Save/Recall
Acquire
Utility
Autoset
Run/Stop
Single
Default
Select
Search
Set/Clear
Zoom
Menu
50 %
Force-Trig
POSITION
POSITION
VERTICAL
MATH
REF
BUS
SCALESCALE
PUSH TO
ZERO
PUSH TO
ZERO
M
R
B
Digital Storage Oscilloscope
GDS-2202E
200 MHz 1 GS/s
Visual Persistence Oscilloscope
Side menu keys
Bottom menu keys
Hardcopy Key
The Hardcopy key is a quick-save or quick-print key, depending on its configuration. For more information see pages 247(save) or 246(print).
Variable Knob and Select Key
Select
VARIABLE
The Variable knob is used to increase/decrease values or to move between parameters.
The Select key is used to make selections.
17
GDS-2000E Series User Manual
Function Keys
The Function keys are used to enter and configure different functions on the GDS-2000E.
Measure
Measure
Configures and runs automatic measurements.
Cursor
Cursor
Configures and runs cursor measurements.
APP
APP
Configures and runs GW Instek applications.
Acquire
Acquire
Configures the acquisition mode, including Segmented Memory acquisition.
Display
Display
Configures the display settings.
Help
Help
Shows the Help menu.
Save/Recall
Save/Recall
Used to save and recall waveforms, images, panel settings.
Utility
Utility
Configures the Hardcopy key, display time, language, probe compensation and calibration. It also accesses the file utilities menu.
Autoset
Autoset
Press the Autoset key to automatically set the trigger, horizontal scale and vertical scale.
18
GETTING STARTED
Run/Stop Key
Run/Stop
Press to Freeze (Stop) or continue (Run) signal acquisition (page 41). The run stop key is also used to run or stop Segmented Memory acquisition (page 89).
Single
Single
Sets the acquisition mode to single triggering mode.
Default Setup
Default
Resets the oscilloscope to the default settings.
Horizontal Controls
The horizontal controls are used to change the position of the cursor, set the time base settings, zoom into the waveforms and search for events.
Horizontal Position
POSITION
PUSH TO
ZERO
The Position knob is used to position the waveforms horizontally on the display screen. Pressing the knob will reset the position to zero.
SCALE
SCALE
The Scale knob is used to change the horizontal scale (TIME/DIV).
Zoom
Zoom
Press Zoom in combination with the horizontal Position knob.
Play/Pause
The Play/Pause key allows you to view each search event in succession – to effectively “play” through each search event. It is also used to play through a waveform in zoom mode.
19
GDS-2000E Series User Manual
Search
Search
The Search key accesses the search function menu to set the search type, source and threshold.
Search Arrows
Use the arrow keys to navigate the search events.
Set/Clear
Set/Clear
Use the Set/Clear key to set or clear points of interest when using the search function.
Trigger Controls
The trigger controls are used to control the trigger level and options.
Level Knob
LEVEL
Used to set the trigger level. Pressing the knob will reset the level to zero.
Trigger Menu Key
Menu
Used to bring up the trigger menu.
50% Key
50 %
Sets the trigger level to the half way point (50%).
Force - Trig
Force-Trig
Press to force an immediate trigger of the waveform.
Vertical POSITION
POSITION
PUSH TO
ZERO
Sets the vertical position of the waveform. Push the knob to reset the vertical position to zero.
Channel Menu Key
CH1
Press the CH1~4 key to set and configure the channel.
20
GETTING STARTED
(Vertical)SCALE Knob
SCALE
Sets the vertical scale of the channel (TIME/DIV).
External Trigger Input
EXT TRIG
Accepts external trigger signals (page 144). Only on 2 channel models.
Input impedance: 1MΩ Voltage input: ±15V(peak), EXT trigger capacitance:16pF.
Math Key
MATH
M
Use the Math key to set and configure math functions.
Reference Key
REF
R
Press the Reference key to set or remove reference waveforms.
BUS Key
BUS
B
The Bus key is used for parallel and serial bus (UART, I2C, SPI, CAN, LIN) configuration.
Channel Inputs
CH1
Accepts input signals. Input impedance: 1MΩ. Capacitance: 16pF CAT I
USB Host Port
2V
TypeA, 1.1/2.0 compatible. Used for data transfer.
21
GDS-2000E Series User Manual
Ground Terminal
2V
Accepts the DUT ground lead for common ground.
Probe Compensation Outputs
2V
The probe compensation output is used for probe compensation. It also has an adjustable output frequency.
By default this port outputs a 2Vpp, square wave signal at 1kHz for probe compensation.
Please see page 188 for details.
Power Switch
POWER
Used to turn the power on/off.
: ON
: OFF
22
Rear Panel
Calibration output
Key lock Power input socketFan
Go/No Go output
LAN portUSB Device port
LINE VOLTAGE
AC 100 240V
RANGE
FREQUENCY 50 60Hz
POWER MAX. 30 Watts
CAUTION
TO AVOID ELECTRIC SHOCK THE POWER CORD PROTECTIVE GROUNDING
DO NOT REMOVE COVERS. REFER SERVICING TO QUALIFIED PERSONNEL.
CONDUCTOR MUST BE CONNECTED TO GROUND.
Ser. No. Label
Calibration Output
Outputs the signal for vertical scale accuracy calibration (page 263).
USB Device Port
The USB Device port is used for remote control.
LAN (Ethernet) Port
The LAN port is used for remote control over a network or when combined with the Remote Disk app, allows the scope to be mounted to a share disk.
GETTING STARTED
23
GDS-2000E Series User Manual
Power Input Socket
Power cord socket accepts AC mains, 100 ~ 240V, 50/60Hz.
For power up sequence, see page
28.
Security Slot
Kensington security slot compatible.
Go-No Go Output
Outputs Go-No Go test results (page 193) as a 500us pulse signal.
24
Display
Memory bar
Reference
waveform
Analog
Waveform
Bus
Channel status Horizontal status
Trigger
configuration
Waveform
frequency
Date and time
Trigger position
Trigger status
Acquistion mode
Trigger level
Channel
Indicators
Memory length
and sample rate
Analog Waveforms
Shows the analog input signal waveforms.
Channel 1: Yellow
Channel 2: Blue
Channel 3: Pink
Channel 4: Green
Bus Waveforms
Shows the bus waveforms for serial buses. The values are displayed in hex or binary.
Channel Indicators
The channel indicators show the zero volt level of the signal waveform for each activated channel. Any active channel is shown with a solid color.
Analog channel indicator
Bus indicator(B)
Reference waveform indicator
Math indicator
Trigger Position
Shows the position of the trigger.
Horizontal Status
Shows the horizontal scale and position.
Date and Time
Current date and time (page 187).
GETTING STARTED
25
GDS-2000E Series User Manual
Trigger Level
Shows the trigger level on the graticule.
Memory Bar
The ratio and the position of the displayed waveform compared to the internal memory (page
104).
Trigger Status
Trig’d
Triggered.
PrTrig
Pre-trigger.
Trig?
Not triggered, display not updated.
Stop
Trigger stopped. Also appears in Run/Stop (page 41).
Roll
Roll mode.
Auto
Auto trigger mode.
For trigger details, see page 144.
Acquisition Mode
Normal mode
Peak detect mode
Average mode
For acquisition details, see page 80.
Signal Frequency
Shows the trigger source frequency.
Indicates the frequency is less than 2Hz (lower frequency limit).
Trigger Configuration
Trigger source, slope, voltage, coupling.
Horizontal Status
Horizontal scale, horizontal position.
For trigger details, see page 144.
Channel Status
Channel 1, DC coupling, 2V/Div.
For channel details, see page 113.
26
Set Up
Tilt
To tilt, pull the legs forward, as shown below.
Stand
To stand the scope upright, push the legs back under the casing as shown below.
Tilt Stand
GETTING STARTED
27
Power Up
Requirements
The GDS-2000E accepts line voltages of 100 ~ 240V at 50 or 60Hz.
Step
1. Connect the power cord to
the rear panel socket.
2. Press the POWER key. The
display becomes active in ~ 30 seconds.
: ON
: OFF
POWER
Note
The GDS-2000E recovers the state right before the power is turned OFF. The default settings can be recovered by pressing the Default key on the front panel. For details, see page 231.
GDS-2000E Series User Manual
28
First Time Use
Background
This section describes how to connect a signal, adjust the scale, and compensate the probe. Before operating the GDS-2000E in a new environment, run these steps to make sure the instrument performs at its full potential.
1. Power On
Follow the procedures on the previous page.
2. Set the Date and Time
Set the date and time.
Page 187
3. Reset System
Reset the system by recalling the factory settings. Press the Default key on the front panel. For details, see page 231.
Default
4. Connect Probe
Connect the probe to the Channel 1 input and to the CAL signal output. This output provides a 2Vp-p, 1kHz square wave for signal compensation by default.
Set the probe attenuation to x10 if the probe has adjustable attenuation.
VARIABLE
POSITION
HORIZONTAL
POSITION
POSITION
POSITION
POSITION
VERTICAL
MATH
REF
BUS
SCALE
TRIGGER
LEVEL
PUSH TO
ZERO
SCALESCALESCALE
PUSH TO ZERO
PUSH TO
ZERO
PUSH TO
ZERO
PUSH TO
ZERO
PUSH TO ZERO
SCALE
POWER
CH1 CH2 CH3 CH4
2V
CAT I
1MW16pF 300Vpk MAX.
Measure
Display
Cursor
Help
Test
Save/Recall
Acquire
Utility
Autoset
Run/Stop
Single
Default
Select
Search
Set/Clear
Zoom
CH1 CH2 CH3 CH4
M
R
B
Menu
50 %
Force-Trig
x1
x10
X 10
X 1
CH1
2V
GETTING STARTED
29
GDS-2000E Series User Manual
5. Capture Signal
(Autoset)
Press the Autoset key. A square waveform appears on the center of the screen. For Autoset details, see page 39.
Autoset
6. Select Vector
Waveform
Press the Display key, and set the display to Vector on the bottom menu.
Display
7. Compensate
Probe
Turn the adjustment point on the probe to make the square waveform edge flat.
Under
Compensation
Normal
Over
Compensation
30