Instek APS-7050, APS-7100 User Manual

Programmable AC Power Source
APS-7000 Series
USER MANUAL
GW INSTEK PART NO. 82PS-71000MA1
ISO-9001 CERTIFIED MANUFACTURER
This manual contains proprietary information, which is protected by copyright. All rights are reserved. No part of this manual may be photocopied, reproduced or translated to another language without prior written consent of Good Will company.
The information in this manual was correct at the time of printing. However, Good Will continues to improve products and reserves the rights to change specification, equipment, and maintenance procedures at any time without notice.
Good Will Instrument Co., Ltd. No. 7-1, Jhongsing Rd., Tucheng Dist., New Taipei City 236, Taiwan.
SAFETY INSTRUCTIONS
Table of Contents
SAFETY INSTRUCTIONS ................................................... 4
GETTING STARTED ........................................................... 8
APS-7000 Series Overview ...................... 9
Appearance .......................................... 13
OPERATION .................................................................... 20
Set Up .................................................. 22
Basic Operation ................................... 43
Advanced Settings ............................... 61
Miscellaneous ...................................... 68
Trigger ................................................. 74
Preset Settings ..................................... 78
Arbitrary Waveform Function ............... 82
Test Mode Function ............................. 89
REAR PANEL SIGNAL OUTPUTS .................................... 126
COMMUNICATION INTERFACE ..................................... 131
Interface Configuration ...................... 132
FAQ ............................................................................... 152
APPENDIX ..................................................................... 153
Firmware Update ............................... 153
APS-7000 Default Settings ................. 155
APS-7000 Specifications ..................... 157
APS-7000 Dimensions ........................ 163
Declaration of Conformity .................. 165
INDEX............................................................................ 166
3
SAFETY INSTRUCTIONS
WARNING
Warning: Identifies conditions or practices that could result in injury or loss of life.
CAUTION
Caution: Identifies conditions or practices that could result in damage to the APS-7000 or to other properties.
DANGER High Voltage
Attention Refer to the Manual
Protective Conductor Terminal
Earth (ground) Terminal
SAFETY INSTRUCTIONS
This chapter contains important safety instructions that you must follow during operation and storage. Read the following before any operation to ensure your safety and to keep the instrument in the best possible condition.
Safety Symbols
These safety symbols may appear in this manual or on the instrument.
4
SAFETY INSTRUCTIONS
Do not dispose electronic equipment as unsorted municipal waste. Please use a separate collection facility or contact the supplier from which this instrument was purchased.
General Guideline
CAUTION
Do not place any heavy object on the APS-7000. Avoid severe impact or rough handling that
leads to damaging the APS-7000.
Do not discharge static electricity to the APS-
7000.
Use only mating connectors, not bare wires, for
the terminals.
Do not block the cooling fan opening. Do not disassemble the APS-7000 unless you are
qualified.
(Measurement categories) EN 61010-1:2010 specifies the measurement categories and their requirements as follows. The APS-7000 falls under category I.
Measurement category IV is for measurement performed at the
source of low-voltage installation.
Measurement category III is for measurement performed in the
building installation.
Measurement category II is for measurement performed on the
circuits directly connected to the low voltage installation.
Measurement category I is for measurements performed on
circuits not directly connected to Mains.
Power Supply
WARNING
AC Input voltage range: 115/230 Vac ± 15% Frequency: 47~63Hz To avoid electrical shock connect the protective
grounding conductor of the AC power cord to an earth ground.
Safety Guidelines
5
APS-7000 Series User Manual
Cleaning the APS­7000
Disconnect the power cord before cleaning. Use a soft cloth dampened in a solution of mild
detergent and water. Do not spray any liquid.
Do not use chemicals containing harsh material
such as benzene, toluene, xylene, and acetone.
Operation Environment
Location: Indoor, no direct sunlight, dust free,
almost non-conductive pollution (Note below)
Relative Humidity: 20%~ 80%, no condensation Altitude: < 2000m Temperature: 0°C to 40°C
(Pollution Degree) EN 61010-1:2010 specifies the pollution degrees and their requirements as follows. The APS-7000 falls under degree
2.
Pollution refers to “addition of foreign matter, solid, liquid, or
gaseous (ionized gases), that may produce a reduction of dielectric strength or surface resistivity”.
Pollution degree 1: No pollution or only dry, non-conductive
pollution occurs. The pollution has no influence.
Pollution degree 2: Normally only non-conductive pollution
occurs. Occasionally, however, a temporary conductivity caused by condensation must be expected.
Pollution degree 3: Conductive pollution occurs, or dry, non-
conductive pollution occurs which becomes conductive due to condensation which is expected. In such conditions, equipment is normally protected against exposure to direct sunlight, precipitation, and full wind pressure, but neither temperature nor humidity is controlled.
Storage environment
Location: Indoor Temperature: -10°C to 70°C
Relative Humidity: 80%, no condensation
Disposal
Do not dispose this instrument as unsorted municipal waste. Please use a separate collection facility or contact the supplier from which this instrument was purchased. Please make sure discarded electrical waste is properly recycled to reduce environmental impact.
6
SAFETY INSTRUCTIONS
Green/ Yellow:
Earth
Blue:
Neutral
Brown:
Live (Phase)
Power cord for the United Kingdom
When using the instrument in the United Kingdom, make sure the power cord meets the following safety instructions.
NOTE: This lead/appliance must only be wired by competent persons
WARNING: THIS APPLIANCE MUST BE EARTHED IMPORTANT: The wires in this lead are coloured in accordance with the following code:
As the colours of the wires in main leads may not correspond with the coloured marking identified in your plug/appliance, proceed as follows:
The wire which is coloured Green & Yellow must be connected to the Earth terminal marked with either the letter E, the earth symbol
or coloured Green/Green & Yellow.
The wire which is coloured Blue must be connected to the terminal which is marked with the letter N or coloured Blue or Black.
The wire which is coloured Brown must be connected to the terminal marked with the letter L or P or coloured Brown or Red.
If in doubt, consult the instructions provided with the equipment or contact the supplier.
This cable/appliance should be protected by a suitably rated and approved HBC mains fuse: refer to the rating information on the equipment and/or user instructions for details. As a guide, a cable of 0.75mm2 should be protected by a 3A or 5A fuse. Larger conductors would normally require 13A types, depending on the connection method used.
Any exposed wiring from a cable, plug or connection that is engaged in a live socket is extremely hazardous. If a cable or plug is deemed hazardous, turn off the mains power and remove the cable, any fuses and fuse assemblies. All hazardous wiring must be immediately destroyed and replaced in accordance to the above standard.
7
APS-7000 Series User Manual
APS-7000 Series Overview ................................................. 9
Series lineup ................................................................................................................ 9
Operating Area ........................................................................................................... 9
Main Features ........................................................................................................... 10
Accessories ............................................................................................................... 11
Appearance ..................................................................... 13
Front Panel ............................................................................................................... 13
Rear Panel ................................................................................................................. 17
GETTING STARTED
This chapter describes the power source in a nutshell, including its main features and front / rear panel introduction.
8
GETTING STARTED
Model name
Max. Output Current
Power Rating
Output Voltage
APS-7050
4.2A/2.1A
500VA
0~310.0 Vrms
APS-7100
8.4A/4.2A
1000VA
0~310.0 Vrms
Voltage (V)
Current (A)
0
1
2
0
3
4
5
100
200
300
400 500
600
120 155 240
310 480
600
4.2
3.23
2.1
1.62
1.05
0.83
APS-7050 Output Operating Area
155V Range 310V Range 600V Range
(APS-003 Option)
APS-7000 Series Overview
Series lineup
The APS-7000 series consists of 2 models, the APS-7050 and the APS-7100, differing only in capacity. Note that throughout the user manual, the term “APS-7000” refers to both the APS-7050 and APS­7100, unless stated otherwise.
Operating Area
9
Voltage (V)
Current (A)
0
2
4
0
6
8
10
100
200
300
400 500
600
120 155 240
310 480
600
8.4
6.45
4.2
3.23
2.1
1.67
155V Range 310V Range 600V Range
APS-7100 Output Operating Area
(APS-003 Option )
Main Features
Performance
Low output ripple and noise Excellent and feature-rich measurement
capacity
Standard maximum output voltage is 310Vrms Maximum output voltage and frequency of
600Vrms(APS-003 Option)/999.9Hz(APS-004 Option)
APS-7000 Series User Manual
10
Features
OVP, OCP and OTP protection Variable voltage, frequency and current limiter Sequence and simulation function Large 4.3 inch TFT panel Globally adjustable power inlet not restricted by
the power supply environment
USB interface is equipped as standard with the
ability to save and recall setup files.
Only 88mm (2U) case height.
Interface
Ethernet port USB host RS-232 / USB CDC (optional) GPIB (optional)
Accessories
Standard Accessories
Part number
Description
CD ROM
User manual, programming manual
4322-30700101 (UL/CSA)
Type I Power cord (APS-7050)
4322-B0700101 (PSE)
Type II Power cord (APS-7050)
4300-31000101 (UL/CSA)
Type I Power cord (APS-7100)
4300-B1000201 (PSE)
Type II Power cord (APS-7100)
62PS-7K0SC401 x1 5302-01613001 x1
Mains terminal cover set (APS-7050)
GETTING STARTED
11
APS-7000 Series User Manual
62PS-7K0SC701 x1 5302-01613001 x2
Mains terminal cover set (APS-7100)
GTL-123
Test leads: 1x red, 1x black
Optional Capacity
Part number
Description
APS-003
Output Voltage Capacity: 0 ~ 600Vrms
APS-004
Output Frequency Capacity: 45 ~ 999.9Hz
Optional Accessories
Part number
Description GRA-423
APS-7000 rack mount kit
APS-001
GPIB interface card
APS-002
RS-232 / USB interface card
Download
Name
Description
gw_aps.inf
USB driver
12
GETTING STARTED
APS-7050
AC Power Source
Display
Programmable
F 1
F 2
F 3
F 4
Output
: Long Push
Menu Test Preset
V
Local
Off Phase
On Phase
0
1
4
7
2
5
8
3
6
9
V-Limit F-Limit IPK-Limit
F I rms
Cancel
Shift
Range
Enter
Lock
ALM CLR
IPK CLR
Unlock
Surge / Dip
Ramp
TriggerARB
USB A port
Power switch
Front voltage output socket
Menu, Test, Preset keys
LCD display
Function keys
Display mode key
Arrow keys, Scroll wheel
Voltage, Frequency, Current setting and limit
Keypad
Range, Shift, Cancel, Enter, Lock keys
Output key
Item
Description
Power Switch
Turns on the mains power.
Front Voltage Output Socket
Neutral
GND
Line
Output voltage terminal using a regional universal plug. There is a Euro and a Universal regional plug.
CAUTION
Maximum allowable output voltage and current are 250Vrms and 10Arms.
For voltages exceeding 250Vrms, please use the rear output terminal.
Appearance
Front Panel
APS-7050, APS-7100
13
APS-7000 Series User Manual
USB A Port
The USB port is used for data transfers and upgrading software.
LCD Screen
Displays the measured values or menu system.
Display Mode Select Key
Display
Selects between Standard mode and Simple mode.
Function Keys
F 1
Assigned to the functions displayed on the right-hand side of the screen.
Menu Key
Menu
Enters the Main menu or goes back to one of the display modes.
Test Key
Test
Puts the instrument into the Sequence, Simulation and Program Control mode.
Preset Key
Preset
Puts the instrument into Preset mode.
Arrow Keys
The arrow keys are used to select the digit power of a value that is being edited.
V
V
V-Limit
Used for setting the output voltage.
V-Limit
(Shift + V)
Used for setting the output voltage limit value.
F
F
F-Limit
Used for setting the output frequency.
F-Limit
(Shift + F)
Used for setting the output frequency limit value.
I rms
I rms
IPK-Limit
Used for setting the maximum output current.
IPK-Limit
(Shift + I rms)
Used to set the peak output current limit value.
14
GETTING STARTED
Range Key
Range
Switches between the 155V, 310V and 600V ranges (the 600V range is an option).
Scroll Wheel
Used to navigate menu items or for incrementing/decrementing values one step at a time.
Lock Key
Lock
Unlock
Locks the number pad to prevent accidentally changing panel settings.
Unlock Key
(Long press)
Disables the key lock.
Enter Key
Enter
Confirms selections / settings
Cancel Key
Cancel
Clears entries that are made in the number entry dialog when a value is edited using the arrow keys or the scroll wheel.
The Cancel key can also be used to cancel function setting menus or dialogs.
Shift Key
Shift
Turns on the shift state, which enables shortcut operations.
Output Key
Output
Turns the output on or off.
Number Pad
0
142
5
8
3
6
9
Lock
Unlock
7
On Phase Surge / Dip IPK CLR
Off Phase Ramp ALM CLR
ARB Trigger
Local
Used to enter values.
Local Mode
(Shift + 0)
Switches operation back to local mode from remote mode.
ARB Mode
(Shift + 1)
Sets the ARB function.
Trigger Mode
(Shift + 2)
Sets the JI port trigger behavior on the rear panel.
15
APS-7000 Series User Manual
Off Phase
(Shift + 4)
Sets the off phase for the output voltage.
RAMP
(Shift + 5)
Quick settings for Ramp control.
ALM CLR
(Shift + 6)
Clears alarms.
On Phase
(Shift + 7)
Sets the on phase for the output voltage.
Surge/Dip
(Shift + 8)
Quick settings for Surge/Dip control.
IPK CLR
(Shift + 9)
Clears peak current protection.
16
Rear Panel
SYNC
Signal Output LAN
OUTPUT INPUT
L N N L
600Vac MAX. 115 / 230V 15%
50 / 60Hz
3.6kVA MAX.
Out On / Off
COM
J1
Trigger Out
Trigger In
115 / 230V 15%
50 / 60Hz
1.8kVA MAX.
SYNC
SER. NO. LB
L N
600Vac MAX.
Signal Output
Trigger Out
Trigger In Out On / Off
COM
LAN
J1
OUTPUT
Fan
GPIB
Line voltage input
GPIB
RS232
RS-232/
USB Device
Sync
Signal Output Rear panel voltage
output terminal
Remote control Ethernet port
Line voltage input
APS-7050
APS-7100
Opt. int. port
Line Voltage Input
APS-7050
115 / 230V 15%
50 / 60Hz
1.8kVA MAX.
Voltage Input: 115/230±15% VAC; Line frequency: 50Hz/60 Hz (Automatically switchable)
GETTING STARTED
17
APS-7000 Series User Manual
APS-7100
OUTPUT INPUT
L N N L
Voltage Input: 115/230±15% VAC ; Line frequency: 50Hz/60 Hz (Automatically switchable)
Rear Voltage Output Socket
Output voltage terminal.
APS-7050
L N
OUTPUT
APS-7100
OUTPUT INPUT
L N N L
Sync Output Socket
SYNC
BNC socket. This socket will output a signal of approximately 10V when the output is on.
Signal Output
Signal Output
Connector for monitoring PASS, FAIL and PROCESSING output signals when using the Program mode.
Remote Control
Trigger Out
Trigger In Out On / Off
COM
J1
Connector for controlling the TRIGGER IN, TRIGGER OUT and OUT ON/OFF states.
Ethernet Port
LAN
The Ethernet port is used for remote control and digital monitoring from a PC.
Optional Interface Slot
Optional GPIB communication, RS-232/USB B communication.
FAN
Temperature controlled fan.
18
GETTING STARTED
Status bar
/
Indicates if the output is ON or OFF.
Indicates the output power as a percentage of full scale.
Indicates that the Surge/Dip function is active.
Indicates that the Ramp function is active.
The alarm icon will appear on the status bar when one of the protection functions are tripped. Applies to Over Power, Over Irms, Over Ipeak and Over Temperature protection.
Indicates that the APS-7000 is in remote mode.
Indicates that a USB drive is detected in the front panel host port.
Indicates that the LAN interface is activated.
Indicates that the panel lock is active.
Status Bar Icons
19
APS-7000 Series User Manual
Set Up ............................................................................. 22
Line Voltage Connection (APS-7100) ................................................................... 22
Power Up .................................................................................................................. 25
Output Terminals .................................................................................................... 26
Installing the Optional Modules ............................................................................ 30
Using the Rack Mount Kit ..................................................................................... 32
How to Use the Instrument ................................................................................... 32
Reset to Default Settings ........................................................................................ 37
View System Version and Build Date ................................................................... 38
LCD Configuration ................................................................................................. 39
USB Driver Installation .......................................................................................... 40
Basic Operation .............................................................. 43
Setting the Voltage Range ....................................................................................... 43
Setting the Voltage Limit ........................................................................................ 44
Setting the Output Voltage ..................................................................................... 45
Setting the Frequency Limit ................................................................................... 47
Setting the Output Frequency ................................................................................ 48
Setting the Peak Current Limit .............................................................................. 49
Setting the Current RMS Level .............................................................................. 52
Setting the On/Off Phase ...................................................................................... 55
Alarm Clear ............................................................................................................... 56
Display Modes .......................................................................................................... 57
Panel Lock ................................................................................................................ 59
Turning the Output On .......................................................................................... 60
Advanced Settings ........................................................... 61
Surge/Dip Control .................................................................................................. 61
Ramp Control ........................................................................................................... 65
Miscellaneous ................................................................. 68
T Ipeak, hold ............................................................................................................ 68
Power ON Output................................................................................................... 70
OPERATION
20
OPERATION
Buzzer ........................................................................................................................ 71
SCPI Emulation ....................................................................................................... 72
Program Timer Unit ................................................................................................ 73
Trigger ............................................................................ 74
Trigger Control Settings .......................................................................................... 74
Preset Settings ................................................................ 78
Save Preset Settings to Local Memory .................................................................. 78
Load Preset Settings to Local Memory ................................................................. 79
Manage Preset Settings ............................................................................................ 80
Arbitrary Waveform Function ........................................... 82
ARB Mode Overview .............................................................................................. 83
Selecting an ARB Waveform .................................................................................. 86
Test Mode Function ......................................................... 89
Simulate Mode Overview ....................................................................................... 91
Simulate Settings ...................................................................................................... 95
Save a Simulation to Local Memory ...................................................................... 98
Recall a Simulation from Local Memory .............................................................. 98
Manage Simulation Settings .................................................................................... 99
Running a Simulation ............................................................................................ 101
Sequence Mode Overview .................................................................................... 103
Sequence Settings ................................................................................................... 107
Save a Sequence to Local Memory ...................................................................... 111
Recall a Sequence from Local Memory .............................................................. 111
Manage Sequence Settings .................................................................................... 112
Running a Sequence .............................................................................................. 114
Program Mode Overview ..................................................................................... 116
Save a Program to Local Memory ....................................................................... 123
Recall a Program from Local Memory ................................................................ 123
21
APS-7000 Series User Manual
Background
The APS-7100 is equipped with an input power terminal that can accept 115V/230V ± 15%. To connect or replace the power cord (GW Instek part number: APS-7100: 4300-31000101, use the procedure below:
Warning
The following procedure should only be attempted by competent persons.
Ensure the AC power cord is not connected to power.
Removal
1. Turn off the power switch.
1
APS-7100
2. Unscrew the power cord protective sheath on
the rear output socket.
3. Remove the 2 screws holding the power cord
cover and remove.
Set Up
Line Voltage Connection (APS-7100)
22
OPERATION
APS-7100
3
2
4. Remove the AC
power cord wires.
4
CAUTION
The power line inputs for the APS-7100 are on the outer cluster of terminals.
The terminals to the inner of the panel are the rear panel outputs.
Installation
1. Connect the AC power cord wires to the AC
input terminals.
White/Blue Neutral (N) Green/Green-yellowGND ( ) Black/Brown Line (L)
23
APS-7000 Series User Manual
OUTPUT INPUT
L N N L
115 / 230V 15%
Line
Neutral
Ground
2. Re-install the power cord cover.
3. Screw the power cord sheath back onto the
cover.
2
3
24
Power Up
Steps
1. Socket type: Connect the
power cord to the rear panel socket.
Input Power Terminal: Connect the power cord to the input power terminals.
Page 22
2. Press the POWER key. The splash screen will
appear momentarily before the continuous mode screen appears with the settings loaded.
CAUTION
The power supply takes around 10 seconds to fully turn on and shutdown.
Do not turn the power on and off quickly.
OPERATION
25
APS-7000 Series User Manual
Background
The output terminals can be output from either the front panel or from the rear panel. The outputs are limited to 4.2A/2.1A (APS-7050) or
8.4A/4.2A(APS-7100).
Supported plugs
Multi-region terminal Socket
Supported Standards
IEC, North America, Japan.
EURO CEE type universal plug
WARNING
Dangerous voltages. Ensure that the power to the instrument is disabled before handling the power supply output terminals. Failing to do so may lead to electric shock.
CAUTION
For the front panel output, the maximum output voltage is 250VAC and current is 10A.
Front Panel Output Connection
1. The front panel has a multi-region power
socket depending on the socket type.
2. Insert the plug from the DUT into the socket.
EURO CEE socket
IEC North America, Japan
Output Terminals
26
OPERATION
3. Turn the power on. The AC power supply is
now ready to power the DUT.
Rear Panel Output Connection
The rear panel output is used to supply higher power DUTs. The rear panel output connection is similar to the universal rear panel line input connection on the APS-7100.
1. Disconnect the unit from the mains power
socket and turn the power switch off.
2. Unscrew the power cord protective sheath.
3. Remove the 2 screws holding the power cord
cover and remove.
APS-7050
3
2
APS-7100
3
2
27
APS-7000 Series User Manual
CAUTION
For the APS-7100, there is a single bank for the input and output terminals. Ensure the correct terminals are connected. The APS-7050 only has a single bank of output terminals on the rear panel.
Installation
4. Connect the output AC power cord wires to
the AC output terminals.
Black Neutral (N) Green GND ( ) Black Line (L)
OUTPUT INPUT
L N N L
115 / 230V 15%
Line
Neutral
Ground
Ground
Neutral
Line
APS-7100 shown. The input terminals are already connected and shows which output terminals are to be connected.
5. Re-install the power cord cover.
6. Screw the power cord sheath back onto the
cover.
28
OPERATION
APS-7100
5
6
APS-7050
5
6
7. Turn the power on. The AC power supply is
now ready to power the DUT.
29
APS-7000 Series User Manual
Background
There are a number of optional modules that can be installed for remote control or for the SCPI compatibility modes.
Optional Modules
APS-001
GPIB Interface card
APS-002
RS-232/USB interface card
WARNING
Dangerous voltages. Ensure that the power to the instrument is disabled before handling the power supply output terminals. Failing to do so may lead to electric shock.
CAUTION
Ensure the power is off before installing any of the optional modules.
Installation
1. Turn off the power switch.
1
2. Unscrew the two screws holding the options
panel plate.
2
Installing the Optional Modules
30
OPERATION
3. Slide the module PCB onto the rails on the
inside of the module slot.
4. Secure the module with the screws that were
removed from step 2.
4
3
5. The module will be recognized upon startup.
31
APS-7000 Series User Manual
Background
The APS-7000 series has an optional Rack Mount Kit (GW Instek part number: GRA-423). The APS-7050 and APS-7100 are designed to fit into a 2U rack height. Please see your distributor for further rack mount details.
Rack mount diagram
CAUTION
Ensure adequate ventilation is provided when using the rack mount. Ensure at a gap of at least 50mm is given for the side air intakes. Failure to do so may cause the instrument to overheat.
Background
The APS-7000 AC power supplies generally use the scroll wheel, arrow keys and Enter keys to edit numerical values or to select menu options.
Menu navigation is performed using the menu keys and function keys on the front panel.
The following section will explain some of these concepts in detail.
Using the Rack Mount Kit
How to Use the Instrument
32
OPERATION
Selecting Menu Items
1. Turn the scroll wheel to select
parameters in menus and lists. The selected parameter will be highlighted in orange. The scroll wheel is also used to increment/decrement setting values.
2. Press the Enter key to edit the
parameter or to enter the selected menu.
Enter
Example
Selected parameter
The following is an example of the menu list that appears when the Menu key is pressed.
Using the Keypad to edit parameter values
When editing a value the keypad can be used to directly enter the desired value.
1. Type the value of the
parameter using the keypad.
0
1
4
2
5
8
3
6
9
Lock
Unlock
7
On Phase Surge / Dip IPK CLR
Off Phase Ramp ALM CLR
ARB Trigger
Local
2. Press the Enter key to confirm the
edit.
Enter
33
APS-7000 Series User Manual
Example
Parameter
Using the Arrow Keys and Scroll wheel to edit parameter values
Use the arrow keys to select a digit power and then use the scroll wheel to edit the value by that power.
1. Use the arrow keys to move the
cursor to the digit of the desired power.
2. Turn the scroll wheel to edit the
value by the resolution of the selected digit.
Cursor
3. Repeat the steps above for all the relevant
digits.
4. Press the Enter key to confirm the
edit.
Enter
34
OPERATION
Note
By default the cursor starts at the lowest power digit.
Using the onscreen keyboard
The onscreen keyboard is only used in the Program Mode. The screenshot below is an example of the onscreen keyboard.
On screen keyboard
Entered characters
Using the Function Keys
The function keys are quick settings keys, the function of which depends on the current menu or operation.
1. Press the function key that corresponds to the
setting directly to its left.
2. The setting or parameter is immediately
executed.
Function keys
APS-7050
AC Power Source
Display
Programmable
Output
: Long Push
Menu Test Preset
V
Local
Off Phase
On Phase
0
1
4
7
2
5
8
3
6
9
V-Limit F-Limit IPK-Limit
F I rms
Cancel
Shift
Range
Enter
Lock
ALM CLR
IPK CLR
Unlock
Surge / Dip
Ramp
TriggerARB
F 1
F2
F3
F4
Corresponding quick
settings
35
APS-7000 Series User Manual
3. Repeat the steps above for all the relevant
digits.
36
OPERATION
Background
The default settings can be restored from the Menu key settings. See page 155 for the default factory settings.
Steps
1. Press the Menu key. The Menu
settings will appear on the display.
Menu
2. Use the scroll wheel to go to item 10, Default
Setting.
3. Press Enter x2 to restore the default settings.
Default settings
Reset to Default Settings
37
APS-7000 Series User Manual
Background
The Menu>System Information setting displays the serial number and version number.
Steps
1. Press the Menu key. The Menu
setting will appear on the display.
Menu
2. The system information should now be listed
on the display.
If not, use the scroll wheel to go to item 1, System Information.
System
Information
View System Version and Build Date
38
OPERATION
Background
The LCD Configuration setting sets the brightness, contrast and saturation level of the LCD display.
Steps
1. Press the Menu key. The Menu
settings will appear on the display.
Menu
2. Use the scroll wheel to go to item 9, LCD
Configuration and press Enter.
3. Set the brightness, contrast and saturation.
Contrast(%)
1 ~ 100% (Default=50%)
Brightness(%)
1 ~ 100% (Default=50%)
Saturation(%)
1 ~ 100% (Default=50%)
Exit
4. Press Exit[F4] to exit from the
Ramp Control settings.
Default Settings
5. Press Default[F3] to set all the LCD settings to
50%.
LCD settings
Default
settings
LCD Configuration
39
APS-7000 Series User Manual
Background
If the USB Type B interface is to be used for remote control, the USB driver needs to be installed.
Note
The USB driver, GW_APS.inf, is located on the CD Rom that accompanied this user manual. Alternatively the driver can be downloaded from the GW Instek website.
For information on the USB interface, see page
132.
Steps
1. Connect the rear panel USB -B port on the APS-
7000 to the PC using a USB Type A to B cable.
2. Go the Windows Device Manager.
For Windows 7:
Start > Control Panel > Hardware and Sound > Device Manager
3. The APS-7000 will be located under Other
Devices in the hardware tree. Right-click the APS-70XX and choose Update Driver Software.
USB Driver Installation
40
OPERATION
4. From the hardware wizard choose Browse my
computer driver software.
5. Set the file path to the location of the USB
driver, click Next and finish the driver installation.
41
APS-7000 Series User Manual
6. In APS-7000 will now be located in the Ports
node of the hardware tree in the Windows Device Manager if the driver installation was successful.
42
OPERATION
Background
The Range setting determines the general outlet voltage range. The ranges available correspond to common mains output voltage standards.
Steps
1. Press Range to access the Range
menu.
Range
Basic Operation
This section describes the basic operations required to operate the power supply.
Setting the Voltage Range → from page 43 Setting the Voltage Limit → from page 44 Setting the Output Voltage → from page 45 Setting the Frequency Limit page 47 Setting the Output Frequency → page 48 Setting the Peak Current Limit → from page 49 Setting the Current RMS Level → from page 52 Setting the On/Off Phase → from page 55 Clearing the Alarm → from page 56 Setting the Display mode → from page 57 Panel lock → from page 59 Turning the Output on/off → from page 60
Before operating the power supply, please see the Getting Started chapter, page 8.
Setting the Voltage Range
43
APS-7000 Series User Manual
2. Set the voltage range with the scroll wheel or
with the F1 ~ F4 soft-keys.
Range
AUTO, 600V(option), 310V, 155V
Soft-keys
AUTO, 600V(option), 310V, 155V
3. Press Enter to confirm the Range setting.
Range setting
Soft-key
settings
F1 F2 F3 F4
Note
If the range is changed from 155V to 600V, the Irms and IPK values will automatically be changed to a lower value. If the range is changed from 600V to 155V, the Irms and IPK values remain the same.
If the voltage range is changed when the output is on, the output will be automatically turned off.
Background
Setting the voltage limit allows the output voltage to be set to any level within the voltage limit (V Limit) range.
Steps
1. Press Shift + V to access the Volt
Limit menu.
Shift
+
V
V-Limit
Setting the Voltage Limit
44
OPERATION
2. Set the voltage limit with the scroll wheel or
with the F3 ~ F4 soft-keys. The MAX and MIN soft-keys set the limit to the maximum and minimum, respectively.
Range
10% of full range ~ full range
Soft-keys
MAX, MIN
3. Press Enter to confirm the voltage limit setting.
Volt Limit
Min/Max
settings F3 F4
Note
Each voltage range (155V, 310V, 600V) has an independent voltage limit.
Background
Before setting the power supply voltage level, set the voltage range and voltage limit.
Steps
1. Press the V key. The ACV
parameter will be editable.
V
V-Limit
2. Set the voltage with the scroll wheel/keypad or
with the F1 ~ F4 soft-keys.
Range
0 volts ~ full range
Soft-keys
DEF1, DEF2, MAX, MIN
Setting the Output Voltage
The voltage setting sets the voltage level of the power supply.
45
APS-7000 Series User Manual
3. Press Enter to confirm the voltage setting.
Preset Settings
The DEF1 and DEF2 preset settings are user defined settings. By default they are set to 0.00 volts. The MAX and MIN soft-keys set the voltage to the maximum and minimum, respectively.
4. Press the V key and set the desired voltage with
the scroll wheel/keypad.
Range
0 volts ~ full scale of voltage range
5. Press and hold the DEF1 or DEF2 soft-key until
“Saved to DEF1/2” is displayed. This will save the voltage setting to the DEF1 or DEF2 soft­key.
Note
Trying to set the voltage outside of the voltage limit/range will result in a voltage setting error being displayed on the screen.
The voltage level can be set when the output is on.
Example
Voltage setting
Preset voltage
settings
F3 F4
F1 F2
46
OPERATION
Background
Setting the frequency limit allows the frequency output to be set to any level within the limit range.
Steps
1. Press Shift + F to access the Freq
Limit menu.
Shift
+
F
F-Limit
2. Set the frequency limit with the scroll
wheel/keypad or with the F3 ~ F4 soft-keys. The MAX and MIN soft-keys set the frequency limit to the maximum and minimum, respectively.
Range
45.00 ~ 500.0Hz (999.9Hz option)
Soft-keys
MAX, MIN
3. Press Enter to confirm the limit setting.
Example
Freq Limit
Min/Max
settings F3 F4
Setting the Frequency Limit
47
APS-7000 Series User Manual
Background
Before setting the frequency, set the frequency limit.
Steps
1. Press the F key. The FREQ
parameter will be editable.
F
F-Limit
2. Set the frequency with the scroll wheel/keypad
or with the F1 ~ F4 soft-keys.
Range
45.00 ~ 500.0Hz (999.9Hz option)
Soft-keys
DEF1, DEF2, MAX, MIN
3. Press Enter to confirm the frequency setting.
Preset Settings
The DEF1 and DEF2 preset settings are user defined settings. By default they are set to
50.00Hz and 60.00Hz, respectively. The MAX and MIN soft-keys set the frequency to the maximum and minimum, respectively.
4. Press the F key and set the desired frequency
with the scroll wheel/keypad.
Range
45.00 ~ 500.0Hz (999.9Hz option)
5. Press and hold the DEF1 or DEF2 soft-key until
“Saved to DEF1/2” is displayed. This will save
the frequency setting to the DEF1 or DEF2 soft­key.
Setting the Output Frequency
The frequency setting sets the frequency of the output.
48
OPERATION
Example
Frequency setting
Preset frequency
settings
F3 F4
F1 F2
Note
Trying to set the frequency outside of the frequency limit will result in a frequency setting error being displayed on the screen.
The frequency can be set when the output is on.
Background
Setting the peak current limit sets a limit on the current that can be sourced by the power supply.
Note
When the peak current limit is tripped, an alarm will sound. Press Shift + 9 to clear the Ipk alarm.
Shift+6 can also clear the Ipk alarm. See page 56 for details.
Steps
1. Press Shift + I rms to access the
Ipeak menu.
Shift
+
I rms
IPK-Limit
Setting the Peak Current Limit
49
APS-7000 Series User Manual
2. Set the peak current with the scroll
wheel/keypad or with the F3 ~ F4 soft-keys. The MAX and MIN soft-keys set the peak current to the maximum and minimum, respectively.
Range
10% ~ 100% peak current value. The peak current value depends on the selected voltage range.
Soft-keys
MAX, MIN
3. Press Enter to confirm the peak current setting.
Example
Ipeak
Min/Max
settings F3 F4
Delay Time Settings
The Delay Time setting essentially defines how long the measurement of the peak current must be sustained for before it is recognized. By default the delay time is turned off.
Load on
Ipk Limit reached
I Pk Limit Delay
I
Ipk Limit acknowledged,
protection triggered
T
50
OPERATION
4. Press Shift + I rms and then press DELAY[F2].
5. Set the desired delay time with the scroll
wheel/keypad or with the F3 ~ F4 soft-keys. The MAX and MIN soft-keys set the delay time to the maximum and minimum, respectively.
Range
0(off) ~ 10 seconds
Soft-keys
MAX, MIN
6. Press Enter to confirm the delay time setting.
Example
Delay Time
Min/Max
settings F3 F4
51
APS-7000 Series User Manual
Background
The I rms setting sets the root mean square current.
Steps
1. Press I rms to access the I rms
menu.
I rms
Ipk-Limit
2. Set the I rms level with the scroll wheel/keypad
or with the F3 ~ F4 soft-keys. The MAX and MIN soft-keys set the I rms level to the maximum and minimum, respectively.
Range
0.00 ~ full scale A (dependant on the voltage range)
Soft-keys
MAX, MIN
3. Press Enter to confirm the current setting.
Example
Irms
Min/Max
settings F3 F4
Note
Setting the I rms level to 0.00 will disable OCP.
Setting the I rms to 0V is dangerous.
Setting the Current RMS Level
52
OPERATION
I rms Delay Time Settings
The Delay Time setting defines how long the I rms measurement must be sustained for before it is recognized. By default the I rms delay time is turned off.
Load on
Irms Limit
reached
Irms Limit Delay
I
Irms Limit
acknowledged,
protection triggered
T
Irms Limt
4. Press I rms and then press DELAY[F2].
5. Set the desired delay time with the scroll
wheel/keypad or with the F3 ~ F4 soft-keys. The MAX and MIN soft-keys set the delay time to the maximum and minimum, respectively.
Range
0(off) ~ 10 seconds
Soft-keys
MAX, MIN
6. Press Enter to confirm the delay time setting.
Example
Delay Time
Min/Max
settings F3 F4
53
APS-7000 Series User Manual
OC Fold Settings
The over current fold back settings allow the APS-7000 to work as either a constant voltage source or a constant current source.
While the unit is sourcing less current than the Irms current limit, the APS-7000 will act as a constant voltage source. In this mode, the voltage level will remain constant while the current level may vary. This is the normal operating mode.
When the current level reaches the Irms limit, the APS-7000 will act as a constant current source. In this mode the current is constant and the voltage level varies. When the current subsides below the Irms limit again, the unit will again act as a constant voltage source. When OC Fold is turned off, the APS-7000 will act as a current limiting power source when the Irms limit has been reached.
Vrms
setting
ACV
I
Irms limit
Irms
Note
OC-FOLD can only be active when the I rms level is greater than 0.
7. Press I rms and then press OC-FOLD[F1] toggle
the OC-Fold function on or off.
54
OPERATION
OC-Fold setting
Background
The on phase setting sets the starting phase of the voltage output. The off phase setting sets the ending phase of the voltage output.
Steps
1. Press Shift + 7 or Shift + 4
the On Phase or Off Phase, respectively.
Shift
+
7
On Phase
or
4
Off Phase
2. Set the On Phase or Off Phase setting with the
scroll wheel/keypad or with the F3 ~ F4 soft­keys. The MAX and MIN soft-keys set the phase to the maximum and minimum, respectively.
Range
0 ~ 359º
Soft-keys
MAX, MIN
Setting the On/Off Phase
55
3. Press Enter to confirm the phase setting.
Example
On Phase
Min/Max
settings F3 F4
Off Phase
Alarm Clear
Background
The ALM CLR (Alarm Clear) function will clear any Over Power, Over Irms, Over Ipeak and Over Temperature alarms.
Applicable Alarms
OVER POWER, OVER IRMS, OVER IPEAK, OVER TEMPERATURE
Steps
1. Press Shift + 6 to clear any alarms.
Shift
+
6
ALM CLR
Example
Alarm message
ALM indicator
APS-7000 Series User Manual
56
OPERATION
Steps
1. Press the Display key.
2. The display mode will toggle each
time the key is pressed.
Display
Standard Mode
Settings Measurements
Configurable
measurements
Hold
measurement
Simple Mode
Measurement Items
Hold
measurement
Configuring the Standard Mode Measurements
1. Press the Meas Item1, Item2 or Item3
soft-key.
Display Modes
The APS-7000 power supply has two display modes. The standard display mode shows the power supply setup on the left and the 3 configurable measurements on the right. The simple display mode shows all measurement items available on the APS-7000.
57
APS-7000 Series User Manual
2. Use the scroll wheel to select a measurement
item and press Enter to confirm.
Sets first measurement item to Voltage
Hold Measurement
The Hold function will “hold” the current measurements on the display. Measurements
won’t be updated on the display until the
function is released.
Press HOLD[F4] to toggle hold on or off.
58
OPERATION
Activate the panel lock
Press the Lock key to active the panel lock. “Keys locked” appears on the display.
A lock icon will appear in the top corner when the panel keys are locked.
Lock
Unlock
Disable the panel lock
Hold the Lock key for ~3 seconds to disable the panel lock. “Keys unlocked” will appear on the display and the padlock icon will disappear.
Lock
Unlock
(hold)
Example
Message Lock icon
Panel Lock
The panel lock feature prevents settings from being changed accidentally. When activated, all keys and knobs except the Lock/Unlock key and the Output key (if active) will be disabled.
If the instrument is remotely controlled via the USB/LAN/RS­232/GPIB interface, the panel lock is automatically enabled. See page 131 for remote control details.
59
APS-7000 Series User Manual
Warning
Both of these outputs are electrically linked. Only one DUT should be connected to any one of the outputs at a time. Using both outputs at the same time is not supported. Using the front and rear outputs are the same time could cause dangerous operating conditions. See page 26 for details about using the output terminals or sockets.
Turn Output On
Press the Output key. The Output key will light up and ON will be displayed in the status bar to indicate that the output is on.
Output
Turn Output Off
Press the Output key. The Output key light will go out and OFF will be displayed in the status bar to indicate that the output is off.
Output
Turning the Output On
When the output is turned on, the DUT can be connected to either the rear panel output or the front panel output.
60
OPERATION
Parameter Settings
Mode
Auto: When the output is on, this mode will automatically generate a trigger at 0º to keep the surge or dip event repeatedly on site.
Manual: When the output is on, this mode will wait for the TRIG[F4] soft­key to be pressed before starting the surge or dip event on site.
OFF: Disables surge/dip control.
Note
The magnitude of the surge/dip part of the resultant waveform depends on the surge/dip ACV setting level.
Advanced Settings
Surge/Dip Control → from page 61 Ramp Control → from page 65
Surge/Dip Control
Surge and dip control allows the power supply to source artificial surges or dips in voltage to a DUT. The surge/dip control feature provides a fast method to generate a surge/dip voltage event on a nominal voltage.
There have 4 parameters for configuring this feature: Mode selection (Mode), surge/dip voltage (ACV), the start time (T1) and testing duration(T2) of the surge/dip voltage period. The nominal voltage and frequency settings are based on the Basic Operation section.
61
APS-7000 Series User Manual
Triggering Example:
Site selection: The surge/dip site is selected as shown below.
Trigger (0º for Auto)
Dip/Surge site
T2T1
ACV
100mS
ACV
Sets the ACV surge/dip level from the 0V level.
T1
Sets the T1 time.
T2
Sets the width of the surge/dip.
100mS
Fixed 100mS delay after triggering.
Steps
1. Press the Menu key. The Menu
setting will appear on the display.
Menu
2. Use the scroll wheel to go to item 2, Surge/Dip
Control and press Enter.
3. Go to the Mode setting using the scroll wheel
and press Enter. Select the desired mode and press Enter again to confirm.
The Manual mode will allow you to manually trigger the surge/dip site. The Automatic setting will automatically trigger the surge/dip site.
Mode
Manual, Auto, OFF
4. Set the remaining parameters.
Note: these parameters are not visible when MODE is set to OFF.
Remaining parameters
ACV, T1, T2
62
OPERATION
Exit
5. Press Exit[F4] to exit from the
Surge/Dip Control settings.
6. After exiting the menu, the
surge/dip control icon will appear in the status bar.
Triggering the Manual Surge/Dip Site
For the manual mode, the surge/dip site is determined by a manual trigger.
1. Configure the nominal voltage and
frequency settings. See the Basic Operation chapter for details.
Page: 45, 47
2. Turn the output on. The nominal
settings above will be output.
Page 60
3. Press Shift to arm the trigger.
Press TRIG[F4] to generate the trigger manually via the front panel.
Shift
Or, Alternatively, pulse the Trigger In pin on the J1 connector high to generate the trigger.
OR Page 127
Note
The TRIG soft-key is only available when the surge/dip control is set to
Manual
.
63
APS-7000 Series User Manual
Example
The following example shows a typical surge/dip output waveform.
Trigger = 0º
ACV
T1 T2100mS
64
OPERATION
Parameter Settings
Time
The Time setting allows you to set the ramp time as ms/Vrms.
Tup = ramp up time/1Vrms Tdn = ramp down time/1Vrms
1Vrms 1Vrms
Tup Tdn
Voltage
The Voltage setting allows you to set the ramp time as Vrms/ms.
Vup = voltage up/1ms Vdn = voltage down/1ms
Vup Vdn
1ms 1ms
Steps
1. Press the Menu key. The Menu
setting will appear on the display.
Menu
2. Use the scroll wheel to go to item 3, Ramp
Control and press Enter.
Ramp Control
The Ramp Control function controls how fast the voltage level ramps up and down. This function allows you to ramp the voltage as a unit of time or as a unit of voltage.
65
APS-7000 Series User Manual
3. Go to the Mode setting using the scroll wheel
and press Enter. Select either Voltage or Time and press Enter again to confirm.
Mode
Voltage, Time
4. For Voltage mode, set Vup and Vdn.
Vup
0.01 ~ 99.99Vrms
Vdn
0.01 ~ 99.99Vrms
5. For Time mode, set Tup and Tdn.
Tup
0.1 ~ 999.9ms
Tdn
0.1 ~ 999.9ms
Exit
6. Press Exit[F4] to exit from the
Ramp Control settings.
7. Upon exiting the ramp icon will
appear in the status bar.
66
OPERATION
Example
Settings: Mode=Time, Tup=1 msec, VAC=100V, Freq=50Hz, Ramp output=on.
Ramp up Tup=1msec
Example:
Settings: Mode=Voltage, Vdn=2Vrms, VAC=100V, Freq=50Hz, Ramp output=off.
Ramp down Vdn =2Vrms
67
APS-7000 Series User Manual
Output on
Ipeak Hold time
I
Begin the peak current
measurement
T
Steps
1. Press the Menu key. The Menu
setting will appear on the display.
Menu
2. Use the scroll wheel to go to item 4, MISC
Configuration and press Enter.
3. Go to the T Ipeak, hold(msec) setting using the
scroll wheel and press Enter. Set the time and press Enter again to confirm.
T Ipeak
1 ~ 60,000 ms
Miscellaneous
The Miscellaneous menu contains miscellaneous parameter settings.
T Ipeak, hold → from page 68 Power ON Output → from page 70 Buzzer → from page 71 SCPI Emulation → from page 72 Program Timer Unit → from page 73
T Ipeak, hold
The T Ipeak, hold function sets the hold time for the peak current measurement. After the output is turned on, the APS-7000 will delay starting the peak current measurement by this hold time.
68
OPERATION
Exit
4. Press Exit[F4] to exit from the
MISC Configuration settings.
Example
Settings
69
APS-7000 Series User Manual
Steps
1. Press the Menu key. The Menu
setting will appear on the display.
Menu
2. Use the scroll wheel to go to item 4, MISC
Configuration and press Enter.
3. Go to the Power ON Output setting using the
scroll wheel and press Enter. Select a setting and press Enter to confirm.
ON
Set Output ON.
OFF
Set Output OFF.
SEQ
Execute the sequence that was loaded before the unit was last turned off.
SIM
Execute the simulation that was loaded before the unit was last turned off.
PROG
Execute the program that was loaded before the unit was last turned off.
Exit
4. Press Exit[F4] to exit from the
MISC Configuration settings.
Power ON Output
The Power ON Output setting allows you to have the output turn on automatically after startup. The settings that are loaded are the last settings that were present in the standard mode before the unit was turned off last.
70
OPERATION
Example
Settings
Steps
1. Press the Menu key. The Menu
setting will appear on the display.
Menu
2. Use the scroll wheel to go to item 4, MISC
Configuration and press Enter.
3. Go to the Buzzer setting using the scroll wheel
and press Enter. Turn the setting on or off and press Enter again to confirm.
Buzzer
ON, OFF
Exit
4. Press Exit[F4] to exit from the
MISC Configuration settings.
Example
Settings
Buzzer
The Buzzer setting turns the buzzer sound on or off for key presses and alarms.
71
APS-7000 Series User Manual
Steps
1. Press the Menu key. The Menu
setting will appear on the display.
Menu
2. Use the scroll wheel to go to item 4, MISC
Configuration and press Enter.
3. Go to the SCPI Emulation setting using the scroll
wheel and press Enter. Choose the SCPI emulation mode and press Enter again to confirm.
SCPI Emulation
GW, EXTECH
Exit
4. Press Exit[F4] to exit from the
MISC Configuration settings.
Example
Settings
SCPI Emulation
The SCPI Emulation setting supports both GW Instek as well as EXTECH SCPI remote commands. When set to EXTECH, the SCPI Emulation setting allows the APS-7000 to be a drop-in replacement for the EXTECH 6700 series by emulating the remote commands of the EXTECH units.
72
OPERATION
Steps
1. Press the Menu key. The Menu
setting will appear on the display.
Menu
2. Use the scroll wheel to go to item 4, MISC
Configuration and press Enter.
3. Go to the Program Timer Unit setting using the
scroll wheel and press Enter. Select the unit and press Enter to confirm.
Program Timer Unit
Hour, Min, Sec
Exit
4. Press Exit[F4] to exit from the
MISC Configuration settings.
Example
Settings
Program Timer Unit
The Program Timer Unit setting is used to configure the timer unit for the Program mode.
73
APS-7000 Series User Manual
Pin Assignment
Trigger Out
Trigger In Out On / Off
COM
J1
Note
The following trigger control settings are not applicable when in the Simulation test mode or the Sequence test mode.
Steps
1. Press Shift + 2 to access the Trigger
Control menu.
Shift
+
2
Trigger
2. The trigger control settings appear. The settings
are divided into Input Pin and Output Pin.
Trigger
The trigger settings are used to turn the output on or off or to set the behavior of the Trigger In and Trigger Out pins of the J1 Connector. The trigger can also be used in conjunction with the Sequence mode, Simulation mode and Surge/Dip mode. See page 127 for further details.
Trigger Control Settings
The trigger control menu sets both the trigger in and the trigger out behavior of the J1 connector.
74
OPERATION
Trigger input
settings
Exit
Trigger output
settings
Trigger Input Pin
1. The Input Pin settings configure what action is
taken when the Trigger In pin on the J1 connector is high.
2. Go to the Mode setting using the scroll wheel
and press Enter. Choose Auto or Manual and then press Enter again to confirm.
Auto
Auto mode will trigger the system when the signal input into the Trigger In pin goes high (+5V).
Manual
The Manual mode will ignore any signal input into the Trigger In pin, but will trigger the system by remote commands
[1]
.
3. Go to the Action setting using the scroll wheel
and press Enter. Choose what will happen when the Trigger In pin is set high(+5V) and then press Enter to confirm.
None
No action is taken.
Output
Turns the output on or off when triggered.
Setting
Sets a user-defined voltage and frequency setting when triggered.
Preset
Loads a preset setting when triggered.
SurgeDip
Triggers the surge/dip control.
75
APS-7000 Series User Manual
4. If Setting was chosen, Set Vset and Fset.
Vset
0 ~ full scale voltage (dependant on the chosen range; 600V optional)
Fset
45.00 ~ 500.0 Hz (999.9 Hz optional)
5. If Preset was chosen, choose which preset
setting should be loaded when the Trigger Input is set to high(+5V).
Preset
0 ~ 9
Trigger Output Pin
1. Go to the Mode setting using the scroll wheel
and press Enter. Choose Auto or Off and then press Enter.
Auto
Auto mode will set the Trigger Out pin to high (+5V) when the conditions are true.
Off
Off mode will set the Trigger Out pin to low (0V).
2. Go to the Source setting to choose what function
or test mode will be able to the trigger the output pin on.
None
No source is able to use the trigger output.
Output
The trigger output goes high (+5V) when the Output is turned on or off.
Setting
When the settings change the trigger output goes high (+5V).
Preset
When a preset is loaded or saved, the trigger output goes high (+5V).
All
The trigger output goes high (+5V) when any of the settings, presets or output are changed.
76
OPERATION
3. Go to Width to set how long the trigger output
is set high for.
Width
0.1 ~ 100.0ms
Exit
4. Press Exit[F4] to exit from the
Trigger Control settings.
[1] Remote Control Commands
The following remote control commands are applicable for the trigger input or output. See the programming manual for usage details.
*TRG :INITiate[:IMMediate]:NAME :INITiate[:IMMediate][:TRANsient] :MEMory:SAV :MEMory:RCL :MEMory:TRIGgered :MEMory:TRIGgered? :OUTPut[:STATe]:TRIGgered :OUTPut[:STATe]:TRIGgered? :SYSTem:CONFigure:TRIGger:INPut:MODE :SYSTem:CONFigure:TRIGger:INPut:MODE? :SYSTem:CONFigure:TRIGger:INPut:SOURce :SYSTem:CONFigure:TRIGger:INPut:SOURce? :SYSTem:CONFigure:TRIGger:OUTPut:MODE :SYSTem:CONFigure:TRIGger:OUTPut:MODE? :SYSTem:CONFigure:TRIGger:OUTPut:SOURce :TRIGger:OUTPut:SOURce :TRIGger:OUTPut:SOURce? :TRIGger:OUTPut[:IMMediate] :TRIGger:MEMory:SOURce :TRIGger:MEMory:SOURce? :TRIGger:MEMory[:IMMediate] :TRIGger[:TRANsient]:SOURce :TRIGger[:TRANsient]:SOURce? :TRIGger[:TRANsient][:IMMediate] [:SOURce]:FREQuency:TRIGgered [:SOURce]:FREQuency:TRIGgered? [:SOURce]:VOLTage[:LEVel]:TRIGgered[:AMPLitude] [:SOURce]:VOLTage[:LEVel]:TRIGgered[:AMPLitude]?
77
APS-7000 Series User Manual
Steps
1. Press Preset and then hold a
number key (0~9) to save the
present settings to the corresponding memory number.
Preset
+
0
~
9
(hold) Presets
M0 ~ M9
2. Press the Preset key again to exit from the preset
mode.
Example
For example, pressing Preset & holding 1 will save the present settings to memory slot 1 (saved to M1).
Note
The preset key will become green when active. A beep will be heard (Buzzer set to ON) and a message will displayed when the settings are saved.
Preset Settings
Save Presets to Local Memory → from page 78 Recall Presets to Local Memory → from page 79 Manage Preset Settings → from page 80
Save Preset Settings to Local Memory
Up to 10 preset settings can be saved to internal memory.
78
OPERATION
Steps
1. Press Preset and press a
number key(0~9) to load the
corresponding memory number.
Preset
+
0
~
9
Presets
M0 ~ M9
2. Press the Preset key again to exit from the preset
mode.
Example
For example, pressing Preset + 1 will recall the saved settings from memory slot 1 (recalled from M1).
Note
The preset key will become green when active. A beep will be heard and a message will be displayed when the settings are recalled.
Load Preset Settings to Local Memory
Any of the 10 preset settings can be recalled from internal memory.
79
APS-7000 Series User Manual
File Format
When files are saved to USB they are saved in the following format: presetX.set, where X is the memory number M0 ~ M9. The files are saved to USB:/gw.
When files are recalled from USB, files must be recalled from the same memory number. For example, the file preset0.set can only be recalled to memory number M0. The files can only be recalled from the USB:/gw directory.
Steps
1. Press the Menu key. The Menu
settings will appear on the display.
Menu
2. Use the scroll wheel to go to item 12, Save/Recall
Files and press Enter.
3. Go to the Type setting using the scroll wheel
and press Enter. Select Preset and press Enter to confirm.
4. Go to the Action setting and choose the file
operation and then press Enter.
MEMUSB
Saves the selected preset memory from the local memory to a USB flash drive.
MEMUSB
Loads a preset memory from a USB flash drive to the selected local memory.
DELETE(MEM)
Deletes the selected preset memory from local memory.
Manage Preset Settings
Preset settings can be easily saved to or from a USB flash drive using the Save/Recall Files utility in the Menu system. Files can also be deleted from local memory using the utility.
80
OPERATION
5. Go to the Memory No. setting and select the
preset memory number to perform the operation on. Press Enter to confirm.
Memory No.
0 ~ 9 (M0 ~ M9)
Execute File Operation
6. Press Exe[F1] to perform the file
operation.
Exit
7. Press Exit[F4] to exit from the
Save/Recall Files settings.
Example
Settings
81
APS-7000 Series User Manual
Note
The output will always be turned off if the ARB waveform is changed. It is not possible to keep the output on whilst the type of ARB waveform is changed.
Arbitrary Waveform Function
The ARB function is able to select a large range of continuous arbitrary waveforms. There are a number of basic waveform shapes to choose from, each of which can be customized to output a waveform with your desired attributes.
The ARB function can be used in conjunction with the Simulate and Sequences modes, but cannot be used with the Program test modes. When the WAVE parameter is set to any waveform other than SINE, ARB will be shown in the status bar when in the BASIC operation, Simulate mode or Sequence mode to indicate that the ARB function is active when in these modes.
ARB Function Overview → from page 83 Selecting an ARB Waveform → from page 86
82
OPERATION
Background
The ARB mode function has 7 basic waveforms and a number of settable attributes, depending on the chosen waveform.
Setting Screen Overview
Waveform type
Attributes for the selected waveform
Visual representation of the waveform shape and its attributes
F4 Exit
ARB Waveform Overview
The following describes each of the 7 basic waveforms.
Sine
This is the default waveform used for the APS-7000. No settable attributes.
Triangle
Basic triangle waveform. No settable attributes.
ARB Mode Overview
83
APS-7000 Series User Manual
Staircase
The staircase waveform has a settable number of step levels.
Attributes: Stairs: 1 ~ 100
Clip
Outputs a clipped sinewave. The degree to which the sine wave is clipped is settable.
Attributes: n*ACV: 1.0 ~ 10.0 (clip level)
CF
Crest factor waveform. The crest factor is settable.
Attributes: CF: 1.4 ~ 10.0
84
OPERATION
Surge
The surge waveform has a settable ACV base level, site size and site shape.
Attributes: Type: Sine, Square (site waveform type) Acv: 0 ~ 100% (base waveform ampl.) Site: 0 ~ 100% (site waveform width)
ARB
The ARB waveform shape function simply adds a number Fourier series terms to create an arbitrary waveform.
Attributes: Type: 1 ~ 22 (Number of selectable ARB waveforms)
The numerical figures shown under the Type parameter shows which Fourier series terms are used and their (bn) coefficients.
85
APS-7000 Series User Manual
Entering the ARB Menu
1. Press Shift + 1 to access the ARB
menu.
Shift
+
1
ARB
2. Set the Wave parameter with the scroll wheel
and confirm the setting with the Enter key.
Wave
Sine, Triangle, Stair, Clip, CF, Surge, ARB
3. Use the scroll wheel and Enter key to select any
other attributes for the selected waveform and press Enter to confirm any settings.
Other Attributes:
Sine Wave:
None
Triangle Wave:
None
Stair:
Stairs: 1 ~ 100
Clip:
N*ACV: 1 ~ 10
CF:
CF: 1.4 ~ 10.0
Surge:
Type: Sine, Square ACV: 0 ~ 100% Site: 0 ~ 100%
Exit
4. Press Exit[F4] to exit from the ARB
settings.
Selecting an ARB Waveform
86
OPERATION
Common Settings
5. Press Menu to go back to the main
menu.
Menu
6. Set the voltage, frequency, Irms
level and so on as you normally would for the standard ACV output. See the Basic Operation chapter for setting details.
Page 43
Note
When the WAVE parameter is set to any waveform other than SINE, ARB will be shown on the display when in the BASIC operation, Simulate mode or Sequence mode to indicate that the ARB function is active when in these modes.
Standard Mode Example
Simple Mode Example
Sequence Example
87
APS-7000 Series User Manual
Simulate Example
88
OPERATION
Test Mode Function
There are three test modes.
Simulate Mode: voltage/frequency fluctuations.
The Simulate function is a quick and easy way to simulate fluctuations in voltage, frequency, phase and other anomalies in mains supply power.
Sequence Mode: creates test sequences.
The Sequence function creates arbitrary AC waveforms based on user defined parameters.
Program Mode: Runs a group of test sequences.
The Program function runs different memory steps sequentially. Each step of a program is a preset memory configuration (M01 ~ M50). The Program function can also perform pass/fail testing.
Simulate Mode Overview → from page 91 Simulate Settings → from page 95 Save Simulation to Local Memory→ from page 98 Recall Simulation from Local Memory→ from page 98 Manage Simulation Settings → from page 99 Run Simulation → from page 101
Sequence Mode Overview → from page 103 Sequence Settings → from page 107 Save Sequence to Local Memory → from page 111 Recall Sequence from Local Memory → from page 111 Manage Sequence Settings → from page 112 Run Sequence → from page 114
89
APS-7000 Series User Manual
Program Mode Overview → from page 116 Save a Program to Local Memory → from page 123 Recall a Program from Local Memory → from page 123 Manage Program Settings → from page 124
90
OPERATION
Background
The Simulate function is used to test power supply fluctuation. This function is able to simulate common abnormalities in mains power such as fluctuations in voltage, phase and frequency. These simulations can be run as one-off anomalies or cyclic anomalies.
Setting Screen Overview
Number of repetitions
Test mode
V RANGE
Recall test
Save test Run test
ON/OFF phase Trigger output
level
Step Frequency Step voltage Step time Step
Step Overview
The Simulate function is comprised of 6 steps. Each step is run sequentially in the following order: Initial, Normal1, Trans1, Abnormal, Trans2, Normal2, Initial.
Initial
The Initial step is used as the initial and final settings of the waveform simulation. This is the standby step before the test starts and the standby step after the test ends.
Normal1
This step configures the normal output conditions that precede the abnormal conditions.
Simulate Mode Overview
91
APS-7000 Series User Manual
Trans1
This step configures the transition from normal to abnormal conditions. This step will linearly interpolate the normal settings to the abnormal settings. This step can be skipped for abrupt state changes.
Abnormal
This step contains the abnormal conditions for the simulation.
Trans2
This step configures the transition from abnormal to normal conditions.
Normal2
This step configures the normal conditions that supersede the abnormal conditions.
AbnormalTrans1Normal1Init Trans2 Normal2 Init
Parameter Overview
The following table shows which parameters are available for each step.
Step\Parameter
Initial
Normal1
Trans1
Abnormal
Trans2
Normal2
Repeat
✓ ✓ ✓ ✓ ✓
Time
✓ ✓ ✓ ✓ ✓
ON Phs
✓ X ✓ X ✓
OFF Phs
✓ X ✓ X ✓
Vset
✓ X ✓ X ✓
Fset
✓ X ✓ X ✓
Trig Out
✓ ✓ ✓ ✓ ✓
92
OPERATION
Repeat
Indicates the number of times the simulation will be run, from Normal1 to Normal2.
A value of 0 indicates infinite repeats. The repeat setting is the same for each step.
Time
Sets the duration time of the step. When the ON Phs=ON, the total duration of the step is equal to the Time setting + ON Phs=ON duration.
ON Phs
Sets the starting phase of the waveform for the step. Not applicable for the Trans 1/2 steps.
OFF Phs
Sets the off phase of the waveform after the output has been turned off. Not applicable for the Trans 1/2 steps.
Fset
Sets the frequency of the step. Not applicable for the Trans 1/2 steps.
Vset
Sets the voltage of the step. Not applicable for the Trans 1/2 steps.
Trig Out
Sets the polarity of the trigger out port for the duration of the step. The trigger out signal is sent at the start of each step (including the first initial step).
93
APS-7000 Series User Manual
The following diagram illustrates the relationship between each of the parameters in a step.
Time
ON Phs
ON Phs
additional time
Output off
OFF Phs
Trigger out
position
Fset
OFF Phs
additional time
Vset
94
OPERATION
Entering the Simulate Menu
1. Press Test.
Test
2. Press Mode[F1] and use the scroll wheel to
select SIMULATE and then press Enter.
Simulate Mode
F1 Mode
Steps
3. Use the scroll wheel to go to the Step setting
and press Enter.
4. Use the scroll wheel to select one of the
simulation steps and press Enter.
Steps
Initial, Normal1, Trans1, Abnormal, Trans2, Normal2
5. Go to the Time setting and set the duration of
the step.
Time
0.01 ~ 99.99s, 0(Trans 1 and Trans2) Note: For Trans1 and Trans2, it supports a value of 0, which will skip the step.
Simulate Settings
95
APS-7000 Series User Manual
6. Go to the ON Phs setting and set the starting
phase of the step. Not applicable for Trans1 or Trans2.
ON Phase
ON, OFF
ON Phase
0~ 359º
Resolution
7. Go to the OFF Phs setting and set the end phase
of the step. Not applicable for Trans1 or Trans2.
OFF Phase
ON, OFF
OFF Phase
0~ 359º
Resolution
8. Press the Range key repeatedly to set the
voltage range for the Vset parameter. The range will be shown in the top corner, which indicates that the simulation will be executed within this voltage range.
Range
Range
LO(155V), MD(310V), HI(600V, optional)
9. Go to the Vset setting and set the Vrms level of
the step. If you input a Vset value that is not within the voltage range, the input value will be ignored. Not applicable for Trans1 and Trans2.
Vset
0.00 ~ 310.0Vrms (range dependent) (600V optional)
96
OPERATION
10. Go to the Fset setting set the frequency of the
step. Not applicable for Trans1 and Trans2.
Fset
45.00 ~ 500.0Hz (999.9Hz optional)
11. Go to the Trig Out setting and set the state of
the trigger out pin at the start of each step.
Trig Out
HI, LO
12. Lastly, go to the Repeat parameter select the
number of times the simulation will repeat the Normal1-Trans1-Abnormal-Trans2-Normal2 sequence of steps. A value of 0 will set the number of repetitions to infinite.
Repeat
1 ~ 9999, 0(infinite)
97
APS-7000 Series User Manual
Saving a Simulation
Simulation settings can be saved to one of 10 memory slots (SIM0 ~ SIM9).
Steps
1. Press Save[F3] and then long press a number key
when prompted.
2. A message will appear when the save is
successful.
Save
SIM0 ~ SIM9
Recall a Simulation
Simulation settings can be recalled from one of 10 memory slots (SIM0 ~ SIM9).
Steps
1. Press Recall[F2] and then press a number key
when prompted.
2. A message will appear when the settings are
recalled successfully.
Recall
SIM0 ~ SIM9
Save a Simulation to Local Memory
Recall a Simulation from Local Memory
98
OPERATION
File Format
When files are saved to USB they are saved in the following format: simX.sim, where X is the memory number 0 ~ 9 (SIM0 ~ SIM9). The files are saved to USB:/gw.
When files are recalled from USB, files must be recalled from the same memory number. For example, the file sim0.sim can only be recalled to memory number SIM0. The files can only be recalled from the USB:/gw directory.
Steps
1. Press the Menu key. The Menu
settings will appear on the display.
Menu
2. Use the scroll wheel to go to item 12, Save/Recall
Files and press Enter.
3. Go to the Type setting using the scroll wheel
and press Enter. Select SIMULATE and press Enter to confirm.
4. Go to the Action setting and choose the file
operation and then press Enter.
MEMUSB
Saves the selected simulation memory from the local memory to a USB flash drive.
MEMUSB
Loads the simulation memory from a USB flash drive to the selected local memory.
Manage Simulation Settings
Simulation settings can be easily saved to or from a USB flash drive using the Save/Recall Files utility in the Menu system. Files can also be deleted from local memory using the utility.
99
APS-7000 Series User Manual
DELETE(MEM)
Deletes the selected simulation memory from local memory.
5. Go to the Memory No. setting and select the
simulation memory number to perform the operation on. Press Enter to confirm.
Memory No.
0 ~ 9 (SIM0 ~ SIM9)
Execute File Operation
6. Press Exe[F1] to perform the file
operation.
Exit
7. Press Exit[F4] to exit from the
Save/Recall Files settings.
Example
Settings
100
Loading...