This manual contains proprietary information, which is protected by
copyright. All rights are reserved. No part of this manual may be
photocopied, reproduced or translated to another language without
prior written consent of Good Will company.
The information in this manual was correct at the time of printing.
However, Good Will continues to improve products and reserves the
rights to change specification, equipment, and maintenance
procedures at any time without notice.
Good Will Instrument Co., Ltd.
No. 7-1, Jhongsing Rd., Tucheng Dist., New Taipei City 236, Taiwan.
Warning: Identifies conditions or practices that
could result in injury or loss of life.
CAUTION
Caution: Identifies conditions or practices that
could result in damage to the APS-7000 or to other
properties.
DANGER High Voltage
Attention Refer to the Manual
Protective Conductor Terminal
Earth (ground) Terminal
SAFETY INSTRUCTIONS
This chapter contains important safety
instructions that you must follow during
operation and storage. Read the following before
any operation to ensure your safety and to keep
the instrument in the best possible condition.
Safety Symbols
These safety symbols may appear in this manual or on the
instrument.
4
SAFETY INSTRUCTIONS
Do not dispose electronic equipment as unsorted
municipal waste. Please use a separate collection
facility or contact the supplier from which this
instrument was purchased.
General
Guideline
CAUTION
Do not place any heavy object on the APS-7000.
Avoid severe impact or rough handling that
leads to damaging the APS-7000.
Do not discharge static electricity to the APS-
7000.
Use only mating connectors, not bare wires, for
the terminals.
Do not block the cooling fan opening.
Do not disassemble the APS-7000 unless you are
qualified.
(Measurement categories) EN 61010-1:2010 specifies the
measurement categories and their requirements as follows. The
APS-7000 falls under category I.
Measurement category IV is for measurement performed at the
source of low-voltage installation.
Measurement category III is for measurement performed in the
building installation.
Measurement category II is for measurement performed on the
circuits directly connected to the low voltage installation.
Measurement category I is for measurements performed on
circuits not directly connected to Mains.
Power Supply
WARNING
AC Input voltage range: 115/230 Vac ± 15%
Frequency: 47~63Hz
To avoid electrical shock connect the protective
grounding conductor of the AC power cord to
an earth ground.
Safety Guidelines
5
APS-7000 Series User Manual
Cleaning the APS7000
Disconnect the power cord before cleaning.
Use a soft cloth dampened in a solution of mild
detergent and water. Do not spray any liquid.
Do not use chemicals containing harsh material
such as benzene, toluene, xylene, and acetone.
Operation
Environment
Location: Indoor, no direct sunlight, dust free,
almost non-conductive pollution (Note below)
Relative Humidity: 20%~ 80%, no condensation
Altitude: < 2000m
Temperature: 0°C to 40°C
(Pollution Degree) EN 61010-1:2010 specifies the pollution degrees
and their requirements as follows. The APS-7000 falls under degree
2.
Pollution refers to “addition of foreign matter, solid, liquid, or
gaseous (ionized gases), that may produce a reduction of dielectric
strength or surface resistivity”.
Pollution degree 1: No pollution or only dry, non-conductive
pollution occurs. The pollution has no influence.
Pollution degree 2: Normally only non-conductive pollution
occurs. Occasionally, however, a temporary conductivity caused
by condensation must be expected.
Pollution degree 3: Conductive pollution occurs, or dry, non-
conductive pollution occurs which becomes conductive due to
condensation which is expected. In such conditions, equipment
is normally protected against exposure to direct sunlight,
precipitation, and full wind pressure, but neither temperature
nor humidity is controlled.
Storage
environment
Location: Indoor
Temperature: -10°C to 70°C
Relative Humidity: ≤80%, no condensation
Disposal
Do not dispose this instrument as unsorted
municipal waste. Please use a separate collection
facility or contact the supplier from which this
instrument was purchased. Please make sure
discarded electrical waste is properly recycled to
reduce environmental impact.
6
SAFETY INSTRUCTIONS
Green/ Yellow:
Earth
Blue:
Neutral
Brown:
Live (Phase)
Power cord for the United Kingdom
When using the instrument in the United Kingdom, make sure the
power cord meets the following safety instructions.
NOTE: This lead/appliance must only be wired by competent persons
WARNING: THIS APPLIANCE MUST BE EARTHED
IMPORTANT: The wires in this lead are coloured in accordance with the
following code:
As the colours of the wires in main leads may not correspond with
the coloured marking identified in your plug/appliance, proceed
as follows:
The wire which is coloured Green & Yellow must be connected to
the Earth terminal marked with either the letter E, the earth symbol
or coloured Green/Green & Yellow.
The wire which is coloured Blue must be connected to the terminal
which is marked with the letter N or coloured Blue or Black.
The wire which is coloured Brown must be connected to the
terminal marked with the letter L or P or coloured Brown or Red.
If in doubt, consult the instructions provided with the equipment
or contact the supplier.
This cable/appliance should be protected by a suitably rated and
approved HBC mains fuse: refer to the rating information on the
equipment and/or user instructions for details. As a guide, a cable
of 0.75mm2 should be protected by a 3A or 5A fuse. Larger
conductors would normally require 13A types, depending on the
connection method used.
Any exposed wiring from a cable, plug or connection that is
engaged in a live socket is extremely hazardous. If a cable or plug is
deemed hazardous, turn off the mains power and remove the cable,
any fuses and fuse assemblies. All hazardous wiring must be
immediately destroyed and replaced in accordance to the above
standard.
7
APS-7000 Series User Manual
APS-7000 Series Overview ................................................. 9
Series lineup ................................................................................................................ 9
Operating Area ........................................................................................................... 9
Main Features ........................................................................................................... 10
This chapter describes the power source in a
nutshell, including its main features and front /
rear panel introduction.
8
GETTING STARTED
Model name
Max. Output Current
Power Rating
Output Voltage
APS-7050
4.2A/2.1A
500VA
0~310.0 Vrms
APS-7100
8.4A/4.2A
1000VA
0~310.0 Vrms
Voltage (V)
Current (A)
0
1
2
0
3
4
5
100
200
300
400500
600
120 155240
310480
600
4.2
3.23
2.1
1.62
1.05
0.83
APS-7050 Output Operating Area
155V Range310V Range600V Range
(APS-003 Option)
APS-7000 Series Overview
Series lineup
The APS-7000 series consists of 2 models, the APS-7050 and the
APS-7100, differing only in capacity. Note that throughout the user
manual, the term “APS-7000” refers to both the APS-7050 and APS7100, unless stated otherwise.
Operating Area
9
Voltage (V)
Current (A)
0
2
4
0
6
8
10
100
200
300
400500
600
120 155240
310480
600
8.4
6.45
4.2
3.23
2.1
1.67
155V Range310V Range600V Range
APS-7100 Output Operating Area
(APS-003 Option )
Main Features
Performance
Low output ripple and noise
Excellent and feature-rich measurement
capacity
Standard maximum output voltage is 310Vrms
Maximum output voltage and frequency of
600Vrms(APS-003 Option)/999.9Hz(APS-004
Option)
APS-7000 Series User Manual
10
Features
OVP, OCP and OTP protection
Variable voltage, frequency and current limiter
Sequence and simulation function
Large 4.3 inch TFT panel
Globally adjustable power inlet not restricted by
the power supply environment
USB interface is equipped as standard with the
ability to save and recall setup files.
Only 88mm (2U) case height.
Interface
Ethernet port
USB host
RS-232 / USB CDC (optional)
GPIB (optional)
Accessories
Standard
Accessories
Part number
Description
CD ROM
User manual,
programming manual
4322-30700101 (UL/CSA)
Type I Power cord
(APS-7050)
4322-B0700101 (PSE)
Type II Power cord
(APS-7050)
4300-31000101 (UL/CSA)
Type I Power cord
(APS-7100)
4300-B1000201 (PSE)
Type II Power cord
(APS-7100)
62PS-7K0SC401 x1
5302-01613001 x1
Mains terminal cover set
(APS-7050)
GETTING STARTED
11
APS-7000 Series User Manual
62PS-7K0SC701 x1
5302-01613001 x2
Mains terminal cover set
(APS-7100)
GTL-123
Test leads: 1x red, 1x black
Optional
Capacity
Part number
Description
APS-003
Output Voltage Capacity:
0 ~ 600Vrms
APS-004
Output Frequency Capacity:
45 ~ 999.9Hz
Optional
Accessories
Part number
Description
GRA-423
APS-7000 rack mount kit
APS-001
GPIB interface card
APS-002
RS-232 / USB interface
card
Download
Name
Description
gw_aps.inf
USB driver
12
GETTING STARTED
APS-7050
AC Power Source
Display
Programmable
F 1
F 2
F 3
F 4
Output
: Long Push
MenuTestPreset
V
Local
Off Phase
On Phase
0
1
4
7
2
5
8
3
6
9
V-Limit F-Limit IPK-Limit
FI rms
Cancel
Shift
Range
Enter
Lock
ALM CLR
IPK CLR
Unlock
Surge / Dip
Ramp
TriggerARB
USB A port
Power
switch
Front voltage
output socket
Menu, Test,
Preset keys
LCD
display
Function keys
Display
mode key
Arrow keys,
Scroll wheel
Voltage, Frequency,
Current setting and limit
Keypad
Range, Shift, Cancel,
Enter, Lock keys
Output key
Item
Description
Power Switch
Turns on the mains power.
Front Voltage
Output Socket
Neutral
GND
Line
Output voltage
terminal using a
regional universal
plug. There is a Euro
and a Universal
regional plug.
CAUTION
Maximum allowable output voltage
and current are 250Vrms and
10Arms.
For voltages exceeding 250Vrms,
please use the rear output terminal.
Appearance
Front Panel
APS-7050, APS-7100
13
APS-7000 Series User Manual
USB A Port
The USB port is used for data
transfers and upgrading software.
LCD Screen
Displays the measured values or
menu system.
Display Mode
Select Key
Display
Selects between Standard mode
and Simple mode.
Function Keys
F 1
Assigned to the functions
displayed on the right-hand side
of the screen.
Menu Key
Menu
Enters the Main menu or goes
back to one of the display modes.
Test Key
Test
Puts the instrument into the
Sequence, Simulation and
Program Control mode.
Preset Key
Preset
Puts the instrument into Preset
mode.
Arrow Keys
The arrow keys are used to select
the digit power of a value that is
being edited.
V
V
V-Limit
Used for setting the output
voltage.
V-Limit
(Shift + V)
Used for setting the output voltage
limit value.
F
F
F-Limit
Used for setting the output
frequency.
F-Limit
(Shift + F)
Used for setting the output
frequency limit value.
I rms
I rms
IPK-Limit
Used for setting the maximum
output current.
IPK-Limit
(Shift + I rms)
Used to set the peak output
current limit value.
14
GETTING STARTED
Range Key
Range
Switches between the 155V, 310V
and 600V ranges (the 600V range
is an option).
Scroll Wheel
Used to navigate menu items or
for incrementing/decrementing
values one step at a time.
Lock Key
Lock
Unlock
Locks the number pad to prevent
accidentally changing panel
settings.
Unlock Key
(Long press)
Disables the key lock.
Enter Key
Enter
Confirms selections / settings
Cancel Key
Cancel
Clears entries that are made in the
number entry dialog when a value
is edited using the arrow keys or
the scroll wheel.
The Cancel key can also be used to
cancel function setting menus or
dialogs.
Shift Key
Shift
Turns on the shift state, which
enables shortcut operations.
Output Key
Output
Turns the output on or off.
Number Pad
0
142
5
8
3
6
9
Lock
Unlock
7
On PhaseSurge / Dip IPK CLR
Off PhaseRampALM CLR
ARBTrigger
Local
Used to enter values.
Local Mode
(Shift + 0)
Switches operation back to local
mode from remote mode.
ARB Mode
(Shift + 1)
Sets the ARB function.
Trigger Mode
(Shift + 2)
Sets the JI port trigger behavior on
the rear panel.
15
APS-7000 Series User Manual
Off Phase
(Shift + 4)
Sets the off phase for the output
voltage.
RAMP
(Shift + 5)
Quick settings for Ramp control.
ALM CLR
(Shift + 6)
Clears alarms.
On Phase
(Shift + 7)
Sets the on phase for the output
voltage.
Surge/Dip
(Shift + 8)
Quick settings for Surge/Dip
control.
IPK CLR
(Shift + 9)
Clears peak current protection.
16
Rear Panel
SYNC
Signal OutputLAN
OUTPUTINPUT
L NN L
600Vac MAX.115 / 230V 15%
50 / 60Hz
3.6kVA MAX.
Out On / Off
COM
J1
Trigger Out
Trigger In
115 / 230V 15%
50 / 60Hz
1.8kVA MAX.
SYNC
SER. NO. LB
L N
600Vac MAX.
Signal Output
Trigger Out
Trigger In Out On / Off
COM
LAN
J1
OUTPUT
Fan
GPIB
Line voltage input
GPIB
RS232
RS-232/
USB Device
Sync
Signal OutputRear panel voltage
output terminal
Remote controlEthernet port
Line voltage input
APS-7050
APS-7100
Opt. int. port
Line Voltage
Input
APS-7050
115 / 230V 15%
50 / 60Hz
1.8kVA MAX.
Voltage Input: 115/230±15% VAC;
Line frequency: 50Hz/60 Hz
(Automatically switchable)
GETTING STARTED
17
APS-7000 Series User Manual
APS-7100
OUTPUTINPUT
LNN L
Voltage Input: 115/230±15%
VAC ; Line frequency:
50Hz/60 Hz (Automatically
switchable)
Rear Voltage
Output Socket
Output voltage terminal.
APS-7050
LN
OUTPUT
APS-7100
OUTPUTINPUT
LNN L
Sync Output
Socket
SYNC
BNC socket. This socket will output
a signal of approximately 10V when
the output is on.
Signal Output
Signal Output
Connector for monitoring PASS,
FAIL and PROCESSING output
signals when using the Program
mode.
Remote Control
Trigger Out
Trigger InOut On / Off
COM
J1
Connector for controlling the
TRIGGER IN, TRIGGER OUT and
OUT ON/OFF states.
Ethernet Port
LAN
The Ethernet port is used for remote
control and digital monitoring from
a PC.
Optional
Interface Slot
Optional GPIB communication, RS-232/USB B
communication.
FAN
Temperature controlled fan.
18
GETTING STARTED
Status bar
/
Indicates if the output is ON or OFF.
Indicates the output power as a percentage of full
scale.
Indicates that the Surge/Dip function is active.
Indicates that the Ramp function is active.
The alarm icon will appear on the status bar
when one of the protection functions are tripped.
Applies to Over Power, Over Irms, Over Ipeak
and Over Temperature protection.
Indicates that the APS-7000 is in remote mode.
Indicates that a USB drive is detected in the front
panel host port.
Indicates that the LAN interface is activated.
Indicates that the panel lock is active.
Status Bar Icons
19
APS-7000 Series User Manual
Set Up ............................................................................. 22
Line Voltage Connection (APS-7100) ................................................................... 22
Power Up .................................................................................................................. 25
Running a Sequence .............................................................................................. 114
Program Mode Overview ..................................................................................... 116
Save a Program to Local Memory ....................................................................... 123
Recall a Program from Local Memory ................................................................ 123
21
APS-7000 Series User Manual
Background
The APS-7100 is equipped with an input power
terminal that can accept 115V/230V ± 15%. To
connect or replace the power cord (GW Instek
part number: APS-7100: 4300-31000101, use the
procedure below:
Warning
The following procedure should only be attempted
by competent persons.
Ensure the AC power cord is not connected to
power.
Removal
1. Turn off the power switch.
1
APS-7100
2. Unscrew the power cord protective sheath on
the rear output socket.
3. Remove the 2 screws holding the power cord
cover and remove.
Set Up
Line Voltage Connection (APS-7100)
22
OPERATION
APS-7100
3
2
4. Remove the AC
power cord wires.
4
CAUTION
The power line inputs for the APS-7100 are on the
outer cluster of terminals.
The terminals to the inner of the panel are the rear
panel outputs.
Input Power Terminal:
Connect the power cord to
the input power
terminals.
Page 22
2. Press the POWER key. The splash screen will
appear momentarily before the continuous
mode screen appears with the settings loaded.
CAUTION
The power supply takes around 10 seconds to fully
turn on and shutdown.
Do not turn the power on and off quickly.
OPERATION
25
APS-7000 Series User Manual
Background
The output terminals can be output from either
the front panel or from the rear panel. The
outputs are limited to 4.2A/2.1A (APS-7050) or
8.4A/4.2A(APS-7100).
Supported plugs
Multi-region terminal Socket
Supported Standards
IEC, North America, Japan.
EURO CEE type universal plug
WARNING
Dangerous voltages. Ensure that the power to the
instrument is disabled before handling the power
supply output terminals. Failing to do so may lead
to electric shock.
CAUTION
For the front panel output, the maximum output
voltage is 250VAC and current is 10A.
Front Panel
Output
Connection
1. The front panel has a multi-region power
socket depending on the socket type.
2. Insert the plug from the DUT into the socket.
EURO CEE socket
IEC North America, Japan
Output Terminals
26
OPERATION
3. Turn the power on. The AC power supply is
now ready to power the DUT.
Rear Panel Output
Connection
The rear panel output is used to supply higher
power DUTs. The rear panel output connection
is similar to the universal rear panel line input
connection on the APS-7100.
1. Disconnect the unit from the mains power
socket and turn the power switch off.
2. Unscrew the power cord protective sheath.
3. Remove the 2 screws holding the power cord
cover and remove.
APS-7050
3
2
APS-7100
3
2
27
APS-7000 Series User Manual
CAUTION
For the APS-7100, there is a single bank for the
input and output terminals. Ensure the correct
terminals are connected. The APS-7050 only has a
single bank of output terminals on the rear panel.
Installation
4. Connect the output AC power cord wires to
the AC output terminals.
Black Neutral (N)
Green GND ()
Black Line (L)
OUTPUTINPUT
LNNL
115 / 230V15%
Line
Neutral
Ground
Ground
Neutral
Line
APS-7100 shown. The input terminals are already
connected and shows which output terminals are
to be connected.
5. Re-install the power cord cover.
6. Screw the power cord sheath back onto the
cover.
28
OPERATION
APS-7100
5
6
APS-7050
5
6
7. Turn the power on. The AC power supply is
now ready to power the DUT.
29
APS-7000 Series User Manual
Background
There are a number of optional modules that
can be installed for remote control or for the
SCPI compatibility modes.
Optional Modules
APS-001
GPIB Interface card
APS-002
RS-232/USB interface card
WARNING
Dangerous voltages. Ensure that the power to the
instrument is disabled before handling the power
supply output terminals. Failing to do so may lead
to electric shock.
CAUTION
Ensure the power is off before installing any of the
optional modules.
Installation
1. Turn off the power switch.
1
2. Unscrew the two screws holding the options
panel plate.
2
Installing the Optional Modules
30
OPERATION
3. Slide the module PCB onto the rails on the
inside of the module slot.
4. Secure the module with the screws that were
removed from step 2.
4
3
5. The module will be recognized upon startup.
31
APS-7000 Series User Manual
Background
The APS-7000 series has an optional Rack
Mount Kit (GW Instek part number: GRA-423).
The APS-7050 and APS-7100 are designed to fit
into a 2U rack height. Please see your
distributor for further rack mount details.
Rack mount
diagram
CAUTION
Ensure adequate ventilation is provided when
using the rack mount. Ensure at a gap of at least
50mm is given for the side air intakes. Failure to
do so may cause the instrument to overheat.
Background
The APS-7000 AC power supplies generally
use the scroll wheel, arrow keys and Enter keys
to edit numerical values or to select menu
options.
Menu navigation is performed using the menu
keys and function keys on the front panel.
The following section will explain some of
these concepts in detail.
Using the Rack Mount Kit
How to Use the Instrument
32
OPERATION
Selecting Menu
Items
1. Turn the scroll wheel to select
parameters in menus and lists. The
selected parameter will be
highlighted in orange. The scroll
wheel is also used to
increment/decrement setting
values.
2. Press the Enter key to edit the
parameter or to enter the selected
menu.
Enter
Example
Selected parameter
The following is an example of the menu list that
appears when the Menu key is pressed.
Using the Keypad
to edit parameter
values
When editing a value the keypad can be used
to directly enter the desired value.
1. Type the value of the
parameter using the
keypad.
0
1
4
2
5
8
3
6
9
Lock
Unlock
7
On PhaseSurge / DipIPK CLR
Off PhaseRampALM CLR
ARBTrigger
Local
2. Press the Enter key to confirm the
edit.
Enter
33
APS-7000 Series User Manual
Example
Parameter
Using the Arrow
Keys and Scroll
wheel to edit
parameter values
Use the arrow keys to select a digit power and
then use the scroll wheel to edit the value by
that power.
1. Use the arrow keys to move the
cursor to the digit of the desired
power.
2. Turn the scroll wheel to edit the
value by the resolution of the
selected digit.
Cursor
3. Repeat the steps above for all the relevant
digits.
4. Press the Enter key to confirm the
edit.
Enter
34
OPERATION
Note
By default the cursor starts at the lowest power
digit.
Using the
onscreen
keyboard
The onscreen keyboard is only used in the
Program Mode. The screenshot below is an
example of the onscreen keyboard.
On screen keyboard
Entered characters
Using the
Function Keys
The function keys are quick settings keys, the
function of which depends on the current
menu or operation.
1. Press the function key that corresponds to the
setting directly to its left.
2. The setting or parameter is immediately
executed.
Function keys
APS-7050
AC Power Source
Display
Programmable
Output
:LongPush
MenuTestPreset
V
Local
Off Phase
On Phase
0
1
4
7
2
5
8
3
6
9
V-LimitF-Limit IPK-Limit
FI rms
Cancel
Shift
Range
Enter
Lock
ALM CLR
IPK CLR
Unlock
Surge / Dip
Ramp
TriggerARB
F 1
F2
F3
F4
Corresponding quick
settings
35
APS-7000 Series User Manual
3. Repeat the steps above for all the relevant
digits.
36
OPERATION
Background
The default settings can be restored from the
Menu key settings. See page 155 for the default
factory settings.
Steps
1. Press the Menu key. The Menu
settings will appear on the display.
Menu
2.Use the scroll wheel to go to item 10, Default
Setting.
3.Press Enter x2 to restore the default settings.
Default settings
Reset to Default Settings
37
APS-7000 Series User Manual
Background
The Menu>System Information setting
displays the serial number and version
number.
Steps
1. Press the Menu key. The Menu
setting will appear on the display.
Menu
2. The system information should now be listed
on the display.
If not, use the scroll wheel to go to item 1,
System Information.
System
Information
View System Version and Build Date
38
OPERATION
Background
The LCD Configuration setting sets the
brightness, contrast and saturation level of the
LCD display.
Steps
1. Press the Menu key. The Menu
settings will appear on the display.
Menu
2.Use the scroll wheel to go to item 9, LCD
Configuration and press Enter.
3. Set the brightness, contrast and saturation.
Contrast(%)
1 ~ 100% (Default=50%)
Brightness(%)
1 ~ 100% (Default=50%)
Saturation(%)
1 ~ 100% (Default=50%)
Exit
4. Press Exit[F4] to exit from the
Ramp Control settings.
Default Settings
5. Press Default[F3] to set all the LCD settings to
50%.
LCD settings
Default
settings
LCD Configuration
39
APS-7000 Series User Manual
Background
If the USB Type B interface is to be used for
remote control, the USB driver needs to be
installed.
Note
The USB driver, GW_APS.inf, is located on the
CD Rom that accompanied this user manual.
Alternatively the driver can be downloaded
from the GW Instek website.
For information on the USB interface, see page
132.
Steps
1. Connect the rear panel USB -B port on the APS-
7000 to the PC using a USB Type A to B cable.
2. Go the Windows Device Manager.
For Windows 7:
Start > Control Panel > Hardware and Sound
> Device Manager
3.The APS-7000 will be located under Other
Devices in the hardware tree. Right-click the
APS-70XX and choose Update Driver Software.
USB Driver Installation
40
OPERATION
4.From the hardware wizard choose Browse my
computer driver software.
5. Set the file path to the location of the USB
driver, click Next and finish the driver
installation.
41
APS-7000 Series User Manual
6. In APS-7000 will now be located in the Ports
node of the hardware tree in the Windows
Device Manager if the driver installation was
successful.
42
OPERATION
Background
The Range setting determines the general outlet
voltage range. The ranges available correspond
to common mains output voltage standards.
Steps
1. Press Range to access the Range
menu.
Range
Basic Operation
This section describes the basic operations required to operate the
power supply.
Setting the Voltage Range → from page 43
Setting the Voltage Limit → from page 44
Setting the Output Voltage → from page 45
Setting the Frequency Limit → page 47
Setting the Output Frequency → page 48
Setting the Peak Current Limit → from page 49
Setting the Current RMS Level → from page 52
Setting the On/Off Phase → from page 55
Clearing the Alarm → from page 56
Setting the Display mode → from page 57
Panel lock → from page 59
Turning the Output on/off → from page 60
Before operating the power supply, please see the Getting Started
chapter, page 8.
Setting the Voltage Range
43
APS-7000 Series User Manual
2. Set the voltage range with the scroll wheel or
with the F1 ~ F4 soft-keys.
Range
AUTO, 600V(option), 310V, 155V
Soft-keys
AUTO, 600V(option), 310V, 155V
3. Press Enter to confirm the Range setting.
Range setting
Soft-key
settings
F1
F2
F3
F4
Note
If the range is changed from 155V to 600V, the
Irms and IPK values will automatically be changed
to a lower value. If the range is changed from 600V
to 155V, the Irms and IPK values remain the same.
If the voltage range is changed when the output is
on, the output will be automatically turned off.
Background
Setting the voltage limit allows the output
voltage to be set to any level within the voltage
limit (V Limit) range.
Steps
1. Press Shift + V to access the Volt
Limit menu.
Shift
+
V
V-Limit
Setting the Voltage Limit
44
OPERATION
2. Set the voltage limit with the scroll wheel or
with the F3 ~ F4 soft-keys. The MAX and MIN
soft-keys set the limit to the maximum and
minimum, respectively.
Range
10% of full range ~ full range
Soft-keys
MAX, MIN
3. Press Enter to confirm the voltage limit setting.
Volt Limit
Min/Max
settings
F3
F4
Note
Each voltage range (155V, 310V, 600V) has an
independent voltage limit.
Background
Before setting the power supply voltage level,
set the voltage range and voltage limit.
Steps
1. Press the V key. The ACV
parameter will be editable.
V
V-Limit
2. Set the voltage with the scroll wheel/keypad or
with the F1 ~ F4 soft-keys.
Range
0 volts ~ full range
Soft-keys
DEF1, DEF2, MAX, MIN
Setting the Output Voltage
The voltage setting sets the voltage level of the power supply.
45
APS-7000 Series User Manual
3. Press Enter to confirm the voltage setting.
Preset Settings
The DEF1 and DEF2 preset settings are user
defined settings. By default they are set to 0.00
volts. The MAX and MIN soft-keys set the
voltage to the maximum and minimum,
respectively.
4. Press the V key and set the desired voltage with
the scroll wheel/keypad.
Range
0 volts ~ full scale of voltage range
5. Press and hold the DEF1 or DEF2 soft-key until
“Saved to DEF1/2” is displayed. This will save
the voltage setting to the DEF1 or DEF2 softkey.
Note
Trying to set the voltage outside of the voltage
limit/range will result in a voltage setting error
being displayed on the screen.
The voltage level can be set when the output is on.
Example
Voltage setting
Preset voltage
settings
F3
F4
F1
F2
46
OPERATION
Background
Setting the frequency limit allows the frequency
output to be set to any level within the limit
range.
Steps
1. Press Shift + F to access the Freq
Limit menu.
Shift
+
F
F-Limit
2. Set the frequency limit with the scroll
wheel/keypad or with the F3 ~ F4 soft-keys.
The MAX and MIN soft-keys set the frequency
limit to the maximum and minimum,
respectively.
Range
45.00 ~ 500.0Hz (999.9Hz option)
Soft-keys
MAX, MIN
3. Press Enter to confirm the limit setting.
Example
Freq Limit
Min/Max
settings
F3
F4
Setting the Frequency Limit
47
APS-7000 Series User Manual
Background
Before setting the frequency, set the frequency
limit.
Steps
1. Press the F key. The FREQ
parameter will be editable.
F
F-Limit
2. Set the frequency with the scroll wheel/keypad
or with the F1 ~ F4 soft-keys.
Range
45.00 ~ 500.0Hz (999.9Hz option)
Soft-keys
DEF1, DEF2, MAX, MIN
3. Press Enter to confirm the frequency setting.
Preset Settings
The DEF1 and DEF2 preset settings are user
defined settings. By default they are set to
50.00Hz and 60.00Hz, respectively. The MAX
and MIN soft-keys set the frequency to the
maximum and minimum, respectively.
4. Press the F key and set the desired frequency
with the scroll wheel/keypad.
Range
45.00 ~ 500.0Hz (999.9Hz option)
5. Press and hold the DEF1 or DEF2 soft-key until
“Saved to DEF1/2” is displayed. This will save
the frequency setting to the DEF1 or DEF2 softkey.
Setting the Output Frequency
The frequency setting sets the frequency of the output.
48
OPERATION
Example
Frequency setting
Preset frequency
settings
F3
F4
F1
F2
Note
Trying to set the frequency outside of the frequency
limit will result in a frequency setting error being
displayed on the screen.
The frequency can be set when the output is on.
Background
Setting the peak current limit sets a limit on the
current that can be sourced by the power
supply.
Note
When the peak current limit is tripped, an alarm
will sound. Press Shift + 9 to clear the Ipk alarm.
Shift+6 can also clear the Ipk alarm. See page 56
for details.
Steps
1. Press Shift + I rms to access the
Ipeak menu.
Shift
+
I rms
IPK-Limit
Setting the Peak Current Limit
49
APS-7000 Series User Manual
2. Set the peak current with the scroll
wheel/keypad or with the F3 ~ F4 soft-keys.
The MAX and MIN soft-keys set the peak
current to the maximum and minimum,
respectively.
Range
10% ~ 100% peak current value. The
peak current value depends on the
selected voltage range.
Soft-keys
MAX, MIN
3. Press Enter to confirm the peak current setting.
Example
Ipeak
Min/Max
settings
F3
F4
Delay Time
Settings
The Delay Time setting essentially defines how
long the measurement of the peak current must
be sustained for before it is recognized. By
default the delay time is turned off.
Load on
Ipk Limit
reached
I Pk Limit Delay
I
Ipk Limit acknowledged,
protection triggered
T
50
OPERATION
4. Press Shift + Irms and then press DELAY[F2].
5. Set the desired delay time with the scroll
wheel/keypad or with the F3 ~ F4 soft-keys.
The MAX and MIN soft-keys set the delay time
to the maximum and minimum, respectively.
Range
0(off) ~ 10 seconds
Soft-keys
MAX, MIN
6. Press Enter to confirm the delay time setting.
Example
Delay Time
Min/Max
settings
F3
F4
51
APS-7000 Series User Manual
Background
The I rms setting sets the root mean square
current.
Steps
1. Press I rms to access the I rms
menu.
I rms
Ipk-Limit
2. Set the I rms level with the scroll wheel/keypad
or with the F3 ~ F4 soft-keys. The MAX and
MIN soft-keys set the I rms level to the
maximum and minimum, respectively.
Range
0.00 ~ full scale A (dependant on
the voltage range)
Soft-keys
MAX, MIN
3. Press Enter to confirm the current setting.
Example
Irms
Min/Max
settings
F3
F4
Note
Setting the I rms level to 0.00 will disable OCP.
Setting the I rms to 0V is dangerous.
Setting the Current RMS Level
52
OPERATION
I rms Delay Time
Settings
The Delay Time setting defines how long the I
rms measurement must be sustained for before
it is recognized. By default the I rms delay time
is turned off.
Load on
Irms
Limit
reached
Irms Limit Delay
I
Irms Limit
acknowledged,
protection triggered
T
Irms Limt
4. Press Irms and then press DELAY[F2].
5. Set the desired delay time with the scroll
wheel/keypad or with the F3 ~ F4 soft-keys.
The MAX and MIN soft-keys set the delay time
to the maximum and minimum, respectively.
Range
0(off) ~ 10 seconds
Soft-keys
MAX, MIN
6. Press Enter to confirm the delay time setting.
Example
Delay Time
Min/Max
settings
F3
F4
53
APS-7000 Series User Manual
OC Fold Settings
The over current fold back settings allow the
APS-7000 to work as either a constant voltage
source or a constant current source.
While the unit is sourcing less current than the
Irms current limit, the APS-7000 will act as a
constant voltage source. In this mode, the
voltage level will remain constant while the
current level may vary. This is the normal
operating mode.
When the current level reaches the Irms limit,
the APS-7000 will act as a constant current
source. In this mode the current is constant and
the voltage level varies. When the current
subsides below the Irms limit again, the unit
will again act as a constant voltage source.
When OC Fold is turned off, the APS-7000 will
act as a current limiting power source when the
Irms limit has been reached.
Vrms
setting
ACV
I
Irms limit
Irms
Note
OC-FOLD can only be active when the I rms level is
greater than 0.
7. Press Irms and then press OC-FOLD[F1] toggle
the OC-Fold function on or off.
54
OPERATION
OC-Fold setting
Background
The on phase setting sets the starting phase of
the voltage output. The off phase setting sets
the ending phase of the voltage output.
Steps
1. Press Shift + 7 or Shift + 4
the On Phase or Off
Phase, respectively.
Shift
+
7
On Phase
or
4
Off Phase
2. Set the On Phase or Off Phase setting with the
scroll wheel/keypad or with the F3 ~ F4 softkeys. The MAX and MIN soft-keys set the
phase to the maximum and minimum,
respectively.
Range
0 ~ 359º
Soft-keys
MAX, MIN
Setting the On/Off Phase
55
3. Press Enter to confirm the phase setting.
Example
On Phase
Min/Max
settings
F3
F4
Off Phase
Alarm Clear
Background
The ALM CLR (Alarm Clear) function will clear
any Over Power, Over Irms, Over Ipeak and
Over Temperature alarms.
Applicable
Alarms
OVER POWER, OVER IRMS, OVER IPEAK,
OVER TEMPERATURE
Steps
1. Press Shift + 6 to clear any alarms.
Shift
+
6
ALM CLR
Example
Alarm message
ALM indicator
APS-7000 Series User Manual
56
OPERATION
Steps
1. Press the Display key.
2. The display mode will toggle each
time the key is pressed.
Display
Standard Mode
SettingsMeasurements
Configurable
measurements
Hold
measurement
Simple Mode
Measurement Items
Hold
measurement
Configuring the
Standard Mode
Measurements
1. Press the Meas Item1, Item2 or Item3
soft-key.
Display Modes
The APS-7000 power supply has two display modes. The standard
display mode shows the power supply setup on the left and the 3
configurable measurements on the right. The simple display mode
shows all measurement items available on the APS-7000.
57
APS-7000 Series User Manual
2. Use the scroll wheel to select a measurement
item and press Enter to confirm.
Sets first measurement item to Voltage
Hold
Measurement
The Hold function will “hold” the current
measurements on the display. Measurements
won’t be updated on the display until the
function is released.
Press HOLD[F4] to toggle hold on or off.
58
OPERATION
Activate the panel
lock
Press the Lock key to active the
panel lock. “Keys locked” appears
on the display.
A lock icon will appear in the top
corner when the panel keys are
locked.
Lock
Unlock
Disable the panel
lock
Hold the Lock key for ~3 seconds to
disable the panel lock. “Keys
unlocked” will appear on the
display and the padlock icon will
disappear.
Lock
Unlock
(hold)
Example
MessageLock icon
Panel Lock
The panel lock feature prevents settings from being changed
accidentally. When activated, all keys and knobs except the
Lock/Unlock key and the Output key (if active) will be disabled.
If the instrument is remotely controlled via the USB/LAN/RS232/GPIB interface, the panel lock is automatically enabled. See
page 131 for remote control details.
59
APS-7000 Series User Manual
Warning
Both of these outputs are electrically linked. Only one
DUT should be connected to any one of the outputs at
a time. Using both outputs at the same time is not
supported. Using the front and rear outputs are the
same time could cause dangerous operating
conditions. See page 26 for details about using the
output terminals or sockets.
Turn Output On
Press the Output key. The Output
key will light up and ON will be
displayed in the status bar to
indicate that the output is on.
Output
Turn Output Off
Press the Output key. The Output
key light will go out and OFF will
be displayed in the status bar to
indicate that the output is off.
Output
Turning the Output On
When the output is turned on, the DUT can be connected to either
the rear panel output or the front panel output.
60
OPERATION
Parameter
Settings
Mode
Auto: When the output is on, this
mode will automatically generate a
trigger at 0º to keep the surge or dip
event repeatedly on site.
Manual: When the output is on, this
mode will wait for the TRIG[F4] softkey to be pressed before starting the
surge or dip event on site.
OFF: Disables surge/dip control.
Note
The magnitude of the surge/dip part of the resultant
waveform depends on the surge/dip ACV setting level.
Advanced Settings
Surge/Dip Control → from page 61
Ramp Control → from page 65
Surge/Dip Control
Surge and dip control allows the power supply to source artificial
surges or dips in voltage to a DUT. The surge/dip control feature
provides a fast method to generate a surge/dip voltage event on a
nominal voltage.
There have 4 parameters for configuring this feature: Mode
selection (Mode), surge/dip voltage (ACV), the start time (T1) and
testing duration(T2) of the surge/dip voltage period. The nominal
voltage and frequency settings are based on the Basic Operation
section.
61
APS-7000 Series User Manual
Triggering
Example:
Site selection: The surge/dip site is
selected as shown below.
Trigger (0º for Auto)
Dip/Surge site
T2T1
ACV
100mS
ACV
Sets the ACV surge/dip level from
the 0V level.
T1
Sets the T1 time.
T2
Sets the width of the surge/dip.
100mS
Fixed 100mS delay after triggering.
Steps
1. Press the Menu key. The Menu
setting will appear on the display.
Menu
2.Use the scroll wheel to go to item 2, Surge/Dip
Control and press Enter.
3. Go to the Mode setting using the scroll wheel
and press Enter. Select the desired mode and
press Enter again to confirm.
The Manual mode will allow you to manually
trigger the surge/dip site. The Automatic
setting will automatically trigger the surge/dip
site.
Mode
Manual, Auto, OFF
4. Set the remaining parameters.
Note: these parameters are not visible when
MODE is set to OFF.
Remaining parameters
ACV, T1, T2
62
OPERATION
Exit
5. Press Exit[F4] to exit from the
Surge/Dip Control settings.
6. After exiting the menu, the
surge/dip control icon will appear
in the status bar.
Triggering the
Manual
Surge/Dip Site
For the manual mode, the surge/dip site is
determined by a manual trigger.
1. Configure the nominal voltage and
frequency settings. See the Basic
Operation chapter for details.
Page: 45, 47
2. Turn the output on. The nominal
settings above will be output.
Page 60
3. Press Shift to arm the trigger.
Press TRIG[F4] to generate the
trigger manually via the front
panel.
Shift
Or,
Alternatively, pulse the Trigger In
pin on the J1 connector high to
generate the trigger.
OR
Page 127
Note
The TRIG soft-key is only available when the surge/dip
control is set to
Manual
.
63
APS-7000 Series User Manual
Example
The following example shows a typical surge/dip
output waveform.
Trigger = 0º
ACV
T1T2100mS
64
OPERATION
Parameter
Settings
Time
The Time setting allows you to set
the ramp time as ms/Vrms.
Tup = ramp up time/1Vrms
Tdn = ramp down time/1Vrms
1Vrms1Vrms
TupTdn
Voltage
The Voltage setting allows you to set
the ramp time as Vrms/ms.
Vup = voltage up/1ms
Vdn = voltage down/1ms
VupVdn
1ms1ms
Steps
1. Press the Menu key. The Menu
setting will appear on the display.
Menu
2.Use the scroll wheel to go to item 3, Ramp
Control and press Enter.
Ramp Control
The Ramp Control function controls how fast the voltage level
ramps up and down. This function allows you to ramp the voltage
as a unit of time or as a unit of voltage.
65
APS-7000 Series User Manual
3. Go to the Mode setting using the scroll wheel
and press Enter. Select either Voltage or Time
and press Enter again to confirm.
3. Go to the T Ipeak, hold(msec) setting using the
scroll wheel and press Enter. Set the time and
press Enter again to confirm.
T Ipeak
1 ~ 60,000 ms
Miscellaneous
The Miscellaneous menu contains miscellaneous parameter settings.
T Ipeak, hold → from page 68
Power ON Output → from page 70
Buzzer → from page 71
SCPI Emulation → from page 72
Program Timer Unit → from page 73
T Ipeak, hold
The T Ipeak, hold function sets the hold time for the peak current
measurement. After the output is turned on, the APS-7000 will
delay starting the peak current measurement by this hold time.
68
OPERATION
Exit
4. Press Exit[F4] to exit from the
MISC Configuration settings.
Example
Settings
69
APS-7000 Series User Manual
Steps
1. Press the Menu key. The Menu
setting will appear on the display.
Menu
2.Use the scroll wheel to go to item 4, MISC
Configuration and press Enter.
3. Go to the Power ON Output setting using the
scroll wheel and press Enter. Select a setting
and press Enter to confirm.
ON
Set Output ON.
OFF
Set Output OFF.
SEQ
Execute the sequence that was
loaded before the unit was last
turned off.
SIM
Execute the simulation that was
loaded before the unit was last
turned off.
PROG
Execute the program that was
loaded before the unit was last
turned off.
Exit
4. Press Exit[F4] to exit from the
MISC Configuration settings.
Power ON Output
The Power ON Output setting allows you to have the output turn
on automatically after startup. The settings that are loaded are the
last settings that were present in the standard mode before the unit
was turned off last.
70
OPERATION
Example
Settings
Steps
1. Press the Menu key. The Menu
setting will appear on the display.
Menu
2.Use the scroll wheel to go to item 4, MISC
Configuration and press Enter.
3. Go to the Buzzer setting using the scroll wheel
and press Enter. Turn the setting on or off and
press Enter again to confirm.
Buzzer
ON, OFF
Exit
4. Press Exit[F4] to exit from the
MISC Configuration settings.
Example
Settings
Buzzer
The Buzzer setting turns the buzzer sound on or off for key presses
and alarms.
71
APS-7000 Series User Manual
Steps
1. Press the Menu key. The Menu
setting will appear on the display.
Menu
2.Use the scroll wheel to go to item 4, MISC
Configuration and press Enter.
3. Go to the SCPI Emulation setting using the scroll
wheel and press Enter. Choose the SCPI
emulation mode and press Enter again to
confirm.
SCPI Emulation
GW, EXTECH
Exit
4. Press Exit[F4] to exit from the
MISC Configuration settings.
Example
Settings
SCPI Emulation
The SCPI Emulation setting supports both GW Instek as well as
EXTECH SCPI remote commands. When set to EXTECH, the SCPI
Emulation setting allows the APS-7000 to be a drop-in replacement
for the EXTECH 6700 series by emulating the remote commands of
the EXTECH units.
72
OPERATION
Steps
1. Press the Menu key. The Menu
setting will appear on the display.
Menu
2.Use the scroll wheel to go to item 4, MISC
Configuration and press Enter.
3. Go to the Program Timer Unit setting using the
scroll wheel and press Enter. Select the unit and
press Enter to confirm.
Program Timer Unit
Hour, Min, Sec
Exit
4. Press Exit[F4] to exit from the
MISC Configuration settings.
Example
Settings
Program Timer Unit
The Program Timer Unit setting is used to configure the timer unit
for the Program mode.
73
APS-7000 Series User Manual
Pin Assignment
Trigger Out
Trigger InOut On / Off
COM
J1
Note
The following trigger control settings are not
applicable when in the Simulation test mode or the
Sequence test mode.
Steps
1. Press Shift + 2 to access the Trigger
Control menu.
Shift
+
2
Trigger
2. The trigger control settings appear. The settings
are divided into Input Pin and Output Pin.
Trigger
The trigger settings are used to turn the output on or off or to set
the behavior of the Trigger In and Trigger Out pins of the J1
Connector. The trigger can also be used in conjunction with the
Sequence mode, Simulation mode and Surge/Dip mode. See page
127 for further details.
Trigger Control Settings
The trigger control menu sets both the trigger in and the trigger out
behavior of the J1 connector.
74
OPERATION
Trigger input
settings
Exit
Trigger output
settings
Trigger Input Pin
1. The Input Pin settings configure what action is
taken when the Trigger In pin on the J1
connector is high.
2. Go to the Mode setting using the scroll wheel
and press Enter. Choose Auto or Manual and
then press Enter again to confirm.
Auto
Auto mode will trigger the system
when the signal input into the
Trigger In pin goes high (+5V).
Manual
The Manual mode will ignore any
signal input into the Trigger In pin,
but will trigger the system by
remote commands
[1]
.
3. Go to the Action setting using the scroll wheel
and press Enter. Choose what will happen
when the Trigger In pin is set high(+5V) and
then press Enter to confirm.
None
No action is taken.
Output
Turns the output on or off when
triggered.
Setting
Sets a user-defined voltage and
frequency setting when triggered.
Preset
Loads a preset setting when
triggered.
SurgeDip
Triggers the surge/dip control.
75
APS-7000 Series User Manual
4. If Setting was chosen, Set Vset and Fset.
Vset
0 ~ full scale voltage (dependant on
the chosen range; 600V optional)
Fset
45.00 ~ 500.0 Hz (999.9 Hz optional)
5. If Preset was chosen, choose which preset
setting should be loaded when the Trigger
Input is set to high(+5V).
Preset
0 ~ 9
Trigger Output
Pin
1. Go to the Mode setting using the scroll wheel
and press Enter. Choose Auto or Off and then
press Enter.
Auto
Auto mode will set the Trigger Out
pin to high (+5V) when the
conditions are true.
Off
Off mode will set the Trigger Out
pin to low (0V).
2. Go to the Source setting to choose what function
or test mode will be able to the trigger the
output pin on.
None
No source is able to use the trigger
output.
Output
The trigger output goes high (+5V)
when the Output is turned on or off.
Setting
When the settings change the
trigger output goes high (+5V).
Preset
When a preset is loaded or saved,
the trigger output goes high (+5V).
All
The trigger output goes high (+5V)
when any of the settings, presets or
output are changed.
76
OPERATION
3. Go to Width to set how long the trigger output
is set high for.
Width
0.1 ~ 100.0ms
Exit
4. Press Exit[F4] to exit from the
Trigger Control settings.
[1] Remote
Control
Commands
The following remote control commands are
applicable for the trigger input or output. See the
programming manual for usage details.
present settings to the
corresponding memory
number.
Preset
+
0
~
9
(hold)
Presets
M0 ~ M9
2. Press the Preset key again to exit from the preset
mode.
Example
For example, pressing Preset & holding 1 will
save the present settings to memory slot 1
(saved to M1).
Note
The preset key will become green when active. A
beep will be heard (Buzzer set to ON) and a
message will displayed when the settings are
saved.
Preset Settings
Save Presets to Local Memory → from page 78
Recall Presets to Local Memory → from page 79
Manage Preset Settings → from page 80
Save Preset Settings to Local Memory
Up to 10 preset settings can be saved to internal memory.
78
OPERATION
Steps
1.Press Preset andpress a
number key(0~9) to load the
corresponding memory
number.
Preset
+
0
~
9
Presets
M0 ~ M9
2. Press the Preset key again to exit from the preset
mode.
Example
For example, pressing Preset + 1 will recall the
saved settings from memory slot 1 (recalled
from M1).
Note
The preset key will become green when active. A
beep will be heard and a message will be displayed
when the settings are recalled.
Load Preset Settings to Local Memory
Any of the 10 preset settings can be recalled from internal memory.
79
APS-7000 Series User Manual
File Format
When files are saved to USB they are saved in
the following format:
presetX.set, where X is the memory number
M0 ~ M9. The files are saved to USB:/gw.
When files are recalled from USB, files must be
recalled from the same memory number. For
example, the file preset0.set can only be recalled
to memory number M0. The files can only be
recalled from the USB:/gw directory.
Steps
1. Press the Menu key. The Menu
settings will appear on the display.
Menu
2.Use the scroll wheel to go to item 12, Save/Recall
Files and press Enter.
3. Go to the Type setting using the scroll wheel
and press Enter. Select Preset and press Enter to
confirm.
4. Go to the Action setting and choose the file
operation and then press Enter.
MEMUSB
Saves the selected preset
memory from the local memory
to a USB flash drive.
MEMUSB
Loads a preset memory from a
USB flash drive to the selected
local memory.
DELETE(MEM)
Deletes the selected preset
memory from local memory.
Manage Preset Settings
Preset settings can be easily saved to or from a USB flash drive
using the Save/Recall Files utility in the Menu system. Files can
also be deleted from local memory using the utility.
80
OPERATION
5. Go to the Memory No. setting and select the
preset memory number to perform the
operation on. Press Enter to confirm.
Memory No.
0 ~ 9 (M0 ~ M9)
Execute File
Operation
6. Press Exe[F1] to perform the file
operation.
Exit
7.Press Exit[F4] to exit from the
Save/Recall Files settings.
Example
Settings
81
APS-7000 Series User Manual
Note
The output will always be turned off if the ARB
waveform is changed. It is not possible to keep the
output on whilst the type of ARB waveform is
changed.
Arbitrary Waveform Function
The ARB function is able to select a large range of continuous
arbitrary waveforms. There are a number of basic waveform shapes
to choose from, each of which can be customized to output a
waveform with your desired attributes.
The ARB function can be used in conjunction with the Simulate and
Sequences modes, but cannot be used with the Program test modes.
When the WAVE parameter is set to any waveform other than
SINE, ARB will be shown in the status bar when in the BASIC
operation, Simulate mode or Sequence mode to indicate that the
ARB function is active when in these modes.
ARB Function Overview → from page 83
Selecting an ARB Waveform → from page 86
82
OPERATION
Background
The ARB mode function has 7 basic waveforms
and a number of settable attributes, depending
on the chosen waveform.
Setting Screen
Overview
Waveform
type
Attributes for the
selected waveform
Visual representation
of the waveform shape
and its attributes
F4 Exit
ARB Waveform
Overview
The following describes each of the 7 basic
waveforms.
Sine
This is the default waveform used for
the APS-7000. No settable attributes.
Triangle
Basic triangle waveform. No settable
attributes.
ARB Mode Overview
83
APS-7000 Series User Manual
Staircase
The staircase waveform has a settable
number of step levels.
Attributes:
Stairs: 1 ~ 100
Clip
Outputs a clipped sinewave. The
degree to which the sine wave is
clipped is settable.
Attributes:
n*ACV: 1.0 ~ 10.0 (clip level)
CF
Crest factor waveform. The crest factor
is settable.
Attributes:
CF: 1.4 ~ 10.0
84
OPERATION
Surge
The surge waveform has a settable
ACV base level, site size and site
shape.
The ARB waveform shape function
simply adds a number Fourier series
terms to create an arbitrary waveform.
Attributes:
Type: 1 ~ 22 (Number of selectable
ARB waveforms)
The numerical figures shown under the
Type parameter shows which Fourier
series terms are used and their (bn)
coefficients.
85
APS-7000 Series User Manual
Entering the ARB
Menu
1. Press Shift + 1 to access the ARB
menu.
Shift
+
1
ARB
2. Set the Wave parameter with the scroll wheel
and confirm the setting with the Enter key.
Wave
Sine, Triangle, Stair, Clip, CF, Surge,
ARB
3. Use the scroll wheel and Enter key to select any
other attributes for the selected waveform and
press Enter to confirm any settings.
Other Attributes:
Sine Wave:
None
Triangle Wave:
None
Stair:
Stairs: 1 ~ 100
Clip:
N*ACV: 1 ~ 10
CF:
CF: 1.4 ~ 10.0
Surge:
Type: Sine, Square
ACV: 0 ~ 100%
Site: 0 ~ 100%
Exit
4. Press Exit[F4] to exit from the ARB
settings.
Selecting an ARB Waveform
86
OPERATION
Common Settings
5. Press Menu to go back to the main
menu.
Menu
6. Set the voltage, frequency, Irms
level and so on as you normally
would for the standard ACV
output. See the Basic Operation
chapter for setting details.
Page 43
Note
When the WAVE parameter is set to any waveform
other than SINE, ARB will be shown on the display
when in the BASIC operation, Simulate mode or
Sequence mode to indicate that the ARB function
is active when in these modes.
Standard Mode
Example
Simple Mode
Example
Sequence
Example
87
APS-7000 Series User Manual
Simulate Example
88
OPERATION
Test Mode Function
There are three test modes.
Simulate Mode: voltage/frequency fluctuations.
The Simulate function is a quick and easy way to simulate
fluctuations in voltage, frequency, phase and other anomalies in
mains supply power.
Sequence Mode: creates test sequences.
The Sequence function creates arbitrary AC waveforms based on
user defined parameters.
Program Mode: Runs a group of test sequences.
The Program function runs different memory steps sequentially.
Each step of a program is a preset memory configuration (M01 ~
M50). The Program function can also perform pass/fail testing.
Simulate Mode Overview → from page 91
Simulate Settings → from page 95
Save Simulation to Local Memory→ from page 98
Recall Simulation from Local Memory→ from page 98
Manage Simulation Settings → from page 99
Run Simulation → from page 101
Sequence Mode Overview → from page 103
Sequence Settings → from page 107
Save Sequence to Local Memory → from page 111
Recall Sequence from Local Memory → from page 111
Manage Sequence Settings → from page 112
Run Sequence → from page 114
89
APS-7000 Series User Manual
Program Mode Overview → from page 116
Save a Program to Local Memory → from page 123
Recall a Program from Local Memory → from page 123
Manage Program Settings → from page 124
90
OPERATION
Background
The Simulate function is used to test power
supply fluctuation. This function is able to
simulate common abnormalities in mains
power such as fluctuations in voltage, phase
and frequency. These simulations can be run as
one-off anomalies or cyclic anomalies.
Setting Screen
Overview
Number of repetitions
Test mode
V RANGE
Recall test
Save test
Run test
ON/OFF phase
Trigger output
level
Step Frequency
Step voltage
Step time
Step
Step Overview
The Simulate function is comprised of 6 steps.
Each step is run sequentially in the following
order: Initial, Normal1, Trans1, Abnormal,
Trans2, Normal2, Initial.
Initial
The Initial step is used as the initial
and final settings of the waveform
simulation. This is the standby step
before the test starts and the standby
step after the test ends.
Normal1
This step configures the normal output
conditions that precede the abnormal
conditions.
Simulate Mode Overview
91
APS-7000 Series User Manual
Trans1
This step configures the transition
from normal to abnormal conditions.
This step will linearly interpolate the
normal settings to the abnormal
settings. This step can be skipped for
abrupt state changes.
Abnormal
This step contains the abnormal
conditions for the simulation.
Trans2
This step configures the transition
from abnormal to normal conditions.
Normal2
This step configures the normal
conditions that supersede the
abnormal conditions.
AbnormalTrans1Normal1InitTrans2 Normal2 Init
Parameter
Overview
The following table shows which parameters
are available for each step.
Step\Parameter
Initial
Normal1
Trans1
Abnormal
Trans2
Normal2
Repeat
✓ ✓ ✓ ✓ ✓
✓
Time
✓ ✓ ✓ ✓ ✓
✓
ON Phs
✓
✓ X ✓ X ✓
OFF Phs
✓
✓ X ✓ X ✓
Vset
✓
✓ X ✓ X ✓
Fset
✓
✓ X ✓ X ✓
Trig Out
✓ ✓ ✓ ✓ ✓
✓
92
OPERATION
Repeat
Indicates the number of times the
simulation will be run, from Normal1
to Normal2.
A value of 0 indicates infinite repeats.
The repeat setting is the same for
each step.
Time
Sets the duration time of the step.
When the ON Phs=ON, the total
duration of the step is equal to the
Time setting + ON Phs=ON
duration.
ON Phs
Sets the starting phase of the
waveform for the step. Not applicable
for the Trans 1/2 steps.
OFF Phs
Sets the off phase of the waveform
after the output has been turned off.
Not applicable for the Trans 1/2
steps.
Fset
Sets the frequency of the step. Not
applicable for the Trans 1/2 steps.
Vset
Sets the voltage of the step. Not
applicable for the Trans 1/2 steps.
Trig Out
Sets the polarity of the trigger out
port for the duration of the step. The
trigger out signal is sent at the start
of each step (including the first initial
step).
93
APS-7000 Series User Manual
The following diagram illustrates the
relationship between each of the parameters in
a step.
0.01 ~ 99.99s, 0(Trans 1 and
Trans2)
Note: For Trans1 and Trans2, it
supports a value of 0, which will
skip the step.
Simulate Settings
95
APS-7000 Series User Manual
6. Go to the ON Phs setting and set the starting
phase of the step. Not applicable for Trans1 or
Trans2.
ON Phase
ON, OFF
ON Phase
0~ 359º
Resolution
1º
7. Go to the OFF Phs setting and set the end phase
of the step. Not applicable for Trans1 or Trans2.
OFF Phase
ON, OFF
OFF Phase
0~ 359º
Resolution
1º
8. Press the Range key repeatedly to set the
voltage range for the Vset parameter. The range
will be shown in the top corner, which indicates
that the simulation will be executed within this
voltage range.
Range
Range
LO(155V), MD(310V), HI(600V,
optional)
9. Go to the Vset setting and set the Vrms level of
the step. If you input a Vset value that is not
within the voltage range, the input value will
be ignored.
Not applicable for Trans1 and Trans2.
10. Go to the Fset setting set the frequency of the
step. Not applicable for Trans1 and Trans2.
Fset
45.00 ~ 500.0Hz (999.9Hz optional)
11. Go to the Trig Out setting and set the state of
the trigger out pin at the start of each step.
Trig Out
HI, LO
12. Lastly, go to the Repeat parameter select the
number of times the simulation will repeat the
Normal1-Trans1-Abnormal-Trans2-Normal2
sequence of steps. A value of 0 will set the
number of repetitions to infinite.
Repeat
1 ~ 9999, 0(infinite)
97
APS-7000 Series User Manual
Saving a
Simulation
Simulation settings can be saved to one of 10
memory slots (SIM0 ~ SIM9).
Steps
1. Press Save[F3] and then long press a number key
when prompted.
2. A message will appear when the save is
successful.
Save
SIM0 ~ SIM9
Recall a
Simulation
Simulation settings can be recalled from one of
10 memory slots (SIM0 ~ SIM9).
Steps
1. Press Recall[F2] and then press a number key
when prompted.
2. A message will appear when the settings are
recalled successfully.
Recall
SIM0 ~ SIM9
Save a Simulation to Local Memory
Recall a Simulation from Local Memory
98
OPERATION
File Format
When files are saved to USB they are saved in
the following format:
simX.sim, where X is the memory number
0 ~ 9 (SIM0 ~ SIM9). The files are saved to
USB:/gw.
When files are recalled from USB, files must be
recalled from the same memory number. For
example, the file sim0.sim can only be recalled
to memory number SIM0. The files can only be
recalled from the USB:/gw directory.
Steps
1. Press the Menu key. The Menu
settings will appear on the display.
Menu
2.Use the scroll wheel to go to item 12, Save/Recall
Files and press Enter.
3. Go to the Type setting using the scroll wheel
and press Enter. Select SIMULATE and press
Enter to confirm.
4. Go to the Action setting and choose the file
operation and then press Enter.
MEMUSB
Saves the selected simulation
memory from the local memory
to a USB flash drive.
MEMUSB
Loads the simulation memory
from a USB flash drive to the
selected local memory.
Manage Simulation Settings
Simulation settings can be easily saved to or from a USB flash drive
using the Save/Recall Files utility in the Menu system. Files can
also be deleted from local memory using the utility.
99
APS-7000 Series User Manual
DELETE(MEM)
Deletes the selected simulation
memory from local memory.
5. Go to the Memory No. setting and select the
simulation memory number to perform the
operation on. Press Enter to confirm.
Memory No.
0 ~ 9 (SIM0 ~ SIM9)
Execute File
Operation
6. Press Exe[F1] to perform the file
operation.
Exit
7.Press Exit[F4] to exit from the
Save/Recall Files settings.
Example
Settings
100
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.