7529_eng.fm Page 1 Wednesday, November 12, 2003 8:57 AM
OPERATORS MANUAL FOR
MODEL ETA2 AND ETA5 ANALYZERS
This ETA Analyzer is used for calibration or data collection of torque tools, power and
hand operated, using external transducer(s). This Analyzer can also be used for joint
development and as a failure analysis instrument (using external transducers).
Ingersoll-Rand is not responsible for customer modification of units for applications on
which Ingersoll-Rand was not consulted.
This symbol is to alert the user and service personnel to the presence of important operating instructions that must be read and understood to prevent personal injury, electrical
shock or damage to the equipment.
IMPORTANT SAFETY INFORMATION ENCLOSED -
SAVE THESE INSTRUCTIONS.
READ AND UNDERSTAND THIS MANUAL BEFORE OPERATING THIS PRODUCT.
IT IS YOUR RESPONSIBILITY TO MAKE THIS SAFETY INFORMATION
AVAILABLE TO OTHERS THAT WILL OPERATE THIS PRODUCT.
FAILURE TO OBSERVE THE FOLLOWING WARNINGS COULD RESULT IN
ELECTRIC SHOCK, FIRE AND/OR SERIOUS PERSONAL INJURY.
04578092
Form P7529
Edition 1
January, 2002
Refer All Communications to the Nearest Ingersoll-Rand
Office or Distributor.
7529_eng.fm Page 2 Wednesday, November 12, 2003 8:57 AM
PLACING ANALYZER IN SERVICE
•Always install, operate, inspect and
maintain this product in accordance
with all applicable standards and
regulations (local, state, country, federal, etc.).
USING THE ANALYZER
•Always wear eye protection when operating or performing maintenance on
this tool.
•Always use Personal Protective Equipment appropriate to the tool used and
material worked. This may include dust
mask or other breathing apparatus,
safety glasses, ear plugs, gloves,
apron, safety shoes, hard hat and other
equipment.
•Keep others at a safe distance from
your work area, or ensure they use
appropriate Personal Protective Equipment.
•This tool is not insulated against electric shock.
•This unit is for indoor use only.
•Store idle units. When not in use, units
should be stored in a dry, high or
locked location, out of the reach of children.
•Power tools can vibrate in use. Vibration, repetitive motions or uncomfortable positions may be harmful to your
hands and arms. Stop using any tool if
discomfort, tingling feeling or pain
occurs. Seek medical advice before
resuming use.
•Keep body stance balanced and firm.
Do not overreach when operating this
tool. Anticipate and be alert for sudden
changes in motion, reaction torques, or
forces during start up and operation.
•Ensure work pieces are secure. Use
clamps or vises to hold work piece
whenever possible.
•Never use a damaged or malfunctioning tool or accessory.
•Keep work area clean, uncluttered,
ventilated and illuminated.
•Do not remove any labels. Replace
any damaged label.
•Maintain unit with care. Keep unit clean
for better and safer performance. Follow instructions for changing accessories. Inspect unit cords periodically and
if damaged, have them repaired by an
authorized service facility.
•Do not operate this product in explosive atmospheres, such as in the presence of flammable liquids, gases or
dust.
•Do not modify this product, safety
devices, or accessories.
•Do not use this product for purposes
other than those recommended.
•Use accessories recommended by
Ingersoll-Rand for your model.
•Don’t abuse the cord. Never use the
cord to pull the plug from an outlet.
Keep cords away from heat, oil, sharp
edges or moving parts. Replace damaged cords immediately.
•Avoid body contact with grounded surfaces such as pipes, metal structures
or other electrical products.
•Stay alert, watch what you are doing
and use common sense when operating this product. Do not use this product while tired or under the influence of
drugs, alcohol, or medication.
•Do not use this product near water or
other liquids.
•Do not attach or place external loads
on any part of the unit.
•Replace battery with the same or equivalent type recommended by the battery
manufacturer. Observe correct polarity
when installing battery. Dispose of
used batteries according to battery
manufacturer’s instructions.
2
7529_eng.fm Page 3 Wednesday, November 12, 2003 8:57 AM
CAUTION
•The unit should be checked periodically to ensure that the torque readings
are accurate.
•DO NOT store the unit in relative
humidity above 85%.
•Have the unit calibrated at least once a
year.
•Keep the exterior of the unit clean and
dry.
•Do not drop or abuse the unit.
•Do not allow chemicals such as ace-
tone, benzene, thinner, ketone or
trichloroethylene to come in contact
with the housings, as damage will
result.
•The instrument should not be subjected to vibration or shock.
Ingersoll-Rand Company makes no warranty or implies any warranty or liabilities due to the
misuse or damage resulting from the application of the information supplied by this manual.
Liabilities due to errors in the manual are only limited to replacement of the manual.
Ingersoll-Rand reserves the right to change information contained in this manual or the program without notice at any time.
All programs and themanual are copyrighted and rights reserved. Reproduction is prohibited
without prior permission of Ingersoll-Rand.
The use of other than genuine Ingersoll-Rand replacement parts may result in safety hazards,
decreased tool performance, and increased maintenance, and may invalidate all warranties.
Repairs should be made only by authorized trained personnel. Consult your nearest IngersollRand Authorized Servicenter.
•For calibration or repair, return to:
Ingersoll-Rand
ETA Dept.
510 Hester Dr.
Whitehouse, TN 37188
Ingersoll-Rand IDC
ETA Dept.
Swan Lane, Hindley Green
Nr. Wigan
Lancashire, WN2 4EZ
U.K.
3
7529_eng.fm Page 4 Wednesday, November 12, 2003 8:57 AM
WARNING SYMBOL IDENTIFICATION
WARNING
Always wear eye protection
when operating or performing
maintenance on this tool.
WARNING
Read this manual before
operating tool.
WARNING
Always wear hearing
protection when operating
this tool.
SAVE THESE INSTRUCTIONS. DO NOT DESTROY.
When the life of the tool has expired, it is recommended that the
tool be disassembled, degreased and the parts be separated by material
so that they can be recycled.
4
7529_eng.fm Page 5 Wednesday, November 12, 2003 8:57 AM
7529_eng.fm Page 7 Wednesday, November 12, 2003 8:57 AM
How to use this Manual
This manual is split into sections describing the steps to be taken to configure the
ETA for use in measuring and recording torque values.
There are a number of other parameters which are set from the Software Reset Menu, some of
which are detailed elsewhere in this manual.
Pressing after making a selection will take the user on to the next screen.
Pressing will give the user the option to exit to the Misc menu or continue to edit by
Enter
MENU
pressing .
Screens can be skipped using the and keys.
The following method is adopted throughout this manual.
Keys to be pressed will be shown as the key legend in large type: Example
Any special instructions or point to note will be shown as:
Enter
Enter
INSTRUCTION
TO BE
FOLLOWED
Refer to the table of contents to find action to be performed.
Follow instructions for key presses required to carry out required action.
Note:It is possible to change the default primary character on dual function
keys. ie: when text is entered into various fields, it is possible to set
the ETA to default to either the numerals or text as the primary character (the alternate character is selected by pressing the shift key
before typing). To set this:
From main menu:
Press then press (Misc)
6
Press then press (Software Reset)
6
Press the 7 times to access the ‘Shift key’ menu
Select between options 1 ‘Characters’ or 2 ‘Numbers’ using:
MENU
Press Screen will prompt “End of setup”
MENU MENU
Press to return to main menu
7
7529_eng.fm Page 8 Wednesday, November 12, 2003 8:57 AM
The following items, except those listed as optional, are included with the ETA unit:
7529_eng.fm Page 9 Wednesday, November 12, 2003 8:57 AM
Care and Storage
When not in use the unit should be returned
to the supplied carry case.
This unit is designed for indoor use only
Operating temperature range
5-40 degrees C
Storage temperature range
0-50 degrees C
Battery Charging
The batteries in the ETA unit are shipped
fully charged.
In continuous use with a transducer connected, the batteries have a life of at least 8
hours.
With the unit switched off from a 10% charge
state, the batteries will fully discharge in 2550 days. To prevent the loss of all setup
data, the unit has additional battery backup
for the internal memory.
Automatic Power Off
To conserve battery life in use, the unit will
switch off automatically after a predefined
period. Pressing the on key will restore the
unit to the last display prior to powering off.
The membrane keypad may be wiped clean
with a soft damp cloth. The unit is not waterproofed and spillages should be avoided.
THIS UNIT CONTAINS NO USER SERVICEABLE PARTS. ONLY QUALIFIED
SERVICE PERSONNEL SHOULD
REPLACE OR FIT PARTS.
Fast Charge
The ETA’s should only be recharged with the
supplied ETA-BC (Model MPP3OUS12) Battery Charger. ETA’s, are supplied with internal NiMH (Nickel Metal Hydride) batteries.
These are environmentally friendly batteries
presenting no problems for safe disposal
and are capable of sustaining a rapid
recharge. From fully discharged, the unit will
require a charging period of approximately
2.5 hours.
To charge the batteries following a period of
use, connect the supplied charger to the
charging socket on the back of the unit and
plug the charger into a suitable supply. The
green LED will come on to indicate the
machine is charging and will start to flash
when the unit is fully charged. If the red LED
comes on, this means there is a problem
and usually indicates a faulty battery.
The Fast Charge method gives an increase
in the effective life of the charged batteries.
9
7529_eng.fm Page 10 Wednesday, November 12, 2003 8:57 AM
Edit Transducer File
Failure to input proper information may result in inaccurate readings.
The following instructions for “Edit Transducer File” must be followed
before any reading or measurements can be taken.
This section may be skipped if you are using an Ingersoll-Rand TSD,
TRD or Crane UTA Series Transducer.
There are 5 types of transducers and inputs
which the ETA will accept.
1.High Output (H/O). These transducers
have an internal amplifier and give an
output signal level of 1-2 Volts.
2.Industry Standard (IS). These trans-
ducers have no internal amplifier. The
exact span or rated torque will be
marked on the transducer nameplate.
The sensitivity of these transducers is 2
mV/V., they are fitted with a 350 Ohm
bridge.
3.Serial. These are digital type transducers and produce an RS232 signal.
4.Keypad (KPD). This is an optional keypad which connects to the ETA.
5.Keyboard (KEY). Keyboard on ETA.
Use of any of the above 5 types of inputs
requires that settings are made in the transducer file before any measurements can
take place. The procedure for entering settings follows.
For ETA5:
At main menu, press (Configure) then
press (Edit Tx)
Screen will show
2
1
Transducer
B (Current set-up)
Complete set-up
The ‘B’ indicates that this is the ‘B’ file in
memory. It is the first of 8 memories (labelled
B to I) reserved for the saving of transducer
set-ups. Your most used transducer setting
should be stored in this memory.
The remaining memories (C to I) may be
accessed by pressing .
Press
Select input type H/O, I/S, SER, KPD or
KEY using ; press .
Enter
Enter
For ETA2:
At main menu, press (Edit Tx)
1
Depending on type of type selected above,
continue set-up by following the appropriate
section.
10
7529_eng.fm Page 11 Wednesday, November 12, 2003 8:57 AM
If you have chosen I/S
Select units of measure - inlb, ftlb etc. using press
Type in transducer span-press
Type in mV/Vpress
Type in Bridge resistancepress
Type in Pulses Per Rev (only for angle type)press
Type in Serial Numberpress
To set-up another transducerpress
Enter
Enter
Enter
Enter
Enter
Enter
MENU
Select another memory (B to I) using the keys and press
To return to the main menu, press
MENU
If you have chosen SER
Select location of decimal point using press
Enter
1
Enter
Select Baud Rate, Data Bits, Stop Bits and Parity in the same manner
To set-up another transducer press
MENU
1
Select another memory (B to I) using the keys and press
To return to main menu press
MENU
11
Enter
7529_eng.fm Page 12 Wednesday, November 12, 2003 8:57 AM
If you have selected H/O
Select units of measure- inlb, ftlb, etc. using
Note: High Output devices usually produce an output voltage of approximately 1-2 Volts. This
analogue voltage signal will not be exact and may not be linear. If you are using a device that
has a nominal 1Volt output, we suggest that you multiply the span of your transducer by 2.5 to
determine the new span setting.
Type in your new span; press
Type in Serial Number; press
Screen will indicate ‘End of Set-up’; press to return to main menu
Enter
Enter
MENU
Connect your high output device and select measure mode
(refer to instruction sheet for measure mode only)
Use an independent measuring device to apply a torque to the transducer and check the reading on the independent unit against the reading shown on the ETA. It will be necessary to return
to the ‘Edit Tx’ menu option and adjust the span setting until the two readings agree. You may
have to repeat the above procedure several times until your high output transducer is correctly
calibrated.
If you have selected KPD
Select location of decimal point using then press
Select Parity- Odd, Even or Disabled using then press
To set-up another transducer press
MENU
1
Select another memory (B to I) using the keys and press
Enter
Enter
Enter
To return to main menu press
MENU
12
7529_eng.fm Page 13 Wednesday, November 12, 2003 8:57 AM
If you have selected KEY
Select location of decimal point using then press
To set-up another transducer press
MENU
1
Select another memory (B to I) using the keys and press
To return to main menu press
MENU
Enter
Enter
Barcode read input for characteristic
names and comments
It is possible to scan a barcode name in ETA5 at certain prompt lines. The input of a barcode is
allowed at the characteristic name prompt in store characteristic, direct characteristic and edit
characteristic modes. In addition it is also possible to scan a barcode in the comment field in
store characteristic only. Edit characteristic and normal store behave in a slightly different manner to direct characteristic. The ETA5 will accept most barcode reading formats, but not all.
General Operational Method for Scanning Bar Code Data
Connect the Bar Code Reader ETA-BAR to the Printer/PC connector at the rear of the ETA5.
Barcode readings can generally be taken at the prompt to input data (Characteristic name or a
comment) when the flashing cursor is on the screen. This can be on a blank line or following a
line of data (an existing characteristic or a comment). When the cursor is flashing, scan the
code with a barcode reader and observe the screen. If the flashing cursor disappears, the code
has been captured - wait a few seconds for the data to appear in the display. If the flashing cursor remains, re-scan the code until it is accepted.
When the data is displayed, press to continue. From this point, follow the instructions
Enter
as for normal data input.
When entering data via the Bar Code Reader into a comment field, make sure you have
reserved enough characters in the comment field to accept the code. If the number of characters scanned in exceeds the number reserved, the ETA5 will beep and refuse to accept the
data. Refer to the instructions on Configuration for more details.
13
7529_eng.fm Page 14 Wednesday, November 12, 2003 8:57 AM
Measure Mode (Up to 200 readings stored)
To start, press
MENU
Press
Press (measure)
2
Plug in transducer; press
On
Enter
select proper transducer using
then press
Enter
If using I/S or H/O type transducers
continue with instructions.
If using SER, KPD or KEY, you are now
ready to take measurements.
Select units of measure using ; press
Select Peak, Track, Impulse or ‘Click Dip’ using ; press
Use Peak for direct drive tools, Track for diagnostics, Impulse for pulse tools, and ‘Click Dip’ for
click wrenches.
Select measurement direction using (Right is clockwise. Left is anti-clockwise);
Enter
Enter
press
Select ‘Cycle end Time’ using ; press (normal is 0.5 sec.)
Select frequency response using ; press
Type in Max torque value; press
Type in Min torque value; press
Enter
Enter
Enter
Enter
Enter
14
7529_eng.fm Page 15 Wednesday, November 12, 2003 8:57 AM
Type in threshold value; press
Enter
Note: When selecting a threshold torque value the ETA will ignore any
torque that is below this value. You cannot take another reading until
the applied torque falls below this value.
Type in amount of Dip; press
Select Second Parameter using ; press
Enter
Enter
If the setup and transducer has not been changed since the ETA was last used you are now
ready to take torque measurements.
Pressing after taking a series of readings will result in statistical information being dis-
#
played (Range, mean and standard deviation) for the stored readings.
To return to measure mode, press
If you have selected ‘Click Dip’ type in the amount of ‘Dip’ to be sensed; press
#
Enter
Note: 5% of torque for 0 to 100 Nm and 10% of torque for over 100 Nm
generally provide excellent results.
You are now ready to take torque measurements. When finished press
MENU
Note: When taking readings you can change the set-up parameters by
pressing the up arrow
Changing any of the set parameters will result in the ETA prompting to erase all data.
To erase stored readings, press
To return to setup screens, press
Enter
MENU
Example: You have been measuring in a right hand direction and you want to
switch to left hand.
Press the up arrow twice, change from right hand to left hand. Screen will prompt to erase
all data.
Press to erase all stored readings and return to the measure screen.
Enter
15
7529_eng.fm Page 16 Wednesday, November 12, 2003 8:57 AM
Printing of stored readings from measure mode
Pressing from the measure mode will result in all stored readings being printed to an
attached printer or computer.
P
Note: Transducer needs to be connected to ETA.
Measure Mode Circle-Store Option
Circle-Store is a function to increase the number of stored readings to extend beyond the normal maximum of 200 readings possible in measure mode. With the circle-store option selected,
readings will be accepted beyond the 200 limit but the first readings taken will be lost. Effectively
the last 200 readings taken will be stored but the user is not limited to taking only 200 readings.
To select this option:
From main menu
Press then press (Misc’)
Press then press (Software Reset)
Press the 6 times to access the Circle-store menu
Select option 2 ‘Circle-store On’ using ; press
Press Screen will prompt: “End of setup”
Press to return to main menu
MENU
MENU MENU
6
6
Enter
16
7529_eng.fm Page 17 Wednesday, November 12, 2003 8:57 AM
Direct Measure Mode - ETA5 ONLY
(Ingersoll-Rand TRD, TSD or
Crane UTA Type Transducers)
Direct measure mode allows an operator to connect different transducers which may have different spans and an angle output, and can switch between them using an external T switch. This is
a display only function and readings are not stored or printed out.
To start, press
ON
press
press(Direct)
press(Direct Measure)
press(Setup)
select units of measure using ; press
select peak, track, Impulse or ‘Click Dip’ using ; press
MENU
4
1
1
Enter
Enter
select measurement direction using (Right is clockwise. Left is anti-clockwise);
press
select ‘Cycle end Time’ using ; press
select frequency response using ; press
type in the Threshold Percent (mx 50.0; mn 0.0), and press
type in amount of Click (mx 100.0 Nm; mn 0.0 Nm), and press
Enter
Enter
Enter
Enter
Enter
select ‘Second Parameter’ using ; press
Screen shows “End of Setup”, press to exit
MENU MENU
press to return to main menu
MENU
17
Enter
7529_eng.fm Page 18 Wednesday, November 12, 2003 8:57 AM
The READ option allows you to read the torque from a connected transducer.
From the main menu, press(Direct)
press(Direct Measure)
press(Read)
If no transducer is connected, display will prompt: “Insert Tx” (transducer)
connect transducer and press
Screen will show the span of the connected transducer and the torque measured in units
selected at setup. The display will show the last reading taken until the next torque input
exceeds the threshold level set.
Note: If password protection has been set on Direct measure, then exit is by
password only. A correct password returns you to the setup read submenu. Entering three incorrect passwords will force the user back
into read mode. See password protection later in this book.
4
1
2
Enter
18
7529_eng.fm Page 19 Wednesday, November 12, 2003 8:57 AM
To set up a Characteristic - ETA5 ONLY
To set up a job
To start, press
press
ON
MENU
plug in transducer
press (Configure)
press (Edit Char)
Type in Job Name ______________________ , press
1
1
Enter
Alternatively, when the cursor is flashing, you may connect a Bar Code Reader to the Printer/PC
connector on the back of the ETA5 and scan the information from a barcode. Scan across the
label (from left or right) and watch the display. If the flashing cursor disappears, the reading has
been captured; wait a few seconds for the reading to be displayed. If not, re-scan until the data
is accepted.
When the information is displayed on the screen, press
Type in number of samples __________________ , press
Type in number of subgroups ________________ , press
Select transducer ID using , press
Enter
Enter
Enter
Enter
Select units of measure using , press
Enter
Select ‘Peak’, ‘Impulse’, or ‘Click’ using; then press
Select measurement direction using ; then press
Select cycle end time using ; then press
19
Enter
Enter
Enter
Loading...
+ 46 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.