IMPACT 326, 326M User manual

4 (3)

6515-01-435-0050

IMPACT Suction Apparatus, Model 326/326M

Impact Instrumentation Inc.

27 Fairfield Place

West Caldwell NJ 07006-6206 (973)882-1212

The following is a list of TMDE required for the complete PMCS of this suction apparatus:

TMDE Item

 

Variable DC Power Supply

Oscilloscope (if calibration is required)

Calibrated stopwatch

Cross-tip screwdriver (if cal required)

RT200, VT Plus, T300, or equivalent

 

FORMS REQUIRED

1. This item does not require a CVC label 2163. 2.

Item may also need extra accessories sent with it. See the contract for specific item or see shop supervisor for details.

PMCS CHECKLIST

VISUAL CHECKS

page 11-1 and 11-2

Perform steps 1-15 as outlined in the manual on the pages above.

Verify revision of literature provided with the aspirator.

Inspect the transporting case for damage if provided.

Inspect the AC power supply case for cracks or damage also.

Verify all accessories provided with item. See page -v- for checklist

PERFORMANCE CHECKS page 11-2

1.Maximum Flow

2.Rate of Vacuum rise.

3.Maximum Vacuum-See attached altitude VS pressure chart for acceptable tolerances.

4.Vacuum Gauge accuracy

Be sure to charge batteries before packing the aspirator to optimize performance of the unit for receiving personnel. Simply connect battery to charging circuit and plug in the AC power supply to proper voltage.

TIMER VERIFICATION (ADDED VERIFICATION)

see pages 9-1 for calibration if needed.

1.Turn on aspirator to intermittent operation.

2.Turn timer knobs to 5 seconds on and off. Verify with stopwatch that the aspirator operates within correct times.

3.Turn timer knobs to 20 seconds on and off. Verify proper operation.

4.Turn timer knobs to 45 seconds on and off. Verify proper operation.