Page
Control panel and display 02 Contents 09 Declaration of conformity 09 Warranty 09 Explication of warning symbols 09 Safety instructions 10 Unpacking 11 Correct use 11 Commissioning 11 Safe temperature limit 11
Setting the safe temperature limit 11 Stirring function 11 Eco mode 12 Operating modes 12 Direction of rotation reversal 12 Function heating 12 Maintenance and cleaning 12 Error codes (RT 5/10/15) 13 Accessories 13 Technical data 14
EN
Declaration of conformity
Warranty
In accordance with IKA® warranty conditions, the warranty period is 24 months. For claims under the warranty please contact your local dealer. You may also send the machine direct to our factory, enclosing the delivery invoice and giving reasons for the claim. You will be liable for freight costs. The warranty does not cover worn out parts, nor does it apply to faults resulting from improper use, insucient care or maintenance not carried out in accordance with the instructions in this operating manual.
Explication of warning symbols
General hazard.
DANGER
WARNING
CAUTION
This symbol identifies informaton that is of absolute importance to ensure health and safety. Failure to observe this information may be detrimental to health or may result in injuries.
This symbol indicates information witch is important for ensuring that the appliance functions without any technical problems. Failure to observe this information could damage the appliance.
This symbol indicates information which is important for proper use of the appliance and / or
ensuring that the appliance functions correctly.
Failure to observe this information can lead to inaccurate results.
ATTENTION
DANGER
ATTENTION - Risk of damage due to magnetism.
DANGER - Reference to the endangerment by a hot surface
9
Safety instructions
For your protection
Read the operating instructions in full before starting
up and follow the safety instructions.
• Keep the operating instructions in a place where they can be accessed by everyone.
• Ensure that only trained staff work with the appliance.
• Follow the safety instructions, guidlines, occupational health and safety and and accident prevention regulations.
• Socket must be earthed (protective ground contact). Attention - Magnetism!
ATTENTION
Effects of the magneticfield have to be into account (e.g. data storage media, cardiac pacemakers ...). RT
DANGER
Risk of burns! Exercise caution when touching the
housing parts and the heating plate. The heating plate can reach temperatures in excess of 125 °C. Pay attention to the residual heat after switching off. RT
WARNING
Ensure that the mains power supply cable does not touch the heating base plate.
• Wear your personal protective equipment in accordance with the hazard category of the media to be processed. Otherwise there is a risk from:
- splashing and evaporati on of liquids
- ejection of parts
- release of toxic or combustable gases.
• Set up the appliance in a spacious area on an even, stable, clean, non-slip, dry and fireproof surface.
• The feet of the appliance must be clean and undamaged.
• Check the appliance and accessories beforehand for damage each time you use them. Do not use damaged components.
• Gradually increase the speed.
• Reduce the speed if
- the medium splashes out of the vessel because the speed is too high
- the appliance is not running smoothly
- the container moves on the base plate.
RT
WARNING
Caution! Only process and heat up any media that has a flash point higher than 180 °C. The safe temperatue limit must always be set to at least 25 °C lower than the fire point of the media used. Beware of hazards due to:
WARNING
- flammable materials
- combustibles media with a low boiling temperature
- glass breakage
- inkorrect container size
- overfilling of media
- unsafe condition of container.
• The appliance may heat up when in use.
• The base plate can heat up due to the action of the drive mag­ nets at high motor speeds, even if the heater is not operational.
• Process pathogenic materials only in closed vessels under a suitable extractor hood. Please contact IKA questions. Do not operate the appliance in explo-
DANGER
sive atmospheres, with hazardous substances or under water. Only process media that will not react
DANGER
dangerously to the extra energy produ­ ced through processing. This also ap­ plies to any extra energy produced in other ways, e.g. through light irridiation.
• Please observe the operating instructions for any accessories used.
• Safe operation is only guaranteed with the accessories descri­ bed in the „Accessories“ chapter.
• Accessories must be securely attached to the device and cannot com off by themselves. The centre of gravity of the assembly must lie within the surface on which it is set up.
• Always disconnect the plug before fitting accessories.
• The appliance can only be disconnected from the mains supply by pulling out the mains plug or the connector plug.
• The socket for the mains cord must be easily accessible.
• It may be possible for wear debris from rotating accessory parts to reach the material being processed.
• When using PTFE-coated magnetic bars, the following has to be noted: Chemical reactions of PTFE occur in contact with molten or
solute alkali metals and alkaline earth metals, as well as with fine powders of metals in groups 2 and 3 of the periodic sys­ tem at temperatures above 300 °C - 400 °C. Only elementary fluorine, chlorotrifluoride and alkalimetals attack it; halogena­ ted hydrocarbons have a reversible swelling effect.
For protection of the equipment
• The appliance may only be opened by experts.
• The voltage stated on the type plate must correspond to the mains voltage.
• Do not cover the device, even partially e.g. with metallic plates or film. This results in overheating.
• Protect the appliance and accessories from bumps and impacts.
• Ensure that the base plate is kept clean.
• Observe the minimum distances between the devices, between device and wall and minimum distances (min. 800 mm) above the assembly, see Fig. 2.
(Source: Römpps Chemie-Lexikon and „Ullmann“ Volume 19)
>100 mm
>100 mm
>100 mm
®
if you have any
Fig. 2