Item Designation
1 Operator panel and display
2 Mains switch
3 Power socket
4 Buoyage
5 Heater
6 Stand
7 Pt 100 + Pt 1000 temperature sensor
8 Handle
9 Fan
10 RS 232 port
11 USB port
12 Clamp
13 Cable clip
13
9
6
Fig. 2
2
10
11
12
7
2
Page 3
Source language: German
EN
Contents
Page
Declaration of Conformity ............................................................................................................................4
Explication of warning symbols ...................................................................................................................4
General information ............................................................................................................................................. 5
Correct use ...................................................................................................................................................7
Use ...................................................................................................................................................................... 7
Range of use (indoor use only) ............................................................................................................................. 7
Setting up ............................................................................................................................................................ 7
Filling and draining ............................................................................................................................................... 7
Setting the safety temperature .................................................................................................................. 10
Menu settings ......................................................................................................................................... .. 11
Menu structure .................................................................................................................................................. 11
Basic guide to setting the menu options ............................................................................................................. 11
Maximum temperature (HI T) ............................................................................................................................. 11
Minimum temperature (LO T) ............................................................................................................................. 11
Maximum speed (HI R) ....................................................................................................................................... 11
Minimum speed (LO R) ....................................................................................................................................... 11
Fluid type (FLUI) ................................................................................................................................................. 11
The type of temperature PID control arithmetic (AUTO) ...................................................................................... 11
Alarm and key tone (BEEP) ................................................................................................................................. 12
Interface and output ..................................................................................................................................13
Maintenance and cleaning .........................................................................................................................15
Technical data ............................................................................................................................................17
We declare under our sole responsibility that this product corresponds to the regulations 2014/35/EU, 2006/42/EC, 2014/30/EU
and 2011/65/EU and conforms with the standards or standardized documents: EN 61010-1, EN 61010-2-010, EN 61010-2-051,
EN 61326-1, EN 60529, DIN 12876-1 und EN ISO 12100.
Explication of warning symbols
Indicates an (extremly) hazardous situation, which, if not avoided, will result in death,
DANGER
serious injury.
Indicates a hazardous situation, which, if not avoided, can result in death, serious injury.
WARNING
Indicates a potentially hazardous situation, which, if not avoided, can result in injury.
CAUTION
Indicates practices which, if not avoided, can result in equipment damage.
NOTICE
Indicates a hazardous situation that cause from a hot surface!
CAUTION
Hot surface!
4
Page 5
Safety instructions
General information:
• Read the operating instructions completely before
starting up and follow the safety instructions.
• Keep the operating instructions in a place where it can be
accessed by everyone.
• Ensure that only trained staff work with the device.
• Follow the safety instructions, guidelines, occupational
health, safety and accident prevention regulations.
• Set up the device in a spacious area on an even, stable,
clean, non-slip, dry and fireproof surface.
• Do not operate the device in explosive atmospheres,
with hazardous substances.
• Protect the device and accessories from bumping and
impacting.
• Check the device and accessories for damage before
each use. Do not use damaged components.
• Safe operation is only guaranteed with the accessories
described in the ”Accessories” section.
• The socket for the mains cord must be easily accessible.
• Socket must be earthed (protective ground contact).
• The voltage stated on the type plate must correspond to
the mains voltage.
• The device can only be disconnected from the mains supply by pulling out the mains plug or the connector plug.
• Disconnect the mains plug before attaching or changing
any accessories.
• Disconnect the mains plug before cleaning, maintenance
and transportation of the unit.
• The device must only be opened by trained specialists,
even during repair. The device must be unplugged from
the power supply before opening. Live parts inside the
device may still be live for some time after unplugging
from the power supply.
NOTICE
be put back on the device again to ensure safe operation, for example to keep foreign objects and liquids,
etc. from getting into the device.
• The device may only be used as prescribed and as described in the operating instructions. This includes operation by instructed specialist personnel.
• When using critical or hazardous materials in your processes, IKA
measures to ensure safety in the experiment. For example, users can implement measures that inhibit fire or
explosions or comprehensive monitoring equipment.
• Process pathogenic material only in closed vessels under
a suitable fume hood. Please contact IKA
support if you have any question.
®
recommends to use additional appropriate
Coverings or parts that can be
removed from the device without the aid of any tools must
®
application
WARNING
• A laboratory circulator heats and circulates fluid according to specified parameters. This involves hazards due to
high temperatures, fire and general hazards due to the
device of electrical energy. The user is largely protected by the device of relevant standards. Further hazard
sources may arise due to the type of tempering fluid,
e.g. by exceeding or undercutting certain temperature
thresholds or by the breakage of the container and reaction with the heat carrier fluid. It is not possible to
consider all eventualities. They remain largely subject to
the judgment and responsibility of the operator. For this
reason, it may become necessary for user to take other
precautionary safety measures.
• Insucient ventilation may result in the formation of explosive mixtures. Only use the device in well ventilated areas.
WARNING
temperature cannot be exceeded even in the event of a
fault. Check the safe temperature circuit on a regular basis (see section “Setting the safety temperature“) .
• Securely fix the
bath, so that it cannot tip over.
• When device is used for external circulation, extra precaution must be taken for hot fluid leakage due to damaged hose.
• Use suitable hoses for connection.
• Secure hoses and tubes against slippage and avoid kinks.
• Check hoses, tubes and bath at regular intervals for possible material fatigue (cracks/leaks).
• Mains cable should not get in contact with hot parts
and fluids.
• If you are using plastic bath, observe the permitted
working temperature range and fluids.
ICC immersion circulator for use at the
If the OFF switch is not within
reach when device is operating,
an EMERGENCY STOP switch
that can be easily accessed must
be installed in the work area.
The safety circuit (safe temperature) must be adjusted so
that the maximum permissible
Do not start up the device if:
DANGER
- It is damaged or leaking
-
Cable (not only supply cable) is damaged.
• Be careful when filling a hot bath.
WARNING
At high operating temperatures, the temperature of housing parts, surfaces and tubes
can exceed 70 ° C.
5
Page 6
WARNING
• After a power failure during operation, the device may
start automatically (depending on operating mode).
• Transport the device with care.
• Do not transport or empty the bath while it is still hot.
This may result in accidents, especially scalding injuries.
NOTICE
cured with the cable clamp (13, see Fig. 2 and Fig. 4).
Fluids:
WARNING
WARNING
aware of the chemical hazards that may be associated
with the bath fluid used. Observe all safety warnings for
the fluids.
• Depending on the bath fluid used and the type of operation, toxic or flammable vapors can arise. Ensure suitable extraction.
• Do not use any fluid which may cause dangerous reactions during processing.
• Only use recommended bath fluid. Only use non-acid
and non corroding fluid.
WARNING
perature limit that has been set. The safe temperature
limit must always be set to at least 25 ºC lower than the
flash point of the fluid used. Examine regularly the function of the safety temperature limiter.
NOTICE
ensure that the minimum clearances and immersion
depths in the fluid are observed. Check the fluid level
detection at a regular basis (see section "Filling and draining").
It is dangerous to touch the
heater. The temperature of the
heater can be very high.
In order to prevent the power
cable from falling into the fluid,
the mains cable must be se-
Beware of the risk of burning
due to delay in boiling!
Only use fluids, which fulfill the
requirements for safety, health
and device compatibility. Be
Only process and heat up any
fluid that has a flash point higher than the adjusted safe tem-
Never operate the device without sucient heat carrier fluid!
You should also be careful to
• Continuous monitoring of the bath and the filling level of
the bath fluid is required, especially at high temperatures.
• For optimum temperature stability, the fluids viscosity
should be 50 mm
perature. This permits good fluid circulation and minimizes heating from the pump.
NOTICE
• Untreated tap water is not recommended. It is recommended to use distilled water or high purity water (ion
exchangers) and add 0.1 g soda (sodium carbonate
Na
) /liter, to reduce corrosive properties.
2CO3
WARNING
NOTICE
NOTICE
2
/s or less at its lowest operating tem-
If water is used at higher temperature, there is heavy loss of
fluid due to the evaporation.
Risk of burning caused by vapor or hot water at the outlet
of the cooling coil (accessory).
Do not use the cooling coil with
water at bath temperatures >
95° C.
For bath temperatures > 60°C
make sure that the flow rate
through the cooling coil is high
enough.
Don't use following fluids:
NOTICE
- Untreated tap water
- Acids or bases
- Solutions with halides: chlorides, fluorides, bromides,
iodides or sulfur
- Bleach (Sodium Hypochlorite)
- Solution with chromates or chromium salts
- Glycerine
- Ferrous water.
NOTICE
100 °C, remove the remaining water from the complete
system
(including hoses and external devices). When do-
ing this, also open the stopper and union nuts caps of
the pump outputs and inputs and blow compressed air
through all the pump outputs and inputs!
When changing the bath fluid
from water to a heat transfer
fluid for temperatures above
6
Page 7
Correct use
Use:
Use ICC (Immersion Circulator Compact) for heating and circulating fluids.
Intended Use: Tabletop device
Range of use (indoor use only):
- Laboratories - Schools
- Pharmacies - Universities
Unpacking
Unpacking:
- Please unpack the device carefully
- In the case of any damage a detailed report must be sent
immediately (post, rail or forwarder).
NOTICE
Transport safety
Remove the transport protection under the
buoyage
(4).
This device is suitable for use in all areas except:
- Residential areas.
- Areas that are connected directly to a low-voltage supply
network that also supplies residential areas.
The safety of the user cannot be guaranteed:
- If the device is operated with accessories that are not supplied or recommended by the IKA
- If the device is operated improperly or in contrary to the
®
IKA
specifications.
- If the device or the printed circuit board are modified by
third parties.
Delivery scope:
- ICC baisc
- Mains cables
- Screwdriver (use for safety circuit)
- Operating instructions
- Warranty card.
®
.
Preparations
Setting up:
- Securely fix the immersion circulator on the bath.
- Place the device on an even, stable, clean, nonslip, dry
and fireproof surface.
- Keep at least 20 cm of open space at the front and rear
side.
- When a plastic bath is used, please ensure that the heater
does not contact the bath.
- The place for installation should be large enough and provide sucient air ventilation to ensure the room does not
warm up excessively because of the heat from device
radiates to the environment.
- Do not set up the device in the immediate vicinity of heat
sources and do not expose to sun light.
Filling and draining:
- Connect the mains plug and turn on the device with
mains switch (2).
- The low level warning message appear on the display.
- Fill fluid to the bath.
Note: Pay attention to the fluid level information on the
display:
Low Level
High Level
For draining, please refer to the bath operating instructions.
7
Page 8
Fluids (Standard information for IKA fluid):
No.
0
1
2
3
4
5
6
7
8
®
IKA
Designation
CF.EG28.N10.80.8
CF.EG39.N20.80.16
CF.EG44.N25.80.19
CF.EG48.N30.80.22
UF.Si.N30.150.10LV
HF.Si.20.200.50
HF.Si.20.250.50A
Water o
Customized p
Viscosiy at 25 °C for Heating Fluid (HF)
Viscosity at -20 °C for Cooling Fluid (CF)
Viscosity at 25 °C for Universal Fluid (UF)
Dynamic viscosity [mPa.s] is a product of kinematic viscosity [mm²/s] and density [kg/m³] of the fluid divided by 1000.
(5) Additional information:
A: Oil Additives
LV: Low Viscosity
n
Note: For open bath application!
o
Note: Tap water may be unsuitable for operation because the calcium carbonate content may cause calcification.
High purity water (from icon exchangers) and distilled or bi-distilled water are unsuitable for operation due to corrosive
properties of these media. High purity water and distillates are suitable as a medium after adding 0.1 g soda (Na
sodium carbonate) per liter of water.
o
Note: The temperature limit values are adjustable in accordance with the fluid used.
2CO3
,
8
Page 9
Operator panel and display
Fig. 3
M
B
A
L
F
C
JKO
GHIDE
Item Designation Function
A On/Off button: Switch on/off the circulator.
B Adjustable safety circuit: Adjust the safety temperature limit with delivered screwdriver.
C LED display:
D “Temp” button:
E “Temp (+)” button:
F “Temp (-)” button:
G “OK/Pump” button:
H “Pump (+)” button:
I “Pump (-)” button:
J “Menu“ button:
K
LED, key button: Indicate that the function of keys are deactivated.
LED, heater:
L
M LED, set: The LED lights up at the same time as the set value is displayed.
O Key button:
Display the settings and actual values.
Start/stop the heating function.
Increase the temperature setting value.
Navigation, selecting the settings in the menu.
Decrease the temperature setting value.
Navigation, selecting the settings in the menu.
Start/stop the pump function.
Confirm the menu settings.
Increase the pump speed setting value.
Change the value setting in the menu.
Decrease the pump speed setting value.
Change the value setting in the menu.
Press it once: menu option is display.
Press it a second time: back to the working screen.
Indicate the heating function is activated.
Lock/unlock keys.
9
Page 10
Commissioning
Check whether the voltage specified on the type plate
matches the mains voltage available.
The power socket used must be
earthed (protective earthing).
If above conditions are met, the device is ready for operation after plugging in the mains plug.
If these conditions are not met, safe operation is not guaranteed and the device could be damaged.
Observe the ambient conditions (temperature, humidity,
etc.) listed under “Technical Data”.
After switching on the mains switch (2) at the back of the
device or pressing the power key (A), all LED segments
light up during the self test.
c
Starting
d
Software version (ver)
e
Operating mode (mode)
f
Safety temperature (safe)
g
Working setting
After that, following two screens appear alternatively.
h
i
Then the device enters standby status and is ready for operation.
Change the temperature setting with “Temp (+)“ button
(E) or “Temp (-)“ button (F).
Change the pump speed setting with “Pump (+)“ button
(H) or “Pump (-)“ button (I).
In standby status, activate the heating function by pressing
the “Temp“ button (D). The device start heating function,
meanwhile the pump start to run.
In working status, press the “OK/Pump“ button (G) to stop
the pump function. The heating function and pump stops.
Note: In running status, the pump speed adjusts automatically according to the load. When the actual speed
doesn’t reach the setting speed, the actual speed and setting speed will appear on the screen alternately.
In standby status, press the “OK/Pump“ button (G) to
start the pump function. The heating function will not be
activated.
In working status, press the “Temp“ button (D) to stop
the heating function, the pump keep running.
The control elements of the device allow to be locked by
pressing the key button (O), so no accidental changes during operation are possible (LED (K) lights up).
By pressing the key button (O) again, the controls are released (LED (K) no longer lights up).
Setting the safe temperature
Adjust the safe temperature with screwdriver delivered
with the device.
Fig. 4
The safe temperature setting will appear on the display.
Safety temperature (safe)
Factory setting: approximate max. value
Adjustment range: 0 – 160 °C
Warning: The safe temperature must always be set to at
least 25 ºC lower than the flash point of the fluid used.
10
Page 11
• Menu structure:
MODE
Menu settings
A
B
C
Default settings
activated
-
-
Maximum temperature (HI T)
Minimum temperature (LO T)
Maximum speed (HI R)
Minimum speed (LO R)
Fluid type (FLUI)
Menu
The type of temperature PID control arithmetic (AUTO)
Alarm and key tone (BEEP)
Calibration (CALI)
• Basic guide to setting the menu options:
) Enter the menu by pressing ”Menu” button (J).
) Press the
(F)
)
Press the“Pump (+)“ button (H) or “Pump (-)“ but-
“Temp (+)“ button (E) or “Temp (-)“ button
to change the menu options.
ton (I) to change the value settings in the menu.
) Confirm the menu settings by pressing the
button (G)
.
“OK/Pump“
AUTO 1
AUTO 0
BEEP 0
BEEP 1
CALI 0
CALI 2
CALI 3
Proportional coecient of PID (Kp 1)
Integrate time of PID (Ti 1)
Differential time of PID (Td 1)
• Minimum temperature (LO T):
The minimum adjustable value: -20° C
• Maximum speed (HI R):
150 °C
-20 °C
3200 rpm
1000 rpm
8
activated
15.0
5.00
5.0
activated
activated
-
-
• Operating mode (MODE):
Mode A: After power on/power failure no automatic
restart of functions.
Mode B: After power on/power failure automatic restart of functions, depending on previous settings.
Mode
C:
Set values (set in A or B) cannot be changed.
After power on/power failure automatic restart of functions, depending on previous settings.
• Maximum temperature (HI T):
The maximum adjustable value: 150° C. This value can
be limited additionally by the selected fluids (see section
“Fluids”).
Note: The maximum value is limited by the set safe
temperature.
The maximum adjustable value: 3200
rpm.
• Minimum speed (LO R):
The minimum adjustable value: 1000 rpm
• Fluid type (FLUI):
The chosen fluid (No.) of the temperature setting is limited. See Section “Fluids”.
• Temperature control type (AUTO):
11
Page 12
AUTO 1:
AUTO 1 is the default setting. The optimum settings are
determined automatically.
Proportional coecient of PID (Kp 1):
Heat-up curve in auto-tuning control (AUTO 1):
°C
Fig. 5
t
AUTO 0:
For special requirements, AUTO 0 can be used with
manual adjustment of the control parameters.
If the PID control (AUTO 0) option is selected, the following items including Kp 1, Ti 1 and Td 1 could be
set. Otherwise they are not displayed in the menu list.
Note: Inappropriate settings may produce the following heat-up curves:
°C
°C
Kp too high
t
Ti too low
The proportional coecient Kp is the controller amplification and determines how strongly the control deviation (the difference between the target temperature and
actual temperature) directly affects the control variable
(on-time of the heater). Kp values that are too large can
lead to the controller overshooting.
Integral time of PID (Ti 1):
The integral time Ti (s) is the correction time and determines how strongly the duration of the control deviation affects the control variable. Ti compensates an existing control deviation. A high Ti means a smaller and
slower effect on the control variable. Ti values that are
too small can lead to instability of the controller.
Differential time of PID (Td 1):
The differential time Td (s) is the derivative time and determines how strongly the rate of change of the control
deviation affects the control variable. Td compensates
for rapid control deviations. A high Td means a smaller
and slower effect on the control variable. Td values that
are too small can lead to instability of the controller.
°C
°C
Fig. 6
t
Kp too low
Ti too high
t
Td too high
t
•
Alarm and key tone (BEEP):
BEEP 0: no alarm and key tone
BEEP 1: alarm and key tone is activated
Dip the temperature sensor of the reference measuring
instrument into the bath fluid:
Select 2-point calibration in the menu. Press the “OK/Pump“ button (G) to start the 2-point calibration.
12
Page 13
Set the first point temperature (e.g. 80 °C). Confirm the
setting by pressing the “OK/Pump“ button (G).
Left area of the display appears the set value (80 °C) and
the right area of the display appears the actual temperature value measured. The set LED (M) flashes briefly.
The unit now starts and controls to the set value. When
the temperature has reached the set value and has become constant, the “Set” LED (M) no longer flashes and
the following display appears.
The temperature measured by the unit appears on the
left hand side.
Input the calibration value from the reference mesuring
instrument (e.g. 80.2 °C) with the “Pump (+)“ button
(H) or “Pump (-)“ button (I).
Confirm the value by pressing the “OK/Pump“ button (G).
Back to previous screen for input a new value by pressing the “Temp“ button (D).
BackOK
The first point calibration is finished now.
Calibration of the other points is performed in the same
way.
Interface and output
The device can be operated in “Remote” mode via the RS 232
interface (10) or the USB interface (11) connected to a PC
and with the laboratory software Labworldsoft
Note: Please comply with the system requirements together with the operating instructions and help section
included with the software.
USB interface:
The Universal Serial Bus (USB) is a serial bus for connecting
the device to the PC. Equipped with USB devices can be connected to a PC during operation (hot plugging). Connected
devices and their properties are automatically recognized.
The USB interface can also be used to update firmware.
USB device drivers:
First, download the latest driver for IKA
interface from:
http://www.ika.com/ika/lws/download/usb-driver.zip.
Install the driver by running the setup file. Then connect
the IKA
®
device through the USB data cable to the PC.
The data communication is via a virtual COM port. Configuration, command syntax and commands of the virtual
COM ports are as described in RS 232 interface.
RS 232 interface:
Configuration
- The functions of the interface connections between the
stirrer machine and the automation system are chosen
from the signals specified in EIA standard RS 232 in accordance with DIN 66 020 Part 1.
-
For the electrical characteristics of the interface and the allocation of signal status, standard RS 232 applies in accordance with DIN 66 259 Part 1.
®
.
®
devices with USB
- Transmission procedure: asynchronous character transmission in start-stop mode.
- Type of transmission: full duplex.
-
Character format: character representation in accordance
with data format in DIN 66 022 for start-stop mode. 1
start bit; 7 character bits; 1 parity bit (even); 1 stop bit.
- Transmission speed: 9600 bit/s.
- Data flow control: none
- Access procedure: data transfer from the stirrer machine to
the computer takes place only at the computer’s request.
Command syntax and format:
The following applies to the command set:
- Commands are generally sent from the computer (Master)
to the device (Slave).
- The device sends only at the computer’s request. Even
fault indications cannot be sent spontaneously from the
device to the computer (automation system).
- Commands are transmitted in capital letters.
- Commands and parameters including successive parameters are separated by at least one space (Code: hex 0x20).
- Each individual command (incl. parameters and data)
and each response are terminated with Blank CR LF
(Code: hex 0x20 hex 0x0d hex 0x20 hex 0x0A) and have
a maximum length of 80 characters.
- The decimal separator in a number is a dot (Code: hex
0x2E).
The above details correspond as far as possible to the recommendations of the NAMUR working party (NAMUR recommendations for the design of electrical plug connections
for analogue and digital signal transmission on individual
items of laboratory control equipment, rev. 1.1).
13
Page 14
The NAMUR commands and the additional specific IKA® co mmands serve only as low level commands for communication
between the device and the PC. With a suitable terminal or communications programme these commands can be transmit-
®
ted directly to the circulator equipment. The IKA
software package, Labworldsoft®, provides a convenient tool for controlling circulating equipment and collecting data under MS Windows, and includes graphical entry features, for pump motor
speed ramps for example.
Read the internal actual temperature
Read the safety actual temperature
Read the pump actual speed
Read the internal setting temperature
Read the safety setting temperature
Read the pump setting speed
Set the internal setting temperature XXX
Set the WD safety temperature with echo of the set (defined) value.
Set the pump speed XXX
Set the WD-safety speed with echo of the set (defined) value.
Start the watchdog mode 1 and set the time for the watchdog to n (20...1500) seconds.
Echo of the Watchdog time.
During a WD1-event, the heating and pump functions are switched off.
This command needs to be send within the watchdog time.
OUT_WD2@n
Start the watchdog mode 2 and set the watchdog time to n (20...1500) seconds.
Echo of the watchdog time.
During a WD2-event, the set temperature is changed to the WD safety temperature and the pump set
speed is set to the WD safety speed.
This command needs to be send within the watchdog time.
RESET
START_1
START_4
STOP_1
STOP_4
Reset the PC control and stop the device functions.
Start the heating function
Start the pump function
Stop the heating function
Stop the pump function
PC 1.1 Cable:
This cable is required to connect RS 232 port (10) to a PC.
6
7
8
9
Fig. 7
1
2
3
4
5
RxD 2
TxD 3
GND 5
RTS 7
CTS 8
1
4
6
9
USB cable A - Micro B 2.0:
This cable is required to connect USB port (11) to a PC.
USB micro B
Fig. 8
5
4
3
2
1
1
2 RxD
3 TxD
4
5 GND
6
7 RTS
8 CTS
9
9
8
7
6
PC
A
USB A
14
Page 15
Maintenance and cleaning
The device is maintenance-free. It is only subject to the
natural wear and tear of components and their statistical
failure rate.
Cleaning:
For cleaning disconnect the mains
plug!
Use only cleaning agents which have been approved by
®
IKA
to clean IKA® devices.
DirtCleaning agent
DyeIsopropyl alcohol
Construction material
Cosmetics
FoodstuffWater containing tenside
FuelWater containing tenside
Water containing tenside/ isopropyl alcohol
Water containing tenside/ isopropyl alcohol
For materials which are not listed, please request
information from IKA
®
application support.
Wear protective gloves while cleaning the device.
Electrical devices may not be placed in the cleansing agent
for the purpose of cleaning.
Do not allow moisture to get into the device when cleaning.
Before using another than the recommended method for
cleaning or decontamination, the user must ascertain with
®
IKA
that this method does not damage the device.
Spare parts order:
When ordering spare parts, please give:
- machine type
- manufacturing number, see type plate
item and designation of the spare part,
see www.ika.com, spare parts diagram and list
- Software version.
Repair:
Please send the device for repair only after it has
been cleaned and is free from any materials which
may constitute a health hazard.
For repair, please request the “Decontamination Certificate” from IKA
®
IKA
website www.ika.com.
®
, or download printout of it from the
If you require servicing, return the device in its original
packaging. Storage packaging is not sucient. Please also
use suitable transport packaging.
Error codes
Any malfunctions during operation will be identified by an error message on the display.
Proceed as follows in such cases:
)
Switch off device using the main switch at the back of the device
)
Carry out corrective measures
)
Restart device
Error codeEffectCauseSolution
Err 02Pump off
Heating off
Err 03Pump off
Heating off
Err 04Pump off
Heating off
Err 05Pump off
Heating off
Err 06Pump off
Heating off
Err 07Pump off
Heating off
Err 08Pump off
Heating off
Motor over current (rate current)- Reduce pump motor speed
- Use fluid with lower viscosity
- Check if the pump impeller is blocked
Motor over current (Max. current)- Reduce pump motor speed
- Use fluid with lower viscosity
- Check if the pump impeller is blocked
Motor hall signal missing- Reduce pump motor speed
- Use fluid with lower viscosity
- Check if the pump impeller is blocked
Too high liquid level- Check the liquid level and buoyage
Too low liquid level- Check the liquid level and buoyage
Too high voltage- Check the mains power
Too low voltage- Check the mains power
15
Page 16
Err 09Pump off
Heating off
Err 10Pump off
Heating off
Err 11Pump off
Heating off
Err 12Pump off
Heating off
Err 13Pump off
Heating off
Err 14Pump off
Heating off
If the actions described fails to resolve the fault or another error code is displayed then take one of the following steps:
- Contact the service department;
- Send the device for repair, including a short description of the fault.
Device internal temperature is too
high
PC communication failure
Temperature difference between
control sensor and safety sensor is
too much
Safety temperature alarm
Heater switched off by safety
circuit
Fan error- Check the fan and clean the grids at the rear side
- Check communication cable
- Check the ambient temperature and let the device cool down
Bridges and Covers:
BS.ICC Small bridge (for IB eco 8, IB pro 9)
BL.ICC Large bridge (for IB eco 18, IB pro 12,
CS.ICC Samll cover (for IB eco 8, IB pro 9)
CM.ICC Medium cover (for IB pro 12)
CL.ICC Large cover (for IB eco 18, IB pro 20)
•
Additional accessories:
PCS.ICC Pump set
PC 1.1 Cable (RS 232)
USB 2.0 cabel A - Micro B
Labworldsoft
See more accessories on www.ika.com.
®
IB pro 20)
16
Page 17
Technical data
Operating voltage
Frequency
Max. input power
Working temperature range (RT+10 °C at 1000 rpm)
Operating temperature range (with forced cooling)
Temperature stability – Internal temperature
control 70 °C, water (according to DIN12876)
Temperature control
Temperature measurement, absolute accuracy
Temperature setting
Temperature setting resolution
Temperature display
Temperature display resolution
Classification according to DIN12876-1
Safety circuit (adjustable)
Safety temperature display
Heating capacity
Pump speed (adjustable)
Max. pump pressure / suction
Max. flow rate (at 0 bar)
Sub-level protection
Interface
Permitted on-time
Protection class according to EN 60529
Protection class
Excess voltage category
Contamination level
Permitted ambient temperature
Permitted ambient humidity
Dimension (W x D x H)
Weight
Operation at a terrestrial altitude
VAC230 ± 10 %
115 ±10 %
100 ± 10 %
Hz50/60
2100 (230 VAC)
W
°CRT + 10 ... 150
°C- 20 ... 150
K± 0.02
K± 0.2
K0.1
K0.1
°C0 ... 160
W
rpm1000 ... 3200
bar0.3 / 0.2
l/min18
%100
°C+ 5 ... + 40
%80
mm145 x 200 x 340
kg3.75
mmax. 2000
1100 (115 VAC)
860 (100 VAC)
PID
Button
LED
Class III (FL) suitable for flammable and non-flammable fluids
LED
2000 (230 VAC)
1000 (115 VAC)
760 (100 VAC)
Yes
USB, RS 232
IP 21
I
II
2
Note: Complies to EN61000-3-11 subject to conditional connection: Zmax = 0.282 Ω. If necessary, consult your electricity supplier.
Subject to technical changes!
17
Page 18
Warranty
In accordance with IKA® warranty conditions, the warranty
period is 24 months. For claims under the warranty please
contact your local dealer. You may also send the machine
direct to our factory, enclosing the delivery invoice and giving reasons for the claim. You will be liable for freight costs.
The warranty does not cover worn out parts, nor does
it apply to faults resulting from improper use, insucient
care or maintenance not carried out in accordance with
the instructions in this operating manual.
Pump performance curve
Pump performance curve measured with water:
(Measurements done according DIN 12876-2 with water at 20°C; pump in a closed-loop circuit).
300
200
100
ΔP [mbar]
-100
-200
-300
Fig. 9
0
3200 rpm
2000 rpm
1000 rpm
05101520
Q [l/min]
18
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.