IKA HRC 2 basic Operating Instructions Manual

Page 1
20000015724a
HRC 2 basic_112017
IKA® HRC 2 basic
Operating instructions EN
Page 2
Device setup
1
2
3
2
Fig. 1
Item Designation 1 Operator panel and display 2 Handle 3 Venting grid 4 Filling opening lid 5 RS 232 port 6 USB port 7 Overflow 8 Backflow 9 Power socket 10 Power switch 11 Pump connector OUT 12 Pump connector IN
4
5 6
12 11
10 9
8 7
2
3
2
2
Page 3
Source language: German
EN
Contents
Page
Device setup ................................................................................................................................................. 2
Declaration of Conformity ............................................................................................................................ 4
Explication of warning symbols ................................................................................................................... 4
Safety instructions ....................................................................................................................................... 5
General information ����������������������������������������������������������������������������������������������������������������������������������������������5
Disposal of the device �������������������������������������������������������������������������������������������������������������������������������������������6 Fluids ��������������������������������������������������������������������������������������������������������������������������������������������������������������������6
Correct use ................................................................................................................................................... 6
Use �����������������������������������������������������������������������������������������������������������������������������������������������������������������������6 Range of use (indoor use only) ������������������������������������������������������������������������������������������������������������������������������6
Unpacking .................................................................................................................................................... 7
Unpacking ������������������������������������������������������������������������������������������������������������������������������������������������������������7 Delivery scope �������������������������������������������������������������������������������������������������������������������������������������������������������7
Preparations ................................................................................................................................................. 7
Setting up �������������������������������������������������������������������������������������������������������������������������������������������������������������7 Connecting the tubings �����������������������������������������������������������������������������������������������������������������������������������������7 Filling and draining ������������������������������������������������������������������������������������������������������������������������������������������������8 Fluid (Standard information for IKA® fluid) ������������������������������������������������������������������������������������������������������������9 Moving the device ����������������������������������������������������������������������������������������������������������������������������������������������10
Operator panel and display ........................................................................................................................ 10
Setting the safety temperature .................................................................................................................. 11
Commissioning ........................................................................................................................................... 11
Menu settings ............................................................................................................................................ 12
Menu structure ���������������������������������������������������������������������������������������������������������������������������������������������������12 Basic guide to setting the menu options ��������������������������������������������������������������������������������������������������������������12 Operating mode (MODE)�������������������������������������������������������������������������������������������������������������������������������������12 Maximum temperature (HI T) ������������������������������������������������������������������������������������������������������������������������������12 Minimum temperature (LO T) ������������������������������������������������������������������������������������������������������������������������������12 Maximum speed (HI R) ����������������������������������������������������������������������������������������������������������������������������������������12 Minimum speed (LO R) ����������������������������������������������������������������������������������������������������������������������������������������12 Fluid type (FLUI) ��������������������������������������������������������������������������������������������������������������������������������������������������12 Temperature control type (AUTO) ������������������������������������������������������������������������������������������������������������������������13
Alarm and button tone (BEEP)
Calibration (CALI) ������������������������������������������������������������������������������������������������������������������������������������������������13
Interface and output .................................................................................................................................. 14
Maintenance and cleaning ......................................................................................................................... 16
Error codes ................................................................................................................................................. 17
Accessories ................................................................................................................................................. 17
Technical data ............................................................................................................................................ 18
Warranty .................................................................................................................................................... 19
Pump performance curve ........................................................................................................................... 19
������������������������������������������������������������������������������������������������������������������������������13
3
Page 4
Declaration of Conformity
We declare under our sole responsibility that this product corresponds to the regulations 2014/35/EU, 2006/42/EC, 2014/30/EU and 2011/65/EU and conforms with the standards or other normative documents: EN 61010-1, EN 61010-2-010, EN 61326-1, EN 60529 and DIN 12876-1�
Explication of warning symbols
Indicates an (extremly) hazardous situation, which, if not avoided, will result in death,
DANGER
serious injury�
Indicates a hazardous situation, which, if not avoided, can result in death, serious injury�
WARNING
Indicates a potentially hazardous situation, which, if not avoided, can result in injury�
CAUTION
Indicates practices which, if not avoided, can result in equipment damage�
NOTICE
Indicates a hazardous situation that cause from a hot surface!
CAUTION
Hot surface!
4
Page 5
Safety instructions
General information:
Read the operating instructions completely before
starting up and follow the safety instructions.
• Keep the operating instructions in a place where it can be accessed by everyone�
• Ensure that only trained staff work with the device�
• Follow the safety instructions, guidelines, occupational health, safety and accident prevention regulations�
• Set up the device in a spacious area on an even, stable, clean, non-slip, dry and fireproof surface�
• Do not operate the device in explosive atmospheres
• Protect the device and accessories from bumping and im­pacting�
• Check the device and accessories for damage before each use� Do not use damaged components�
• Safe operation is only guaranteed with the accessories de­scribed in the ”Accessories” section�
• The device must only be operated with the original power cable�
• The socket for the power cable must be easily accessible�
• Socket must be earthed (protective ground contact)�
• The voltage stated on the type plate must correspond to the power voltage�
The device can only be disconnected from the power sup­ply by pulling out the power plug or the connector plug�
• Disconnect the power plug before attaching or changing any accessories�
• Disconnect the power plug before cleaning, maintenance or moving the device�
• The device must only be opened by trained specialists, even during repair� The device must be unplugged from the power supply before opening� Live parts inside the device may still be live for some time after unplugging from the power supply�
NOTICE
safe operation, for example to keep foreign objects and liquids, etc� from getting into the device�
• The device must only be used as prescribed and as de­scribed in the operating instructions� This includes opera­tion by instructed specialist personnel�
• When using critical or hazardous materials in your pro­cesses, IKA® recommends to use additional appropriate measures to ensure safety in the experiment� For example, users can implement measures that inhibit fire or explo­sions or comprehensive monitoring equipment�
• Process pathogenic material only in closed vessels under a suitable fume hood� Please contact IKA® application sup­port if you have any question�
Coverings or parts that can be re­moved from the device must be put back on the device again to ensure
CAUTION
can be easily accessed must be installed in the work area�
• A laboratory circulator heats/refrigerates and circulates fluid according to specified parameters� This involves haz­ards due to high and low temperatures, fire and general hazards due to electrical energy� The user safety can not be ensured simply with design requirements on the part of the device� Further hazard sources may arise due to the type of tempering fluid, e�g� by exceeding or undercutting certain temperature thresholds or by the breakage of the container and reaction with the carrier fluid� It is not pos­sible to consider all eventualities� They remain largely sub­ject to the judgment and responsibility of the operator� For this reason, it may become necessary for user to take other precautionary safety measures�
• Insucient ventilation may result in the formation of ex­plosive mixtures� Only use the device in well ventilated areas�
WARNING
not be exceeded even in the event of a fault� Check the safe temperature circuit on a regular basis (see section "Setting the safety temperature")�
• When device is used for external circulation, extra pre­caution must be taken for hot/cold liquid leakage due to damaged hose�
- Use suitable hoses for connection�
- Secure hoses and tubes against slippage and avoid kinks�
- Check hoses, tubes and bath at regular intervals for pos­sible material fatigue (cracks/leaks)�
• Power cable should not get in contact with hot parts and fluids�
WARNING
• Be careful when filling a hot bath�
CAUTION
• After a power failure during operation, the device may start automatically (depending on operating mode)�
• Transport the device with care (see section “Moving the device“)�
If the power switch is not within reach when device is operating, an EMERGENCY STOP switch that
The safety circuit (safe temperature) must be adjusted so that the maxi­mum permissible temperature can-
Do not start up the device if:
- It is damaged or leaking
- Cable (not only power cable) is damaged�
At high operating temperature, the temperature of housing parts, sur­faces and tubes can exceed 70 ° C�
5
Page 6
• Do not transport or empty the bath while it is still hot/ cold� This may result in accidents, especially scalding inju­ries or frostbite�
Disposal of the device:
• The device must be disposed of in accordance with na­tional or local regulations�
Fluids:
• Continuous monitoring of the bath and the filling level of the bath fluid is required, especially at high temperatures�
• To ensure a sucient fluid circulation, the viscosity of the bath fluid must not exceed of 50mm2/s at the lowest op­erating temperature�
NOTICE
If water is used at higher tempera­ture, there is heavy loss of fluid due to the evaporation of the steam�
CAUTION
WARNING
the chemical hazards that may be associated with the bath fluid used� Observe all safety warnings for the fluids�
• Depending on the bath fluid used and the type of opera­tion, toxic or flammable vapors can arise� Ensure suitable extraction�
• Do not use any fluid which may cause dangerous reac­tions during processing�
• Only use recommended bath fluids� Only use non-acid and non corroding fluids�
WARNING
that has been set� The safe temperature limit must always be set to at least 25 ºC lower than the flash point of the fluid used� Examine regularly the function of the safety temperature limiter�
NOTICE
basis (see the chapter “Filling and draining”)�
Beware of the risk of burning due to delay in boiling!
Only use fluids, which fulfill the re­quirements for safety, health and device compatibility� Be aware of
Only process and heat up any fluid that has a flash point higher than the adjusted safe temperature limit
Never operate the device without sucient heat carrier fluid! Check the fluid level detection at a regular
Untreated tap water is not recommended� It is recom­mended to use distilled water or high purity water (ion ex­changers) and add 0�1 g soda (sodium carbonate Na2CO3) /liter, to reduce corrosive properties�
CAUTION
Risk of burning caused by vapor or hot water at the outlet of the cool­ing coil�
Don’t use following fluids:
NOTICE
- Untreated tap water
- Acids or bases
- Solutions with halides: chlorides, fluorides, bromides, iodides or sulfur
- Bleach (Sodium Hypochlorite)
- Solution with chromates or chromium salts
- Glycerine
- Ferrous water�
NOTICE
ing hoses and external devices)� When doing this, also open the stopper and union nuts caps of the pump out­puts and inputs and blow compressed air through all the pump outputs and inputs! Beware of the risk of burning due to delay in boiling!
When changing the bath fluid type, must remove the remaining water from the complete system (includ-
Correct use
• Use:
Use HRC 2 (Heated Recirculating Chiller) for tempering and circulating fluids�
Intended Use: Tabletop device
• Range of use (indoor use only):
- Laboratories - Schools
- Pharmacies - Universities
This instrument is suitable for use in all areas except:
- Residential areas
-
Areas that are connected directly to a low-voltage supply network that also supplies residential areas�
The safety of the user cannot be guaranteed:
- If the device is operated with accessories that are not sup­plied or recommended by IKA® �
- If the device is operated improperly or in contrary to the
IKA® specifications�
- If the device or the printed circuit board are modified by third parties�
6
Page 7
Unpacking
• Unpacking:
- Unpack the device carefully�
- Any damage should be notified immediately to the shipping agent (post oce, railway network or logistics company)�
• Delivery scope:
- HRC 2 basic
- Power cables
- Hose olive NW 8 (2 pieces) see Fig� 2
- Hose olive NW 12 (2 pieces) see Fig� 2
- Screwdriver (use for safety circuit) see Fig� 3
- Plastic cap (for “Overflow” connector)
- Plastic cap (for “Backflow” connector)
- USB 2�0 cabel A–micro B
- User guide
- Warranty card�
Preparations
Fig. 3Fig. 2
• Setting up:
- Place the unit on an even, stable, clean, nonslip, dry and fireproof surface�
- Keep at least 20 cm of open space on the front and rear side�
- The place for installation should be large enough and provide sucient air ventilation to ensure the room does not warm up excessively because of the heat from de­vice radiates to the environment�
- Do not set up the device in the immediate vicinity of heat sources and do not expose to sun light�
- Cooling machine, pump motor and electronics produce intrinsic heat that is dissipated via the venting grids (3)! Never cover these venting grids!
Note: After setting up the device, wait at least one hour before starting the operation to avoid the damage to the cooling system�
• Connecting the tubings:
- Unscrew the union nuts and stoppers using a wrench (SW 19) from the pump connector IN (12) and OUT (11)�
Stopper
- Screw the hose olive to the pump connection IN and OUT onto the olives� The hoses must be secured with suit­able clamps�
- Via a hose to the “Overflow” connector (7), overflow­ing fluid can be directed into a suitable vessel� The ves­sel should be positioned lower than the “Overflow” connector�
- Connect the “Backflow” connector (8) to IKA® calo­rimeter with a suitable hose�
(11)
with union nuts� Slide the hoses (NW 12)
Olives
Union nut
External
system
(12)
Fig. 5
Union nut
Fig. 4
- Connect the hoses for circulating the external system to the pump connectors M 16 x 1 for IN directly or with the olives�
(12)
and OUT
(11)
Fig. 6
IKA®
Calorimeter
7
Page 8
Note: Please note the permissible temperature range of hoses� For hot fluids we recommend the IKA hoses� When the external system is not necessary, please seal the pump connector IN and OUT with the existing union nuts and stoppers� When IKA® calorimeter is not connected, please close the “Backflow” connector with included backflow cap� In case of closed external system, if there is in any case no risk of exceeding the maximum fluid level, you can close the “Overflow” connector with the included overflow cap� This minimizes the energy loss and protects, in case of long-term applications with low temperatures and high humidity, against unwanted water input and the as­sociated volume increase� Check fluid heat expansion!
• Filling and draining:
- Before filling the fluid into the bath, open the venting grid as indicated in following figure�
®
LT 5�20
Drain port
Drain valve
Bracket
Fig. 8
Note: Pay attention to the fluid level information:
Low Level
Fig. 7
- Check and make sure that the drain valve is closed (ro­tate clockwise to the stop position, see Fig� 8)�
Please note information in section “Commissioning”�
Note:
- Connect the power plug and turn on the device with power switch (10)�
- The low level warning message appear on the display� Meanwhile, the bottom LED segment of fluid level indi­cator lights up in red color to indicate the low fluid level (see Fig� 9)
- Open the filling opening lid (4) and remove the stopper from the filling opening� Then, fill fluid to the bath�
High Level
Red (High Level)
Red (Low Level)
Fig. 9
- To drain the fluid from the bath, connect a hose to the drain port and turn the drain valve in counter clockwise direction with a straight screwdriver�
Note: The stopper must be kept being plugged in dur­ing during operation except for filling fluids�
Fig. 10
Warning: Do not empty the bath while the fluid is still hot/
cold, there is a risk of burning/freezing�
8
Page 9
• Fluids (Standard information for IKA® fluid):
No.
0 1 2 3 4 5 6 7
8
®
IKA
Designation
CF�EG28�N10�80�8 CF�EG39�N20�80�16 CF�EG44�N25�80�19 CF�EG48�N30�80�22 UF�Si�N30�150�10LV HF�Si�20�200�50 HF�Si�20�250�50A
Water Customized
Operating temperature
range for open bath
application
(°C)
-10 ��� 80 -10 … 80 90 115
-20 … 80 -20 … 80 90 115
-25 … 80 -25 … 80 90 115
-30 … 80 -30 … 80 90 115
-30 … 130 -30 … 150 20 … 200 20 … 200 255 > 280 20 … 200 20 … 250 255 > 280
5 … 95 5 … 95 - -
Operating temperature
range for closed bath
applications
(°C)
Safety temperature
(°C)
145
The listed fluid temperature range may deviate from the device operating temperature ranges and settings�
Nomenclaturefor IKA® fluids: CF�EG28�N10�80�8 --
(5) Addition information
mm²/s
(4) Viscosity (8
) (3) Temperature range (-10–80 °C) (2) Chemical composition (Ethylene Glycol 28%) (1) Classification (Cooling fluid)
Flash point
(°C)
> 170
(1) Classification:
HF: Heating Fluid CF: Cooling Fluid UF: Universal Fluid
(2) Chemical composition:
Si: Silicone oil EG: Ethylene Glycol
(3) Temperature range: (
Minimum temperature� Maximum temperature)
N: Negative Temperature
(4) Viscosity:
Viscosity at 25 °C for Heating Fluid (HF) Viscosity at -20 °C for Cooling Fluid (CF) Viscosity at 25 °C for Universal Fluid (UF)
Dynamic viscosity [mPa•s] is a product of kinematic viscosity [mm²/s] and density [kg/m³] of the fluid divided by 1000�
(5) Additional information:
A: Oil Additives LV: Low Viscosity
Note: for open bath application!
Note: Tap water may be unsuitable for operation because the calcium carbonate content may cause calcification� High purity water (from icon exchangers) and distilled or bi-distilled water are unsuitable for operation due to corrosive properties of these media� High purity water and distillates are suitable as a medium after adding 0�1 g soda (Na2CO3, sodium carbonate) per liter of water�
Note: The temperature limit values are adjustable in accordance with the fluid used�
9
Page 10
• Moving the device:
Empty all fluid in the bath before moving device from one
place to other place�
The device can be lifted up and moved by using the top
handles or the bottom handles�
It can also be moved on flat surface by lifting and pushing
the front of the device� It is easy to move the device with
the help of the wheels�
The angle of inclination should never be more than 10 º
at any direction when move the device!
Note: The device must not be moved during operation�
After moving the device, you must wait at least one hour
before restarting the unit�
Operator panel and display
A
P
O
Fig. 12
Fig. 11
E
L JK
max. 10 ˚
F H
GCB D
IMN
Item Designation Function A On/Off button: Switch on/off the circulator� B LED, set: The LED lights up at the same time as the set value is displayed� C LED, heating: D LED, E LED display: F “Menu” button: Press it once: menu option is display�
Press it once again: back to working screen�
G LED, lock: Indicate the buttons are H Lock button: I “OK / Pump” button: Start/stop the pump function�
Confirm the menu options� J “Pump (+)” button: Increase the pump speed setting� Change the menu value setting� K “Pump (-)” button: Decrease the pump speed setting� Change the menu value setting�
L “Temp” button: Start/stop the tempering function� M “Temp (+)” button: Increase the temperature setting�
N “Temp (-)” button: Decrease the temperature setting�
O Adjustable safety circuit: Adjust the safety temperature limit� P Fluid level indicator: Indicate the fluid level (when the fluid level is too low or too high, the bottom or the
refrigerating
: The LED indicates the
Indicate the heating function is activated�
refrigerating
Display the settings and actual values�
deactivated�
Lock/unlock buttons�
Navigation, selecting the settings in the menu
Navigation, selecting the settings in the menu�
top LED segment change into red color)�
function is activated�
10
Page 11
Setting the safety temperature
Setting the safety temperature with screwdriver delivered with the device�
+
-
Fig. 13
Commissioning and operating
Note:
Before commissioning, make sure that the device has
not been moved in one hour�
Check whether the voltage specified on the type plate matches the power voltage available�
The power socket used must be earthed (protective earthing)�
If above conditions are met, the instrument is ready for op­eration after plugging in the power plug� If these conditions are not met, safe operation is not guar­anteed and the device could be damaged�
Observe the ambient conditions (temperature, humidity, etc�) listed under “Technical Data”�
After switching on the power switch (10) at the back of the device or pressing the on/off button (A), all LED segments light up during the self test�
The safety temperature setting will appear on the display�
Safety temperature
Factory setting: approximate max� value Adjustment range: 0 to 110 °C�
Warning: The safety temperature must always be set to at least 25 ºC lower than the flash point of the fluid used�
After that, following two screens appear alternately�
Then the device enters standby status and is ready for op­eration�
Change the temperature setting with “Temp (+)“ button (M) or “Temp (-)“ button (N)� Change the pump speed setting with “Pump (+)“ button (J) or “Pump (-)“ button (K)�
In standby status, activate the tempering function by press­ing the “Temp“ button (L)� The device starts tempering function and pumping function� After pressing the power switch (10) the start-up of the compressor is delayed (about 3 minutes)�
Display field
Software version (ver)
Operating mode (mode)
Safety temperature
Working setting
In working status, press the “OK/Pump“ button (I) to stop pumping and tempering function�
Note: In running status, the pump speed adjusts automati­cally according to the load� When the actual speed doesn’t reach the setting speed, the blinking set speed and the ac­tual speed appears alternately�
In standby status, press the “OK/Pump“ button (I) to start the pump function� The tempering function will not be ac­tivated� In working status, press the “Temp“ button (L) to stop the re tempering function, the pump keep running�
Pressing the lock button (H) for about 2 seconds, the device control elements can be locked to prevent accidental chang­es that can be done during operation (LED (G) will light up)� To unlock the control elements, press the lock button again for about 2 seconds (LED (G) goes out)�
11
Page 12
• Menu structure:
MODE
Menu settings
A B C
Default settings
Activated
-
-
Maximum temperature (HI T) Minimum temperature (LO T) Maximum speed (HI R) Minimum speed (LO R) Fluid type (FLUI)
Menu
The type of temperature PID control arithmetic (AUTO) (Heating)
Alarm and button tone (BEEP)
Calibration (CALI)
Basic guide to setting the menu options:
F Enter the menu by pressing ”Menu” button (F)�
F Press the
“Temp (+)“ button (M) or “Temp (-)“ button (N)
to change the menu options�
F Press the
“Pump (+)“ button (J) or “Pump (-)“ button (K)
to change the menu value settings�
F Confirm the menu settings by pressing the
key (I)
Operating mode (MODE):
“OK/Pump“
AUTO 1 AUTO 0
BEEP 0 BEEP 1
CALI 0 CALI 2 CALI 3
Minimum temperature (LO T):
Proportional coecient of PID (Kp 1) Integrate time of PID (Ti 1) Differential time of PID (Td 1)
The minimum adjustable value: - 20 °C
Maximum speed (HI R):
The maximum adjustable value: 4000 rpm
100 °C
- 20 °C 4000 rpm 2000 rpm 8
Activated 60�0 5�00 0�3
­Activated
Activated
-
-
Mode A: After power on/power failure no automatic re-
start of functions�
Mode B: After power on/power failure automatic restart
of functions, depending on previous settings�
Mode
C:
Set values (set in mode A or B) cannot be changed� After power on/power failure automatic restart of functions, depending on previous settings�
Maximum temperature (HI T):
The maximum adjustable value: 100 °C� This value can be limited additionally by the selected fluid (No�, see “Fluid (Standard information for IKA® fluid)”�
Note: The maximum value is limited by the set safety temperature�
Minimum speed (LO R):
The minimum adjustable value: 2000 rpm�
Fluid type (FLUI):
The chosen fluid (No�) limits the maximum and minimum temperature� Customized fluid (No�: 8) enables the whole working temperature range� See Section “Fluids (Stan-
®
dard information for IKA
fluid)”�
12
Page 13
Temperature control type (AUTO) (heating):
°C
t
°C
t
°C
t
°C
°C
t
Proportional coecient of PID (Kp 1)
AUTO 1:
AUTO 1 is the default setting� The default PID settings are used automatically�
Heating up curve in AUTO 1 control (AUTO 1):
Fig. 14
AUTO 0: For special requirements, AUTO 0 can be used with man­ual adjustment of the control parameters�
If the PID control (AUTO 0) option is selected, the follow­ing items including Kp 1, Ti 1 and Td 1 could be set� Otherwise they are not displayed in the menu list�
Note: Inappropriate settings may produce the following heating up curves:
Kp too high
The proportional coecient Kp is the controller ampli­fication and determines how strongly the control devia­tion (the difference between the target temperature and actual temperature) directly affects the control variable (cooling down time)� Kp values that are too large can lead to the controller overshooting�
Integral time of PID (Ti 1)
The integral time Ti (s) is the correction time and deter­mines how strongly the duration of the control deviation affects the control variable� Ti compensates an existing control deviation� A high Ti means a smaller and slower effect on the control variable� Ti values that are too small can lead to instability of the controller�
Differential time of PID (Td 1)
The differential time Td (s) is the derivative time and de­termines how strongly the rate of change of the control deviation affects the control variable� Td compensates for rapid control deviations� A high Td means a smaller and slower effect on the control variable� Td values that are too small can lead to instability of the controller�
Fig. 15
Ti too low
Kp too low Ti too high
Td too high
t
Alarm and button tone (BEEP):
BEEP 0: no alarm and button tone BEEP 1: alarm and button tone is activated
Calibration and adjustment (CALI):
CALI 0: Reset calibration CALI 2: 2-point calibration CALI 3: 3-point calibration
13
Page 14
Example: 2-point calibration:
Dip the temperature sensor of the reference measuring instrument into the bath fluid�
Select 2-point calibration in the menu� Press the “OK/ Pump“ button (I) to start the 2-point calibration�
The unit now starts and controls to the set value� When the temperature has reached the set value and has be­come constant, the “Set” LED (C) no longer flashes and the following display appears� The temperature measured by the unit appears on the left hand side�
Set the first point temperature (step 1, e�g� 10 °C) with “Pump (+)” button (J) or “Pump (-)” button (K)� Confirm the setting by pressing “OK/Pump“ button (I)�
Left area of the display appears the set value (10 °C) and the right area of the display appears the actual tempera­ture value measured� The “Set” LED (C) flashes briefly�
Interface and output
The device can be operated in “Remote” mode via the RS 232 port (5) or the USB port (6) connected to a PC and with the laboratory software Labworldsoft®�
Note: Please observe the system requirements together with the operating instructions and help section included with the software�
USB interface:
The Universal Serial Bus (USB) is a serial bus for connecting the device to the PC� Equipped with USB devices can be connected to a PC during operation (hot plugging)� Con­nected devices and their properties are automatically recog­nized� The USB port can also be used to update firmware�
USB device driver:
First, download the latest driver for IKA® devices with USB port from: http://www.ika.com/ika/lws/download/usb-driver.zip. Install the driver by running the setup file� Then connect the
IKA® device through the USB data cable to the PC�
The data communication is via a virtual COM port� Configu­ration, command syntax and commands of the virtual COM ports are as described in RS 232 interface�
Device software update: For device software update, visit IKA® website www.ika. com and enter the “Service” menu� Download and run the
Firmware Update Tool�
Input the calibration value from the reference measuring instrument (e�g� 10�1 °C) with “Pump (+)” button (J) or “Pump (-)” button (K)�
Confirm the value by pressing the “OK/Pump“ button (I)� Back to previous screen for input a new value by pressing “Temp” button (L)�
Back OK
The first point calibration is finished now� Calibration of the other points is performed in the same way�
Find and click the Firmware Update Tool in your PC after installation� Register your E-mail and password� Connect the device to your PC via the USB cable� Then, you can update the device software according to instructions of the Firmware Update Tool�
RS 232 interface:
Configuration:
- The functions of the interface connections between the stirrer machine and the automation system are chosen from the signals specified in EIA standard RS 232 in ac­cordance with DIN 66 020 Part 1�
- For the electrical characteristics of the interface and the al­location of signal status, standard RS 232 applies in accor­dance with DIN 66 259 Part 1�
- Transmission procedure: asynchronous character transmis­sion in start-stop mode�
- Type of transmission: full duplex�
- Character format: character representation in accordance with data format in DIN 66 022 for start-stop mode� 1 start bit; 7 character bits; 1 parity bit (even); 1 stop bit�
- Transmission speed: 9600 bit/s�
- Data flow control: none
- Access procedure: data transfer from the stirrer machine to the computer takes place only at the computer’s request�
14
Page 15
Command syntax and format:
The following applies to the command set:
- Commands are generally sent from the computer (Master) to the device (Slave)�
- The device sends only at the computer’s request� Even fault indications cannot be sent spontaneously from the device to the computer (automation system)�
- Commands are transmitted in capital letters�
The above details correspond as far as possible to the recommendations of the NAMUR working party (NAMUR recom­mendations for the design of electrical plug connections for analogue and digital signal transmission on individual items of laboratory control equipment, rev� 1�1)� The NAMUR commands and the additional specific IKA® co mmands serve only as low level commands for communica­tion between the device and the PC� With a suitable terminal or communications programme these commands can be transmitted directly to the circulator equipment� The IKA® software package, Labworldsoft®, provides a convenient tool for controlling circulating equipment and collecting data under MS Windows, and includes graphical entry features, for pump motor speed ramps for example�
Commands:
NAMUR Commands Function
IN_PV_2 Read the internal actual temperature IN_PV_3 Read the safety actual temperature IN_PV_4 Read the pump actual speed IN_SP_1 Read the internal setting temperature IN_SP_3 Read the safety setting temperature IN_SP_4 Read the pump setting speed OUT_SP_1 xxx Set the internal setting temperature XXX OUT_SP_12@n Set the WD safety temperature with echo of the set (defined) value� OUT_SP_4 xxx Set the pump speed XXX OUT_SP_42@n Set the WD-safety speed with echo of the set (defined) value� OUT_WD1@n Start the watchdog mode 1 and set the time for the watchdog to n (20���1500) seconds�
Echo of the Watchdog time� During a WD1-event, the tempering and pump functions are switched off� This command needs to be send within the watchdog time�
OUT_WD2@n Start the watchdog mode 2 and set the watchdog time to n (20���1500) seconds�
Echo of the watchdog time� During a WD2-event, the set temperature is changed to the WD safety temperature and the pump set speed is set to the WD safety speed�
This command needs to be send within the watchdog time� RESET Reset the PC control and stop the device functions� START_1 Start the tempering function� START_4 Start the pump function� STOP_1 Stop the tempering function� STOP_4 Stop the pump function�
- Commands and parameters including successive param­eters are separated by at least one space (Code: hex 0x20)�
- Each individual command (incl� parameters and data) and each response are terminated with CR LF (Code: hex 0x0d hex 0x0A) and have a maximum length of 80 characters�
- The decimal separator in a number is a dot (Code: hex 0x2E)�
15
Page 16
Connections between device and external devices:
9
6 7 8 9
PC 1�1 Cable: This cable is required to connect RS 232 interface (5) to a PC�
Fig. 16
1 2 3 4 5
RxD 2 TxD 3
GND 5
RTS 7 CTS 8
9 8 7 6
1
4
6
9
1 2 RxD
3 TxD 4 5 GND 6 7 RTS 8 CTS
PC
5 4 3 2 1
USB 2�0 cable A - micro B: This cable is required to connect USB interface (6) to a PC�
A
USB micro B
Fig. 17
USB A
Maintenance and cleaning
To avoid contamination, it is essential to change the bath fluid regularly� If water is used for bath fluid, we recommend to add Wa­ter bath protective media� The protective media stops the growth of algae, bacteria and other microorganisms�
To keep the full cooling performance, the dust filter of the cooler must be checked regularly and cleaned if necessary�
• Switch off the device and disconnect power cable�
• Open the front venting grid (3)�
• Remove the bracket�
• Clean condenser dust protection sheet with a vacuum cleaner or wash the sheet in the water and dry the sheet before assembly�
Note: Don’t touch the condenser surface with hard parts�
Cleaning:
Disconnect main plug prior to cleaning!
Use only cleaning agents which have been approved by
®
to clean the device�
IKA
Dirt Cleaning agent
Dye Isopropyl alcohol Construction material
Water containing tenside/ isopropyl alcohol
Cosmetics
isopropyl alcohol Foodstuff
Water containing tenside/ isopropyl alcohol
Water containing tenside Water containing tenside
Fuel Water containing tenside
For materials which are not listed, please request in­formation from IKA® application support.
Wear protective gloves while cleaning the devices� Electrical devices may not be placed in the cleansing agent for the purpose of cleaning� Do not allow moisture to get into the device when cleaning� If a different cleaning or decontamination method than the method defined by IKA® is planned, the user must ascertain with IKA® that this method does not damage the device�
Spare parts order:
When ordering spare parts, please give:
- machine type
- manufacturing number, see type plate
- item and designation of the spare parts,
see www.ika.com
- software version�
Repair: Please send in device for repair only after it has been cleaned and is free from any materials which may constitute a health hazard.
For repair, please request the “Decontamination Clearance Certificate” from IKA®, or download printout of it from the
IKA® website www.ika.com�
If you require servicing, return the device in its original pack­aging� Storage packaging is not sucient� Please also use suitable transport packaging�
16
Page 17
Error codes
Any malfunctions during operation will be identified by an error message on the display� Proceed as follows in such cases:
F Switch off device using the main switch at the back of the device F Carry out corrective measures F Restart device
Error code Effect Cause Solution Err 02 Pump off
Heating/refrigerating off
Err 04 Pump off
Heating/refrigerating off
Err 06 Pump off
Heating/refrigerating off
Err 09 Pump off
Heating/refrigerating off
Err 11 Pump off
Heating/refrigerating off
Err 12 Pump off
Heating/refrigerating off
Err 13 Pump off
Heating/refrigerating off
Err 15 Pump off
Heating/refrigerating off
Err 16 Pump off
Heating/refrigerating off
Err 17 Pump off
Heating/refrigerating off
Err 20 Pump off
Heating/refrigerating off
Motor over current (rate current)
Motor hall signal missing - Reduce pump motor speed
Too low liquid level - Check the fluid level and buoyage
Device internal temperature is too high
Temperature difference between control sensor and safety sensor is too much
Safety temperature alarm
Heater switched off by safety circuit
Safety shutdown of the compressor
Errors at the compressor (suction side)
Errors at the compressor (pressure side)
Fan cooling system error - Restart the device
- Reduce pump motor speed
- Use fluid with lower viscosity
- Check if the pump impeller is blocked
- Use fluid with lower viscosity
- Check if the pump impeller is blocked
- Check the ambient temperature and let the device cool down
- Check the fan and clean the grids at the rear side
- Check safety temperature circuit and bath fluid
- Check the bath temperature measurement
- Check safety temperature set value, fluid level
- Restart the device
- Restart the device
- Restart the device
If the actions described fails to resolve the fault or another error code is displayed then take one of the following steps:
- Contact the service department
- Send the device for repair, including a short description of the fault�
Accessories
Tubing and hoses: LT 5.20 Metal hose (isolated M16 x 1) LT 5.21 PTFE hose (isolated M16 x 1) H.PVC.8 PVC tube (nominal width 8 mm) H.PVC.12 PVC tube (nominal width 12 mm) H.SI.8 Silicone tube (nominal width 8 mm) H.SI.12 Silicone tube (nominal width 12 mm) H.PUR.8 PUR tube (nominal width 8 mm) H.PUR.12 PUR tube (nominal width 12 mm) H.FKM.8 FKM tube (nominal width 8 mm) H.FKM.12 FKM tube (nominal width 12 mm)
See more accessories on www.ika.com
Tubing insulations: ISO. 8 Insulation (8 mm) ISO.12 Insulation (12 mm)
Additional accessories: PC 1.1 Cable (RS 232) Labworldsoft
®
17
Page 18
Technical data
Operating voltage VAC Frequency Hz 50 / 60 Max� input power W 1800 (230 VAC) / 1500 (115 VAC) / 1210 (100 VAC) Working temperature range °C - 20 ��� + 100 Operating temperature range Temperature stability – internal temperature control
70
°C, water Temperature control PID Temperature measurement, absolute accuracy K ± 0�5 Temperature setting Button Temperature setting resolution K 0�1 Temperature display LED Temperature display resolution K 0�1 Classification according to DIN 12876-1 Class III (FL) suitable for flammable and non-flammable fluids Safety circuit (adjustable) °C 0 ��� + 110 Safety temperature display LED Heating capacity W 1500 (230 VAV) / 1200 (115 VAC) / 910 (100 VAC) Cooling capacity according to DIN 12876 (at 4000 rpm):
+ 20 °C + 10 °C 0 °C
- 10 °C
- 20 °C Refrigerant
Refrigerant quantity g 230 Max� refrigerating system pressure bar 20 Pump speed (adjustable) rpm 2000 ��� 4000 Max� pump pressure/suction bar 0�5 / 0�25 Max� flow rate (at 0 bar) l/min 21 Bath volume I 1�4–4�0 Fluid maximum viscosity mm Low fluid level protection Yes Interface USB, RS 232 Permitted on-time % 100 IP code according to EN 60 529 Protection class Excess voltage category II Contamination level 2 Permitted ambient temperature °C + 5 ��� + 32 Permitted ambient humidity % 80 Dimension (W x D x H) mm 220 x 525 x 475 Weight kg 28�5 Operation at a terrestrial altitude m max� 2000
(according to DIN 12876)
°C - 20 ��� + 100 K ± 0�1
W 400
230 ± 10 % / 115 ± 10 % / 100 ± 10 %
370 320 240 130
R134a
2
/s 50
IP 21 I
Note: Refrigerant must be disposed of in accordance with local and national regulations�
Subject to technical changes!
18
Page 19
Warranty
0 5 10 15 20
500
-
-200
-300
In accordance with IKA® warranty conditions, the warranty period is 24 months� For claims under the warranty please contact your local dealer� You may also send the machine direct to our factory, enclosing the delivery invoice and giv­ing reasons for the claim� You will be liable for freight costs�
The warranty does not cover worn out parts, nor does it ap­ply to faults resulting from improper use, insucient care or maintenance not carried out in accordance with the instruc­tions in this operating manual�
Pump performance curve
Pumpenkennlinie gemessen mit Wasser:
(Messung nach DIN 12876-2 mit Wasser bei 20°C, geschlossener Pumpen-Kreislauf)�
400
300
200
100
ΔP [mbar]
0
100
Fig. 18
4000 rpm
3200 rpm
2000 rpm
Q [l/min]
19
Page 20
IKA®-Werke GmbH & Co.KG
Janke & Kunkel-Str� 10 D-79219 Staufen Tel� +49 7633 831-0 Fax +49 7633 831-98 sales@ika�de
www.ika.com
Loading...