This manual applies directly to any HP 85062A Calibration Kit
having serial prefix US3444 and above.
HEWLETT
PACKARD
HP Part Number:
Printed in USA
First Edition
0
Copyright 1995 Hewlett-Packard Company
85062-90001
1995
NUMBERS
CERTIFICATION
Hewlett-Packard Company certifies that this product met its published specifications at the time of
shipment from the factory. Hewlett-Packard further certifies that its calibration measurements are
traceable to the United States National Institute of Standards and Technology (NIST, formerly NBS), to
the extent allowed by the institutes calibration facility, and to the calibration facilities of other
International Standards Organization members.
WARRANTY
This Hewlett-Packard instrument product is warranted against defects in material and workmanship for
a period of one year from date of delivery. During the warranty period, Hewlett-Packard will, at its
option, either repair or replace products which prove to be defective. For warranty service or repair, this
product must be returned to a service facility designated by HP. Buyer shall prepay shipping charges
HP and HP shall pay all shipping charges to return the product to Buyer. However, Buyer shall pay all
shipping charges, duties, and taxes for products returned to HP from another country. HP warrants that
its software and firmware designated by HP for use with an instrument will execute its programming
instructions when properly installed on that instrument. HP does not warrant that the operation of the
instrument, or software, or firmware will be uninterrupted or error free.
LIMITATION OF WARRANTY
to
I
The foregoing warranty shall not apply to defects resultin,
Buyer-supplied software or interfacing, unauthorized modification or misuse, operation outside of the
environmental specifications for the product, or improper site preparation or maintenance.
NO OTHER WARRANT
IMPLIED WARRANTIE
PURPOSE. EXCLUSIVE REMEDIES THE REMEDIES PROVIDED HEREIN ARE BUYER’S
SOLE AND EXCLUSIVE REMEDIES. HP SHALL NOT BE LIABLE FOR ANY DIRECT,
INDIRECT, SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGES, WHETHER BASED
ON CONTRACT, TORT, OR ANY OTHER LEGAL THEORY.
Y IS EXPRESSED OR IMPLIED. HP SPECIFICALLY DISCLAIMS THE
S
OR MERCHANTABILITY AND FITNESS FOR A PARTICULAR
0 from improper maintenance by Buyer,
ASSISTANCE
Product maintenance agreements and other customer assistance agreements are available for HewlettPackard Products. For any assistance, contact your nearest Hewlett-Packard Sales and Service Office.
Addresses are provided at the back of this manual.
Typical 3.5 mm Connector Gage
Figure 4-2. Gaging a 3.5 mm Male Connector Using a Push-on Type Gage
Figure 4-3. Gaging a 3.5 mm Male Connector Using a Screw-on Type Gage
Figure 4-4. Gaging a 3.5 mm Female Connector Using a Push-on Type Gage
Figure 4-5. Gaging a 3.5 mm Female Connector Using a Screw-on Type Gage
Figure 4-6. Using the Torque Wrench
The HP 85062A 3.5 mm Electronic Calibration Kit is designed to work with the HP 8506OC Electronic
Calibration (ECal) Control Unit to calibrate network analyzer systems (such as the HP 85 10) for measurement of components with 3.5 mm connectors up to 26.5 GHz. See the Operating, Programming, and Service Manual for the HP 8506OC control unit for specific network analyzer models.
The calibration kit consists of the following:
l
A 1 to 26.5 GHz Calibration Module (Option 001 adds a 45 MHz to 2 GHz Calibration Module)
.
A Specification
l
An Operating and Service Manual
l
A 5116 inch, 90
l
A Female-to-Female Adapter (not included with Option OOF or Option OOM)
l
A Male-to-Male Adapter (not included with Option OOF or Option OOM)
/
Verification Program Disk
N-cm (8
in-lb) torque wrench for use on the 3.5 mm connectors
This manual describes the HP 85062A Electronic Calibration Kit and provides replacement part numbers,
specifications, and procedures for using, maintaining, and troubleshooting the kit. The available calibration types, the best application for each type, and the associated calibration procedures are found in the
Operating, Programming, and Service Manual for the HP 8506OC control unit.
This manual assumes you know proper connector care. If not, refer to
“Principles of Microwave Connector Care-Quick Reference Card”, located in the back of this manual.
Refer to Chapter 7, “Replaceable Parts”, for the HP part number if another copy is needed.
Note
Or, contact your nearest HP sales office for the customer training course:
“Understanding Connectors Used With Network Analyzers”.
43P
8505OA
4-J.P 8505OA
+ 24A (on-site)
+ 24D (at HP sales office)
AvailableOptions
l
Option 001 adds a low band (45 MHz to 2 GHz) calibration module.
l
Option OOM replaces the standard 3.5 mm insertable module with a non-insertable male module.
l
Option OOF replaces the standard 3.5 mm insertable module with a non-insertable female module.
Options may be combined. For example, ordering Option 001 with Option OOF would provide both a high
band and a low band calibration module with non-insertable female connectors.
EQUIPMENTREQUIREDBUTNOTSUPPLIED
Connector gage sets are required for measuring connector pin depth. Gages and various connector cleaning
supplies are not provided in this kit. Refer to Chapter 7,
HP 85062A
“Replaceable Parts”, for ordering information.
General information
l-1
SERIALNUMBERS
A serial number label is attached to this calibration kit. The serial number is in the form; 0000000000. The
first six digits form the serial number prefix (the first two digits are a country code, the next
four
are a year/
week designation). The last four digits form the suffix (a simple counting sequence).
)
The contents
of
this manual apply directly to calibration kits having the serial number prefix(es) listed un-
der SERIAL NUMBERS on the title page.
A typical serial number label is shown in Figure l-l.
SERIAL NUMBER
f
PREFIX SUFFIX
Figure I-1. Typical Calibration Kit Serial Number Label
-l
In addition to the kit serial number, the calibration modules in the HP 85062A calibration kit are individually serialized. These serial numbers should be recorded in Table l-l when the
purpose
with U.S.
of
this is to help maintain the integrity
MLSTD-45662A,
should your compliance be required.
of
your kit. Kit integrity is one important part
kit is
first received. The
of
compliance
1-2
General Information
HP 85062A
Table I-I. Serial Numbers of Individual Calibration Devices
Serialized Device
HP 85062A 3.5 mm Electronic Calibration Kit
HP 85062-6000 1 Electronic Calibration Module
(low band, insertable, Option 001)
HP 85062-60002 Electronic Calibration Module
(high band, insertable, standard)
HP 85062-60003 Electronic Calibration Module
(low band, non-insertable male, Option 001 and Option OOM)
Refer to Figure 7-1, Figure 7-2, or Figure 7-3 (depending on your calibration
plete shipment. Record the serial numbers of all serialized devices in your
electrical performance
of
the devices in this
kit,
refer to Chapter 5, “Performance Verification.” The foam-
kit
options) to verify a com-
kit
using Table l-l. To verify
lined storage case provides protection during shipping. If the case or any device appears damaged, contact
the nearest Hewlett-Packard sales and service office (see inside the back cover of this manual). HewlettPackard will arrange for repair or replacement of incomplete or damaged shipments without waiting for
settlement from the transportation company.
If
the instrument is being returned to Hewlett-Packard for service, please include service information as
described in “Returning a Kit or Device to HP” in Chapter 6, “Troubleshooting” (attach to the instrument
if possible).
PREVENTWEMAINTENANCE
The best techniques for maintaining the integrity of the devices in this kit include routine visual inspection
and cleaning, and proper gaging and connection techniques. Failure to detect and remove dirt or metallic
particles on a connector mating plane surface can degrade repeatability and accuracy and can damage any
connector mated to it. Improper connections resulting from pin depth values being out
of
the observed lim-
its (see Table 2-2) or from using poor connection techniques can also damage these devices.
Visual inspection, cleaning and connection techniques, and proper gaging for pin depth are all described
in Chapter 4, “Preparation for Use - Gaging and Making Connections.”
HP 85062A
General Informationl-3
Chapter 2
SPECIFICATIONSTERMINOLOGY
Definitions
Specifications and Characteristics
The definitions explained here help to clarify the terms used in the remainder of this chapter. These definitions refer to the HP 85062A 3.5 mm Electronic Calibration Kit only and are not necessarily valid definitions for other HP products.
Operating Temperature Range.
This is the temperature range over which the calibration standards main-
tain performance to their specifications.
Accuracy-Enhanced Temperature Range.
This is the allowable network analyzer ambient temperature
drift during measurement calibration and during measurements when the network analyzer correction is
turned on. It is also the range over which the network analyzer maintains its specified performance while
correction is turned on.
If
this temperature range is exceeded, the accuracy-enhanced performance of the
network analyzer will be degraded.
Measurement Calibration.
This is the calibration performed to determine the corrections necessary for
accuracy-enhanced (correction on) measurements.
Connector Pin Depth VaZues.
Refer to Figure
2-l
and Table 2-2. Pin depth is a relative measurement be-
tween the center conductor and outer conductor mating surfaces. All references to pin depth in this manual
treat positive (+) values as protrusions, and negative (-) values as recessions.
ReferencestoConnectorSex
In this manual, adapters, calibration modules, and gage masters are referred to by the sex of their connector. For example, a male adapter has a male connector. A gage is referred to by the sex of the connector
that
it
measures. For example, a male gage
measures
male connectors; a male gage has a female connector.
HP 85062A
Specifications and Characteristics
2-1
ENVIRONMENTALSPEClFlCATlONS
Table 2-l lists the environmental specifications for the modules in the HP 85062 Calibration Kit.
Table 2-1. Environmental Speci$ications
Specifications
Operating Temperature Range
Error-Corrected Temperature Range
Storage Temperature
+20 to +26 “C (+68 to +79 o F)
*l “C of measurement calibration temperature
-40 to +75 “C (-40 to +167 OF)
Barometric Pressure (Altitude)
Operation
Storage
Relative Humidity
Operation
Storage
<4,500 meters (15,000 feet)
<15,000 meters (50,000 feet)
Non-condensing at all times
0 to 80% (26 “C maximum dry bulb)
0 to 95%
EM1
Conducted Susceptibility
Radiated Susceptibility
Radiated Emissions
Magnetic Emissions
i
OperatingTemperatureandAccuracyEnhancement
CETM 765
EN 50082-l/IEC 801-3
CISPRl
CETM 765
Limits
/
1
Temperature of the calibration modules is important because device dimensions (and therefore electrical
characteristics) change with temperature. The temperature of the calibration devices and all connectors
must be stable before use and within the operating limits shown above.
A measurement calibration of the network analyzer can be made at any temperature within the operating
temperature range
of
the calibration kit. The measurement calibration temperature of the
network
analyzer
must be maintained within the error corrected temperature range of the network analyzer. The error corrected operating temperature range for most HP network analyzers is the initial measurement calibration
temperature &l “C (21.8 “F). See the appropriate network analyzer User’s Manual.
Measurement calibration, performance verification, and actual device measurements must be made within
the error corrected operating temperature range specification of the network analyzer. Part
of
the error corrected operating temperature range can fall outside of the calibration temperature window. For example,
if
measurement calibration is performed at +20 “C (+68 “F), verification and measurements must be made
between t-19 “C (+66.2 OF) and +21 “C (+69.8 “F). Also,
if
the network analyzer ambient operating temperature drift exceeds the allowable error corrected temperature range, a new measurement calibration
must be performed to assure optimum accuracy.
Remember that your fingers are a heat source, so avoid handling the devices unnecessarily during calibration.
2-2
Specifications and Characteristics
HP
85062A
Loading...
+ 28 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.