HP 83491A, 83492A, 83493A User Manual

Page 1
User’s Guide
HP 83491/2/3A Clock Recovery Modules
Page 2
© Copyright Hewlett-Packard Company 1999 All Rights Reserved. Repro­duction, adaptation, or trans­lation without prior written permission is prohibited, except as allowed under copy­right laws.
HP Part No. 83491-90011 Printed in USA June 1999
Hewlett-Packard Company Lightwave Division 1400 Fountaingrove Parkway Santa Rosa, CA 95403-1799, USA (707) 577-1400
Notice.
The information contained in this document is subject to change without notice. Com­panies, names, and data used in examples herein are ficti­tious unless otherwise noted. Hewlett-Packard makes no warranty of any kind with regard to this material, includ­ing but not limited to, the implied warranties of mer­chantability and fitness for a particular purpose. Hewlett­Packard shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material.
Restricted Rights Legend.
Use, duplication, or disclo­sure by the U.S. Government is subject to restrictions as set forth in subparagraph (c) (1) (ii) of the Rights in Technical Data and Computer Software clause at DFARS 252.227-7013 for DOD agencies, and sub­paragraphs (c) (1) and (c) (2) of the Commercial Computer Software Restricted Rights clause at FAR 52.227-19 for other agencies.
Warranty.
This Hewlett-Packard instru­ment product is warranted against defects in material and workmanship for a period of one year from date of ship­ment. During the warranty period, Hewlett-Packard Com­pany will, at its option, either repair or replace products which prove to be defective. For warranty service or repair, this product must be returned to a service facility desig­nated by Hewlett-Packard. Buyer shall prepay shipping charges to Hewlett-Packard and Hewlett-Packard shall pay shipping charges to return the product to Buyer. However, Buyer shall pay all shipping charges, duties, and taxes for products returned to Hewlett­Packard from another coun­try.
Hewlett-Packard warrants that its software and firmware designated by Hewlett-Pack­ard for use with an instrument will execute its programming instructions when properly installed on that instrument. Hewlett-Packard does not warrant that the operation of the instrument, or software, or firmware will be uninter­rupted or error-free.
Limitation of Warranty.
The foregoing warranty shall not apply to defects resulting from improper or inadequate maintenance by Buyer, Buyer­supplied software or interfac­ing, unauthorized modifica­tion or misuse, operation outside of the environmental specifications for the product, or improper site preparation or maintenance.
No other warranty is expressed or implied. Hewlett-Packard specifically
disclaims the implied warran­ties of merchantability and fit­ness for a particular purpose.
Exclusive Remedies.
The remedies provided herein are buyer's sole and exclusive remedies. Hewlett-Packard shall not be liable for any direct, indirect, special, inci­dental, or consequential dam­ages, whether based on contract, tort, or any other legal theory.
Safety Symbols.
CAUTION
The hazard. It calls attention to a procedure which, if not cor­rectly performed or adhered to, could result in damage to or destruction of the product. Do not proceed beyond a cau­tion sign until the indicated conditions are fully under­stood and met.
WAR NI NG
The hazard. It calls attention to a procedure which, if not cor­rectly performed or adhered to, could result in injury or loss of life. Do not proceed beyond a warning sign until the indicated conditions are fully understood and met.
sign denotes a
caution
warning
sign denotes a
The instruction man­ual symbol. The prod­uct is marked with this warning symbol when it is necessary for the user to refer to the instructions in the manual.
The laser radiation symbol. This warning symbol is marked on products which have a laser output.
The AC symbol is used to indicate the required nature of the line module input power.
The ON symbols are
|
used to mark the posi­tions of the instrument power line switch.
The OFF symbols
are used to mark the positions of the instru­ment power line switch.
The CE mark is a reg­istered trademark of the European Commu­nity.
The CSA mark is a reg­istered trademark of the Canadian Stan­dards Association.
The C-Tick mark is a registered trademark of the Australian Spec­trum Management Agency.
This text denotes the
ISM1-A
instrument is an Industrial Scientific and Medical Group 1 Class A product.
ii
Page 3

General Safety Considerations

General Safety Considerations
WARNING
WARNING
WARNING
CAUTION
CAUTION
Light energy can radiate from the front panel HP 83492A and HP 83493A modules. The light emitted from these connectors is the slightly attenuated light that is input to the front-
Input
panel
If this instrument is not used as specified, the protection provided by the equipment could be impaired. This instrument must be used in a normal condition (in which all means for protection are intact) only.
NO OPERATOR SERVICEABLE PARTS INSIDE. Refer servicing to qualified Hewlett-Packard personnel. To prevent electrical shock, do not remove covers.
This product is designed for use in INSTALLATION CATEGORY II and POLLUTION DEGREE 2, per IEC 1010 and 664 respectively.
Electrostatic discharge (ESD) on or near input connectors can damage circuits inside the instrument. Repair of damage due to misuse is warranty. Before connecting any cable to the electrical input, momentarily short the center and outer conductors of the cable together. Personnel should be properly grounded, and should touch the frame of the instrument before touching any connector.
This product has been designed and tested in accordance with IEC Publica­tion 1010, Safety Requirements for Electronic Measuring Apparatus, and has been supplied in a safe condition. The instruction documentation contains information and warnings which must be followed by the user to ensure safe operation and to maintain the product in a safe condition.
connector.
Output
connectors on
covered under
not
iii
Page 4
General Safety Considerations
iv
Page 5

Contents

General Safety Considerations iii
1 Installation
Installation 1-2
2Operation
HP 83491/2/3A Modules—at a glance 2-2 Front-Panel Features 2-4 Block Diagrams 2-7 To display a signal 2-8 To compensate for module insertion loss 2-9 Using Probes with an HP 83491A 2-10
3 Specifications and Regulatory Information
HP 83491A Specifications 3-3 HP 83492A Specifications 3-4 HP 83493A Specifications 3-6 HP 83491/2/3A Operating Specifications 3-7 Declaration of Conformity 3-8
4 Reference
Front-Panel Optical Adapters 4-2 In Case of Difficulty 4-3 Error Messages 4-5 Electrostatic Discharge Information 4-8 Cleaning Connections for Accurate Measurements 4-10 Returning the Module for Service 4-21 Hewlett-Packard Sales and Service Offices 4-24
Contents-1
Page 6
Contents
Contents-2
Page 7
1
To install the module 1-3 To connect cables to an HP 83492A 1-7

Installation

Page 8

Installation

Installation
Installation
HP 83491/2/3A modules require that firmware revision A.06.25 or later be installed in the HP 83480A. If you wish to install the module in an HP 54750A digitizing oscilloscope, you must first install the HP 83480K communications firmware upgrade kit.
To check the HP 83480A’s firmware revision code
CAUTION
CAUTION
1
Press the
2
The firmware revision number is listed under the Frame section of the display.
Fiber-optic connectors are easily damaged when connected to dirty or damaged cables and accessories. The HP 83492A and HP 83493A front-panel input connectors are no exception. When you use improper cleaning and handling techniques, you risk expensive instrument repairs, damaged cables, and compromised measurements. Before you connect any fiber-optic cable to an HP 83492A or HP 83493A module, refer to “Cleaning Connections for
Accurate Measurements” on page 4-10.
The circuits on electrical inputs and outputs can be damaged by electrostatic discharge (ESD). Therefore, avoid applying static discharges to any front or rear-panel electrical connector. Before connecting any coaxial cable to a front­panel connector, momentarily short the center and outer conductors of the cable together. Avoid touching the front-panel connectors without first touching the frame of the instrument. Be sure that the instrument is properly earth-grounded to prevent buildup of static charge. Refer to “Electrostatic
Discharge Information” on page 4-8.
Utility
key and then the
System config
softkey.
1-2
Page 9
Installation
Installation

To install the module

1
Verify that all system components ordered have arrived by comparing the shipping forms to the original purchase order. Inspect all shipping containers. The shipment includes:
An HP 83491/2/3A Clock Recovery Module
Fiber-optic adapter cable
Two RF adapter cables
HP 83491/2/3A Clock Recovery Modules User’s Guide (this book)
If your shipment is damaged or incomplete, save the packing materials and notify both the shipping carrier and the nearest Hewlett-Packard Sales and Service office. HP will arrange for repair or replacement of damaged or incom­plete shipments without waiting for a settlement from the transportation com­pany. Notify the HP customer engineer of any problems.
1
Make sure that the serial number listed on the module’s rear-panel label matches the serial number listed on the shipping document.
(HP 83492/3A module only)
(HP 83491A module only)
NOTE
Figure 1-1. Serial number label
2
Install the HP 83491/2/3A module into the HP 83480A mainframe’s left slot. Finger-tighten the knurled screw on the front panel of the plug-in module to ensure that the module is securely seated in the mainframe.
Installing the module into the left slot ensures that the supplied adapter cable will fit. See Figure 1-2.
1-3
Page 10
Installation
Installation
Figure 1-2. Position of modules in the mainframe
3
Clean all optical interfaces as described in “Cleaning Connections for Accurate
Measurements” on page 4-10, before making measurements
4
Perform the following steps if you’re installing an HP 83492/3A module:
.
WARNING
a
Unscrew and remove the fiber-optic adapter that is located on the optical module’s front-panel optical input connector.
b
Connect the adapter that was removed in the previous step onto the HP 83492/3A module’s front-panel
Input
connector.
On HP 83492A module’s, the input connector used depends on the wavelength of the input signal. Refer to “To connect cables to an
HP 83492A” on page 1-7.
5
Connect the supplied adapter cable as shown in Figure 1-3.
Light energy can radiate from the front panel
Output
connectors on HP 83492A and HP 83493A modules. The light emitted from these connectors is the slightly attenuated light that is input to the front-
Input
panel
6
Turn on the HP 83480A, and connect a modulated signal to the HP 83491/2/3A module’s
1-4
connector.
Input
connector.
Page 11
CAUTION
Installation
Installation
Figure 1-3. The adapter cable
HP 83491A Modules:
HP 83491A Modules: Maximum safe signal input level is ±5V. The input circuits
HP 83491A Modules: HP 83491A Modules: can also be damaged by electrostatic discharge (ESD). Before connecting any coaxial cable to the connectors, momentarily short the center and outer conductors of the cable together. Avoid touching the front-panel input connectors without first touching the frame of the instrument. Be sure that the instrument is properly earth-grounded to prevent buildup of static charge.
CAUTION
HP 83492/3A Modules:
7
On the HP 83480A, press the
trigger 2
select
8
On the HP 83491/2/3A module, repeatedly press the
(the HP 83491/2/3A module) for the trigger source.
Maximum safe signal input level is +3 dBm.
Trigger
key and then the
Source
softkey. Then,
SELECT
key until a front-
panel light indicates the data rate of the signal. See Figure 1-4 on page 1-6.
Green and red data-rate lights
The data-rate indicator lights change color between red and green to show which data rate is selected. A red light does
indicate a problem. A red light shows that the adja-
not
cent red data rate label is selected. A green light shows that the adjacent green data rate label is selected. Repeatedly pressing the SELECT key cycles through the selections in one color before switching to the opposite color. On HP 83491A modules for example, the first selection cycle shows 155 Mb/s selected. The second section cycle shows 1062 Mb/s selected.
1-5
Page 12
Installation
Installation
Figure 1-4. Front-panel lights
Clock
Unlocked
Data
and
is off.
light
outputs on an oscilloscope. Waveforms should be
9
Confirm that the
10
Observe the present. The instrument is now ready for you to begin making measurements.
1-6
Page 13
Installation
Installation
To connect cables to an HP 83492A
On HP 83492A modules, the front-panel fiber-optic connectors reverse input and output roles depending on the wavelength of the signal. Signals in the 750 nm to 860 nm wavelength range are input to the left connector and output from the right connector. Signals in the 1000 nm to 1600 nm wavelength range are input to the right connector and output from the left connector.
Figure 1-5. Input and output connections versus wavelength
1-7
Page 14
Installation
Installation
1-8
Page 15
2
HP 83491/2/3A Modules—at a glance 2-2 Front-Panel Features 2-4 Block Diagrams 2-7 To display a signal 2-8 To compensate for module insertion loss 2-9 Using Probes with an HP 83491A 2-10
To compensate for a passive probe 2-11 To compensate an HP 54701A active probe 2-11 To compensate for other devices 2-11

Operation

Page 16
Operation
HP 83491/2/3A Modules—at a glance
HP 83491/2/3A Modules—at a glance
The HP 83491/2/3A Clock Recovery Modules are designed to operate in an HP 83480A Digital Communications Analyzer. These modules recover clock and data information at standard telecom and datacom rates. The resulting trigger signal is made available to the HP 83480A mainframe via a connector located on the module’s rear-panel. An external front-panel cable passes the data signal, with some insertion loss, to the receiver module.
Table 2-1. Module Features
Module Input Connector Selectable Rates (Mb/s)
HP 83491A
HP 83492A
HP 83493A
50Ω electrical
Multimode fiber (62.5/125
Single-mode fiber (9/125
µ
m)
µ
m)
155, 622, 1060, 1250, 2120, 2488, 2500
155, 622, 1060, 1250, 2120, 2488, 2500
155, 622, 1250, 2488, 2500
Figure 2-1. An HP 83493A installed in an HP 83480A mainframe
2-2
Page 17
Operation
HP 83491/2/3A Modules—at a glance
WARNING
Light energy can radiate from the front panel
Output
connectors on HP 83492A and HP 83493A modules. The light emitted from these connectors is the slightly attenuated light that is input to the front-
Input
panel
connector.
Use with HP 71603B or HP 71612B Error Performance Analyzers
The front-panel
Data
and
Clock
outputs provide electrical recovered clock and regenerated data signals for simultaneous testing with other instruments, such as the HP 71603B or HP 71612B error performance analyzers.
Multimode module and single-mode reference receivers
HP does
recommend using the HP 83492A multimode module with single-
not
mode reference receivers such as the HP 83481A, HP 83482A, or HP 83485A,B modules. Connecting multimode to single-mode fibers causes large reflections and insertion loss because of the reduction of the optical fiber’s core from 62.5 µm to 9 µm.
Single-mode module and multimode reference receivers
It is acceptable to use an HP 83493A single-mode module with a multimode reference receiver such as the HP 83486A module. This is true provided that single-mode fiber is connected to the HP 83493A module’s front-panel
Input
connector.
What you won’t find on these modules . . .
Unlike other modules designed to be used with the HP 83480A digital communications analyzer, the HP 83491/2/3A modules do not include Channel keys or menus. Also, there are no GPIB programming commands for these modules.
2-3
Page 18
Operation

Front-Panel Features

Front-Panel Features
SELECT
Figure 2-2. HP 83491/2/3A front panels
key
key Pressing this key changes the modulation rate of the input signal. The recov-
key key
ered and retimed clock trigger is sent to the mainframe. The selection is a bypass mode where the data stream directly triggers the main­frame. Refer to “Block Diagrams” on page 2-7 to view a schematic of the nor­mal and bypass paths.
2-4
Trigger On Data
Page 19
Operation
Front-Panel Features
Green and red data-rate lights
The data-rate indicator lights change color between red and green to show which data rate is selected. A red light does
indicate a problem. A red light shows that the adja-
not
cent red data rate label is selected. A green light shows that the adjacent green data rate label is selected. Repeatedly pressing the SELECT key cycles through the selections in one color before switching to the opposite color. On HP 83491A modules for example, the first selection cycle shows 155 Mb/s selected. The second section cycle shows 1062 Mb/s selected.
UNLOCKED
indicator
Auxiliary outputs
Input and Output connectors
This light shows when clock recovery cannot be established on the signal. If a clock rate is selected, the trigger output to the mainframe is disabled to pre­vent free-run triggering. However in bypass mode ( triggering is
disabled. When the
not
UNLOCKED
Trigger On Data
selected),
light is on, you can establish a
trigger on the data input to the reference receiver.
connector:
connector: This connector provides a fully regenerated version of the
DATA
connector:connector:
input signal. It is intended for monitoring purposes only and
for rigorous
not
eye mask compliance testing. The frequency response does not conform to the requirements for eye mask testing as described in ITU-T G.957 and Bellcore GR-253-CORE. On HP 83492A and HP 83493A modules, this port is amplitude stabilized for input signals greater than approximately –23 dBm.
onnector:
CLOCK c
onnector: This connector provides the recovered clock signal. You can
onnector:onnector: use this signal to measure jitter transfer, because this output can track and fol­low input data with very fast jitter; it has a wide bandwidth jitter transfer func­tion when compared to the recovered clock signal which is routed through a rear-panel connector to the mainframe for triggering. Note that the
iliary Output
errors at the
remains synchronized to input signals several dB below the onset of
DATA Auxiliary Output
.
CLOCK Aux-
The input connectors pass the digitally modulated signal to the receiver mod­ule. The input signal, slightly attenuated and available at the
Output
connector, is connected to the input of any of the HP 83481,2,3,4,5,6, or 7 modules. The connectors on optical modules include adapters which can easily be changed to match the type of connectors that are used on your fiber-optic cables. Refer
to “Front-Panel Optical Adapters” on page 4-2 for a description of the avail-
able adapters.
2-5
Page 20
Operation
Front-Panel Features
Multimode and single-mode connections
HP 83492A modules use multimode fiber. Connecting the output to the Optical Input con­nector on HP 83481/2/5 single-mode modules results in large reflections and insertion loss.
HP 83493A modules use 9/125 µm single-mode fiber. Connecting multimode fiber to the Optical Input connector results in large reflections and insertion loss.
Recovered Clock
The recovered clock signal is routed directly to the HP 83480A mainframe through the module’s rear panel. This output has a lower jitter modulation bandwidth than the front-panel
CLOCK Auxiliary Output
.
Because of the reduced jitter modulation bandwidth on the mainframe trigger signal, a more complete view of the jitter on the waveform data is obtained.
2-6
Page 21

Block Diagrams

Operation
Block Diagrams
Figure 2-3. HP 83491A Block Diagram
Figure 2-4. HP 83492A and HP 83493A Block Diagram
2-7
Page 22
Operation

To display a signal

To display a signal
1
Install the module as described in “To install the module” on page 1-3. Be sure to connect all of the cables as described in the procedure.
2
Repeatedly press the panel light indicates the proper data rate of the signal.
Green and red data-rate lights
The data-rate indicator lights change color between red and green to show which data rate is selected. A red light does cent red data rate label is selected. A green light shows that the adjacent green data rate label is selected. Repeatedly pressing the SELECT key cycles through the selections in one color before switching to the opposite color. On HP 83491A modules for example, the first selection cycle shows 155 Mb/s selected. The second section cycle shows 1062 Mb/s selected.
• If the
• Avoid selecting a data rate that is a multiple of the input signal. For example,
• If you cannot get the clock recovery module to lock on the signal, make sure
• Signals displayed using a data trigger are less reliable than using a recovered
UNLOCKED
don’t select a 622 Mb/s data rate if the signal is really at 155 Mb/s.
that you have selected the correct data rate and that the HP 83480A (or HP 54750A) mainframe trigger level is adjusted appropriately.
clock. Signals triggered on data can also vary depending upon the trigger level.
SELECT
key on the clock recovery module until the front-
not
is on, clock recovery
light
indicate a problem. A red light shows that the adja-
cannot be established on the signal.
2-8
Page 23
Operation

To compensate for module insertion loss

To compensate for module insertion loss
The following steps allow you to enter an offset to compensate for the inser­tion loss of the clock recovery module. This provides accurate amplitude mea­surements at the input to the clock recovery module.
1
Disconnect the cable from the clock recovery module’s
2
Measure the signal using a power meter. You can use either the HP 83480A’s built-in power meter or an external power meter.
3
Reconnect the cable to the clock recovery module.
4
Disconnect the cable from the reference receiver module’s input connector.
5
Measure the signal using a power meter. You can use either the HP 83480A’s built-in power meter or an external power meter.
6
Subtract the two measurements to determine the insertion loss of the module.
Insertion loss:. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ____________
7
On the
8
Press
9
Press
reference receiver module
External scale
Attenuation
, and set the
, and enter the value calculated in Step 6.
, press the front-panel channel
Atten units
to “decibel”.
Input
connector.
SETUP
key.
2-9
Page 24
Operation
Using Probes with an HP 83491A
Using Probes with an HP 83491A
You can use external passive and active probes with the HP 83491A electrical clock recovery module. The procedures in this section generate vertical scale factors. These factors are applied to the calibration of the reference receiver module’s electrical channel. When selecting a probe, keep in mind that the input impedance of the HP 83491A is 50Ω.
If the probe being calibrated has an attenuation factor that allows the instru­ment to adjust the gain to produce even steps in the vertical scale factors, the instrument will do so. Typically, probes have standard attenuation factors such as divide by 10, divide by 20, or divide by 100.
Because the following procedures include compensation for insertion loss of the clock recovery module, do not perform the procedure “To compensate for
module insertion loss” on page 2-9.
The following probes are available for use with HP 83491A clock recovery modules:
• HP N1020A TDR probe
HP N1020A TDR probe. This passive probe (1:1, 50Ω) provides a fixture for po-
HP N1020A TDR probeHP N1020A TDR probe sitioning and holding the probe tip on the device being tested.
• HP 54701A 2.5 GHz active probe
HP 54701A 2.5 GHz active probe. This is a 100kΩ, 10:1, probe.
HP 54701A 2.5 GHz active probeHP 54701A 2.5 GHz active probe
• HP 54006A 6 GHz handheld low-impedance probe
HP 54006A 6 GHz handheld low-impedance probe. This passive probe (10:1,
HP 54006A 6 GHz handheld low-impedance probeHP 54006A 6 GHz handheld low-impedance probe 500Ω, 20:1, 1kΩ) has an input capacitance of 0.25 pf.
• HP 1163A 1 GHz resistive-divider probe
HP 1163A 1 GHz resistive-divider probe. This passive 500Ω probe has an input
HP 1163A 1 GHz resistive-divider probeHP 1163A 1 GHz resistive-divider probe capacitance of 1.5 pf.
2-10
Page 25

To compensate for a passive probe

Operation
Using Probes with an HP 83491A
1
Connect the probe to the module.
2
Attach the probe tip to the
3
Press the
4
Press
reference receiver module’s
Calibrate
and then
Input
connector on the HP 83491A clock recovery
CAL
hook that is located near the floppy disk drive.
SETUP
Calibrate probe
front-panel channel
.
key.
To compensate an HP 54701A active probe
1
Connect the HP 83491A output to the electrical measurement channel input.
2
Connect the probe to the module.
3
Connect the probe power cable to the receiver module.
4
Attach the probe tip to the
5
Press the
6
Press
reference receiver module’s
Calibrate
and then
Input
connector on the HP 83491A clock recovery
CAL
hook that is located near the floppy disk drive.
Calibrate probe
Probe Power
front-panel channel
.
connector on the reference
SETUP
key.

To compensate for other devices

The information in this section applies to both optical and electrical measure­ments. Since the mainframe’s CAL signal is a voltage source, it cannot be used to calibrate to the probe tip when the units are set to Ampere, Watt, or Unknown. Instead, set the external gain and external offset to compensate for the actual characteristics of the device. If you do not know the actual charac­teristics, you can refer to the typical specifications that came with the device.
1
Press the reference receiver module’s front-panel channel
2
3
External scale
Press
Atten units Ratio
Press Unknown).
.
Attenuation 1:1
,
, and then
Units Ampere
SETUP
key.
(Volt, Watt, or
2-11
Page 26
Operation
Using Probes with an HP 83491A
4
5
Press
Press
Ext gain
, and enter the actual gain characteristics of the device.
Ext offset
, and enter the offset introduced by the device.
2-12
Page 27
3
HP 83491A Specifications 3-3 HP 83492A Specifications 3-4 HP 83493A Specifications 3-6 HP 83491/2/3A Operating Specifications 3-7 Declaration of Conformity 3-8

Specifications and Regulatory Information

Page 28

Specifications and Regulatory Information

Specifications and Regulatory Information
Specifications and Regulatory Information
This chapter lists specifications and characteristics of the HP 83491/2/3A Clock Recovery Modules. Specifications apply over the temperature range +15°C to +35°C (unless otherwise noted) after the instrument’s temperature has been stabilized after 60 minutes of continuous operation.
Specifications
Characteristics
Calibration cycle
Specifications
Characteristics
tions and performance of the instrument.
italics.
HP warrants instrument specifications over the recommended calibration interval. To maintain specifications, periodic recalibrations are necessary. We recommend that the HP 83491/2/3A Clock Recovery Modules be calibrated at an HP service facility every 24 months.
described warranted performance.
provide useful, nonwarranted, information about the func-
Characteristics are printed in
3-2
Page 29
Specifications and Regulatory Information

HP 83491A Specifications

HP 83491A Specifications
Table 3-1. HP 83491A Specifications
Clock recovery rates (NRZ coding)
155.52 Mb/s
622.08 Mb/s
1062.50 Mb/s 1250 Mb/s
2125.00 Mb/s
2488.32 Mb/s
2500.00 Mb/s
Data triggering (characteristic) 50 Mb/s to 2500 Mb/s
Operating input power level
a b
Triggering operation, all rates
–10
10
BER, all rates
c
±0.1% ±0.1% ±0.1% ±0.1% ±0.1% ±0.1% ±0.1%
–10 dBm to 3 dBm –10 dBm to 3 dBm
Insertion loss (through path)
DC through 2500 MHz
Output jitter, all rates
d
7 dB
0.0125 UI
Maximum continuous electrical power before damage (characteristic) 1W peak
DATA and CLOCK output amplitude, all rates (characteristic) 0.5Vp-p
INPUT electrical return loss
DC through 1250 MHz (characteristic) 1250 MHz through 2500 MHz (characteristic)
20 dB
15 dB
DATA and CLOCK electrical return loss
50 MHz through 2000 MHz (characteristic) 2000 MHz through 2500 MHz (characteristic)
a. Source extinction ratio ≥ 8.2 dB when measured per TIA/EIA OFSTP-4A. b. Operating power level applies over temperature range 25 c. Better than 10 d. Measured on an oscilloscope eye diagram with PRBS 223–1 test pattern.
–10
BER when tested with PRBS 223–1 pattern.
°C ± 5°
.
10 dB
6 dB
rms
3-3
Page 30
Specifications and Regulatory Information

HP 83492A Specifications

HP 83492A Specifications
Table 3-2. HP 83492A Specifications
Wavelength range (characteristic) 750 nm to 860 nm and
1000 nm to 1600 nm
Optical INPUT and OUTPUT fiber (characteristic) 62.5/125 multimode
Optical insertion loss (through path)
750 nm to 860 nm 1000 nm to 1600 nm
Optical return loss
b
Clock recovery rates (NRZ coding)
155.52 Mb/s
622.08 Mb/s
1062.50 Mb/s 1250 Mb/s
2125.00 Mb/s
2488.32 Mb/s
2500.00 Mb/s
a
5.0 dB
5.0 dB
28 dB
±0.1% ±0.1% ±0.1% ±0.1% ±0.1% ±0.1% ±0.1%
Data triggering (characteristic) 50 Mb/s to 2500 Mb/s
Operating input power level
c d
750 nm to 860 nm
Triggering operation, all rates
–10
BER, all rates
10
e
–10 dBm to 3 dBm –10 dBm to 3 dBm
1000 nm to 1600 nm
Triggering operation, all rates
–10
BER, all rates
10
Output jitter, all rates
f
g
–13 dBm to 3 dBm –13 dBm to 3 dBm
0.0125 UI
rms
Maximum continuous optical power before damage (characteristic) 10 mW peak
DATA and CLOCK output amplitude, all rates (characteristic) 0.5Vp-p
DATA and CLOCK electrical return loss
50 MHz through 2000 MHz (characteristic) 2000 MHz through 2500 MHz (characteristic)
10 dB
6 dB
3-4
Page 31
Specifications and Regulatory Information
HP 83492A Specifications
a. Minimum loss in 850 nm window. b. Single-mode backreflection tested with FC/PC adapter and single-mode fiber. Optical output terminated
with > 33 dB return loss. Return loss with fully filled 62.5 c. Source extinction ratio d. Operating power level applies over temperature range 25 e. Better than 10 f. Better than 10
–10
8.2 dB when measured per TIA/EIA OFSTP-4A.
–10
BER when tested with PRBS 223–1 pattern.
BER when tested with PRBS 223–1 pattern.
g. Measured on an oscilloscope eye diagram with PRBS 2
m core multimode fiber may be slightly lower.
µ
.
°C ± 5°
23
–1 test pattern.
3-5
Page 32
Specifications and Regulatory Information

HP 83493A Specifications

HP 83493A Specifications
Table 3-3. HP 83493A Specifications
Wavelength range (characteristic) 1000 nm to 1600 nm
Optical INPUT fiber (characteristic) 9/125 single mode
Optical insertion loss (through path)
Optical return loss
a
1.5 dB
28 dB
Clock recovery rates (NRZ coding)
155.52 Mb/s
622.08 Mb/s 1250 Mb/s
2488.32 Mb/s
2500.00 Mb/s
±0.1% ±0.1% ±0.1% ±0.1% ±0.1%
Data triggering (characteristic) 50 Mb/s to 2500 Mb/s
Operating input power level
Triggering operation, all rates
–10
10
BER, all rates
Output jitter, all rates
b c
–20 dBm to 3 dBm
d
e
–17 dBm to 3 dBm
0.0125 UI
rms
Maximum continuous optical power before damage (characteristic) 10 mW peak
DATA and CLOCK output amplitude, all rates (characteristic) 0.5Vp-p
DATA and CLOCK output
50 MHz through 2000 MHz (characteristic) 2000 MHz through 2500 MHz (characteristic)
a. Tested with FC/PC adapter. Optical output terminated without > 33 dB return loss. b. Source extinction ratio ≥ 8.2 dB when measured per TIA/EIA OFSTP-4A. c. Operating power level applies over temperature range 25 d. Better than 10 e. Measured on an oscilloscope eye diagram with PRBS 223–1 test pattern.
electrical return loss
–10
BER when tested with PRBS 223–1 pattern.
°C ± 5°
10 dB
6 dB
.
3-6
Page 33
Specifications and Regulatory Information
HP 83491/2/3A Operating Specifications
HP 83491/2/3A Operating Specifications
Table 3-4. HP 83491/2/3A Operating Specifications
Use Iindoor
Temperature
Operating Non-operating
Altitude
Operating Non-operating
Humidity
0°C to +55°C
°
–40
C to +70°C
4600 m (15,000 ft) 15,300 m (50,000 ft)
Operating Non-operating
Net weight approximately 1.2 kg (2.6 lb.)
Shipping weight approximately 2.1 kg (4.6 lb.)
Power Requirements Supplied by mainframe
up to 90% relative humidity at <35 up to 90% relative humidity at <35
°
C
°
C
3-7
Page 34
Specifications and Regulatory Information

Declaration of Conformity

Declaration of Conformity
3-8
Page 35
4
Front-Panel Optical Adapters 4-2 In Case of Difficulty 4-3 Error Messages 4-5 Electrostatic Discharge Information 4-8 Cleaning Connections for Accurate Measurements 4-10 Returning the Module for Service 4-21 Hewlett-Packard Sales and Service Offices 4-24

Reference

Page 36
Reference

Front-Panel Optical Adapters

Front-Panel Optical Adapters
Figure 4-1. Front-panel adapters for input and output connectors
4-2
Page 37

In Case of Difficulty

In Case of Difficulty
This section provides a list of suggestions for you to follow if the plug-in mod­ule fails to operate. A list of messages that may be displayed is also included in this chapter. Before calling Hewlett-Packard or returning the unit for service, a few minutes spent performing some simple checks may save waiting for your instrument to be repaired.

If the mainframe does not operate

Is the line fuse good?
Does the line socket have power?
Reference
Is the unit plugged in to the proper ac power source?
Is the mainframe turned on?
Is the rear-panel line switch set to on?
Will the mainframe power up
If the mainframe still does not power up, refer to the optional
HP 54750A Service Guide
department.
without
or return the mainframe to a qualified service
the plug-in module installed?
HP 83480A,

If the plug-in does not operate

Is the plug-in module firmly seated in the mainframe slot?
Are the knurled screws at the bottom of the plug-in module finger-tight?
Is the clock recovery module set to the modulation rate of the input signal?
If other equipment, cables, and connectors are being used with the plug-in module, are they connected properly and operating correctly?
Review the procedure for the test being performed when the problem ap-
4-3
Page 38
Reference
In Case of Difficulty
peared. Are all the settings correct? Can the problem be reproduced?
Are the connectors clean? See “Cleaning Connections for Accurate Measure-
ments” on page 4-10 for more information.
Perform the following procedures:
1
Make sure that the instrument is ready to acquire data by pressing
2
Find any signals on the channel inputs by pressing
3
See if any signals are present at the channel inputs by pressing
Autoscale.
Run
Trigger,
.
Sweep,
Freerun.
4
After viewing the signal, press
5
Make sure Channel Display is on by pressing
6
Make sure the channel offset is adjusted so the waveform is not clipped off the
triggered.
Channel
Display on off, on
,
.
display.
7
Make sure the mainframe identifies the plug-in module by pressing
System config
....
Utility
The calibration status of the plug-in modules is listed near the bottom of the display, in the box labeled
“Plug-ins”.
If the model number of the plug­in module is listed next to the appropriate slot number, then the mainframe has identified the plug-in.
“~known”
If
is displayed instead of the model number of the plug-in
module, remove and reinsert the plug-in module in the same slot.
“~known”
If
is still displayed, the mainframe may need to have the latest operating system firmware installed. Options 001 and 002 provide this firmware on a 3.5 inch diskette. To load new firmware, follow the instructions provided with this diskette. If you do not have the optional diskette, contact your local HP Sales and Service Office (refer to “Hewlett-
Packard Sales and Service Offices” on page 4-24).
, then
If the mainframe firmware is current and the plug-in module is correctly installed, then the memory contents of the plug-in module are corrupt. Contact a qualified service department.
4-4
Page 39
Reference

Error Messages

Error Messages
The following error messages are for the plug-in module. Typically, the error messages indicate there is a problem with either the plug-in or the mainframe. This section explains what the messages mean and offers a few suggestions that might help resolve the error condition. If the suggestions do not eliminate the error message, then additional troubleshooting is required that is beyond the scope of this book. Additional error messages are listed in the
HP 54750A User’s Guide
for the mainframe.
Memory error occurred in plug-in_:Try reinstalling plug-in
HP 83480A,
The mainframe could not correctly read the contents of the memory in the plug-in.
Remove and reinstall the plug-in module. Each time a plug-in is installed, the mainframe re-reads the memory in the plug-in module.
Verify the plug-in module is firmly seated in the mainframe slot.
Verify the knurled screws at the bottom of the plug-in module are finger-tight.
Install the plug-in in a different slot in the mainframe.
4-5
Page 40
Reference
Busy timeout occurred with plug-in_:Try reinstalling plug-in
Busy timeout occurred with plug-in_:Try reinstalling plug-in
The mainframe is having trouble communicating with the plug-in module. Make sure there is a good connection between the mainframe and the plug-in module.
Remove and reinstall the plug-in module.
Verify the plug-in module is firmly seated in the mainframe slot.
Verify the knurled screws at the bottom of the plug-in module are finger-tight.
Install the plug-in in a different slot in the mainframe.
Communications failure exists at slot_:Service is required
An illegal hardware state is detected at the mainframe-to-plug-in module interface of the specified slot.
• If the slot is empty, there is a mainframe hardware problem. Refer to the
HP 83480A, HP 54750A Service Guide
• If a plug-in is installed in the slot, there is a plug-in module hardware problem. Return the plug-in module to a qualified service department.
.
ID error occurred in plug-in_:Service is required
The information read from the memory of the plug-in module does not match the hardware in the plug-in module. This can be caused by a communication problem between the mainframe and the plug-in module. Make sure there is a good connection between the mainframe and the plug-in.
Remove and re-install the plug-in module.
Verify the plug-in module is firmly seated in the mainframe slot.
Verify the knurled screws at the bottom of the plug-in module are finger tight.
The standard HP 54750A mainframe does not accept the HP 83491/2/3A Light­wave Clock Recovery Modules module. To use the module, a firmware upgrade
4-6
Page 41
Plug-in is not supported:System firmware upgrade is needed
must first be installed. Order the HP 83480K communications firmware kit and install according to the instructions.
The HP 83480A, HP 54750A mainframes do not accept plug-in modules de­signed for use with the HP 54710A, 54720A.
Plug-in is not supported:System firmware upgrade is needed
The mainframe may need to have the latest operating system firmware installed. Options 001 and 002 provide this firmware on a 3.5 inch diskette. To load the new firmware, follow the instructions provided with the diskette. If you do not have the optional diskette, contact your local HP Sales and Service Office.
Reference
4-7
Page 42
Reference

Electrostatic Discharge Information

Electrostatic Discharge Information
Electrostatic discharge (ESD) can damage or destroy electronic components. All work on electronic assemblies should be performed at a static-safe work station. The following figure shows an example of a static-safe work station using two types of ESD protection:
• Conductive table-mat and wrist-strap combination.
• Conductive floor-mat and heel-strap combination.
Figure 4-2. Static-safe work station
4-8
Page 43
Reference
Electrostatic Discharge Information
Both types, when used together, provide a significant level of ESD protection. Of the two, only the table-mat and wrist-strap combination provides adequate ESD protection when used alone.
To ensure user safety, the static-safe accessories must provide at least 1 MΩ of isolation from ground. Refer to Table 4-1 for information on ordering static­safe accessories.
WARNING
These techniques for a static-safe work station should not be used when working on circuitry with a voltage potential greater than 500 volts.

Reducing ESD Damage

The following suggestions may help reduce ESD damage that occurs during testing and servicing operations.
• Personnel should be grounded with a resistor-isolated wrist strap before re­moving any assembly from the unit.
• Be sure all instruments are properly earth-grounded to prevent a buildup of static charge.
Table 4-1. Static-Safe Accessories
HP Part Number Description
9300-0797
Set includes: 3M static control mat 0.6 m × 1.2 m (2 ft× 4 ft) and 4.6 cm (15 ft) ground wire. (The wrist-strap and wrist-strap cord are not included. They must be ordered separately.)
9300-0980 Wrist-strap cord 1.5 m (5 ft)
9300-1383 Wrist-strap, color black, stainless steel, without cord, has four adjustable
links and a 7 mm post-type connection.
9300-1169 ESD heel-strap (reusable 6 to 12 months).
4-9
Page 44
Reference

Cleaning Connections for Accurate Measurements

Cleaning Connections for Accurate Measurements
Today, advances in measurement capabilities make connectors and connec­tion techniques more important than ever. Damage to the connectors on cali­bration and verification devices, test ports, cables, and other devices can degrade measurement accuracy and damage instruments. Replacing a dam­aged connector can cost thousands of dollars, not to mention lost time! This expense can be avoided by observing the simple precautions presented in this book. This book also contains a brief list of tips for caring for electrical connec­tors.

Choosing the Right Connector

A critical but often overlooked factor in making a good lightwave measure­ment is the selection of the fiber-optic connector. The differences in connector types are mainly in the mechanical assembly that holds the ferrule in position against another identical ferrule. Connectors also vary in the polish, curve, and concentricity of the core within the cladding. Mating one style of cable to another requires an adapter. Hewlett Packard offers adapters for most instru­ments to allow testing with many different cables. The Figure 4-3 on page 4-11 shows the basic components of a typical connectors.
The system tolerance for reflection and insertion loss must be known when selecting a connector from the wide variety of currently available connectors. Some items to consider when selecting a connector are:
• How much insertion loss can be allowed?
• Will the connector need to make multiple connections? Some connectors are better than others, and some are very poor for making repeated connections.
• What is the reflection tolerance? Can the system take reflection degradation?
• Is an instrument-grade connector with a precision core alignment required?
• Is repeatability tolerance for reflection and loss important? Do your specifica-
4-10
Page 45
Reference
Cleaning Connections for Accurate Measurements
tions take repeatability uncertainty into account?
• Will a connector degrade the return loss too much, or will a fusion splice be re­quired? For example, many DFB lasers cannot operate with reflections from connectors. Often as much as 90 dB isolation is needed.
Figure 4-3. Basic components of a connector.
Over the last few years, the FC/PC style connector has emerged as the most popular connector for fiber-optic applications. While not the highest perform­ing connector, it represents a good compromise between performance, reli­ability, and cost. If properly maintained and cleaned, this connector can withstand many repeated connections.
However, many instrument specifications require tighter tolerances than most connectors, including the FC/PC style, can deliver. These instruments cannot tolerate connectors with the large non-concentricities of the fiber common with ceramic style ferrules. When tighter alignment is required, HP instru­ments typically use a connector such as the Diamond HMS-10, which has con­centric tolerances within a few tenths of a micron. HP then uses a special universal adapter, which allows other cable types to mate with this precision connector. See Figure 4-4 on page 4-12.
4-11
Page 46
Reference
Cleaning Connections for Accurate Measurements
Figure 4-4. Universal adapters to Diamond HMS-10
The HMS-10 encases the fiber within a soft nickel silver (Cu/Ni/Zn) center which is surrounded by a tough tungsten carbide casing, as shown in
Figure 4-5.
Figure 4-5. Cross-section of the Diamond HMS-10 connector.
The nickel silver allows an active centering process that permits the glass fiber to be moved to the desired position. This process first stakes the soft nickel silver to fix the fiber in a near-center location, then uses a post-active staking to shift the fiber into the desired position within 0.2 µm. This process, plus the keyed axis, allows very precise core-to-core alignments. This connector is found on most HP lightwave instruments.
4-12
Page 47
Reference
Cleaning Connections for Accurate Measurements
The soft core, while allowing precise centering, is also the chief liability of the connector. The soft material is easily damaged. Care must be taken to mini­mize excessive scratching and wear. While minor wear is not a problem if the glass face is not affected, scratches or grit can cause the glass fiber to move out of alignment. Also, if unkeyed connectors are used, the nickel silver can be pushed onto the glass surface. Scratches, fiber movement, or glass contamina­tion will cause loss of signal and increased reflections, resulting in poor return loss.

Inspecting Connectors

Because fiber-optic connectors are susceptible to damage that is not immedi­ately obvious to the naked eye, bad measurements can be made without the user even being aware of a connector problem. Although microscopic exami­nation and return loss measurements are the best way to ensure good connec­tions, they are not always practical. An awareness of potential problems, along with good cleaning practices, can ensure that optimum connector perfor­mance is maintained. With glass-to-glass interfaces, it is clear that any degra­dation of a ferrule or the end of the fiber, any stray particles, or finger oil can have a significant effect on connector performance.
Figure 4-6 shows the end of a clean fiber-optic cable. The dark circle in the
center of the micrograph is the fiber’s 125 µm core and cladding which carries the light. The surrounding area is the soft nickel-silver ferrule. Figure 4-7 shows a dirty fiber end from neglect or perhaps improper cleaning. Material is smeared and ground into the end of the fiber causing light scattering and poor reflection. Not only is the precision polish lost, but this action can grind off the glass face and destroy the connector.
Figure 4-8 shows physical damage to the glass fiber end caused by either
repeated connections made without removing loose particles or using improper cleaning tools. When severe, the damage on one connector end can be transferred to another good connector that comes in contact with it.
The cure for these problems is disciplined connector care as described in the following list and in “Cleaning Connectors” on page 4-16.
Use the following guidelines to achieve the best possible performance when making measurements on a fiber-optic system:
• Never use metal or sharp objects to clean a connector and never scrape the connector.
• Avoid matching gel and oils.
4-13
Page 48
Reference
Cleaning Connections for Accurate Measurements
Figure 4-6. Clean, problem-free fiber end and ferrule.
Figure 4-7. Dirty fiber end and ferrule from poor cleaning.
Figure 4-8. Damage from improper cleaning.
While these often work well on first insertion, they are great dirt magnets. The oil or gel grabs and holds grit that is then ground into the end of the fiber. Also, some early gels were designed for use with the FC, non-contacting con-
4-14
Page 49
Reference
Cleaning Connections for Accurate Measurements
nectors, using small glass spheres. When used with contacting connectors, these glass balls can scratch and pit the fiber. If an index matching gel or oil must be used, apply it to a freshly cleaned connector, make the measurement, and then immediately clean it off. Never use a gel for longer-term connections and never use it to improve a damaged connector. The gel can mask the extent of damage and continued use of a damaged fiber can transfer damage to the instrument.
• When inserting a fiber-optic cable into a connector, gently insert it in as straight a line as possible. Tipping and inserting at an angle can scrape material off the inside of the connector or even break the inside sleeve of connectors made with ceramic material.
• When inserting a fiber-optic connector into a connector, make sure that the fi­ber end does not touch the outside of the mating connector or adapter.
• Avoid over tightening connections.
Unlike common electrical connections, tighter is
better. The purpose of
not
the connector is to bring two fiber ends together. Once they touch, tightening only causes a greater force to be applied to the delicate fibers. With connec­tors that have a convex fiber end, the end can be pushed off-axis resulting in misalignment and excessive return loss. Many measurements are actually improved by backing off the connector pressure. Also, if a piece of grit does happen to get by the cleaning procedure, the tighter connection is more likely to damage the glass. Tighten the connectors just until the two fibers touch.
• Keep connectors covered when not in use.
• Use fusion splices on the more permanent critical nodes. Choose the best con­nector possible. Replace connecting cables regularly. Frequently measure the return loss of the connector to check for degradation, and clean every connec­tor, every time.
All connectors should be treated like the high-quality lens of a good camera. The weak link in instrument and system reliability is often the inappropriate use and care of the connector. Because current connectors are so easy to use, there tends to be reduced vigilance in connector care and cleaning. It takes only one missed cleaning for a piece of grit to permanently damage the glass and ruin the connector.
Measuring insertion loss and return loss
Consistent measurements with your lightwave equipment are a good indica­tion that you have good connections. Since return loss and insertion loss are key factors in determining optical connector performance they can be used to determine connector degradation. A smooth, polished fiber end should pro-
4-15
Page 50
Reference
Cleaning Connections for Accurate Measurements
duce a good return-loss measurement. The quality of the polish establishes the difference between the “PC” (physical contact) and the “Super PC” con­nectors. Most connectors today are physical contact which make glass-to-glass connections, therefore it is critical that the area around the glass core be clean and free of scratches. Although the major area of a connector, excluding the glass, may show scratches and wear, if the glass has maintained its polished smoothness, the connector can still provide a good low level return loss con­nection.
If you test your cables and accessories for insertion loss and return loss upon receipt, and retain the measured data for comparison, you will be able to tell in the future if any degradation has occurred. Typical values are less than 0.5 dB of loss, and sometimes as little as 0.1 dB of loss with high performance con­nectors. Return loss is a measure of reflection: the less reflection the better (the larger the return loss, the smaller the reflection). The best physically contacting connectors have return losses better than 50 dB, although 30 to 40 dB is more common.
Visual inspection of fiber ends
Visual inspection of fiber ends can be helpful. Contamination or imperfections on the cable end face can be detected as well as cracks or chips in the fiber itself. Use a microscope (100X to 200X magnification) to inspect the entire end face for contamination, raised metal, or dents in the metal as well as any other imperfections. Inspect the fiber for cracks and chips. Visible imperfec­tions not touching the fiber core may not affect performance (unless the imperfections keep the fibers from contacting).

Cleaning Connectors

The procedures in this section provide the proper steps for cleaning fiber­optic cables and HP universal adapters. The initial cleaning, using the alcohol as a solvent, gently removes any grit and oil. If a caked-on layer of material is still present, (this can happen if the beryllium-copper sides of the ferrule retainer get scraped and deposited on the end of the fiber during insertion of the cable), a second cleaning should be performed. It is not uncommon for a cable or connector to require more than one cleaning.
4-16
Page 51
Reference
Cleaning Connections for Accurate Measurements
CAUTION
Hewlett-Packard strongly recommends that index matching compounds
not
be applied to their instruments and accessories. Some compounds, such as gels, may be difficult to remove and can contain damaging particulates. If you think the use of such compounds is necessary, refer to the compound manufacturer for information on application and cleaning procedures.
Table 4-2. Cleaning Accessories
Item HP Part Number
Isopropyl alcohol 8500-5344
Cotton swabs 8520-0023
Small foam swabs 9300-1223
Compressed dust remover (non-residue) 8500-5262
Table 4-3. Dust Caps Provided with Lightwave Instruments
Item HP Part Number
Laser shutter cap 08145-64521
CAUTION
FC/PC dust cap 08154-44102
Biconic dust cap 08154-44105
DIN dust cap 5040-9364
HMS10/HP dust cap 5040-9361
ST dust cap 5040-9366
To clean a non-lensed connector
Do not use any type of foam swab to clean optical fiber ends. Foam swabs can leave filmy deposits on fiber ends that can degrade performance.
1
Apply pure isopropyl alcohol to a clean lint-free cotton swab or lens paper.
Cotton swabs can be used as long as no cotton fibers remain on the fiber end after cleaning.
4-17
Page 52
Reference
Cleaning Connections for Accurate Measurements
2
Clean the ferrules and other parts of the connector while avoiding the end of the fiber.
3
Apply isopropyl alcohol to a new clean lint-free cotton swab or lens paper.
4
Clean the fiber end with the swab or lens paper.
Do
scrub during this initial cleaning because grit can be caught in the
not
swab and become a gouging element.
5
Immediately dry the fiber end with a clean, dry, lint-free cotton swab or lens paper.
6
Blow across the connector end face from a distance of 6 to 8 inches using filtered, dry, compressed air. Aim the compressed air at a shallow angle to the fiber end face.
Nitrogen gas or compressed dust remover can also be used.
CAUTION
Do not shake, tip, or invert compressed air canisters, because this releases particles in the can into the air. Refer to instructions provided on the compressed air canister.
7
As soon as the connector is dry, connect or cover it for later use.
If the performance, after the initial cleaning, seems poor try cleaning the con­nector again. Often a second cleaning will restore proper performance. The second cleaning should be more arduous with a scrubbing action.
To clean an adapter
The fiber-optic input and output connectors on many HP instruments employ a universal adapter such as those shown in the following picture. These adapt­ers allow you to connect the instrument to different types of fiber-optic cables.
Figure 4-9. Universal adapters
1
Apply isopropyl alcohol to a clean foam swab.
Cotton swabs can be used as long as no cotton fibers remain after cleaning. The foam swabs listed in this section’s introduction are small enough to fit into
4-18
Page 53
Reference
Cleaning Connections for Accurate Measurements
adapters.
Although foam swabs can leave filmy deposits, these deposits are very thin, and the risk of other contamination buildup on the inside of adapters greatly out­weighs the risk of contamination by foam swabs.
2
Clean the adapter with the foam swab.
3
Dry the inside of the adapter with a clean, dry, foam swab.
4
Blow through the adapter using filtered, dry, compressed air.
Nitrogen gas or compressed dust remover can also be used. Do not shake, tip, or invert compressed air canisters, because this releases particles in the can into the air. Refer to instructions provided on the compressed air canister.

Caring for electrical connections

The following list includes the basic principles of microwave connector care. For more information on microwave connectors and connector care, consult the
Hewlett-Packard Microwave Connector Care Manual
08510-90064.
, HP part number
Handling and Storage
• Keep connectors clean
• Extend sleeve or connector nut
• Use plastic endcaps during storage
•Do
•Do
• Inspect all connectors carefully before every connection
• Look for metal particles, scratches, and dents
•Do
• Try cleaning with compressed air first
• Clean the connector threads
•Do
•Do
touch mating plane surfaces
not
set connectors contact-end down
not
Visual Inspection
use damaged connectors
not
Cleaning
use abrasives
not
get liquid onto the plastic support beads
not
4-19
Page 54
Reference
Cleaning Connections for Accurate Measurements
Making Connections
• Align connectors carefully
• Make preliminary connection lightly
• To tighten, turn connector nut
•Do
•Do
•Do
•Do
apply bending force to connection
not
overtighten preliminary connection
not
twist or screw in connectors
not
tighten past the “break” point of the torque wrench
not
only
4-20
Page 55
Reference

Returning the Module for Service

Returning the Module for Service
The instructions in this section show you how to properly return the module for repair or calibration. Always call the HP Instrument Support Center first to initiate service that the repair (or calibration) can be properly tracked and that your module will be returned to you as quickly as possible. Call this number regardless of where you are located. Refer to “Hewlett-Packard Sales and Service Offices”
on page 4-24 for a list of service offices.
HP Instrument Support Center . . . . . . . . . . . . . . . . . . . . . . . . . (800) 403-0801
If the module is still under warranty or is covered by an HP maintenance con­tract, it will be repaired under the terms of the warranty or contract (the war­ranty is at the front of this manual). If the module is no longer under warranty or is not covered by an HP maintenance plan, Hewlett-Packard will notify you of the cost of the repair after examining the unit.
When an module is returned to a Hewlett-Packard service office for servicing, it must be adequately packaged and have a complete description of the failure symptoms attached. When describing the failure, please be as specific as pos­sible about the nature of the problem. Include copies of additional failure information (such as the instrument failure settings, data related to module failure, and error messages) along with the original cal data disks and the module being returned.
returning your module to a service office. This ensures
before

Preparing the module for shipping

1
Write a complete description of the failure and attach it to the module. Include any specific performance details related to the problem. The following
4-21
Page 56
Reference
Returning the Module for Service
information should be returned with the module.
• Type of service required.
• Date module was returned for repair.
• Description of the problem:
• Whether problem is constant or intermittent.
• Whether module is temperature-sensitive.
• Whether module is vibration-sensitive.
• Module settings required to reproduce the problem.
• Performance data.
• Company name and return address.
• Name and phone number of technical contact person.
• Model number of returned module.
• Full serial number of returned module.
• List of any accessories returned with module.
2
Cover all front or rear-panel connectors that were originally covered when you first received the module.
CAUTION
CAUTION
Cover electrical connectors to protect sensitive components from electrostatic damage. Cover optical connectors to protect them from damage due to physical contact or dust.
Module damage can result from using packaging materials other than the original materials. Never use styrene pellets as packaging material. They do not adequately cushion the module or prevent it from shifting in the carton. They may also cause module damage by generating static electricity.
3
Pack the module in the original shipping containers. Original materials are available through any Hewlett-Packard office. Or, use the following guidelines:
• Wrap the module in antistatic plastic to reduce the possibility of damage caused by electrostatic discharge.
• For modules weighing less than 54 kg (120 lb), use a double-walled, corru­gated cardboard carton of 159 kg (350 lb) test strength.
• The carton must be large enough to allow approximately 7 cm (3 inches) on all sides of the module for packing material, and strong enough to accommo­date the weight of the module.
• Surround the equipment with approximately 7 cm (3 inches) of packing ma­terial, to protect the module and prevent it from moving in the carton. If packing foam is not available, the best alternative is S.D-240 Air Cap™ from Sealed Air Corporation (Commerce, California 90001). Air Cap looks like a
4-22
Page 57
Returning the Module for Service
plastic sheet filled with air bubbles. Use the pink (antistatic) Air Cap™ to reduce static electricity. Wrapping the module several times in this material will protect the module and prevent it from moving in the carton.
4
Seal the carton with strong nylon adhesive tape.
5
Mark the carton “FRAGILE, HANDLE WITH CARE”.
6
Retain copies of all shipping papers.
Reference
4-23
Page 58
Reference

Hewlett-Packard Sales and Service Offices

Hewlett-Packard Sales and Service Offices
Before returning an instrument for service, call the HP Instrument Support Center at (800) 403-0801.
Hewlett-Packard Sales and Service Offices (1 of 2)
U.S. FIELD OPERATIONS
Headquarters
Hewlett-Packard Company 19320 Pruneridge Avenue Cupertino, CA 95014 U.S.A. (800) 752-0900
Colorado
Hewlett-Packard Company 24 Inverness Place, East Englewood, CO 80112 (303) 649-5000
New Jersey
Hewlett-Packard Company 150 Green Pond Road, Dock 1 Rockaway, NJ 07866 (201) 586-5910
EUROPEAN FIELD OPERATIONS
Headquarters
Hewlett-Packard S.A. 150, Route du Nant-d’Avril 1217 Meyrin 2/Geneva Switzerland (41 22) 780.8111
Great Britain
Hewlett-Packard Ltd. Eskdale Road, Winnersh Triangle Wokingham, Berkshire RG11 5DZ
California, Northern
Hewlett-Packard Company 301 East Evelyn Mountain View, CA 94041 (415) 694-2000
Georgia
Hewlett-Packard Company 2000 South Park Place Atlanta, GA 30339 (404) 955-1500
Texas
Hewlett-Packard Company 930 East Campbell Road Richardson, TX 75081 (214) 231-6101
France
Hewlett-Packard France 1 Avenue Du Canada Zone D’Activite De Courtaboeuf F-91947 Les Ulis Cedex France (33 1) 69 82 60 60
California, Southern
Hewlett-Packard Company 1421 South Manhatten Ave. Fullerton, CA 92631 (714) 999-6700
Illinois
Hewlett-Packard Company 5201 Tollview Drive Rolling Meadows, IL 60008 (708) 342-2000
Germany
Hewlett-Packard GmbH Hewlett-Packard Strasse 61352 Bad Homburg Germany (+49 6172) 16-0
4-24
Page 59
Hewlett-Packard Sales and Service Offices (2 of 2)
INTERCON FIELD OPERATIONS
Headquarters
Hewlett-Packard Company 3495 Deer Creek Rd. Palo Alto, California 94304-1316 (415) 857-5027
Australia
Hewlett-Packard Australia Ltd. 31-41 Joseph Street Blackburn, Victoria 3130 (61 3) 895-2895
Reference
Hewlett-Packard Sales and Service Offices
Canada
Hewlett-Packard Ltd. 17500 South Service Road Trans-Canada Highway Kirkland, Quebec H9J 2X8 Canada (514) 697-4232
China
China Hewlett-Packard Company 38 Bei San Huan X1 Road Shuang Yu Shu Hai Dian District Beijing, China (86 1) 256-6888
Taiwan
Hewlett-Packard Taiwan 8th Floor, H-P Building 337 Fu Hsing North Road Taipei, Taiwan (886 2) 712-0404
Japan
Hewlett-Packard Japan Ltd. 9-1, Takakura-Cho, Hachioji-Shi, Tokyo, 192, Japan (81 427) 59-1311
Singapore
Hewlett-Packard Singapore Ltd. Pte. Ltd. Alexandra P.O. Box 87 Singapore 9115 (65) 271-9444
4-25
Page 60
Reference
Hewlett-Packard Sales and Service Offices
4-26
Page 61

Index

A
adapter cable, 1-3, 1-4
B
Bellcore GR-253-CORE, 2-5 block diagram, 2-7 bypass mode, 2-4, 2-5
C
calibration cycle, 3-2 care of fiber optics, 1-2 channel
key, 1-2
setup, 1-2 Channel key, 2-3 characteristics, defined, 3-2 cleaning
adapters, 4-18
fiber-optic connections, 4-10, 4-17
non-lensed connectors, 4-17 CLOCK Auxiliary Output connector, 2-5 clock recovery input connector, 4-21 compensation
HP 54701A active probe, 2-11
insertion loss, 2-9
other devices, 2-11
passive probe, 2-11 compressed dust remover, 4-17 connector
care, 4-10
care manual, 4-19 cotton swabs, 4-17 customer assistance, iii
D
damaged shipment, 1-3 DATA Auxiliary Output connector, 2-5 data rate
indicator lights, 1-5 data rates, 2-2
multiple of, 2-8 declaration of conformity, 3-8 dust caps, 4-17
E
electrostatic discharge, 1-2, 4-8 error messages, 4-5 error performance analyzers, use with, 2-3 ESD
reducing damage caused by ESD, 4-9 static-safe work station, 4-9
extender cables, 1-3
F
fiber
adapters, 4-2 care of, 1-2 cleaning connections, 4-10 connections on HP 83492A, 1-7 connectors, covering, 4-22 multimode, 2-2 reflections, 2-6 single-mode, 2-2 single-mode connections, 2-6
firmware
upgrade kit, 1-2
version required, 1-2 foam swabs, 4-17 front panel, 2-4
adapters, 4-2
features, 2-4
G
GPIB programming, 2-3 green light, 1-5
H
HP 1163A probe, 2-10 HP 54006A low impedance probe, 2-10 HP 54701A active probe, 2-10, 2-11 HP 54750A digitizing oscilloscope, 1-2 HP 83492A
connections, 1-7
maximum input level, 1-5 HP Instrument Support Center, 4-21 HP N1020A TDR probe, 2-10 HP offices, 4-24
Index-1
Page 62
Index
I
IEC 1010, iii IEC 664, iii Input connector, 1-2, 1-7, 4-10 input connector, 2-2 input signal, maximum safe, 1-5 insertion loss, 2-3, 2-6
compensation, 2-9 INSTALLATION CATEGORY II, iii isopropyl alcohol, 4-17 ITU-T G.957, 2-5
M
mainframe troubleshooting, 4-3 maximum input level, 1-5 module
returning for service, 4-21
O
Output connector, 2-5
P
packaging for shipment, 4-22 plug-in module
serial number, 1-3 pollution degree 2, iii programming, 2-3
R
red light, 1-5 regulatory information, 3-2 responsivity, testing, 1-4 returning for service, 4-21
procedure, 4-21 single-mode fiber, 2-6 specifications, 3-2
defined, 3-2
HP 83491A, 3-3
HP 83492A, 3-4
HP 83493A, 3-6
operating, 3-7
temperature range, 3-2 swabs, 4-17
T
technical assistance, iii testing, responsivity, 1-4 trigger bypass mode, 2-4, 2-5 Trigger On Data light, 2-4 troubleshooting, 4-3
U
Unlocked light, 1-6, 2-5, 2-8
S
safety information, iii, 1-3, 2-3 sales and service offices, 4-24 SELECT key, 1-5, 2-4 serial number, 1-3 service, 4-21
returning for, 4-21 sales and service offices, 4-24
shipping
Index-2
Loading...