HP 745a schematic

*TB 9-4931-700-50
CHANGE 6
DEPARTMENT OF THE ARMY TECHNICAL BULLETIN
CALIBRATION PROCEDURE FOR
AC VOLTAGE STANDARD,
HEWLETT-PACKARD MODEL 745A
HEWLETT-PACKARD MODEL 746A
INCLUDING OPTIONS C90, C91, AND C93
Headquarters, Department of the Army, Washington, DC
11 September 2001
Approved for public release; distribution is unlimited.
TB 9-4931-700-50, 9 December 1977, is changed as follows:
1. Remove old pages and insert new pages as indicated below. New or changed material is indicated by a vertical bar in the margin of the page.
Remove pages Insert pages
1 and 2 1 and 2
13 and 14 13 and 14
2. File this change sheet in front of the publication for reference purposes.
By Order of the Secretary of the Army:
0121914
Distribution: To be distributed in accordance with Std IDS No. RLC-1500, 11 May 1992, requirements for calibration procedure TB 9-4931-700-50.
PIN NO: 009082-006
*TB 9-4931-700-50
Change 5
DEPARTMENT OF THE ARMY TECHNICAL BULLETIN
CALIBRATION PROCEDURE FOR
AC VOLTAGE STANDARD,
HEWLETT-PACKARD MODEL 745A
AND HIGH VOLTAGE AMPLIFIER,
HEWLETT-PACKARD MODEL 746A
INCLUDING OPTIONS C90, C91, AND C93
Headquarters, Department of the Army, Washington, DC
15 May 1989
TB 9-4931-700-50, 9 December 1977, is changed as follows:
1. Remove old pages and insert new pages as indicated below. New or changed material is indicated by a vertical bar in the margin of the page.
Remove pages Insert pages
3 through 6 3 through 6
2. File this change sheet in front of the publication for reference purposes. By Order of the Secretary of the Army:
CARL E. VUONO
General, United States Army Chief of Staff
Official:
WILLIAM J. MEEHAN ii
Brigadier General, United States Army The Adjutant General
Distribution: To be distributed in accordance with Std IDS No. RLC-1500, 25 January 1989.
PIN NO: 009082-005
*TB 9-4931-700-50
DEPARTMENT OF THE ARMY TECHNICAL BULLETIN
CALIBRATION PROCEDURE FOR
AC VOLTAGE STANDARD,
HEWLETT-PACKARD MODEL 745A
AND HIGH VOLTAGE AMPLIFIER,
HEWLETT-PACKARD MODEL 746A
INCLUDING OPTIONS C90, C91, AND C93
Headquarters, Department of the Army, Washington, DC
19 December 1988
TB 9-4931-700-50, 9 December 1977, is changed as follows:
Change 4
1. Remove old pages and insert new pages as indicated below. New or changed material is indicated by a vertical bar in the margin of the page.
Remove pages Insert pages
4 and 5 4 and 5
2. File this change sheet in front of the publication for reference purposes. By Order of the Secretary of the Army:
CARL E. VUONO
General, United States Army Chief of Staff
Official:
WILLIAM J. MEEHAN ii
Major General, United States Army The Adjutant General
Distributed: To be distributed in accordance with STD IDS No. RLC-1500, 18 November 1987.
PIN NO: 009082-004
This is a REPRINT, including changes 1 through 3.
............................
..................
.......
...........
..........
....
...............
...........
...........
.....
*TB 9-4931-700-50
DEPARTMENT OF THE ARMY TECHNICAL BULLETIN
CALIBRATION PROCEDURE FOR
AC VOLTAGE STANDARD,
HEWLETT-PACKARD MODEL 745A
AND HIGH VOLTAGE AMPLIFIER,
HEWLETT-PACKARD MODEL 746A
INCLUDING OPTIONS C90, C91, AND C93
Headquarters, Department of the Army, Washington, DC
9 December 1977
REPORTING OF ERRORS
You can help improve this publication. If you find any mistakes or if you know of a way to improve the procedure, please let us know. Mail your letter or DA Form 2028 to: Commander, U. S. Army Aviation and Missile Command, ATTN: AMSAM-MMC­MA-NP, Redstone Arsenal, AL 35898-5230. A reply will be furnished to you. You may also send in your comments electronically to our e-mail address: ls-
lp@redstone.army.mil or FAX 256-842-6546/DSN 788-6546.
Paragraph Page
SECTION I. IDENTIFICATION AND DESCRIPTION
Test instrument identification Calibration data card, DA Form 2416
Calibration description................................
II. EQUIPMENT REQUIREMENTS
Equipment required ................................
Accessories required................................
III. PRELIMINARY OPERATIONS
Preliminary instructions................................
Equipment Setup................................
IV. CALIBRATION PROCESS
Frequency accuracy ................................
Error measurement................................
Voltage accuracy, regulation, and frequency
response (Model 745A) ................................
__________ *This bulletin supersedes TB 9-4931-700-60, 6 October 1976. CHANGE 6
1 2 2 2 3 2
4 3 5 3
6 6 7 7
8 7 9 12
10 14
TB 9-4931-700-50
.....
..........................
................................
................................
..................
Paragraph Page Voltage accuracy, regulation, and frequency
response (Model 746A) ................................
Distortion ................................
Power supply (Model 745A) Power supply (Model 746A)
Final procedure................................
11 20 12 26 13 28 14 30 15 31

SECTION I

IDENTIFICATION AND DESCRIPTION
1. Test Instrument Identification. This bulletin provides instructions for the
calibration of Ac Voltage Standard, Hewlett-Packard Model 745A and High Voltage Amplifier, Hewlett-Packard Model 746A including Options C90, C91, and C93. The manufacturer's instruction manuals were used as the prime data source in compiling these instructions. The equipment being calibrated will be referred to as the TI (test instrument) throughout this bulletin.
a. Model Variations. There are no variations among models which affect the calibration of the TI. The C90 option adds slides to the instrument. The C91 option adds a blower to the instrument. The C93 option adds slides and a blower to the instrument.
b. Time and Technique . The time required for this calibration is approximately 8 hours, using the dc (direct current) and low frequency technique.
2. Calibration Data Card, DA Form 2416
a. Forms, records, and reports required for calibration personnel at all levels arc prescribed by TM 38-750. DA Form 2416 must be annotated in accordance with TM 38­750 for ea ch calibration performed.
b. Adjustments to be reported on DA Form 2416 are designated (R) at the end of the sentence in which they appear. When adjustments are in tables, the (R) follows the designated adjustment. Report only those adjustments made and designated with (R).
3. Calibration Description. TI parameters and performance specifications which pertain to this calibration are listed in table 1.
Table 1. Calibration Description
Test instrument parameters Performance specifications
Output frequency range Continuously adjustable from 10 Hz to 110 kHz in 4
decade ranges with 10% overlap Frequency accuracy ±2% of setting Error measurement range ±0.3% and ±0.3% with zero center dial Error measurement accuracy ±0.51% of FS Distortion and noise (0.05% of setting + 10 µ V)
See footnote at end of table.
CHANGE 6 2
TB 9-4931-700-50
Table 1. Calibration Description - Continued.
Test Instrument Parameters Performance Specifications
Line regulation ±0.001% of setting change in output voltage for a 10%
change in line voltage (included in accuracy specifications)
Voltage stability 1 mV to 100 V: short term; ±0.002% of setting for 24
hrs. 1000 V: short term; ±0.005% of setting for 24 hrs
50 Hz to 20
kHz
1 -mV 10-mV, and 100-mV
ranges
50 Hz to 20
Output
voltage
50 Hz to 20
1
Combined accuracy and load regulation.
1-V, 10-V, and 100-V ranges 20 to 50 Hz
1000-V range
20 to 50 Hz
and 20 to 110 kHz 10 to 20 Hz
kHz
and 20 to 110 kHz 10 to 20 Hz
kHz 20 to 50 Hz
and 20 to 50 kHz 10 to 20 Hz 50 to 110 kHz
±(0.04% of setting +0.002% of range
+ 10µ V)1 ±(0.07% of setting +0.005% of range +50µ V) ±(0.22% of setting +0.005% of range
+50µ V) ±(0.032% of setting +0.002% of
range +10µ V) ±(0.062% of setting +0.005% of
range +50µ V) ±(0.212% of setting +0.005% of
range +50µ V)
±0.04% of setting
±0.08% of setting
±(0.2% of setting +0.005% of range) ±0.15% of setting

SECTION II

EQUIPMENT REQUIREMENTS
4. Equipment Required. Table 2 identifies the specific equipment used in this
calibration procedure. This equipment is issued with Secondary Reference Standards Calibration Set NSN 4931-00-621-7878 and is to be used in performing this procedure. Alternate items may be used by the calibrating activity when the equipment listed in table 2 is not available. The items selected must be verified to perform satisfactorily prior to use and must bear evidence of current calibration. The equipment must meet or exceed the minimum use specifications listed in table 2. The accuracies listed in table 2 provide a four-to-one ratio between the standa rd and TI
5. Accessories Required. The accessories listed in table 3 are issued as indicated in paragraph 4 above and are to be used in this calibration procedure. When necessary, these items may be substituted by equivalent items unless specifically prohibited.
3
TB 9-4931-700-50
Table 2. Minimum Specifications of Equipment Required
Item
A1 AC/DC VOLTMETER
A2 AUTOTRANSFORMER
A3 DECADE RESISTOR Range: 22 to 3010 ohms Biddle Gray, Model 601147-1
A4 DISTORTION
A5 FREQUENCY
A6 LOAD (CALIBRATOR
A7 OSCILLOSCOPE Sensitivity: 0.01 to 20 V/cm
A8 RATION
A9 THERMAL TRANSFER
A10 VARIABLE
1
Combined accuracy of A1 and A9: ±0.0087%.
Common Name
(DIFFERENTIAL VOLTMETER)
(VARIABLE POWER TRANSFORMER)
ANALYZER (SPECTRUM ANALYZER)
COUNTER (ELECTRONIC DIGITAL COUNTER)
LOAD)
TRANSFORMER (AC RATIO STANDARD)
STANDARD
CAPACITOR (VARIABLE CAPACITANCE STANDARD)
Item
B1 ADAPTER (POWER ADAPTER) 2 to 2-wire isolation plug (7912356) B2 ADAPTER (ELECTRICAL PLUG
CONNECTOR)
B3 ADAPTER (CONNECTOR
ADAPTER)
B4 ADAPTER (TIP JACK
ASSEMBLY)
B5 CABLE (RF CABLE ASSEMBLY) 30-in., RG -58( )/U with BNC plug terminations
B6 CABLE (RF CABLE ASSEMBLY) 36-in., RG -58( )/U with BNC plug to alligator clips
Common Name
(Official Nomenclature)
Minimum Use
Specifications Range: 0.002 to 1850 V dc Accuracy: 1 Capability: Use as null detector Range: 105 to 125 V ac Accuracy: ±1%
Range: 10 Hz to 100 kHz Accuracy: Measure less than
.06% (Must have amplifier
capability) Range: 9.8 Hz to 119 kHz Accuracy: ±0.5%
Range: 2200 ohms Accuracy: ±5%
Sweeptime: 2 ms to 10 ms/cm
Capability: Divide to 7 places Accuracy: ±0.0085%
Range: 1 to 100 V dc 1 to 100 V
ac Accuracy: 1 Range: 100 and 200 pF General Radio, Model 1422D
Table 3. Accessories Required
(Part Number)
Double banana plug to BNC jack terminations
(7907592)
Double banana jack to phone plug adapter (7907566)
GR plug to binding posts 874Q2 (874Q2)
(7907467)
(7909410)
Manufacturer and Model
(Part Number)
John Fluke, Models
887ABAN (887 ABAN) and 80E10 (80E10)
General Radio, Model
W10MT3AS3 (7910809)
(7910328)
Hewlett-Packard, Model
334A (7911957)
Systron-Donner, Model
1037M (7910823) w/Timer Interval Unit, Model 1926A (7910824)
MIRCOM, Model SKA-4850-
101
Tektronix, Model RM561A
(7910655-2) w/3A6 plug-in (7911441-1) and 3B4 (7912040-1)
Gertsch, Model 1000
(7907060-1)
Ballantine Laboratories,
Model 1600A (MIS-10554-
2)
(8579475)
Description
4
B7 CABLE1 (RF CABLE
ASSEMBLY)
TB 9-4931-700-50
30-in., RG -58( )/U with double banana plug
terminations (7907470)
5
TB 9-4931-700-50
Table 3. Accessories Required - Continued.
Item
B8 CABLE (TEST LEAD) 36-in., RG -58( )/U with BNC plug and double banana
B9 HOOK-UP WIRE (ELECTRICAL
POWER CABLE)
B10 ISOLATION BOX (ADAPTER
BOX)
B11 PROBE Tektronix, Type 010-0128-00 (010-0128-00)
1
Three required.
Common Name
(Official Nomenclature)
Description (Part Number)
plug terminations (7907471)
Shielded pair No. 18 AWG, solid copper (MIS-10312)
Double banana jack terminals (SKD4850-3)

SECTION III

PRELIMINARY OPERATIO NS
6. Preliminary Instructions a. The instructions outlined in this section are preparatory to the calibration
process. Personnel should become familiar with the entire procedure before beginning the calibration.
b. Items of equipment used in this procedure are referenced within the text by common name and item identification number as listed in tables 2 and 3. For the identification of equipment referenced by item numbers prefixed with A, see table 2, and for prefix B, see table 3.
WARNING
HIGH VOLTAGE is used during the performance of this calibration. DEATH ON CONTACT may result if personnel fail to observe safety precautions.
NOTE
Unless otherwise specified, verify the results of each test and, whenever the test requirement is not met, take corrective action before continuing with the calibration. Adjustments required to calibrate the TI are included in this procedure. Additional maintenance information is contained in the manufacturer's manual, TM 9-4931-700-14/1, and TM 9­4931-700-14-2 for this TI.
NOTE
Unless otherwise specified, all controls and control settings refer to the TI.
6
TB 9-4931-700-50
7. Equipment Setup
a. Connect TI to autotransformer (A2). b. Connect autotransformer to 115-volt ac (alternating current) and adjust its output
control to obtain a 115-volt output. c. Set LINE switches to ON and allow 1 hour for equipment to warm-up and
stabilize.
SECTION III
CALIBRATION PROCESS
NOTE
When indications specified in paragraphs 8, 9, 10, or 12 are not within tolerance, perform the power supply check in paragraph 13 prior to making adjustments. After adjustments are made, repeat paragraphs 8, 9, 10, and 12. If indications specified in paragraph 11 are not within tolerance, the power supply check in paragraph 14 prior to making adjustments. Do not perform power supply checks if all other parameters are within tolerance.
8. Frequency Accuracy a. Performance Check
(1) Connect TI front panel output to input of frequency counter (A5), using cable (B8).
(2) Position TI controls as listed in (a) through (d) below: (a) Press FREQUENCY RANGE X10 pushbutton.
(b) Adjust FREQUENCY control to obtain FREQ scale indication of 1. (c) Press VOLTAGE RANGE 1V pushbutton. (d) Set voltage controls to obtain indication of 1.000000 V. (3) If frequency counter does not indicate between 98.0392 and 102.041 ins (milliseconds), perform b below.
(4) Repeat technique of (2) and (3) above , using settings and indications listed in table 4
7
TB 9-4931-700-50
Test Instrument Frequency Counter Indications
Frequency FREQUENCY RANGE
Settings Switch Setting Minimum Maximum
2 49.0196 ms 51.0204 ms 3 32.6797 ms 34.0136 ms 5 X10 19.6078 ms 20.4082 ms 7 14.0056 ms 14.5773 ms
10 9.8039 ms 10.2041 ms
1 9.80392 ms 10.2041 ms 2 4.90196 ms 5.10204 ms 3 X100 3.26797 ms 3.40136 ms 5 1.96078 ms 2.04082 ms 7 1.40056 ms 1.45773 ms
10 980 Hz 1020 Hz
1 980 Hz 1020 Hz 2 1960 Hz 2040 Hz 3 X1K 2940 Hz 3060 Hz 5 4900 Hz 5100 Hz 7 6860 Hz 7140 Hz
10 9.8 kHz 10.2 kHz
1 9.8 kHz 10.2 kHz 2 19.6 kHz 20.4 kHz 3 X10K 29.4 kHz 30.6 kHz 5 49.0 kHz 51.0 kHz 7 68.6 kHz 71.4 kHz
10 98.0 kHz 102.0 kHz
b. Adjustments
Table 4. Frequency Accuracy
(1) Press FREQUENCY RANGE X10K pushbutton and adjust FREQUENCY control for FREQ scale indication of 1.0.
(2) Adjust A12C2 (fig. 1) to obtain indication of 10.000 kHz (kilohertz) on frequency counter (R).
(3) Adjust FREQUENCY control slowly clockwise while observing frequency counter indication. If frequency increases, proceed to b(6) below. If frequency decreases, proceed to b(4) below.
(4) Repeat (1) above. (5) While observing frequency counter indication, adjust A12C2 (fig. 1) to obtain a
second 10 kHz output frequency (R). If a second 10 kHz cannot be obtained by adjusting A12C2, readjust A12C2 to obtain original 10 kHz frequency. While observing frequency counter indication, adjust A12L1 (fig. 1) in direction that causes output to decrease and pass through 0 (zero) Hz (hertz) and increase to 10 kHz. Adjust A12L1 to obtain indication of 10.000 kHz on frequency counter (R).
8
Loading...
+ 26 hidden pages