CANNOT FEED BACK THROUGH CONNECTING LEADSTOGROUND.
Figure
2-2.
0
C:::-
FROM
Desirable
ments
chassis
Calibration
for
:::::::::~
OTHER INSTRUMENTSINRACK
and
undesirable
grounding
when
using
Generator
§
~~
TO
THE OTHER.
0 0
arrange-
instrument
the
Voltmeter
Page 12
Sect.IIPage
The
accuracy
The
1)
4
precision
accuracyofthe
2)
The
accuracy
The
3)
4)
The
per
After
put
ceeding
If
quate
reached
turning
amplitude
The
value
738A
reaches
week
after
this
warmup,
voltageisabout 0.1%
this
amount
the
738A
has
accuracy
after
only 2
on
and
accuracyofthe
2-5
OPERATING
To
use
the
738Atocalibrate
voltmeter,
Turn
1)
to
30
proceedasfollows:
on
the
minutes,
of
the
738A
with whichitis
calibrating
of
the
meter).
output
stabilityvstime.
of
the
external
its
maximum
about
the
30
minutes
normal
per
indicates
been
calibrated
for
most
average
minutes
off
does
not
output
voltage.
INSTRUCTIONS
738A
and
allow to
depending on
is
determined
calibrated
attenuators
load
stability
(0.1%).
resistance.
of 0.1%
warmup
driftofthedcout-
week. Any
subnormal
drift
operation.
within a week,
uses
(1%)
warmup.
noticeably
an
oscilloscope
Repeated
affect
warmupfor
accuracy
required.
by:
(the
time.
ex-
ade-
the
or
2)
Set
MULTIPLIER
3)
C.onnect the
by
two
shield
post.
2-3
and
4)
Set
the
desired
2-6
There
ments:
volt
panel;
on
the
the
rms
rear
is
must
second,
important
rately
Make
cedure,asshown in
Figure
3
Figure
Figure
the
FUNCTION
to
O.
voltmeter
plain
wires
connected
Observe
precautions
2-5.
the
MULTIPLIER
voltage.
CALIBRATING
are
three
output-voltage
the
300-volt
ac
output
and
the
300-volt
left
side
of
of
the
attenuator
be
set
correctly
and the
before
all
three
peak-to-peak
that
the
any
calibration
other
2-3.
2-4.
2-5.
switchtoCALIBRATE, and
inputtothe
orashielded
to
the
bottom
738A output
cable
OUTPUT binding
discussedinparagraphs
and RANGE
THE
738A
selectors
calibrating
dc output
adjustment,
peak-to-peak
the
adjustment,
electronics
section.
first,
the
ac
both on the
acadjustment
chassis
Thedcvoltage
rms
VOltage
300-voltdcoutput be
adjustments
adjustments
the
following
Calibration
Calibration
Calibration
are
in one
Figures:
I
II
1II
Model738A
with the
for
adjust-
the
300-
front
just
to
ac
voltage
last.Itis
set
accu-
attempted.
pro-
Page 13
Model 738A
Sect.IIPage
5
CALIBRATION
o
o
I
o
RO
1.
Connect a 3OO-volt
standard
or
OUTPUT
2. Set the
quality, having an
better
and
tenninals
controls
meter
appropriate
of the 738A
on the 738A
of
calibration
accuracy
of 0.1%
resolution,tothe
as
shown.
as
shown.
Figure
3.
Turn
30
minutes.
4.Ifnecessary,
cessible
to
obtain exactly 300 volts. Disconnect
standard
2-3
on 738A and allow to
adjust
potentiometer
through the panel hole
meter.
wann
marked
up
R63,
for
ac-
OC,
the
Page 14
Sect.IIPage
6
Model738A
CALIBRATION
o
o
II
o
o
RO
This
shown in
TION
1.
procedure
I,
With
the
preceding
has
been
OUTPUT
eitheranelectrostatic
300-volt
accuracy
ac-dc
to
738Aasshown.
assumes
that
illustration,
perfonned.
SELECTOR
on OFF,
or
thermocouple
meterofcalibration
the
OUTPUT
terminals
the
procedure
CALIBRA-
connect
type
standard
of
the
Figure
2. Switch
3.
4.
2-4
DC-
and
the
readings.
Set
the
If
necessary,
accessible
ed
AC, to
step
in
2.
the
OUTPUT
DC+
and
OUTPUT
adjust
through
obtain
SELECTOR
note
the
midpoint
SELECTOR
potentiometer
the
front
the
midpoint
to400'"RMS.
panel
hole
reading
between
between
R40,
mark-
noted
Page 15
Model 738A
v
Sect,IIPage
7
0
W=
0
~
7
-
.~~TOI
..
@J@J
CALIBRATION
III
4
o
o
r!iJ
400H
o
00
·J
This
procedure
shown in
BRATION I
performed.
1.
Switch
2. Connect
the
3.
Set
On
4.
the
the
and
tfiI;
Model 400Htothe
the
738A OUTPUT SELECTOR
controlsasfollows:
RANGE
MULTIPLIER
OUTPUT SELECTOR
FUNCTION
the400H
meter
switchto1
reads
assumes
preceding
CALIBRATION II,
738Atothe
switchto100
switchtoCALIBRATE
adjust.00lV-40OCPS
exactly
that
illustrations,
100-volt
400Hasshown, with
to
400'"RMS
100.
the
procedure
has
settoOFF.
control
This
control
CALI-
been
range.
until
Figure
mounted
on
from
5. While 100
the
pointer
6.
Set
PTP
7.
The
necessary
meter
the
8.
Repeat
zero-set
is
calibrate
2-5
on a
bracket
the
left
side
the
from.
volts
remain
mechanical
indicates
the
738A OUTPUT SELECTORto400
and
the
400H
Calibrator
attenuator
step
until
the
meter
RANGE
should
adjust
section).
5, but
the
400H.
near
the
of
the
chassis,
across
zero
exactly94volts.
switchto3.
now
read
R67 (on
chassis
turn
the
meter
the400H,
until
100
left
justtothe
mechanical
indicates
from
panel
as
viewed
turn
the
meter
'"
exactly.
sideofVolt-
100 to
rear
meter
If
of
re-
Page 16
4DO~
RC
VIaV2
S2A
0000
RECTIFIER
FIXED
IBOT.
VI4
OSC
SECT.I[;."\
I
OUTPUT
SELECTOR
I--
\
r.7 "
\
1\
SERIES
REGULATOR
VI3
DIFFERENCE
AMPLIFIER
VIIaVI2
REFERENCE
TUBE
VIO
LEVEL
CONTROL
AMPLIFIER
V3
\
AGC
\
-4TO-6.5V
\
\
\
\
\
\
400-
I--
VGtTAGE
CRI
BAND
PASS
FILTER
...fi..D..
/I
\
I
I
I
I
I
I
I
I
l~
-~.~~
R67
PEAKTOPEAK
AC
~-ts9
__-__
S2A
(TOP
I--
20DB
PRE-AMPLIFIERINVERTER
V9
AGC
AMPLIFIER
ADJUST
SECT.)
I--
V4
J:l...[}
I--
~
66
~~https://manualmachine.com/I~
S2B
(BOT.
SECT}/
II
/
/
/
I
I
/
J/
PHASE
v5
VB
AGC
DIODE
C~6
..
I-
I
--
'=
FEEDBACK
~
fg)
V6
POWER
AMPLIFIER
V7
R40
~
I-
IJC
"
ltjf,
______
"7
~
en
(1)
n
....
......
......
......
I-
OG
'1:l
Il>
(1)
o
c~
c~
ri
~
I
I
II
I
1I
I.
I
J
I
I
MULTIPLIER I
I
I
I
I
I
I
§J
GJ
I
I
I
I
IE]
: !
ITRACKING I
NOTE:
ATTENVATOfi
FOR
ATTENtJATOF;
SCHEMATIC.
IFUNCTION I
'-'
lETAIU
!OUTPUT!
~
tl
10"'
'E
'""
'0'
t
Figure
3-1.
Block
Diagram
•
Model738A
~
8-
(1)
......
'-I
CJ.)
00
Page 17
Model 738 A
Sect.
III
Page
I
..
3-1
INTRODUCTION
The
Model 738A
the
major
gram
generatorisexplained
circuits
The
300 volt working
the
circuits
the
300
of
all
almost
tube
supply.
adjusted
sible
circuit
in
Figure
are
regulated
voltsdcprimarily
calibrating
entirely
which
The
potentiometer
through the
Voltmeter
sections
3-1.
explainedingreater
dc
power
standard
in
theacgenerator.
voltages.
by
the
suppliesareference
300vdcisadjustable
front
Calibrator
shown in
The
overall
first,
then
supply
and dc
determines
Stabilityisdetermined
gaseous
R63,
panel.
the
operation
certain
detail.
provides
powertomost
The
voltage
voltage
by a
screwdriver-
marked
consists
block
important
both
accuracy
the
accuracy
reference
for
DC,
acces-
dia-
of the
the
the
of
of
dc
of
THEORY
3-2
The
former
electronic
gulator
reference
which
two-stage
plified
tube.
If
this
and
against
amplifies
appliesitto
If
through
act
DC
GENERATOR
dc
power
and
tube which
the
difference
The
the
300 volt output
changeiscoupledtodifference
V12.
the
the
300
V13 will
the
tendencyofthe
supply
rectifier
regulator
tube
providingastable
300
voltsiscontinuously
difference
voltage
This
outputof
and
inverts
the
volts
level
consists
which
consisting
acts
amplifier
voltagetothe
regulating
tendstochange, a
amplifier
the
Voltage
the
gridofSeries
tendstodecrease,
increase
output voltagetodecrease.
SECTION
OF
asavariable
and will
OPERATION
SECTION
of a
power
supply
compares
difference
voltages
of a
series
resistor,
voltage
compared,
applying
series
processisas
amplifier
the
Reference
voltage
Regulator
the
instantly
trans-
for
against
and
an
regulator
follows:
portion
change
tube VIO,
Tube
V13.
current
counter-
III
an
re-
a
a
am-
of
VII
and
Fordcmeter
supplyisconnected
TOR
switch
FUNCTION
The
internal
of
the
output
point of
ages
are
the
floating dc
grounded
the
positive
required.
For
ac
LECTOR
generatortothe
of
theacvoltage
loadisalso
and
peak-to-peak
ing
within
calibration,
S2, to
switch
impedanceofthedcgenerator
attenuator
overload.
obtainedbyreversing
only
after
or
negative
voltmeter
switch
theacgenerator
connects
attenuator.
generator
nearly
the
output of
through the
the
attenuator
S102,tothe
is
Negative
power
voltagesisdeterminedbyswitch-
supply.
the
switching
lead
calibration,
0 ohm.
OUTPUT
OUTPUT
nearly
and
the
can
the
the
output
The
internal
uptothe
The
generationofrms
andisdescribed
the
300
volt
SELEC-
and
through
terminals.
0 ohm uptothe
positivedcvolt-
output
The
supply
so
that
be
grounded
OUTPUT
from
impedance
point of
ahead
leads
either
the
over-
later.
the
as
SE-
ac
A
difference
the
insensitive
supply
sioninparticular).
amplifierisas
of
Tube
is
thedcoutput with
Reference
plished
gridofVIlA
of Voltage
appliedtothe
currentinVllA
common
will
will
out of
to
the
type of
degenerative
to
voltages
Vll,
the
Tube
by applying a
whichisnormally
Reference
gridofVllA
cathode
decrease.
have
changedbyequal
phase.
grids
The
of VI2A& B whichisanother
amplifier
feedback
changes
(heater
The
follows:
input
stage,
the
constant
VIO.
portionofthe
Tube
will
increase,
resistor,
The
plate
platesofVII
is
usedtoamplify
becauseitis
in tube
voltages
The
characteristics
and
cathode
operationofthe
comparesasample
voltage
comparisonisaccom-
from
+300 vdc
equaltothe
VlO.
Assume
tends
togo
positive,
and
duetothe
the
current
voltagesofVIlA
amounts
and will be 180
are
direct-coupled
relatively
and
emis-
difference
Voltage
to
the
output
the
voltage
the
large
in
VllB
and
difference
of
B
0
Page 18
Sect.
III
Page
2
amplifier
just
polarity
the
plateofVI2B.
3-3
Theacvoltage
rc
oscillator
a
filter,apre-amplifier,
an
automatic
lator
signal
resonant
follows.
to
Level
kept
variable-mu
Control
introducedbythe
The
feedback
The
bothtothe
to
the
control
cuit
followed by
Reference
voltage.
this
andisthen
for
Level
amplifier
output
whose
described.
for
AC
The
Series
GENERATOR
generator
followed by a
gain
generates
whose
frequency
frequency
The
signal
Control
constant
by
control
Tube
is
variable-mu
signal
is
then
stabilized
signal
from
the
function
output
consists
circuit.
attenuator,
The
of
amplifier
Clipping
Any
portionoftheacwave
voltageispassed
rectified
Control
and
output
level
would
not held down by
of
V3.
Consequently,
put
level
to
decrease
be
corrected
always
the
Reference
level
tendstorise,
ac
which, when
trying
automatically,
to
maintain
Clipping
rectified,
operation
control
Regulator
is
similar
voltageofthe
Vl3isobtained
SECTION
consistsofa
level
an
output
control
a highly
Tube
application
filtered
of
from
grid.
(AGC)
is
the
the
V3.
circuit.
stable,
adjusted
400-cycle
oscillatorisapplied
The
output
of
AGC
The
outputofthe
to
remove
grid
amplified
output
output
selector
and
fed
amplifier,
amplifierisconnected
for
further
andtothe
automatic
Reference
V9
and
gain
Clipping
rectifier
Diodeisbiasedbyan
by the
diodetoamplifier
by CR1toproduce
Tube
V3.
The
gainofthe
amplifierissuch
always
the
AGC
the
any
tendency
(due to
Diode
diode
tend
bias
appliedtothe
loading,
as
the
level
bias.
passes
to
amplifier
develops
to
the
desired
from
conventional
control
tube,
amplifier
The
oscil-
lOW-distortion
to
the
exact
filter
which
from
V3
biastoits
level
harmonics
characteristic.
through
V5,
V6and
application
automatic
control
gain
cir-
Diode, V8,
CR1.
The
exact
thatisabove
AGC
bias
pre-
that
the
rise
were
grid
for
theacout-
etc)
would
V9
established
If
the
output
moreofthe
more
bias.
one
and
is
the
V7.
dc
V9
ac
is
by
it
The
increaseinbias
fier
and
holds
diode
the
is
To
turned
to
the
made
tiometer
the
also
the
with a
ment
result
The
or
switch
automatically
two
constant
produce
+300
cathode
intoavoltage
same
reduces
desired
small
avoids
with
3-4
OUTPUT
ac
diodes
through
volts.
R67toground
time
OUTPUT
output to
selects
MULTIPLIER
or
from
the1or3volt
with
rms
return
the
reductioninoutput
increase
the
the
SELECTOR
(S103)
reduces
the
output
divides
so
that
varying
voltages,
series
To
produce
resistors
OUTPUT
levelofthe
increaseindistortion
greater
ATTENUATOR
the
attenuator
the
output
and
(SlO1) SWitch.
The
attenuator
each
dividerisreturnedtoground,
divider
vious
are
The
from
selects
voltage
is
divider,
1/10
MUL
TIP
oneofthese
the
divider
consists
connectedtothe
each
divider
the
voltage
LIER
switch
dividers.
output
from
selected
switch.
The
tracking
propriate
0.1
volt
vides
0.5
attenuators
parallel
sistors
voltages
attenuator
divisions
volt
divisions
are
combination
that
are
taps.
from0to
simple,
very
stable.
the
gainofthe
constant.
the
current
the
bias
on the
signal
resistors
levels.
the
cathodeofV8isre-
R67, R68
peak-to-peak
R67, R68
divider
by
through
returning
R66
SELECTOR
400'\,
inputto
voltageisobtained
in
AGC
bias.
grid
bias.
connects
The
voltage
either
RANGE (S104)
positionsofthe
of five
voltage
each
lowest
produces
from
the
divider
selects
the
The
oneofthe
by
are
obtained
One
five
the
position
1 volt,
from0to3volts.
uncompensated
of
precision
Model 738A
ampli-
The
second
between
signal
and
voltages
and
R69
poten-
and
S2A. At
switch
V3.
This
arrange-
which would
either
the
FUNCTION
from
switches,
TRACKING
dividers,
subsequent
taponthe
voltages
before
output
voltage
RANGE
switch
tapsonthe
MUL
TIPLIER
from
the
provides
the
other
series
wirewound
diode
R69
are
S2
Thus
dc
the
pre-
that
it.
ap-
pro-
The
and
re-
Page 19
Model 738A
Sect.
IV
Page
1
4-1
This
ment
shooting
To
cludedatthe
procedure
localizing
the
on
assemblies
Hewlett-Packard
4-2
The
TubesV8and
teristics
voltages.
erator;
put
the
thercoscillator
quency.
frequency
justedtothe
inputtopreamplifier
replaced
output
servicing.
INTRODUCTION
section
and
facilitate
useoftest
the
TUBE REPLACEMENT
738A
voltages
dc
contains
for
adjusting,
the Model 738A
service,
endofthe
gives
a fault to a
schematic
does
directly
Tube
V8
affects
voltages.
After
shouldbechecked
without
voltages
a quick
points.
shouldberemoved
Company
not
VIa
are
affect
VlO
only
and
canbereplaced
may
tube
center
adjustment.
should be
instructions
calibrating,
Voltmeter
location
section.Atrouble
small
These
diagram.
require
the
affects
the
Replacement
causeaminor
replacement,
frequencyofthe
V4.
illustrations
systematic
circuit
test
Damaged
and
for
repair.
specially
onlytubes whose
the
accuracyofthe
all
voltagesinthe
accuracyoftheacout-
and,ifnecessary,
The
The
checked
for
tube
replace-
and
trouble
Calibrator.
are
shooting
method
area
through
points
of a tube within
other
are
shown
attenuator
returned
selected
without affecting
changeinfrethe
accuracyofthe
tubes.
charac-
output
oscillator
filteratthe
tubes
can
after
in-
to the
gen-
ad-
be
any
by
Research
at
hardware
ing and
4-4
of
Itisimportant
output be
filter
function of
which follows
distortion
and
of thercoscillator
resonant
frequencyisadjusted
the
way to
resonant
maximum
by a
Model4l0B.
In
the
and
input and output
requires
l50A
Proceedasfollows:
air-conditioning
ADJUSTING
AC
in
maximum
bottom
set
dc
the
more
oscillatoristunedsothat
therefore
with a Model 152 Dual Channel
MAINTENANCE
Products
stores
OUTPUT
that
the
sameasthe
order
frequency
frequencyofthe
vacuum-tube
an
to obtain
the
high-Q 400
Level
introduced
signal-to-noise
compartment.Asimple
the
oscillator
dc
voltageattest
exact
there
signals.
oscilloscope,
Corp.
and
other
stores
supplies.
THE FREQUENCY OF THE
the
frequencyofthe
resonant
proper
cycle
Control
by V3.
output
of
tuning
is
TubeV3istoremove
mustbethe
the
filter.
by
capacitor
output
filteristo
pointD,as
voltmeter,
procedure
no
phase
The
such
SECTION
This
oilisavailable
carrying
oscillator
frequency
operation.
filter
(LI,
For
maximum
ratio,
the
frequency
sameasthe
The
oscillator
C3,
located
and
satisfactory
frequencytothe
adjustC3for
measured
such
as
given below,
the
filterisresistive
shift
between the
follOWing
as
procedure
the
t[ij;
Amplifier.
IV
heat-
of the
The
CIa)
output
in
the
I/i$
Model
4-3
This
nation
for
asitappears
instrument
pulling
element,
recoat
A
CLEANING
instrument
for
the
fan tobeeffective.
it
washinwarm
with
suitable
THE
containsafan-air
cooling.
dirty.Itis
cabinet
out
special
coating
and
of
its
oilisR-P
oil
AIR
FILTER
The
filter
mustbekept
Clean
locatedatthe
mayberemoved
holder.
water
available
To
and
Filter
filter
the
filterassoon
rearofthe
simply
clean
detergent;
for
this
Coat
combi-
clean
the
filter
purpose
#3,
made
by
then
1) Remove the
turn
on, and allow a
2) Connect one
other
vertical
the
oscilloscope
loscope
avoid
oscillatororfilter.
connections
shifting
metal
vertical
input to
externally
the
dust
five-minute
input to
test
are
madeatthese
resonant
cover
from
frequencyofeither
point
test
from
the 738A,
warmup.
test
point B,
E.
Synchronize
pointF.Oscil-
points to
the
the
Page 20
Sect.
IV
Page
3)
Adjust
impose
Adjust
amplitudeofthe
4)
If
traces,
the
two
is
now
filter,
achieved.
the
the
horizontal
the
vertical
necessary,
Le.toreduce
signalstozero.
at
the
and
best
2
vertical
two
peakofthe
stability
position
centerlinesofthe
sensitivity
traces.
adjust
the
C3 to
phase
The
oscillator
resonance
and
greatest
control
two
to
equalize
superimpose
difference
curveofthe
output
to
super-
traces.
the
the
between
frequency
are
is
designedtowork
ohms
To
or
apply
and will not
pedance
meter
voltages
the
meters
prevent
a
meterisdifferent
damage
which
differential-type
except
the
correction
into
read
(see
paragraph
to
hasatleast
the
300
Model 738A
an
impedanceof3
accurately
the
attenuator,
1000
voltmeter
volts.Ifthe
from3megohmsbesure
as
given in
with
2-2).
ohms/volt
to
impedance
paragraph
other
useavolt-
impedance
measure
meg-
im-
all
of
to
2-2.
4-5
To
brator,
factorily
OUTPUT
the
schematic
correct,
ly,
A.
1)
toapower
2)
switchestomaximum
3) With
lutionofat
voltageatthe
voltage
remain
the
V12
of
300 volt
mately
is
TESTING
THE
test
or
first
by
terminals.
key
voltages
diagram.
assume
and
proceedtothe
OC
CHECKS
Connect
Set
range,
the
too
the
the OUTPUT SELECTORtoOC+,
an
from
constant.
check
and
Vll,
regulator.
bus.
0.8
high,
AND
CALIBRATION
trouble
determine
measuring
at
the
738A
source,
electronic
least
± 0.1%
OUTPUT
103to127
V14
then
The
millivolts
check
Vll,
TROUBLE
GENERATOR
shoot
the
738A
if
it
is
its
output
To
locateafailure,
the
test
points
When a
circuittobe
next
throughavariable
tumonand
clockwise
voltmeter,
Ifitvaries
for
check
Check
ripple
peak-to-peak.
V12
measured
step.
at
300 vdc,
terminals.
volts.
The
over
low output;
for
ripple
level
shouldbeapproxi-
and
C29.
allowtowarm
positions.
excessive
SHOOTING
Voltmeter
operating
voltages
shown on
operating
transformer
havingareso-
measure
Vary
VOltage should
all
voltage
If
measure
VOltage
all
or
check
the
Calisatisat
proper-
other
the
part
loading
on
ripple
the
the
up.
the
line
V13,
the
Switch
through
brated
dc
incorrect,
6)
positionstodetermineifallofthe
put
is
correct,
rect,
The
satisfactory
circuit
B.
1)
2) With a
voltageattest
adjust
of
low and
then
too high
V7,
termines
the
output
is
correct,
Switch
voltages
the
above
failure.
AC
Set
the
to
other
V8
and
MULTIPLIER
their
positionstodetermineifall
voltages
but
anyoneofthe
the
attenuator
the
FUNCTION
are
correct.
but
anyoneofthe
attenuator
steps
and,ifnot, which
CHECKS
OUTPUTto400", RMS.
precision
obtain
cannot
tubesinthe
and
the
the
bias
sectionisdefective.
should
point
F.Ifitistoo highortoo low,
the
be
adjusted
cannot
voltage-divider
for
and RANGE
are
correct.Ifthe
section
switchtothe
If
the
lower
voltagesisincor-
indicateifoperation
section
voltmeter
correct
amplifier.
be
adjusted
V8.
measure
voltage.
up,
check
circuit
lower
is
calibrated
300
If
the
down,
selectors
the
cali-
300
volts
voltages
defective.
TRACKING
out-
voltsdcis
contains
the
signal
If
itistoo
V6
and
V7,
voltage
check
which
V6,
de-
is
is
a
..
is
4) Switch
RMS
position
as
in
the
voltmeters.
5)Ifthe
with a low
300
may
the
lose
dc
volts
be
damaged.
precision
their
the
OUTPUT
and
preceding
output of
impedance
(straight
resistors
accuracy.
check
paragraph.
through)
If too
SELECTORtothe
the
line
voltage
Disconnect
the
attenuator
voltmeter
much
may
In addition
on
range,
the
currentisdrawn
be
overheated
this
400
regulation
is
measured
any
but
attenuator
instrument
the
the
and
3) With a high
'"
such
cycle
The
25
reduced,
low
It
check
(0.1%).
distortion.
impedanceacelectronic
as
the
IJ$
Model
voltageatpOint C
voltageatpoint C
volts.
and
may
If
itisexcessively
replace
cannot
alsobedesirabletocheck
point C which
If
lamp
be
raised,
shouldbeat
necessary,
400D/H/L
in
the
should
RTI.
check
select
voltmeter,
measure
oscillator
be
between20and
high and
If
itisexcessively
RTl,
the
least60db
tubes
cannot
VI
and
distortion
to
the
400
circuit.
V2.
down
reduce
be
at
Page 21
Model738A
Sect.
IV
Page
3
5) With a
dc
electronic
Model 410B Vacuum
bias
voltageattest
-4
and
-6.5
volts.
the
amplifier
is
low.
6)
Measure
test
point
V6
andV7andifthe
distortioninthe
about
since
0.15%.
it
F.
will
tubes
the
distortionatthe
If
good
output should be
The
affect
driftintheaclevel.
When
measuring
the
330B with
contains
need
you
in
the
out
appreciable
balancing.
are
not
unit.
the
leadsofthe
pieceofwire
60", hum on
there
tubes
Now
toroid
of
wire.
wire
the
hum
rotate
is
etc.
place
to
the
toroid
Make
possible.
toroid
pickup.
for
hum in
Then
junction of C9
coil.
observing
the
minimum
the
distortion,
an
However,
just
introducing
To
determineifthe
possible.
the
remove
short
is
this
This
Now
To
hum.
voltmeter,
Tube
Voltmeter,
point
D:itshould
If too low,
V3, V4, V5,
tubes
output
tubes
distortion
the
AGC
oscilloscope.
60", hum
when doing
toroid
coilLlwith
Noteifthe
screen
the
of
unit
removeitby
ths
from
short
the
and
the
lead
short
with
will
remove
rotate
rotate
hum,
The
the
the
and
hum
suchasthe
the
gain
V6
output
are
used
are
balanced,
reducedtobelow
must
voltage
monitor
If
the
the
toroid
humtobuck
unit
has
output
the
oscilloscope.
across
ground
R18.
The
that
goestothe
the
shortest
all
hum
toroid
for
coil
loosen
tighten while
should
measure
be between
on one of
and/or
terminals,
for
V4, V5,
be
kept low
and
cause
the
output of
distortion
coil
may
thisbesure
out
hum
hum,
short
the
shortest
contains
replacing
the
toroid.
sideofthe
ground
side
black
piece
ahead
minimum
the
nut,
still
be down to
t$
the
V7
the
of
of
a
If
about 0.03%.
up,
replace
removed
cover
tween
and
the
Remove
V3. If
be
suretoretain
the
coil.
cover
and
the
operation.
4-6
Accuracy
accuracy
the
methods
tiometers,
used.
only one
voltmeter.
will
maximum
a
There
ments:
volts
panel;
on
be
peak-to-peak
the
other