Agilent Model 6010A; Serials US37110171 and above
Agilent Model 6011A; Serials US35460156 and above
Agilent Model 6012B; Serials US35430336 and above
Agilent Model 6015A; Serials US37050146 and above
For instruments with higher serial numbers, a change page may be included.
Microfiche Part No. 5964-8276 Printed in USA: July 2001
CERTIFICATION
Agilent Technologies certifies that this product met its published specifications at time of shipment from the factory. Agilent
Technologies further certifies that its calibration measurements are traceable to the United States National Institute of
Standards and Technology, to the extent allowed by the Institute’s calibration facility, and to the calibration facilities of
other International Standards Organization members.
WARRANTY
This Agilent Technologies hardware product is warranted against defects in material and workmanship for a period of three
years from date of delivery. Agilent Technologies software and firmware products, which are designated by Agilent
Technologies for use with a hardware product and when properly installed on that hardware product, are warranted not to
fail to execute their programming instructions due to defects in material and workmanship for a period of 90 days from date
of delivery. During the warranty period Agilent Technologies will, at its option, either repair or replace products which
prove to be defective. Agilent Technologies does not warrant that the operation of the software, firmware, or hardware shall
be uninterrupted or error free.
For warranty service, with the exception of warranty options, this product must be returned to a service facility designated
by Agilent. Technologies. Customer shall prepay shipping charges by (and shall pay all duty and taxes) for products
returned to Agilent Technologies. for warranty service. Except for products returned to Customer from another country,
Agilent Technologies s hall p a y for return of products to Customer.
Warranty services outside the country of initial purchase are included in Agilent Technologies’ product price, only if
Customer pays Agilent Technologies international prices (defined as destination local currency price, or U.S. or Geneva
Export price).
If Agilent Technologies is unable, within a reasonable time to repair or replace any product to condition as warranted, the
Customer shall be entitled to a refund of the purchase price upon return of the product to Agilent Technologies.
LIMITATION OF WARRANTY
The foregoing warranty shall not apply to defects resulting from improper or inadequate maintenance by the Customer,
Customer-supplied software or interfacing, unauthorized modification or misuse, operation outside of the environmental
specifications for the product, or improper site preparation and maintenance. NO OTHER WARRANTY IS EXPRESSED
OR IMPLIED. AGILENT TECHNOLOGIES SPECIFICALLY DISCLAIMS THE IMPLIED WARRANTIES OF
MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE.
EXCLUSIVE REMEDIES
THE REMEDIES PROVIDED HEREIN ARE THE CUSTOMER’S SOLE AND EXCLUSIVE REMEDIES. AGILENT
TECHNOLOGIES SHALL NOT BE LIABLE FOR ANY DIRECT, INDIRECT, SPECIAL, INCIDENTAL, OR
CONSEQUENTIAL DAMAGES, WHETHER BASED ON CONTRACT, TORT, OR ANY OTHER LEGAL THEORY.
ASSISTANCE
The above statements apply only to the standard product warranty. Warranty options, extended support contracts, product
maintenance agreements and customer assistance agreements are also available. Contact your nearest Agilent
Technologies Sales and Service office for further information on Agilent Technologies’ full line of Support Programs.
2
SAFETY SUMMARY
The following general safety precautions must be observed during all phases of operation, service and repair of this
instrument. Failure to comply with these precautions or with specific warnings elsewhere in this manual violates safety
standards of design, manufacture, and intended use of the instrument. Agilent Technologies, Inc. assumes no liability for the
customer's failure to comply with these requirements.
BEFORE APPLYING POWER.
Verify that the product is set to match the available line voltage and the correct fuse is installed.
GROUND THE INSTRUMENT.
This product is a Safety Class 1 instrument (provided with a protective earth terminal). To minimize shock hazard, the instrument chassis
and cabinet must be connected to an electrical ground. The instrument must be connected to the ac power supply mains through a threeconductor power cable, with the third wire firmly connected to an electrical ground (safety ground) at the power outlet. For instruments
designed to be hard wired to the ac power lines (supply mains), connect the protective earth terminal to a protective conductor before any
other connection is made. Any interruption of the protective (grounding) conductor or disconnection of the protective earth terminal will
cause a potential shock hazard that could result in personal injury. If the instrument is to be energized via an external autotransformer for
voltage reduction, be certain that the autotransformer common terminal is connected to the neutral (earth pole) of the ac power lines
(supply mains).
INPUT POWER MUST BE SWITCH CONNECTED.
For instruments without a built-in line switch, the input power lines must contain a switch or another adequate means for disconnecting
the instrument from the ac power lines (supply mains).
DO NOT OPERATE IN AN EXPLOSIVE ATMOSPHERE.
Do not operate the instrument in the presence of flammable gases or fumes.
KEEP AWAY FROM LIVE CIRCUITS.
Operating personnel must not remove instrument covers. Component replacement and internal adjustments must be made by qualified
service personnel. Do not replace components with power cable connected. Under certain conditions, dangerous voltages may exist even
with the power cable removed. To avoid injuries, always disconnect power, discharge circuits and remove external voltage sources before
touching components.
DO NOT SERVICE OR ADJUST ALONE.
Do not attempt internal service or adjustment unless another person, capable of rendering first aid and resuscitation, is present.
DO NOT EXCEED INPUT RATINGS.
This instrument may be equipped with a line filter to reduce electromagnetic interference and must be connected to a properly grounded
receptacle to minimize electric shock hazard. Operation at the line voltage or frequencies in excess of those stated on the data plate may
cause leakage currents in excess of 5.0mA peak.
SAFETY SYMBOLS.
DO NOT SUBSTITUTE PARTS OR MODIFY INSTRUMENT.
Because of the danger of introducing additional hazards, do not install substitute parts or perform any unauthorized modification to the
instrument. Return the instrument to a Agilent Technologies, Inc. Sales and Service Office for service and repair to ensure that safety
features are maintained.
Instruction manual symbol: the product will be marked with this symbol when it is necessary for the user to refer to the
instruction manual (refer to Table of Contents) .
Indicates hazardous voltages.
Indicate earth (ground) terminal.
The WARNING sign denotes a hazard. It calls attention to a procedure, practice, or the like, which, if not correctly
performed or adhered to, could result in personal injury. Do not proceed beyond a WARNING sign until the indicated
conditions are fully understood and met.
The CAUTION sign denotes a hazard. It calls attention to an operating procedure, or the like, which, if not correctly
performed or adhered to, could result in damage to or destruction of part or all of the product. Do not proceed beyond
a CAUTION sign until the indicated conditions are fully understood and met.
Instruments which appear damaged or defective should be made inoperative and secured against unintended operation until they can be
repaired by qualified service personnel.
3
Safety Symbol Definitions
SymbolDescriptionSymbolDescription
Direct currentTerminal for Line conductor on permanently
installed equipment
Alternating currentCaution, risk of electric shock
Both direct and alternating currentCaution, hot surface
Three-phase alternating currentCaution (refer to accompanying documents)
Earth (ground) terminalIn position of a bi-stable push control
Protective earth (ground) terminal
(Intended for connection to external
protective conductor.)
Frame or chassis terminalOn (supply)
Terminal for Neutral conductor on
permanently installed equipment
Terminal is at earth potential
(Used for measurement and control
circuits designed to be operated with
one terminal at earth potential.)
Printing HistoryThe edition and current revision of this manual are indicated below. Reprints of this manual containing minor corrections
and updates may have the sa me p rinting date. Revised editions are identi fied by a new printing date. A revised edition
incorporates all new or corrected material since the previous printing date. Changes to the manual occurring between
revisions are covered by change sheets shipped with the manual. Also, if the serial number prefix of your power supply is
higher than those listed on the title page of this manual, then it may or may not include a change sheet. That is because
even though the higher serial number prefix indicates a design change, the change may not affect the content of the manual.
Out position of a bi-stable push control
Off (supply)
Standby (supply)
Units with this symbol are not completely
disconnected from ac mains when t his switch is
off. To completely disconnect the unit from ac
mains, either disconnect the power cord or have
a qualified electrician install an external switch.
may be photocopied, reproduced, or translated into another language without the prior consent of Agilent Technologies,
Inc. The information contained in t his d ocument is subject to change witho ut notice.
Calibration and Verification .............................................................................................................................................7
Principles of Operation ..................................................................................................................................................... 7
Calibration and Verification.................................................................................................................................................. 9
Test Equipment Required ..................................................................................................................................................... 9
Constant Voltage (CV) Tests.......................................................................................................................................... 18
Constant Current (CC) Tests........................................................................................................................................... 24
Repair and Replacement ..................................................................................................................................................... 29
A2 Control Board Removal............................................................................................................................................ 30
A4 FET Board Removal ................................................................................................................................................. 30
A3 Front Panel Board Removal...................................................................................................................................... 31
A1 Main Board Removal................................................................................................................................................ 31
Using the Tables ............................................................................................................................................................. 33
Main Troubleshooting Setup .......................................................................................................................................... 33
Front Panel Troubleshooting........................................................................................................................................... 36
Power Section Blocks ..................................................................................................................................................... 40
Troubleshooting Down Programmer .............................................................................................................................. 42
Troubleshooting CC Circuit............................................................................................................................................44
Principles of Operation ........................................................................................................................................................47
System Description............................................................................................................................................................. 47
Regulation & Control Subsystem ....................................................................................................................................... 48
Input Power Subsystem ......................................................................................................................................................53
5
DC Power Conversion Subsystem...................................................................................................................................... 54
The Front Panel Board........................................................................................................................................................ 54
Component Location and Circuit Diagrams ...................................................................................................................... 79
System Option 002 (6010A, 6011A, 6012B) ........................................................................................................................ 91
General Information............................................................................................................................................................ 91
Status Indicators............................................................................................................................................................ 102
AC Dropout Buffer Circuit........................................................................................................................................... 105
Multiple Supply System Shutdown ..............................................................................................................................105
Troubleshooting Resistance and Voltage Programming............................................................................................... 107
Troubleshooting Current Programming........................................................................................................................ 107
This manual contains information for troubleshooting the Agilent Models 6010A, 6011A, 6012B, or 6015A 1000W
Autoranging Power Supply to the component level. Wherever applicable, the service instructions given in this manual refer
to pertinent information provided in the Operation Manual. Both manuals cover Agilent Models 6010A/11A/12B/15A;
differences between models are described as required.
The following information is contained in this manual.
Calibration and Verification
Contains calibration procedures for Agilent Models 6010A/11A/12B/15A. Also contains verification procedures that check
the operation of the supplies to ensure they meet the specifications of Chapter 1 in the Operating Manual.
Troubleshooting
Contains troubleshooting procedures to isolate a malfunction to a defective component on the main circuit board or to a
defective assembly (front panel, power transformer, or cable assembly). Board and assembly level removal and replacement
procedures are also given in this section.
Principles of Operation
Provides block diagram level descriptions of the supply's circuits. The regulation and control, protection, input power, dc
power conversion and output circuits are described. These descriptions are intended as an aid in troubleshooting.
Replaceable Parts
Provides a listing of replaceable parts for all electronic components and mechanical assemblies for Agilent Models
6010A/11A/12B/15A.
Circuit Diagrams
Contains functional schematics and component location diagrams for all Agilent 6010A/11A/12B/15A circuits. The names
that appear on the functional schematics also appear on the block diagrams in Chapter 4. Thus, the descriptions in Chapter 4
can be correlated with both the block diagrams and the schematics.
Safety Considerations
This product is a Safety Class 1 instrument, which means that it is provided with a protective earth terminal. Refer to the
Safety Summary page at the beginning of this manual for a summary of general safety information. Safety information for
specific procedures is located at appropriate places in the manual.
7
Manual Revisions
Agilent Technologies instruments are identified by a 10-digit serial number. The format is described as follows: first two
letters indicate the country of manufacture. The next four digits are a code that identify either the date of manufacture or of
a significant design change. The last four digits are a sequential number assigned to each instrument.
ItemDescription
USThe first two letters indicates the country of manufacture, where US = USA; MY = Malaysia.
3648This is a code that identifies either the date of manufacture or the date of a significant design change.
0101The last four digits are a unique number assigned to each power supply.
If the serial number prefix on your unit differs from that shown on the title page of this manual, a yellow Manual Change
sheet may be supplied with the manual. It defines the differences between your unit and the unit described in this manual.
The yellow change sheet may also contain information for correcting errors in the manual.
Note that because not all changes to the product require changes to the manual, there may be no update information
required for your version of the supply.
Older serial number formats used with these instruments had a two-part serial number, i.e. 2701A-00101. This manual also
applies to instruments with these older serial number formats. Refer to Appendix B for backdating information.
8
2
Calibration and Verification
Introduction
This section provides test and calibration procedures. The operation-verification tests comprise a short procedure to verify
that the unit is performing properly, without testing all specified parameters. After troubleshooting and repair of a defective
power supply you can usually verify proper operation with the turn-on checkout procedure in the Operating Manual.
Repairs to the A1 main board and the A2 control board can involve circuits which, although functional, may prevent the
unit from performing within specified limits. So, after A1 or A2 board repair, decide if recalibration and operation
verification tests are needed according to the faults you discover. Use the calibration procedure both to check repairs and
for regular maintenance.
When verifying the performance of this instrument as described in this chapter, check only those specifications for which a
performance test procedure is included.
Test Equipment Required
Table 2-1 lists the equipment required to perform the tests of this section. You can separately identify the equipment for
performance tests, calibration and troubleshooting using the USE column of the table.
Operation Verification Tests
To assure that the unit is performing properly, without testing all specified parameters, first perform the turn-on checkout
procedure in the Operating Manual. Then perform the following performance tests, in this section.
CV Load Effect
CC Load Effect
Calibration Procedure
Calibrate the unit twice per year and when required during repair. The following calibration procedures which follow
should be performed in the sequence given. Table 2-2 describes in detail these calibration procedures and lists the expected
results to which each adjustment must be made.
Note: Some of the calibration procedures for this instrument can be performed independently, and some
procedures must be performed together and/or in a prescribed order. If a procedure contains no references
to other procedures, you may assume that it can be performed independently.
To return a serviced unit to specifications as quickly as possible with minimal calibration, the technician
need only perform calibration procedures that affect the repaired circuit. Table 2-3 lists various power
supply circuits with calibration procedures that should be performed after those circuits are serviced.
No saturation at:
6010A 20Adc
6011A 100Adc
6012B 51Adc
6015A 51Adc
Bandwidth: 20Hz to 20MHz
Open and short switches
6010A
Voltage range: 200Vdc
Current range: 20Adc
6011A
Voltage range: 30Vdc
Current range: 120Adc
6012B
Voltage range: 65Vdc
Current range: 55Adc
6015A
Voltage range: 200Vdc
Current range: 5Adc
PAgilent 3400A
P,A,TAgilent 3456A
PTektronix P6303
Probe/AM503 Amp/
TM500 Power Module
P,ATransistor Devices Model
DLP 130-50-2500
DLR-400-15-2500
DLP 50-150-3000
DLP 130-50-2500
DLR-400-15-2500
CC PARD Test
Resistive Load
10
Value:
6010A
3.5 ohms >1000W
Accuracy: 1%
6011A
0.058 ohms >1000W
Accuracy 1%
6012B
0.4 ohms >1000W
Accuracy: 1%
Rheostat or Resistor Bank
P,A
TYPEREQUIRED CHARACTERISTICSUSERECOMMENDED MODEL
Load Resistors
(6015A)
Table 2-1. Test Equipment Required (continued)
40Ω, ±1%, 1000W
P,A
250Ω, ±1%, 1000W
Current-Monitoring
Resistors
Calibration and Test
Resistors
Terminating
Resistors (4)
Blocking
Capacitors (2)
Value:
6010A
100mV @ 10A (10mΩ must
be capable of 20Amps)
Accuracy: 0.02% **
TC: 10ppm/°C
6011A
50mV @100A (0.5mΩ)
Accuracy: 0.05% **
TC: 30ppm/°C
6012B
50mV @ 50A (1.0mΩ)
Accuracy: 0.02% **
TC: 30ppm/°C
6015A
0.1Ω, 15A, ±0.04%**
Value: 50Ω, 5%, 40W
2KΩ, 0.01%, ¼W
Value: 50Ω ± 5%, noninductive
Value: 0.01µF, 600Vdc
P,A
A,T
P
P
Common-Mode
Toroidal Core
≥3.7µH/turn
≅23mm I.D
2
PFerrox-Cube
500T600-3C8,
Agilent 9170-0061
.
DC Power SupplyVoltage range: 0-60Vdc
T,PAgilent 6012B
Current range: 0-50Adc
Variable Voltage
Transformer
(autotransformer)
Range greater than -13% to +6% of
nominal input AC voltage
4KVA
P,A
P = performance testing A = calibration adjustmentsT = troubleshooting
* Resistors may be substituted for test where an electronic load is not available.
** Less accurate, and less expensive, current-monitor resistors can be used, but the accuracy to which current programming
and current meter reading can be checked must be reduced accordingly.
11
Initial Setup
Maintenance described herein is performed with power supplied to the instrument, and protective covers
removed. Such maintenance should be performed only by service trained personnel who are aware of the
hazards involved (for example, fire and electrical shock). Turn off ac power when making or removing
connections to the power supply. Where maintenance can be performed without power applied, the power
should be removed.
a. Unplug the line cable and remove the top cover by removing the two screws.
b. Slide the cover to the rear.
c. Plug a control board test connector A2P7 onto the A2J7 card-edge fingers.
d. Turn OVERVOLTAGE ADJUST control A3R97 fully clockwise.
e. Disconnect all loads from output terminals.
f. Connect power supply for local sensing, and ensure that MODE switches are set as shown below.
g. Reconnect the line cable and turn on ac power.
h. Allow unit to warm up for 30 minutes.
i. At the beginning of each calibration procedure, the power supply should be in its power-off state, with no external
circuitry connected except as instructed.
j. The POWER LIMIT adjustment (A2R25) must be adjusted at least coarsely before many of the calibration procedures
can be performed. If you have no reason to suspect that the Power Limit circuit is out of adjustment, do not disturb its
setting. Otherwise, center A2R25 before you begin to calibrate the power supply.
Table 2-2. Calibration Procedure
TESTTESTED
VARIABLE
Meter F/S
Adjust.
Resistance
Programming
F/S
Adjust.
Meter Ref.
Voltage
Prog. VoltageVP ( + )
TEST POINTSTEST SEQUENCE AND ADJUSTMENTSEXPECTED
6010A, 6012B
A2J3 pin 7 ( + )
A2J3 pin 10 (-)
6011A, 6015A
A2J3 pin 6 ( + )
A2J3 pin 9 (-)
P ( - )
a. Connect DVM across test points and turn on
ac power.
b. Adjust A2R24 to obtain the voltage range
specified in the results.
a. Connect a 2KΩ 0.01%, ¼W resistor and
DVM between test points.
b. Set MODE switch as in Figure 2-1 and turn on
ac power.
c. Adjust A2R23 to obtain the voltage range
specified in the results.
RESULTS
0.5V ± 50µV
2.5V ±4mV
12
Table 2-2. Calibration Procedure (continued)
TESTTESTED
VARIABLE
V-MON
Zero
Adjust.
Common
Mode
Adjust.
I-MON
Zero
Adjust.
I-MON
F / S
Adjust.
V-MON VM ( + )
Residual
Output
Voltage
VM( + )
I-MONIM ( + )
I-MONIM ( + )
TEST POINTSTEST SEQUENCE AND ADJUSTMENTSEXPECTED
RESULTS
M ( - )
VM ( + )
M ( - )
M (-)
M ( - )
Rm ( + )
Rm ( - )
a. Set voltage and current controls to minimum
settings.
b. Disable power supply as in Initial Setup step i.
c. Short circuit output terminals and connect the
DVM between test points. Turn on power
supply.
d. Adjust V-MON Zero trim pot A2R22 to
voltage range specified in the results.
a. Set voltage and current controls to minimum
and short the unit's sense terminals
( + S & - S).
b. Attach the DVM across test points and disable
power supply as Initial Setup step i.
c. Turn on ac power and record the initial
voltage (IR) with DVM across test points.
d. Remove the local sensing straps and connect a
1Vdc power supply between - S( + ) and –
OUT( - ). See Figure 2-1.
e. Adjust A2R21 to the voltage range specified.
f. Remove the 1V supply and replace jumpers.
a. Set voltage and current controls to minimum.
b. Disable power supply as in Initial Setup step I
and short output terminals. Turn on ac power.
c. Connect DVM across test points and adjust
I-MON Zero trim pot A2R8 as shown in
results.
a. Perform I-MON Zero Adjust before
proceeding .
b. Connect a 0.010Ω (6010A), 0.0005Ω (6011A)
0.0001Ω (6012B), current monitoring resistor
Rm across the output terminals.
c. Turn on ac power and using the “Display
Setting”, set current control to 17A (6010A),
120A (6011A), 50A (6012B), 5A (6015A),
and voltage control to 5V.
d. Connect DVM across test points and take an
initial reading (IR).
e. Connect DVM across Rm monitoring
terminals and adjust A2R9 as shown in the
results.
0 ± 80µV
IR* ±80µV
IR* ±40µV
(6015A)
0± 100µV
IR*
0.034 IR*±
33.5µV (6010A,
6015A)
0.012 IR*
±40µV (6011A,
6012B)
*IR = Initial Reading
13
Table 2-2. Calibration Procedure (continued)
TESTTESTED
VARIABLE
Power
Limit
Adjust.
V(OUT)
I(OUT)
TEST POINTSTEST SEQUENCE AND ADJUSTMENTSEXPECTED
RESULTS
a. Perform I-MON F/S Adjust before
proceeding.
b. Connect the unit to the ac power line via a
variable transformer. Set input power rail to
240Vdc; DVM ( + ) on rear of A1R3 and
DVM (-) to rear of A1R1. Note that power rail
must be maintained at 240Vdc during
calibration.
WARNING
The inner cover must be removed to connect the
voltmeter. Disconnect the power line and wait two
minutes before connecting or disconnecting the
voltmeter.
c. Connect a 3.8Ω (6010A), 0.066Ω (6011A),
0.44Ω (6012B), 40Ω (6015A) resistor or an
electronic load across the unit's output
terminals.
d. Set the load for 18A (6010A), 120A (6011A),
50A (6012B), 5A (6015A), in CC mode, and
turn A2R25 (lower knee) fully counter
clockwise.
e. Turn on power supply and set voltage at 65V
(6010A), 8V (6011A), 22V (6012B), 204V
(6015A), and current at 17.5A (6010A), 121A
(6011A), 51A (6012B), 5.1A (6015A), using
DISPLAY SETTINGS.
f. Turn A2R25 clockwise until CV LED lights.
Output should be 65V ± 0.6V (6010A), 8
±0.08V (6011A), 22 ±0.2V (6012B), 204V
(6015A), and 17A (6010A), 120A (6011A)
51A (6012B), 5.1A (5015A) in CV mode.
g. Turn off ac power and replace the 3.8Ω
(6010A), 0.066Ω (6011A), 0.44Ω (6012B),
40Ω (6015A), resistor with a 38Ω (6010A),
0.36Ω (6011A), 3.3Ω (6012B), 250Ω
(6015A), resistor or reset electronic load for
5.5A (6010A), 55A (6011A), 18.2A (6012B)
in CC mode.
14
Table 2-2. Calibration Procedure (continued)
TESTTESTED
VARIABLE
Power Limit
Adjust
(continued)
TEST POINTSTEST SEQUENCE AND ADJUSTMENTSEXPECTED
RESULTS
h. Turn A2R26 (upper knee) fully counter
clockwise. Turn on the supply and set voltage
at 200V (6010A), 20V (6011A), 60V
(6012B), 500V (6015A), and current at 5.25A
(6010A), 56A (6011A), 19A (6012B) 2.25A
(6015A), using DISPLAY SETTINGS.
i. Turn A2R26 (upper knee) clockwise until CV
LED lights. Output should be 200 ± 2V
(6010A), 20 ±0.5V (6011A), 60 ±0.4V
(6012B), and 5.25A (6010A), 55A (6011A),
18.2A (6012B), 2.2A (6015A), in CV mode.
Figure 2-1. Common Mode Setup
15
Table 2-3. Guide to Recalibration After Repair
Printed Circuit
Board
A1 Main BoardR11
A1 Main BoardT1, T25
A5 Diode BoardCR4
A2 Control BoardConstant Voltage
A2 Control BoardConstant Voltage
A2 Control BoardConstant Current
A2 Control BoardPower Limit
A2 Control BoardBias Power Supplies
A2 Control BoardU7, R84, R85, R247
1. V-MON Zero Calibration
2. Common-Mode Calibration
3. I-MON Zero Calibration
Block NameCircuit WithinRef.
CR5, CR1 (6011A)
All Except Current
(CV) Circuit
(CV) Circuit
(CC) Circuit
Comparator
* Code To Calibration Procedure To Be Performed
Source
Current SourceAll6
± 15V Supplies
4. I-MON Full Scale (F/S) Calibration
5. Power Limit Calibration
6. Resistance Programming Full Scale (F/S) Calibration
7. Meter Full Scale (F/S) Calibration
Designator
R13 (6011A)
All1 then 2
All3 then 4
All5
AllAll
Performance Tests
Perform These
Procedures*
3 then 4
5
The following paragraphs provide test procedures for verifying the unit's compliance with the specifications of Table 1-1 in
the Operating Manual. Please refer to CALIBRATION PROCEDURE or TROUBLESHOOTING if you observe
out-of-specification performance.
Measurement Techniques
Setup For All Tests. Measure the DC output voltage directly at the + S and - S terminals. Connect unit for local sensing,
and ensure that MODE switches are set as shown below. Select an adequate wire gauge for load leads using the procedures
given in the Operating Manual for connecting the load.
Electronic Load. The test and calibration procedures use an electronic load to test the unit quickly and accurately. If an
electronic load is not available, you may substitute:
CV Load Effect (Load Regulation)
CV PARD (Ripple and Noise)
CC Source Effect (Line Regulation)
CC PARD (Ripple and Noise)
The substitution of the load resistor requires adding a load switch to open and short the load in the CC or CV load
regulation tests. The load transient recovery time test procedure is not amenable to modification for use with load resistors.
An electronic load is considerably easier to use than a load resistor. It eliminates the need for connecting resistors or
rheostats in parallel to handle the power, it is much more stable than a carbon-pile load, and it makes easy work of
switching between load conditions as is required for the load regulation and load transient-response tests.
Current-Monitoring Resistor Rm. To eliminate output current measurement error caused by voltage drops in the leads
and connections, connect the current-monitoring resistor between -OUT and the load as a four-terminal device. Figure 2-2
shows correct connections. Select a resistor with stable characteristics:
0.010, 0.02% accuracy, 30 ppm/°C (6010A)
0.0005Ω, 0.05% accuracy, 30ppm/°C (6011A)
0.0010Ω, 0.05% accuracy, 30ppm/°C (6012B)
0.010Ω, 0.02% accuracy, 30ppm/°C (6015A)
or lower temperature coefficient and a current rating of:
CV Setup. If more than one meter or a meter and an oscilloscope are used, connect each to the + S and - S terminals by a
separate pair of leads to avoid mutual coupling effects. Connect only to + S and -S (except for peak-to-peak PARD)
because the unit regulates the output voltage between + S and - S, not between + OUT and -OUT. Use coaxial cable or
shielded 2-wire cable to avoid pickup on test leads. For all CV tests set the output current at full output to assure CV
operation.
Load Effect (Load Regulation). Constant-voltage load effect is the change in dc output voltage (Eo) resulting from a
load-resistance change from open-circuit to full-load. Full-load is the resistance which draws the maximum rated output
current at voltage Eo. Proceed as follows:
a. Connect the test equipment as shown in Figure 2-3. Operate the load in constant resistance mode (Amps/Volt) and set
resistance to maximum.
b. Turn the unit's power-on, and turn up current setting to full output.
c. Turn up output voltage to:
60Vdc (6010A)
7.0Vdc (6011A)
20.0Vdc (6012B)200Vdc (6015A)as read on the digital voltmeter.
Figure 2-3. Basic Test Setup
d.Reduce the resistance of the load to draw an output current of:
17.0Adc (6030A)120Adc (6011A)50Adc (6012B)
5.0Adc (6015A)Check that the unit's CV LED remains lighted.
e. Record the output voltage at the digital voltmeter.
f. Open-circuit the load.
18
g. When the reading settles, record the output voltage again. Check that the two recorded readings differ no more than:
Source Effect (Line Regulation). Source effect is the change in dc output voltage resulting from a change in ac input
voltage from the minimum to the maximum value as specified in Input Power Requirements in the Specifications Table, in
the Operating Manual. Proceed as follows:
a. Connect the test equipment as shown in Figure 2-3. Operate the load in constant resistance mode (Amps/Volt) and set
resistance to maximum.
b. Connect the unit to the ac power line through a variable autotransformer which is set for low line voltage (104Vac for
120Vac).
c. Turn the unit's power-on, and turn up current setting to full output.
d. Turn up output voltage to:
60.0Vdc (6010A)
20.0Vdc (6011A)
20.0Vdc (6012B)500Vdc (6015A)as read on the digital voltmeter.e.Reduce the resistance of the load to draw an output current of:
17.0Adc (6010A)50Adc (6011A)50Adc (6012B)
2.0Adc (6015A)Check that the unit's CV LED remains lighted.
f. Record the output voltage at the digital voltmeter.
g. Adjust autotransformer to the maximum for your line voltage.
h. When the reading settles record the output voltage again. Check that the two recorded readings differ no more than:
PARD (Ripple And Noise). Periodic and random deviations (PARD) in the unit's output-ripple and noise-combine to
produce a residual ac voltage superimposed on the dc output voltage. Constant-voltage PARD is specified as the
root-mean-square (rms) or peak-to-peak (pp) output voltage in a frequency range of 20Hz to 20MHz (10MHz, 6010A).
RMS Measurement Procedure. Figure 2-4 shows the interconnections of equipment to measure PARD in Vrms. To
ensure that there is no voltage difference between the voltmeter's case and the unit's case, connect both to the same ac
power outlet or check that the two ac power outlets used have the same earth-ground connection.
Use the common-mode choke as shown to reduce ground-loop currents from interfering with measurement. Reduce noise
pickup on the test leads by using 50Ω coaxial cable, and wind it five turns through the magnetic core to form the
common-mode choke. Proceed as follows:
a. Connect the test equipment as shown in Figure 2-4. Operate the load in constant resistance mode (Amps/Volt) and set
resistance to maximum.
b. Turn the unit's power-on, and turn up current setting to full output.
c. Turn up output voltage to:
d.Reduce the resistance of the load to draw an output current of:
17.0Adc (6010A)120Adc (6011A)
17.5Adc (6012B)
5.0Adc (6015A)Check that the unit's CV LED remains lighted.
e. Check that the rms noise voltage at the true rms voltmeter is no more than:
22mV rms (6010A)
8.0mV rms (6011A)
8.0mV rms (6012B)
50mV rms (6015A)
Figure 2-4. RMS Measurement Test Setup, CV PARD Test
Peak Measurement Procedure. Figure 2-5 shows the interconnections of equipment to measure PARD in Vpp. The
equipment grounding and power connection instructions of PARD rms test apply to this setup also. Connect the
oscilloscope to the + OUT and - OUT terminals through 0.01µF blocking capacitors to protect the oscilloscope's input from
the unit's output voltage. To reduce common-mode noise pickup, set up the oscilloscope for a differential, two-channel
voltage measurement. To reduce normal-mode noise pickup, use twisted, 1 meter or shorter, 50Ω coaxial cables with
shields connected to the oscilloscope case and to each other at the other ends. Proceed as follows:
a. Connect the test equipment as shown in Figure 2-5. Operate the load in constant resistance mode (Amps/Volt) and set
resistance to maximum.
b. Turn the unit's power-on, and turn up current setting to full output.
c. Turn up output voltage to:
60Vdc (6010A)
7.0Vdc (6011A)
20
60Vdc (6012B)200Vdc (6015A)d.Reduce the resistance of the load to draw an output current of:
17.0Adc (6010A)120Adc (6011A)
17.5Adc (6012B)
5.0Adc (6015A)Check that the unit's CV LED remains lighted.
e. Set the oscilloscope's input impedance to 50Ω and bandwidth to 20MHz. Adjust the controls to show the 20KHz and
higher frequency output-noise waveform of Figure 2-6.
Figure 2-5. Peak-To-Peak Measurement Test Setup, CV PARD Test
Load Transient Recovery Time. Specified for CV operation only; load transient recovery time is the time for the output
voltage to return to within a specified band around its set voltage following a step change in load.
Use the equipment setup of Figure 2-3 to display output voltage transients while switching the load between 10% with the
output set at:
60Vdc (6010A)
7Vdc (6011A)
20Vdc (6012B)
200Vdc (6015A)
21
6010A6011A
NOT APPLICABLE
6012B6015A
Figure 2-6. 20KHz Noise, CV Peak-to-Peak PARD
Proceed as follows:
a. Connect the test equipment as shown in Figure 2-3. Operate the load in constant-current mode and set for minimum
current.
b. Turn the unit's power-on, and turn up current setting to full output.
c. Turn up output voltage to:
60Vdc (6010A)
7.0Vdc (6011A)
20.0Vdc (6012B)200Vdc (6015A)as read on the digital voltmeter.d.Set the load to vary the load current between:15 and 17Adc (6010A)108 and 120Adc (6011A)45 and 50Adc (6012B)
4.5 and 5.0Adc (6015A)at a 30Hz rate for the 10% RECOVERY TEST.
e. Set the oscilloscope for ac coupling, internal sync and lock on either the positive or negative load transient.
f. Adjust the oscilloscope to display transients as in Figure 2-7.
22
g. Check that the pulse width of the transient pulse is no more than:
Temperature Coefficient. Temperature coefficient (TC) is the change in output voltage for each °C change in ambient
temperature with constant ac line voltage, constant output voltage setting and constant load resistance. Measure temperature
coefficient by placing the unit in an oven, varying the temperature over a range within the unit's operating temperature range,
and measuring the change in output voltage. Use a large, forced air oven for even temperature distribution. Leave the unit at
each temperature measurement for half hour to ensure stability in the measured variable. Measure the output voltage with a
stable DVM located outside the oven so voltmeter drift does not affect the measurement accuracy. To measure offset TC,
repeat the procedure with output voltage set to 0.10Vdc.
Proceed as follows:
a. Connect DVM between +S and -S.
b. Place power supply in oven, and set temperature to 30°C.
c. Turn the unit's power-on and turn up current setting to full output.
d. Turn up output voltage to the following:
23
200Vdc (6010A)
20.0Vdc (6011A)
60.0Vdc (6012B)
500Vdc (6015A)
as read on the DVM.
e. After 30 minutes stabilization, record the temperature to the nearest 0.1°C. Record the output voltage on the DVM.
f. Set oven temperature to 50°C.
g. After 30 minutes stabilization, record the temperature to the nearest 0.1°C. Record output voltage.
h. Check that the magnitude of the output voltage change is no greater than
620mV.(6010A)
80mV (6011A)
176mV (6012B)
1.6V (6015A)
Drift (Stability). Drift is the change in output voltage beginning after a 30-minute warm-up during 8 hours operation with
constant ac input line voltage, constant load resistance and constant ambient temperature. Use a DVM and record the output at
intervals, or use a strip-chart recorder to provide a continuous record. Check that the DVM's or recorder's specified drift during
the 8 hours will be no more than 0.001%. Place the unit in a location with constant air temperature preferably a large forced-air
oven set to 30°C and verify that the ambient temperature does not change by monitoring with a thermometer near the unit.
Typically the drift during 30 minute warm-up exceeds the drift during the 8-hour test. To measure offset drift, repeat the
procedure with output voltage set to 0.10Vdc.
a. Connect DVM between + S and - S.
b. Turn the unit's power-on and turn up current setting to full output.
c. Turn up output voltage to:
200Vdc (6010A)
20Vdc (6011A)
60.0Vdc (6012B)
500Vdc (6015A)
as read on the digital voltmeter.
d. After a 30 minute warmup, note reading on DVM.
e. The output voltage should not deviate more than
77mV (6010A)
9mV (6011A)
23mV (6012B)
190mV (6015A)
from the reading obtained in step d over a period of 8 hours.
Constant Current (CC) Tests
CC Setup. Constant-current tests are analogous to constant-voltage tests, with the unit's output short circuited and the
voltage set to full output to assure CC operation. Follow the general setup instructions on Page 16.
Load Effect (Load Regulation). Constant current load effect is the change in dc output current (Io) resulting from a
load-resistance change from short-circuit to full-load, or full-load to short-circuit. Full-load is the resistance which develops
the maximum rated output voltage at current Io. Proceed as follows:
a. Connect the test equipment as shown in Figure 2-3. Operate the load in constant resistance mode (Amps/Volt) and set
resistance to minimum.
b. Turn the unit's power-on, and turn up voltage setting to full output.
c. Turn up output current to:
5.0Adc (0.050Vdc across Rm) (6010A) Check that the AMPS display reads about 5 amps.
50Adc (0.25Vdc across Rm) (6010A) Check that the AMPS display reads about 50 amps.
24
17.5Adc (0.0175Vdc across Rm) (6012B) Check that the AMPS display reads about 17.5 amps.
2Adc (0.20Vdc across Rm) (6015A) Check that the AMPS display reads about 2 amps.
d. Increase the load resistance until the output voltage at +S and -S increases to:
200Vdc (6010A)
20Vdc (6011A)
60Vdc (6012B)
500Vdc (6035A)
Check that the CC LED is lighted and AMPS display still reads ≈ current setting.
e. Record voltage across Rm.
f. Short circuit the load.
g. When the reading settles (≈ 10s), record the voltage across Rm again. Check that the two recorded readings differ no
Source Effect (Line Regulation). Constant current source effect is the change in dc output current resulting from a
change in ac input voltage from the minimum to the maximum values listed in the Specifications Table in the Operating
Manual. Proceed as follows:
a. Connect the test equipment as shown in Figure 2-3. Operate the load in constant resistance mode (Amps/Volt) and set
resistance to minimum.
b. Connect the unit to the ac power line through a variable autotransformer set for low line voltage (e.g. 104Vac for
120Vac).
c. Switch the unit's power-on and turn up output voltage setting to full output.
d. Turn up output current to:
17.0Adc (6010A)120Adc (6011A)50Adc (6012B)
5.0Adc (6015A)Check that the AMPS display reads ≈ current setting.e.Increase the load resistance until the output voltage between + S and - S increases to:60Vdc (6010A)
7.0Vdc (6011A)
20.0Vdc (6012B)200Vdc (6035A)Check that the CC LED is still on and the AMPS display still reads ≈ current setting.
f. Record the voltage across Rm.
g. Adjust autotransformer to the maximum for your line voltage.
h. When the reading settles record the voltage across Rm again. Check that the two recorded readings differ no more than:
PARD Ripple And Noise. Periodic and random deviations (PARD) in the unit's output (ripple and noise) combine to
produce a residual ac current as well as an ac voltage super-imposed on the dc output. The ac voltage is measured as
constant-voltage PARD. Constant-current PARD is specified as the root-mean-square (rms) output current in a frequency
range 20Hz to 20MHz with the unit in CC operation. To avoid incorrect measurements, with the unit in CC operation,
caused by the impedance of the electronic load at noise frequencies, use a:
0.4Ω (6010A)
0.058Ω (6011A)
0.4Ω (6012B)
40Ω (6015A)
25
load resistor that is capable of safely dissipating 1000 watts. Proceed as follows:
a. Connect the test equipment as shown in Figure 2-8.
b. Switch the unit's power-on and turn the output voltage all the way up.
c. Turn up output current to:
17.0Adc (6010A)120Adc (6011A)50Adc (6012B)
5.0Adc (6015A)Check that the unit's CC LED remains lighted.
d. Check that the rms noise current measured by the current probe and rms voltmeter is no more than:
Maintenance described herein is performed with power supplied to the instrument, and protective covers
removed. Such maintenance should be performed only by service-trained personnel who are aware of the
hazards involved (for example, fire and electrical shock). Where maintenance can be performed without
power applied, the power should be removed.
Introduction
Before attempting to troubleshoot this instrument, ensure that the fault is with the instrument itself and not with an
associated circuit. The performance test enables this to be determined without having to remove the covers from the supply.
The most important aspect of troubleshooting is the formulation of a logical approach to locating the source of trouble. A
good understanding of the principles of operation is particularly helpful, and it is recommended that Chapter 4 of this
manual be reviewed before attempting to troubleshoot the unit. Often the user will then be able to isolate a problem simply
by using the operating controls and indicators. Once the principles of operation are understood, refer to the following
paragraphs.
Table 2-1 lists the test equipment for troubleshooting. Chapter 6 contains schematic diagrams and information concerning
the voltage levels and waveforms at many of the important test points. Most of the test points used for troubleshooting the
supply are located on the control board test "fingers", which are accessible close to the top of the board. See Table 3-1.
3
If a component is found to be defective, replace it and re-conduct the performance test. When a component is replaced,
refer to Calibration Procedure (Chapter 2). It may be necessary to perform one or more of the adjustment procedures after a
component is replaced.
Initial Troubleshooting Procedures
If a problem occurs, follow the steps below in sequence:
a. Check that input power is available, and check the power cord and rear-panel circuit breaker.
b. Check that the settings of mode switch A2S1 are correct for the desired mode of operation. (See Operating Manual).
c. Check that all connections to the power supply are secure and that circuits between the supply and external devices are
not interrupted.
d. If the power supply fails turn-on self-test or gives any other indication of malfunction, remove the unit from the
operating system before proceeding with further testing.
Some circuits on the power mesh are connected directly to the ac power line. Exercise extreme caution
when working on energized circuits. Energize the supply through an isolation transformer to avoid
shorting ac energized circuits through the test instrument's input leads. The isolation transformer must
have a power rating of at least 4KVA. During work on energized circuits, the safest practice is to
disconnect power, make or change the test connections, and then re-apply power.
Make certain that the supply's ground terminal (┴) is securely connected to an earth ground before applying
power. Failure to do so will cause a potential shock hazard that could result in personal injury.
The following caution outlines important precautions which should be observed when working with static sensitive
components in the power supply.
This instrument uses components which can be damaged by static charge. Most semiconductors can
suffer serious performance degradation as a result of static charges, even though complete failure may
not occur. The following precautions should be observed when handling static-sensitive devices.
a. Always turn power off before removing or installing printed-circuit boards.
b.
Always stored or transport static-sensitive devices (all semiconductors and thin-film devices) in conductive material.
Attach warning labels to the container or bag enclosing the device.
c.
Handle static-sensitive devices only at static-free work stations. These work stations should include special conductive
work surfaces (such as Agilent Part No. 9300-0797) grounded through a one-megohm resistor. Note that metal table
tops and highly conductive carbon-impregnated plastic surfaces are too conductive; they can act as large capacitors and
shunt charges too quickly. The work surfaces should have distributed resistance of between 10
d.
Ground all conductive equipment or devices that may come in contact with static-sensitive devices or subassemblies
containing same.
e.
Where direct grounding of objects in the work area is impractical, a static neutralizer should be used (ionized air
blower directed at work). Note that this method is considerably less effective than direct grounding and provides less
protection for static-sensitive devices.
f.
While working with equipment on which no point exceeds 500 volts, use a conductive wrist strap in contact with skin.
The wrist strap should be connected to ground through a one-megohm resistor. A wrist strap with insulated cord and
built-in resistor is recommended, such as 3M Co. No. 1066 (Agilent Part No. 9300-0969 (small) and 9300-0970
[large]).
6
and 10l2 Ω per square.
Do not wear a conductive wrist strap when working with potentials in excess of 500 volts; the one-megohm
resistor will provide insufficient current limiting for personal safety.
g. All grounding (device being repaired, test equipment, soldering iron, work surface, wrist strap, etc.) should be done to
the same point.
h.
Do not wear nylon clothing. Keep clothing of any kind from coming within 12 inches of static-sensitive devices.
i.
Low-impedance test equipment (signal generators, logic pulsers, etc.) should be connected to static-sensitive inputs
only while the components are powered.
j.
Use a mildly activated rosin core solder (such as Alpha Metal Reliacor No. 1, Agilent Part No. 8090-0098) for repair.
The flux residue of this type of solder can be left on the printed circuit board. Generally, it is safer not to clean the
printed-circuit board after repair. Do not use Freon or other types of spray cleaners. If necessary, the printed-circuit
board can be brushed using a natural-bristle brush only. Do not use nylon-bristle or other synthetic-bristle brushes. Do
not use high-velocity air blowers (unless ionized).
k.
Keep the work area free of non-conductive objects such as Styrofoam-type cups, polystyrene foam, polyethylene bags,
and plastic wrappers. Non-conductive devices that are necessary in the area can be kept from building up a static
charge by spraying them with an anti-static chemical (Agilent Part No. 8500-3397).
l.
Do not allow long hair to come in contact with static-sensitive assemblies.
m.
Do not exceed the maximum rated voltages specified for the device.
Repair and Replacement
Repair and replacement of most components in the power supply require only standard techniques that should be apparent
to the technician. The following paragraphs provide instructions for removing certain assemblies and components for which
the procedure may not be obvious upon inspection.
29
To avoid the possibility of personal injury, remove the power supply from operation before opening the
cabinet. Turn off ac power and disconnect the line cord, load, and remote sense leads before attempting
any repair or replacement.
When replacing any heatsink-mounted components except thermostat, smear a thin coating of heatsink
compound between the component and heatsink. If a mica insulator is used, smear a thin coating of
heatsink compound on both sides of the mica insulator.
Do not use any heatsink compound containing silicone, which can migrate and foul electrical contacts
elsewhere in the system. An organic zinc oxide cream, such as American Oil and Supply Company
Heatsink Compound #100, is recommended.
Most of thc attaching hardware in this unit is metric. The only non-metric (sometimes called English or
inch) fittings are listed below. Be careful when both types of screws are removed not to get them
mixed up.
a. Screws that secure the input and output capacitors to A1 main board and output bus.
b.
Rear-panel circuit breaker.
c.
Rear-panel ground binding post.
Top Outside Cover Removal. Remove the two top rear screws using a Size 2, Pozidriv screwdriver. A Phillips head
screwdriver does not fully seat into Pozidriv screws and risks stripping the heads. Remove the top cover by sliding it to the
rear and lifting at the front.
Bottom Cover Removal. Remove the handles from both sides of the unit and remove the bottom cover by sliding it to the
rear. Use a Phillips head #2 screwdriver to remove the handle screws. You do not need to remove the unit's feet.
Inside Top Cover Removal. The unit includes an inside cover which secures the vertical board assemblies. Remove the
inside cover for repair but not for calibration. Remove the nine mounting screws (Pozidriv, M4x7) – two in the left side,
three on the right side, and four on top. Remove the inside cover by lifting at the front edge.
When installing the inside cover, insert it first at the right side. While holding it tilted up at the left, reach through the
cutouts in the cover and fit the top tabs of the A2 control board into the mating slots in the cover. Then repeat the process
for the A4 FET board, and the A5 Diode board. Press the inside cover down firmly while tightening screws that secure
cover to chassis. Be careful when replacing printed-circuit assemblies and covers not to bend any boards or components.
A2 Control Board Removal
After removing the inside cover, unplug the W1 ribbon cable at the front edge of the A2 control board and unplug the W7
and W8 ribbon cables from the lower center of the A2 control board. Remove the A2 board by lifting first at the front edge
and than pulling it up and out of the unit.
When installing the A2 board, insert it first at the rear of the unit. While holding it tilted up at the front, fit the A2TB1
terminal strip into the mating cutout in the rear panel. Then lower the A2 board's bottom tabs into the mating slots on the
chassis. Re-install the W1, W7, and W8 ribbon cables.
A4 FET Board Removal
After removing the inside cover, remove the A4 FET board by lifting, using the large aluminum heatsink as a handle. One
connector and one tab holds the A4 board at its bottom edge.
30
Loading...
+ 96 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.