This guide describes how to service the HP 53131A and HP 53132A
Universal Counters, and the HP 53181A Frequency Counter.
The information in this guide applies to instruments having the number
prefix listed below, unless accompanied by a “Manual Updating Changes”
package indicating otherwise.
SERIAL PREFIX NUMBER:3711 AND ABOVE (HP 53131A)
3710 AND ABOVE (HP 53132A)
3711 AND ABOVE (HP 53181A)
HP 53131A/132A/181A 225 MHz
Counter
Page 4
Copyright Hewlett- P ac ka rd
Company 1994, 1995, 1997,
1999
All Rights Reserved.
Reproduction, adaptation, or
translations without prior
written permission is
prohibited, except as allowed
under the copyright laws.
Printed: March 1999
Printed in USA
Manual part number
53131-90039
Certification
and Warranty
Certification
Hewlett-Packard Company
certifies that this product met
its published specification at the
time of shipment from the
factory. Hewlett-Packard
further certifies that its
calibration measurements are
traceable to the United States
National Institute of Standards
and Technol ogy (formerl y
National Bureau of Standards),
to the extent allowed by the
Institute’s calibration facility,
and to the calibration facilities
of other International
Standards Organization
members.
Warranty
HP warrants HP hardware,
accessories and supplies against
defects in materials and
workmanship for a period of
three years from date of
shipment. If HP receives notice
of such defects during the
warranty period, HP will, at its
option, either repair or replace
products which prove to be
defective. Replacement products
may be either new or like-new.
HP warrants that HP software
will not fail to execute its
programming instructions, for
the period specified above, due
to defects in material and
workmanship when properly
installed and used. If HP
receives notice of such defects
during the warranty period, HP
will replace software media
which does not execute its
programming instructions due
to such defects.
For detailed warra nty
information, see back matter.
Safety Considerations
General
This product and related
documentation must be
reviewed for familiarizat ion
with this safety marki ngs and
instructions before operation.
Before Cleaning
Disconnect the product from
operating power be fore
cleaning.
Warning Symbols That May
Be Used In This Book
Instruction manual symbol; the
product will be marked with
this symbol when it is necessary
for the user to refer to the
instruction manual.
Indicates hazardous voltages.
Indicates earth (ground)
terminal.
or
Indicates terminal is connected
to chassis when such connection
is not apparent.
Indicates Alternating current.
Indicates Direct current.
Safety Considerations
(contd)
WARNING
BODILY INJUR Y OR DEATH
MAY RESULT FROM
FAILURE TO HEED A
WARNING. DO NOT
PROCEED BEYOND A
WARNING UNTIL THE
INDICATED CONDITIONS
ARE FULLY UNDERSTOOD
AND MET .
CAUTION
Damage to equipment, or
incorrect measurement data,
may result from failure to
heed a caution. Do not
proceed beyond a CAUTION
until the indicated conditions
are fully understood and met.
Safety Earth Ground
An uninterruptible safety earth
ground must be maintained
from the mains power source to
the product’s ground circuitry.
WARNING
WHEN MEASURING POWER
LINE SIGNALS, BE
EXTREMELY CAREFUL AND
ALWAYS USE A
STEP-DOWN ISOLATION
TRANSFORMER WHICH
OUTPUT IS COMPATIBLE
WITH THE INPUT
MEASUREMENT
CAPABILITIES OF THIS
PRODUCT. THIS PRODUCT’S
FRONT AND REAR PANELS
ARE TYPCIALLY AT EARTH
GROUND.
TO MEASURE AC POWER
LINE SIGNALS WITHOUT AN
ISOLATION TRANSFORMER.
For additional safety and
acoustic noise information, see
back matter.
THUS, NEVER TRY
Hewlett-Packard Company7.NC.NL.A.11.03.97.R1.P.CW6FC
Santa Clara Division
5301 Stevens Creek Boulevard
Santa Clara, California 95052-8059
Page 5
Contents
Preface
How to Use This Guidexiii
Repair Strategyxiii
Instrument Identificationxiii
Instruments Covered by this Guidexiv
Assembly-Level Service Guide Organizationxiv
How to Order Guidesxv
Description of the 225 MHz Countersxvi
Optionsxviii
Equipment Required1-4
HP 53131A/132A/181A Operational Verification1-5
Power-On Self Tests1-5
Run Self Test1-8
Termination Check1-8
External Arm Test (HP 53131A/132A Only)1-10
Equipment1-10
Assembly-Level Service Guideiii
Page 6
Contents
Counter Setup1-10
Procedure1-11
External Timebase Tests1-12
Equipment1-12
1 MHz External Timebase Input (HP 53131A and HP
53181A Only)1-13
5 MHz External Timebase Input (HP 53131A and HP
53181A Only)1-14
10 MHz External Timebase Input (HP 53131A and HP
53181A Only)1-14
10 MHz External Timebase Input for the HP
53132A1-15
HP 53131A/132A Complete Performance Tests1-16
Test 1: Time Interva l (H P 53 13 1A/132A Only)1-1 7
Equipment1-17
Counter Setup1-17
Procedure1-18
Test 2: Trigger Level (HP 53131A/132A Only)1-19
Equipment1-19
Counter Setup1-19
Procedure1-20
Test 3: Channels 1 and 2 Frequency Sensitivity
(HP 53131A/132A Only)1-22
Equipment1-22
Counter Setup1-22
100 kHz to 100 MHz Sensitivity for HP 53131A/132A1-23
100 MHz to 200 MHz Sensitivity for HP 53131A/
132A1-25
200 MHz to 225 MHz Sensitivity for HP 53131A/
132A1-26
Test 4: Channels 1 and 2 Frequency Accuracy
(HP 53131A/132A Only)1-27
Equipment1-27
ivAssembly-Level Service Guide
Page 7
Contents
Counter Setup1-27
Procedure1-28
Test 5: Option 030/050/124 Channel 3 Frequency
Sensitivity (HP 53131A/132A Only)1-30
Equipment1-30
Counter Setup1-30
100 MHz to 2.5 GHz Sensitivity for HP 53131A/132A Option
030 only1-31
2.7 to 3.0 GHz Sensitivity for Option 030 only1-32
200 MHz to 5.0 GHz Sensitivity for HP 53131A/132A
(Option 050 only)1-33
200 MHz to 12.4 GHz Sensitivity for HP 53131A/132A
(Option 124 only)1-34
Test 6: Option 030/050/124 Channel 3 Frequency
Accuracy (HP 53131A/132A Only)1-35
Equipment1-35
Counter Setup1-35
Procedure1-35
Test 7: Peak Volts, Channels 1 and 2 (HP 53131A/132A
Only)1-37
Equipment1-37
Counter Setup1-37
Volt Peak 11-38
Volt Peak 21-39
HP 53131A/132A HP-IB Verification Program
(Optional)1-40
HP 53131A/132A Pe rformance Test Record 1-41
Assembly-Level Service Guidev
Page 8
Contents
HP 53181A Complete Performance Tests1-47
Test 1: Trigger Level (HP 53181A Only)1-48
Equipment1-48
Counter Setup1-48
Procedure1-49
Test 2: Channel 1 Frequency Sensitivity (HP 53181A
Only)1-51
Equipment1-51
Counter Setup1-51
100 kHz to 100 MHz Sensitivity for HP 53181A1-52
100 MHz to 200 MHz Sensitivity for HP 53181A1-54
200 MHz to 225 MHz Sensitivity for HP 53181A1-54
Test 3: Channel 1 Frequency Accuracy (HP 53181A
Only)1-55
Equipment1-55
Counter Setup1-55
Procedure1-56
Test 4: Option 015/030/050/124 Channel 2 Frequency
Sensitivity1-58
Equipment1-58
Counter Setup1-58
100 MHz to 1.5 GHz Sensitivity for Option 015 Only1-59
100 MHz to 2.7 GHz Sensitivity for Option 030 Only1-60
2.8 to 3.0 GHz Sensitivity for Option 030 Only1-61
200 MHz to 5.0 GHz Sensitivity for Option 050 only1-61
200 MHz to 12.4 GHz Sensitivity for Option 124 only1-62
100 MHz to 1.5 GHz Accuracy Test for Option 0151-63
Equipment1-63
Counter Setup1-63
Procedure1-64
viAssembly-Level Service Guide
Page 9
Contents
Minimum to Maximum Accuracy Test for Options 030/050/
1241-65
Test 6: Peak Volts, Channel 1 (HP 53181A Only)1-68
Equipment1-68
Counter Setup1-68
Procedure1-69
HP 53181A HP-IB Verification Program (Optional)1-70
HP 53181A Performance Test Record 1-71
2Service
Introduction2-2
Returning the Instrument to Hewlett-Packard for
Service2-3
To Provide Repair Information2-3
To Pack in the Original Packaging Materials2-4
To Pack in the Commercially Available Materials2-5
About the HP 53131A/132A Calibration Menu2-6
Overview of the HP 53131A/132A Calibration Menu2-6
The HP 53131A/132A Calibration Menu Tree2-8
To View the Calibration Menu and Security Status2-9
To Unsecure for Calibration2-9
To Initiate the Calibration Routines2-9
To Secure Against Calibration2-13
To Change to a New Security Code2-13
To View the Calibration Count2-14
To Get Help with the Calibration Menu2-14
The HP 53131A/132A Calibration Procedures2-15
First Determine the Counter Firmware Revision2-15
Equipment1-65
Counter Setup1-65
Procedure1-66
Assembly-Level Service Guidevii
Page 10
Contents
To Calibrate the Offset for Channels 1 and 22-15
To Calibrate the Gain for Channels 1 and 22-18
To Calibrate Time Interval2-19
CAL: TI QUIK? Calibration2-20
Equipment2-20
CAL: TI FINE? Calibration2-21
Equipment2-21
To Calibrate the Standard Timebase2-23
To Calibrate the High Stability Timebase Option (Medium,
High, or Ultra-High)2-24
About the HP 53181A Calibration Menu2-25
Overview of the HP 53181A Calibration Menu2-25
The HP 53181A Calibration Menu Tree2-27
To View the Calibration Menu and Security Status2-28
To Unsecure for Calibration2-28
To Initiate the Calibration Routines2-29
To Secure Against Calibration2-29
To Change to New the Security Code2-30
To View the Calibration Count2-30
To Get Help with the Calibration Menu2-30
The HP 53181A Calibration Procedures2-31
First Determine the Counter Firmware Revision2-31
To Calibrate the Offset for Channel 12-31
To Calibrate the Gain for Channel 12-33
To Calibrate the Standard Timebase2-34
To Calibrate the High Stability Timebase Option (Medium,
High, or Ultra-High)2-35
Pre-Troubleshooting Information2-36
Safety Considerations2-36
Recommended Test Equipment2-37
Repair Considerations2-38
Electrostatic Discharge2-38
viiiAssembly-Level Service Guide
Page 11
Contents
Surface Mount Repair2-38
Disassembly and Reassembly Specifics2-38
Power Supp ly Ch e ck2-45
Overview of the Self-Test Routines2-47
Diagnosing the Faulty Assembly by Using the
Self Tests2-49
To Run the Test-All Self Test2-49
To Run the Individual Self Tests2-50
3Replacing Assemblies
Introduction3-2
Tools Required3-3
Do This First3-3
To Remove the Cover3-4
To Remove the Front Bezel3-6
To Remove A1 Motherboard Assembly3-9
To Remove the A2 Display Board, Keypads,
and Window3-11
To Remove A3 1.5/3.0/5.0/12.4 GHz Channel Assembly
(Option 015/ 03 0/ 05 0/ 12 4 )3-13
To Remove A4 AC Power Supply Assembly3-15
To Remove A5 DC Power Input Assembly
(Option 002)3-17
To Remove A6 High Stability Timebase Assembly
(Options 001, 010, and 012)3-19
To Remove the Rear Terminals (Option 060)3-21
Assembly-Level Service Guideix
Page 12
Contents
4Retrofitting Options
Introduction4-2
Tools Required4-2
Do This First4-2
To Retrofit A3 1.5/3.0/5.0/12.4 GHz Channel Assembly
(Option 015/ 03 0/ 05 0/ 12 4 )4- 3
To Retrofit A5 DC Power Input Assembly (Option
002)4-4
Option 002 DC Power Input Assembly Parts4-4
Preliminary Procedure4-4
Retrofitting Procedure4-5
To Retrofit A6 High Stability Timebase Assembly
(Options 001, 010, and 012)4-11
Option 001 Medium Stability Timebase Assembly
Parts4-11
Option 010 High Stability Timebase Assembly Parts4-11
*
Option 012 Ultra-High Stability Timebase Assembly
Parts4-11
Procedure4-12
To Retrofit the Rear Terminals (Option 060)4-14
Option 060 Rear Terminals Parts4-14
5Replaceable Parts
Introduction5-2
Exchange Assemblies5-2
Reference Designations5-3
Replaceable Parts5-3
How To Order A Part5-4
Parts Identification5-4
Contacting Hewlett-Packard5-5
Instrument Inputs7-2
Time Base7-4
Measurement Specifications7-5
Measurement Definitions7-12
Measurement Arming and Processing7-14
General Information7-16
8HP 53181A Specifications
Introduction8-2
Instrument Inputs8-2
Time Base8-4
Measurement Specifications8-5
Measurement Definitions8-9
Measurement Arming and Processing8-10
General Information8-11
Index
Assembly-Level Service Guidexi
Page 14
Contents
xiiAssembly-Level Service Guide
Page 15
Preface
This guide provides assembly-level service information for the HP 53131A
and HP 53132A Universal Counters, and the HP 53181A Frequency
Counter.
How to Use This Guide
Repair Strategy
This service guide is designed to isolate failures to the assembly level only.
The HP 53131A/132A/181A Counter can be returned to Hewlett-Packard
(HP) for all service work, including troubleshooting, and verifying
specifications. Contact your nearest HP Sales and Service Office for more
details.
Note: there is an Express Repair/Performance Calibration Service for US A
customers if downtime is critical. You can receive your repaired Counter
via overnight shipment. Just call 800-403-0801 and ask for Express
Repair/Performance Calibration Service. When your Counter is repaired,
it will be returned via overnight shipment at no extra charge.
If you decide to service the Counter yourself, use the troubleshooting
procedures in Chapter 2 (Service) and the disassembly and reassembly
procedures in Chapter 3 (Replacing Assemblies). Then use the calibration
instructions in Chapter 2 to calibrate the Counter for peak-performance
operation, and finally perform the complete performance tests in
Chapter 1 to verify the Counter’s specifications.
Instrument Identification
Instrument identification is made from the serial number on the rear
panel of the Counter. HP uses a two-part serial number w ith the first part
(prefix) identifying a series of instruments and the second part (suffix)
identifying a particular instrument within a series. An HP-assigned alpha
character between the prefix and suffix identifies the country in which the
instrument was manufactured.
Assembly-Level Service Guidexiii
Page 16
Preface
Instruments Covered by this Guide
This guide applies directly to HP 53131A, HP 53132A, and HP 53181A
Counters that have the same serial number prefix(es) shown on the title
page. If the serial number prefix of your Counter differs from that listed
on the title page of this guide, then there may be differences between this
guide and your instrument.
Instrument s having a higher serial prefix are covered when required by
one or more manual-change sheets included with this guide. If a required
change sheet is missing, contact your nearest HP Sales Office listed at the
back of this guide. Instruments having a serial prefix lower than that
listed on the title page are covered in Chapter 6, “Backdating.”
Assembly-Level Service Guide Organization
This assembly-level service guide consists of a table of contents, preface,
eight chapters and an index. The page running headers identify the
chapters and sections of this manual. The chapter contents are
summarized as follows:
Chapter 1, “Performance Tests,” provides procedures that verify the
Counter operates properly and meets the HP 53131A/132A or HP 53181A
specifications given in Chapter 7, “HP 53131A/132A Specifications,” or
Chapter 8, “HP 53181A Specifications,” in this guide.
Chapter 2, “Service,” is divided into seven main sections that provide
instructions for returning the Counter to HP for service, calibrating the
Counter, and troubleshooting the assemblies in the Counter.
Chapter 3, “Replacing Assemblies,” provides procedures for replacing
defective assemblies and/or modules in the Counter.
Chapter 4, “Retrofitting Options,” provides field-installation procedures
that help you install options into the Counter.
Chapter 5, “Replaceable Parts,” lists the replaceable parts contained in
the Counter, and explains how to order replacement parts for your
Counter.
xivAssembly-Level Service Guide
Page 17
Preface
Chapter 6, “Backdating,” contains information required to adapt this
manual for older instruments.
Chapter 7, “HP 53131A/132A Specifications,” lists all the specifications
and operating characteristics for the HP 53131A/132A Universal Counter.
Chapter 8, “HP 53181A Specifications,” lists all the specifications and
operating characteristics for the HP 53181A Frequency Counter.
How to Order Guides
The part number for this guide is listed in the Certification and Warranty
page (located on the back of the title page) and on the rear cover of this
guide. Preface will always follow table of contents. Page numbers will be
in roman lower case.
Assembly-Level Service Guidexv
Page 18
Preface
Description of the 225 MHz Counters
The HP 53131A and HP 53132A are universal counters capable of
measuring frequencies to 225 MHz on Channels 1 and 2. With an optional
Channel 3, this frequency is extended to 3.0 GHz (Option 030); 5.0 GHz
(Option 050) or 12.4 GHz (Option 124).”
For the HP 53131A, frequency and time interval resolutions are 10 d igits
in one second and 500 picoseconds, respectively. The HP 53131A provides
users with an HP-IB measuring speed of 200 measurements per second,
and is suitable for bench-top operation and lower-volume ATE operation.
The frequency and time interval resolutions for the HP 53132A are up to
12 digits in one second and 150 picoseconds, respectively. The HP 53132A
provides users with exceptional resolution, and is ideal for ATE systems
operation.
The HP 53131A/132A basic measurement functions include Frequency,
Period, Pulse Width, Duty Cycle, Rise/Fall Time, Time Interval,
Frequency Ratio, Totalize, Phase, and Peak Voltage.
The HP 53181A is a frequency counter capable of measuring frequencies
to 225 MHz on Channel 1. Depending on which optional Channel 2 the
counter contains, this capability is extended to 1.5 GHz (Option 015) or
3.0 GHz (Option 030). The HP 53181A has a frequency resolution of
10 digits in one second.
The HP 53181A provides users with an HP-IB measuring speed of
200 measurements per second, and is suitable for bench-top and ATE
operation.
The HP 53181A basic measurement functions include Frequency, Period,
Peak Voltage, and Frequency Ratio (if Channel 2 is installed).
The HP 53131A/132A/181A include additional measurement functions
and features that are designed specifically for manufacturing and service
applications:
xviAssembly-Level Service Guide
Page 19
Preface
•1, 5, 10 MHz external reference capability—to match customer’s house
standard (however, the HP 53132A’s external reference capability is
10 MHz only),
•optional ultra-high stability , high stability, or medium stability oven
oscillators for high accuracy needs,
•external gati ng,
•statistics,
•automatic limit testing,
•SCPI programming capability, and
•analog display mode limit testing
Programmable control is performed via an HP-IB. The HP-IB and a
talk-only RS-232C serial port are standard for the HP 53131A/132A/181A.
The serial port is for printing measured and analyzed data on serial
printers, or for outputting an out-of-limit signal.
Assembly-Level Service Guidexvii
Page 20
Preface
Options
The options available for the HP 53131A/132A/181A Counter are listed
following this paragraph. Specifications for the options are listed in
Chapter 7, “HP 53131A/132A Specifications,” and Chapter 8, “HP 53181A
Specifications.” If you’ve purchased an option with the initial order, it will
be installed at the factory and ready for operation at delivery. Refer to the
“Retrofitting Options” chapter in this guide for instructions on field
installation of the options.
NOTE
The “0’s” and “1’s” in the following option numbers are numeric characters
(that is, they are not letters).
Hardware
•Medium Stability Oven Timebase, Option 001
•DC Power Input, Option 00 2
•High Stability Oven Timebase, Option 010
•Ultra-High Stability Oven Timebase, Option 012 (HP 53132A and
HP 53181A Only)
•Rack Mount Kit, Option 1CM (HP part number 5063-9240)
*
For the HP 53131A/132A Option 030, front and rear terminals can exist for Channel 1 and 2. The optional
Channel 3 can be configured as rear terminals only or front terminals only. For the HP 53181A, front and rear
terminals can exist for Channel 1. The optional Channel 2 can be configured as rear terminals only or front
terminals only. Option 050 and Option 124 can be configured only as front terminals.
xviiiAssembly-Level Service Guide
Page 21
Preface
Support
•5-year Return to HP for Repair, Option W50
•5-year Return to HP for Calibration, Option W52
Accessories Supplied and Available
Accessories Supplied
•Power cord, 2.3 meters
Accessories Available
•HP 34161A Accessory Pouch
•Printer RS-232 Interface cables, HP 24542G or HP 24542H
•HP-IB cables, HP 10833A/B/C/D
Assembly-Level Service Guidexix
Page 22
Preface
xxAssembly-Level Service Guide
Page 23
1
Performance Tests
Verifying Specifications
Page 24
Chapter 1 Performance Tests
Introduction
Introduction
1
This chapter provides procedures to test the electrical performance of the
HP 53131A/132A Universal Counter and HP 53181A Frequency Counter,
using the specifications listed in Chapter 7, “HP 53131A/132A
Specifications,” and Chapter 8, “HP 53181A Specifications,” of this guide.
Three types of testing are provided:
•Operational Verification
•Complete Performance Tests
•Optional HP-IB Verification
This chapter is organized as follows:
•HP 53131A/132A HP-IB Verification Program (Optional) page 1-40
•HP 53131A/132A Performance Test Record (Page 1 of 6)page 1-41
•HP 53181A Complete Performance Testspage 1-47
•HP 53181A HP-IB Verification Program (Optional)page 1-70
•HP 53181A Performance Test Record (Page 1 of 4)page 1-71
Operational Verification
The operational verification is an abbreviated series of checks that may be
performed to give a high degree of confidence that the instrument is
operating properly without performing the complete performance tests.
An operational verification is useful for incoming inspection, routine
maintenance, and after instrument repair.
1-2Assembly-Level Service Guide
Page 25
Chapter 1 Performance Tests
Introduction
Complete Performance Tests
The complete performance tests verify the specifications listed in
Chapter 7, “HP 53131A/132A Specifications,” and Chapter 8, “HP 53181A
Specifications.” All tests can be performed without accessing the inside of
the instrument.
Optional HP-IB Verification
The HP-IB verification program, described on page 1-40
(HP 53131A/132A) or page 1-70 (HP 53181A), exercises the Counter via
the HP-IB interface. The program is written for an HP 9000 series 200 or
300 Desktop Computer as the controller. If the instrument successfully
completes all phases of the verification program, there is a very high
probability that the HP-IB interface is working properly.
The HP-IB program is available on disks as indicated in the following
listing:
•For HP 53131A/132A Counters—HP part number 53131-13501
(3 1/2-inch LIF disk) and HP part number 53131-13502 (5 1/4-inch
LIF disk).
1
•For HP 53181A Counters—HP part number 53181-13501
(3 1/2-inch LIF disk) and HP part number 53131-13502 (5 1/4-inch
LIF disk).
Recommended Calibration Cycle
The Counter requires periodic verification of operation. Depending on the
use and environmental conditions, aging, and measurement accuracy
required, the Counter should be checked using the operational verification
procedure at least once every year.
Test Record
The results of the operational verification, complete performance tests,
and HP-IB verification test should be recorded on a copy of the
Performance Test Record, located at the end of each (i.e., HP 53131A/132A
and HP 53181A) complete performance tests section in this chapter.
SynthesizerDC to 13 MHzHP 3325BOV, P, T
Synthesized Signal Generator 100 kHz to 2560 MHzHP 8663AOV, P, T
Pulse Generator5 ns pulse widthHP 8130A or equivalentOV, C
Sweep Oscillator3.0 GHzHP 8340BP, T
Primary Frequency Standard
(10 MHz House Standard)
DC Power SupplyAdjustable to 5.000VHP 6234AC
Digital VoltmeterMicrovolt accuracyHP 3458AOV, C
Time Interval CalibratorEffective edge repeatability:
Absolute accuracy >1 x 10
<10 ps
-10
HP 5071AC
HP 59992A J06C
Type N-to-BNC Connector—HP 1250-0780OV, P
BNC-to-Dual Banana PlugBNC(m) to banana plugHP 11001-60001OV, P, C T
50Ω Feedthrough (3)—HP 10100COV, P, C
50Ω Coaxial Cable with
BNC connectors (5)
Coaxial BNC Tee Connector—1250-0781P, C
P = Performance Tests
OV = Operational Verification
C = Calibration
T = Troubleshooting
BNC(m) to BNC(m), 48 inches HP 10503AOV, P, C, T
Some of this equipment may no longer be available as new. Use an equivalent model number.
1-4Assembly-Level Service Guide
Page 27
Chapter 1 Performance Tests
HP 53131A/132A/181A Operational Verification
HP 53131A/132A/181A Operational
NOTE
Verification
The operational verification is an abbreviated series of checks that may be
performed to give a high degree of confidence that the instrument is
operating properly without performing the complete performance tests.
If you are unfamiliar with the operation of the Counter, you should review
the Chapter 1, “Getting Started,” in the Operating Guide. However, the
procedures in this chapter are written so that little experience is
necessary. The procedures should be followed in the order in which they
appear.
Power-On Self Tests
1Inspect the Counter for any damage.
2Make sure no cables are connected to the input channels (rear
and front panels) of the Counter.
It is normal for the fan in the Counter to continue to run after the Counter
is placed in Standby mode. Power to the timebase is continuous to
maintain long term measurement reliability, and the fan helps maintain
timebase temperature stability.
1
NOTE
The power supply will automatically sense the input voltage; thus, there is
no need to change the ac voltage setting.
3Connect t he power cord to the Counter, a nd connect the other en d
of the power cord to the primary power source.
4Press POWER key to turn on the Counter.
Assembly-Level Service Guide1-5
Page 28
Chapter 1 Performance Tests
HP 53131A/132A/181A Operational Verification
5Verify that the following happens:
•All segments of the front-panel display light up,
•
1
SELFTST: PASS
•
HP-IB AT n
is displayed,
is quickly displayed—where “n” represents the HP-IB
address (Note that some earlier versions of the Counter’s firmware di d
not display the HP-IB address at power-up.), and
•dashes are displayed.
The Counter is now ready to measure frequency of a signal applied to
CHANNEL 1 input as indicated by illumination of the Freq and Ch1
annunciators.
If a fail message is displayed, refer to the troubleshooting section in
Chapter 2, “Service,” of this guide.
6For HP 53131A/132A, mark Pass or Fail in the HP 53131A/132A
Performance Test Record on page 1-41, Test 1.
6’For HP 53181A, mark Pass or Fail in the HP 53181A Performance
Test Record on page 1-71, Test 1.
7Connect the Counter’s rear-panel 10 MHz Out as shown in
Figure 1-1A or Figure 1-1B, depending on which Counter you are
testing.
HP 53131A/132A
Counter
Channel 1
10 MHz
Out
(rear panel)
Channel 2
BNC Tee
Figure 1-1A. Operational Verification Setup for HP 53131A/132A
1-6Assembly-Level Service Guide
Page 29
Chapter 1 Performance Tests
HP 53131A/132A/181A Operational Verification
HP 53181A
Counter
10 MHz
Out
(rear panel)
Channel 1
Figure 1-1B. Operational Verification Setup for HP 53181A
8Depending on which Counter you are testing, verify the
appropriate reading as shown below:
For HP 53131A and HP 53181A, verify a reading of 10.000,000,00 MHz.
For the HP 53132A, verify a reading of 10.000,000,000 MHz.
9For HP 53131A/132A, mark Pass or Fail in the HP 53131A/132A
Performance Test Record on page 1-41, Test 2.
9’For HP 53181A, mark Pass or Fail in the HP 53181A Performance
Test Record on page 1-71, Test 2.
1
NOTE
The remaining steps apply to the HP 53131A and HP 53132A only
since the HP 53181A does not have a standard Channel 2 input.
For the HP 53181A, proceed to the following “Run Self Test” section.
10Press Freq & Ratio key until
FREQUENCY 2
is displayed to select
Channel 2 for frequency measurements.
For the HP 53131A, verify a reading of 10.000,000,00 MHz.
For the HP 53132A, verify a reading of 10.000,000,000 MHz.
11For HP 53131A/132A, mark Pass or Fail in the HP 53131A/132A
Performance Test Record on page 1-41, Test 3.
Assembly-Level Service Guide1-7
Page 30
Chapter 1 Performance Tests
HP 53131A/132A/181A Operational Verification
Run Self Test
1Disconnect the input signals from the Counter.
1
2Power down (turn off the Counter), press and hold Recall (Utility)
key, then press POWER key.
3Press Recall key until
TEST: ALL ?
is displayed.
4Press Enter key.
Observe that
If
ALL: FAIL
ALL: PASS
is displayed or a failure is indicated, refer to the
is displayed after the self test is completed.
troubleshooting section in Chapter 2, “Service,” of this guide.
5For HP 53131A/132A, mark Pass or Fail in the HP 53131A/132A
Performance Test Record on page 1-41, Test 4.
5’For HP 53181A, mark Pass or Fail in the HP 53181A Performance
Test Record on page 1-71, Test 3.
Termination Check
1Cycle the POWER key to preset the Counter.
2Connect a digital voltmeter (DVM) to CHANNEL 1 as shown in
Figure 1-2, and set it to measure ohms (set DVM to appropriate
Ω
range).
Counter
HP 53131A/132A is shown in abo ve figure. This setup fo r the HP53181A requires a connection to Channel 1 only.
HP 3458A
Digital Multimeter
BNC - to - Dual
Banana Plug
Figure 1-2. Termination Check Setup
1-8Assembly-Level Service Guide
Page 31
Chapter 1 Performance Tests
HP 53131A/132A/181A Operational Verification
3Press Channel 1 50Ω/1MΩ key.
The LED turns on, indicating a 50Ω input impedance for Channel 1.
4Press Channel 1 DC/AC key.
The LED turns on, indicating DC input coupling for Channel 1.
5Verify the DVM reads 51.1Ω ± 3Ω.
6For HP 53131A/132A, mark Pass or Fail in the HP 53131A/132A
Performance Test Record on page 1-41, Test 5a.
6’For HP 53181A, mark Pass or Fail in the HP 53181A Performance
Test Record on page 1-71, Test 4a.
7Press Channel 1 50Ω/1MΩ key.
The LED turns off, indicating a 1 MΩ input impedance for Channel 1.
1
8Verify the DVM reads approximately 1.0000000 MOHM.
9For HP 53131A/132A, mark Pass or Fail in the HP 53131A/132A
Performance Test Record on page 1-41, Test 5b.
9’For HP 53181A, mark Pass or Fail in the HP 53181A Performance
Test Record on page 1-71, Test 4b.
10Repeat steps 3 through 9 for Channel 2 for the HP 53131A/132A
Counter ONLY since the HP 53181A Counter does not have a
standard Channel 2.
11For HP 53131A/132A, mark Pass or Fail in the HP 53131A/132A
Performance Test Record on page 1-41, Tests 6a and 6b.
12Disconnect the test setup.
Assembly-Level Service Guide1-9
Page 32
Chapter 1 Performance Tests
HP 53131A/132A/181A Operational Verification
External Arm Test (HP 53131A/132A Only)
This test verifies the External Arm port of the HP 53131A and HP 53132A
Counter is operational.
1
Equipment
HP 8663A Synthesized Signal Generator
HP 8130A Pulse Generator (or equivalent)
HP 10100C 50Ω Feedthrough
HP 10503A BNC Cables (2)
HP 1250-0780 N-to-BNC Connector
Counter Setup
1Cycle the POWER key to preset the Counter.
2Set the Counter as follows:
Other Meas,
Gate & ExtArm,
3Set CHANNEL 1 input conditions to:
Trigger/Sensitivity,
Trigger/Sensitivity,
50Ω/1MΩ, 50
DC/AC, DC
X10 Attenuator, Off
100kHz Filter, Off
Figure 1-3. External Arm Test Setup (HP 53131A/132A Only)
Assembly-Level Service Guide1-11
Page 34
Chapter 1 Performance Tests
HP 53131A/132A/181A Operational Verification
3Set the HP 8663A to output a 200 MHz, 100 mV sine wave signal to
Channel 1.
1
4On the Counter, Press Run key.
5Send a single pulse by pressing the MAN key on the HP 8130A, and
observe that with each press the
The Counter should display approximately 1000.
6For HP 53131A/132A, mark Pass or Fail in the HP 53131A/132A
Performance Test Record on page 1-41, Test 7.
7Disconnect the test setup.
Gate
annunciator flashes.
External Timebase Tests
These tests verify the External Timebase 1, 5, 10 MHz specifications of
the HP 53131A and HP 53181A Counters by checking that TIMEBASE
AUTO allows the Counter to be synchronized to the external reference.
Note that the HP 53132A only operates with the 10 MHz external
timebase; therefore, perform the procedure in the section titled
“10 MHz External Timebase Input for the HP 53132A” on page 1-15
to verify that TIMEBASE AUTO allows the Counter to be synchronized to
the external reference.
Equipment
HP 3325B Synthesizers (2)
HP 10100C 50W Feedthrough
HP 10503 BNC Cables (2)
1-12Assembly-Level Service Guide
Page 35
Chapter 1 Performance Tests
HP 53131A/132A/181A Operational Verification
1 MHz External Timebase Input (HP 53131A and HP 53181A
Only)
1Connect the equipment as shown in Figure 1-4.
HP 3325B
Synthesizer #1
50Ω
Feedthrough
Ref. In
(rear panel)
HP 3325B
Synthesizer #2
Counter
Channel 1
Output
Output
Figure 1-4. External Timebase Test Setup
2Cycle the POWER key to preset the Counter.
3Set Synthesizer #1 (i.e., the HP 3325B connected to the rear-panel
Ref In input of the Counter via a 50Ω feedthrough) to output a
1 MHz, 200 mV rms sine wave signal.
1
4Set Synthesizer #2 (i.e., the HP 3325B connected to Channel 1) to
output a 2 MHz, 100 mV rms square wave signal.
Verify the Counter displays a reading of approximately 2 MHz and the
ExtRef annunciator is lit.
5For HP 53131A, mark Pass or Fail in the HP 53131A/132A
Performance Test Record on page 1-41, Test 8a.
Assembly-Level Service Guide1-13
Page 36
Chapter 1 Performance Tests
HP 53131A/132A/181A Operational Verification
5’For HP 53181A, mark Pass or Fail in the HP 53181A Performance
Test Record on page 1-71, Test 5a.
1
Only)
1Change the frequency of Synthesizer # 1, which is connected to
the Counter’s rear-panel Ref In input to 5 MHz.
2On the Counter, press Run key.
Verify the Counter displays a reading of approximately 2 MHz and the
ExtRef annunciator is lit.
3For HP 53131A, mark Pass or Fail in the HP 53131A/132A
Performance Test Record on page 1-41, Test 8b.
3’For HP 53181A, mark Pass or Fail in the HP 53181A Performance
Test Record on page 1-71, Test 5b.
10 MHz External Timebase Input (HP 53131A and HP 53181A
Only)
1Change the frequency of Synthesizer #1, which is connected to the
Counter’s rear-panel Ref In input to 10 MHz.
2On the Counter, press Run key.
5 MHz External Timebase Input (HP 53131A and HP 53181A
Verify the Counter displays a reading of approximately 2 MHz and the
ExtRef annunciator is lit.
3For HP 53131A, mark Pass or Fail in the HP 53131A/132A
Performance Test Record on page 1-41, Test 8c.
3’For HP 53181A, mark Pass or Fail in the HP 53181A Performance
Test Record on page 1-71, Test 5c.
4Disconnect the test setup.
This completes the Operational Verification for the HP 53131A and
HP 53181A.
1-14Assembly-Level Service Guide
Page 37
Chapter 1 Performance Tests
HP 53131A/132A/181A Operational Verification
10 MHz External Timebase Input for the HP 53132A
1Connect the equipment as shown in Figure 1-4.
2Set Synthesizer #1 (i.e., the HP 3325B that is connected to the
rear-panel Ref In input of the Counter via a 50Ω feedthrough) to
output a 10 MHz, 200 mV rms sine wave signal.
3Set Synthesizer #2 (i.e., the HP 3325B that is connected to
Channel 1) to output a 2 MHz, 100 mV rms square wave signal.
Verify the Counter displays a reading of approximately 2 MHz and the
ExtRef annunciator is lit.
4Mark Pass or Fail in the HP 53131A/132A Performance Test
Record on page 1-41, Test 8d.
5Disconnect the test setup.
This completes the Operational Verification for the HP 53132A.
1
Assembly-Level Service Guide1-15
Page 38
Chapter 1 Performance Tests
HP 53131A/132A Complete Performance Tests
HP 53131A/132A Complete Performance
1
Tests
The specifications of the HP 53131A/132A Universal Counter can be
verified by performing the performance tests provided in this section.
Table 1-2 lists a summary of the HP 53131A/132A performance tests.
Record the results of the performance tests in the appropriate
place on HP 53131A/132A Performance Test Record, which starts
on page 1-41.
(Note: the complete performance tests for the HP 531 81A Frequency
Counter begin on page 1-47 of this chapter.)
NOTE
To perform valid testing of the specifications, warm up the Counter and
test equipment for 30 minutes.
Table 1-2. The HP 53131A/132A Performance Tests
Page Number*Test Description
page 1-17Test 1: Time Interval (HP 53131A/132A Only)
page 1-19Test 2: Trigger Level (HP 53131A/132A Only)
page 1-22Test 3: Channels 1 and 2 Frequency Sensitivity
(HP 53131A/132A Only)
page 1-27Test 4: Channels 1 and 2 Frequency Accuracy
(HP 53131A/132A Only)
page 1-30Test 5: Option 030/050/124 Channel 3 Frequency Sensitivity
(HP 53131A/132A Only)
page 1-35Test 6: Option 030/050/124 Channel 3 Frequency Accuracy
* Other Counter measurement functions (e.g., Period) are mathematically derived by the microprocessor from
the parameters verified by these performance tests. If the Counter passes the performance tests, the other
measurement functions are functioning to specifications.
1-16Assembly-Level Service Guide
Page 39
Chapter 1 Performance Tests
Test 1: Time Interval (HP 53131A/132A Only)
Test 1: Time Interval (HP 53131A/132A Only)
This test verifies the Time Interval specifications between Channels 1 and 2.
Equipment
HP 8130A Pulse Generator (or equivalent)
HP 10503A BNC Cable
Counter Setup
1Cycle the POWER key to preset the Counter.
2Press Time & Period key until
3Set CHANNEL 1 input conditions to:
Trigger/Sensitivity,
Trigger/Sensitivity,
Trigger/Sensitivity,
50Ω/1MΩ, 50
DC/AC, DC
X10 Attenuate, Off
100kHz Filter, Off
4Set CHANNEL 2 input conditions to:
Ω
AUTO TRG: OFF
LEVEL: .500V (Enter)
COMMON 1: ON
TI 1 TO 2
is displayed.
1
Trigger/Sensitivity,
Trigger/Sensitivity,
Trigger/Sensitivity,
50Ω/1MΩ, 50
DC/AC, DC
X10 Attenuate, Off
100kHz Filter, Off
Assembly-Level Service Guide1-17
Ω
AUTO TRG: OFF
LEVEL: .500V (Enter)
SLOPE: NEG
Page 40
Chapter 1 Performance Tests
Test 1: Time Interval (HP 53131A/132A Only)
Procedure
1Connect the pulse generator si gnal to Channel 1 of the C ounter as
shown in Figure 1-5.
1
HP 8130A
HP53131A/132
Counter
Pulse Generator
Channel 1
Figure 1-5. Time Interval (Channel 1 TO 2) Test Setup
2Set the Pulse Generator to the following:
PERIOD: 10 µs
WIDTH: 150 ns
High: 1.00 VLow : 0.00 V
Input Mode: Normal
3On the Counter, Press Single/Stop key.
The HP 53131A Counter should display 0.1500 µs ±0.0040 µs.
The HP 53132A Counter should display 0.1500 µs ±0.0025 µs.
4Record this value in the appropriate place in the Performance
Test Record (Test 1, Line 1 for HP 53131A or Test 1, Line 2 for
HP 53132A).
5Disconnect the test setup.
Output
1-18Assembly-Level Service Guide
Page 41
Chapter 1 Performance Tests
Test 2: Trigger Level (HP 53131A/132A Only)
Test 2: Trigger Level (HP 53131A/132A Only)
This test verifies the Trigger Level accuracy of the HP 53131A and
HP 53132A Universal Counters.
Equipment
HP 3325B Synthesizer
HP 10503A BNC Cable
Counter Setup
1Cycle the POWER key to preset the Counter.
2Set CHANNEL 1 input conditions to:
Trigger/Sensitivity,
Trigger/Sensitivity,
50Ω/1MΩ, 50
DC/AC, DC
X10 Attenuate, Off
100kHz Filter, Off
3Set CHANNEL 2 input conditions to:
Trigger/Sensitivity,
Trigger/Sensitivity,
50Ω/1MΩ, 50
DC/AC, DC
X10 Attenuate, Off
100kHz Filter, Off
Ω
Ω
AUTO TRG: OFF
LEVEL: 0.000V
AUTO TRG: OFF
LEVEL: 0.000V
1
Assembly-Level Service Guide1-19
Page 42
Chapter 1 Performance Tests
Test 2: Trigger Level (HP 53131A/132A Only)
Procedure
1Connect the output of the HP 3325B to Channel 1 of the Counter
as shown in Figure 1-6.
1
HP 3325B
HP 53131A/132A
Counter
Synthesizer
Channel 1
Channel 2
Output
Figure 1-6. Trigger Level Test Setup
2Set the HP 3325B to output a 1 MHz, 80 mVp-p s quare wave signal.
3Set the HP 3325B DC OFFSET to −60 mV.
4On the Counter, press Run key.
5Now, increment the dc offset on the HP 3325B by +1 mV until the
Counter’s
Gate
annunciator flashes and continue incrementing
until the Counter displays a stable reading of approximately
1MHz.
Observe the offset value on the display of the HP 3325B.
Record the dc offset value__________ mV.
6Add the upper peak voltage (40 mV) of the 80 mV p-p signal to the
offset value in step 5 (For example, −24 mV + 40 mV = 16 mV).
This is the upper hysteresis level.
Record the upper hysteresis l evel__________ mV.
1-20Assembly-Level Service Guide
Page 43
Chapter 1 Performance Tests
Test 2: Trigger Level (HP 53131A/132A Only)
7In the Counter’s Channel 1 Trigger/Sensitivity menu, change the
SLOPE to NEG.
8Set the HP 3325B DC OFFSET to +60 mV.
9On the Counter, press Run key.
10Now, decrement the offset on the HP 3325B by −1 mV until the
Counter’s
until the Counter displays a stable reading of approximately
1MHz.
Observe the offset value on the display of the HP 3325B.
Record the value__________ mV.
11Add the lower peak voltage (−40mV) of the 80 mVp-p signal to the
offset value in step 10 (for example, 21 mV − 40 mV = −19 mv.)
This result is the lower hysteresis level.
Record the lower hysteresis level __________ mV.
12Now, add the upper hysteresis value (recorded in step 6) and
lower hysteresis value (recorded in step 11).
This is the trigger level; it should be 0.0 V ±15 mV.
Gate
annunciator flashes and continue decrementing
1
13Record the trigger level value in the appropriate place in the
Performance Test Record (Test 2, Line 1).
14Repeat steps 1 through 13 for Channel 2.
15Record the trigger level value in the appropriate place in the
Performance Test Record (Test 2, Line 2).
16Disconnect the test setup.
Assembly-Level Service Guide1-21
Page 44
Chapter 1 Performance Tests
Test 3: Channels 1 and 2 Frequency Sensitivity (HP 53131A/132A Only)
Test 3: Channels 1 and 2 Frequency
1
Sensitivity (HP 53131A/132A Only)
This set of tests verifies frequency sensitivity specifications of the
HP 53131A and 53132A Universal Counters.
Equipment
HP 8663A Synthesized Signal Generator
HP 1250-0780 Type N-to-BNC Connector
HP 10100C 50Ω Feedthrough
HP 10503A BNC Cables (2)
Counter Setup
1Cycle the POWER key to preset the Counter.
2Set CHANNEL 1 input conditions to:
Trigger/Sensitivity,
Trigger/Sensitivity,
50Ω/1MΩ, 50
DC/AC, DC
X10 Attenuate, Off
100kHz Filter, Off
Ω
AUTO TRG: OFF
LEVEL: 0.000 V
3Set CHANNEL 2 input conditions to:
Trigger/Sensitivity,
Trigger/Sensitivity,
50Ω/1MΩ, 50
DC/AC, DC
X10 Attenuate, Off
100kHz Filter, Off
1-22Assembly-Level Service Guide
Ω
AUTO TRG: OFF
LEVEL: 0.000 V
Page 45
Chapter 1 Performance Tests
Test 3: Channels 1 and 2 Frequency Sensitivity (HP 53131A/132A Only)
100 kHz to 100 MHz Sensitivity for HP 53131A/132A
1Connect the timebase output of the unit with the better timebase
to the timebase reference input of the other unit as shown in
Figure 1-7.
For this test, make sure you always use the unit (HP 53131A/132A or
HP 8663A) that contains the better 10 MHz timebase as the output source
as shown in Figure 1-7.
HP 8663A
HP 53131A/132A
Counter
50Ω
Feedthrough
Ref. In
(rear panel)
Synthesized
Signal Generator
Output
(rear panel)
1
N-to-BNC
Channel 1
Output
Connector
OR
HP 8663A
Synthesized
Signal Generator
HP 53131A/132A
Counter
Channel 1
50Ω
Feedthrough
10 MHz
Output
(rear panel)
Figure 1-7. Channels 1 and 2 Frequency Sensitivity Test Setup
Assembly-Level Service Guide1-23
Input
(rear panel)
Output
N-to-BNC
Connector
Page 46
Chapter 1 Performance Tests
Test 3: Channels 1 and 2 Frequency Sensitivity (HP 53131A/132A Only)
2Set the HP 8663A Signal Generator to output a 100 MHz,
−
35.0 dBm sine wave signal (−20.0 dBm for Option 060 Rear
Terminals) to the Counter’s Channel 1 input.
1
3On the Counter, press Run key.
4In 0.1 dB steps, increase the power level until the Counter
displays a stable reading of 100 MHz.
The Counter should display 100 MHz when the signal generator level is
≤−
21.0 dBm (20 mVrms). Counters with Option 060 should display
100 MHz when the signal generator level is
5Sweep the frequency from 100 MHz to 100 kHz. For Option 060,
sweep frequencies from 225 MHz to 100 kHz.
The Counter should read frequencies from 100 MHz to 100 kHz at an
input power level of
Option 060 should read frequencies from 225 MHz t o 100 kH z at
9.5 dBm (75 mVrms).
6Record the actual reading in the appropriate place in the
Performance Test Record (Test 3, Line 1).
If you are testing a Counter with front-panel terminals (Standard),
record the reading on Line 1 of the Standard portion of the test record.
If you are testing a Counter with rear terminals (Option 060),
record the reading on Line 1 of the Option 060 portion of the test record.
≤ −
21.0 dBm (20 mVrms). Count ers with
≤ −
9.5 dBm (75 mVrms).
≤−
NOTE
DO NOT test the front terminals if rear terminals are installed. The front
terminal performance is not specified when the rear terminals are installed.
7Connect the HP 8663A to Channel 2 of the Counter.
8On the Counter, press Freq & Ratio key until
displayed.
9Repeat steps 2 through 5 for Channel 2.
1-24Assembly-Level Service Guide
FREQUENCY 2
is
Page 47
Chapter 1 Performance Tests
Test 3: Channels 1 and 2 Frequency Sensitivity (HP 53131A/132A Only)
10Record the actual reading in the appropriate place in the
Performance Test Record (Test 3, Line 2).
NOTE
If you are testing a Counter with front-panel terminals (Standard),
record the reading on Line 2 of the Standard portion of the test record.
If you are testing a Counter with rear terminals (Option 060),
record the reading on Line 2 of the Option 060 portion of the test record.
The remaining sensitivity procedures are NOT required for Counters that
contain Option 060 Rear Terminals; thus, this completes the sensitivity
portion of the Frequency Performance Test. Go to the Frequen cy Accuracy
Test on page 1-27.
100 MHz to 200 MHz Sensitivity for HP 53131A/132A
1Leave the signal generator connected to Channel 2. Change the
signal generator settings to 200 MHz at −30 dBm.
2On the Counter, press Run key.
3Increase the power level by 0.1 dB steps until the Counter
displays a stable 200 MHz reading.
The Counter should display 200 MHz when the input signal level is
≤−
17.5 dBm (30 mVrms).
1
4Sweep the signal generator frequency from 200 MHz to 100 MHz.
The Counter should read frequencies from 200 MHz to 100 MHz when the
input signal level is
5Record the actual reading in the Performance Test Record
(Test 3, Line 3).
6Connect the signal generator to Channel 1.
7On the Counter, press Freq & Ratio key until
displayed.
Assembly-Level Service Guide1-25
≤ −
17.5 dBm (30 mVrms).
Frequency 1
is
Page 48
Chapter 1 Performance Tests
Test 3: Channels 1 and 2 Frequency Sensitivity (HP 53131A/132A Only)
8Repeat steps 1 through 4 for Channel 1.
9Record the actual reading in the Performance Test Record
(Test 3, Line 4).
1
200 MHz to 225 MHz Sensitivity for HP 53131A/132A
1Leave the signal generator connected to Channel 1. Change the
signal generator settings to 225 MHz at −30 dBm.
2On the Counter, press Run key.
3Increase the power level by 0.1 dB steps until the Counter
displays a stable reading of 225 MHz.
The Counter should read 225 MHz when the input signal level is
≤ −
15.0 dBm (40 mVrms).
4Sweep the signal generator frequency from 225 MHz to 200 MHz.
The Counter should read frequencies from 225 MHz to 200 MHz when the
input signal level is
5Record the actual reading in the Performance Test Record
(Test 3, Line 5).
6Connect the signal generator to Channel 2.
7On the Counter, press Freq & Ratio key until
displayed.
8Repeat steps 1 through 4 for Channel 2.
9Record the actual reading in the Performance Test Record
(Test 3, Line 6).
10Disconnect the test setup.
1-26Assembly-Level Service Guide
≤ −
15.0 dBm (40 mVrms).
Frequency 2
is
Page 49
Chapter 1 Performance Tests
Test 4: Channels 1 and 2 Frequency Accuracy (HP 53131A/132A Only)
Test 4: Channels 1 and 2 Frequency Accuracy
(HP 53131A/132A Only)
This set of tests verifies the frequency accuracy specifications of the
HP 53131A and HP 53132A Universal Counters.
Equipment
HP 8663A Synthesized Signal Generator
HP 1250-0780 Type N-to-BNC Connector
HP 10100C 50Ω Feedthrough
HP 10503A BNC Cables (2)
Counter Setup
1Cycle the POWER key to preset the Counter.
2Using the Gate & ExtArm key, set the gate time to
3Set CHANNEL 1 input conditions to:
Trigger/Sensitivity,
Trigger/Sensitivity,
50Ω/1MΩ, 50
DC/AC, DC
X10 Attenuate, Off
100kHz Filter, Off
Ω
AUTO TRG: OFF
LEVEL: 0.000 V
TIME 1.000 s
1
.
4Set CHANNEL 2 input conditions to:
Trigger/Sensitivity,
Trigger/Sensitivity,
50Ω/1MΩ, 50
DC/AC, DC
X10 Attenuate, Off
100kHz Filter, Off
Assembly-Level Service Guide1-27
Ω
AUTO TRG: OFF
LEVEL: 0.000 V
Page 50
1
HP 53131A/132A
Counter
Chapter 1 Performance Tests
Test 4: Channels 1 and 2 Frequency Accuracy (HP 53131A/132A Only)
Procedure
1Connect the equipment as shown in Figure 1-8 with the signal
generator connected to Channel 1 of the Counter.
50Ω
Feedthrough
Ref. In
(rear panel)
HP 8663A
Synthesized Signal Generator
Channel 2
(rear panel)
Input
Channel 1
Figure 1-8. Channels 1 and 2 Frequency Accuracy Test Setup
2Set the Signal Generator to output 1 MHz at −7 dBm.
3On the Counter, press Run key.
Verify the HP 53131A Counter reads 1 MHz ±1 mHz.
Verify the HP 53132A Counter reads 1 MHz ±200 µHz.
4Record the actual reading in the appropriate place in the
Performance Test Record (Test 4, Line 1 for HP 53131A or Test 4,
Line 3 for HP 53132A.
1-28Assembly-Level Service Guide
Output
N-to-BNC
Connector
Page 51
Chapter 1 Performance Tests
Test 4: Channels 1 and 2 Frequency Accuracy (HP 53131A/132A Only)
5Change the signal generator frequency to 100 MHz.
Verify the HP 53131A Counter reads 100 MHz ±70 mHz.
Verify the HP 53132A Counter reads 100 MHz ±15 mHz.
6Record the actual reading in the appropriate place in the
Performance Test Record (Test 4, Line 5 for HP 53131A or Test 4,
Line 7 for HP 53132A).
7Change the signal generator frequency to 200 MHz.
Verify the HP 53131A Counter reads 200 MHz ±130 mHz.
Verify the HP 53132A Counter reads 200 MHz ±30 mHz.
8Record the actual reading in the appropriate place in
Performance Test Record (Test 4, Line 9 for HP 53131A or Test 4,
Line 11 for HP 53132A).
1
9Change the signal generator frequency to 225 MHz.
Verify the HP 53131A Counter reads 225 MHz ±150 mHz.
Verify the HP 53132A Counter reads 225 MHz ±35 mHz.
10Record the actual reading in the appropriate place in the
Performance Test Record (Test 4, Line 13 for HP 53131A or Test 4,
Line 15 for HP 53132A).
11Connect the signal generator to Channel 2.
12On the Counter, press Freq & Ratio key until
displayed.
13Repeat steps 2 through 9 for Channel 2.
14Record the actual readings in the appropriate places in the
Performance Test Record.
15Disconnect the test setup.
Assembly-Level Service Guide1-29
Frequency 1
is
Page 52
Chapter 1 Performance Tests
Test 5: Option 030/050/124 Channel 3 Frequency Sensitivity (HP
53131A/132A Only)
Test 5: Option 030/050/124 Channel 3
1
Only)
This test verifies the frequency range, and sensitivity of the optional
3-GHz and 5-GHz Channel 3 (Option 030/050/124) for the HP 53131A and
the HP 53132A Universal Counters.
Equipment
HP 8663A Synthesized Signal Generator (not needed for Option 050)
HP 8340B Sweep Oscillator
HP 1250-0780 Type N-to-BNC Connector (not needed for Option 050)
HP 10503A BNC Cable(2)
HP 10100C 50Ω Feedthrough
HP 11500D N-N Cable (Option 050 only)
HP 1250-1250 N(M) to SMA(F)
HP 1250-0777 N(F) to N(F)
Counter Setup
1Cycle the POWER key to preset the Counter.
2Set the Counter as follows:
Frequency Sensitivity (HP 53131A/132A
Freq & Ratio,
FREQUENCY 3
1-30Assembly-Level Service Guide
Page 53
Chapter 1 Performance Tests
Test 5: Option 030/050/124 Channel 3 Frequency Sensitivity (HP
53131A/132A Only)
100 MHz to 2.5 GHz Sensitivity for HP 53131A/132A
Option 030 only
HP 53131A/132A
Counter
1Set the HP 8663B Synthesized Signal Generator to output a
2500 MHz, −40 dBm sine wave, and connect the signal to Channel 3
of the Counter as shown in Figure 1-9.
Frequency Standard
(switch set to EXT)
Output
Input
(rear panel)
50Ω
Feedthrough
Ref. In
(rear panel)
HP 8340B
Sweep Oscillator
HP 8663A
Synthesized Signal Generator
1
Channel 3
Output
N-to-BNC
Connector
Figure 1-9. Option 030/050/124 Channel 3 Frequency Sensitivity and Accuracy Test Setup
2On the Counter, Press Run key.
Assembly-Level Service Guide1-31
Page 54
Chapter 1 Performance Tests
Test 5: Option 030/050/124 Channel 3 Frequency Sensitivity (HP
53131A/132A Only)
3Increase the power level in 5 d B steps until the Counter displays a
stable reading of approximately 2.5 GHz.
The Counter should display 2.5 GHz at
≤ −
27.0 dBm (10 mVrms).
1
4Sweep the frequencies from 2.5 GHz to 200 MHz.
The Counter should read frequencies from 2.5 GHz to 200 MHz at
≤ −
27.0 dBm (10 mVrms).
5Record the actual reading in the Performance Test Record
(Test 5, Line 1).
2.7 to 3.0 GHz Sensitivity for Option 030 only
1Disconnect the HP 8663A Synthesized Signal Generator from the
Counter (see Figure 1-9), and connect the HP 8340B Sweep
Oscillator to Channel 3 of the Counter.
2Change the settings of the Sweep Oscillator to 3.0 GHz, −40 dBm.
3Increase the power level in 1 d B steps until the Counter displays a
stable reading of approximately 3.0 GHz.
The Counter should display 3.0 GHz at
4Sweep the frequencies from 3.0 GHz to 2.7 GHz.
≤ −
21.0 dBm.
The Counter should read frequencies from 3.0 to 2.7 GHz at
5Record the actual reading in the Performance Test Record
(Test 5, Line 2).
6Disconnect the test setup.
1-32Assembly-Level Service Guide
≤ −
21.0 dBm.
Page 55
Chapter 1 Performance Tests
Test 5: Option 030/050/124 Channel 3 Frequency Sensitivity (HP
53131A/132A Only)
200 MHz to 5.0 GHz Sensitivity for HP 53131A/132A
(Option 050 only)
NOTE
1Connect the 8340B Sweep Oscillator to Channel 3 of the counter.
2Change the settings of the Sweep Oscillator to 5.0 GHz, −40 dBm.
3Increase the power level in 1 d B steps until the Counter displays a
stable reading of approximately 5.0 GHz.
The Counter should display 5.0 GHz at
Use a power meter and sensor to ensure the accuracy of the power being
input.
4Sweep the frequencies from 200 MHz to 5.0 GHz.
The Counter should read frequencies from 200 MHz to 5.0 GHz at ≤−3dBm.
5Record the actual reading in the Performance Test Record
(Test 5, Line 3).
6Disconnect the test setup.
≤ −
23 dBm.
1
Assembly-Level Service Guide1-33
Page 56
Chapter 1 Performance Tests
Test 5: Option 030/050/124 Channel 3 Frequency Sensitivity (HP
53131A/132A Only)
200 MHz to 12.4 GHz Sensitivity for HP 53131A/132A
(Option 124 only)
1Connect the 8340B Sweep Oscillator to Channel 3 of the counter.
1
2Change the settings of the Sweep Oscillator to 12.4 GHz, −40 dBm.
3Increase the power level in 1 d B steps until the Counter displays a
stable reading of approximately 12.4 GHz.
NOTE
The Counter should display 12.4 GHz at
Use a power meter and sensor to ensure the accuracy of the power being
input.
4Sweep the frequencies from 200 MHz to 12.4 GHz.
The Counter should read frequencies from 200 MHz to 12.4 GHz
at ≤−3dBm.
5Record the actual reading in the Performance Test Record
(Test 5, Line 4).
6Disconnect the test setup.
≤ −
23 dBm.
1-34Assembly-Level Service Guide
Page 57
Chapter 1 Performance Tests
Test 6: Option 030/050/124 Channel 3 Frequency Accuracy
(HP 53131A/132A Only)
Test 6: Option 030/050/124 Channel 3
Frequency Accuracy (HP 53131A/132A Only)
This test verifies the frequency accuracy of the optional 3-GHz or 5-GHz
Channel 3 (Option 030).
Equipment
HP 8663A Synthesized Signal Generator (Option 030 only)
HP 8340B Sweep Oscillator
HP 1250-0780 Type N-to-BNC Connector (Option 030 only)
HP 10503A BNC Cable (2)
HP 10100C 50Ω Feedthrough
HP 11500D N-N Cable (Option 050 only)
HP 1250-1250 N(M) to SMA(F)
HP 1250-0777 N(F) to N(F)
Counter Setup
1Cycle the POWER key to preset the Counter.
2Set the Counter as follows:
Freq & Ratio,
Gate & ExtArm,
FREQUENCY 3
TIME: TIME 1.000 s
1
Procedure
1Connect equipment as shown in Figure 1-9.
(NOTE: For Options 050 and 124, skip steps 2 through 5.
Proceed to step 6.)
2Set the HP 8663A Synthesized Signal Generator to output
100 MHz at −17.0 dBm.
Assembly-Level Service Guide1-35
Page 58
Chapter 1 Performance Tests
Test 6: Option 030/050/124 Channel 3 Frequency Accuracy
(HP 53131A/132A Only)
3On the Counter, press Run key.
Verify the HP 53131A Counter reads 100 MHz ±70 mHz.
1
Verify the HP 53132A Counter reads 100 MHz ±20 mHz.
4Record the actual reading in the Performance Test Record
(Test 6, Line 1 for HP 53131A or Test 6, Line 2 for HP 53132A).
5Disconnect the HP 8663A output from Channel 3 of the Counter,
and connect the HP 8340B Sweep Oscillator output to Channel 3
of the Counter as shown in Figure 1-9.
6Set the HP 8340B to output 3.0 GHz at −17.0 dBm.
Verify the HP 53131A Counter reads 3.0 GHz ±2 Hz.
Verify the HP 53132A Counter reads 3.0 GHz ±0.4 Hz.
7Record the actual reading in the Performance Test Record
(Test 6, Line 3 for HP 53131A or Test 6, Line 4 for HP 53132A).
8Set the HP 8340B to output 5.0 GHz at −17 dBm.
Verify the HP 53131A Counter reads 5.0 GHz ±3 Hz.
Verify the HP 53132A Counter reads 5.0 GHz ±0.7 Hz.
9Record the actual reading in the Performance Test Record
(Test 6, Line 5 for HP 53131A or Test 6, Line 6 for HP 53132A).
10Set the HP 8340B to output 12.4 GHz at −17 dBm. (This step is for
Option 124 only.)
Verify the HP 53131A Counter reads 12.4 GHz ±8 Hz.
Verify the HP 53132A Counter reads 12.4 GHz ±2 Hz.
11Record the actual reading in the Performance Test Record
(Test 6, Line 7 for HP 53131A or Test 6, Line 8 for HP 53132A.)
12Disconnect the test setup.
1-36Assembly-Level Service Guide
Page 59
Chapter 1 Performance Tests
Test 7: Peak Volts, Channels 1 and 2 (HP 53131A/132A Only)
Test 7: Peak Volts, Channels 1 and 2
(HP 53131A/132A Only)
This test verifies the Peak Volts accuracy specification of the HP 53131A
and HP 53132A Universal Counters.
Equipment
HP 3325B Synthesizer
HP 10100C 50Ω Feedthrough
HP 10503A BNC Cable
Counter Setup
1Cycle the POWER key to preset the Counter.
2Press Other Meas key until
3Set CHANNEL 1 input to:
50Ω/1MΩ, 50
X10 Attenuate, Off
100kHz Filter, Off
4Set CHANNEL 2 input to:
Ω
VOLT PEAKS 1
is displayed.
1
50Ω/1MΩ, 50
X10 Attenuate, Off
100kHz Filter, Off
Assembly-Level Service Guide1-37
Ω
Page 60
Chapter 1 Performance Tests
Test 7: Peak Volts, Channels 1 and 2 (HP 53131A/132A Only)
Volt Peak 1
1Connect the equipment as shown in Figure 1-10.
1
HP 3325B
HP 53131A/132A
Counter
Channel 2
Synthesizer
Channel 1
Figure 1-10. Peak Volts Test Setup
2Set the HP 3325B to output a 2 MHz, 2 Vp-p sine wave.
The Counter should display −1.00 ±0.12V for the negative peak of the sine
wave. This reading is displayed on the left side of the display.
Record the actual reading in the Performance Test Record
(Test 7, Line 1a).
Also, the Counter should display +1.00 ±0.12V for the positive peak of the
sine wave. This reading is displayed on the right side of the display.
Record the actual reading in the Performance Test Record
(Test 7, Line 1b).
Output
1-38Assembly-Level Service Guide
Page 61
Chapter 1 Performance Tests
Test 7: Peak Volts, Channels 1 and 2 (HP 53131A/132A Only)
Volt Peak 2
1Remove the signal from Channel 1 of the Counter and connect it
to Channel 2.
1
2Press Other Meas key until
The Counter should display -1.00 ±0.12V for the negative peak of the sine
wave. This reading is displayed on the left side of the display.
Record the actual reading in the Performance Test Record
(Test 7, Line 2a).
Also, the Counter should display +1.00 ±0.12V for the positive peak of the
sine wave. This reading is displayed on the right side of the display.
Record the actual reading in the Performance Test Record
(Test 7, Line 2b).
3Disconnect the test setup.
VOLT PEAK 2
is displayed.
Assembly-Level Service Guide1-39
Page 62
Chapter 1 Performance Tests
HP 53131A/132A HP-IB Verification Program (Optional)
HP 53131A/132A HP-IB Verification Program
1
(Optional)
The HP-IB Verification program exercises the HP 53131A/132A and
through various operating modes via the its HP-IB interface. If the
Counter successfully completes all phases of the verification program,
there is a high probability that the HP-IB interface is operating correctly.
If the Counter fails the verification program, refer to the troubleshooting
section in Chapter 2, “Service,” of this guide.
The HP-IB Verification program may be loaded into the HP 9000
series 200 or 300 Desktop Computer from the 3 1/2-inch disk
(HP P/N 53131-13501) or the 5 1/4-inch disk (HP P/N 53131-13502).
1To run the program on the disk, insert the disk into the Desktop
Computer, load the program by typing Load “VER31”, and press
RUN key on the computer’s keyboard.
2Follow the instructions displayed in the computer’s screen.
3After running this program, record the result (Pass or Fail) on the
last page (page 1-46) of the HP 53131A/132A Performance Test
Record.
1-40Assembly-Level Service Guide
Page 63
Chapter 1 Performance Tests
HP 53131A/132A Performance Test Record (Page 1 of 6)
HP 53131A/132A Performance Test Record
(Page 1 of 6)
Hewlett-Packard Model 53131A/132A Universal Counter
Serial Number: _________________________ Repair/Work Order No. __________________
Test Performed By: _____________________ Temperature: __________________________
Date: __________________________________ Relative Humidity: _____________________
Notes: _________________________________________________________________________
7External Arm Testqq
8a1 MHz External Timebase Input Test (HP 53131A Only)qq
1
8b5 MHz External Timebase Input Test (HP 53131A Only)qq
8c10 MHz External Timebase Input Test (HP 53131A Only)qq
8d10 MHz External Timebase Input Test for the HP 53132Aqq
Assembly-Level Service Guide1-41
Page 64
Chapter 1 Performance Tests
HP 53131A/132A Performance Test Record (Page 2 of 6)
HP 53131A/132A Performance Test Record (Page 5 of 6)
HP 53131A/132A Performance Test Record
(Page 5 of 6)
Complete Performance Tests (Continued)
Test
NumberTest Descri p tionMinimum
6Option 030
Channel 3
Frequency
Accuracy
(HP 53131A/132A
Only)
100 MHz Test
(HP 53131A):
Channel 3
100 MHz Test
(HP 53132A):
Channel 3
63.0 GHz Test
(HP 53131A):
Channel 3
3.0 GHz Test
(HP 53132A):
Channel 3
65.0 GHz Test
(HP 53131A):
Channel 3
5.0 GHz Test
(HP 53132A):
Channel 3
99.999,999,93 MHz
99.999, 999,980 MHz
2,999.999,998 MHz
2,999.999,999,60 MHz
4,999.999,997,0 MHz
4,999.999,999,3 MHz
Actual
ReadingMaximum
1. __________
2. __________
3. __________
4. __________
5. __________
6. __________
100.000,000,07 MHz
100.000,000,020 MHz
3,000.000,002 MHz
3,000.000, 000,4 MHz
5,000.000,003,0 MHz
5,000.000, 000,7 MHz
1
Assembly-Level Service Guide1-45
Page 68
Chapter 1 Performance Tests
HP 53131A/132A Performance Test Record (Page 6 of 6)
HP 53131A/132A Performance Test Record
1
(Page 6 of 6)
Complete Performance Tests (Continued)
Test
NumberTest Descri p tionMinimum
612.4 GHz Test
(HP 53131A):
Channel 3
12.4 GHz Test
(HP 53132A):
Channel 3
612.4 GHz Test
(HP 53131A):
Channel 3
12.4 GHz Test
(HP 53132A):
Channel 3
7Peak Volts,
Channels 1 and 2
(HP 53131A/132A
Only):
Volt Peak 1
Channel 1
Volt Peak 2
Channel 2
12.399.999 992 MHz
12,399.999 998 MHz
12,399.999,982,0 MHz
12,399.999,992,0 MHz
−1.12 V
+0.88 V
−1.12 V
+0.88 V
Actual
ReadingMaximum
7. __________
8. __________
7. __________
8. __________
1 a.___________
b.___________
2 a.___________
b.___________
12,400.000 008 MHz
12,400.000 002 MHz
12,400.000 018, MHz
12,400.000, 008,0 MHz
−0.88 V
+1.12V
−0.88 V
+1.12V
Complete Performance Tests (Continued)
HP-IB Verification (Optional) Test Results
Pass Fail
qq
1-46Assembly-Level Service Guide
Page 69
Chapter 1 Performance Tests
HP 53181A Complete Performance Tests
HP 53181A Complete Performance Tests
NOTE
The specifications of the HP 53181A Frequency Counter can be verified
by performing the performance tests provided in this section. Table 1-3
lists a summary of the HP 53181A performance tests.
Record the results of the performance tests in the appropriate
place on HP 53181A Performance Test Record, which starts on
page 1-71.
(Note: the complete performance tests for the HP 531 31A/132A Universal
Counter begin on page 1-16 of this chapter.)
To perform valid testing of the specifications, warm up the Counter and
test equipment for 30 minutes.
Table 1-3. The HP 53181A Performance Tests
Page Number*Test Description
page 1-48Test 1: Trigger Level (HP 53181A Only)
page 1-51Test 2: Channel 1 Frequency Sensitivity (HP53181A Only)
page 1-55Test 3: Chan nel 1 Frequency Accuracy (HP 53181A Only)
page 1-58Test 4: Option 015/030/050/124 Channel2 Frequency
Sensitivity
1
page 1-63Test 5: Option 015/030/050/124 Channel 2 Frequency
* Other Counter measurement functions (e.g., Period) are mathematically derived by the microprocessor from
the parameters verified by these performance tests.
Assembly-Level Service Guide1-47
Page 70
Chapter 1 Performance Tests
Test 1: Trigger Level (HP 53181A Only)
Test 1: Trigger Level (HP 53181A Only)
1
This test verifies the Trigger Level accuracy of the HP 53181A Frequency
Counter.
Equipment
HP 3325B Synthesizer
HP 10503A BNC Cable
Counter Setup
1Cycle the POWER key to preset the Counter.
2Set CHANNEL 1 input condition to:
Trigger/Sensitivity,
Trigger/Sensitivity,
50Ω/1MΩ, 50
DC/AC, DC
X10 Attenuate, Off
100kHz Filter, Off
Ω
AUTO TRG: OFF
LEVEL: 0.000V
1-48Assembly-Level Service Guide
Page 71
Chapter 1 Performance Tests
Test 1: Trigger Level (HP 53181A Only)
Procedure
1Connect the output of the HP 3325B to Channel 1 of the Counter
as shown in Figure 1-11.
HP 3325B
HP 53181A
Counter
Synthesizer
1
Channel 1
Figure 1-11. Trigger Level Test Setup
2Set the HP 3325B to output a 1 MHz, 80 mVp-p s quare wave signal.
3Set the HP 3325B DC OFFSET to −60 mV.
4On the Counter, press Run key.
5Now, increment the dc offset on the HP 3325B by +1 mV until the
Counter’s
until the Counter displays approximately 1 MHz.
Observe the offset value on the display of the HP 3325B.
Record the dc offset value__________ mV.
6Add the upper peak voltage (40 mV) of the 80 mV p-p signal to the
offset value in step 5 (For example, −24 mV + 40 mV = 16 mv).
This is the upper hysteresis level.
Gate
annunciator flashes and continue incrementing
Output
Record the upper hysteresis l evel__________ mV.
Assembly-Level Service Guide1-49
Page 72
Chapter 1 Performance Tests
Test 1: Trigger Level (HP 53181A Only)
7In the Counter’s Channel 1 Trigger/Sensitivity menu, change the
SLOPE to NEG.
1
8On the Counter, press Run key.
9Set the HP 3325B DC OFFSET to +60 mV.
10Now, decrement the offset on the HP 3325B by −1 mV until the
Counter’s
until the Counter displays approximately 1 MHz.
Observe the offset value on the display of the HP 3325B.
Record the value__________ mV.
11Add the lower peak voltage (−40mV) of the 80 mVp-p signal to the
offset value in step 10 (for example, 21 mV − 40 mV = −19 mV.)
This result is the lower hysteresis level.
Record the lower hysteresis level __________ mV.
12Now, add the upper hysteresis value (recorded in step 6 )and
lower hysteresis value (recorded in step 11).
This is the trigger level; it should be 0.0 V ±15 mV.
13Record the trigger level value in the Performance Test Record
(Test 1, Line 1).
Gate
annunciator flashes and continue decrementing
1-50Assembly-Level Service Guide
Page 73
Chapter 1 Performance Tests
Test 2: Channel 1 Frequency Sensitivity (HP 53181A Only)
Test 2: Channel 1 Frequency Sensitivity
(HP 53181A Only)
This set of tests verifies frequency sensitivity specifications of the
HP 53181A Frequency Counter.
Equipment
HP 8663A Synthesized Signal Generator
HP 1250-0780 Type N-to-BNC Connector
HP 10100C 50Ω Feedthrough
HP 10503A BNC Cables (2)
Counter Setup
1Cycle the POWER key to preset the Counter.
2Set CHANNEL 1 input conditions to:
Trigger/Sensitivity,
Trigger/Sensitivity,
50Ω/1MΩ, 50
DC/AC, DC
X10 Attenuate, Off
100kHz Filter, Off
Ω
AUTO TRG: OFF
LEVEL: 0.000 V
1
Assembly-Level Service Guide1-51
Page 74
Chapter 1 Performance Tests
Test 2: Channel 1 Frequency Sensitivity (HP 53181A Only)
100 kHz to 100 MHz Sensitivity for HP 53181A
1Connect the equipment as shown in Figure 1-12 with the signal
generator connected to Channel 1 of the Counter.
1
Input
(rear panel)
BNC Tee
HP 8663A
Synthesized Signal Generator
HP 53181A
Counter
50Ω
Feedthrough
Ref. In
(rear panel)
Channel 1
Output
N-to-BNC
Connector
Figure 1-12. Channels 1 Frequency Sensitivity Test Setup
2Set the HP 8663A Signal Generator to output a 100 kHz, −35.0 dBm
sine wave signal (−20.0 dBm for Option 060) to the Counter’s
Channel 1 input.
3In 0.1 dB steps, increase the power level until the Counter
displays a stable reading of 100 kHz.
The Counter should display 100 kHz when the signal generator level
≤ −
is
21.0 dBm (20 mVrms). Counters with Option 060 Rear Terminals
should display 100 kHz when the signal generator level is
≤ −
9.5 dBm
(75 mVrms).
1-52Assembly-Level Service Guide
Page 75
Chapter 1 Performance Tests
Test 2: Channel 1 Frequency Sensitivity (HP 53181A Only)
4Sweep the frequency from 100 kHz to 100 MHz. For Option 060,
sweep frequencies from 100 kHz to 225 MHz.
NOTE
NOTE
The Counter should read frequencies from 100 kHz to 100 MHz at an
input power level of
read frequencies from 100 kHz to 225 MHz at
5Record the actual reading in the Performance Test Record
(Test 2, Line 1).
If you are testing a Counter with front-panel terminals (Standard),
record the reading on Line 1 of the Standard portion of the test record.
If you are testing a Counter with rear terminals (Option 060),
record the reading on Line 1 of the Option 060 portion of the test record.
DO NOT test the front terminals if rear terminals are installed. The front
terminal performance is not specified when the rear terminals are
installed.
The remaining sensitivity procedures are NOT required for Counters that
contain Option 060 Rear Terminals; thus, this completes the sensitivity
portion of the Frequency Performance Test. Go to the Frequen cy Accuracy
Test on page 1-55.
≤ −
21.0 dBm (20 mVrms). Option 060 Counters should
≤ −
9.5 dBm (75 mVrms).
1
Assembly-Level Service Guide1-53
Page 76
Chapter 1 Performance Tests
Test 2: Channel 1 Frequency Sensitivity (HP 53181A Only)
100 MHz to 200 MHz Sensitivity for HP 53181A
1Change the signal generator settings to 110 MHz at −30 dBm.
1
2Increase the power level by 0.1 dB steps until the Counter
displays a stable 110 MHz reading.
The Counter should read 200 MHz when the input signal level
≤ −
is
17.5 dBm (30 mVrms).
3Sweep the signal generator frequency from 100 MHz to 200 MHz.
The Counter should read frequencies from 100 MHz to 200 MHz when the
input signal level is
4Record the actual reading in the Performance Test Record
(Test 2, Line 2).
≤ −
17.5 dBm (30 mVrms).
200 MHz to 225 MHz Sensitivity for HP 53181A
1Change the signal generator settings to 200 MHz at −30 dBm.
2Increase the power level by 0.1 dB steps until the Counter
displays a stable 200 MHz reading.
The Counter should read 200 MHz when the input signal level is
≤ −
15.0 dBm (40 mVrms).
3Sweep the signal generator frequency from 200 MHz to 225 MHz.
The Counter should read frequencies from 200 MHz to 225 MHz when the
input signal level is
4Record the actual reading in the Performance Test Record
(Test 2, Line 3).
5Disconnect the test setup.
1-54Assembly-Level Service Guide
≤ −
15.0 dBm (40 mVrms).
Page 77
Chapter 1 Performance Tests
Test 3: Channel 1 Frequency Accuracy (HP 53181A Only)
Test 3: Channel 1 Frequency Accuracy
(HP 53181A Only)
This set of tests verifies the frequency accuracy specifications of the
HP 53181A Frequency Counter.
Equipment
HP 8663A Synthesized Signal Generator
HP 8340B Sweep Oscillator
HP 1250-0780 Type N-to-BNC Connector
HP 10100C 50Ω Feedthrough
HP 10503A BNC Cables (2)
Counter Setup
1Cycle the POWER key to preset the Counter.
2Using the Gate & ExtArm key, set gate time to
3Set CHANNEL 1 input conditions to:
Trigger/Sensitivity,
Trigger/Sensitivity,
50Ω/1MΩ, 50
DC/AC, DC
X10 Attenuate, Off
100kHz Filter, Off
Ω
AUTO TRG: OFF
LEVEL: 0.000 V
TIME 1.000 s
1
.
Assembly-Level Service Guide1-55
Page 78
Chapter 1 Performance Tests
Test 3: Channel 1 Frequency Accuracy (HP 53181A Only)
Procedure
1Connect the equipment as shown in Figure 1-13 with the signal
generator connected to Channel 1 of the Counter.
1
HP 53181A
Counter
Channel 1
50Ω
Feedthrough
Ref. In
(rear panel)
HP 8340B
Sweep Oscillator
Output
HP 8663A
Synthesized Signal Generator
Output
Frequency Standard
(switch set to EXT)
Input
(rear panel)
Figure 1-13. Channels 1 Frequency Accuracy Test Setup
2Set the HP 8663A Signal Generator to output 1 MHz at −7 dBm.
3On the Counter, press Run key.
Verify the Counter reads 1 MHz ±1 mHz.
1-56Assembly-Level Service Guide
N-to-BNC
Connector
Page 79
Chapter 1 Performance Tests
Test 3: Channel 1 Frequency Accuracy (HP 53181A Only)
4Record the actual reading in the Performance Test Record
(Test 3, Line 1).
5Change the signal generator frequency to 100 MHz.
Verify the Counter reads 100 MHz ±70 mHz.
6Record the actual reading in the Performance Test Record
(Test 3, Line 2).
7Change the signal generator frequency to 200 MHz.
Verify the Counter reads 200 MHz ±130 mHz.
8Record the actual reading in the Performance Test Record
(Test 3, Line 3).
9Change the signal generator frequency to 225 MHz.
Verify the Counter reads 225 MHz ±150 mHz.
10Record the actual reading in the Performance Test Record
(Test 3, Line 4).
11Disconnect the test setup.
1
Assembly-Level Service Guide1-57
Page 80
Chapter 1 Performance Tests
Test 4: Option 015/030/050/124 Channel 2 Frequency Sensitivity
Test 4: Option 015/030/050/124 Channel 2
1
Frequency Sensitivity
This test verifies the frequency range and sensitivity of the optional
1.5-GHz (Option 015), 3-GHz (Option 030), 5.0-GHz (Option 050), or
12.4-GHz (Option 124) Channel 2 for the HP 53181A Frequency Counter.
Equipment
HP 8663A Synthesized Signal Generator (Not needed for Option 050/124)
HP 8340B Sweep Oscillator
HP 1250-0780 Type N-to-BNC Connector (Not needed for Option 050/124)
HP 10503A BNC Cable (2)
HP 10100C 50Ω Feedthrough
Counter Setup
1Cycle the POWER key to preset the Counter.
2Set the Counter as follows:
Freq & Ratio,
FREQUENCY 2
1-58Assembly-Level Service Guide
Page 81
Chapter 1 Performance Tests
Test 4: Option 015/030/050/124 Channel 2 Frequency Sensitivity
100 MHz to 1.5 GHz Sensitivity for Option 015 Only
1Set the HP 8663B Synthesized Signal Generator to output a
500 MHz, −60 dBm sine wave and connect the signal to Channel 2
of the Counter as shown in Figure 1-14.
1
HP 53181A
Counter
Channel 2
50Ω
Feedthrough
Ref. In
(rear panel)
HP 8340B
Sweep Oscillator
Output
HP 8663A
Synthesized Signal Generator
Output
Frequency Standard
(switch set to EXT)
Input
(rear panel)
Figure 1-14. Option 015/030/050 Channel 2 Frequency Test
2On the Counter, Press Run key.
Assembly-Level Service Guide1-59
N-to-BNC
Connector
Page 82
Chapter 1 Performance Tests
Test 4: Option 015/030/050/124 Channel 2 Frequency Sensitivity
3Increase the power level in 1 dB steps until the Counter reads
500 MHz.
The Counter should display 500 MHz at
≤ −
27.0 dBm (20 mVrms).
1
4Sweep the frequencies from 100 MHz to 1.5 GHz.
The Counter should read frequencies from 100 MHz to 1.5 GHz
≤ −
at
27.0dBm (20 mVrms).
5Record the actual reading in the Performance Test Record
(Test 4, Line 1).
100 MHz to 2.7 GHz Sensitivity for Option 030 Only
1Set the HP 8663B Synthesized Signal Generator to output a
500 MHz, −60 dBm sine wave, and connect the signal to Channel 3
of the Counter as shown in Figure 1-14.
2On the Counter, Press Run key.
3Increase the power level in 1 dB steps until the Counter reads
500 MHz.
The Counter should display 500 MHz at
4Sweep the frequencies from 100 MHz to 2.7 GHz.
≤ −
27.0 dBm (20 mVrms).
The Counter should read frequencies from 100 MHz to 2.7 GHz
≤ −
at
27.0 dBm (20 mVrms).
5Record the actual reading in the Performance Test Record
(Test 4, Line 2).
1-60Assembly-Level Service Guide
Page 83
Chapter 1 Performance Tests
Test 4: Option 015/030/050/124 Channel 2 Frequency Sensitivity
2.8 to 3.0 GHz Sensitivity for Option 030 Only
1Disconnect the HP 8663A Synthesized Signal Generator from the
Counter (see Figure 1-14), and connect the HP 8340B Sweep
Oscillator to Channel 3 of the Counter.
2Change the settings of the Sweep Oscillator to 3.0 GHz, −40 dBm.
3Increase the power level in 1 dB steps until the Counter displays
3.0 GHz.
1
NOTE
The Counter should display 3.0 GHz at
4Sweep the frequencies from 3.0 GHz to 2.8 GHz.
The Counter should read frequencies from 3.0 to 2.8 GHz at
5Record the actual reading in the Performance Test Record
(Test 4, Line 3).
6Disconnect the test setup.
≤ −
21.0 dBm.
≤ −
21.0 dBm.
200 MHz to 5.0 GHz Sensitivity for Option 050 only
1Connect the 8340B Sweep Oscillator to Channel 2 of the counter.
2Change the settings of the Sweep Oscillator to 5.0 GHz, −30 dBm.
3Increase the power level in 1 d B steps until the Counter displays a
stable reading of approximately 5.0 GHz.
The Counter should display 5.0 GHz at
Use a power meter and sensor to ensure the accuracy of the power being
input.
4Sweep the frequencies from 200 MHz to 5.0 GHz.
≤ −
23 dBm.
The Counter should read freque ncies from 200 MHz to 5.0 GHz at ≤ −23 dBm.
5Record the actual reading in the Performance Test Record
(Test 4, Line 4).
Assembly-Level Service Guide1-61
Page 84
Chapter 1 Performance Tests
Test 4: Option 015/030/050/124 Channel 2 Frequency Sensitivity
200 MHz to 12.4 GHz Sensitivity for Option 12 4 only
1Connect the 8340B Sweep Oscillator to Channel 2 of the counter.
1
NOTE
2Change the settings of the Sweep Oscillator to 12.4 GHz, −30 dBm.
3Increase the power level in 1 d B steps until the Counter displays a
stable reading of approximately 12.4 GHz.
The Counter should display 12.4 GHz at
Use a power meter and sensor to ensure the accuracy of the power being
input.
4Sweep the frequencies from 200 MHz to 12.4 GHz.
The Counter should read frequencies from 200 MHz to 12.4 GHz
≤ −
at
23 dBm.
5Record the actual reading in the Performance Test Record
(Test 4, Line 5).
6Disconnect the test setup.
≤ −
23 dBm.
1-62Assembly-Level Service Guide
Page 85
Chapter 1 Performance Tests
Test 5: Option 015/030/050/124 Channel 2 Frequency Accuracy (HP
53181A Only)
Test 5: Option 015/030/050/124 Channel 2
Frequency Accuracy (HP 53181A Only)
This test verifies the frequency accuracy of the optional 1.5-GHz
(Option 124) Channel 2 for the HP 53181A Frequency Counter.
The tests are organized as follows:
•Option 015 Channel 2 Frequency Accuracy Test, starts on page 1-63
•Option 030 Channel 2 Frequency Accuracy Test, starts on page 1-65
•Option 050 Channel 2 Frequency Accuracy Test, starts on page 1-65
•Option 124 Channel 2 Frequency Accuracy Test, starts on page 1-67
100 MHz to 1.5 GHz Accurac y Test for Op tion 015
Equipment
HP 8663A Synthesized Signal Generator
HP 8340B Sweep Oscillator
HP 1250-0780 Type N-to-BNC Connector
HP 10503A BNC Cable (2)
HP 10100C 50Ω Feedthrough
HP 11500D N-N Cable (Option 050 only)
HP 1250-1250 N(M) to SMA(F)
HP 1250-0777 N(F) to N(F)
1
Counter Setup
1Cycle the POWER key to preset the Counter.
2Set the Counter as follows:
MEASURE/Freq & Ratio,
Gate & ExtArm,
Assembly-Level Service Guide1-63
TIME: TIME 1.000 s
FREQUENCY 2
Page 86
Chapter 1 Performance Tests
Test 5: Option 015/030/050/124 Channel 2 Frequency Accuracy (HP
53181A Only)
Procedure
1Connect equipment as shown in Figure 1-15.
1
HP 8340B
Sweep Oscillator
HP 8663A
Synthesized Signal Generator
Frequency Standard
(switch set to EXT)
Output
Input
(rear panel)
HP 53181A
Counter
Channel 2
50Ω
Feedthrough
Ref. In
(rear panel)
Output
Figure 1-15. Option 015/030/050/124 Channel 2 Frequency Accuracy Test Setup
2Set the HP 8663A Synthesized Signal Generator to output
100 MHz at −17.0 dBm.
Verify the Counter reads 100 MHz ±70 mHz.
1-64Assembly-Level Service Guide
N-to-BNC
Connector
Page 87
Chapter 1 Performance Tests
Test 5: Option 015/030/050/124 Channel 2 Frequency Accuracy (HP
53181A Only)
3Record the actual reading in the Performance Test Record
(Test 5, Line 1).
4Change the synthesized signal generator’s frequency to 1.5 GHz.
Verify the Counter reads 1.5 GHz ±1 Hz
5Record the actual reading in t he Performa nce Test (Tes t 5, Line 2).
6Disconnect the test setup.
Minimum to Maximum Accuracy Test for Options
030/050/124
Equipment
HP 8663A Synthesized Signal Generator (not needed for Option 050)
HP 8340B Sweep Oscillator
HP 1250-0780 Type N-to-BNC Connector (not needed for Option 050)
HP 10503A BNC Cable (2)
HP 10100C 50Ω Feedthrough
HP 11500D N-N Cable (Option 050 only)
HP 1250-1250 N(M) to SMA(F)
HP 1250-0777 N(F) to N(F)
Counter Setup
1Cycle the POWER key to preset the Counter.
1
2Set the Counter as follows:
MEASURE/Freq & Ratio,
Gate & ExtArm,
Assembly-Level Service Guide1-65
TIME: TIME 1.000 s
FREQUENCY 2
Page 88
Chapter 1 Performance Tests
Test 5: Option 015/030/050/124 Channel 2 Frequency Accuracy (HP
53181A Only)
Procedure
1Connect equipment as shown in Figure 1-15.
1
(NOTE: For Option 050/124, skip steps 2 through 7. Proceed to
step 8.)
2Set the HP 8663A Synthesized Signal Generator to output
200 MHz at −17.0 dBm.
3On the Counter, press Run key.
Verify the HP 53181A Counter reads 100 MHz ±70 mHz.
4Record the actual reading in the Performance Test Record
(Test 5, Line 3).
5Disconnect the HP 8663A output from Channel 2 of the Counter,
and connect the HP 8340B Sweep Oscillator output to Channel 2
of the Counter as shown in Figure 1-15.
6Set the HP 8340B to output 3.0 GHz at −17.0 dBm.
Verify the HP 53181A Counter reads 3.0 GHz ±2 Hz.
7Record the actual reading in the Performance Test Record.
(Test 5, Line 4).
1-66Assembly-Level Service Guide
Page 89
Chapter 1 Performance Tests
Test 5: Option 015/030/050/124 Channel 2 Frequency Accuracy (HP
53181A Only)
(NOTE: For Options 050/124)
8Set the HP 8340B to output 5.0 GHz at −17 dBm.
Verify the HP 53181A Counter reads 5.0 GHz ±3 Hz.
9Record the actual reading in t he Performance Test Record (Test 5,
Line 5).
(NOTE: For Option 124)
10Set the HP 8340B to output 12.4 GHz at –17 dBm.
Verify the HP 53181A Counter reads 12.4 GHz ±8 Hz.
11Record the actual reading in the Performance Test Record (Test 5,
Line 6).
12Disconnect the test setup.
1
Assembly-Level Service Guide1-67
Page 90
Chapter 1 Performance Tests
Test 6: Peak Volts, Channel 1 (HP 53181A Only)
Test 6: Peak Volts, Channel 1 (HP 53181A
1
Only)
This test verifies the Peak Volts accuracy specification of the HP 53181A
Frequency Counter.
Equipment
HP 3325B Synthesizer
HP 10100C 50Ω Feedthrough
HP 10503A BNC Cable
Counter Setup
1Cycle the POWER key to preset the Counter.
2Press Other Meas key until
3Set CHANNEL 1 and CHANNEL 2 inputs to:
50Ω/1MΩ, 50
X10 Attenuate, Off
100kHz Filter, Off
Ω
VOLT PEAKS 1
is displayed.
1-68Assembly-Level Service Guide
Page 91
Chapter 1 Performance Tests
Test 6: Peak Volts, Channel 1 (HP 53181A Only)
Procedure
1Connect the equipment as shown in Figure 1-16.
HP 3325B
HP 53181A
Counter
Synthesizer
1
Channel 1
Figure 1-16. Peak Volts Test Setup
2Set the HP 3325B to output a 2 MHz, 2 Vp-p sine wave, and connect
the signal to Channel 1 of the Counter.
The Counter should display −1.00 ±0.12V for the negative peak of the
sine wave. This reading is displayed on the left side of the display.
Record the actual readings in the Performance Test Record
(Test 6, Line a).
Also, the Counter should display +1.00 ±0.12V for the positive peak of the
sine wave. This reading is displayed on the right side of the display.
Record the actual readings in the Performance Test Record
(Test 6, Line b).
3Disconnect the test setup.
Output
Assembly-Level Service Guide1-69
Page 92
Chapter 1 Performance Tests
HP 53181A HP-IB Verification Program (Optional)
HP 53181A HP-IB Verification Program
1
(Optional)
The HP-IB Verification program exercises the HP 53181A through various
operating modes via the its HP-IB interface. If the Counter successfully
completes all phases of the verification program, there is a high
probability that the HP-IB interface is operating correctly. If the Counter
fails the verification program, refer to the troubleshooting section in
Chapter 2, “Service,” of this guide.
The HP-IB Verification program may be loaded into the HP 9000
series 200 or 300 Desktop Computer from the 3 1/2-inch disk
(HP P/N 53181-13501) or the 5 1/4-inch disk (HP P/N 53181-13501).
1To run the program on the disk, insert the disk into the Desktop
Computer, load the program by typing Load “VER81”, and press
RUN key on the computer’s keyboard.
2Follow the instructions displayed in the computer’s screen.
3After running this program, record the result (Pass or Fail) on the
last page (page 1-74) of the HP 53181A Performance Test Record.
1-70Assembly-Level Service Guide
Page 93
Chapter 1 Performance Tests
HP 53181A Performance Test Record (Page 1 of 4)
HP 53181A Performance Test Record
(Page 1 of 4)
Hewlett-Packard Model 53181A Frequency Counter
Serial Number: _________________________ Repair/Work Order No. __________________
Test Performed By: _____________________ Temperature: __________________________
Date: __________________________________ Relative Humidity: _____________________
Notes: _________________________________________________________________________
for Option 124:12,399.999,9926. __________12.400.000,008 MHz
Assembly-Level Service Guide1-73
Page 96
Chapter 1 Performance Tests
HP 53181A Performance Test Record (Page 4 of 4)
HP 53181A Performance Test Record
1
(Page 4 of 4)
Complete Performance Tests
Test
NumberTest Descri p tionMinimum
6Peak Volts,
Channel 1
(HP 53181A Only):
Channel 1−1.12 V
+0.88 V
HP-IB Verification (Optional) Test Results
Actual
ReadingMaximum
a. __________
b. __________
Pass Fail
−0.88 V
+1.12V
qq
1-74Assembly-Level Service Guide
Page 97
2
Service
Page 98
Chapter 2 Service
Introduction
Introduction
This chapter provides service information for your HP 53131A/132A/181A
and is divided into seven major sections:
•Returning the Instrument to Hewlett-Packard for Servic e
(page 2-3). This sect ion prov id es y ou wi th step- by -step i ns tructi ons on
how to return the instrument for service.
•About the HP 53131A/132A Calibration Menu (page 2-6).
2
This section contains reference information that gives a complete
overview of the calibration menu of the HP 53131A/132A.
•The HP 53131A/132A Calibration Procedures (page 2-15).
This section provides step-by-step procedures for calibrating the
HP 53131A/132A.
•About the HP 53181A Calibration Menu (page 2-25). This section
contains reference information that gives a complete overview of the
calibration menu of the HP 53181A.
•The HP 53181A Calibration Procedures (page 2-31). This section
provides step-by-step procedures for calibrating the HP 53181A.
•Pre-Troubleshooting Information (page 2-36). This section
provides you with pertinent information such as safety considerations,
recommended test equipment, repair and after-service cons iderations,
and assembly identification and location.
•Troubleshooting the Counter (page 2-45). This section provides
you with troubleshooting procedures that isolate the faulty assembly
or module.
If the instrument is under warranty, return the instrument to
Hewlett-Packard for service. Refer to the first section of this chapter titled
“Returning the Instrument to Hewlett-Packard for Service.” If you decide
to troubleshoot the instrument yourself, refer to the section titled
“Troubleshooting the Counter.”
2-2Assembly-Level Service Guide
Page 99
Chapter 2 Service
Returning the Instrument to Hewlett-Packard for Service
Returning the Instrument to Hewlett-Packard
for Service
To Provide Repair Information
If you are shipping the instrument to an HP office for service or repair,
call your nearest Hewlett-Packard Sales Office to make arrangements.
Then tag and package the HP 53131A/132A/181A for shipment.
Note there is an Express Repair/Performance Calibration Service for USA
customers if downtime is critical. You can receive your repaired Counter
via overnight shipment. Just call 800-403-0801 and ask for Express
Repair/Performance Calibration Service. When your Counter is repaired,
it will be returned via overnight shipment at no extra charge.
1Write the following information on a tag:
•Owner’s name and address
•Instrument model number
•Complete serial number
•Description of service required or failure indications
2Attach the tag to the instrument.
3Pack the instrument.
If you have the original packaging materials, refer to the following section
titled “To Pack in the Original Packaging Materials.” If the original
packaging material are not available, you can order new materials
through an HP Sales Office. The new materials are identical to those used
by the factory when packaging the instrument. If you want to use
commercially available materials, refer to the section titled “To Pack in
the Commercially Available Materials.”
2
Assembly-Level Service Guide2-3
Page 100
Chapter 2 Service
Returning the Instrument to Hewlett-Packard for Ser vice
To Pack in the Original Packaging Materials
Whenever possible, repack the instrument in its original packaging for
shipment. In any correspondence, refer to the instrument by the model
number and complete serial number.
1Disconnect the power cord, probes, cables, or other accessories
attached to the instrument.
2Make sure the folded corrugated sp acer (which normally contains
the manuals) is in the box to ensure proper fitting.
2
3Make sure one of th e pol yst yrene bl ock is in i ts prop er p osit ion i n
the box.
4Place the instrument, front panel end first, on the polystyrene
block.
5Place the other polystyrene block on top of the rear end of the
instrument to secure it.
6Do not return the manuals with the instrument. Return an
accessory only when it is a part of the failure symptoms.
7Seal the shipping container securely.
2-4Assembly-Level Service Guide
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.