No part of this manual may be reproduced
in any form or by any means (including
electronic storage and retrieval or translation into a foreign language) without prior
agreement and written consent from Agilent Technologies, Inc. as governed by
United States and international copyright
laws.
This manual covers the following
Agilent equipment:
• E6198B Switch/Load Unit
• E6175A 8-Channel High Current
Load Card
• E6176A 16-Channel High Current
Load Card
• E6177A 24-Channel Medium Current
Load Card
• U1777A 24-Channel Medium
Current Load Card
• U7178A 8-Channel Heavy Duty Load
Card
• U7179A 16-Channel High Current
Load Card
• E6178B 8-Channel Heavy Duty Load
Card
• N9377A 16-Channel Dual-Load Load
Card
• N9378A 24-Channel Low Resistance
Load Card
• N9379A 48-Channel High Density
Load Card
• E8792A Pin Matrix Card
• E8793A Pin Matrix Card
• E8782A 24-Instrument,
40-Measurement Matrix Card
• E8783A 64-Pin Matrix Card
• E8794A Custom Card
Manual Part Number
E6198-90050
Edition
9th Edition, August 2012
Printed in Malaysia
Agilent Technologies Microwave Products
(Malaysia) Sdn. Bhd.
Bayan Lepas Free Industrial Zone
11900 Penang, Malaysia
Warranty
The material contained in this document is provided “as is,” and is subject to being changed, without notice,
in future editions. Further, to the maximum extent permitted by applicable
law, Agilent disclaims all warranties,
either express or implied, with regard
to this manual and any information
contained herein, including but not
limited to the implied warranties of
merchantability and fitness for a particular purpose. Agilent shall not be
liable for errors or for incidental or
consequential damages in connection
with the furnishing, use, or performance of this document or of any
information contained herein. Should
Agilent and the user have a separate
written agreement with warranty
terms covering the material in this
document that conflict with these
terms, the warranty terms in the separate agreement shall control.
Technology Licenses
The hardware and/or software described
in this document are furnished under a
license and may be used or copied only in
accordance with the terms of such license.
Restricted Rights Legend
defined in FAR 2.101(a) or as “Restricted
computer software” as defined in FAR
52.227-19 (June 1987) or any equivalent
agency regulation or contract clause. Use,
duplication or disclosure of Software is
subject to Agilent Technologies’ standard
commercial license terms, and non-DOD
Departments and Agencies of the U.S.
Government will receive no greater than
Restricted Rights as defined in FAR
52.227-19(c)(1-2) (June 1987). U.S. Government users will receive no greater than
Limited Rights as defined in FAR 52.227-14
(June 1987) or DFAR 252.227-7015 (b)(2)
(November 1995), as applicable in any
technical data.
Safety Notices
A CAUTION notice denotes a hazard. It calls attention to an operating procedure, practice, or the like
that, if not correctly performed or
adhered to, could result in damage
to the product or loss of important
data. Do not proceed beyond a
CAUTION notice until the indicated
conditions are fully understood and
met.
A WARNING notice denotes a
hazard. It calls attention to an
operating procedure, practice, or
the like that, if not correctly performed or adhered to, could result
in personal injury or death. Do not
proceed beyond a WARNING
notice until the indicated conditions are fully understood and
met.
If software is for use in the performance of
a U.S. Government prime contract or subcontract, Software is delivered and
licensed as “Commercial computer software” as defined in DFAR 252.227-7014
(June 1995), or as a “commercial item” as
Contents
1Legal Information
Legal Information 1-2
Warranty 1-2
Technology Licenses 1-2
Restricted Rights Legend 1-2
Service and Support 1-3
Agilent on the Web 1-3
Agilent By Phone 1-3
2Safety and Regulatory Information
Declaration of Conformity 2-2
Safety Information 2-3
Safety Summary 2-3
Safety Notice 2-3
General Safety Information 2-3
Environmental Conditions 2-4
Before Applying Power 2-4
Ground the System 2-5
Fuses 2-5
Operator Safety Information 2-5
Safety Symbols and Regulatory Markings 2-6
End of Life: Waste Electrical and Electronic Equipment (WEEE) Directive
2002/96/EC 2-8
3Switch/Load Unit and Plug-In Cards
E6198B Standalone/Integrated Switch/Load Unit Rating 3-2
Electrical 3-2
Mechanical 3-2
E6198B Standalone Switch/Load Unit Description 3-3
USB Interface 3-5
Digital I/O 3-5
Digital Input 3-5
Open Drain Digital Output 3-6
DAC Channels 3-7
+5 Vdc, +12 Vdc, –12 Vdc, Spare Power 3-7
Current Sense 3-7
Power Bus Sense 3-7
Switch/Load Unit Block Diagram 3-7
E6198B Integrated Switch/Load Unit Description 3-9
E6198B/E6218A Switch/Load Unit User Manualiii
Backplane And Breakout Board Connectors and LEDs 3-10
Backplane Connector Name 3-12
Backplane J1, J2 and J3 Connectors 3-13
Breakout Board Connector Name (for E6198B Standalone
Differentiating E6198B Standalone Option and System Integrated
Option 3-22
Fuse Ratings 3-24
DC Power Bus Connections 3-24
4Configuring the Switch/Load Unit
Card Location Recommendations 4-2
Load Card Types and IDS 4-4
Connecting an Additional (Spare) Power Supply 4-5
Configuring the Power Busses 4-6
Setting the UUT Power Supply Remote/Local Sense Jumpers 4-8
Configuring the Current-Sense Jumpers 4-10
Connecting E6198B to the Computer via USB Interface or Parallel
Port 4-12
Adding a Second Switch/Load Unit 4-13
Load Box Installation 4-14
5Using Load Cards and Loads
Load Card Capabilities 5-2
Loads Overview 5-4
Using the Load Cards 5-5
Load Card Type and Configuration ID 5-5
Using the Agilent E6175A 8-Channel Load Card 5-7
Selecting a Power Supply Configuration 5-9
Selecting a Current-Sense Method 5-10
Installing a LEM Current Transducer 5-11
Selecting and Loading Flyback Protection 5-14
Protection Devices 5-16
ivE6198B/E6218A Switch/Load Unit User Manual
Selecting a Load Fuse 5-17
Sample Load Configurations 5-17
Setting up a Bridge Configuration 5-19
Connecting Loads 5-20
Load Wiring 5-21
Current Sharing 5-22
UUT Connections 5-23
Using the Agilent E6176A 16-Channel Load Card 5-25
Selecting a Power Supply Configuration 5-27
Selecting a Current-Sense Resistor Value 5-28
Selecting a Load Fuse 5-28
Selecting and Loading Flyback Protection 5-28
Protection Devices 5-30
Connecting Loads 5-32
Load Wiring 5-32
Current Sharing 5-34
External Load Mounting Options 5-35
UUT Connections 5-36
Using the Agilent E6177A 24-Channel Load Card 5-38
Card Layout 5-40
Selecting a Power Supply Configuration 5-41
Using the Power Switches as General Purpose Relays 5-41
Connecting Loads 5-42
Load Wiring 5-44
UUT Connections 5-45
Using the Agilent U7177A 24-Channel Load Card 5-47
Selecting a Power Supply Configuration 5-49
Selecting a Current-Sense Resistor Value 5-50
Using the Power Switches as General Purpose Relays 5-50
Connecting Loads 5-51
Load Wiring 5-52
UUT Connections 5-54
Using the Agilent E6178B 8-Channel Load Card 5-55
Selecting a Power Supply Configuration 5-57
Local / Remote Sensing 5-58
Selecting a Load Fuse 5-58
Current Monitor 5-59
Selecting and Loading Flyback Protection 5-59
Protection Devices 5-60
Load and UUT Connections 5-61
Using the Agilent U7178A 8-Channel Heavy Duty Load Card 5-63
E6198B/E6218A Switch/Load Unit User Manualv
Selecting a Power Supply Configuration 5-65
Local/Remote Sensing 5-66
Selecting a Load Fuse 5-66
Current Monitor 5-66
Selecting and loading flyback protection 5-67
Protection Devices 5-68
Load and UUT connections 5-69
Using the Agilent U7179A 16-Channel High Current Load Card 5-71
Selecting a Power Supply Configuration 5-73
Selecting a Current-Sense Resistor Value 5-74
Selecting a Load Fuse 5-74
Selecting and Loading Flyback Protection 5-74
Protection Devices 5-75
Load Wiring 5-77
External Load Mounting Options 5-79
UUT Connections 5-81
Using the Agilent N9377A 16-Channel Dual-Load Load Card 5-83
Selecting a Power Supply Configuration 5-86
Selecting a Current-Sense Resistor Value 5-86
Selecting a Load Fuse 5-86
Selecting and Loading Flyback Protection 5-87
Protection Devices 5-88
Connecting Loads 5-90
Load Wiring 5-90
Recommended Connectors 5-94
Current Sharing 5-95
External Load Mounting Options 5-95
UUT Connections 5-96
Using the Agilent N9378A 24-Channel Low Resistance Load Card 5-97
Selecting a Power Supply Configuration 5-100
Using the Power Switches as General Purpose Relays 5-100
Connecting Loads 5-101
Installing and Removing Mezzanine Cards 5-102
Wiring the Mezzanine Cards 5-103
UUT Connections 5-103
Using the Agilent N9379A 48-Channel High-Density Load Card 5-105
Card Layout 5-107
Load Circuit Details 5-109
Selecting a Power Supply Configuration 5-109
Using the Power Switches as General Purpose Relays 5-110
Connecting Loads 5-110
viE6198B/E6218A Switch/Load Unit User Manual
6Using the Pin Matrix Cards
Using the 32-Pin Matrix Cards 6-2
Conceptual Overview 6-2
Features 6-2
Detailed Block Diagram Descriptions 6-3
Differences Between the Cards 6-3
Features Common to Both Cards 6-3
Relay Timer 6-7
Column Disconnect Relay Control 6-7
OAR 6-7
Reset 6-7
Protection Bypass 6-8
Reset State 6-8
User Connectors and Pinouts 6-8
J1 Connector Pinouts 6-9
J1 Instrument Connections 6-11
P2 Connector Pinouts 6-14
Installing in the Switch/Load Unit 6-15
Using the 64-Pin Matrix Cards 6-16
Conceptual Overview 6-16
Features 6-16
Detailed Block Diagram Descriptions 6-17
Differences Between the Cards 6-17
Features Common to Both Cards 6-17
OAR 6-21
Reset 6-21
Protection Bypass 6-21
Reset State 6-22
User Connectors and Pinouts 6-22
J1 Connector Pinouts 6-23
P2 Connector Pinouts 6-27
Installing in the Switch/Load Unit 6-29
Instrument Multiplexer (Agilent E8792A Only) A-13
General Specifications (Agilent E8792A and E8793A) A-13
Relay Characteristics (Agilent E8792A and E8793A) A-14
Agilent E8782A and E8783A Specifications A-15
Instrument Multiplexer (Agilent E8782A Only) A-15
General Specifications (Agilent E8782A and E8783A) A-15
Relay Characteristics (Agilent E8782A and E8783A) A-16
Relay Life A-16
End-of-Life Detection A-16
Replacement Strategy A-17
BRegister Definitions
Address Space B-2
Base Address B-3
Switch/Load Unit Register Definitions B-4
Card Type (R)Base + 0
Card Configuration (R) Base + 1
Status Register (R) Base + 2
Fixture ID (R) Base + 3
Digital Input (R) Base + 4
DAC1 Output MSB (W) Base + 8
h B-5
h B-5
h B-5
h B-6
h B-6
h B-6
viiiE6198B/E6218A Switch/Load Unit User Manual
DAC1 Output LSB (W) Base + 9
h B-6
DAC Scaling B-7
Control Register (W) Base + A
Open Collector Output (W) Base + B
Digital Output (W) Base + C
DAC2 Output MSB(W) Base + D
DAC2 Output LSB (W) Base + E
Figure 3-1. Agilent E6198B Standalone Switch/Load Unit Front View 3-4
Figure 3-2. Agilent E6198B Standalone Switch/Load Unit Back View 3-4
Figure 3-3. Digital Input Example 3-6
Figure 3-4. Digital Output Example 3-6
Figure 3-5. Switch/Load Unit Block Diagram 3-8
Figure 3-6. E6198B Integrated SLU Back View 3-9
Figure 3-7. Switch/Load Unit Backplane Connectors (USB port adaptor board Not
Table B-1. Standard Registers B-2
Table B-2. Standard Registers B-2
Table B-3. Switch/Load Unit Registers B-4
Table B-4. DAC Scaling Examples B-7
Table B-5. OCout
Table B-6. Summary of Load Card Register Definitions B-10
States B-8
x
E6198B/E6218A Switch/Load Unit User Manualxv
CRepair Information
Table C-1. Measurement Matrix Relays C-5
Table C-2. Instrument Matrix Relays C-7
Table C-3. Bypass and Disconnect Relays C-7
Table C-4. Miscellaneous Relays C-7
Table C-5. Measurement Matrix Relays for E8782A C-8
Table C-6. Measurement Matrix Relays for E8783A C-10
Table C-7. Bypass and Disconnect Relays for E8782A and E8783A C-12
AGlossary Of Terms
xviE6198B/E6218A Switch/Load Unit User Manual
Agilent TS-5000
E6198B/E6218A Switch/Load Unit User Manual
1
Legal Information
Legal Information 1-2
Warranty 1-2
Technology Licenses 1-2
Restricted Rights Legend 1-2
Service and Support 1-3
Agilent on the Web 1-3
Agilent By Phone 1-3
Agilent Technologies
1-1
1Legal Information
Legal Information
Warranty
The material contained in this document is provided “as is,” and
is subject to being changed, without notice, in future editions.
Further, to the maximum extent permitted by applicable law,
Agilent disclaims all warranties, either express or implied, with
regard to this manual and any information contained herein,
including but not limited to the implied warranties of
merchantability and fitness for a particular purpose. Agilent
shall not be liable for errors or for incidental or consequential
damages in connection with the furnishing, use, or performance
of this document or of any information contained herein.
Should Agilent and the user have a separate written agreement
with warranty terms covering the material in this document
that conflict with these terms, the warranty terms in the
separate agreement shall control.
Technology Licenses
The hardware and/or software described in this document are
furnished under a license and may be used or copied only in
accordance with the terms of such license.
Restricted Rights Legend
If software is for use in the performance of a U.S. Government
prime contract or subcontract, Software is delivered and
licensed as “Commercial computer software” as defined in
DFAR 252.227-7014 (June 1995), or as a “commercial item” as
defined in FAR 2.101(a) or as “Restricted computer software” as
defined in FAR 52.227-19 (June 1987) or any equivalent agency
regulation or contract clause. Use, duplication or disclosure of
Software is subject to Agilent Technologies’ standard
commercial license terms, and non-DOD Departments and
Agencies of the U.S. Government will receive no greater than
Restricted Rights as defined in FAR 52.227-19(c)(1-2)(June
1987). U.S. Government users will receive no greater than
Limited Rights as defined in FAR 52.227-14 (June 1987) or
DFAR 252.227-7015 (b)(2)(November 1995), as applicable in any
technical data.
1-2E6198B/E6218A Switch/Load Unit User Manual
Service and Support
Agilent on the Web
Agilent By Phone
Legal Information1
Any adjustment, maintenance, or repair of this product must be
performed by qualified personnel. Contact your customer
engineer through your local Agilent Technologies Service
Center.
You can find information about technical and professional
services, product support, and equipment repair and service on
the Web: http://www.agilent.com/
Double-click the link to Test & Measurement. Select your country
from the drop-down menus. The Web page that appears next has
contact information specific for your country.
If you do not have access to the Internet, call one of the
numbers in Table 1-1
Table 1-1 Agilent Sales And Support Contact Numbers
AmericasCanada(877) 894-4414
Latin America305 269 7500
United States(800) 829-4444
Asia PacificAustralia1 800 629 485
China800 810 0189
Hong Kong800 938 693
India1 800 112 929
Japan0120 (421) 345
Korea080 769 0800
Malaysia1 800 888 848
Singapore1 800 375 8100
Taiwan0800 047 866
Thailand1 800 226 008
E6198B/E6218A Switch/Load Unit User Manual1-3
1Legal Information
Table 1-1 Agilent Sales And Support Contact Numbers (continued)
Agilent TS-5000
E6198B/E6218A Switch/Load Unit User Manual
2
Safety and Regulatory Information
Declaration of Conformity 2-2
Safety Information 2-3
Safety Summary 2-3
Safety Notice 2-3
General Safety Information 2-3
Environmental Conditions 2-4
Before Applying Power 2-4
Ground the System 2-5
Fuses 2-5
Operator Safety Information 2-5
Safety Symbols and Regulatory Markings 2-6
End of Life: Waste Electrical and Electronic Equipment (WEEE)
Directive 2002/96/EC 2-8
Agilent Technologies
2-1
2Safety and Regulatory Information
NOTE
Declaration of Conformity
The Declaration of Conformity (DoC) for this product is
available on the Agilent Technologies website. You can search
for the DoC by its product model or description at the following
web address:
If you are unable to locate the DoC, please contact your local Agilent
representative.
2-2E6198B/E6218A Switch/Load Unit User Manual
Safety Information
CAUTION
WARNING
WARNING
Safety Summary
Safety Notice
Safety and Regulatory Information2
The following general safety precautions must be observed
during all phases of operation of this instrument. Failure to
comply with these precautions or with specific warnings
elsewhere in this manual violates safety standards of design,
manufacture, and intended use of the instrument. Agilent
Technologies, Inc. assumes no liability for the customer's failure
to comply with these requirements.
A CAUTION notice denotes a hazard. It calls attention to an
operating procedure, practice, or the like, that, if not correctly
performed or adhered to, could result in damage to the product or
loss of important data. Do not proceed beyond a CAUTION notice
until the indicated conditions are fully understood and met.
A WARNING notice denotes a hazard. It calls attention to an
operating procedure, practice, or the like that, if not correctly
performed or adhered to, could result in personal injury or death.
Do not proceed beyond a WARNING notice until the indicated
conditions are fully understood and met.
General Safety Information
This product is provided with a protective earth terminal. The
protective features of this product may be impaired if it is used
in a manner not specified in the operation instructions.
DO NOT OPERATE IN AN EXPLOSIVE ATMOSPHERE. Do not
operate the product in the presence of flammable gases or
flames.
E6198B/E6218A Switch/Load Unit User Manual2-3
2Safety and Regulatory Information
WARNING
WARNING
CAUTION
Environmental Conditions
DO NOT REMOVE RACK PANELS OR INSTRUMENT COVERS.
Operating personnel must not remove any rack panels or
instrument covers. Component replacement and internal
adjustments must be made only by qualified service personnel.
Products that appear damaged or defective should be made
inoperative and secured against unintended operation until they
can be repaired by a qualified service personnel.
The protection provided by the Agilent TS-5000 system may be
impaired if the system is used in a manner not specified by
Agilent.
The Automotive Electronics Functional Test System is designed
for indoor use only. Table 2-1shows general environmental
requirements.
Table 2-1 Environment Requirements
Environment ConditionsRequirements
Maximum Altitude2000 meters
Temperature (Operation)5 °C to 40 °C
Maximum Relative HumidityThe test system is designed to
operate in the range from 5% to 80%
relative humidity (non-condensing).
This product is designed for use in Installation Category II and
Pollution Degree 2, per IEC 61010-1.
Before Applying Power
Verify that the product is set to match the available line voltage
and all safety precautions are taken. Note the external markings
of the instruments described in “Safety Symbols and Regulatory
Markings”.
2-4E6198B/E6218A Switch/Load Unit User Manual
Ground the System
WARNING
WARNING
WARNING
To minimize shock hazard, the instrument chassis and cover
must be connected to an electrical protective earth ground. The
instrument must be connected to the ac power mains through a
grounded power cable, with the ground wire firmly connected
to an electrical ground (safety ground) at the power outlet. Any
interruption of the protective (grounding) conductor or
disconnection of the protective earth terminal will cause a
potential shock hazard that could result in personal injury.
Fuses
Use only fuses with the required rated current, voltage, and
specified type (normal blow, time delay). Do not use repaired
fuses or short-circuited fuse holders. To do so could cause a
shock or fire hazard.
For AC inlet fuse, use only 5A/250 Vac fuse with Time-Lag and
Low Breaking capacity. For other fuse ratings, refer Table 3-11.
Safety and Regulatory Information2
In order to avoid electrical hazards, all system internal fuses must
be replaced by trained and qualified personnel.
Operator Safety Information
Module connectors and Test Signal cables connected to them
cannot be operator accessible.
M
Cables and connectors are considered inaccessible if a tool (e.g.
screwdriver, wrench, socket, etc.) or a key (equipment in a
locked cabinet) is required to gain access to a conductive
surface connected to any cable conductor (High, Low or Guard).
Assure the equipment under test has adequate insulation
between the cable connections and any operator-accessible parts
(doors, covers, panels shields, cases, cabinets, etc.)
Verify there are multiple and sufficient protective means (rated
for the voltages you are applying) to assure the operator will
NOT come into contact with any energized conductor even if
one of the protective means fails to work as intended. For
E6198B/E6218A Switch/Load Unit User Manual2-5
2Safety and Regulatory Information
Safety Symbols and Regulatory Markings
example, the inner side of a case, cabinet, door cover or panel
can be covered with an insulating material as well as routing the
test cables to the front panel connectors of the module through
non-conductive, flexible conduit such as that used in electrical
power distribution.
Symbols and markings on the system, in manuals and on
instruments alert you to potential risks, provide information
about conditions, and comply with international regulations.
Table 2-2 defines the symbols and markings you may find in a
manual or on an instrument.
Table 2-2 Safety Symbols and Regulatory Markings
SymbolsDescription
Safety symbols
Warning: risk of electric shock.
Caution: refer to accompanying documents.
Alternating current.
Both direct and alternating current.
Earth (ground) terminal
Protective earth (ground) terminal
Frame or chassis terminal
Terminal is at earth potential. Used for measurement and
control circuits designed to be operated with one terminal at
earth potential.
Switch setting indicator. O = Off, | = On.
2-6E6198B/E6218A Switch/Load Unit User Manual
Safety and Regulatory Information2
Table 2-2 Safety Symbols and Regulatory Markings (continued)
SymbolsDescription
Standby (supply); units with this symbol are not completely
disconnected from ac mains when this switch is off. To
completely disconnect the unit from ac mains, either
disconnect the power cord, or have a qualified electrician
install an external switch.
Regulatory markings
This text indicates that the ISM device complies with
Canadian ICES-001.
Cet appareil ISM est conforme à la norme NMB-001 du
Canada.
The CSA mark is a registered trademark of the Canadian
Standards Association. A CSA mark with the indicators “C”
and “US” means that the product is certified for both the
U.S. and Canadian markets, to the applicable American and
Canadian standards.
The C-tick mark is a registered trademark of the Spectrum
Management Agency of Australia. This signifies compliance
with the Australia EMC Framework regulations under the
terms of the Radio Communication Act of 1992.
This instrument complies with the WEEE Directive
(2002/96/EC) marking requirement. This affixed product
label indicates that you must not discard this
electrical/electronic product in domestic household waste.
The CE mark is a registered trademark of the European
Community. This CE mark shows that the product complies
with all the relevant European Legal Directives.
E6198B/E6218A Switch/Load Unit User Manual2-7
2Safety and Regulatory Information
End of Life: Waste Electrical and Electronic Equipment (WEEE) Directive
2002/96/EC
This product complies with the WEEE Directive (2002/96/EC)
marking requirement. The affixed product label (see below)
indicates that you must not discard this electrical/electronic
product in domestic household waste.
Product Category:
With reference to the equipment types in the WEEE directive
Annex 1, this product is classified as a “Monitoring and Control
Instrumentation” product.
Do not dispose in domestic household waste
To return unwanted products, contact your local Agilent office,
or see:
http://www.agilent.com/environment/product
for more information.
2-8E6198B/E6218A Switch/Load Unit User Manual
Agilent TS-5000
E6198B Switch/Load Unit User Manual
3
Switch/Load Unit and Plug-In Cards
E6198B Standalone/Integrated Switch/Load Unit Rating 3-2
E6198B Standalone Switch/Load Unit Description 3-3
E6198B Integrated Switch/Load Unit Description 3-9
Backplane And Breakout Board Connectors and LEDs 3-10
Differentiating E6198B Standalone Option and System Integrated
Option 3-22
This chapter gives an overview of the Agilent E6198B
(Standalone and Integrated) Switch/Load Unit (SLU), load
cards, and other associated equipment.
Agilent Technologies
3-1
3Switch/Load Unit and Plug-In Cards
E6198B Standalone/Integrated Switch/Load Unit Rating
Electrical
Tab le 3- 1 Switch/Load Unit Rating (Electrical)
ParameterValue
Input Voltage100–240 VAC nominal
Frequency50/60 Hz
Power325 W maximum
Mechanical
Tab le 3- 2 Switch/Load Unit Rating (Mechanical)
ParameterValue
Width484.64 mm
Height373.61 mm
Depth520.52 mm
Weight15.0 kg
3-2E6198B/E6218A Switch/Load Unit User Manual
Loading...
+ 316 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.