The information contained in this booklet
i3 intended for the operation and maintenance of Hewlett-Packard equipment and
is not to be used otherwise or reproduced
without the writtenconsent
Packard Company.
PAL0
of
the Hewlett-
ALTO,
CALIFORNIA, U.S.A.
'
15OAOO1-1
INTERNAL SWEEP:
MAGNIFICAT ION:
SYNCHRONIZATION:
SYNC CONTROL:
150A SPECIFICATIONS
SWEEP GENERATOR
24
calibrated ranges provide sweep speeds from 0.1
psec/cm to 5 sec/cm; accurate to within
A vernier control provides continuous adjustment of
sweep speed between calibrated ranges and extends
slowest sweep to 15 sec/cm.
5 calibrated ranges, X1, X5, X10, X50 and XlOO magni-
fies center portion of unmagnified trace; increases
fastest sweep speed to
magnification and
ranges retain accuracy of original sweep.
Internal, from line power or vertical input signal which
causes 1/2 cm or more vertical deflection. External,
either capacitive or direct coupled with 1/2V p-p or
more.
Sweep can be triggered from either
negative going voltage; the triggering voltage level of
external sync signals
-30
to
+30
volts.
Switch position automatically provides optimum sync
stability for majority of uses.
2p
.02
XlOO
is
psec/cm except lp X50
magnification. X1 and X5
continuously adjustable from
39%.
a
positive- or
a
SINGLE SWEEP:
SAWTOOTH OUTPUT:
GATE OUTPUT:
BAND WIDTH:
.
SENSITIVITY:
INPUT IMPEDANCE:
OUTPUT:
Switch in top access provides single-sweep operation.
+20
to
-20
volt sawtooth output available from connec-
tor in top access.
20-volt positive pulse for duration of sweep available
from connector in top access.
HORIZONTAL AMPLIFIER
Direct current to 500 kilocycles.
5 calibrated ranges provide sensitivities from
volt/cm to 5 volts/cm. A vernier control provides
continuous adjustment between calibrated ranges and
extends the minimum sensitivity to 15 volts/cm.
1 megohm shunted by
CALIBRATOR
1000-cycle square wave having 1-psec rise and decay
time available
18 calibrated ranges provide from
volts peak-to-peak, accurate to within
at
40
pfl.
front-panel connector
0.2
.
millivolt to 100
370.
0.2
I
I
,
I
I
,
!
t
I
!
TYPE:
CATHODE RAY TUBE
5 AMP-
P1,
potential
rr-cno-accelerator,
P2,
P7
.
or P11 screen. 5000-volt accelerating
flat face, available with
\
I
i
I
1
I
DEFLECTION PLATE
CONNECTIONS:
INTENSITY MODULATION:
REPLACEABILITY:
POWER REQUIREMENTS:
SIZE:
WEIGHT:
BANDWIDTH:
SENSITIVITY:
Both screw and pin-type terminals in top access re-
or
ceive wires
tion to plates.
Terminals inside top access to receive +20 pulse for
blanking CRT trace of normal intensity.
CRT bezel removes with
graticule or CRT from the front panel.
CRT bezel provides firm mounting for standard oscilloscope cameras
115/230 VAC
watts.
13-1/2" wide
Net weight
@l52 DUAL TRACE AMPLIFIER
Direct Current to
microsecond.
9
calibrated ranges provide sensitivities from
20 volt/cm, 57'0. Vernier control provides continuous
adjustment between calibrated ranges and extends the
minimum sensitivity to 50 volts /cm.
special connector assembly for connec-
.
+lo%,
x
17-1/411
80
pounds; shipping weight
50/60
high x 25" deep.
10
megacycles; rise time 0.035
150
twist for replacement of
cycles, approximately500
104
pounds.
0.05
to
INPUT IMPEDANCE:
I
I
,
1
I
I
I
i
DUAL TRACE
PRESENTATIONS
1'
i
!
I
i
i
I
I
i
i
I
1
megohm shunted by
Simultaneous traces obtained either by alternate sweep
or
ing
ping lines between traces are blanked.
Channels are completely independent and have identical operating controls
Pos-Up-Pos Down switch provided.
by 100-kilocycle chopping,
40
.
ppf
each channel.
as
selected. Chop-
TABLE OF CONTENTS
FOR
MODEL 150A AND MODEL 152A
HIGH FREQUENCY OSCILLOSCOPES
Par
1-1
1-2
1-3
1 -4
1-5
1-6
2-1
2 -2
2-2A
2-2B
2-2c
2-2D
2-2E
2-3
2 -4
2-5
3-1
3-2
3 -3
3 -4
3-5
3-6
3-7
3-8
-9
3
3-10
3-1 1
3-12
3-13
4-1
a
.
SECTION
GENERAL DESCRIPTION
General Information.
Damage
Power Line Voltages
Power Cord
Cooling
Overload Relay
OPERATING INSTRUCTIONS
High Voltage Time Delay
Controls and Terminals
Internal Sweep Controls
Horizontal Input Controls.
Dual Channel Amplifier
Controls
CRT Panel Controls
Top Access Controls
Operating Procedures
'Vertical
Use of Chopped or
Alternate Sweep
THEORY
General Content
Overall Operation
Synchronizing Circuit
Sweep Generator
Feedback Integrator and
Replacing the Air Filter
Removing the Cabinet
Connecting for 230-Volt
Power Lines
Servicing Etched Circuits
Tube Replacement Chart
Isolating Troubles to
Major Sections
Adjusting the Low-Voltage
Power Supplies
Adjusting the High-Voltage
Power Supply
Replacing and Adjusting
theCRT
Adjusting the Calibrator
Adjusting the Main
Horizontal Amplifier.
Main Horizontal Gain
Adjustment
Calibration
Centering and X1 Centering.
Frequency Response
5-AMP mono-accelerator type cathode ray
tube, with unitized, plug-in construction for
maximum accessibility and flexibility. It
can be used with either internal or external
sweeps which can be either internally or externally synchronized. The horizontal amplifier incorporates magnification circuitry
is
which
sweeps up
sweep range extends from .02 bsec/cm to
15 sec/cm.
The Model 150A has been designed for use
with
units to perform many different functions
but for the purpose of explaining scope operation in this manual,
the Model 152A Dual Channel Amplifier
employed, Since the control layout on the
front panel has been carefully organized and
labeled for convenient operation, most controls will be self-explanatory. However
major controls will be discussed completely
in the operating section of the manual. A
large fold outview of the panel
Section I1 with
s
cr ibed
capable of expanding basic internal
to
100 times. Thus the internal
a
variety of vertical amplifier plug-in
it
will be assumed that
is
included in
all
operating controls de-
.
is
is
The three conductor power cable supplied
withthe instrument
a
ized three prong male connector recommend-
ed by the National Electrical Manufacturers'
Association. The third contact
round pin added to
plug which grounds the instrument chassis
when used with the appropriate receptacle.
To use this plug in
outlet an adapter should be used to connect
the NEMA plug to the two contact system.
When the adapter
tion becomes
which can be connected to the outlet mounting
box for the protection of operating personnel.
,
1-5
COOLING
The Modell5OA employs a forced draft cooling system to maintain satisfactory operating
temperatures within the case. The
and
air
filter
case, and adequate cooling will take place
,
long
as
the case
ing obstructs the filter. Generally, the height
of the cabinet feet provide such clearance.
Thus the Model150A can be used in
fined bench set-up
of the cabinet
tures are not extreme.
is
is
is
terminated in a polar-
is
a
standard two-blade ac
a
standard two contact
is
used the ground connec-
a
short lead from the adapter
located under the instrument
is
on the instrument and noth-
as
long
as
the underside
clear, and ambient tempera-
an offset,
air
intake
as
a
con-
1-2
DAMAGE IN TRANSIT
After unpacking this instrument should any The Model
shipping damage become evident, refer to the reduces the output voltages
1lClaim for Damage in Shipment" paragraph ply when:
on the warranty sheet in this manual.
1-3
POWER LINE VOLTAGES
The Oscilloscope, like other @ instruments,
is
shippedfrom thefactory wiredfor 115 volt
ac line operation unless otherwise specified
in the order. However, the instrument may
also be operated from a 230 volt ac line source
if
the proper conversion
transformer. This conversion
is
described in the Maintenance section.
is
made to the power
is
simple, and
1-6
OVERLOAD RELAY
150A
1.
)
Any series tube on the d'-c heater
string
strument
2.)
A plug-in vertical amplifier
removed while the instrument
is
on.
3.)
A short circuit or excessive
loading of any positive-voltage
supply occurs.
has an overload-relay which
is
removed while the in-
is
on.
1-1
of
the power
SUP-
is
i
SECTION
OPERATING INSTRUCTIONS
2-1
HIGH VOLTAGE TIME DELAY
The Model 150A applies voltage only to
tube filaments for about
it
turn
warm up before tube plate voltages and CRT
high voltages are applied.
is
recycle before the instrument returns to operation.
2-2
The front panel arrangement and operating
controls are shown in the fold out illustration, Figure 2-1. The description of the operating controls should enable you to operate
the instrument
of oscilloscope technique. However,
plete description of control function follows:
on. This delay provides an adequate
turned off for any reason this delay will
CONTROLS AND TERMINALS
if
you have a basic knowledge
30
seconds after you
If
the instrument
2-ZA INTERNAL SWEEP CONTROLS
SWEEP TIME/CM
Amulti-position switchwhich selects any one
of twenty four calibrated sweep speeds from
0.1 pec/cm to5 sec/cm. This switchis
sociated with a concentric VERNIER which
provides continuous adjustment of sweep
speed between steps, and in the full counterclockwise position extends the sweep time to
is
15 sec/cm. When the VERNIER
clockwise position (GAL),
cuit and the sweep time
k370
as
read
on
the SWEEP TIME/CM dial.
INT. SWEEP MAGNIFICATION
Afive position magnification control,
X10, X50, and X100. The Xlposition provides
calibrated sweep speeds
TIME /CM switch. Other positions indicate
degree of expansion taking place about the
vertical centerline on the scope graticule.
Whenever the sweep
the SWEEP MAGNIFIED lamp will light. The
fastest calibrated sweep time of
is
obtained by setting the SWEEP TIME/CM
switch to
.1
or
.2
psec/cm and setting the
it
is
is
calibrated within
as
read onthe SWEEP
is
magnified beyond X1
in the full
out of the cir-
0.02
pec/cm
a
X1,
its
com-
as-
X5,
II
magnifier on X5 or
exceed the fastest calibrated sweep time, the
MAGNIFIER UNCALIBRATED reminder lamp
will light. It will also light with 1 psec/cm
X50 and 2 psec/cm X100.
SYNC
Afour position switch which enables the sweep
to be triggered either internally or externally. Internal triggering can be accomplished
a
from
applied vertical input signal of sufficient amplitude to produce
nal triggering can be effected by signals having amplitudes greater than
the low frequency cut
sync input circuit
it
low frequency sync signals. Concentric with
the SYNC switch
SWEEP MCDE
A
bias of the sweep generator. As the control
is
tion, the sweep generator will pass from
free running (FREE) condition to a condition
where triggering
the extreme counterclockwise position the
control switches into
PRESET provides optimum triggering bias
for nearly
TRIGGER LEVEL
A continuous control which selects the ampli-
tude level of the sync signal where trigger-
ing occurs. For external synchronizing sig-
nals therange of this controlis from
-30
at
complex waveforms. See Figure 2-2.
TRIGGER SLOPE
A two position switch concentric with TRIGGER LEVELwhich permits triggering to occur on either the positive
of internal, external or line voltage sync signals. See Figure 2-2.
EXT
A
which are connected in parallel and receive
external synchronizing signals.
line voltage signal or (INT) from an
is
advisable to use
continuous control which adjusts the input
rotated from the extreme clockwise posi-
all
sync applications.
volts, allowing the sweep to be triggered
any point within this range on simple or
SYNC
pair of binding posts and a BNC connector
INPUT
X1O
respectively. If you
a
5mm deflection. Exter-
0.
5 volt. Since
off
of the ac coupled
is
approximately 200 cps,
EXT
DC for external
is
the SWEEP MODE control.
is
required (TRIGGER). At
a
PRESET position.
or
negative slope
+30
to
a
2-1
2-2C
DUAL
VERTICAL PRESENTATION
A four position switch which selects the desired vertical channel or desired dual channel presentation.
CHANNEL AMPLIFIER CONTROLS
This switch
is
labeled:
FIGURE 2-2
2-2B
HORIZONTAL INPUT CONTROLS
EXT HORIZONTAL INPUT
a
A pair of binding posts and
which are connected in parallel and receive
externally applied sweeping signals to drive
the horizontal amplifier. The amplifier pass
is
band
from dc to
500
KC.
BNC connector
A ONLY
B ONLY
ALTERNATE
CHOPPED
POLARITY
A four position switch which allows the input
signal polarity to be displayed either up or
down on the scope, and which selects either
AC or
VERTICAL POSITION
A continuously adjustable control, concentric
with POLARITY, which adjusts the vertical
position of the input signal. One for each
channel.
DC
The output of the A channel
applied to the vertical amplifier.
The output of the B channel
applied to the vertical ampli-
fier.
The two channels are alter-
at
nated
paragraph 2-5.
The two channels
at
a
KC.
composed of five microsecond
segments spaced five microseconds apart. See paragraph
2-5.
coupling. One for each channel.
the sweep rate. See
are
switched
free running rate
The
two
traces are each
of
100
is
is
EXT INPUT VOLTS/CM
A five position attenuator calibrated from .2
volts/cm to 5 volts/cm which establishes the
input
This switch
VERNIER which provides continuous ad just
ment between steps and reduces the input
sensitivity to 15 volts /cm in the full counterclockwise position. When the VERNIER
the full clockwise (CAL) position,
the circuit and the VOLTS/CM switch
brated with
HORIZONTAL POSITION
A multi-turn control which changes the horiz-
ontalpositionof the trace. It covers the range
in several turns to let you make fine adjustments under high magnifications of an internal
sweep.
sensitivity of the horizontal amplifier.
is
associated with a concentric
is
it
is
out of
is
cali-
+30/0.
in
VOLTS /CM
A nine step attenuator which selects the input
sensitivity of the vertical amplifier.
tivities range from. 05 trolts/cm to 20 volts/
a
cm in
-
VERNIER
‘CM
for each channel.
A-BAL and B-BAL
Screwdriver adjustments for adjusting the dc
balance of each vertical amplifier.
ure
A-CAL and B-CAL
Screwdriver adjustments for setting the gain
of each vertical amplifier.
2-2
5-1-2 sequence.
is
in the CAL position the VOLTS/
attenuator is calibrated within
2-11.
Whenthe concentric
Sensi-
+3oJO.
One
See Fig-
2-2D
CRT PANEL CONTROLS
FOCUS
01s
Contr
trace resolution,
RESET
A terminal which can receive an external
pulse to rearm the single sweep circuit. Pulse
characteristics required
peakwithno overshoot;
=
+15 to +25 volts
2
to 4 pseconds width.
INTENSITY
Controls trace brightness.
SCALE LIGHT
Adjusts brightness of graticule lines.
CALIBRATOR
a
at
1000
the
rise
A multi-position switch which adjusts
cps square wave from 0.2 millivolts to 100
volts in
+3
of
adjacent terminal and can be used to calibrate the deflection sensitivity of the vertical
and horizontal amplifier
and decay times are each less than 1 micro-
second. The square wave output may also be
used to adjust the divider probe for frequency
response.
a
2-5-10 sequence with an accuracy
70.
The output voltage appears
.
Square wave
2-2E TOP ACCESS CONTROLS
The following controls and terminals are accessible through the top access door of the
instrument cabinet.
Horizontal and vertical deflection plates.
A terminal for CRT intensity (Z-axis) modulation
FUSE
Line fuse
Thermal cutout with resetting button.
2-3
OPERATING PROCEDURES
Basic operating procedures are described by
illustrations
steps in eachcase. The
are complete. Others are arranged to supplement the
tions possible in using the oscilloscope.
index to these illustrations follows:
Capacitive Coupling to CRT Plates
Single Sweep Operation
Delayed Sweep Operation
Vertical VOLTS /CM Calibration
External Intensity Modulation
--
Internal Sync
--
External Sync
--
Dual Trace
CRT
Plates
An
SAWTOOTH OUT
A terminal which provides
voltage corresponding to the sweep.
GATE OUT
A terminal which provides
voltage for the duration of the sweep.
SINGLE-NORMAL
A switch which selects either normal or sin-
gle sweep operation. In single sweep posi-
tion the circuits are arrangedto sweep once
a
after
tive until manually or
The ARMED lamp lights when the single sweep
circuit
trigger signal and then remain inopera-
is
ready for a trigger.
a
sawtooth output
a
positive gate
electronically reset.
2-4
VERTICAL AC OR DC COUPLING
Under most conditions AC coupling will be
used.
without regard for the dc levels involved. In
the AC position the input signal (vertical or
external sync)
through
component from the input wave. This coup-
ling circuit has
cps; however, to avoid degrading input pulses
or square waves below 10 cps
to use dc coupling.
When you want to look
to
rundown in circuit work, or in mechanical
work where the output from
a
It permits high gain to be employed
is
coupled to the instrument
a
capacitor which removes the dc
a
low frequency cut off
it
is
advisable
at
waveforms relative
a
dc level; for example, observing a Miller
a
transducer has
dc component; use dc coupling.
2-3
at
2
2-5
USE
OF
CHOPPED
OR
ALTERNATE SWEEP
CHOPPED and ALTERNATE VERTICAL PRG
two
SENTATIONare used topresent
electrical phenomona, which are related in
frequency or rate of recurrence, to the
cilloscope CRT for simultaneous viewing.
ALTERNATE may be used whenever sweep
time and rate of recurrence are rapid enough
is
and screenpersistence
vent objectionable flicker. CHOPPED PRESENTATION should be used
NATE position
on the screen as the inputs are switched after
a
flickering effect
long enoughto pre-
if,
separate
os-
in the ALTER-
is
noticeable
each sweep. When two related signals are
being presented to the oscilloscope for
multaneous viewing using either CHOPPED
or ALTERNATE VERTICAL PRESENTATION
the 150A should be synchronized externally.
Insome cases
different input waves not related infrequency.
In this instance the 150A may be synchronized
internally, triggering automatically
one signal then from the other
is
dure
may cause noticeable deterioration of the
quality
may be improved byadjusting the VERTICAL
POSITION controls
not ordinarily recommended since
of
it
may be desirable to view two
first
.
This proce-
the presentation. Quality generally
so
that the traces overlap.
si-
from
it
2-4
INTERNAL SWEEP - INTERNAL SYNCHRONIZATION
-0-
I
1.
Place vertical input signal into plug-in am-
plifier.
2.
If dual trace amplifier
CAL PRESENTATION selector to
put; assume one input is used.
3
.
Adjust VOLTS/CM selector for desired sensitiv it
y
.
4.
Set SYNC selector to INT.
is
used, set VERTI-
A
or
B
in-
FIGURE
5.
Set HORIZONTAL SENSITIVITY to INT
x1.
6.
Select desired sweep speed.
7.
Set TRIGGER LEVEL to zero.
8.
Set TRIGGER SLOPE for triggering on positive or negative slope
sired.
9.
Set SWEEP MODE to PRESET.
2-3
of
input wave
SWEEP:
as
de-
INTERNAL SWEEP - EXTERNAL SYNCHRONIZATION
1.
Place vertical input signal into plug-in am-
plifier.
2.
If dual trace amplifier
CALPRESENTATION switch toproper input.
Assume one input
3
.
Place sync signal into EXT SYNC INPUT terminals
4.
Place HORIZ SENSITIVITY to INT SWEEP X1.
.
is
is
used (A or
used, set VERTI-
B).
FIGURE
5.
Select desired sweep speed with SWEEPTIME/CM switch.
6.
Set TRIGGER LEVEL to zero.
7.
Adjust SYNC selector to AC or DC
quired (see paragraph
8.
Place SWEEP MODE in PRESET.
2-4
2-4).
as
re-
INTERNAL SWEEP MAGNIFICATION
I
1.
Select sweep speed with SWEEP TIME/CM
switch.
Place
2.
of SWEEP TIME/CM switch
3.
Set HORIZ SENSITIVITY to INT SWEEP X1
(X1
4.
Adjust horizontal position of
want
ticular wave in
nification. UNCAL lamp stays off)
VERNIER in CAL when direct reading
is
the unmagnified sweep position).
to
magnify a portionof a wave or a par-
a
train, place wave
SWEEP TIME/CM
MAGNIFICATION
Actual sweeptime/cm=
.
Olpsec / cm;uncalibr ated lamp will light. In-
crease sweep time or reduce degree of mag-
=
=
is
desired.
of
G
If you
inter-
X50=
trace.
.5 psec/CM
X50 MAGNIFICATION
psec/cm
.5
est under vertical graticule center line with
HORIZ POSITION control.
5. Switch HORIZ SENSITIVITY to desired magnification (X5 or above).
6.
SWEEP MAGNIFIED indicator will light.
7.
If
combination of sweep
magnification exceeds maximum calibrated
sweep speed
FIER UNCALIBRATED indicator will light.
SWEEP TIME/CM
Actual Sweep Time
(Sweep
time
>
.02pec/CM,MAGNIFIER
speed and degree of
(.02
peconds/cm) the MAGNI-
=
.5 psec/CM
=
X10
=
.5 psec/CM $(XlO)
=
.05
psec/CM
FIGURE
2-5
EXTERNAL HORIZONTAL INPUT
(if
VLINILI
1.
Insert external horizontal signal.
2.
Select desired
side
of
HORIZONTAL SENSITIVITY control,
sensitivity
on
EXT
INPUT
FIGURE
3.
Adjust horizontal position.
4.
Turn SYNC selector to an EXT position.
5.
For vertical input see Figures
2-3
2-6
and
2-4.
VERTICAL INPUT - DUAL TRACE
1.
Set to A only.
2.
Place one of two input signals intoINPUT A.
3.
Adjust deflection sensitivity with VOLTS /CM
switch. When VERNIER reads CAL, accuracy of VOLTS/CM switch
4.
Select polarity of presentation.
negative or positive portion of input wave
displayed up or down.)
is
+3%.
(
Whether
is
FIGURE
Adjust vertical position of trace near upper
(or lower) portion of scope face.
5.
Switch to INPUT B and repeat above procedure on
6.
Switch to ALTERNATE or CHOPPED for dual
trace presentation.
B
side of panel.
(See paragraph
2-7
2-5.
)
..
Connect theAC-21AProbe to the desired ver-
lo
tical input, and set the VERTICAL PRESENTATION selector to the corresponding input.
Set the CALIBRATOR selector to
2.
vertical VOLTS/CM selector to
Set the SWEEP TIME/CM selector to. 5 MIL-
3.
2.
.05.
LISECOND/CM; set the SYNC selector to INT,
the SYNC control to PRESET.
Touch the probe to the CALIBRATOR connector
4.
and observe the 1-kc square wave.
I
Set the
.-
Loasen probe locknut by unscrewing.
5.
Tune probe to obtain flattest top on square
6.
wave by turning rear flange
Tighten locknut to retain adjustmenti
7.
on
probe.
t:
/
II
'
FIGURE
2-8
REMOVING CRT BEZEL
LOCK
RELEA
The CRT bezel contains the filter and graticule shields;
cept oscilloscope camera equipment without
adapters of any kind.
cathode ray tubes the bezel must be removed,
it
has also been designed to ac-
To
change filters or
FIGURE
To
remove bezel:
1
.
Depress lock release button, and twist bezel
counter clockwise about 15 degrees
2.
Pull bezel straight from panel
2-9
.
as
shown.
ALIGNING SCOPE TRACE
WITH
GRATICULE
Alignment
complished through access hatch without removing instrument from case.
of
CRT
with graticule may be ac-
FIGURE
(2)
Fiber lever
itudinal positioning
clamp
(l),
controls both radial and
using screwdriver.
2-10
of
CRT
and
is
locked by
long-
VERTICAL BALANCE ADJUSTMENT
Procedure given for B channel, but
A
for
1. Set SWEEP MODE control to FREE RUN.
2
channel.
.
Set VERNIER to minimum (full counterclock
.
wise)
is
same
FIGURE
3
.
Center trace with VERTICAL POSITION
control.
4.
Switch VERNIER to
5.
Adjust B-BAL control to return trace to
center of scope.
2
Repeat steps
2-11
to 5 until trace remains centered.
CAL
(full clockwise).
DIRECT CONNECTION
//
EXTERNAL INPUT SIGNAL
I
TO
DEFLECTION
PLATES
'R282
IM
CAUTION
oscilloscope operate
approximately +250 volts. Therefore there
canbe no commonchassis connection between
the signal source and the oscilloscope. In
most cases the signal source chassis
assume the deflection plate potential. Turn
off the
to deflection plate terminals.
To connect an external signal directly to ver
tical deflection plates:
A. Balanced Input
1,Connect signal to terminals A and
-
The deflection plates of the
at
a
d-c potential of
will
'
instrument before making connection
,
E.
FIGURE
-
2.Relocate leads from
as shown by dashed lines.
B.
Single-Ended Input
1.Connect the signal to A for + up deflection (to E for + down deflection).
2.
Connect an appropriate bypass capacitor betweenC and
(between C and Afor
Signal source return connects to
A, whichever is bypassed.
3.
Relocate leads from
as
shown by dashed lines.
2-12
F,
I,
D3
E
for t up deflection
+
down deflection).
F,
I,
D3
and
and
E
D4
or
D4
CAPACITIVE CONNECTION
EXTERNAL
INPUT
SIGNAL^
.
(SINGLE-
/
TO
DEFLECTION
-R282
IM
PLATES
CAUTION
-
Turn off instrument before
making connection to deflection plate terminals.
ac
couple a signal to the vertical deflection
To
plate
:
A. Single-Ended Input
1.
Connect external signal to B for t up
deflection (to D for
and the signal return to
2.
Connect an appropriate bypass capaci-
tor betweenC and E for
t
down deflection)
C.
t
up deflection
(between C and Afor t down deflection).
3. Connect an appropriate d-c blocking
FIGURE
capacitor betweenA and B for
flection (between D and
deflection).
F,
4.Relocate leads from
as
shown by dashed lines.
I,
B. Balanced Input
1.
Connect external balanced signal to
minals B and D.
2.
Connect appropriate d-c blocking
pacitors between terminals A and B
and between terminals
3.Relocate leads from
as
shown
by dashed lines.
D and
F,
I,
2-13
t
E
up de-
for t down
D3, and D4
E.
D3, and D4
ter-
ca-
SINGLE SWEEP OPERATION
PP
THERMAL OVERLOAD SWITCH
(MANUAL RESET
1
1.
Set selector inside top access to SINGLE.
2.
Select sweep speed.
3.
Set SYNC switch to INT. and SWEEP MODE MODE control
to PRESET.
4.
Select desired vertical sensitivity.
5.
Adjust TRIGGER LEVEL to ZERO and
SLOPE
it
TRIGGER LEVEL in NORMAL operation.)
as
may be convenient to establish desired
desired.
(In
some applications
FIGURE
6.
Switch SWEEP MODE control out ofPRESET and back to PRESET to arm sweep
circuit
7.
Indicator should light
When input signal
fire
once, indicator
and sweep will remain locked out until re-
set
as described in step
8.
Resetting can be accomplished by placing
1-4
pec
into RES ET terminal
2-14
.
width,
as
soon
as
is
returned to PRESET.
is
received, sweep will
(7)
will extinguish,
6.,
or
t15
to
t25
volt peak pulse
.
SWEEP
4
DELAYED SWEEP OPERATION
D
OUTPUT
DELAY GENERATOR
-hp-
2
OR
I2
INPUT
J3
1.
Set selector inside top access to NORMAL.
2.
Select sweep speed.
3.
Set SYNC switch to INT and SWEEP MODE
to PRESET.
4.
Select desired vertical sensitivity.
5.
Adjust TRIGGER LEVEL and SLOPE for
desired triggering.
6.
Switch selector to SINGLE.
7.
Connect output of DELAY GENERATOR
to RESET terminal inside top
access.
8.
Connect signal to be observed to scope in-
(Channel A or
put.
9.
Connect trigger source to input of delay
generator. Trigger-source signal in some
applications may be same
observed(step
signal. This signal serves
sweep retriggering source. (See SINGLE
SWEEP OPERATION.
10.
It may be desired to have the sweep
immediately
of awaiting the arrival of signal to be observed.
1 through
SWEEP MODE control to FREE RUN.
at
To
accomplish this, follow steps
9,
except that after step
8)
end
B
as
desired.)
as
or
it
may be some other
)
of
delayperiod instead
signal
as
a
to
be
single
start
6,
set
FIGURE
2-
15
VERTICAL VOLTS/CM CALIBRATION
This procedure assumes that vertical . alance
adjustments (Figure 2- 11) have been made.
1.
Connect CALIBRATOR output (or any accurate ac voltage) to the vertical INPUT
(assume B in this case).
2. Set CALIBRATOR to the .2 volt position
if
2 volt position
(use
AC-21A probe
used).
3.
Set VOLTS/CM switch to
.05,
and switch
FIGURE
VER
TIER to CAL.
4.
Sync scope internally.
5.
Use convenient sweep speed (app. .5 MS/
CM).
6.
is
Adjust R589 (accessible through frontpanel
as shown) to obtain a verticaldeflection of
exactly
4
centimeters on scope graticule.
Repeat procedure for channel A.
2-16
EXTERNAL INTENSITY MODULATION
I
To intensity
signals
:
1.
Set
INT.
modulate the CRT with external
Z
-
EXT. Z switch to EXT.
FIGURE
Z.
Connect modulating signal to input ter-
2.
minals.
peak will
normal intensity.
brighten the trace.
2-17
A
positive voltage of
blank
the CRT trace from
A
negative input will
20
volts
1
INT.
SYNC
I
VERTICAL
CRT
PLATES
\
HORIZONTAL
AMPLIFIER
HORIZONTAL
FIGURE
CAT
PLATES
3-1
TO
ASTlCMATlSY
3
-0
SECTION
111
THEORY
3-1
GENERAL CONTENT
This section contains a brief description of
the overall operationof the 150A Oscilloscope,
descriptions of each major section and detailed explanations of the Feedback Integrator
Sawtooth Generator andSchmitt Trigger. The
is
description for each major section
ported by
voltage
and when necessary, a switch detail diagram,
at
the rear of the manual. The material in
this section
3-2 Overall Operation and Circuit
3-3 Synchronizing Circuit
3-4 Sweep Generator
3-5 FeedbackIntegrator and Diode
3-6
3-7
3-8 Main Vertical Amplifier
3-9
3- 10 Calibrator
3-
3- 12 Regulated High-Voltage Power
3-13 Single Sweep And Delayed
3-2
OVERALL OPERATION
The simplified block diagram in Figure 3-1
shows the basic signal circuits in the Model
150A Oscilloscope: the
Horizontal Amplifier, Sweep Generator and
Cathode Ray Tube.
a.
Vertical Amplifier - The verticalamplifier receives the input signal, amplifies
it,
and drives the vertical deflection plates.
It
provides attenuation of the input signal;
determines the direction of spot deflection
a
for
the vertical positionof the spot on the screen;
The complete Vertical Amplifier circuit
in three separate parts: the final amplifiers,
located on the upper board; the intermediate
amplifiers on the lower board on the instru-
ment chassis,
input-preamplifier attenuators, phase inverters and beam positioners are on the etched
board in the Vertical Amplifier Plug-In Unit.
All connections to the plug-in unit are made
through two multiple-contact connectors
,
the plug-in
connections to the Main Vertical Amplifier
on the instrument chassis are made through
a
tube-socket connector, while the signal path
between the intermediate and power amplifier
stages is made thrbugh the special coaxialcable delay lines.
b. Horizontal Amplifier
amplifier receives the sweep voltage
either from the HORIZONTAL INPUT jack or
from the internal sweep generator, amplifies
it,
and drives the horizontal deflection plates.
It provides attenuation of the HORIZONTAL
INPUT signal or MAGNIFICATION of the internal sweep and determines the horizontal
position of the spot on the screen.
The complete Horizontal Amplifier consists
of the main Amplifier, and
(used only for external signals applied to the
HORIZONTAL INPUT connector). The preamplifier stage includes the VOLTS /CMportion of the HORIZONTAL SENSITIVITY selector switch and
is
set
to one of the MAGNIFICATIONpositions.
The MAGNIFICATION circuit of the HORIZONTAL SENSITIVITY selector
Horizontal Amplifier. The sawtooth sweep
is
applied to the Main Horizontal Amplifier
while external signals
the HORIZONTAL SENSITIVITY switch and
the preamplifier.
The complete Horizontal Amplifier circuit
located on the single etched board on the lefthand, swing-out chassis. Connections to the
chassis
connectors on the board.
are
as
shown in Figure
is
pushed into position. Power
-
The horizontal
a
preamplifier
is
not effective when the switch
is
in the Main
are
first fed through
made through two tube-socket
3-2,
is
the
as
is
-1
3
INPUT
LTTENUATORS
-
VERTICAL
AMPLIFIER
(PLUG-
(51W
TIUIICU
IN
I
)
VtRUltR
10
MAIN
VERTICAL
AMPLIFIER
VI-V6
INTERNAL-SYNC
AMPLIFIER
TRIGGER
-
YlRTlCIL
CRT
-
t4lSVOC
lUIRtCl
t
400
t
260
t
I30
-
82
-
150
UII8RIIOR
VOC
YOC
VOC
VQC
VOC
REGULATED
UIW-VOLTAGE
POWER
SUPPLIES
SEE
mirn
survu
8LOCl
OIIUU
I
SWEEP
GENERATOR
VlO-Vl7
IStl
Otllll
BLOCI:
UITWTU
SlttP
01IORA1I
UUBLIIIIXt PULSt
CRI
INTlMSlll
WWLl
10
INltISlTl
-5OOOVDC
1,,,,1$
HICH
VOLTAGE
POWFR SUPPLY
V27-VJO
d’(lllTtlSITII
PREAMPLIFIER LMPLlFlER
HORIZONTAL
FIGURE
3-2
3-2
L
UORIZQITIL
cnr
-
c.
Sync Circuit - The Synchronizing Circuit
receives a sync signal either from the
Vertical Amplifier for internal synchroniza-
from the EXT. SYNC INPUT connector
tion;
for external synchronization, or from an internal 6.3-volt source for line frequency syn-
chronization. The Sync Circuit amplifies
input signals, determines the input voltage
level and polarity of input sync signal which
start
will
reliable sync pulse for operation of the Saw-
tooth Generator.
a
sweep; and supplies a fast
all
and
switch.
fied and inverted in V8A, or are amplified
without inversion in V8B. The desired polarity of sync signal that
the sweep
plate by the TRIGGER SLOPE control. A
negative-going signal
lowing Schmitt Trigger
positive input-sync signals
from the plate of V8B while negative signals
are
All
incoming sync signals
is
then taken from the appropriate
is
required for the fol-
taken uninverted from
is
tobeusedto
V9;
consequently,
are
taken inverted
V8
A.
are
ampli-
start
-
d. Sweep Generator
receives a negative starting pulse from
the Sync Circuit and generates
be fed to the Horizontal Amplifier. The Feedback Integrator determines the basic sweep
time per centimeter, the Retriggering Bias
Controldetermines the sensitivity of the generator to incoming sync signals and provides
either single or repetitive sweeps. The sweep
generator also supplies an unblanking pulse
a
to the CRT and
sawtooth flyback) to the Dual Channel Verti-
cal
Amplifier Plug-In Unit for ALTERNATE
operation.
The complete Sweep Generator and Sync Cir-
is
cuit
the right-hand, swing-out chassis.
nections to this chassis
three tube-socket connectors and three pin
connectors on the board.
e. The CRT
operated at -5000 volts. The 5AMP may be
obtained with four different phospors:
green medium;
plied with the 150A)
short. All are interchangeable with little
readjustment and the tube
through the front panel. The mono-accelerator anode makes possible
matism adjustment (located inside the access
hole) which requires no resetting when adjusting the FOCUS or INTENSITY. The deflection plate terminals located on the peri-
phery of the tube
movable jumpers directly to the Main Vertical Amplifier.
3-3
The Synchronizing Circuit consists
put Amplifier-Phase Inverter (V8) and
Schmitt Trigger wave shaper
ceives synchronizing signals either from the
EXT. SYNC INPUT connector on the panel,
the Main Vertical Amplifier, or from the line
frequency,
located on the single etched board on
mono-accelerator tube with the cathode
SYNCHRONIZING CIRCUIT
as
timing signal (during each
-
The CRT
2,
green long; (normally sup-
are
selected by the SYNC selector
The Sweep Generator
a
sawtooth to
are
made through
is
a
type SAMP,
7,
blue long; the
is
easily changed
a
simple astig-
connected through
(V9).
All
11,
of
V8
con-
1,
blue
re-
an In-
re-
The TRIGGER LEVEL control shifts
V8's operating potentials in apositive or nega-
so
tive direction
nal meets the trigger-voltage level of the
following Schmitt Trigger with more or less
sync voltage input.
Sync Schmitt Trigger
going input voltage to
tion, and produces
start
by the negative-going portionof the sync pulse,
V9
(which normally follows on
form) before
start
age moved only once in one direction
produce
V9, and no more sweeps would be possible
untilV9 was returned to somewhat beyond
original
tion of the hysteresis in
a
The negative output pulse from V9B
entiated, producing
which
Triggering Sensitivity Ad jus tment R 66 varies
the gain of V9A to vary the hysteresis of the
trigger circuit, thus determining the input
voltage level which will cause the trigger to
switch
Symmetry Adjustment
bias level on V9B to shift the hysteresis
of the trigger
tive direction. This adjustment positions the
hysteresis voltage limits equally above and
below the 0-signal input voltage level. Since
both controls affect the gain of the circuit,
a
adjustment of either one
however, varying the plate load resistance
of V9A varies the degree of hysteresis predominantly, while varying the grid bias on
V9B varies the voltage position of the hyster-
esis
the Sweep Generator. Once triggered
must be reset by a positive-going voltage
another sweep.
a
single sweep, but would not reset
state
starts
state,
area
predominantly.
an amplified input sync sig-
V9
requires a negative-
start
the desired opera-
a
reliable output pulse to
a
repetitive wave-
it
can again be triggered and
If
the input sync volt-
(see para. 3-6 for an explana-
a
Schmitt Trigger).
is
a
5-volt negative spike
the sawtooth sweep.
(see para. 3-6). Triggering
e2
varies the grid
circuit in
a
positive or nega-
affects
the other;
all
it
would
differ-
area
of
its
3
-3
Loading...
+ 100 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.