Model 130B, Cabinet Mount: $650.00
Model 130BR, Rack Mount: $650.00
Opt ions
:
2. P-2 CRT (installed)
3. Internal graticule CRT (installed)
7. P
-
-
6
i
J
1-19-1
Ik
130
BR
11.
7 CRT (installed)
P-11 CRT (installed)
00013-2
V
Sect. I Page0
Model 130B
)
Figure
1-1.
Model 13OB/BR Oscilloscope
00013-2
,I
Model
130B
Sect.1
Page
1
i
1-1
GENERAL
The Hewlett-Packard Model 130B Oscilloscope
general purpose oscilloscope.
either internal
either internally
it
can
be
type mounting. Because
balanced input, the Model 130B may often
directly with transducers, enabling you
direct
having
Some of the
are
A.
tained from
insures
type of sweep generator,
independent
of sweep generator.
B.
sients by expanding
the
detail.
all
sweep time
C. CALIBRATED AMPLIFIERS
are
*5%.
in
ization
Phase shift measurements can
with this oscilloscope over
frequencies.
00013-2
presentation of phenomena
to
resort
as
follows:
LINEAR INTEGRATOR SWEEP GENERATOR
The accurate direct reading sweeps
a
X5 SWEEP EXPANSION
You speed observation and analysis
trace
This X5 sweep expander, may
sweep time settings and expands the fastest
Voltage measurements of various waveforms
quickly made with the 130B, accurate within
A built-in calibrator which
*2%
permits quick verification and standard-
of
the amplifier gain.
or
external sweeps which can
or
externally synchronized and
obtained in either the cabinet
of
to
preamplifiers.
special
a
Miller-integrator sweep circuit which
high
of
tube characteristics than other types
to
10
to
.2
features
order
linearity and stability.
a
two centimeter segment
centimeters
microsecond/cm.
it
can
be
used with
or
its
high sensitivity and
be
to
desired
of
this oscilloscope
are
is
more reliable and
of
for
easy viewing
be
used on
is
accurate with-
be
made accurately
a
wide range of input
without
is
be
rack
used
see
ob-
This
tran-
SECTION
GENERAL DESCRIPTION
1-2
a
This instrument should
when
refer
graph
1-3
a
The Oscilloscope
wired
otherwise specified. However, the instrument
may
source
power transformer. This conversion
in the Maintenance Section (Section
1-4
The three conductor power
instrument
prong male connector recommended by the National
Electrical
contact
two-blade
chassis when used with the appropriate receptacle.
of
An adapter should
of
plug
adapter
short lead from the adapter which should be connected
operating personnel.
1-5
The @ 130BR
ported in
usual manner;
mounted in the rack with brackets
Figure
supported by the dust cover and may
DAMAGE
it
is
to
the “Claim
on
POWER LINE VOLTAGES
for
also
be
if
the proper conversion
POWER CORD
Manufacturers’ Association. The third
is
an offset round pin, added to a standard
ac
to
a
standard two contact output.
is
used, the ground connection becomes
to
a
INSTALLATION
a
1-1.
IN
TRANSIT
be
thoroughly inspected
received. If any damage
for
Damage in Shipment”
the Warranty sheet in this manual.
is
shipped from the factory
115
volts
ac
line operation, unless
operated from a 230 voIts
is
IV).
cable
supplied with the
is
terminated in a polarized three
plug, which grounds the instrument
be
used
to
connect the NEMA
suitable ground
is
designed
19 inch rack by the front panel in the
or,
the dust cover may
In the
OF
latter
for
the protection
RACK MOUNT
so
that
it
case,
the chassis
is
evident,
made
is
described
When the
can
be
be
as
shown in
be
slipped in
para-
ac
line
to
the
a
of
sup-
rigidly
is
I
Sect. I Page
or
out easily; the screws through the front panel
merely holding the chassis in place.
mount the
1)
Mount the bracket
2
130BR
using the brackets:
as
shown in Figure
To
1-3
rack
with
screws through the outside holes of the brackets.
The length of these screws may
the front panel from the panel
The brackets
at
the
rear
are
installations but can be used
required. These
brackets
Hewlett-Packard Company
2)
Remove the dust cover from the
The
cathode
ny
date
not Included
You?
glad to process
Whenever
oui
rival,
ray
bbes
prchased
01
sale by
ln
local Hewletl-Packard
a
bbe is
in
full
and
since m eredlt
1)
Carefully wrnp
pddlng material.
2)
Wrap the
3)
Pack
the
4)
Prmund
almllir
tlght ail around
5)
mbes
be
E)
Shipprepaid preferably
We
CATHODf RAY
tube
supplied
from
theHewlett-PackrdCompany.
this
guarantee.
your
warnnty clalm
returned
returned with the tube. Follawshi&lng instructions carefuilyto
can
be
above
In
a
rlgid
tube In each dimension.
the
tube
shockabsorbing
returned
packed
I"
a
do
not
recommend
TUB1
In
your
Hnvlett-Packud
9
are
-ranteed
representative
for
allowed
SHIPPING
the
tube
in
heavy krdt
mntalner
withal leaat four lmhes afpekedexcelslor
the
from
wooden
maintalna
for
you. Please
L
warranty claim the
on
broken
INSTRUCTIONS
In
1/4"
thick
pper
which
Is
material.
bbe.
outside
the
eontlnental
box.
by
AIR
FRElGHTor
parcel
p~st
be
chosen
rails
to
as
desired.
notnecessaryin most
if
added support
are
available from the
as
an accessory item.
130BR
WARRANTY
Oselllodeope
agllmt
electrical
Broken
bbes
a
stock
conwl1
reverse
tubes.
cotton
batting
at
least
4
Be
certain
United
RAILWAY EXPRESS.
or
air
preel post shlpment.
nnd
replacement
failure
for
or
tubes withtmrned phosphor
of
him.
slde
Inches
that
thepcklng
Sate8
one
replacementtubes and
of
this sheet must
or
other
soft
larger
ULVl
or
Is
should
insure
space
and
csWa
year
from
be
de
is
wlll
filled
mount it in the brackets with the
10-32
screws provided.
3)
Slip the
130BR
into the dust cover and fasten in
place with screws through the front panel.
1-6
CATHODE RAY TUBE WARRANTY
The cathode ray tube (crt) supplied with the oscil-
loscope and replacement crt'
Hewlett-Packard company
s
purchased from
are
guaranteed against
electrical failure for one year from the date of sale
by Hewlett-Packard. Cathode Ray Tube Warranty
sheet is illustrated in figure
use
is included in the appendix of this manual.
FROM:
NAME:
COMPANY:
ADDRESS:
Perm"
to contact
for
the
are
NAHE:
TITLE:
kt
COMPANY:
ADDRESS:
ar-
rn
pceaa
1)
@MsTRuMENT
2)
TUBE TYPE SERIAL
3)
ORIGINAL TUBE REPLACEMENT TUBE
4)
YOUR
5)
DATE PURCHASED
8)
PURCHASED
7) COMPIAINT (Plawe describe mbre
8)
OPERATING CONDITIONS:
further
-
govr
clDlm
quickly please enter the
MODEL
PURCHASE ORDER
FROM
ZIT
Irdormatlon:
NO.
(Please
deacrlbe condltlona prlor
WARRANTY
of
Information
trouble)
1-2.
CLAIM
indicated
SERUL
A
sheet for your
below
to
nnd
at
Model
trusshead
DATE
___
Ume
of faflure
130B
-
HEWLETT-PACKARD
CO.
PAGE
MILL
ROAD,
PAL0
Figure
ALTO, CALIF. U.S.A.
1-2.
Cathode Ray Tube Warranty
i
00013-2
Model
130B
Sect.11 Page
1
\
1
OPERATING INSTRUCTIONS
2-1
Front panel operation controls
ure
enables you to operate
a
tailed operating procedures
ating plates.
INTERNAL SWEEP CONTROLS
SWEEP TIME/CM
This switch determines the speed at which the Horizontal or Sync INPUT
crt beam crosses the screen. HORIZ. SENSlTIVITY
tion or internal sweeps are not generated. Associ- sweeping voltages. On
ated with the SNEEP TIME/CM switch
centric VERNIER which provides continuous ad- rear of
justment of sweep speed
sweep magnifier operates
CONTROLS AND TERMINALS
2-1.
This description of the operating controls
the
instrument
basic
knowledge of oscilloscope technique.
are
-
switch must be in
an
INT. SWEEP posi- external
between
on
all ranges.
are
shown in Fig-
if
given
in
steps.
you have
De-
the
oper-
is
a con-
A
X5
TRIGGER LEVEL
This continuous control selects the level
sync waveform where triggering
the TRIGGER LEVEL control
trigger circuits
TRIGGER SLOPE
This two-position switch, concentric with TRIGGER
LEVEL, permits triggering to occur on either
positive or negative slope of internal, external or
line voltage sync signals.
A
set
of three binding posts used for receiving
sync
a
3-conductor receptacle
the
with the binding posts.
-
are
the most sensitive.
-
voltages and
instrument,
the
SECTION
on
is
to occur. When
is
set
to zero, the
-
external
rack mount model only,
5102,
is
connected
generated
mounted at the
in
parallel
II
the
the
SYNC
This three position switch lets
gered
triggering can be accomplished from a line
quency signal or from an applied
signal of sufficient amplitude to produce a one- Horizontal and vertical deflection plates, and a
half centimeter deflection. External triggering terminal for crt intensity (Z-axis) modulation.
can be produced by signals havingamplitude greater
than
SWEEPMODE
As
wise
an
dition through a condition
operation
in
counterclockwise position the control switches
into
optimum triggering bias for nearly all waveforms.
-
the
sweep be trig-
either
0.5
internally or externally. Internal
vertical
volt, peak-to-peak.
-
this control
position,
un-synchronized free-running (FREE
which sweeps
a
PRESET position.
is
rotated from the extreme clock-
the
sweep
is
possible (TRIGGERED) to a position
will
generator will pass from
where
not occur.
This
only triggered
At
the
position provides
input
RUN)
extreme
fre-
con-
2-2
The following terminals
the
See
2-3
When
the
at high sensitivities,
immediately following turn-on, becoming slower
as
drift,
not be attempted until the instrument
warm. For most purposes a
will be adequate.
REAR-ACCESS TERMINALS
------DANGER-
rear access plate of
Figures
trace
the
2-8
WARM-UP DRIFT
the
oscilloscope
will
instrument warms up, Because of this
fine
adjustment of amplifier balance should
HIGH
VOLTAGE-----
are
accessible through
the
instrument cabinet:
and
2-10.
is
first
turned
be
quite noticeable, particularly
the
trace drift
5
on,
is
thoroughly
minute warm-up
is
drift
fastest
in
Sect.11 Page
2-4
AC OR DC COUPLING
2
AC coupling permits high gain to be employed
without regard for
the
dc level involved.
In
the
AC position the input signal (vertical or horizontal)
is
coupled to
which removes the dc component from
the
amplifier through a capacitor
the
input.
This coupling circuit has a low frequency cut-off
at
2
cps. To avoid degrading input pulses or square
waves below
ling. WHEN USING
AGE VALUE
POSITION
200
cps it
is
advisable to
DC
COUPLING THE AVER-
OF
THE DC DETERMINES THE
OF
THE SWEEP ON THE OSCILLO-
use
dc coup-
SCOPE. IF YOU ARE UNABLE TO FIND THE
TRACE WITH THE VERTICAL POSITION CONTROL WHEN USING
DC
COUPLING, TRY AC
COUPLING. When AC coupled the maximum dc
that may be applied
2-5
BALANCED INPUTS
is
600
volts.
The instrument will accept balanced input signals
on
the
six
most sensitive ranges. This arrange-
ment
is
shown
in
Figure
2-7.
Driving the instrument from a balanced source
can be
stray pickup that would otherwise obscure
desired information. To take advantage of
noise reduction that
input, you must be
the source
conductor shielded cable
oscilloscope. The input cable shield must
connected to a suitable ground,
loscope or some other
cautions
very
effective in removing
is
possible
sure
that neither terminal of
is
connected to ground, and
between
point.
in
the external input circuit, any stray
the
with
a balanced
use
the source and
either
at the oscfl-
With these pre-
unwanted
the
the
double
be
signals (noise, hum, etc.) will be coupled equally
to
the
two input terminals, and be cancelled by
the
differential amplifiers. Since
formation
is
applied between
nals, it will be amplified and displayed
normal manner. Since
the
the
noise
the
two
is
desired
input
termi-
in
a problem
in-
the
Model
at
mainly
is
available only on the most sensitive ranges
low level, the fact that balanced input
generally not a serious limitation.
The common-mode signal rejection will be at
least
40
db
(1/100
of
the
input signal).
using a balanced input certain limitations must
be considered. The proper operating levels must
be maintained
on
the
input
amplifier: The COMMON-MODE SIGNAL VOLTAGE MUST NOT EXCEED
TIVE,
this
1.5
VOLTS EITHER POSITIVE ORNEGA-
ON
EITHER INPUT TERMINAL. Note that
is the sum of all voltages (dc plus peak ac).
NOTE
If
balanced ac coupling
to connect a capacitor
is
desired, it is necessary
in
the external signal path
to the middle terminal, since a dc voltage on this
terminal only unbalances
arrangement
is
shown
in
Figure
the
amplifier. This
2-7.
-----------
2-6
Basic operating procedures are described
following illustrations. Positions of controls are
different on
are identical to those of the rack model.
Short vertical input terminals together.
Set AC-DC switch to
Set VERT. SENSITIVITY to CAL.
VERNIER to CAL.
base
line
is
formed.
50
MILLISECONDS/CM
DC.
so
(Any
RUN.
INT.
that a consweep
time
is
satis-
Turn
RO
b
7.
Center bottom portion of calibration signal trace using VERT. POS. control.
8.
Set VERT. SENSITIVITY to 1 MILLIVOLTS/
CM.
9.
Center trace with coarse (screwdriver)
VERT.
control if unbalance
10.Repeat steps
NOTE:
etched board)
VERNIER.
DC
BAL, control or with
is
slight.
6,
7,8
and
A
separate adjustment (Bal. Adj. on
is
provided to balance
9
if
fine
necessary.
/
(knob)
the
the
Figure
2-2
Sect.
I1
Page
4
HORIZONTAL BALANCE ADJUSTMENT
Model
130B
After warm-up:
1.
Short together the horizontal INPUT terminals.
2.
Set AC-DC switch
3.
Set HORIZ. SENSITIVITY to CAL. Turn
VERNIER
4.
Adjust the HORIZ. POS. control
to
left edge of the calibrating signal
CAL.
to
DC.
to
place the
trace
the major vertical axis.
JUMPER
5.
6.
7.
NOTE: A separate adjustment (Bal. Adj. on the
etched
on
VERNIER.
Set HORIZ. SENSITIVITY
VOLT/CM.
Return the spot
with the
coarse
BAL. control
if the unbalance
Repeat steps
board)
to
the major vertical axis
(screwdriver) HORIZ.
or
with the fine (knob) control
is
slight.
2,
3
4,
and 5 if necessary.
is
provided
RO
1
to
to
balance the
MILLI-
DC
Figure
2-3
Model 130B
@-
0-
Sect.11 Page 5
INTERNAL SWEEP - INTERNAL SYNCHRONIZATION
1.
Set HORIZ. SENSITIVITY switch
SWEEP
2.
Set SYNC switch to INT.
3.
Set SWEEP MODE to PRESET.
4.
Connect vertical input signal into vertical
input terminals.
5.
Set AC-DC switch for
6.
Adjust VERT. SENSITIVITY
sensitivity.
X1
(or
to
X5
for
magnified sweeps)
type
coupling desired.
to
for
INT.
desired
Figure
7.
Set TRIGGER SLOPE switch for triggering
on
positive
as desired.
8.
Set TRIGGER LEVEL control
9.
Select desired sweep speed with SWEEP
TIME/CM switch.
10.Adjust TRIGGER LEVEL
desired level. In some cases, it may
necessary to switch SWEEP MODE from
PRESET
particular trace being viewed.
2-4
or
negative
to
an individual adjustment
slope
of
to
input signal,
to
start trace at
0.
for
be
the
Sect.11 Page
6
Model 130B
INTERNAL SWEEP - EXTERNAL SYNCHRONIZATION
1.
Set HORIZ. SENSITIVITY switch
SWEEP
2.
Set SYNC switch to EXT.
3.
Set SWEEP MODE
4.
Feed synchronizing signal
more) to the horizontal input terminals.
5.
Set AC-DC switch
6.
Set TRIGGER LEVEL to
7.
Feed vertical input signal into vertical input
terminals.
X1
(or
to
X5
for magnified sweeps).
.
to
PRESET.
(0.5
for
type coupling desired.
0.
volts p-p or
to
IfiT.
Figure
8.
Adjust VERT. SENSITIVITY for desired
sensitivity.
9.
Select desired sweep speed with SWEEP
TIME/CM switch.
10. Set TRIGGER SLOPE for triggering on posi-
or
tive
11.Adiust TRIGGER LEVEL
desired level. In some cases, it may
found necessary to switch SWEEP MODE
from PRESET to an individual adjustment for
the particular trace being viewed.
negative slope, as desired.
to
start trace at
2-5
be
Model
130B
EXTERNAL HORIZONTAL INPUT
Sect.
I1
Page
7
?
1.
Feed horizontal signal to horizontal input
terminals.
2.
Set AC-DC switch
des ired.
3.
Set
HORIZ.
sensitivity. ing Lissajous patterns,
SENSITIVITY switch
for
type of input coupling
for
desired
Figure
4.
Adjust horizontal position
HORIZ.
This type
POS. control.
of
input
2-6
will
be
etc.
found
of
pattern with
useful
for
view-
Sect.11 Page
8
Model
130B
AC COUPLING BALANCED INPUT
VERT
POS
The following procedure
input, but
1.
Set VERT. SENSITIVITY to
is
the same for the horizontal input.
is
for the vertical
50
MILLIVOLTS/
CM, input not balanced on higher ranges.
2.
Set AC-DC switch to AC.
3.
Disconnect shorting strap.
4.
Connect
0.1
microfarad capacitor to mid-
terminal.
5.
Connect input signal
6.
Ground input at the black terminal.
The capacitor must
to
A
and
be
used to block any dc.
RO
B.
Figure
2-7
Model
130B
CONNECTION TO CRT DEFLECTION PLATES
Sect.
I1
Page 9
The following procedure
ternal signals to
but
is
the
1.
Remove rear access plate fastened by four
the
same for the horizontal plates.
is
for connecting ex-
vertical deflection plates,
screws.
2.
Remove
cal
replace them with
the
shorting bars between
Amplifier and terminals
1
megohm,
D3
the
and
1/2
resistor.
Verti-
D4
and
watt
Figure
For balanced AC coudinp:
3.
Connect balanced signal through appropriate
capacitor to
D3
and
D4.
For single-ended AC coupling:
4.
Bypass
D4
to chassis
with
capacity.
5.
Connect
the
signal to
D3
through an appro-
priate capacitor.
NOTE:
deflect
and connect
If
the
it
is
desired to have positive voltage
beam
downward, bypass
the
signal to
D4.
2-8
an
adequate
D3
tochassis
Sect.11
Page
10
EXTERNAL INTENSITY MODULATION
Z
AXIS
02
POSlTNE
DEFLECTS
RT
d
D3
POSITIVE DEFLECTS UP
VERT
I
RO
CA
Model
L
130B
d
CAUTION: Dangerous Voltages
this terminal board. Be sure the instrumekt
turned
To
signals:
1.
off
when making this connection.
intensity modulate the CRT with external
Remove
small screws
rear
access
at
rear
plate fastened
of
are
present-on
dust cover.
by
four
is
2.
Remove shorting
3.
Connect modulating signal
A
positive voltage
the CRT trace from normal intensity.
bar.
of
to
20
volts peak will blank
these terminals
Figure
2-
9
Model
130B
Sect.11 Page
11
ALIGNING SCOPE TRACE
WITH
GRATICULE
RO
CAUTION: DANGEROUS VOLTAGES ARE
PRESENT INSIDE THE INSTRUMENT
Remove two screws at rear of dust cover and
slide cover off to rear. Fiber lever
both
radial and longitudinal positioning of CRT
and
is
locked by clamp
(1).
(2)
controls
Figure
To align sweep trace with graticule loosen clamF
(1)
with a screwdriver. Rotate fiber arm
until
the
trace
is
parallel to horizontal lines
graticule. Tighten clamp
(1)
after adjustment
has been made.
2-10
(2)
on
Sect.
111
Page
0
Model
130B
r--
I
I
J
I
c
0
L
a
FW
3-
a5
+-I
3n
or
a
E
(d
k
bn
(d
8
c
0.
-I
a
4
al
k
3
bn
iz
OU
I
I
Model 130B
J
Sect.111 Page
1
3-1
GENERAL CONTENT
This section contains a brief description of the
over-all operation of the Model 130B Oscilloscope,
description of each major section and detailed
description
3-2
The block diagram in Figure
circuits of the Model 130B Oscilloscope.
A.
VERTICAL AMPLIFIER
The Vertical Amplifier receives the input
signal, amplifies it, and drives the vertical deflection plates of the cathode ray tube. In addition,
this
amplifier determines the vertical position of
the spot on
chronizing
B. HORIZONTAL AMPLIFIER
The Horizontal Amplifier receives
either from
Sweep Generator, amplifies
zontal deflection plates of
Except for
fier
for amplifying the internally- generated sawtooth voltage,
Vertical Amplifier.
SflEEPGENERATOR
The Sweep Generator forms a sawtooth volt-
to control
across
Generator
generator,
the
sawtooth. The trigger generator controls
allow
sawtooth sweep begins.
of
a Schmitt trigger.
OVER-ALL OPERATION
3-1
the
screen
the
sweep
the
the
provisions in
the
the
face of
is
divided into two parts:
2)
the
operator to choose
and supplies a signal for syn-
with
the vertical input signal.
horizontal INPUT jack or from the
it
and drives
the
cathode ray
the
Horizontal Ampli-
it
is
essentially
horizontal movement of the spot
the
cathode ray tube. The Sweep
a
trigger generator, which starts
the
point at which the
shows the basic
its
signal
the
hori-
tube.
the
same as
1)
a
sawtooth
the
SECTION
THEORY
In addition to forming the 'internal sweep of the
oscilloscope,
the required unblanking pulse
trace during
D.
CALIBRATOR
An
nominal frequency of 300 cps,
ting
the
either
to CAL., turns
and connects
E. CATHODE
accelerator type. It
P1 phosphor screen but
P11 phosphors also and P2 upon special order.
All are electrically interchangeable and the tube
easily changed.
makes possible
which requires no resetting when adjusting the
FOCUS or INTENSITY controls.
plate terminals are connected through removable
jumpers at
rect connections to the plates
3-3
The vertical amplification channel consists of
three parts:
ator,and the amplifier section proper.
A.
is
capacitor
DC
attenuator.
the VERT. or HORIZ. SENSITIVITY switches
The cathode ray tube
VERTICAL AMPLIFICATION CHANNEL
AC - DC
The signal comes into the input terminals and
fed to
position, the signal goes directly to
the
each
sweep.
internal square-wave calibrator, with
basic gain of the amplifiers. Turning
on
its
output to the appropriate amplifier.
RAY
a
the
rear of the instrument
the
SWITCH
the
AC-DC switch. For ac coupling,
is
switched into
OF
Sweep Generator also supplies
the
calibrator supply voltage
TUBE
is
normally supplied
is
The
simple astigmatism adjustment
OPERATION
which
brightens
is
provided for
is
a SAQP - mono-
available
mono-accelerator anode
in
The
the
deflection
so
can
be made easily.
AC-DC
switch,
the
signal path.
the
input attenu-
set-
with
P7 and
that di-
In
the
input
111
the
a
a
is
a
the
Loading...
+ 46 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.